Introduction

Brain injury is common in pediatric critical care as a result of either a primary neurological or neurosurgical process, secondary brain injury or as a consequence of systemic disease. The ability to identify children at risk of neurologic injury and detect early signs of neurologic dysfunction is key to providing good critical care. Pediatric neuromonitoring has traditionally focused on serial neurologic examinations, electroencephalography (EEG), and neuroimaging. Similar to adult neurocritical care, there is an emerging interest in other noninvasive neuromonitoring modalities, particularly automated pupillometry, near-infrared spectroscopy (NIRS), and transcranial Doppler (TCD) ultrasonography. These technologies may provide additional insights into important physiological changes, improve early recognition of acute neurologic deterioration, and help guide management in a meaningful way that improves patient outcomes [1, 2]. In this article, we aim to review pupillometry, NIRS, and TCD ultrasonography in the pediatric intensive care unit (PICU) setting. For each technology, we will discuss the basic methods and available devices, use in clinical care and research, and limitations and future directions, realizing that normative data for the PICU population and standardized protocols, data reporting, and interpretation are needed for their safe and effective use in critically ill children.

Pupillometry

The pupillary light reflex (PLR) is a complex physiologic process involving the sympathetic and parasympathetic nervous system. When light is shined into the eye, retinal photoreceptors are activated via the optic nerve. The optic nerve traverses the optic chiasm, where nerves from the nasal retina cross to the contralateral side and nerves from the temporal retina continue ipsilaterally [3]. After exiting the optic tract, they synapse within the pretectal area of the midbrain. Axons then travel from the pretectal area bilaterally and synapse at both Edinger–Westphal nuclei. The parasympathetic fibers travel via the oculomotor nerves to synapse at the ciliary ganglion and ultimately innervate the iris sphincter, resulting in pupillary constriction [3,4,5]. In contrast to the parasympathetic nervous system, the sympathetic nervous system is responsible for pupillary dilation. Lesions at any point along these pathways, therefore, may impact the pupillary examination. Pupil size or PLR may be confounded by a variety of medications, alcohol, delirium, and acute brain injury itself [6, 7]. Medications that may impact the static or dynamic evaluation of the pupil include but are not limited to atropine, barbiturates, diphenhydramine, naloxone, opioids, propofol, and selective serotonin reuptake inhibitors [4].

Pupillary assessment is a key component of the neurologic examination of any critically ill child that may be performed qualitatively or quantitatively. Traditionally, this has been examiner dependent and subjective and would include an estimation of pupillary size and reactivity assessed as “brisk,” “sluggish,” or “nonreactive.” Prior research has shown that qualitative assessments of pupils using a penlight have the potential to be inaccurate, with limited interrater reliability, and may not identify changes in pupillary size or reactivity [8,9,10]. Automated quantitative pupillometry, on the other hand, has allowed for objective data reporting, evaluation of multiple dynamic components of the PLR, and digital visualization of trends that may signify acute neurologic dysfunction.

Pupillometer Methods and Devices

Although there are a variety of pupillometers reported in the literature, there are two main companies that manufacture devices applicable to the intensivist. Until recently, NeurOptics was the only manufacturer available in the United States. However, as of March 2022, the pupillometer made by ID Med is now available (Table 1). Handheld pupillometers have their own light source, a self-contained digital camera, and a liquid crystal display screen. The pupillometers made by these two manufacturers have a 510 K US Food and Drug Administration (FDA) exemption and are available in both the United States and Europe. As with any commercial device, individual pupillometer parameters and results may vary across manufacturers. An adult study comparing two commercially available pupillometers (NeurOptics NPi-200 and the NeuroLight Aligscan) showed significant correlation for pupil size and constriction velocity but no correlation for pupillary latency [11].

Table 1 Commercially available handheld pupillometers reported in the ICU literature

When using the pupillometer, the examination begins with a flash of light of fixed intensity and duration that stimulates the PLR [14]. Infrared light rays are used to illuminate the eye, and images are acquired by the digital camera [15]. As an example, the NeurOptics NPi-200 captures 90 images in a 3-s period (NeurOptics.com). The results are immediately displayed on the screen and can be trended over time. As such, its use has been associated with reduced nursing assessment time when the pupillometer synchronizes with the electronic medical record (penlight examination: 96 s versus 37 s with the pupillometer) [16].

