Introduction

Magnetic resonance imaging (MRI) is indispensable in the planning of spinal cancer surgery. Tumors, metastases, and especially pathologic fractures of the spine, neoplastic or not, are clearly visible on MRI [1]. Tumor infiltration of the surrounding tissues can be quantified, and therapeutic decisions can be made. In fractures, fracture plane and edema can be analyzed, and the age of the fracture can thus be determined. The same comprehensive information on the pathology cannot be matched by any diagnostic tool using x-ray [2]. MRI-guided musculoskeletal interventions have been proven to be feasible [3]. The detection of soft tissues and other delicate structures, such as nerves, vessels, and spinal cord, improves the safety of these interventions [4]. Fast interventional MRI-sequences generate short repetition rates and allow real-time imaging during the intervention.

To ensure feasible MRI-guided interventions, dedicated MRI-compatible instruments and materials must be developed for such procedures [5]. Currently these instruments are identified indirectly, by the lack of signal because MRI can only visualize protons, i.e., water. Signal-inducing materials are advantageous for such interventions.

Because cementoplasty is considered the standard treatment for most pathological spinal fractures, whether osteoporotic or neoplastic [68], simultaneous visualization of the cement and the pathologic lesion is beneficial during the cement’s injection. In** kyphoplasty, in which a bone cavity lacks signal results, the cement’s signal would permit a controlled defect-filling.

Hydroxyapatite (HA)-based bone substitutes contain water for MRI signal and have an osteoconductive potential [9]. They provide a biologic bone–cement interface [10], and mixing them with conventional polymethylmethacrylate cements (PMMA) increases bioavailability of the PMMA [11].

A contrast agent (CA) further enhances the MRI signal of the PMMA–HA compound.

The aim of this study was to develop and analyze an MRI signal–inducing bone cement based on PMMA, HA, and CA mixture. The signal properties and the compressive strength of this cement were assessed.

Materials and Methods

PMMA (Bon Os; aap Biomaterials, Dieburg, Germany) is made up of a powdery polymer (12 g polymethylmethacrylate) and a liquid monomer (5 ml methylmethacrylate) and cures after mixing. We added 4 ml HA (Ostim 35; aap Biomaterials) and a gadolinium-based CA (0.5 mmol/ml Dotarem [gadoterate meglumine]; Guerbet, Paris, France) to the cement. First the CA was added to the HA; then the two components were mixed with the PMMA in a plastic cup with a spatula.

Signal Evaluation

We evaluated the cement’s signal with increasing CA concentrations (0–60 μl gadoterate meglumine/mm HA) in the cement compound. The added increments of CA were 0.5, 1, 1.5, 2, 4, 10, 15, 20, 30, 40, and 60 μl. To determine the amount of HA necessary for an adequate signal, we decreased the HA in 1-ml increments from 4 to 2 ml using CA concentrations of 1 and 4 μl/ml HA. We tested the cured cements 24 h after mixing, in 10-ml sealed syringes, which were placed into a rack of a double-walled Plexiglas box. The space in between the box’s walls was filled with water and served as a reference signal. We tested five specimens of each cement composition as follows:

  • Two milliliters HA + 5 ml methylmethacrylate + 12 ml polymethylmethacrylate + CA μl (0.5/1/1.5/2/4/10/15/20/30/40/60 μl)

  • Three milliliters HA + 5 ml methylmethacrylate + 12 ml polymethylmethacrylate + CA μl (0.5/1/1.5/2/4/10/15/20/30/40/60 μl)

  • Four milliliters HA + 5 ml methylmethacrylate + 12 ml polymethylmethacrylate + CA μl (0.5/1/1.5/2/4/10/15/20/30/40/60 μl)

Applicability in the Spine

We tested the cement`s applicability in cadaveric human spines by simulating vertebroplasty. This procedure should prove general feasibility, which can hardly be quantified. In addition, it should answer some critical questions such as the following: Can the MRI cement be injected from a conventional 5-ml syringe through the vertebroplastic needle in the vertebral body? Is the cement`s visibility sufficient in the fast T1W Turbo-Spin-Echo (TSE) pulse sequences, and is the image-repetition rate fast enough? Are there any artifacts that hinder the procedure? Finally, it should rule out unforeseen adverse events. The cement was injected transpedicularly with an MRI-compatible vertebroplastic needle (Somatex Medical Technologies, Teltow, Germany) in 20 vertebral bodies from 4 lumbar spines (all female) with surrounding muscle and soft tissue. The spines’ ages were 82, 87, 88, and 88 years. The needle had an outer diameter of 3.0 mm, an inner diameter of 2.45 mm, and a length of 100 mm. A 5-ml syringe was used to inject cement into the vertebral bodies under MRI guidance in an open high-field MRI scanner. The injected cement was made up of 3 ml HA, 4 μl CA, and PMMA. Because the CNRs between cement and bone are reproducibly measurable, they were the only hard data recorded (see later text).

