Abstract
This article is a general overview of the potential capacities of plastic scintillators in radionuclide activity determination. Plastic scintillation (PS) behaves as liquid scintillation does, but with some differences related to the solid state of plastic scintillators. These differences are the base of some drawbacks and some advantages, related to the use of PS. This article describes how these capacities are affected by PS composition, sample preparation, scintillation mechanisms and quenching calibration procedures. It also describes the capabilities for alpha and beta emitter determination and discrimination through the use of PS microspheres and PS resins and their application to different types of samples and radionuclide determination.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
The first paper published on scintillation was probably “Scintillation counting with solutions” by H. Kallmann in 1950 [1]. Other articles followed [2, 3], a few of which focused specifically on plastic scintillators, such as the study by F. D. Brooks in 1960 “Pulse shape discrimination in a plastic scintillator” [4]. However, the book by J. B. Birks “The Theory and Practice of Scintillation Counting”, published in 1964 [5], can be considered the consolidation of scintillation as a potential technique for real and extensive use in laboratories to determine radionuclides in all kind of samples. In that book, Birks describes both liquid and plastic scintillation at the same level; however, their evolution in the following decades was quite different.
The potential of liquid scintillation was concentrated on beta-emitter radionuclide determinations favoured by the development of liquid scintillation counters (later spectrometers) and liquid scintillation cocktails [6,7,8]. Needs and interests converge on the development of nuclear activities together electronics and companies producing counters. Easy sample preparation and large numbers of applications described using the technique contributed to the spread of liquid scintillation [9].
Applications of plastic scintillation focused on the field of large area detectors [10,11,12], dosimetry [13,14,15], scintillation proximity assays [16] and more recently, neutron detection [17,18,19]. These fields developed later and their applications are important but reduced on number. Recently, interest in the application of plastic scintillation to radionuclide activity determination has increased, probably due to the fact that it avoids mixed waste production [20,21,22,23] and its capacity to integrate separation and detection processes.
This article presents a general overview of the capacities of plastic scintillation in radionuclide activity determination, based on our experience. This overview will include Plastic Scintillators composition and sample preparation; scintillation mechanisms, including the different types of quenching and the consequences on alpha/beta discrimination capacities; calibration; and different examples of direct and selective detection of alpha and beta emitters.
Composition
The composition of Plastic Scintillators (PS) is totally equivalent to that of liquid scintillation cocktails. PSs are solids that include a polymeric solvent and both primary and secondary scintillators, such as the classical PPO, POPOP, p-T or bis-MSB. Fluor molecules remain encapsulated inside the PS solid [24].
For water insoluble PSs, the solvents include linear polymeric chains based on styrene or vinyltoluene [25]. For PSs that are insoluble in organic and aggressive media, the polymer has to include a cross linker, as divinylbenzene, to create a three-dimensional structure [26]. The addition of the cross linker decreases the scintillation yield.
Liquid scintillation may also include a secondary solvent, as naphthalene or disopropilnaphthalene, to produce a temporal delay in the emission signal distribution. This addition is especially interesting for PS composition, to improve its capacity for alpha/beta signal discrimination [27].
PSs are solids that could be produced in any form. The most common for radionuclide determination in solution are microspheres (PSm), foils (PSf) and pellets (Fig. 1). Detection capacities depend, among other variables, on their size. Orientative values could be: for PSm, diameters of 10–300 µm, for PSf, widths of 50–100 µm, and for pellets, 3 mm. There are few suppliers for the production of the different compositions and formats [28, 29].
Sample preparation and reuse of PS
In plastic scintillation, sample preparation for measurement also follows the same steps as in liquid scintillation. A vial is filled with the appropriate amount of PSm and the active solution is added; finally, the vial content is homogenized using a shaking procedure such as a Vortex (Fig. 2). This last step is the main difference with regard to sample preparation for liquid scintillation and it makes plastic scintillation a bit longer.
The relationship PSm:sample solution is important for optimum detection conditions. The solution level should not be above the PSm level, because emissions in that part would not be detected; or below it, because the excess PSm would increase the background unnecessarily. As an example, for PSm of 60 µm diameter, the recommended proportion is 1.5 g:0.625 mL, irrespective of the vial volume.
The measurement procedure in liquid scintillation counters is performed in a totally equivalent way to that in liquid scintillation samples, including the use of external standard sources to determine quenching parameters.
Once measurements have been completed, sample solution and PSm can be segregated by filtration avoiding the production of mixed waste and allowing sample recovery and PSm reuse.
For most radionuclides, cleaning procedure is easy by simply washing the PSm with a solution of citric acid, 0.1 M and pH = 3 (for instance, in a flask and by shaking the mixture in an end-over-end system for 24 h) [30]. For some isotopes, such as americium, citric acid has to be substituted by 1 M hydrochloric acid. The main drawback of reuse is the time consumed in the washing process.
A comparison of the capacities of plastic and liquid scintillation is summarized in Table 1.
Mechanism
The transmission of the energy corresponding to the emitted particles to produce a final cascade of detectable photons follows a similar process in both liquid and plastic scintillation. The only difference is two additional quenching interferences: particle quenching and optical quenching.
The general process of energy transmission is shown in Fig. 3.
In a vial including an active sample solution and PSm, the beta or alpha particle emitted in the middle of the sample solution must pass through this phase for some distance before reaching the scintillator material. Once the PSm are reached, energy is transmitted and finally the fluorescent molecule emits the corresponding photons. The photons must then pass through a heterogeneous medium, changing from plastic to solution media on several occasions, before being detected by the photomultiplier tube. This series of successive changes of refraction index produces the corresponding optical processes of reflexion, refraction, dispersion and absorption that may reduce the number of photons finally detected.