Automated pupillometry allows for quantitative measurement of multiple variables that cannot be quantified with the human eye. Measured variables vary based on the manufacturer but often include static and dynamic parameters that may be age related and impacted by environmental light conditions (Table 2) [4, 14]. For example, in a study of 90 children aged 1 to 18 years old in ambient light, the mean maximum pupil size was 4.11 mm, the mean minimum size was 2.65 mm, the mean constriction velocity was 2.34 mm/s, the mean dilation velocity was 2.2 mm/s, and the mean pupillary size reduction was 36% following light stimulus [12]. The authors noted larger pupils in children aged 2 to 6 years old, which may be related to myelination [12]. In contrast, under scotopic light conditions, 101 children had a mean resting pupil diameter of 6.6 ± 0.74 mm, a mean minimum size of 4.7 ± 0.77 mm, a mean constriction velocity of 3.7 ± 0.744 mm/s, a mean dilation velocity of 0.88 ± 0.25 mm/s, and a mean pupillary size reduction of 30% ± 6.2% [13].

Table 2 Pupillometer variables

NeurOptics pupillometers employ a proprietary algorithm to quantify pupillary reactivity called the Neurological Pupil Index (NPi), first introduced in 2011 [14, 17]. The NPi incorporates multiple PLR variables, including size, latency, constriction velocity, and dilation velocity [14]. Each individual variable is compared against the mean reference value of healthy individuals and the difference is standardized to the standard deviation, resulting in a z-score that becomes the NPi [14]. The NPi is scored from 0 to 5, in which 0 indicates an absent PLR and 5 indicates a “brisk” PLR. Within the adult literature, an NPi ≥ 3 suggests that pupil reactivity falls “within the boundaries of the normative pupil behavior distribution” [14]. An NPi < 3 has been termed “abnormal.” There are no data in children to indicate which NPi values are cause for concern.

Use of Pupillometry in Clinical Care and Research

In adults, pupillometry-derived PLR has been studied in a variety of applications (e.g., screening for neurologic complications, neuroprognostication) and conditions (e.g., traumatic brain injury [TBI], aneurysmal subarachnoid hemorrhage/intracerebral hemorrhage). In adults with traumatic and nontraumatic intracerebral hemorrhage, an NPi < 3 was associated with increased intracranial pressure (ICP) (average 30.5 mmHg), whereas an NPi ≥ 3 was associated with normal ICP (average 19.6 mmHg). In a subset of these patients, pupillometry was able to identify changes in pupillary reactivity preceding intracranial hypertension [14]. Findings in adults have also suggested that pupillary constriction velocity < 0.6 mm/second, an NPi < 3, and pupil size reduction < 10% have been associated with intracranial hypertension [4]. In a small cohort of adults with aneurysmal subarachnoid hemorrhage, an NPi < 3 was associated with the development of delayed cerebral ischemia [18]. The pupillometer has been reported to aid in neuroprognostication following a variety of acute neurologic conditions [4]. In a study of 103 adults following cardiac arrest, 48-h quantitative PLR percentage reactivity ≥ 13% distinguished survivors from nonsurvivors (PLR < 13% had a 100% positive predictive value for mortality with 100% specificity) [19]. In a multicenter trial of 456 adults following cardiac arrest, an NPi ≤ 2 at any time between days 1 and 3 was associated with unfavorable outcome with 32% sensitivity, 100% specificity, and a 100% positive predictive value, exceeding performance metrics of the manual pupillary examination [20].

In children, the pupillometer has been used in TBI, stroke, central nervous system infections or malignancies, seizures, encephalopathy, septic shock, cerebral edema, and extracorporeal membrane oxygenation [1]. In a recent survey of neuromonitoring practices in 52 PICUs, 17 (33%) intensive care units reported using pupillometry [1]. Pupillometry use was guided by institutional protocols (ten hospitals, 59%), a pediatric intensivist (ten hospitals, 59%), the neurosurgical team (four hospitals, 24%), and/or the neurology team (six hospitals, 35%). In this survey, the pupillometer was most frequently used in children with severe hypoxic-ischemic brain injury following cardiac arrest (29%), severe TBI (27%), and cerebral edema with no clear underlying etiology (25%) [1]. There are no known specific thresholds or values derived from pupillometry in children that have been clearly associated with unfavorable patient outcomes or acute change in management, and trends in data may be helpful. Freeman and colleagues [21] attempted to evaluate whether abnormal pupillary measurements would precede increases in ICP in a cohort of 28 children with acute brain injury and ICP monitors. When ICP was < 20 mmHg (n = 856 measurements), right and left NPis were 4.4 and 4.4, whereas when ICP was ≥ 20 mmHg (n = 315 measurements), the right NPi was 3.4 and the left NPi was 3.3. There was an inverse correlation with NPi, pupillary size, constriction velocity, maximum constriction velocity, and dilation velocity with ICP but no significant correlation with latency [21]. These authors suggested that individual measured parameters that contribute to the NPi may better correlate with ICP than the absolute NPi score itself [21]. Although normative data exist for a number of components of the PLR in healthy children using the pupillometer [12, 13], it is unclear whether there is a specific NPi cutoff in children that indicates “normal” or “abnormal” PLR and what defines a clinically meaningful change in the NPi for each manufacturer.

