Introduction

Percutaneous vertebral cementoplasty (vertebroplasty) is used for pain palliation and structural stabilization [15] as the sole therapeutic procedure or in combination with vertebral body ablation techniques [610]. Fluoroscopy and computed tomography (CT) guidance techniques are commonly used [5, 1113], but they result in variable radiation doses, which may lead to high exposure levels of ionizing radiation [11, 14, 15]. Furthermore, the low soft-tissue contrast resolution of both techniques may limit the conspicuity of anatomic structures and vertebral body lesions with little or no osseous destruction.

As potential alternative guidance technique, interventional magnetic resonance imaging (MRI) imaging offers favorable attributes for vertebroplasty, such as no ionizing radiation, true multiplanar cross-sectional imaging, and unparalleled contrast resolution [1621]. MRI guidance has also the potential to facilitate temperature mapping of concomitant thermal vertebral body ablation [22, 23].

Successful MRI-guided vertebroplasty require accurate targeting and needle placement, which are challenging because of limited space inside the bore of widely available 1.5-Tesla MRI systems. Augmented reality navigation employing image overlay technology may solve this problem by providing navigation outside the bore [24]. Therefore, the purpose of this study was to prospectively evaluate the feasibility MRI-guided vertebroplasty using augmented reality image overlay navigation in human cadavers.

Materials and Methods

System Description

A commercially available 1.5-Tesla MRI system (MAGNETOM Espree; Siemens Healthcare, Erlangen, Germany), a MRI-compatible augmented reality image overlay navigation system (Fig. 1) [25], and 10 cm-long, 11G MRI-compatible vertebroplasty needle (Somatex, Teltow, Germany) were used. 3D Slicer (The Brigham and Women’s Hospital, Inc., Boston, MA, USA) open-source medical image analysis and visualization software (http://www.slicer.org) and the loadable module PerkStation (Queen’s University, Kingston, ON, Canada) were used for display of MR images, identification of targets, planning of virtual needle paths, and needle navigation [26, 27] (Fig. 2A).

Fig. 1
figure 1

Interventional setup with image overlay navigation system. Computer-aided design shows a clinical 1.5-Tesla MRI system and the stand-alone augmented reality image overlay navigation system consisting of a free-standing frame and an LCD display (white arrow) used for passive projection of the MR images on the semitransparent mirror (gray arrow), which reflects the image back into the operator’s line of sight. This creates apparent hybrid views of the projected MR image (black arrow) and the subject that appear fused. The black line indicates a transpedicular virtual needle path. The white circle symbols the skin entry site formed by the intersection of the laser line (red) and the projected virtual needle path (Color figure online)

Fig. 2
figure 2

Transpedicular MR-guided lumbar vertebroplasty with augmented reality image overlay navigation. A PerkStation interface shows endplate compression deformity of the L3 vertebral body (white arrow) and the planned needle path (line) with target point and skin entry point. B Intraprocedural photograph of the operator view shows the hybrid view of the subject and projected MR image (white arrow), transpedicular virtual needle path (gray arrow) with superimposed vertebroplasty needle, and the operator’s hand (black arrow) holding an MRI-compatible hammer. Axial C and sagittal D T2-weighted turbo spin echo MR image after needle placement shows the tip of the needle (black arrows) at the target. E Sagittal T2-weighted 3D turbo spin echo (SPACE) MR image after injection of 3 ml of polymethylmethacrylate cement shows the hypointense cement deposit centrally (white arrow) (Color figure online)

The navigation system provided MRI guidance through the projection of combined axial MR images and operator-defined virtual needle paths, which created an apparent fusion of the projected axial MR images and the patient under the system [28]. Needle placement consisted of alternating steps of image overlay navigated needle advancement outside the bore and under the navigation system as well as MRI control with the patient inside the bore. The automatic table function was used to translate the subject from the bore of the MRI system under the navigation system in appropriate location to match the projected MR image with the predefined target location and vice versa. The needle was placed by maneuvering the needle along the virtual needle path (Figs. 1, 2B).

