Introduction

Among the possible approaches to perform total hip arthroplasty (THA), the direct anterior approach (DAA) has the advantage of being a muscle preserving anatomic approach with rapid early recovery and low risk of dislocation [1,2,3,4]. The DAA is typically performed supine which facilitates the use of intra-operative fluoroscopy, allowing for more accurate assessment of acetabular component positioning, and anatomically restore leg length [1,2,3,4].

The reported disadvantages of this procedure are the steep learning curves for surgeons, the risks of injury to the lateral femoral cutaneous nerve, and the risk of femoral fracture due to the difficulties with adequate femoral exposure [4,5,6]. Surgeons transitioning from other approaches can also potentially increase the risk of malpositioning of the acetabular cup, due to familiarity with the orientation of the acetabulum of their patients placed in the lateral position on the operating table [7]. And yet, many factors depend on this acetabular position (orientation and restoration of the centre of rotation (COR)), such as stability, stem-cup and ilio-psoas impingement, function and survival [8,9,10,11,12].

Foissey et al. have already studied the importance of instrumentation in DAA and found that instrumentation without offset was sufficient to achieve good cup positioning [7]. However, there was a learning curve to overcome, and many outliers of cup placement and position were due to an excess of anteversion. In order to make the procedure reproducible and to control as many of the parameters of the cup positioning as possible, enabling technologies and robotic assisted surgery have become a widespread method of improving component accuracy [13,14,15,16,17]. However, to our knowledge, there is very few series to date that focuses solely on the contribution of robotics in DAA without traction table [18].

The aim of this study was to investigate the contribution of image-based robotics assisted THA through DAA in improving radiographic precision primarily, but also the potential differences in functional results and complications.

Methods

Patients

This retrospective case control study compared two cohorts of patients operated upon by two experimented surgeons (SL, PK) in two different centres. 100 consecutive primary THA operated conventionally (cTHA) through DAA from January 2015 to May 2019 were compared to 50 primary THA operated robotically (rTHA) through DAA between January and December 2019. The DAA is performed routinely in those departments unless in patients with BMI ≥ 40 kg/m2, with abnormal hip anatomy requiring complex THA (e.g. congenital hip dysplasia), in elderly patients (over 85 years old) with osteoporosis or when there was previous hip surgery (e.g. femoral or pelvic osteotomy). All patients with a follow-up lower than one year were considered as lost to follow-up.

Pre-operative templating

Both conventional and CT-based templating were made by the surgeon to define the level of the femoral cut, the size and the position of the implants to recreate the native anatomy (offset, length, COR), and if necessary, to correct a leg length discrepancy or a change in the COR. Both groups were templated on standard pelvic anteroposterior radiographs using Traumacad® software (Traumacad®, Petach-Tikva, Israel). The rTHA group was templated as well on a pre-operative pelvic and proximal femur CT scan using the V3 MAKOplasty total hip application software (Mako Surgical Corporation, Kalamazoo, Michigan). In addition to allowing the surgeon to plan in 3D the abovementioned data, this system makes it possible to control the position of the cup in the transverse plane, to adjust its anteversion and to check the absence of any anterior overhang (Fig. 1). The objectives of the cup position were 40° ± 10 of inclination and 20° ± 10 of anteversion, restoration of the COR ± 5 mm in the horizontal plane (HCOR) and ± 3 mm in the vertical plane (VCOR). In case of modification of the anatomical landmarks due to osteoarthritis, the references of the healthy contralateral hip were used. When using 3D planning, particular attention was paid to the anterior overhang; if this was present with a restoration of the COR and anteversion of 20°, the anteversion was increased and the COR was medialized (Fig. 2).

Fig. 1
figure 1

Contribution of 3D template: planning of the neck resection site, femoral size and version, leg length and global offset (A), cup size and positioning, COR restitution, anterior overhang (B)

Fig. 2
figure 2

Anterior overhang observed during 3D planning with COR restoration and 20° anteversion (A). Application of a 24° anteversion and a 3 mm medialization of the COR to eliminate this overhang (B)

Operative technique

All THAs were performed through DAA in supine position without any traction table as already describe by Foissey et al. [7]. In both groups, stability and leg length were assessed manually.

Cups in cTHA patients were put with standard straight instrumentation and controlled by per-operative fluoroscopic assessment. Implant position was guided by both pre-operative templating and intra-operative anatomical landmarks (anterior and posterior walls, transverse acetabular ligament). The size of the reamers was progressively increased to remove first the cartilage until the desired size is achieved with satisfactory circumferential contact and exposure of bleeding subchondral bone. In case of anterior overhang, the cup was medialised and anteriorised. All shells were cementless cups (Quattro (Lepine®, Genay, France)), and all stems were cementless as well (Targos (Lepine®, Genay, France)).