The current evidence base regarding the use of automated pupillometry in critically ill children is limited. The World Brain Death Project suggests using a pupillometer during the clinical examination for determination of brain death [22]. Outside of this recommendation, there are no other societal guidelines or consensus statements with clear recommendations as to which critically ill children would benefit from the use of automated pupillometry as a routine portion of their care. Therefore, the use of automated pupillometry is likely best integrated based on individual institutional standards for neurologic examinations of the critically ill child as recommended by a pediatric intensivist, neurologist, or neurosurgeon.

Pupillometry Limitations and Future Directions

Automated pupillometry has the potential to impact clinical care for the critically ill child. The objective data obtained from the pupillometer may allow for early detection of impending neurologic deterioration, may serve as a marker that one could trend pre-intervention and post-intervention, may assist in the noninvasive detection of intracranial hypertension, and may have utility in aiding in neuroprognostication. Prospective studies are needed to better understand these and other clinical applications for pupillometry in the PICU and how pupillometry may impact clinical decision-making at the bedside. Standardized data reporting and interpretation that does not focus solely on analyzing proprietary algorithms is needed to be most generalizable to the global field. Performance metrics of pupillometry for outcomes of interest also need to be better understood.

NIRS

Optimizing oxygen delivery to various tissues is a key goal in critical care. The noninvasive measurement of blood oxygenation via pulse oximetry revolutionized patient management in the 1970s because it gave clinicians a tool to measure systemic oxygen levels, to recognize hypoxemia before it was otherwise obvious, and to respond with changes in treatment [23]. Around the same time in 1977, NIRS was described as a noninvasive tool to detect the oxygenation state of living tissue by Jobsis [24]. Since then, the role for NIRS has been a focus of research, resulting in thousands of publications and review articles, and its use has been integrated into clinical practice in various ways, importantly to assess cerebral regional oxygenation (rSO2).

NIRS Methods and Devices

NIRS is a reasonably priced, portable, relatively small, noninvasive monitor that does not require ionizing radiation and can be quickly applied at the bedside without the need for ongoing hands-on operation. A list of FDA-approved NIRS devices that are commercially available in the United States is shown in Table 3. A probe, which contains both the light emission source and the detection sensor, is applied to the area of interest (e.g., forehead for cerebral monitoring), and data are transmitted to a processor with a display (Fig. 1a). Most machines have neonatal or pediatric probes, which differ from adult probes by the size of the footprint and different depth of light penetration to compensate for the thinner skull of infants and children. Near-infrared light of various wavelengths is emitted from the probe, passes through skin and bone, and diffuses through the tissue, where it encounters hemoglobin in its different states of oxygenation in the arteries, veins, and capillaries in a manner that is not flow dependent. Importantly, deoxyhemoglobin and oxyhemoglobin absorb light at different wavelengths (Fig. 1b). Reflected light is measured by the detection sensor, and these data are processed through a series of complex mathematical transformations with a basis in the modified Beer–Lambert law to provide information about tissue oxygenation. A typically proprietary algorithm that includes compensation for the light penetration depth and other factors is used by each device to provide a regional cerebral oxygen saturation (rSO2), which is expressed as a percentage (oxyhemoglobin/total hemoglobin × 100) displayed on the monitor, along with the trend over time.

Table 3 FDA-approved NIRS devices commercially available in the United States
Fig. 1
figure 1

a Basic model of near infrared spectroscopy showing the light emission source and detector on a single probe. From Denault et al. [25]. b Absorption spectra for deoxyhemoglobin (HHb), oxyhemoglobin (HbO2), cytochrome c oxidase (Cyt), and water (H2O) in the near-infrared range. Most oximeters use wavelengths in the 730–900-nm range. From Shaaban-Ali et al. [26]

