Introduction

Selective decontamination of the digestive tract should decrease the burden of nosocomial infectious agents attributable to the translocation of enteric flora, endotoxin absorption, and subsequent cytokinemia [1]. In France, the combination of two antibiotics, gentamicin sulfate and colistin sulfate, is recommended by the ANSM (the French National Agency for Medicine and Health Products Safety) for selective digestive decontamination [2]. Decontamination of the digestive tract is selective because there is no digestive tract absorption of these two antibiotics [2]. The approach is based on reducing the concentration of Gram negative bacilli (enterobacteriaceae and Pseudomonas aeruginosa) by in situ decontamination of the digestive tract. This drug is given orally in a hard capsule form which containing the both active substances (gentamicin sulfate and colistin sulfate) and the excipients (pregelatinized maize starch, microcrystalline cellulose, magnesium stearate and colloidal silica).

Gentamicin sulfate is a mixture of aminoglycoside antibiotic substances produced by Micromonospora purpurea and it is composed primarily of five closely related compounds [3] designated C1, C1a, C2, C2a and C2b (Fig. 1). On the other hand, colistin sulfate antibiotic substance produced by Bacillus polymyxa var. colistinus is a complex mixture of closely related polypeptides and is active against Gram negative bacteria. General structure of these polypeptides is a cyclic heptapeptide moiety bonded to a tripeptide side chain with an N-terminal fatty acyl residue (Fig. 2). It is a complex mixture of at least 30 different compounds, the main components of which are colistin A (polymyxin E1) and colistin B (polymyxin E2) that differ only in the fatty acid side chain. The sum of polymyxins E1, E2, E3, E1-1 and E1-7MOA accounts for a minimum of 77 % of the dried substance.

Fig. 1
figure 1

Structure of gentamicin sulfate components

Fig. 2
figure 2

Structure of colistin sulfate components

Consequently, simultaneous analysis of gentamicin sulfate and colistin sulfate is a by far real challenge, due to the multicomponent mixture of highly polar, non-volatile and non UV absorbing chromophores.

Detection

The chemical structures of gentamicin and colistin reveal the lack of a chromophore in the molecules. For colistin, direct low-UV detection at 212 nm is described in the European [4] and the US [5] pharmacopoeias and has been previously used to assay colistin in pharmaceutical formulations [6]. Derivatization methods with a chromophore or fluorophore were also described to increase sensitivity of the method [7] and mass spectrometry has been extensively used for stability studies [8].

Considerable work has been conducted to enhance the detection of gentamicin [9]. Refractive index detection proved to be insufficiently sensitive and incompatible with gradient elution. UV-detection after pre-column derivatization with o-phthalaldehyde is described in the US Pharmacopeia [10]. In most cases, pre- and post-column derivatizations with UV or fluorescence detection were found to be tedious, time-consuming and furthermore quantitative analysis is difficult due to the instability of derivatives or incomplete reaction of derivatization. Mass spectrometry is valuable for identifying and quantifying major and minor components, as well as for impurities or degradation products, but needs high cost instrument for routine analysis of pharmaceutical formulations [11]. Direct detection of gentamicin with electrochemical detection is the current official method prescribed in the European Pharmacopoeia [12]. However, this method is difficult in routine use and requires specific and costly instrumentation. Other direct detection methods can also be used, such as evaporative light scattering detection (ELSD) [9, 13, 14] or more recently, charged aerosol detector [15]. One drawback of the ELSD is that the detector response is exponential, rather than linear and explains why the range of linearity is restricted. Charged aerosol detector provides good linear signal and sensitivity with but this detector is relatively recent and is not yet commonly found in quality control laboratories. Direct capillary electrophoresis with low-UV detection of underivatized gentamicin has been conducted at 195 nm but was limited to the detection of gentamicin concentration range of 2–6 mg mL−1 [16].