The rTHA patients were operated using the RIO robotic arm (Mako Surgical Corporation) using the “express” procedure. This express procedure helps to ream the acetabulum, to impact the final cup, and to help to restore the ideal leg length and global offset (Fig. 3) but does not guide femoral neck resection and does not give stem anteversion and COR. One array was placed on the contralateral iliac crest. One checkpoint was placed on the proximal femur, one on the acetabular side and one distal landmark was placed at the level of the ipsilateral patella as part of the express procedure. Femur was prepared manually after a registration of the femoral checkpoint and of the distal landmark. Then, the position of the pelvis was confirmed by registering and verifying the position of patient-specific anatomical landmarks. The robotic arm with haptic control was used to guide the reaming and the impaction of the cup according to the template. After the reduction of the hip with the broach or with the final stem in place, global offset restitution and leg length restitution were assessed. All implants used in the rTHA group were cementless cups (Trident, Stryker®) and an uncemented Anato stem (Stryker®).

Fig. 3
figure 3

Per-operative contribution of robotic “express” procedure: reaming assistance (A), cup impaction assistance (B) and control of the leg length and of the global offset (C)

Radiological assessment

All measurements were recorded on the operated side by two independent surgeons on the standing antero-posterior X-rays at two months after surgery using the picture archive and communication (PACS) software. Magnification was corrected using the implanted acetabular component size as the reference value. Radiographic analysis included inclination and anteversion of the cup, VCOR, HCOR, acetabular offset (AO), femoral offset (FO) and leg length (LLA). Combined offset (CO) was defined as addition of AO and FO [19]. The anteversion was determined by the method described by Widmer [20]. Cup abduction was the angle between the cup axis and parallel between the inter-teardrop line. All the other assessments were performed according to the method described by Kayani et al. [21] (Fig. 4). Cup orientation was classified whether inside or outside the safe zone (SZ) concerning the inclination (30–50°), the anteversion (10–30°) or both position (combined SZ) [7]. Outliers were defined as medial displacement of the COR (ΔHCOR) > 5 mm, vertical displacement of the COR > 3 mm superiorly (ΔVCOR), as these parameters have been associated with greater wear and instability [22]. A difference of LLA > 10 mm compared to the contralateral side was considered as an outlier [19].

Fig. 4
figure 4

Radiographic assessments: HCOR, horizontal centre of rotation; VCOR, vertical centre of rotation; LLA, radiographic leg-length assessment; ML, midline; ITL, inter-teardrop line; VTL, vertical trans-teardrop line; AO, acetabular offset; FO, femoral offset; FA, femoral anatomical axis

Clinical assessment

During the pre-operative consultation, the demographic data, pre-operative Harris hip score (HHS) were obtained. Operative time, peri and early post-operative complications and length of hospitalisation were documented during the stay. Routine follow-ups were planned at two months, one year and every two to five years. This study focused on the results that were collected at one year (HHS, complications).

Statistical analysis

Mean value, standard deviation and minimum and maximum values are presented for continuous variables that were normally distributed. Their distribution was evaluated using a Shapiro–Wilk test. The following statistical tests were used for continuous variables: Student’s t test when the variable of interest was independent and normally distributed; paired Student’s t test when the variable of interest was paired, independent and normally distributed; the Mann–Whitney U test for independent variables that were not normally distributed; and the Wilcoxon signed rank test for paired variables that were not normally distributed. Qualitative variables were summarized as percentages. A chi-square test or Fisher’s exact test was used to ascertain any difference among these. The significance threshold was set at 5%. The statistical analysis was performed using XLSTAT™ software (version 2021, AddInsoft, Paris, France).

Ethical approval

All procedures were performed in accordance with the ethical standards of the institutional and/or national research committee, the 1964 Helsinki Declaration and its later amendments, or comparable ethical standards. The Advisory Committee on Research Information Processing in the Field of Health (CCTIRS) approved this study in Paris on February 17, 2016, under number 16–140. As per institutional standards, formal patient consent was not required for this type of study.

Results

All patients in both groups reached the one year follow-up time, and there were no lost to follow-up. Pre-operative demographic and clinical data are summarized in Table 1. The two groups were comparable, 66% (66/100) were women in the cTHA group versus 46% (23/50) in the rTHA group but this difference was not statistically significant (p = 0.22).