Measurements in the Open MRI

The cement was scanned in an open high-field MRI (1.0-Tesla Panorama; Philips Medical System, Eindhoven, The Netherlands). Because CAs are T1 specific, we used T1-weighted (T1W) fast (turbo) spin echo pulse sequences for the scans. The pulse sequences were an interventional fast T1W TSE (TR 100 ms, TE 5 ms, α 110°, slice thickness 5 mm) and a T1W TSE (TR 400 ms, TE 10 ms, α 90°, slice thickness 4 mm) sequence. The spinal interventions were conducted under image control with fast T1W TSE sequences. The fast sequence had an acquisition time of 2.6 s, and the image orientation was flipped alternately 90º between a sagittal and an axial plane. All images were evaluated without postprocessing or image filters.

The scans of the cement specimens were performed with a solenoid head coil and those of the spinal specimens with an interventional surface coil (multipurpose L). The MRI signal was evaluated by determining the CNR of the cement to air (CNRAir) and to bone (CNRBone). CNR is calculated by measuring the cement’s image signal in the region-of-interest, the surrounding’s image signal (background), and the SD of the background noise as described by Hendrick [12]. We defined equal areas (20 × 20 pixels, 133 mm2) in the cements and their surroundings and measured the signal in eight slices per specimen for five specimens of each cement composition. Due to lower amounts of cement in the vertebral bodies, we measured only three slices for five specimens. The CNRAir was plotted over the CA concentration to determine the maximal MRI signal.

Compressive Strength

Different amounts of HA (2, 3, and 4 ml) were added to the PMMA to evaluate the compressive strength of the cement. Standard PMMA served as the control group. Ten microliters of CA/ml HA were added to the cements for testing because the amounts for optimal signal are smaller (see Results). We tested compressive strength in accordance with the International Organization for Standardization with the method described in ISO 5833:2002 (E) [13]. The cements were filled in different cylindrical polytetrafluoroethylene forms and stored under standardized environmental conditions (23°C [SD 1°C]) for at least 2 h before testing as follows:

  • Two milliliters HA + 5 ml methylmethacrylate + 12 ml polymethylmethacrylate + 10 μl CA

  • Three milliliters HA + 5 ml methylmethacrylate + 12 ml polymethylmethacrylate + 10 μl CA

  • Four milliliters HA + 5 ml methylmethacrylate + 12 ml polymethylmethacrylate + 10 μl CA

The PTFE forms were 12-mm long (SD 0.1 mm) and 6 mm in diameter (SD 0.1 mm). The cured specimens were visually controlled for trapped air or surface defects, polished, and then packed into aluminum bags to avoid dehydration. After being conditioned for 24 h (SD 2 h) at 23°C (SD 1°C), the cements were compressed on an Instron universal testing-machine (Instron no. 3366; Norwood, MA). The specimens were placed upright between the platens and compressed at a rate of 22.70 mm/min until failure occurred. The compressive strength at maximum load or at 2% offset load of the upper yield point, whichever occurred first, were determined. Compressive strength (MPa) and modulus of elasticity were calculated for 6 specimens. The compressive strength was calculated according to formulas provided in the ISO specifications.

Results

The MRI cement`s manufacturing was feasible and similar to that of conventional PMMA. Because PMMA is hydrophobic and the added HA is water-based, the compound requires at least 2 min of stirring until the PMMA–HA–CA mixture is homogenous.

Signal Evaluation

The signal (CNRAir) of the cement depended on the CA concentration, the amount of HA, and the pulse sequence used (Figs. 1, 2, 3; Table 1). The CNRAir curve had the typical sequence-dependent shape of paramagnetic substances showing a signal increase up to a maximum followed by a decrease to signal extinction [14]. Fast T1W TSE sequences had greater peaks at different CA concentrations than standard T1W TSE. The maximal CNRAir in fast T1W TSE was 87.3 (SD 2.9) for cements with 4 μl CA/ml HA (16 μl gadoterate meglumine [4 ml HA]). In T1W TSE, CNRAir was 60.8 (SD 2.4) for cements with 1 μl CA/ml HA (4 μl gadoterate meglumine [4 ml HA]). A decreasing amount of HA (4–2 ml) resulted in relatively linear signal loss (Table 1).