Separation

Ion-pairing Chromatography

Colistin sulfate and gentamicin sulfate are both mixture of basic and polar species and are difficult to separate using reverse phase liquid chromatography. In an attempt at reaching successful separation, ion-pairing agents can be used to form ion pairs with the multiple amine functional groups of both gentamicin and colistin. The major compounds of gentamicin contain five amino groups (Fig. 1). The amino groups on ring I are the most basic (pKa ~ 9.6), whereas the pKa values of those on ring II (pKa ~ 5.6–8.0) and ring III (pKa ~ 7.5) are closer to physiological pH [17]. Polymyxin complexes (i.e. colistin sulfate) contain a polar moiety of five unmasked amino groups in each 2,4-diaminobutanoic acid with pKa about 10 [18]. The pKa of basic compounds such as gentamicin and colistin are shifted to lower values with increasing organic contents [19], while acidic species such as heptafluorobutyric acid (HFBA) are shifted to higher values. The shift is approximately 0.3 pH units per 10 % v/v organic phase and these basic compounds containing amino groups are positively charged in low pH mobile phases. Consequently, these compounds are necessarily retained by a combination of electrical (charge–charge) and hydrophobic interactions with the stationary phase and with the ions of the mobile phase. The mixed mechanism involves ion-pairing, ion exchange and hydrophobic partitioning [20, 21]. The mechanism of ion-pairing chromatography can vary between two extremes, i.e. ion exchange and partitioning similar to that in reverse phase chromatography. A mixed mode mechanism seems likely for many types of ion-pairing chromatography [22].

Elution Gradient

Polymyxin complexes are amphiphilic compounds containing a hydrophobic fatty acid moiety. They are less polar than gentamicin. Polarity gradient elution was in fact necessary to conduct their simultaneous determination as described for the separation of tobramycin and colistin in a pharmaceutical formulation [23]. The elution gradient enabled rapid and simultaneous analysis of two active substances of this pharmaceutical form. The goal of the present development was to quantitatively assay each drug substance after separation of colistin and gentamicin without the separation of their different constituents.

The aim of this work was to develop a rapid and simple chromatographic method for simultaneous quantitative determination in hard capsule of both gentamicin sulfate and colistin sulfate with low-UV detection after ion-pairing chromatographic gradient. The method was validated using criteria according to the ICH guidelines [24] and was evaluated in routine use.

Materials and Methods

Chemicals and Reagents

Hexafluorobutyric acid (HFBA) was used as the pairing reagent for HPLC separation was obtained from Sigma (Steinheim, Germany). Acetonitrile (Hipersolv® Chromanorm®, super-gradient grade) was obtained from Prolabo-VWR International SAS (Fontenay-sous-Bois, France). Water for HPLC analysis and dilutions was produced in-house with a Milli-Q water purification system (Millipore, Bedford, MA, USA). Gentamicin sulfate was supplied by Unipex (La Defense, France) and colistin sulfate by Inresa (Bartenheim, France). Both formulations (pediatric and adult) contained the same excipients: pregelatinized maize starch (Lycatab C®, from IMCD, Saint Denis, France) and microcrystalline cellulose (Emcocel 50 M®, Rettenmaier, Saint-Germain-en-Laye, France) as diluent, magnesium stearate (Univar, Fontenay-sous-Bois, France) as glidant and hydrophilic fumed silica (AerosIl 200®, Azelis, Paris, France) as lubricant.

Liquid Chromatographic Apparatus and Software

Chromatographic separations were carried out on an UltiMate® 3000 system (Dionex Corporation, USA) and consisting of a LPG-3400 A quaternary pump, an AS 3000 auto sampler fitted with a 300 μL loop, a SCM 400 vacuum degasser and a UV spectrophotometric multi-diode array detector. The detection wavelength was set at 215 nm and injection volume was 20 μL. The UltiMate® 3000 system was connected to a Chromeleon® 6.8 data acquisition and data processing software system. The analytical column was a Waters X-Terra® C18 MS (Wexford, Ireland) 2.5 μm, measuring 50 mm × 4.6 mm and was maintained at 35 °C. The mobile phase was a binary mixture of 0.05 % HFBA in water (phase A) and 0.05 % HFBA in acetonitrile (phase B) in a gradient elution mode at a flow-rate of 1.5 mL min−1. The mobile phases were filtered through 0.45 μm HVLP hydrophilic Millipore filters (Millipore, Molsheim, France) before use and gradient elution in final conditions is shown in Table 1.

Table 1 Gradient elution program used in the proposed method of determination

Preparation of Standard and Assay Solutions

Standard solutions were prepared daily by dissolving the appropriate amount of gentamicin sulfate and colistin sulfate in water to obtain final concentrations in the range of 0.6–1.4 mg mL−1 and 0.3–1.3 mg mL−1 for colistin sulfate and gentamicin sulfate, respectively. When excipients were added to the spiked calibration standard solutions or were present in real samples, the corresponding solution was sonicated for 1 min and then filtered through an Acrodisc® syringe filter with a GHP membrane (13 mm diameter, 0.45 μm pore size).