Ionization, colour and chemical quenching are the three best-known processes that decrease the scintillation yield in a liquid scintillation detection process [31]. These are also present in plastic scintillation [32].
In both cases, ionization quenching, which is the reduction in the number of radiation-excited scintillation cocktail molecules as a consequence of ionization generated by the nuclear radiation with the associated reduction in photon intensity [9] depends on the scintillator/solvent composition and it represents a constant decrease for a defined measurement medium.
Colour quenching is also clearly an interference process when the sample medium shows some absorbance in the wavelength of the photons emitted by the scintillator [33]. Meanwhile, chemical quenching is related to interference in the solvent–solvent, solvent–scintillator or scintillator–scintillator energy transmission process that other, mainly electronegative, compounds present in the sample solution may produce.
Taking into account that this transmission process happens in the interior of the PS once the particle has reached its surface, it is reasonable to have some doubts about the extent of chemical quenching in plastic scintillation. The experimental results show that plastic scintillation is affected by chemical quenching just as liquid scintillation is [34]. This could be related to the fact that PS materials are quite “porous” and some of the compounds responsible of chemical quenching could penetrate the spaces in the polymer chains, at least close to the surface where transmission is initiated, and produce interference.
Particle quenching is related to the path that the emitted particle follows before arriving at the PS surface [33, 35]. Along that path, interactions with the medium decrease the energy of the less energetic emissions that finally are not detected (Fig. 4).
The length of the path increases as the diameter of the PSm does and consequently, the larger the PSm used, the lower the radionuclide detection efficiency.
Optical quenching is at the opposite extreme of the detection process [35]. This quenching is a consequence of the successive changes of refraction index encountered by the photons before they exit the vial. The number of solution-PS changes increases as the diameter of the PSm decreases. The consequence in this case is that detection efficiency decreases together with PSm size.
Thus, in terms of detection efficiency, particle and optical quenching act in opposite directions and a compromise is required with regard to the diameter of the PSm (or the structural organization of the sample solution and PS materials when foils or pellets are involved). Our experience has taught us that a convenient diameter range for PSm is 10–300 µm; in this range, PSm detection efficiencies for most alpha and beta emitters are similar to those for liquid scintillation, except for low-energy beta emitters such as 3H.
This is probably the main drawback of using PSm: the low detection efficiency for 3H. For a low activity level and short counting time, liquid scintillation is the recommended option.
Values of detection efficiency for different radionuclide and quenching calibration procedures are given in the following sections.
Detection capacities of plastic scintillators
PSs are sensitive to gamma, beta and alpha emissions, but only beta and alpha show detection efficiency values and spectra that are adequate for direct activity determination in solution samples at environmental activity levels.
Gamma and neutron detectors made of PSs are also used, but for other purposes that are beyond the scope of this overview.
Detection efficiencies of PSs depend mainly in their shape (microspheres or foils) and how they interact with the sample.
Plastic scintillators in the form of microspheres (PSm)
Beta emitters can be detected and quantified by plastic scintillation. Detection efficiencies corresponding to the measurement of aqueous solutions of these radionuclides using 20 mL polyethylene vials and PSm made of polystyrene + PPO + POPOP of diameters 15–150 µm are shown in Table 2. Spectrum distributions can be observed in Fig. 5.
From the table, it can be seen that detection efficiencies for high- and medium-energy beta emitters (90Sr/90Y and 14C) are maximal at around a 60 µm diameter, and similar to the values obtained in liquid scintillation [35].
For 3H, the values are clearly lower, as pointed out previously, mainly due to particle quenching. In spite of this decrease, detection is possible and can be used if activity is not very low. Some proposals, such as the use of pellets by E. Furuta, may increase the efficiency, but with some restrictions on the volume of the sample [36].
The effect of particle quenching becomes lower as the energy of the radionuclide increases. Thus, detection efficiency remains almost constant for 90Sr/90Y, whereas it decreases from 75 to 45% for 14C.
For all three radionuclides, spectrum distributions are similar and located approximately at the same energy position as for the corresponding results obtained by liquid scintillation. These positions are evidence of similar scintillation yields for PSs and liquid scintillation cocktails, and allow similar integration procedures to be applied for activity quantification.
Alpha emitter detection by PSs is also possible. Table 3 shows detection efficiencies for two alpha emitters of different energy (241Am at 5552 keV; and 236U at 4573 keV) using PSs of polystyrene + PPO + POPOP.
It is worth noting that, as in the case of beta emitters, detection efficiency decreases with the decrease of the energy of the particle emitted and with the increase of the PSm diameter. Both variations are related to particle quenching, as already mentioned. Detection efficiencies for PSm of 60 µm or smaller are high and equivalent to the values obtained by liquid scintillation since the mean path of alpha particles in water is around 50 µm (Table 4).
Spectrum distributions of the isotopes 241Am and 236U are shown in Fig. 6. The distributions are similar in shape, quite broad, and energy is in the same position as that obtained by liquid scintillation except for the tail at low energies. The origin of these low-energy signals is not clear, but it could be related to the distance between the emission point and PS and how this affects the energy of the alpha particle. When the diameter of the PSm is reduced (Fig. 7), the tail disappears and the spectrum is more similar to that obtained by liquid scintillation.
When the solution is evaporated to dryness onto PSm, the distance to the scintillator is greatly reduced for some of the alpha emissions, whereas for other, the density of the medium that they have to cross is also reduced significantly. Consequently, detection efficiency increases to values close to 100% and the spectrum band at low energies disappears while the peak at high energies becomes more sharp (Table 5; Fig. 8).