Fig. 1
figure 1

Rack of cement-filled syringes in fast T1W TSE. CA was dosed from 0 to 60 μl/ml HA in 4 ml HA, 5 ml methylmethacrylate, and 12 g polymethylmethacrylate. The pulse-sequence was fast T1W TSE, and a head coil was used. Gadoterate meglumine doses were 0, 0.5, 1, 1.5, 2, 4, 10, 15, 20, 30, 40, and 60 μl

Fig. 2
figure 2

Rack of cement-filled syringes in T1W TSE. CA was dosed from 0 to 60 μl/ml HA in 4 ml HA, 5 ml methylmethacrylate, and 12 g polymethylmethacrylate. The pulse-sequence was T1W TSE, and a head coil was used. Gadoterate meglumine doses were 0, 0.5, 1, 1.5, 2, 4, 10, 15, 20, 30, 40, and 60 μl

Fig. 3
figure 3

CNRAir of cements in T1W TSE and fast T1W TSE. Data are from the cement racks in Figures. 1 and 2. CNRAir represents contrast to the surrounding air. Gadoterate meglumine was dosed in μl per ml HA in 4 ml HA, 5 ml methylmethacrylate, and 12 g polymethylmethacrylate

Table 1 CNRAir and compressive strength based amount of HA added to PMMAa

Applicability in the Spine

The MRI-guided cement injection into the vertebral body through a vertebroplastic needle was feasible (Fig. 4). Injection with a 5-ml syringe through the vertebroplastic provided sufficient pressure. The cement’s signal and contrast to air and to bone were clear. The exact detection of the cement permitted controlled injection and visualization of the procedure under the fast T1W TSE guidance. The CNRBone in fast T1W TSE images was 11.8 (SD 0.1). The cement injection speed had to be adapted to the frame rate of 2.6 s to control the injection. This acquisition time provided sufficient resolution of skeletal structures and soft tissues. The needle’s artifacts were strictly limited to the needle itself and did not disturb the procedure. After injection, the CNRBone in T1W TSE images was 30.7 (SD 7.5).

Fig. 4
figure 4

Transversal vertebral body after cement filling in T1W TSE. The cement consisted of 4 μl gadoterate meglumine, 4 ml HA, 5 ml methylmethacrylate, and 12 g polymethylmethacrylate. The pulse-sequence was T1W TSE, and a body coil was used. C = cement

Compressive Strength

Increasing the amount of HA resulted in lower compressive strength (Table 1). Although the first 2 ml HA reduced the initial compressive strength of conventional PMMA for nearly 50 percent, every further ml HA decreased the strength for only 10 MPa.

Discussion

In this study, we found that it is possible to induce an adequate MR-signal in PMMA bone cements with HA and CA. The main principle to induce an MRI signal in PMMA cement is to mix it with water (protons) and CA.

The signal intensity of the MRI cement can be manipulated by changing the amount of HA, the amount of CA, or the pulse sequence. These physical variables determine the signal. Pulse sequences can be modified by the MRI user to maximize the signal in the region of interest [15]. The MRI signal depends linearly on the amount of HA due to the amount of protons. Every CA has its own signal peak before extinction occurs depending on its paramagnetic properties [16]. Both substances are essential in gaining the maximum signal.

Among the existing CAs, gadolinium induces the highest MRI signal [17]. However, there has been some concern about the use of gadolinium due to cases of nephrogenic systemic fibrosis [18]. Despite the fact that we used gadolinium in this study, generally any CA can be used in the cement to induce a signal; this MRI cement merely demonstrates the principle. Furthermore, toxicological thresholds of gadoterate meglumine were not reached with the miniscule doses used in this study [19]. We used gadolinium because it is readily available in our institution.

In this study, all MRI cements produced a clear signal. The CNRBone was smaller than the CNRAir because the signal of bone is greater than the signal of air but was far greater than 4. According to the Rose model [20], CNR should be >4 to guarantee a sufficient contrast to the surrounding structures for visual differentiation. The high CNRBone permitted direct control of the injection procedure in cadaveric spines with fast T1W TSE, and visualization on T1W TSE images after the procedure. We focused on T1W sequences due to their CA sensitivity. In T2W sequences the MRI cements could be used as well, but the signals would have lower CNR peaks at different CA concentration [14].