Validation Procedure

The HPLC–UV analytical method was validated for specificity, precision (repeatability and intermediate precision), linearity and accuracy.

The validation protocol was conducted on three consecutive days by the same operator and fresh solutions were prepared daily. Two calibration curves with five calibration standards have been prepared in the presence or absence of excipients (with or without matrix). The calibration points have been prepared with respect of the next protocol routinely applied. The matrix was the addition of excipients to the pediatric hard capsules and was the worst case, i.e. with higher excipients percentages in the capsule composition (16.5 and 32.9 %, for adult and pediatric formulations, respectively). The range of the calibration curve was determined to enable the determination of colistin and gentamicin from 80 to 120 % of the capsule contents for pediatric and adult forms. Nine quality controls were prepared daily with excipients at three levels (0.6, 1.0 and 1.4 mg mL−1 for colistin sulfate and 0.3, 0.7 and 1.3 mg mL−1 for gentamicin sulfate) and were injected once. The e-noval software V2.0. (Arlenda, Liège, Belgium) was used to compute all validation results.

Results and Discussion

Development of the Chromatographic Method

Compatible Mobile Phase Composition with Low-UV Detection

UV detection is one of the most widespread detection techniques used with liquid chromatography, but its use is obviously limited since it requires a chromophore group in the analyte structure. Colistin sulfate and more so especially gentamicin sulfate lack highly absorbing π-electrons on their structures. Indeed, the specific absorbance (\( A_{{1{\text{cm}}}}^{1\% } \)), corresponding to the absorbance of a 10 g L−1 solution in a 1 cm cell [25] of colistin sulfate and gentamicin sulfate at 215 nm, was 48.2 and 0.40, respectively (Fig. 3). The wavelength was chosen for its highest signal/noise ratio.

Fig. 3
figure 3

UV spectra of colistin sulfate (66.8 mg L−1), gentamicin sulfate (10.3 mg mL−1) and 0.05 % HFBA (v/v) in water and UV spectrum of pure acetonitrile (blank: ultra pure water)

Consequently, these molecules are barely detected and the sensitivity of UV-detection is very low. Since gentamicin fractions exhibit a very low specific absorbance and because absorbance is additive, co-elution of the five fractions should be envisaged to improve the detection sensitivity even with a large tailing peak. However, in our case, gradient elution is necessary for a shorten analysis run. Unfortunately, gradient elution caused drastic baseline drifts with low-UV detection. These drifts result from modifications of the composition of the mobile phase. Acetonitrile, which is a transparent mobile phase at 215 nm, exhibits moreover sufficient elution strength for paired biomolecules (Fig. 3) and was used in this study. HFBA is the ion-pairing agent which enables good retention of the two drug substances. The retention times of the two drugs increase with increasing hydrophobicity of the acid additive, explaining why HFBA was chosen because of the high polarity of gentamicin.

This ion-pairing has a specific absorbance at 215 nm (Fig. 3). The absorbance of an aqueous solution of 0.05 % (v/v) HFBA was 0.50 UA (\( A_{{1{\text{cm}}}}^{1\% } \) = 10) and concentration fluctuations during gradient elution resulted from baseline drift. The HFBA concentration and gradient elution were optimized as detailed in the following section.

Comparison of Two Gradient Elution Profiles

In the initial gradient (Fig. 4), the binary mixture consisted of aqueous solution with 0.1 % (v/v) HFBA, and ACN was HFBA free. First, gentamicin was eluted without baseline drift (Fig. 5a), then colistin was then eluted during baseline drift. These absorbance fluctuations of the mobile phase were related to variations of the HFBA concentration in the mobile phase. The hydrophobicity of HFBA anionic ion-pairing is due to hydrocarbon chain length (C3) and exhibits some amphiphilic properties [26]. Adsorption of HFBA on the stationary phase occurred and the hydrophobic chain was probably inserted in the bonded organic layer with the polar functional group protruding toward the mobile phase [26]. During gradient elution, the % of ACN increased after 3 min. The result on baseline drift was an increase of absorbance in the 3.7–4.5 min interval and then a linear decrease until 12 min. The increase of absorbance could be resulted from the change in HFBA partitioning between the stationary and mobile phases due to increasing elution. The HFBA concentration then decreased because ACN was HFBA free and could explain the decrease of absorbance after 4.5 min.