Specimen

Five nonembalmed, full spine torso human cadavers (four men and one woman; age range at death range 65–82 years; mean age at death 73 years) were used. Two of five (40 %) cadavers were of small size [16–18.5 kg/m2 living body mass index (BMI),]; 2 of 5 (40 %) were of medium size (18.5–25 kg/m2 BMI), and 1 of 5 (20 %) was of large size (25–30 kg/m2 BMI) [29]. All subjects were obtained and handled in accordance with our institutional rules and in accordance with the Health Insurance Portability and Accountability Act. The frozen cadaveric subjects were allowed to thaw for 24 h at room temperature (~20–22 °C) before MRI.

Research Plan

A total of 25 unilateral vertebroplasties were prospectively planned, including 5 of 25 (20 %) thoracic and 20 of 25 (80 %) lumbar vertebroplasty (Fig. 2). In each subject, 5 vertebroplasties were attempted. The study was performed on 5 different days (one subject per day) over a time period of 6 weeks. Procedures were performed by a single operator with 10 years of experience in percutaneous musculoskeletal procedures and interventional MRI. Initial experience and training was accomplished before the study with image overlay navigated puncture of 5 vertebral bodies of a cadaveric lumbar spine specimen.

Because the presence of vertebral body fractures in the obtained specimens was unknown, each procedure started with MRI of the entire thoracic and lumbar spine using an isotropic, T2-weighted MRI pulse sequence (three-dimensional Sampling Perfection with Application optimized Contrasts using different flip angle Evolutions [SPACE]) with the following parameters: repetition time (TR) 1,400 ms; echo time (TE) 100 ms; flip angle (FA) 120°; averages (Av) 2; echo train length (ETL) 117; voxel size 1 × 1 × 1 mm; field of view (FOV) 192 × 168 mm; base resolution (BR) 192 pixel; phase resolution (PR) 100 %; and bandwidth (BW) 744 Hz. If present, vertebral bodies with compression deformities were selected (Fig. 2). Otherwise, vertebral bodies were selected randomly according to a computer algorithm.

Workflow

All cadavers were placed prone on the table of the MRI system. Parallel MRI was employed using table coil elements and a flexible loop-shaped radiofrequency surface receiver coil (Siemens Healthcare) with a diameter of 19 cm, which was freely movable and centered over the target site.

The initially acquired high-resolution 3D SPACE MRI data set was loaded into 3D Slicer software for planning and navigation. The operator determined the most suitable access to the previously defined vertebral body, prescribed the virtual needle path, and defined the final needle tip position facilitated through simultaneous display of axial, sagittal, and coronal reformats of the isotropic MRI data set by the PerkStation module in 3D Slicer (Fig. 2A). Depending on the individual anatomy, a transpedicular or parapedicular access was chosen.

For each vertebroplasty target, the projected MR image contained the virtual needle path with calculated insertion depth (Fig. 2B). The automatic table of the MRI system was moved under the navigation system in appropriate location. The skin entry point was chosen as indicated by the virtual intersection of the laser of the overlay system and the displayed virtual needle path (Fig. 2B). If necessary, the surface coil was shifted to include the skin entry site centrally. The surface loop coil was fixed with tape and stayed in place for the entire procedure, which facilitated intermittent MRI acquisition for monitoring of the needle placement, which was similar to previously described techniques in patients [16, 20, 30].

An appropriate skin incision was created to allow introduction of the needle. The operator maneuvered the MRI-compatible vertebroplasty needle along the displayed virtual needle path into the defined entry site of the vertebral body (Fig. 2B), and the cortex was penetrated using an MRI-compatible hammer. Using intermittent MRI control of the needle position (axial turbo spin echo MR sequence TR/TE 1800/60, FA 100, Av 1, ETL 14, slice thickness 3 mm, number of slices 7, FOV 349 × 209 mm, BR 320 pixels, PR 58 %, BW 388 Hz, and acquistion time 34 s), the needle was advanced step-wise to the target point.