Plastic scintillators in the form of foils (PSf)
In this case, two scenarios can be described: when the radioactive solution is deposited on one foil or when the radioactive solution is placed between two foils. Values of detection efficiency for the situations for a beta emitter (36Cl) and an alpha emitter (241Am) are shown in Table 6.
It can be seen that in the case of one foil, the efficiency is less than 100%, since there are particles that are emitted in the opposite direction to that of the foil, Detection efficiency is higher than 50% probably due to the fact that some of radionuclide have penetrate a little inside the foil. However, detection efficiencies are high and make measurement viable. When the solution is enclosed between two foils, efficiency increases and almost all events are detected.
Detection efficiency calibration
The different quenching processes that may reduce signal detected in plastic scintillation have been discussed previously. These effects on detection efficiency can be calibrated following the same procedures applied in liquid scintillation: the use of experimental quenching parameters or following computation methods.
Experimental quenching calibration
In plastic scintillation calibration, the quenching parameters used are the same as in liquid scintillation [33]. Most of them are derived from the spectrum obtained from the Compton electrons produced by an external gamma source.
Colour and chemical quenching effects are calibrated using a series of standards of known activity and increasing the amount of the corresponding interfering compound. For colour quenching, Fig. 9 shows the decrease of detection efficiency for 36Cl and the SQP(E) parameter for increasing amounts of methyl orange. It can also be observed that the spectra shift to low energies as a consequence of the quenching effect.
Similar changes are produced by chemical quenching: the detection efficiency of 3H decreases encompassed within SQP(E) values when the amount of nitromethane in the solution increases (Fig. 10).
In contrast to this chemical and colour quenching, particle quenching shows totally different behaviour. As can be observed in Fig. 11a, the position of the 3H spectra remain almost constant despite changes in the concentration of Barium Chloride; whereas in Fig. 11b, the relationship between detection efficiency and salt concentration is clear. According to the constant spectrum positions, the values of the SQP(E) parameter also remains constant at 671 ± 3. This behaviour suggests that particle quenching totally avoids that the emitted particle was detected. As a consequence of this result, the classical quenching parameters, such as SQP(E), cannot be used to model the detection efficiency variation and they must be substituted for parameters related to concentration and easily determined as density.
Optical quenching effects are also different from those observed for colour and chemical quenching.
In this case, there is an apparent contradiction between detection efficiency and quenching parameter values (Table 7). Thus, whereas 36Cl detection efficiency decreases when PSm size increases, the associated SQP(E) values and the spectra distribution location (Fig. 12) follow the opposite trend, achieving higher values. This behaviour is explained by the combined effect of particle quenching and optical quenching. Particle quenching produces a decrease in the detection efficiency when the path through the solution before reaching the PSm increases. In contrast, optical quenching is a consequence of the multiple refraction index changes along the path of the photons. Thus, higher PSm diameters increase particle quenching (reducing detection efficiency) but reduce optical quenching (increasing the number of photons from each pulse and the SQP(E) value).
In summary, colour and chemical quenching can be calibrated in plastic scintillation as in liquid scintillation by using quenching parameters based on the use of an external standard source, whereas particle quenching can only be corrected for by using the solution density as a quenching parameter.
Optical quenching can be considered constant whenever the samples and calibration standards are prepared under the same conditions.
Computation calibration methods
Several attempts have been made in all radioactivity analytical techniques to develop computational models capable of simulating the detection process. Such simulations have not only been useful to provide better understanding of the physical phenomena associated with detection, but they have also been applied to determine, in combination with certain experimental data, the detection efficiency of detection systems. This is the case, in liquid scintillation, of the Ciemat-Nist and the TDCR methods [37,38,39,40].
The application of these methods to PSm measurements requires a good description of the scintillation phenomena and of the particular geometry of the PSm-solution system.
In an initial approach, a TDCR counter was used to determine the activity of several pure beta emitters using the same strategy as in liquid scintillation [41]. The results obtained showed a very high deviation from the reference value for low-energy beta emitters (Table 8). This was due to the fact that triple double detection efficiency (TDCR) computation methods used the theoretical beta spectrum and assumed that all beta particles interacted with the scintillator. This is valid for liquid scintillation, but in PSm measurements, not all beta particles reach the scintillator due to particle quenching and, as consequently, the energy distribution of the beta particles that reach the scintillator is different, especially for low- energy beta emitters. Therefore, the strategy used was to develop several geometries that simulate the system PSm-solution system and compute, through Monte Carlo simulation using the Penelope software, which was the energy distribution of spectra that achieves the scintillator (Fig. 13) and the probability of achieving the microspheres (Table 8). It can be seen how the original spectrum and the spectrum that arrive at the scintillator are different. To take into account this behaviour, a tracer radionuclide, 63Ni, was used to fit the correct geometry. The application of this model allows the activity of the rest of the radionuclide to be quantified successfully with a deviation lower than 3%.
In a second approximation, the Ciemat-Nist method was used to quantify several beta emitters using a Quantulus 1220. Moreover, a very-low-energy beta emitter was used (3H) to stress the robustness of the method [34]. The data provided by the Monte Carlo simulation with the Penelope software (efficiency and spectrum) was input into the Micelle2 code to compute the detection efficiencies and perform the activity quantifications. The results showed errors lower than 10% for 63Ni and lower that 3% for the rest, 3H included (Table 9). These results demonstrate that Monte Carlo simulation of particle quenching was adequate.
