Abstract
Purpose
The purpose of this study was to determine the inter- and intraobserver variability of intraarticular landmark identification for tunnel position calculation in image-free anterior cruciate ligament (ACL) navigation.
Methods
In a test/retest scenario, thirteen experienced ACL surgeons (>50 reconstructions year) experienced in image-free ACL navigation were asked to identify the landmarks required for image-free ACL navigation in the same cadaver knee. Landmark positions were registered using a fluoroscopic ACL navigation system. Positions were determined using validated radiological measurement methods. For variability analysis, mean positions, deviations between the test/retest positions, standard deviations (SD) and range were calculated.
Results
Interobserver analysis showed a mean variability (SD) for the tibial landmark positions of 3.0 mm with deviations of up to 24.3 mm (range). Mean femoral landmark variability was 2.9 mm (SD) with deviations of up to 11.3 mm (range). Intraobserver analysis showed a tibial reproducibility of 2.2 mm (SD 2.0 mm; range 10.9 mm) and a femoral of 1.9 mm (SD 1.9 mm; range 10.4 mm).
Conclusion
The data of the presented study suggest that a considerable inter- and intraobserver variability in intraarticular landmark identification exists. Reasonable ranges were found that have to be considered as a potential risk for miscalculation of tunnel positions in image-free ACL reconstruction.
Clinical relevance
Landmark acquisition affects tunnel calculation in image-free ACL.
Level of evidence
IV.
Similar content being viewed by others
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.Avoid common mistakes on your manuscript.
Introduction
Despite increasing knowledge about knee anatomy and biomechanics as well as technical improvements in ACL reconstruction, tunnel misplacement is still one of the most common reasons for graft failure after anterior cruciate ligament (ACL) reconstruction [9, 14, 15, 19, 42, 45]. Radiological studies indicate that approximately 10–40 % of drill holes in primary ACL reconstructions have been incorrectly placed and that tunnel misplacement highly correlates with poor results after ACL reconstruction [6, 10, 12, 14, 17, 19, 28, 36].
Although the importance of correct tunnel positioning is well accepted and there is general agreement that tunnel positions within the anatomical insertion areas of the ACL are fundamental to successful ACL reconstruction and long-term stability [2, 19, 24, 27, 47], the implementation of this knowledge into intraoperative tunnel positioning seems to remain difficult since reliable landmarks, allowing correct tunnel identification, do not consistently exist [8, 13, 20, 29, 31, 32, 38, 40, 44].
To improve reproducibility and accuracy of anatomical tunnel positioning, navigation technology was introduced to ACL reconstruction. Based on pre-/intraoperative image acquisition (image-based systems) or on intraarticular landmark acquisition (image-free systems), these systems enable the surgeon to plan, identify and drill optimal tunnel positions, to check isometry, to perform a virtual impingement analyse and to measure anterior tibia translation and rotational stability [18, 20, 21, 23, 25, 29, 39, 43, 46]. Depending on the surgeon’s preferred reconstruction technique, navigation systems allow for transtibial and anteromedial portal techniques as well as for single- and double-bundle reconstruction.
Studies on image-free ACL navigation indicate the feasibility of image-free ACL navigation. Other studies have shown that image-free ACL navigation enables for reliable translation and rotation measurements as well as for isometry analysis [3, 5, 16, 21, 22, 25, 26, 29, 30, 33]. Thus, the reliability (accuracy and reproducibility) of tunnel planning has never been validated.
For tunnel position planning, the software of image-free navigation systems [11, 16, 21, 25, 33, 43] creates a 3D model of the knee joint. Therefore, the system requires the acquisition of intraarticular landmarks and surfaces. Based on kinematic analysis and the identification of special landmarks and areas, such as the medial tubercle spine, the anterior margin of the posterior cruciate ligament (PCL), the anterior horn of the lateral meniscus and the o’clock positions at the femur, a software algorithm calculates and suggests an optimized anatomical and isometric tunnel positions.
Since the quality of the calculated 3D model and the suggested tunnel positions strongly rely on the reproducibility and accuracy of intraarticular landmark identification, there obviously seems to be a reasonable potential for miscalculation. Consequently, one wrong landmark, as well as many not very precise acquired landmarks, could create wrong 3D models suggesting wrong tunnel positions.
Therefore, the aim of this study was to determine the inter- and intraobserver variability of intraarticular landmark identification in navigated ACL reconstruction. The hypothesis of this study was that intraarticular landmark identification has a reasonable variability, which has to be considered in image-free ACL navigation.