There are no well-validated normative thresholds, and there is no true gold standard for rSO2, making the trend in data likely more valuable than any individual point in time. However, rSO2 obtained using NIRS has been shown to correlate with jugular bulb saturation (the closest to a gold standard for the measurement of the balance between cerebral oxygen delivery and use that exists) [28,29,30,31,32,33,34]. There is also variability between each NIRS machine because of variations in probe size, depth of beam penetration, signal processing, and data analysis. Although there is no FDA standard for accuracy as there is with pulse oximetry, for example, which must have an accuracy root mean square within 2–3% of arterial blood gas values, comparative studies of several commercially available systems have shown that they are generally equivalent within ± 10% in both real-world and laboratory settings [35,36,37,38,39]. Furthermore, the calculation of rSO2 requires assumptions that may not be true in all clinical scenarios. For example, most systems assume around 10% intracapillary/20% intraarterial/70% intravenous blood volume distribution, and calculations are weighted appropriately. However, there may be individual variability leading to inaccurate results for that specific patient [40]. There are also physiologic or pathophysiologic changes in the arterial/venous ratio that occur, which may lead to inaccuracy of cerebral oximetry [41]. For example, Schober et al. [42] compared the difference between measured rSO2 and that calculated from jugular venous and radial arterial hemoglobin saturations using the manufacturer-provided ratios in healthy volunteers over a range of fraction of inspired oxygen (FiO2) and two partial pressure of carbon dioxide in arterial blood (PaCO2) levels. They showed that acute hypocapnia, which presumably changes the arterial/venous ratio, led to an overestimate of rSO2, and the difference was exacerbated with lower FiO2 [42]. Additionally, it is assumed that most of the data collected by the probe come from the brain, with minimal contribution from the skin and tissues. However, this may not always be the case; for example, different forehead shapes, larger sinuses, tissue edema, or scarring may create artifact or interference with the probe or add a more than minimal contribution to the rSO2 calculation [43]. Because of these limitations, NIRS-derived rSO2 has variability both between patients and within the same patient, making quantitative analysis less valuable than trend monitoring [36, 44]. Although studies using different machines in different populations have generated different normative data, a baseline rSO2 of around 55–75% in adults and children without cyanotic heart disease is commonly accepted [45,46,47,48]. Kurth et al. [49] described the effect of hypoxia on the brain in a piglet model and showed that impairment occurred in a graded fashion and established general thresholds for concern (Fig. 2). Slowing in the EEG and increased brain tissue lactate levels were seen at a NIRS rSO2 of 40–45%, and this progressed to flat EEG results at a NIRS value around 30% [49].

Fig. 2
figure 2

In a piglet model of hypoxia, signs of neurophysiologic impairment were identified in a graded fashion. Electroencephalography (EEG) slowing occurred first, followed by an increase in brain lactate levels around a cerebral regional oxygenation (rSO2) of 40–45%. As hypoxia progressed, brain failure was evident with a flat EEG and decreased ATP. A “buffer” was identified between normal regional cerebral oxygen saturation and 45%, at which physiologic changes became evident. The figure was made from data shown in Kurth et al. [49]

Use of NIRS in Clinical Care and Research

The body of literature evaluating the use of cerebral NIRS in a variety of neonatal and pediatric conditions in the intensive care unit suggests that it may be a useful adjunct with multimodal monitoring, but the impact on patient management and outcomes from NIRS alone remains unclear. In a systematic review from the Cochrane library that examined the use of NIRS in premature infants born 12 weeks early and outcomes, including mortality and later neurologic development, one randomized clinical trial was included involving 166 premature infants randomized to continuous cerebral NIRS monitoring until 72 h after birth and appropriate interventions if NIRS was out of normal range according to a guideline versus conventional monitoring with blinded NIRS. The trial was powered, however, to detect a difference in cerebral oxygenation and not morbidities or mortality. No consistent effects of NIRS along with a guideline on the assessed clinical outcomes were seen. Validity was hampered by a small sample, lack of blinding, lack of power, and indirectness of outcomes [50]. Assessment of cerebral autoregulation using NIRS in the neonatal population with hypoxic-ischemic encephalopathy is beyond the scope of this review. Another systematic review on the use of cerebral NIRS in congenital heart disease reviewed available data from 47 case series, four randomized trials, and three retrospective studies. The data suggested that this literature failed to find an association between the use of NIRS and neurologic outcomes, and the authors concluded that this limits the support for widespread implementation [51]. A recent survey, however, of 52 North American PICUs showed that 87% of respondents reported using NIRS in clinical care at their institution. Patients were chosen for monitoring primarily at the choice of the PICU attending (89%) or based on institutional protocol (56%), and primary reasons for monitoring included ECMO (69%), severe hypoxic-ischemic brain injury after cardiac arrest (56%), and severe TBI (48%) [1]. This study does not provide information on frequency of usage or how NIRS was used, and it does not represent usage outside of North America.

In pediatric TBI, small studies support the use of NIRS [52,53,54,55]. In a small pilot study by Kampfl et al. [52] involving eight patients with ICP monitors post head injury, rSO2 was lower in the high ICP group (> 25 mmHg) than in the low ICP group (< 25 mmHg). Another study by Kirkpatrick et al. [53] of 14 ventilated patients with closed head injury also showed that NIRS changes correlated with changes in cerebral perfusion pressure identified by multimodal monitoring 97% of the time. Studies by Zuluaga et al. and Lang et al. showed that rSO2 trended lower with increased ICP in patients with brain tumors, trauma, and hydrocephalus pre- and post cerebrospinal fluid (CSF) diversion [54, 55].

