Introduction

Polycyclic aromatic hydrocarbons (PAHs) are produced as byproducts during fuel burning or carbon-containing organic substances and found in oil, coal and tar deposit [1]. Natural emissions including volcano eruptions and forest fires can also emit these compounds [2]. Due to their mutagenic, carcinogenic, and endocrine-disrupting properties, these compounds are included in the European Union (EU) and United States Environmental Protection Agency (USEPA) priority pollutant list [3]. In this way, identification and determination of PAHs in environmental samples is an important topic because of their adverse effects on humans and on soil organisms and plants, even at low concentrations of these compounds. However, in most cases, the amounts of PAHs are below detection limit of many analytical techniques, and matrix interferences come along with their determination. Hence, preliminary separation and pre-concentration of these compounds is usually required.

Current methods for the sample preparation and pre-concentration, in terms of extractant phase can be subdivided into solid phase based extraction [46], liquid phase based microextraction [7, 8] and solid-phase based microextraction [9, 10].

The most preferred method for sample preparation and pre-concentration of PAHs is solid-phase extraction (SPE) [11, 12] because of its high enrichment, simplicity, reduction of sample matrix effects and high recovery [13]. The bottleneck of SPE is the sample loading step, which can be performed by gravity flow or pressure/vacuum-assisted. Depending on the instrumental limit of detection, and therefore, on the amount of sample to be extracted, sample loading can require a relatively long period of time. Besides, in most cases, the aggregation tendency of solid particles reduces the active area that limits the full exploitation of the potential extractability.

Dispersive solid phase extraction (D-SPE) is a modified version of SPE that considerably reduces the time consumed, and simplifies the extraction process. In this method, extraction is not carried out in a cartridge, column or disk but in the bulk solution, which leads to more rapidity and ease of operation compared with the conventional SPE. The method consists of two critical steps: i) dispersion, and ii) phase separation. The second step is usually performed by centrifugation, which is very effective. However, it makes the overall procedure time-consuming. In this sense, development of a D-SPE method which could avoid the use of centrifugation is of great importance.

Nanoscale carbon-based materials, such as graphene (G) [14, 15], graphene oxide (GO) [16, 17] and carbon nanotubes (CNTs) are used as solid sorbent materials due to their ultrahigh surface area and high chemical stability. g-C3N4 sheets consist of a hexagonal ring-based carbon-nitrogen network, and they have a strong affinity for aromatic compounds based on large delocalized π-electron structure [14]. g-C3N4 has received vast interest and was applied in various fields, such as photocatalysis [18], hydrogen storage [19], lithium-ion battery [20], electrogenerated chemiluminescence [21] and fluorescent sensor. Notably, g-C3N4 can be used as a novel sorbent for SPE method.

Magnetic nanoparticles (MNPs) have attracted much attention in preconcentration of target analytes from different real samples, due to their gentle separation and non-destructive effects on biological analytes. As a main advantage, target analytes captured to them can be easily and selectively removed from the sample with an external magnet [22]. After washing out the contaminants, the isolated target compounds can be eluted and used for further analysis [23]. In this way, simple collection of the magnetic materials and chemical affinity of PAHs are combined. This results in the facile, fast and efficient preconcentration of the interested analytes. As a consequence, the g-C3N4/Fe3O4 nanocomposites may be favorably used as sorbents for preconcentration of PAHs.

In the present study, an air-assisted dispersive micro-solid phase extraction method (AA-dμ-SPE) (based on using magnetic graphitic carbon nitride nanocomposites as an adsorbents) is presented to extract the PAHs with a low level consumption of adsorbent and without using a toxic organic solvent. A few micrograms of the g-C3N4/Fe3O4 nanocomposites were transferred into aqueous sample solution in a glass vial, and then the mixture was repeatedly sucked into a glass syringe and then injected into the glass vial. By this action, the interaction between magnetic adsorbent and analytes increased more and more [24]. After performing predetermined cycles, the magnetic adsorbents were separated from water by a magnetic field. The adsorbed analytes were eluted with ethanol/acetone and 1 μL of the solution was injected into a gas chromatography with flame ionization detection (GC–FID) system for further analysis. Effective parameters such as amount of sorbent, number of extraction cycles, pH value, ionic strength, type and volume of eluting solvent were investigated to achieve the best conditions for extraction of the analytes from bio-fluid and water samples.