Materials and methods
This study was performed in one fresh female cadaver knee, in which all ligaments were intact before intervention. The study was performed, imitating an operating room (OR) situation. After positioning of the cadaver and draping the leg in an OR like way, the endoscope (Smith & Nephew, Andover, MA) was introduced over a standard anterolateral portal. Then the ACL was cut in its midsubstance over a deep anteromedial portal with a punch, and the fibres were then removed with a shaver. At the tibia, the stump of the ACL was resected in a way that the tibial insertion area of the ACL could be clearly identified. At the femur, the ACL fibres were removed to a degree that the insertion area at the lateral notch wall could be clearly identified. Further soft tissue resection was performed, so that the surgeons could clearly identify the anterior horns of both menisci, the transverse ligament, the medial tubercle spine and the PCL as well as the over the top position and the inner notch wall geometry.
Navigation system
For later analysis, a fluoroscopic-based ACL navigation system (Brainlab, Feldkirchen, Germany) was used [39]. Therefore, one bicortical Schanz Screw (5 mm) was fixed at the distal lateral femur and the distal medial third of the tibia. Further, a metal scaling ball was fixed to the knee, to allow for later measurement of the positions. Next, dynamic reference bases (DRB’s) containing light emitting diodes (LEDs) were rigidly fixed at the Schanz screws. Fluoroscopic images in ap and lateral projections were then obtained using a standard fluoroscope (Ziehm, Nürnberg, Exposcop 7000) which was connected to a registration kit, containing LED’s. During image acquisition, an optoelectronic Polaris® camera (NDI Polaris, Waterloo, Canada) overlooking the surgical field, digitized the reflecting infrared flashes from the DRB’s and the registration kit, creating an individual infrared reflection image. The software of the navigation system used the fluoro image and the LED informations to calculate the three-dimensional position of the knee. Using navigated instruments, intra- and extraarticular landmarks, positions and surface could precisely been tracked and displayed on the fluoroscopic images throughout the operation.
Surgical procedure
Thirteen experienced ACL surgeons (>50 reconstructions year) trained in image-free ACL reconstruction were asked to identify the specific intraarticular landmarks required for image-free ALC navigation in the same knee under the same OR like conditions. According to the workflow of image-free ACL navigation [21, 25], one surgeon after the other had to identify the above-mentioned landmarks arthroscopically using a navigated palpation hook. The display of the navigation system was blinded for the surgeons. As in image-free navigation, they had to position the tip of the navigation hook on the identified landmarks. Of each identified landmark (landmark 1: anterior horn of the lateral meniscus; landmark 2: medial tubercle spine; landmark 3: anterior border of the PCL; landmark 4: 12 o’clock position; landmark 5: 10:30 o’clock position), screenshots of the identified position were taken. Using a fluoroscopic-based navigation system, the positions were indicated in the referenced fluoroscopic anterior–posterior (ap) and lateral projections. A retest for intraobserver analysis was performed, approximately four ours later.
Measurement of the landmark positions
All screenshot images of all surgeons, identifying the above-mentioned landmark positions, were transferred into a professional CAD drawing and measurement software (Canvas 9.0, ACD Systems). To determine the position and to measure deviations and variability of the different landmark positions, reliable radiological measurement methods were used. At the femur, Bernard and Hertel’s quadrant method and Staubli’s method for the tibial measurement were used (Fig. 1).
For femoral analysis, a quadrant was projected on the overlapping condyles of the femur according to the method of Bernard and Hertel [4]. The quadrant was positioned so that the proximal line was overlapping Blumensaat’s line, the projection of the intercondylar roof in lateral X-rays. The total sagittal depth and total height of the intercondylar surface determined the size of the quadrant.
According to the femur, tibial measurement followed in relation to a line described by Staubli [41]. In the lateral radiograph, a line was drawn parallel to the sagittal tibial joint line. The length of the line is limited ventrally by the tibial cortex and dorsally by the descending eminentia intercondylaris.
Then, the different positions of each landmark were measured along the femoral quadrant and the tibial line using a calibrated measurement tool. With the metal scaling ball, all acquired positions of the different landmarks could be determined in millimetres and were collected in an excel data sheet for further analysis. Data were collected in an excel (Microsoft Excel 14.2.5, USA) data sheet. The Study was approved by the local ethics committee of the university of cologne.
Statistical analysis
For both, tibial and femoral analysis, mean positions of each landmark were calculated in two planes to describe their three-dimensional position. The variability of each landmark was then analysed by calculating the standard deviations (SD) of the mean positions in two planes. Further the range between the different landmark positions was calculated, indicating outliers. The variability of the tibial and femoral landmark identification was calculated by calculating the mean SD for the three tibial and the two femoral landmarks. For determination of the reproducibility (intraobserver analysis) of the landmark identification, the differences between the positions of the test and retest were calculated for each landmark of each surgeon in two planes. Further SD and range were calculated to determine outliers.