The small size and portability of NIRS makes it feasible to use in a prehospital setting, and it has been successfully studied by noncivilian groups as a tool to identify traumatic intracranial hemorrhage in the field [56,57,58]. Pediatric research teams have looked at the use of NIRS for detection of intracranial hemorrhage in children because of these promising results. Studies by Salonia et al., Bressan et al., Lewartowska-Nyga et al., and Kirschen et al. each found that the Infrascan, a handheld NIRS device, identified intracranial hemorrhage that was confirmed by head computed tomography with reasonable sensitivity and specificity [59,60,61,62]. This device does not use rSO2 but rather determines differences in the optical density, which is higher in areas of hemorrhage because of increased hemoglobin levels and therefore more absorbent of NIRS light. These studies each show that the handheld NIRS device has a high negative predictive value in their cohorts, suggesting it could act in an adjunctive role with current clinical decision rules to predict risk of intracranial hematoma in children with head trauma. Further research in large cohorts with a focus on effectiveness will be valuable in clarifying the role for this technology, which has the potential to decrease radiation exposure via computed tomography scans and shorten length of observation for patients who may have previously required these interventions.

In post-cardiac-arrest monitoring, studies in adults have suggested a role for cerebral NIRS during cardiac arrest, with higher rSO2 values correlating with return of spontaneous circulation (ROSC) and neurologically favorable survival to hospital discharge [63,64,65]. Several small case series have described use of NIRS in pediatric cardiac arrest [66,67,68]. A small prospective study by Francoeur et al. [69] in 2022 showed that NIRS placement during pediatric in-hospital cardiac arrest was feasible and that the median rSO2 was higher for events with ROSC than those with no ROSC, but there was no association between rSO2 and survival to discharge. As the field of cardiac arrest research focuses on goal-directed cardiopulmonary resuscitation, cerebral NIRS can noninvasively assess rapid changes in cerebral oxygenation, which may correlate with more effective cardiopulmonary resuscitation and improved neurologic outcomes, but larger studies are required to determine its role in this clinical scenario.

Lastly, small pediatric studies suggest that use of NIRS-derived optimal blood pressure determination is feasible post cardiac arrest [70, 71]. Kirschen et al. [72] showed that a larger amount of time with the mean arterial pressure (MAP) below the NIRS-derived optimal MAP (MAPopt) during the first 24 h after cardiac arrest correlated with unfavorable outcomes (Pediatric Cerebral Performance Categorization ≥ 3 with ≥ 1 change from baseline at the time of hospital discharge). Similarly, Lee et al. [73] showed that a larger time with MAP below MAPopt or a larger deviation below MAPopt was associated with new tracheostomy or gastrostomy, brain death, or withdrawal of technological support for neurologic futility. This is an area that requires further research to clarify the role for NIRS in the care of children post arrest.

As highlighted previously, there is a large breadth of studies using NIRS in critically ill pediatric patients. Although there is insufficient evidence to support formal guidelines surrounding its use, there is also no evidence showing harm through the use of this noninvasive device. When the device can be placed around other clinically indicated equipment (e.g., EEG leads), NIRS could be used as a component of multimodality monitoring of patients with neurologic injury [51]. As with other noninvasive monitoring devices, trends in NIRS data may be the most clinically useful. To use cerebral NIRS clinically in any scenario, it is important to remember that cerebral rSO2 is a result of cerebral tissue oxygen supply (a factor of cerebral blood flow and arterial O2 content) minus the demand (Fig. 3a). There are many physiologic alterations that can affect this number, and different interventions may be required. For example, low rSO2 may be due to low oxygen delivery from decreased cardiac output and hypotension. In this case, recognizing and treating the hypotension should improve the rSO2. Alternatively, if cerebral oxygen demand is high because of seizures, antiseizure medication is required to normalize the rSO2. Although there are no accepted protocols or guidelines for when to use NIRS and how to interpret the data in the PICU population that have been shown to result in improved patient-specific outcomes, algorithms have been created by various institutions to help clinicians use the information obtained from NIRS in addition to other physiologic parameters to modify clinical care (Fig. 3b) [74, 75]. Further research is needed to validate these algorithms in the PICU for NIRS to be safely and effectively incorporated into clinical care.