Experimental

Reagents and solvents

All chemicals used were of analytical grade. The PAHs standard, including naphthalene (Naph), fluorene (Fl), phenanthrene (Phen), anthracene (An), pyrene (Pyr), were purchased from Fluka (Buchs, Switzerland). Biphenyl (as internal standard) was purchased from Sigma (Steinheim, Germany). Solvents including methanol, ethanol, acetone, acetonitrile, ammonia and hydrochloric acid (HCl, 36 %) were obtained from Merck (Darmstadt, Germany). Sodium chloride, ferric chloride (FeCl3·6H2O), ferrous chloride (FeCl2·4H2O) and melamine were also obtained from Merck.

Standard stock solutions of five PAHs were prepared at 1000 μg mL−1 level in methanol for each compound. All solutions were stored in the dark at 4 °C. All working solutions were prepared by appropriately diluting the stock standard solution with ultra-pure water.

Instrumentation

Separation and detection of the analytes were performed by a gas chromatograph (GC-17 A, Shimadzu, Japan) equipped with a splitless/split injector and a flame ionization detector. Helium (purity 99.999 %) was used as the carrier gas at the constant flow rate of 4 mL min−1. The temperatures of injector and detector were set at 280 °C and 320 °C, respectively. The injection port was operated at splitless mode and with sampling time 1 min. For FID, hydrogen gas was generated with a hydrogen generator (OPGU-2200S, Shimadzu, Japan). A 30 m BP-10 SGE fused-silica capillary column (0.32 mm i.d. and 0.25 μm film thickness) was applied for separation of PAHs. Oven temperature program was: started from 50 °C, held for 1 min, increased to 190 °C at 30 °C min−1, held for 0 min, increased to 240 °C at 10 °C min−1 and then held for 3 min. A 10.0 μL ITO (Fuji, Japan) micro-syringe applied for the collection of sedimented organic solvent and injection into the GC.

A ultrasonic bath (SW3, Switzerland), was employed at a frequency of 50/60 kHz for speed up desorption of PAHs on the sorbents. The phase identification of g-C3N4/Fe3O4 nanocomposite power was obtained by X-ray diffraction (XRD) on a D8 Bruker XRD (USA) with Cu Kα radiation (35 kV, 15 mA, λ = 1.54051°A). Fourier transform infrared spectroscopy (FT-IR) spectra were recorded on a Shimadzu 8400 s (Japan) spectrophotometer. Thermogravimetric analysis (TGA) of g-C3N4/Fe3O4 was performed on a LINSEIS TG/DTA (STA PT 1600, Germany) thermogravimetric analyzer, the morphology of g-C3N4 and g-C3N4/Fe3O4 nanocomposite was observed an XL30 Philips (USA) scanning electron microscope.

Synthesis of g-C3N4/Fe3O4 nanocomposites

The g-C3N4 was synthesized by pyrolysis of melamine according to a reported procedure [25]. Typically, 20 g of melamine was put into an alumina crucible with a cover and heated at 550 °C for 4 h in muffle furnace with a ramp rate of 2.5 °C min−1. The obtained yellow product was the g-C3N4, and then the bulk g-C3N4 was ground to powder. The g-C3N4 / Fe3O4 nanocomposites were prepared by an in situ precipitation method [26]. In a typical procedure, g-C3N4 (125 mg) was dispersed in 500 mL of ethanol/water (1:2) and ultrasonicated for 5 h at room temperature. FeCl3·6H2O (1.838 g) and FeCl2· 4H2O (0.703 g) were dissolved separately in 20 mL of double-distilled water and added to the suspension of g-C3N4. The mixture was stirred at 80 °C for 30 min, and then 10 mL of ammonia solution (NH4OH) was quickly injected into the reaction mixture. The resulting mixture was stirred for another 30 min, after which the reaction mixture was cooled and washed several times with double-distilled water and absolute alcohol. Finally, obtained precipitate was dried in air at 80 °C for further characterization (Fig. 1).