Table 1 Demographic and clinical data

Radiographical results

There was no significant difference in the mean orientation of the cups in the coronal and sagittal planes. The mean anteversion of the cups was greater than 23° in both groups (Table 2). The overall orientation of the cups is shown in Fig. 5. The robotic cups were better oriented with 98% (49/50) of the cups being in the global SZ in the rTHA group versus 68% (68/100) in the cTHA group (p = 0.0002) (Fig. 6). The main reason for poor positioning was excess of anteversion (21% (21/100) in the cTHA group versus 2% (1/50) in the rTHA group, p = 0.005) (Fig. 6).

Table 2 Radiological results
Fig. 5
figure 5

Diagram depicting the number of cups inside the safe zone in the conventional group (A) and in the robotic group (B). Sixty-eight percent (68/100) of the conventional cups are in the global safe zone versus 98% (49/50) of the robotic cups (p < 0.0001)

Fig. 6
figure 6

Rates of outliers for the cup positioning for both groups

The COR was on average better restored in the robotic group in both the horizontal and vertical planes, although the cups were on average medialized by more than 3 mm in both groups (Table 2). There were fewer outliers in the rTHA group concerning VCOR (10% (5/50) versus 28% (28/100), p = 0.03) (Fig. 7).

Fig. 7
figure 7

Rates of outliers for the restoration of the centre of the rotation and the leg length for both groups. HCOR horizontal centre of rotation; VCOR, vertical centre of rotation; LLA, radiographic leg-length assessment

There was no significant difference in terms of restitution of the length of the lower limbs nor in terms of the restitution of the different offsets (Table 3, Fig. 7).

Table 3 Per- and post-operative data and complications

Clinical results

The robotic procedure was significantly longer by 19 min (p < 0.0001). No complication relative to the robot was reported. The length of hospital stay was similar as well as the HHS at 1 year (Table 3). There was no significant difference in terms of complications between the two groups. No iliopsoas impingement or dislocation was reported (Table 3).

Discussion

Our results favour the use of image based robotic system when performing a THA by DAA. The control of the cup placement by the robotic arm allowed to obtain a placement in the SZ in 98% of the cases as well as a better control of the COR restitution.

To our knowledge, there is only one study that focused solely on the use of the robot via DAA and found excellent results with 100% of the cups in the Lewinnek SZ [18]. These results echo those found in the main series using the same robotic assistance, as they found an accuracy in terms of orientation greater than 97% [21, 23, 24]. This is an advantage that would be all the more interesting at the beginning of the DAA experience because it has been shown that changes in spatial references during the transition to the DAA in supine position could lead to an increased risk of malposition [7], whereas robotics does not present a learning curve in terms of positioning the cup [25].

Robotic restoration of the COR has been less studied than orientation. Kayani et al. also found a better restoration of the COR in the robotic group [21]. This is an important advantage because until now, traditional navigation systems have not proven their effectiveness [14, 26], wherein this restoration theoretically allows a better longevity and a better function [27,28,29]. Although this restoration of the COR has been improved, it remains medialized by 3 mm on average in the robotic group. Also the average anteversion of 23° seems important because it is the upper limit in several safe zone descriptions [30, 31]. This can be explained by the deliberate choice of few towards anterior overhangs even if it means medializing and anteverting the cup (Fig. 2). This may also explain the high outlier rates in the conventional group except that in the absence of robotic control the cup placement tended to be exaggerated with greater anteversion and medialization outside the thresholds. Also, the higher rate of women may be a secondary explanation for this high rate of outliers. Indeed, women tend to have a more marked hypoplasia of the anterior wall [32], making anteversion and greater medialization mandatory in order to avoid an outlier. This strategy seems to be successful because there was no iliopsoas-impingement diagnosed at 1 year. Finally, it should be noted that the pre-operative COR measurements were performed on hips with degenerative changes, which may artificially increase the outlier rate, especially when the surgeon wanted to restore the native COR [33].

Even if this better positioning did not result in a better clinical outcome, it is reasonable to believe that it will have a long-term benefit in terms of wear [22]. There remains an absence of high-quality level I evidence to answer this question whether or not the robot allows an improvement of functional scores after THA but a recent meta-analysis showed a significant improvement of HHS after the use of the robot assisted surgery [24].

There are several limitations to this study. First, it is a retrospective, non-randomized, unmatched study. Even if the difference in the proportion of women was not significant, this may have hindered the comparison of the groups. Secondly, the cup used in each group was not the same, which may have changed the way the implant was positioned. Finally, the measurements of the cup positioning were on radiographs and not CT scans. But a CT scan with irradiations is not recommended post-operatively.

Conclusion

The use of image-based robotic system when performing a THA by DAA provided an advantage in controlling the orientation of the cup and the restoration of its centre of rotation. Thanks to the 3D planning on pre-operative scanner, it allowed to respect the thresholds while avoiding the anterior overhangs.