Alpha/beta discrimination
Alpha and beta emissions can be discriminated in liquid scintillation [42, 43]. Alpha particles produce more triplet excited states than beta particles during energy transfer to the scintillation cocktail. This situation leads to a temporal pulse distribution that is longer for alpha than for beta emissions and this is used for discrimination purposes.
This difference is not so important in plastic scintillation, probably due to the non-radiative transmission of the energy through the polymer chains that decreases the local concentration of triplet states [4, 44, 45].
The different behaviour of liquid and plastic scintillation for alpha and beta classification can be observed in Fig. 14, where misclassification for 3H, 90Sr/90Y and 241Am are represented for different pulse shape analysis (PSA) parameters. PSm composition is polystyrene + PPO + POPOP (120–150 µm) [46]. The higher misclassification of plastic scintillation measurements at low PSA values is related to the low-energy tails observed in the alpha spectrum (see Figs. 6, 7).
It has been reported that the addition of a secondary solvent produces a delay in the emissions [47]. In Fig. 15 it is possible to observe this effect in PSm including polystyrene + naphthalene + PPO + POPOP (120–150 µm). Both alpha and beta pulses are delayed and the final compromise in the PSA value leads to a reduction of misclassification to values slightly higher than those achieved with plastic scintillation [27].
The use of detectors with 3D spectrum capacities (intensity, energy and duration time) allows the results for PSm of the mentioned composition to be improved. The addition of a secondary solvent rotates the pulse time distribution in the opposite, anticlockwise direction, but more for alphas than for betas (Fig. 16).
Looking at this behaviour it is clear that pulse duration depends on the energy of the emission and that the use of a constant discrimination value is not the best option to separate the two distributions. When integrated areas of the 3D distributions are selected by using algorithms that take into account this dependence, the discrimination capacities of plastic scintillation (and probably liquid scintillation) improve significantly, as can be observed in Fig. 17. There, for the beta emitters (90Sr/90Y) misclassification is lower than 2% with a detection efficiency of 90%, and for alpha emitters (241Am) it remains under 1% with a detection efficiency of 25%.
Plastic scintillating resins: extractive scintillating materials
One of the most important drawbacks of scintillation techniques is their lack of selectivity as a consequence of the energy distribution of the beta and alpha emissions, and of the limited resolution and quenching interferences in the scintillation techniques.
With regard to liquid scintillation, plastic scintillation introduces a solid platform into the measurement vial useful to implement separation procedures.
Plastic scintillating resins (PSr)—extractive scintillating materials take advantages of this possibility to include, associated with the PS, an extractive material that allows the selective separation and determination of defined radionuclides [48,49,50].
This approach unifies into a single step the processes of separation and sample preparation for measurement, introducing some advantages such as reductions of time for sample processing, the manpower required and waste production. Among the disadvantages, it is worth mentioning that development of PSr is still in progress and few are commercially available, and that the cost could be higher than that of non scintillating extractive resins.
Composition and preparation approaches
PSr—extractive scintillating materials are composed of a solid support of PS microspheres coated with an extractant that is selective for the radionuclide of interest. To prepare these resins, several strategies can be applied: immobilization, impregnation, imprinted polymers and covalent bonding.
The most common strategy used to prepare the PSr is immobilization. In this strategy, the scintillating support is suspended in a volatile solvent, usually methanol, in which the selective extractant is dissolved. The methanol is slowly evaporated to deposit the extractant as quite a homogeneous and thin layer on the support (Fig. 18). This strategy has the advantages of being an easy process, easy to scale-up and capable of depositing a large amount of extractant and therefore obtaining a high capacity [51]. However, the resin stability depends on the extractant solubility in the sample solution and therefore PSr can present some breakthrough volume and, at the moment, are only suitable for aqueous samples [52]. Among others, this approach has been used to prepare a resin selective for 90Sr, to analyse water samples using a crown ether as the selective extractant.
Another strategy similar to the previous one is impregnation [51]. In this case, the support is suspended in a solution of the extractant and shaken for several hours to allow the absorption of the extractant onto the support. Later, the excess of extractant is removed by filtration and the resin is rinsed. This strategy has the advantages of being even easier than the previous one, but it also presents the same drawbacks together with the fact that it uses large amounts of extractant for the preparation and also that it is more difficult to control the conditions of the resin preparation. This approach has been used to prepare a resin selective for Sn for analysis of 126Sn in radioactive wastes using tributyl phosphate as the extractant.
Other approaches can be adopted to prepare PSr; more are still in development. Among them are imprinted polymers [53, 54]. In this strategy, a non-selective extractant, with capacity to be polymerized and to complex the stable isotope of the radionuclide of interest (called template), is added to the monomer solution and polymerized together with the other components of the PS support. The stable isotope is removed from the polymer leaving a cavity that is selective for the radionuclide of interest, due to its sterical hindrance characteristics and the disposition of the complexing groups. In some cases, the disposition of the polymer chains during the preparation in the absence of any template makes the support selective for some radionuclides such as radon; something that can be called a template-less imprinted polymer. In general, the imprinting strategy has the advantages of (i) not using selective extractants and therefore selective resins can be produced from generic complexant molecules, (ii) not having breakthrough volume and (iii) being suitable for different kinds of samples, included in organic or aggressive media. However, to apply this strategy, a stable isotope of the radionuclide needs to be used because large amounts of template are needed to produce the polymer and not all of the template can be removed from the polymer, increasing the background count rate of the PSr if the isotope used was radioactive. Also, the production of imprinted polymers for small molecules, such as metals, is tricky and several conditions and compositions need to be studied to produce the resin.