Fig. 1
figure 1

The photographic pictures of depositing Fe3O4 nanoparticles on a g-C3N4

Extraction procedure

g-C 3 N 4 /Fe 3 O 4 (15 mg) as a adsorbent, 5 g NaCl and 100 mL of the analytes standard mixture solution (0.1 μg mL−1 of each analyte) were placed into a glass vial and solution pH adjusted (pH = 7) with 0.1 M HCl and 0.1 M NaOH. By using a glass syringe (10 mL), solution was repeatedly sucked and injected into the glass vial. By this action the interaction between adsorbent and analytes increased more and more. After performing 30 cycles, the adsorbent was separated from solution by a magnetic field during 2 min. 1 mL ethanol/acetone was added to the adsorbent and the glass vial was subsequently subjected to ultrasonic irradiation for 5 min to speed up desorption of the analytes. Afterwards, 1 μL of the collected eluent was injected into a GC for further analysis. The results were reported based on relative response that is defined as proportion of analyte peak area to peak area of internal standard (biphenyl). Figure 2 shows the schematic representation of AA-dμ-SPE.

Fig. 2
figure 2

Schematic representation of AA-dμ-SPE. a 100 mL sample solution and 15 mg of adsorbents. b Suction and injection of solution using glass syringe was caused to increase of interaction between adsorbents and analytes c Adsorbents were separated from water by a magnetic field d Aqueous phase removed easily from glass vial e Adsorbed analytes eluted and then injected to GC

Sample preparation

Saliva samples were collected from smoker volunteers (aged between 25 to 35 years) in our laboratory. Two types of saliva samples were prepared from a healthy volunteer in our lab: i) blank saliva samples (analytes-free) and ii) model saliva samples. Blank saliva samples were prepared from the smoker volunteers that not smoked cigarettes at least for 1 week, whereas model saliva samples were prepared from the smoker volunteers 30 min after smoking (addicted to smoking cigarettes, at least for 1 month).

An aliquot of 10.0 mL saliva sample was directly transferred to a volumetric flask and diluted to 100.0 mL. Then, the extraction was performed with the AA-dμ-SPE method.

Blood samples were collected from smoker volunteers into Pyrex centrifuge tubes. The tubes were stored on ice for a maximum of 15 min before centrifugation. Blood samples were centrifuged at 10,000 rpm. After separation of plasma in the collection tube with centrifuge, 2.0 mL of upper phase was immediately transferred to a volumetric flask, diluted to 100.0 mL and the extraction performed with the AA-dμ-SPE method.

Results and discussion

The method is based on the withdrawing and pushing out a mixture of sample solution and adsorbent in a glass tube using a single syringe. This simple action leads to efficient dispersion of adsorbent in the sample solution. Collection of the adsorbent through the extraction (adsorption and desorption steps), was easily performed using a magnet. At the end, the best conditions were selected based on the highest peak areas (or recoveries).

Choice of materials

g-C3N4 has a stacked two-dimensional structure (2D sheets of tri-s-triazine connected via tertiary amines) and is made from simple precursors under ambient conditions with a low cost. However, dispersity of g-C3N4 in water is much better than graphene (G). These properties suggest the possibility of g-C3N4 as a new sorbent for AA-dμ-SPE for the extraction and preconcentration of polycyclic aromatic hydrocarbons. On the other hand, for the simple collection of nanomaterials, immobilizing Fe3O4 particles on the surface of nanomaterials is an effective solution, as it can be easily separated from solution with an external magnet.

Characterization of g-C3N4/Fe3O4

The XRD patterns of the samples are shown in Fig. 1S. The XRD peaks of g-C3N4 and Fe3O4 were observed in the XRD pattern of the synthesized g-C3N4/Fe3O4. Diffraction peaks of Fe3O4 appeared at the Bragg angles of about 30.02o, 35.61o, 36.20o, 43.28o, 53.16o, 57.18o, 62.78o and 73.57o are respectively ascribed to the (220), (311), (222), (400), (422), (511), (440) and (533) that were in good agreement with the standard magnetite Fe3O4. The XRD pattern of the synthesized g-C3N4 shows two broad peaks at 2θ = 13.2 o and 27.3 o. It is well known that g-C3N4 is based on tri-s triazine building blocks. The strongest peak at 27.3 o is close to the peak position of graphite (002), and is due to the interlayer stacking of aromatic systems. The small angle peak at 13.2° is associated with interlayer stacking [27].