Fig. 4
figure 4

Gradient elution program used in the initially tested gradient and optimized chromatographic conditions

Fig. 5
figure 5

Comparison of chromatographic profiles obtained with the initial gradient (a) and the optimized gradient (b). For experimental conditions, see text

After optimization, the HFBA content was 0.05 % (v/v) (versus 0.1 % in chromatogram a). HFBA was added to ACN and the elution gradient profile was modified: baseline drift and analysis run time were minimized with good separation of gentamicin and colistin species (Fig. 5b).

The optimized gradient profile exhibited a number of improvements. First, the retention of gentamicin compounds was better (k = 3.2 versus k = 1.8 with the initial gradient). Second separation was more rapid (less than 10 min between two injections) and baseline fluctuations were controlled during colistin and gentamicin elution.

Interpretation of Negative Peaks

Three negative peaks (a, b and c) were observed on the chromatogram (Fig. 5b). The dwell time was estimated at 0.5 min by an injection of acetone, corresponding to the peak b. Peak a (tr = 0.3 min) was observed before the dwell time and the injection of aqueous sodium sulfate confirmed the presence of a negative sulfate peak at 0.3 min. This negative peak was proportional to the concentrations of colistin and gentamicin injected, since they are both sulfate salts. The chromatographic exclusion of sulfate (doubly negatively charged ion) probably resulted from the negative electrostatic repulsive force between sulfate anions and HFBA anions bonded to the stationary phase. Sulfate is a UV-transparent anion at 215 nm and indirect detection resulted from the absorbance of HFBA, reacting as a chromophoric probe. Peak c corresponded to the decrease of HFBA content in the mobile phase, since the solution injected is HFBA free. This negative peak disappeared when the injection solvent contained the same HFBA concentration than the mobile phase. On contrary, it became positive if the HFBA concentration in the injection solvent was higher than that of the mobile phase. The peak was eluted after the dwell time and illustrated the retention of HFBA on the stationary phase (k = 0.4).

Optimization of the Gradient Elution

The objectives of optimization were to minimize baseline drift and noise by optimizing the HFBA concentration, the proportion of ACN and the gradient profile; the retention factors of gentamicin and colistin depend on these parameters. Retention is in fact a combination of electrical (charge–charge) and hydrophobic interactions with the stationary phase and with the ions present in the mobile phase. The mixed mechanisms involved are (i) ion-pairing, (ii) ion exchange, and (iii) hydrophobic partitioning. The various equilibria were disrupted by variations in HFBA concentration as well as ACN content.

The quantities of HFBA adsorbed on the stationary phase increased with increasing HFBA concentration in the mobile phase [27], and thus increasing the baseline drift. It was important to note that equilibration time was not changed by the HFBA concentration and was short for HFBA ion-pairing due to the short chain length (C3) [27]. This short equilibration time was of interest in this gradient application to minimize the time of each run. Retentions of the two solutes increased with increasing HFBA concentration (from 0.01 to 0.1 % v/v) and decreasing of proportion of ACN.

From the point of view of detection, the ion-pairing agent affected detection because of its absorbance at 215 nm. The level of the ion-pairing agent may in fact have to be minimized to limit baseline noise, and baseline drift was primarily due to variations of the HFBA concentration during the analysis run.

Optimization of this separation had to take into account both constraints related to retention of the solutes and their detection. There are no simple rules to select the optimal concentration of HFBA and the gradient elution profile. To minimize baseline drift, the HFBA concentration had to be reduced with the risk to elute gentamicin in the void volume.

Initially, HFBA was added to the organic phase at the same concentration as that of the aqueous phase. The gradient profile and different HFBA concentrations were then evaluated.

Because of the differences in the retention behavior of gentamicin, colistin and HFBA, changing the gradient slope may alter adsorption of the ion-pairing reagent in the stationary phase to a different degree than gentamicin and colistin. HFBA selectivity may be explained by the concentration-dependent coating of the stationary C18 alkyl chain [28]. Gentamicin and colistin may elute at slightly different HFBA concentrations.