Once the final needle position was confirmed, 3 ml of polymethylmethacrylate cement (Somatex, Teltow, Germany) was injected while the cadaveric specimen was outside the bore. Finally, 3D SPACE MR images (sequence parameters as previously described) of the treated level were acquired.

Assessment of Outcome Variables

The type of vertebral body access was categorized into transpedicular versus parapedicular.

The number of intermittent MRI control steps required for needle placement was defined as pairs of needle advancement, and MRI control was assessed.

The target error of the needle tip, defined as the Euclidean distance between the planned and final position of the tip of the needle, was assessed. PerkStation was used for calculations by comparing the planned location of the needle tip with the location of the needle tip as manually determined on the final axial turbo spin echo MR images used for needle control [27]. Measurements were performed three times. The median was used.

The intravertebral cement location was assessed by measuring the Euclidean distance between the planned needle tip position and the geometric axial plane center of the cement deposit, which was defined as the intersection of the longest axial and shortest axial diameter. Measurements were carried in similar fashion as for the target error of the needle tip using PerkStation [31]. Measurements were performed three times. The median was used. The presence of extravertebral cement spread was recorded as well.

The length of time for vertebroplasty of a single level—including planning, operator calibration, needle placement and intermittent MRI control, cement injection, and acquisition of the final 3D SPACE MR images of the treated level—was calculated from the time information of the DICOM headers.

Statistical and Quantitative Assessments

Statistical analysis was performed using statistical software (JMP, version 7.01; SAS Institute Cary, NC). Categorical variables were expressed as frequencies and proportions. Quantitative variables were expressed as median with minimum and maximum values. Intrarater variability was expressed by use of the coefficient of variation (CV) as CV = σ/μ, where σ is the first SD, and μ is the arithmetic mean. A p value of <0.05 was considered to indicate a significant difference.

Results

All planned procedures (25 of 25, 100 %) were performed, including 5 of 25 thoracic vertebroplasties and 20 of 25 lumbar vertebroplasties (Fig. 2). Of those, 16 of 25 (64 %) were performed with a transpedicular access (Fig. 2) and 9 of 25 (36 %) with a parapedicular access route. Procedures were feasible in all three cadaver sizes. For the large cadaver, the height of the navigation system was individually adjusted to create a sufficient amount of space for access and contact-free handling. A median average of 6 (range 3–9) MRI control were required to place a needle into the targeted vertebral body. The target error of the final needle tip position was 6.1 ± 1.9 mm (range 0.3–8.7; CV = 11.2 ± 7.9 %; range 6.7–14.8 %). The distance between the planned needle tip position and the center of the cement deposit was 4.3 mm (range 0.8–6.8). The center of all cement deposits was located inside the targeted vertebral body (Fig. 2E). There was no extravertebral cement leakage, including no leakage into the central spinal canal or neural foramina. The median length of time for one vertebroplasty level was 16 min (range 11–21).

Discussion

The results of our study show that MRI-guided vertebroplasty is feasible to allow for sufficiently accurate transpedicular and parapedicular vertebral body access and intravertebral cement delivery. Image overlay navigation was useful for the assessment of anatomy, determination of vertebral body access, and guidance of needle placement. An optimized turbo spin echo pulse sequence is capable of creating a well-defined artifact of an MRI-compatible vertebroplasty needle for accurate localization of the needle tip and to avoid obscuring critical anatomic structures, such as vertebral body cortex, thecal sac, and spinal nerves. Intermediate-weighted MRI creates a high contrast-to-noise ratio between the hypointense cement and the hyperintense vertebral body bone marrow signal, thus making the cement deposit inside the vertebral body well visible.