Figure 2S shows the FT-IR spectra of pure g-C3N4 and the g-C3N4/Fe3O4 nanocomposites that confirmed the successful formation of g-C3N4/Fe3O4. The broad Fe-O band in the region from 550 to 650 cm−1 is clearly visible at all compositions [26]. The strong absorption peaks can be observed in the range of 1200–1600 cm−1, which corresponded to the stretching mode of C = N bonds. The broad peaks in the range of 3000–3700 cm−1 were attributed to the adsorbed H2O molecules and NH stretching vibration. Additionally, the absorption peak at 810 cm−1 was assigned to the characteristic breathing mode of the triazine units [25]. The results indicate that the aromatic g-C3N4 owns specific structure leading to its unique absorption ability.

The SEM and TEM images, as shown in Fig. 3, exhibited that Fe3O4 nanoparticles were successfully deposited on the surface of the g-C3N4 sheets. The surface morphology of the synthesized g-C3N4 is seemed to be smooth, after in situ deposition of Fe3O4 nanoparticles onto the surface of the g-C3N4 (Fig 3c and d). The TEM images (Fig 3e and f) were obvious showed that the distribution of Fe3O4 on the surface of g-C3N4 is uniform and prevented the agglomeration of Fe3O4 nanoparticles. Also, no free Fe3O4 nanoparticles were found outside of the g-C3N4 sheet. According to the Fig. 3, the particle size of g-C3N4 and Fe3O4 are 70–180 and 8–12 nm, respectively.

Fig. 3
figure 3

SEM images (pure g-C3N4 (a, b) and g-C3N4/Fe3O4 (c, d)) and TEM images (g-C3N4/Fe3O4 e, f)

TGA curve of g-C3N4/Fe3O4 was shown in Fig. 3S. TGA analysis was used to determine the thermal stability of the g-C3N4/Fe3O4 nanocomposite. The g-C3N4/Fe3O4 began to decompose when the temperature was over 500 °C, and the total decomposition occurred at the temperature of 690 °C.

The porosity and effective surface area measurements for the g-C3N4/Fe3O4 nanocomposites were performed by the N2 equilibrium adsorption isotherms at 77 K (Table 1). The BET surface area of the adsorbent used was 137.713 m2g−1, with a pore volume of 0.388 cm3g−1. Also the average pore diameter was found to be 2.371 nm, indicating an appreciable narrow micro-porosity for the g-C3N4/Fe3O4 nanocomposites.

Table 1 Surface area and porosity of g-C3N4/Fe3O4 nanocomposites

Magnetic properties of the g-C3N4/Fe3O4 nanocomposite

To collect and reuse the adsorbent in this study, its preparation with good super-paramagnetism is essential. The magnetic properties of g-C3N4/Fe3O4 nanocomposites were characterized by a vibrating sample magnetometer (VSM) at room temperature and saturation magnetization value is 44.49 emu g−1. The magnetization curve of g-C3N4/Fe3O4 nanocomposites is shown in Fig. 4S. This nanocomposite has been successfully applied as a magnetic adsorbent in air-assisted dispersive micro solid phase extraction method.

Optimization of air-assisted dispersive micro-solid phase extraction conditions

In order to evaluate the application ability of the g-C3N4/Fe3O4 nanocomposites in extraction of PAHs in bio-fluid and water samples, the effective parameters on the adsorption and desorption should be investigated and optimized. A 0.1 μg mL−1 of solution mixture was used in the optimization of variables. Effective parameters on the adsorption of PAHs on the g-C3N4/Fe3O4 surfaces such as amount of adsorbent, number of extraction cycles, ionic strength, pH, and desorption such as type and volume of elution solvent were studied. To achieve the optimum extraction conditions, one-at-a-time (univariate) optimization strategy was used.