Fig. 3
figure 3

a Examples of physiologic factors that impact cerebral oxygenation. b An example of a management algorithm using cerebral near-infrared spectroscopy (NIRS) data during cardiac surgery as shown by Trafidlo et al. [75]. rSO2, regional cerebral oxygen saturation, HgB, hemoglobin, CBF, cerebral blood flow, SpO2, oxygen saturation, CPP, cerebral perfusion pressure, MAP, mean arterial blood pressure, ICP, intracranial pressure, PaO2, partial pressure of oxygen in arterial blood, PaCO2, partial pressure of carbon dioxide in arterial blood, SvO2, venous oxygen saturation, CMRO2, cerebral metabolic rate of oxygenation consumption

NIRS Limitations and future directions

Cerebral NIRS monitoring has several limitations, some of which are related to the physics behind the machine’s function and the physiology of the brain in normal and pathologic states. As discussed previously, the lack of validated normative thresholds may make a trend more meaningful than individual data points. Additionally, the data obtained are subject to physiologic or pathophysiologic changes of noncerebral origin. A small laboratory-based study evaluated the impact of varying levels of skin pigmentation and melanin on NIRS oxygenation readings and found that as the level of pigmentation increased, the reported saturation decreased [76]. Clinically, this was assessed by Sun et al. [77] in a retrospective registry study in which they found that self-identified African American patients had a lower rSO2 when compared with White patients. This is a potential source of bias that must be addressed in research going forward, and clinicians must be aware of this limitation in the interim. NIRS monitors measure regional brain tissue oxygenation in the frontal regions, and therefore any change seen reflects changes in this part of the brain only. Measurements also reflect only the tissue oxygenation in the outer cortex because beams are unable to reach deeper segments of the brain. Additionally, a local change may occur in a nonfrontal brain region that would not be detected by the NIRS monitor in the frontal region. Logistical challenges exist for smaller children who have smaller forehead sizes, and the probes may not fit well or may interfere with other forms of monitoring (e.g., EEG). Interference may also arise from other equipment at the bedside, such as respiratory or cardiac oscillations. Lastly, a tightly adherent probe is necessary for appropriate data collection, and frequent readjustment may be necessary because of probe dislodgement.

To further highlight the limitations and challenges of NIRS, we can look to the adult literature, in which recent studies have directly compared NIRS monitoring with invasive brain tissue oxygen (PbtO2) monitoring and found various levels of correlation in various pathologies. de Courson et al. [78] found no correlation between the two methods in a cohort of 51 patients with nontraumatic subarachnoid hemorrhage. These data were collected manually each hour, and there was no global correlation over time. When looking at patients with severe TBI, presumably a more global injury, studies have shown variable but present correlation between invasive and NIRS monitoring [79,80,81]. The differential information obtained from invasive and noninvasive monitoring may be because invasive PbtO2 monitors assess oxygen concentration of the interstitial tissue in the deeper white matter, whereas NIRS measures oxygen in intravascular space closer to the brain surface [82]. These two monitoring sources may have differential baseline values from different parts of brain tissue. Additionally, the sampled areas may also be affected differently by brain pathology; for example, if the invasive monitor is placed into injured tissue, local microvascular injury may lead to decreased diffusion of O2 that is not occurring globally in the whole brain monitored by NIRS. Together with other forms of multimodal neuromonitoring, invasive PbtO2 and NIRS monitoring may be complementary to each other, and the utility of each may be different depending on the type of injury.

Large prospective studies with patient-centered outcomes are needed to better understand the role of NIRS as a component of multimodal neuromonitoring in the PICU. Uniform data collection and reporting standards are needed to compare results across studies and sites, and adjustment for variation in proprietary algorithms is needed.

TCD Ultrasonography

Transcranial Doppler was introduced in the early 1980s by Dr. Aaslid, who described the ability of this device to measure cerebral blood flow velocity (CBFV) in the basal large cerebral arteries around the Circle of Willis through an intact skull bone using low-frequency pulsed Doppler ultrasound [83]. Because cerebral hemodynamic information from TCD is obtained in real-time at the bedside, and a single study can be done rapidly and repeated, TCD is not only a diagnostic tool but also an ideal tool for neuromonitoring [84]. Indeed, technological advances have improved TCD technology since its inception, leading to a broad range of potential applications in the PICU setting. TCD has been used in many different pediatric and adult patient populations in the intensive care unit and perioperative setting in the qualitative assessment of raised ICP, cerebral circulatory arrest, and high-intensity transient signals/microemboli as well as in the quantitative assessment of cerebral vasospasm, cerebral pressure autoregulation, and cerebral vasomotor reactivity [85].