Initially, the gradient consisted of two main steps (Fig. 4): the first enabled isocratic elution of gentamicin and the second enabled the elution of colistin compounds, but retention of gentamicin was insufficient (k = 1.8). The isocratic elution of gentamicin with a lower proportion of ACN was associated with a tailing of the different peaks of gentamicin and a loss of sensitivity. For this reason, gradient elution was started by the injection with a lower ACN proportion, thereby eluting gentamicin in the gradient. Gentamicin could then be correctly retained (k = 3.2) and the various peaks of gentamicin tailing could also be reduced.

Influence of the Gradient Slope

Increasing the slope of the gradient increased baseline drift. Concomitantly, peak height increased and peak area remained constant.

Influence of the HFBA Concentration

With the optimized gradient slope described in Fig. 4, HFBA concentrations were evaluated from 0.01 to 0.10 % (v/v). Figure 6 illustrates the effect of HFBA concentration. It is known that the increase in retention resulted from interaction with acidic counter-anions. Increased retention depends on the concentration of the counter-ion in the mobile phase [19]. Retention factors, baseline drift and the background noise increased with increasing HFBA concentrations. The choice of an optimal HFBA concentration was a compromise between drift and the retention and is proposed at 0.05 % (v/v).

Fig. 6
figure 6

Influence of HFBA concentration (1: 0.025 % v/v; 2: 0.04 % v/v; 3: 0.07 % v/v) on baseline drift and retention

Validation of the Proposed Method

Specificity

The method was developed to measure the content of drug substances in the presence of other ingredients and excipients in a pharmaceutical product. The results showed no interferences between gentamicin and colistin and none of the excipients interfered with the drug substances.

Response Function

The response function is the relationship between the response observed and the concentration of the analyte in the sample. The best response function with or without excipients was obtained with linear regression without weighting. The response functions obtained by applying these regression models are shown in Table 2.

Table 2 Regression parameters for gentamicin and colistin in the presence and absence of excipients (n = 3)

The slopes of the calibration curves in water or in the presence of excipients were not significantly different (α = 5 %) for gentamicin and for colistin. The Y intercepts with or without excipients were not different (α = 5 %) but were significantly different from zero for gentamicin (α = 5 %).

No matrix effects due to excipients were observed. The standards for calibration curves and quality controls were prepared in water in routine use for analysis of real samples.

Precision

The precision of the method was determined by computing the relative standard deviations (RSDs, %) for repeatability and intermediate precision at each concentration level of quality control. The precision at each concentration level did not exceed 3.5 and 2.1 % for colistin and for gentamicin, respectively (Table 3), confirming that the method is convenient for quality control use.

Table 3 Precision parameters for gentamicin and colistin at three levels (n = 9)

Accuracy

Accuracy was determined by measuring the relative bias of quality control and mean recovery according to ICH recommendations (2005). As can been seen in Table 4, the relative bias (trueness) was lower than 2.4 % and thus acceptable. The confidence intervals (α = 5 %) of mean recovery included 100 %.

Table 4 Relative bias and confidence interval of mean recovery (n = 9)

Analysis of Real Samples

Six home-made batches of pediatric (n = 3) and adult (n = 3) capsules were evaluated by the validated analytical method (Table 5). The three pediatric batches complied with the test of uniformity of dosage units (n = 10) [29]. Two of the three batches tested for adults complied with the test after 10 or 30 dosage units. Batch A for adult was out of specification according to the specification of the European pharmacopeia (2.09.40): means and RSDs were equivalent for gentamicin and colistin, indicating a suitable blend of drug substances and excipients but incorrect capsule filling.

Table 5 Gentamicin and colistin sulfate contents expressed in % of the theory for pediatric and adult capsules

Conclusion

A very rapid and simple method for the simultaneous assay of gentamicin and colistin in hard capsules by ion-pairing gradient chromatography with low-UV detection has been developed. The method has been validated for specificity, precision, accuracy and linearity. The chromatographic method has been shown suitable for the quantification of gentamicin and colistin in routine use, for pediatric and adult hard capsule formulations.

In pharmaceutical analysis, the detection and quantification of analytes with weak chromophores is highly problematic and requires working at a low-UV wavelength. This application is an example of troubleshooting, i.e. baseline drift, due to gradient elution and absorbance of the ion-pairing agent and illustrates that the mixed mechanism involves ion-pairing, ion exchange and hydrophobic partitioning. Baseline drift was minimized by optimizing the HFBA concentration and the gradient slope and could be applied to other substances with low-UV absorbing chromophores.