Amount of adsorbent

Varying amounts of g-C3N4/Fe3O4 in the range of 5.0–30.0 mg were evaluated by extracting 100 mL of sample solution, the result being shown in Fig. 5S As it can be observed, by increasing adsorbent amount, analytical signals increase till 15.0 mg and then remain constant. Hence, 15.0 mg of g-C3N4/Fe3O4 was selected for the further studies.

Number of extraction cycles

In this work, the sample solution was rapidly sucked into a 10 mL glass syringe and then was injected into the tube. The numbers of suction/injection cycles are considered as number of extraction cycles. To reach equilibrium status, numbers of extraction was studied in the range of 5–40 times. The results obtained in Fig. 6S show that by increasing extraction numbers, analytical signals also increase till 30th extraction and then remain constant. Hence, 30 times of extraction was selected for the further studies. It is noted that this step was performed in less than two minute.

Effect of ionic strength

To evaluate effect of the ionic strength in this study, efficiency of extraction from aqueous solution containing 0.1 μg mL−1 of each target analyte was investigated in presence of different amounts of sodium chloride (0–10 %, w/v). Generally, salt addition can decrease the solubility of analytes in the aqueous phase (salting-out effect), thereby improving the extraction efficiency. However, with the increase of ionic strength, the viscosity of the solution increases and the efficiency of the mass transfer process subsequently decrease due to the viscous resistance effect. The results in Fig. 7S indicate, recovery of the analytes increased up to 5 % (w/v) sodium chloride and after that remained nearly constant in the range of 5–10 % (w/v) sodium chloride. Finally, 5 % w/v sodium chloride was chosen as a suitable concentration.

Effect of pH

The pH value of the sample also affects extraction efficiency. The effect of sample pH on the extraction efficiency of analytes was investigated in the range of 3 to12 (Fig. 8S) with the adjustment with HCl or NaOH. Although PAHs exist as neutral molecules under ordinary conditions and are unlikely to be influenced by a change in the solution pH, the pH of the solution may affect the g-C3N4/Fe3O4 formation and dispersity in water. The results indicated that with decreasing in pH, the analytes recoveries were decreased in the pH range of 3–7. Acidic conditions may cause some leaching of Fe3O4, thereby decreasing the adsorption of PAHs on the g-C3N4/Fe3O4 nanocomposite. In the pH range of 7–12, recovery remained nearly constant. Hence, pH = 7 was selected for the further studies.

Type of solvent elution

Selecting a suitable elution solvent is critical in the AA-dμ-SPE process to achieve high analytical signal and extraction efficiency depending on the properties of eluent. In order to find the suitable elution solvent, four organic solvents and mixture including, acetonitrile, acetone, methanol and ethanol/acetone were compared for their eluting efficiency. As shown in Fig. 9S, desorption efficiency was significantly affected by the eluent and ethanol/acetone resulted the highest elution efficiency for the target analytes. Therefore, ethanol/acetone was selected as an optimum solvent elution.

Volume of elution solvent

To examine effect of volume of elution solvent, experiments involving different volumes of elution solvent (0.75, 1.0, 1.5, and 2.0 mL) were studied. Fig. 10S shows the influence of volume of elution solvent on the analytes extraction. It can be observed that recovery of analytes increased with increasing volume of elution solvent up to 1.0 mL and then decreased. In lower elution solvent volumes, elution of analytes are not completed. High recoveries along with good repeatability were obtained when 1.0 mL of elution solvent was used. Therefore, 1.0 mL was selected as the optimum volume of the elution solvent in this study. The best experimental conditions were found to be as: amount of adsorbent: 15.0 mg; number of extraction cycles: 30 times; ionic strength: 5 %, NaCl; pH value: 7.0; elution solvent: 1.0 mL of ethanol: acetone (1:1).