A final strategy is to covalently bond the selective extractant to the PS support. In this approach, different reactions are applied to functionalise the PS support with the selective extractant. In this way, the resin obtained has no breakthrough volume, but still uses selective extractants, with specific groups capable of reacting, to prepare the PSr. Additionally, the quantity of extractant that can be bonded to the PS support is smaller than in the case of immobilization and therefore it yields less capacity for the PSr [55]. This approach has been used to prepare an extractive scintillating polymer for the determination of 233U using phosphonic acid as the extractant.
Protocol of use
Once the PSr is prepared, its use is very similar to any other separation resin. First, the PSr is packaged inside a column or a cartridge. Then, the column or cartridge is connected to a vacuum chamber or a peristaltic pump to allow the solutions to flow throw it. The column or cartridge is first conditioned with the adequate medium, next, the sample is passed through and finally it is rinsed with the required medium. In the case of cartridges, just 2 ml of solution is enough to rinse and condition them. Finally, the column or cartridge is emptied of solution, disconnected and directly measured in any scintillation counter (Fig. 19).
Applications
Radionuclide quantification
The use of PSm as a routine alternative to the liquid scintillation cocktails has had limited success, probably due to the additional homogenisation step in the sample preparation process and because mixed waste production is not considered an important issue in practical terms.
However, PSs play an interesting role for specific determinations where liquid scintillation cocktails are not stable. For example, the determination of radiotracers in oil reservoirs that involves the activity determination in samples in highly salty matrices as could also be the case of other samples from biochemical studies.
The study of radiotracers was performed in collaboration with the Institute for Energy Technology (IFE, Norway) and the objective was to determine the activity of 14C (in the form of S14CN−) in samples, after a separation process involving high concentrations of NaClO4 [56]. The samples showed phase separation when they were mixed with liquid scintillation cocktails.
The samples were affected by particle and colour quenching and a double calibration using SQP(E) and density quenching parameters (Fig. 20) was applied. In the Fig. 20a, the double quenching effect can be observed: spectrum shifting (colour quenching) and area decrease in the same position (particle quenching).
The analysis of blind samples provided totally comparable results: IFE-LS: 4.23 dpm g−1 versus UB-PS: 4.32 (0.02) dpm g−1 and IFE-LS: 0.55 dpm g−1 versus UB-PS 0.59 (0.04) dpm g−1.
Plastic scintillation resins: extractive scintillating materials
Different PSr were prepared to resolve several analytically complex scenarios.
The first PSr developed was for the determination of 90Sr in water samples. In this case, the PS support was coated with a crown ether dissolved in 1-octanol as the selective extractant. Different parameters were optimized such as the separation media or the extractant proportion. The limit of detection for 10 ml of sample was 0.46 Bq L−1 with a quantitative retention. The resins were applied to the quantification of drinking, sea and river water samples with relative determination errors lower than 4% in all cases [57].
This same resin was also applied to develop a new procedure for emergency quantification of radiostrontium in milk. In this case, the column recovery and the effect of the different milk types and pre-treatment procedures were studied. The total time of the procedure was 5 h (including milk pre-treatment) and the resin presented good reproducibility and recovery (65%). The method was validated with the analysis of reference materials gave errors lower than 2%. Also, the PSr was applied to the quantification of different kinds of milk samples containing 90Sr and 89Sr, obtaining relative biases lower than 7% in practically all cases when total activity was determined.
A second part of the study of oil radiotracers performed with the IFE was to apply a PSr for the separation and analysis of the S14CN− radiotracer. In this case, the samples were seawater containing residues of oil without any previous separation step involving perchloride as in the previous example. In this case, Aliquat·336 was selected as the selective extractant. The samples were decanted and filtered and then passed through the columns without further treatment. For 100 mL of sample, the detection limit was 0.08 Bq L−1 and the retention was quantitative. The quantification of blind samples showed good agreement with the activities determined by the IFE using their method. Also the quantification errors for the spiked oil radiotracer samples were lower than 5% [58].
For the determination of 99Tc in water and urine, a new PSr was developed. In this case, Aliquat 336 was used as the selective extractant. First, the PSr production was evaluated showing good reproducibility of the preparation and storage stability. Also, common interferences, radioactive and non-radioactive, were evaluated and showed little influence on the 99Tc measurement. The PSr showed a detection limit of 0.36 Bq L−1 when 10 ml of sample was analysed and quantitative retention. Finally, the resin was applied to the measurement of seawater and urine samples showing quantification bias lower than 8% in all cases [59].
An additional resin was developed for the measurement of 210Pb, using a crown ether dissolved in 1-octanol as the selective extractant. In this case, the amount of extractant was optimized as well as the optimum separation medium being selected, showing a complete separation from its daughter products, 210Bi and 210Po, with a retention of 91%. This PSr was applied to the measurement of underground and river water samples with errors lower than 9% in all cases [60]. Additionally, this same resin was applied to the development of a dating procedure for bronze sculptures (Benin) using the decay of 210Pb isotope. The study proved the analytical viability of the method in terms of column performance and amount of sample required.
222Rn analysis
Another potential application of PSs, in the form of microspheres or foils, is the absorption and detection of 222Rn. Polymers have been widely described to be capable of absorbing 222Rn very efficiently. They can be considered template-less imprinted polymers. This capacity, together with the scintillation properties of the PSs, makes them attractive for the concentration and detection of 222Rn in several types of samples. PSs in the form of microspheres present very high surface areas, if compared to PS in the form of blocks or sheets. This fact enhances the absorption capacities of the scintillators.