Certainly, MRI-guided cementoplasty must be tested in a clinical setting to rule out unpredictable factors, such as patient breathing or other patient-dependent artifacts. Long-term testing is needed to investigate the cement’s signal behavior over time and to determine whether the cement would be detectable in postoperative diagnostics in vivo.

Because PMMA is biologically inert, the cured cement is later surrounded by a biomembrane within the bone, as is typical for foreign bodies [21]. Hydroxyapatite has better biocompatibility [22, 23], which leads to its increased application for cementoplasty [24]. In our study, HA are hydroxyapatite crystals as a suspension in water, which can easily be mixed with PMMA. HA contains enough water to generate an MRI signal and has a better biocompatibility than conventional polymethylmethacrylate cement. Thus, HA seems to be the ideal additive for PMMA.

The increasing amount of HA decreases the compressive strength. The mechanical properties of cements have been discussed in the literature extensively [25], but their outcomes remain inconclusive. The results among the different studies on the fracture risk after vertebroplasty and kyphoplasty remain controversial. Although some studies show a greater risk of adjacent spinal fractures after vertebroplasty and kyphoplasty [26], but others do not [27]. It is unclear how strong a cement must be [28] be safe for clinical application. The compressive strength is the main factor for the design of kyphoplastic and vertebroplastic cements and is quite often the only tested parameter [29]. Calcium phosphate cements, which are currently being used for cementoplasty, have compressive strengths of 70 MPa [30] in vitro; however, in an aqueous environment they only withstand 5–25 MPa [31]. The MRI cement matches these values for all amounts of HA (2–4 ml) added to the PMMA. Although the properties of bone cements used in spinal surgery should be determined, dynamic long-term tests are needed to show the MRI cement’s clinical applicability.

The increasing amount of HA impairs the cement’s compressive strength but increases the cement’s MRI signal. A high MRI signal requires high amounts of HA; a mechanical resistant one requires low amounts. This means that the signal’s advantage of an increasing HA amount is limited by the mechanical disadvantage. Because the cement’s compressive strength is the crucial parameter for cementoplasty, we suggest a PMMA cement (5 ml methylmethacrylate + 12 ml polymethylmethacrylate) containing 2–3 ml HA with 8–12 μl gadolinium for cementoplasty.

Our results support what has been shown earlier for the enhancement of CA in MRI [32]: A CA concentration-dependant signal curve in the MRI-cements. Concerning the biomechanical properties, the results of the mechanical tests of HA–PMMA compounds support the published data [33] and advocate their use due to their better biocompatibility [34].

Along with the increasing availability of MRI, preoperative MRI diagnostics of the spine increase [35, 36] and the number of MRI-guided interventions is growing [37]. MRI-guided surgery is limited by the dictates of MRI sequences, the induced instrument artifacts and MRI safety. The present MRI-cement overcomes these limitations, as it is MR safe, induces its own signal, and is not visualized as an artifact. In the treatment of neoplasms or fractures with MRI-guided cementoplasty, direct visualization of the pathology and the anatomical structures is possible throughout the procedure. This enables control of the cement, visualization of the neoplasm or fracture and spinal structures, such as nerves, vessels, and soft tissues during the cement application. Furthermore, the cement’s high signal increases safety because it prevents cement leakage. Indirect visualization of the cement would only detect the compression of soft tissues and structures at risk approximately the vertebral body.

MRI-guided interventions allow the treatment of pathologic lesions, which are not detectable with conventional imaging under direct visualization. MRI cement is necessary because MRI-guided cementoplasty requires clearly visible cement [38], which can be differentiated from the surrounding structures. This cement would allow precise and controlled injection of cement into pathologic lesions of the bone, thereby increasing the precision of MRI-guided cementoplasty. In bone-defect filling, such as kyphoplasty, the cement’s signal becomes crucial because it permits complete filling.

Although the resolution of real-time imaging is improving rapidly, and frame rates are becoming shorter, image quality must be improved further to warrant the precision of MRI-guided interventions. A fast-moving grid sequence should be developed to generate a three-dimensional picture of the region of interest. Sequence modulation, such as spectral fat saturation inversion recovery sequences, could further improve the cement’s visibility by suppressing the bone marrow’s signal and thereby increasing the contrast.

In the future, this MRI-compatible cement could broaden the field of MRI-guided cementoplasty to the treatment of other osseous neoplasms to provide both ideal intervention control and postinterventional follow-up.

Conclusion

We developed a PMMA-based signal-inducing bone cement with HA and CA the use in MRI interventional procedure. The cement produces a clearly detectable MRI signal and equals the compressive strength of existing cements for spinal cementoplasty.