Analytical performance

In order to investigate the practicality of the developed method, validation parameters including the linear ranges of calibration graphs, limits of detection (LODs), limits of quantification (LOQs), square of correlation coefficients (r2), relative standard deviations (RSDs) and enrichment factors (EFs) were studied under optimized extraction conditions. The results obtained are summarized in Table 2. Linear ranges are between 1.0 and 100 ng mL−1 for all target analytes with square of correlation coefficients ranging from 0.9977 to 0.9990. Limits of detection (S/N = 3) and quantification (S/N = 10) were in the range of 0.3–0.6 and 1.0–2.0 ng mL−1, respectively. The precision (in terms of RSD) of the method, calculated for six replicate measurements at concentration level 10.0 ng mL−1 of mixture PAHs, were in the range of 3.49 % to 5.94 %. The enrichment factors (EFs), defined as the ratio between the analyte concentration in the final extracted phase and the initial concentration of analytes, were in the range of 59.0 to 98.5.

Table 2 Analytical figures of merit for the AA-dμ-SPE method

Effect of foreign species

To test the applicability of this method in different matrix, the effects of the possible interfering foreign species on the simultaneous preconcentration and determination of PAHs were considered under the optimum conditions. The results showed that the presence of 1000-fold benzoic acid, salicylic acid, chlorophenols (including of 2-chlorophenol, 3-chlorophenol, 4-chlorophenol, 2,4,5-trichlorophenol) and 500-fold sedative drugs (including tramadol and acetaminophen) have no interference on the analytes recoveries (Table 1s).

Reusability of g-C3N4/Fe3O4

In this work, to investigate the recyclability of the adsorbent, g-C3N4/Fe3O4 was reused in AA-dμ-SPE method. After seven times of recycling, there was no obvious decrease or increase in the analytes relative response. Based on the obtained results, g-C3N4/Fe3O4 did not show any memory effects during successive extraction (up to seven times), indicating good reusability of this compound.

Comparison of the method with other works

This method was compared with other previous extraction methods that were used for the determination of PAHs in real samples [2, 3, 17, 2833]. Table 3 demonstrates the amount of adsorbent, linear range, limits of detection, enrichment factor, and extraction time of some analytical methods along with this work. As can be seen in Table 2, the present method requires a lower amount of adsorbent in comparison to most of other methods and lower extraction time (3 min) in comparison to others. In general, linear range of method is wider than those of the mentioned methods. Hence, this method has several advantages over the other reported techniques, while it is also very simple, rapid and less hazardous for environment.

Table 3 Comparison of different methods for PAHs extraction

Real sample analysis

In the optimal conditions, the method was applied for determination of PAHs in the different real samples including saliva, blood and waste waters of Semnan city. The results are listed in Table 4. The recoveries of PAHs for spiked samples were in the range from 94.94 % to 98.60 %. None of the analytes were detected in blood and waste water. Target analytes was found in the saliva sample. The concentrations of naphthalene, fluorene, phenanthrene, anthracene and pyrene were calculated as 68.7 ± 3.8, 89.3 ± 4.9, 65.0 ± 4.0, 31.3 ± 2.1 and 23.0 ± 1.5 ng mL−1 (n = 3), respectively. The typical chromatograms for saliva samples are shown in Fig. 4.

Table 4 Results for the determination of PAHs in saliva and blood sample
Fig. 4
figure 4

GC-FID chromatograms of standard (with 10 μg mL−1 of each analyte), saliva and blank. In all cases, AA-dμ-SPE method was performed and 1 μL of the extraction phase was injected into GC

Conclusion

In this study, air-assisted dispersive micro-solid phase extraction as a sample preparation method coupled to GC-FID was developed for the pre-concentration/determination of PAHs in waste water, saliva and blood samples. Different techniques including FTIR, XRD, TGA, SEM and TEM confirmed the successful synthesis and preparation of the adsorbent (g-C3N4/Fe3O4). The method is simple, efficient and environmental-friendly because a little amount of adsorbent was used, without using a toxic organic solvent as eluent. The main advantage of this method is simplicity of collection of g-C3N4/Fe3O4 nanocomposites based on magnetic properties, short extraction time in comparison to other methods. The results revealed that the developed method was suitable for determination of PAHs at ng mL−1 levels in complex matrices such as waste water, saliva and blood samples.