In the test performed, PSm were exposed to an atmosphere with a known 222Rn activity for a controlled period of time. Later, the PSm filled a 20 mL glass vial and were measured in a commercial counter. The results showed a decrease in the count rate which presented very good correlation with the decay of 222Rn. The detection efficiency was 314% (a maximum of 500% due to 222Rn and its progeny could be expected) and the spectrum presented several broad bands (Fig. 21), compared to that of liquid scintillation with polycarbonate pellets to absorb 222Rn. These results can be attributed to optical quenching when a 20 mL vial filled with PSm is measured and can be highly improved in terms of efficiency and spectrum shape by the use of a thinner glass vial.
Regarding the absorption capacities of the PSm, the partition coefficients between the microspheres and the ambient medium (K) are 4.95 ± 0.54 in air and 37.2 ± 1.9 in water [61]. Comparing these results with other polymers, PSm present an acceptable capacity for 222Rn absorption, but lower than polycarbonates; and the absorption and desorption process is very fast. These properties make PSm very promising for use in fast systems of 222Rn detection, for example in soil–gas samples.
Finally, the diffusion model permits us to determine that PSm of 90 µm diameter should be the most appropriate for 222Rn absorption, since all the volume of the microsphere will be filled with 222Rn and no excess polymer or 222Rn will be present.
Sensors and automatic analysis
PSs and PSr—extractive scintillating materials may have several applications due to their specific capacities.
Continuous detection of alpha and beta emitters in a fluid (aqueous or non-aqueous stream) using liquid scintillation requires a homogeneous mixture of the sample and scintillation cocktail in the measurement cell. This mixture is sometimes unstable and the procedure continually consumes the cocktail and continuously produces mixed waste.
The use of PSs in the cell would be a good option to improve the stability of the measurements, but would probably decreasing the detection efficiency achieved.
This approach is already applied in some chromatography detectors and also in some sensors for water activity determination in environmental analysis [62,63,64,65,66,67,68,69].
Automatic analysis is a continuous demand as a consequence of the increasing amount of radiochemical analysis performed. Many proposals for commercial equipment have been introduced onto the market in the last few years [70, 71].
The introduction of PSs in measurement cells or PSr cartridges in the separation scheme may be an important improvement for reducing the resources required for the different determinations.
Challenges
Plastic scintillation is not a new field in scintillation research but its application for radionuclide determination opens new challenges for the future.
Within this research field, it would be interesting:
-
to know more about the scintillation mechanism in PSs;
-
to continue developing new plastic scintillating formats such as foils or monoliths;
-
to develop procedures for the preparation of selective scintillating and controlled porous materials.
In the application field, more studies would help:
-
to implement new options for automatic analysis designs or continuous monitoring systems;
-
to apply the developments to many areas of analysis such as routine monitoring, emergencies, decommissioning, medical uses and so on.
Conclusions
Plastic scintillation is becoming a useful technique for radionuclide analysis and determination, with some advantages and drawbacks.
Plastic scintillation is capable of determining alpha and beta emitting radionuclides and to perform alpha/beta discrimination based on pulse shape analysis. Among its limitations, the most important is the low detection efficiency for low-energy beta emitters and secondarily, the requirement for an additional step in sample preparation. Among the advantages, sample and scintillator can be segregated after measurement and mixed waste production is thus avoided.
One important potential contribution of PSs to radionuclide analysis is the possibility of developing selective plastic scintillating materials taking advantage of their solid state. This possibility offers a great opportunity for future work on radionuclide determination, because it opens the way for several potential improvements in the design of analytical procedures and new fields of application.
References
Kallmann H (1950) Scintillation counting with solutions. Phys Rev 78:621–622
Steinberg D (1958) Radioassay of carbon-14 in aqueous solutions using a liquid scintillation spectrometer. Nature 182:740–741
Schorr MG, Torney FL (1950) Solid non-crystalline scintillation phosphors. Phys Rev 80:474
Brooks FD, Pringle RW, Funt BL (1960) Pulse shape discrimination in a plastic scintillator. IRE Trans Nucl Sci 7:35–38. https://doi.org/10.1109/TNS2.1960.4315733
Birks JB (1964) The theory and practice of scintillation counting. Pergamon Press, Oxford
Reynolds GT, Harrison FB, Salvine G (1950) Liquid scintillation counters. Phys Rev 78:488
Gibsson JAB, Lally AE (1971) Liquid scintillation counting as an analytical tool. Analyst 96:681–688
Horrocks DL (1974) Applications of liquid scintillation counting. Academic Press, New York San Francisco London
L’Annunziata MF (2013) Handbook of radioactivity analysis. Academic Press, San Diego
Abel KH, Schilk AJ, Brown DP et al (1995) Characterization and calibration of a large area beta scintillation detector for determination of Sr-90. J Radioanal Nucl Chem 193:99–106