Interventional MRI has evolved into clinically beneficial and practically useful musculoskeletal procedures and techniques [1619, 3243]. Although placement of spinal needles inside a bore of a magnet is feasible, the spatial limitations and distance to the isocenter, even inside wide-bore MRI systems [44], likely prevents accurate and sterile advancement of a vertebroplasty needle inside the vertebral body because two hands are needed at the skin entry site, one to hold the needle and one for the hammer [45]. Assisted real-time MRI guidance could be helpful for the determination of the skin entry site and needle advancement in the soft tissues dorsal to the vertebral body inside the bore [46]. Vertebral body puncture could occur outside with indirect free-hand navigation by a separate in-room monitor [47]; however, the visual separation of target and image information may complicate this task. Image-overlay navigation facilitated accurate vertebral body puncture outside the bore by providing virtual reality MRI guidance through the simultaneous visualization of the MR images and the vertebral body target.

We used axial images for navigation, although the actual needle paths for transpedicular access was often oblique in craniocaudally direction. This was not found to limit the operator’s ability of successful puncture because once the needle penetrated the cortex at the defined target point in correct alignment with the displayed virtual path, the needle advancement performed in a stepwise, rotating fashion was guided by the pedicle cortex and was safely achieved in all cases as confirmed by axial and sagittal, intermittent MRI control. The addition of nonaxial image navigation capabilities would be expected to improve the versatility of this navigation technology.

Cement visibility is important for the characterization of cement spread and extravertebral leakage [5, 48]. The use of intermediate echo time pulse sequences resulted in high contrast conspicuity between the hypointense injected cement and the hyperintense vertebral body tissues. However, because of flow-related void of signal in vessels with fast spin echo technique, intravascular uptake of hypointense-appearing cement may not have been detectable in our study [48]. More recently described cement with T1-shortening capabilities appears hyperintense on T1-weighted images [49], which can be beneficial for the visualization of cement spread into hypointense tissues and structures, thus potentially rendering intravascular uptake detectable.

A limitation of MRI-guided vertebroplasty may be the difficulty of real-time imaging monitoring of cement injections. For the purpose of this initial investigation, cement injections were not monitored real-time but were performed outside the MRI system. Similar to our current practice of CT-guided vertebroplasty, however, intermittent MRI monitoring after every 0.5 ml of injected cement can be employed as an alternative approach for potentially early detection of extravertebral spread.

Spinal neoplasms may be treated with a combination of percutaneous ablation of the vertebral body lesion and subsequent cement augmentation [68, 10]. MRI-guided ablation of vertebral body lesions has been shown to be feasible [50] and may be advantageous because of high lesion conspicuity and temperature-monitoring capabilities of the ablation zone and surrounding susceptible structures [22, 23]. With the described technique, the combination of MRI-guided vertebral body ablation and cement augmentation of the ablation zone seems possible.

Our study had limitations. First, because we used cadaveric specimens, patient motion and respiration were not present, which may influence the performance of static image-overlay navigation. However, vertebroplasty is often performed with the patient under general anesthesia with controlled respiration and limited motion. Second, because the presence and location of vertebral body compression deformities was unknown, the MRI data set, which was later used for planning, covered the entire thoracolumbar spine. The given vertebroplasty time therefore did not include the acquisition time of a dedicated MRI data set of the respective vertebral body level, which is approximately 3 min. Third, because the presence of vertebral body lesions and compression fractures of cadavers was not known, the sample size only included two vertebral body compression fractures, whereas the remainder of vertebroplasties were performed in normal-appearing vertebral bodies. Fourth, although high-resolution, intermediate-weighted MR images visualized the intravertebral cement well, subtle cement leakage may have gone undetected. This could be assessed with CT correlation in a future study.

Based on the results of this human cadaver study, we conclude that MRI-guided vertebroplasty with image-overlay navigation is feasible with sufficient accuracy for cement augmentation of thoracic and lumbar vertebral bodies. We believe that MRI-guided vertebroplasty could be a beneficial technique in a properly selected group of patients as well as a supplement to CT and fluoroscopy as the principal imaging-guidance techniques, thus warranting further investigation.