TCD Methods and Devices

A low-frequency ultrasound probe (1–3 MHz) is placed over an acoustic window, which is a region of the skull that is thin and permits penetration of the ultrasound. The most common acoustic windows are transtemporal, suboccipital, and orbital (Fig. 4, Table 4). In infants under 12 months of age, the transfontanelle window can also be used. The vessel segments around the Circle of Willis insonated using TCD are shown in Fig. 5. TCD studies may be obtained using imaging (transcranial color-coded duplex ultrasonography [TCCD]) or nonimaging TCD devices depending on local availability and user preference (Table 5). Both modalities use pulsed wave Doppler to generate cerebral hemodynamic information through waveform analysis (Fig. 6). The conventional approach using nonimaging devices requires vessel identification to occur by knowledge of the typical depth, speed and direction of flow in relation to the probe, waveform morphology, and sound/pitch for each insonated segment. This approach does not require angle correction and assumes a zero angle between the probe and direction of blood flow (angle of insonation). Modern TCD devices have motion (M) Doppler mode, which uses 33 overlapping Doppler samples over 6 cm, which allows the user to visualize the depth at which flow is present and facilitating localization of the signal [86, 87]. TCCD provides anatomical B-mode images of insonated segments, allowing users to directly visualize the cerebral arteries. The TCCD transducer is a phase array probe operating at frequencies of 2 to 3.5 MHz. The TCCD user can identify landmarks such as the midbrain for orientation to the anterior and posterior circulations using color flow Doppler. CBFV is superimposed on the anatomical images of the blood vessels and flow is color coded with red, indicating flow toward the probe, and blue, indicating flow away from the probe. Once the vessels are identified, a defined region of measurement called a gate within the B-mode image is identified, from which Doppler shifts are recorded, creating a black and white pulse wave Doppler waveform from the sampled section. Pulse wave Doppler images from gated samples are often displayed alongside the color-coded B-mode image (Fig. 7). TCCD allows the user to visualize both the vessel and segment from which CBFVs are sampled. Angle correction can be applied with imaging devices, but generally this is not done with data acquired from TCCD.

Fig. 4
figure 4

Acoustic windows for imaging and nonimaging transcranial Doppler ultrasonography (illustration by Dr. Marina Mir)

Table 4 Insonated vessels in each acoustic window [88,89,90, 103]
Fig. 5
figure 5

Vessel segments around the Circle of Willis that can be insonated with transcranial Doppler ultrasound. Insonation of the large intracranial arteries with the ultrasound probe placed over the transtemporal window is demonstrated in the figure (illustration by Dr. Marina Mir)

Table 5 Imaging (TCCD) compared to nonimaging (TCD) devices
Fig. 6
figure 6

Waveform patterns in middle cerebral artery with imaging Doppler. a Bilateral high flow velocity pattern consistent with hyperemia in a school-aged patient with traumatic brain injury (TBI). b Unilateral high flow velocity pattern with a Lindergaard ratio > 3 suggestive of possible vasospasm in a school-aged patient after tumor resection. c Bilateral low-flow velocity high-resistance pattern in a teenager with severe TBI, suggestive of increased intracranial pressure (ICP). d Bilateral low-flow velocity pattern with systolic spikes and absent end-diastolic velocity (EDV) in a drowning infant suggestive of cerebral circulatory arrest. MCA, middle cerebral artery, PSV, peak systolic velocity, EDV, end diastolic velocity, Mean, mean velocity, PI, pulsatility index

Fig. 7
figure 7

Sample image from color duplex transcranial Doppler ultrasound (TCCD). MCA, middle cerebral artery, PS, peak systolic velocity, ED, end diastolic velocity, TAMAX, time-averaged maximum velocity (green envelope tracing), TAMEAN, time-averaged mean velocity (blue envelope tracing)

Both TCD modalities measure CBFVs in the large intracerebral arteries around the Circle of Willis and display spectral waveforms on the monitor. Under nonpathologic conditions, cerebral blood flow moves in one direction within the vessel lumen and demonstrates flow acceleration in systole and stepwise deceleration during diastole. Quantitative variables derived from TCD include peak systolic velocity (PSV), end-diastolic velocity (EDV), and mean flow velocity (MFV) (Fig. 8). Parameters derived from these measurements are the resistive index (RI) and the pulsatility index (PI), which reflect downstream vascular resistance. Of note, cardiac disease states, such as aortic insufficiency or persistent ductus arteriosus, may also influence measurements of vascular resistance.

  • PI = (PSV − EDV)/MFV: Normal values for adults are 0.6 to 1.2. Higher PI values indicate higher cerebrovascular resistance downstream from the site of insonation.

  • RI = (PSV − EDV)/PSV: Commonly used in the neonatal population, normal is < 0.8.

Fig. 8
figure 8

Normal transcranial Doppler ultrasound waveform with sharp systolic flow acceleration, stepwise deceleration, and end-diastolic flow observed above the baseline. Additionally, end-diastolic flow velocity values in the cerebral arteries are normally between 20 and 50% of peak systolic flow velocity values

Qualitative information that can be obtained by visual inspection of TCD waveforms includes the following:

  • Waveform morphology: Low-resistance waveforms are characterized by sharp systolic upstroke, stepwise deceleration, and robust EDV. High-resistance vascular beds are noted by sharp systolic upstroke, stepwise deceleration, and low EDV [91]. In nonpathologic conditions, the cerebral vascular bed is a low-resistance bed.