Nilsson J, Isaksson M (2011) A comparison between Monte Carlo-calculated and -measured total efficiencies and energy resolution for large plastic scintillators used in whole-body counting. Radiat Prot Dosimetry 144:555–559
Siciliano ER, Ely JH, Kouzes RT et al (2005) Comparison of PVT and NaI(Tl) scintillators for vehicle portal monitor applications. Nucl Instrum Methods Phys Res Sect A 550:647–674
Kirov AS, Hurlbut C, Dempsey JF et al (1999) Towards two-dimensional brachytherapy dosimetry using plastic scintillator: new highly efficient water equivalent plastic scintillator materials. Med Phys 26:1515–1523
Lambert J et al (2006) A plastic scintillation dosimeter for high dose rate brachytherapy. Phys Med Biol 51:5505
Beddar AS (2006) Plastic scintillation dosimetry and its application to radiotherapy. Radiat Meas 41:S124–S133
Bosworth N, Towers P (1989) Scintillation proximity assay. Nature 341:167–168
Zaitseva N, Rupert BL, PaweLczak I et al (2012) Plastic scintillators with efficient neutron/gamma pulse shape discrimination. Nucl Instrum Methods Phys Res Sect A 668:88–93
Pozzi SA, Mullens JA, Mihalczo JT (2004) Analysis of neutron and photon detection position for the calibration of plastic (BC-420) and liquid (BC-501) scintillators. Nucl Instrum Methods Phys Res Sect A 524:92–101
Hamel M, Simic V, Normand S (2008) Fluorescent 1,8-naphthalimides-containing polymers as plastic scintillators. An attempt for neutron/gamma discrimination. React Funct Polym 68:1671–1681
EPA Environmental Protection Agency (2001) Rule 40 CFR part 266: storage, treatment, transportation, and disposal of mixed-waste
Hsu J, Krieger JK (1991) Mixed waste: a review from a generators perspective. In: Ross H, Noakes JE, Spaulding JD (eds) Liq. Scintill. Count. Org. Scintill. Int. Conf. New Trends Liq. Scintill. Org. Scintill. Lewis Publishers., pp 557–600
Kalbhen DA, Tarkkanen VJ (1984) Review of the evolution of safety, ecological and economical aspects of LSC materials and techniques. In: Adv. scintill. counting. University of Alberta, pp 66–70
Tarancon A, García JF, Rauret G (2002) Mixed waste reduction in radioactivity determination by using plastic scintillators. Anal Chim Acta 463:125–134
Bertrand GHV, Hamel M, Sguerra F (2014) Current status on plastic scintillators modifications. Chemistry 20:15660–15685
Santiago LM, Bagán H, Tarancón A, Garcia JF (2013) Synthesis of plastic scintillation microspheres: evaluation of scintillators. Nucl Instrum Methods Phys Res Sect A 698:106–116
Bagán H, Tarancón A, Ye L, García JF (2014) Crosslinked plastic scintillators: a new detection system for radioactivity measurement in organic and aggressive media. Anal Chim Acta 852:13–19
Santiago LM, Bagán H, Tarancón A, Garcia JF (2014) Synthesis of plastic scintillation microspheres: alpha/beta discrimination. Appl Radiat Isot 93:18–28
Saint-Gobain Ceramics & Plastics I (2017). http://www.crystals.saint-gobain.com/products/plastic-scintillators
Technology eljen (2017). http://www.eljentechnology.com/products/plastic-scintillators
Tarancón A, García JF, Rauret G (2003) Reusability of plastic scintillators used in beta emitter activity determination. Appl Radiat Isot 59:373–376
Broda R, Cassette P, Kossert K (2007) Radionuclide metrology using liquid scintillation counting. Metrologia 44:S36–S52
Tarancón A, Alonso E, García JF, Rauret G (2002) Comparative study of quenching correction procedures for 90Sr/90Y determination by Cerenkov, liquid scintillation and plastic scintillation techniques. Anal Chim Acta 471:135–143
Santiago LM, Bagán H, Tarancón A et al (2013) Systematic study of particle quenching in organic scintillators. Nucl Instrum Methods Phys Res Sect A 698:26–36
Tarancón A, Barrera J, Santiago LM et al (2015) Application of the CIEMAT–NIST method to plastic scintillation microspheres. Appl Radiat Isot 98:13–22
Santiago LM, Tarancón A, García JF (2016) Influence of preparation parameters on the synthesis of plastic scintillation microspheres and evaluation of sample preparation. Adv Powder Technol 27:1309–1317
Furuta E, Iwasaki N, Kato Y, Tomozoe Y (2016) A new tritiated water measurement method with plastic scintillator pellets. Isot Environ Health Stud 52:560–566
Nähle O, Kossert K (2010) Comparison of the TDCR method and the CIEMAT/NIST method for the activity determination of beta emitting nuclides. Presentation in conference LSC2010, Adv. Liq. Scintill. Spectrom. 6–10 Sept. 2010, Paris
Cassette P, Broda R, Hainos D, Terlikowska T (2000) Analysis of detection-efficiency variation techniques for the implementation of the TDCR method in liquid scintillation counting. Appl Radiat Isot 52:643–648
Grau Malonda A (1999) Free parameter models in liquid scintillation counting. Colecc. Doc. CIEMAT
Pochwalski K, Broda R, Radoszewski T (1988) Standardization of pure beta emitters by liquid-scintillation counting. Appl Radiat Isot 39:165–172
Sanz AT, Kossert K (2011) Application of a free parameter model to plastic scintillation samples. Nucl Instrum Methods Phys Res Sect A 648:124–131
Jobbágy V, Waetjen U, Meresova J (2010) Current status of gross alpha/beta activity analysis in water samples: a short overview of methods. J Radioanal Nucl Chem 286:393–399
Pates JM, Cook GT, MacKenzie AB, Passo CJ (1998) Implications of beta energy and quench level for alpha/beta liquid scintillation spectrometry calibration. Analyst 123:2201–2207
Hawkes NP, Taylor GC (2013) Analysis of the pulse shape mechanism in a plastic scintillator with efficient neutron/gamma pulse shape discrimination. Nucl Instrum Methods Phys Res Sect A 729:522–526
Ranucci G (1995) An analytical approach to the evaluation of the pulse shape discrimination properties of scintillators. Nucl Instrum Methods Phys Res A 354:389–399
Bagán H, Tarancón A, Rauret G, García JF (2010) Alpha/beta pulse shape discrimination in plastic scintillation using commercial scintillation detectors. Anal Chim Acta 670:11–17. https://doi.org/10.1016/j.aca.2010.04.055
Rodriguez Barquero L, Grau Carles A (1998) The influence of the primary solute on alpha/beta discrimination. Appl Radiat Isot 49:1065–1068
DeVol TA, Egorov OB, Roane JE et al (2001) Extractive scintillating resin for 99Tc quantification in aqueous solutions. J Radioanal Nucl Chem 249(181):189
Duval CE, DeVol TA, Wade EC et al (2016) Stability of polymeric scintillating resins developed for ultra-trace level detection of alpha- and beta-emitting radionuclides. J Radioanal Nucl Chem 310:583–588
Roane JE, DeVol TA (2002) simultaneous separation and detection of actinides in acidic solutions using an extractive scintillating resin. Anal Chem 74:5629–5634
Song Y, Du Y, Lv D et al (2014) Macrocyclic receptors immobilized to monodisperse porous polymer particles by chemical grafting and physical impregnation for strontium capture: a comparative study. J Hazard Mater 274:221–228
McLain DR, Mertz CJ, Sudowe R (2016) A performance comparison of commercially available strontium extraction chromatography columns. J Radioanal Nucl Chem 307:1825–1831
Ye L, Ramström O, Mosbach K (1998) Molecularly imprinted polymeric adsorbents for byproduct removal. Anal Chem 70:2789–2795
Alexander C, Andersson HS, Andersson LI et al (2006) Molecular imprinting science and technology: a survey of the literature for the years up to and including 2003. J Mol Recognit 19:106–180
Duval CE, DeVol TA, Husson SM (2016) Extractive scintillating polymer sensors for trace-level detection of uranium in contaminated ground water. Anal Chim Acta 947:1–8
Bagán H, Hartvig S, Tarancón A et al (2009) Plastic vs. liquid scintillation for 14C radiotracers determination in high salt matrices. Anal Chim Acta 631:229–236
Bagán H, Tarancón A, Rauret G, García JF (2011) Radiostrontium separation and measurement in a single step using plastic scintillators plus selective extractants. Application to aqueous sample analysis. Anal Chim Acta 686:50–56
Bagán H, Tarancón A, Stavsetra L et al (2012) Determination of oil reservoir radiotracer (S14CN-) in a single step using a plastic scintillator extractive resin. Anal Chim Acta 736:30–35
Barrera J, Tarancón A, Bagán H, García JF (2016) A new plastic scintillation resin for single-step separation, concentration and measurement of technetium-99. Anal Chim Acta 936:259–266
Lluch E, Barreda J, Tarancón A et al (2016) Analysis of 210Pb in water samples with plastic scintillation resins. Anal Chim Acta 940:38–45
Mitev K, Dimitrova I, Tarancón A et al (2016) Pilot study of the application of plastic scintillation microspheres to Rn-222 detection and measurement. IEEE Trans Nucl Sci 63:1209–1217
Hofstetter KJ, Cable PR, Beals DM et al (1998) Field deployable tritium analysis system for ground and surface water measurements. J Radioanal Nucl Chem 233:201–205
Egorov OB, O’Hara JWMJJWG (2006) Equilibration-based preconcentrating minicolumn sensors for trace level monitoring of radionuclides and metal ions in water without consumable reagents. Anal Chem 78:5480–5890
O’Hara MJ, Burge SR, Grate JW (2009) Quantification of technetium-99 in complex groundwater matrixes using a radiometric preconcentrating minicolumn sensor in an equilibration-based sensing approach. Anal Chem 81:1068–1078
Grate JW, Egorov OB, O’Hara MJ, DeVol TA (2008) Radionuclide sensors for environmental monitoring: from flow injection solid-phase absorptiometry to equilibration-based preconcentrating minicolumn sensors with radiometric detection. Chem Rev 108:543–562
Schilk AJ, Knopf MA, Thompson RC et al (1994) Real-time, in situ detection of 90Sr and 238U in soils via scintillating-fiber-sensor technology. Nucl Instrum Methods Phys Res Sect A 353:477–481
Tarancon A, Padro A, Garcia JF, Rauret G (2005) Development of a radiochemical sensor, Part 2: application to liquid effluents. Anal Chim Acta 538:241–249
Wenzel U (1996) Online scintillation counting on Meltilex basis. J Radioanal Nucl Chem 203:87–96
Lochny M, Ullrich W, Wenzel U (1998) Simple on-line monitoring of α- and β-emitters by solid scintillation counting. J Alloys Compd 271–273:31–37
Villar M, Borràs A, Avivar J et al (2017) Fully automated system for 99Tc monitoring in hospital and urban residues: a simple approach to waste management. Anal Chem 89:5857–5863
Chung KH, Kim H, Lim JM et al (2014) Rapid determination of radiostrontium in milk using automated radionuclides separator and liquid scintillation counter. J Radioanal Nucl Chem 304:293–300
Acknowledgements
We thank the Spanish Ministerio de Economia y Competitividad (MINECO) for financial support under CTM2014-02020 HAR2014-56526-R, and the Catalan Agència de Gestió d’Ajuts Universitaris i de Recerca (AGAUR) for financial support under 2014-SGR-1277.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Tarancón, A., Bagán, H. & García, J.F. Plastic scintillators and related analytical procedures for radionuclide analysis. J Radioanal Nucl Chem 314, 555–572 (2017). https://doi.org/10.1007/s10967-017-5494-5
Received:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10967-017-5494-5