  • Pattern of flow: High-flow velocity pattern may indicate hyperemia (e.g., due to reperfusion injury following cardiac arrest, infection, hypercapnia) or arterial narrowing (e.g., vasospasm following subarachnoid hemorrhage or tumor resection, or vascular injury in TBI). Low-flow velocity pattern may suggest cerebral hypoperfusion, perfusion limiting increased ICP, or low cardiac output.

The integration of TCD into multimodal neuromonitoring may allow for the evaluation and analysis of model-based indices of cerebrovascular dynamics, including cerebral autoregulation, critical closing pressure, arterial time constant, and compliance of the cerebrovascular bed, among others. Supplemental Table 1 reviews concepts and calculations for model-based indices of cerebral hemodynamics. Noninvasive ICP estimation in children using a circuit model-based algorithm in a coupled analysis with time-synchronized arterial blood pressure waveforms obtained via radial artery catheterization and TCD-derived CBFV measurements may enable continuous, calibration-free, patient-specific ICP estimation in clinical scenarios in which invasive ICP monitoring is impractical [92, 93]. Improving in-depth knowledge of these factors calculated and observed in real time at the bedside may facilitate management of cerebrovascular pathologies.

It will be important to standardize the performance and reporting of TCD data to minimize interoperator variability [93,94,95] and reduce variability in results within and between centers. In 2020, a multidisciplinary group of TCD experts undertook a modified Delphi study and developed 34 recommendations in four domains for TCD practice in pediatric critical care (indication and request for examination standards, technical performance standards, data interpretation standards, and data reporting standards). A standard basic TCD protocol and a reporting worksheet were created based on the consensus statements [94]. If incorporated into multimodal monitoring, standardization of TCD data elements and reporting processes may also inform clinical decision-making [96].

Use of TCD Ultrasonography in Clinical Care and Research

Despite the lack of high-quality evidence supporting the utility of TCD in the PICU, the use of TCD has been explored in a variety of clinical disease states in critically ill children. A recent survey of 27 centers found that TCD was most commonly used in the evaluation and management of children in the PICU with intracranial/subarachnoid hemorrhage (20 hospitals), arterial ischemic stroke (14 hospitals), and TBI (10 hospitals) [97]. In these and other conditions, 74% (20 of 27) of the respondents reported that TCD was used to inform clinical care, such as when to perform neuroimaging, when to guide cerebral perfusion pressure management, and/or when to consider a surgical intervention. A review of the current pediatric research literature on TCD in pediatric critical care can be found in Supplemental Table 2. The majority of these studies investigated the use of TCD as a diagnostic tool. Investigations of the role of TCD as a neuromonitoring tool included its use in noninvasive estimation of ICP [93], its use in guiding resuscitation after cardiac arrest [98], and its use in guiding cerebral perfusion pressure augmentation in high-risk children with central nervous system infection [99].

To date, the use of TCD remains investigational in children, with the exception of those with sickle cell disease. For example, the use of TCD for ICP estimation is hypothesis-generating because research to date has not yet demonstrated in children how model-based estimation of ICP using TCD would perform when ICP is elevated above the threshold for treatment. At the present time, clear clinical indications for when TCD should be used in critically ill children remain unknown. In these authors experiences, TCD is used on an ad hoc basis when specific questions around neurovascular pathology arise that cannot be answered with traditional imaging techniques. Furthermore, societal-based guidelines in adults or children that delineate who should perform TCD do not exist. Future research directions should focus on guidelines for specific clinical applications for TCD in pediatric critical care, who should perform TCD (e.g., radiologists, sonographers and/or others), and which data elements are important to collect and how often.

TCD Limitations and Future Directions

TCD is operator dependent, and from the perspective of the clinician, mastery in the ICU setting requires expertise in the interpretation of findings in the context of underlying clinical conditions and treatments commonly encountered in the ICU. TCD-derived CBFV measurements are a surrogate measure of cerebral blood flow as long as vessel diameter and blood viscosity remain constant (flow equals velocity times area) [100]. Recently it has been reported in adults with delayed neurologic deterioration after subarachnoid hemorrhage that TCD-derived CBFVs in the large cerebral arteries were not reflective of cerebral perfusion disturbances in the downstream smaller arteries by positron emission tomography scan [101], potentially challenging the concept of cerebral vasospasm as a factor in this outcome. More work may be needed to better extrapolate when such circumstances may occur in children. There is ongoing work exploring how imaging TCD can provide an estimate of lumen diameter [102], which may allow for an improved understanding of actual cerebral blood flow and volume. Similar to the other noninvasive devices previously reviewed, further studies are needed to delineate how and when TCD may be incorporated into pediatric critical care and how it may impact clinical decision-making for the betterment of patient outcomes.