Results
Interobserver results (Table 1)
The interobserver analysis showed a mean variability of 3.0 mm for the tibial and 2.9 mm for the femoral landmarks. Different variability results were found in the statistical analysis between the five landmarks (Figs. 2, 3). Landmark 1, the anterior horn of the lateral meniscus, showed the lowest variability results with 1.9 mm and also the lowest range of 6.6 mm, indicating the highest accuracy of the five landmarks. In contrast, landmark 2, the medial tubercle spine, showed the highest variability of the required landmarks for image-free ACL navigation. Deviations of 4.8 mm and a range of up to 20.1 mm were found for landmark 2. Landmark 3 (anterior border of the PCL), landmark 4 (12 o’clock position) and landmark 5 (10:30 o’clock position) showed variability results between 2.2 and 3.0 mm with ranges of 10.2–11.3 mm.
Intraobserver results (Table 2)
The intraobserver analysis showed mean deviations of 2.2 mm and 1.9 mm between the test and retest’s positions of the tibia and the femur, with ranges of up to 10.9 mm. Deviations between the positions were found in the ap as well as in the lateral plane (Fig. 4).
Mean deviations of 1.8 mm (SD 1.7; range 6.3 mm) for landmark 1, 2.9 mm (SD 2.1; range 7.3) for landmark 2 and 1.9 mm (SD 2.3; range 10.9) for landmark 3 were found for the tibia. At the femur, mean deviations of 1.7 mm (SD2.2; range 10.4) were found for landmark 4 and 2.2 mm (SD 1.9; range 7.1) for landmark 5.
Discussion
The most important finding of the presented study is that a considerable inter- and intraobserver variability of intraarticular landmark identification exists. Therefore, the hypothesis of the presented study can be accepted. The variability of the intraarticular landmarks positions required in navigated ACL reconstruction to calculate the tunnel positions was found to be 1.9–4.8 mm. Intraobserver analysis showed a mean reproducibility of 1.8–2.9 mm. Next to the variability and reproducibility, ranges between 6.6 and 20.0 mm were found, indicating that outliers exist. The presented results therefore indicate that a considerable variability in landmark identification exists that may influence the calculation of tunnel positioning in image-free ACL navigation.
Thus, these results have to be critically discussed, since a reasonable number of studies seem to confirm the reliability of image-free ACL navigation systems [6, 7, 11, 21, 29].
Picard demonstrated in a study on “Navigated Anterior Cruciate Ligament Reconstruction: Radiographic Validation of a Nonimage-based System” that navigation resulted in more accurate tunnel placements compared to traditional arthroscopic techniques. Thus, in this study, tunnel planning was performed according to the surgeons desired and indicated location by preoperatively inserting radiopaque fiducials [29]. Since tunnel positions were planned by the surgeon and not by the navigation system, the validation of Picard’s study is limited to the accuracy of the navigated drilling process.
A more recent validation study on image-free ACL navigation compared the navigated measurement and determination of tunnel positions with radiological measurement methods of tunnel positions. In this study, the tibial and femoral tunnel positions were “placed into the joint according to the surgeon’s judgment of the optimal location” [16]. Then, navigated and radiological measurements of these positions followed and both methods were compared. The validation of this study is therefore limited to a comparison between the two measurement methods. Tunnel planning did not follow the navigation protocol.
In a cadaver study, Plaweski et al. [33] compared the variability of conventional and image-free navigated tunnel positions. The results of their qualitative analysis showed more anatomical tunnel positions in the navigated group; thus, a quantitative analysis of the tunnel positions was not performed.
Angelini et al. analysed the differences between conventional and navigated tunnel positions using isometry measurement methods [3]. Tunnel positions in the navigation group were calculated by the navigation system according to the above-mentioned landmark acquisition. The navigation group showed significant better results concerning isometry. Tunnel positions, accuracy and reproducibility were not addressed in this study.
In a prospective, randomized, double-blind study, Hart et al. [11] compared stability results, IKDC and Lysholm score as well as radiological tunnel positions between navigated and conventional ACL reconstruction. There were no significant differences between both groups except for the variability of femoral tunnel positioning, which was smaller in the navigation group.
Schep et al. analysed the intraobserver variability between conventional and navigated tunnel positioning in ACL reconstruction. Their results showed that navigation can reduce the intraobserver variability compared to conventional ACL reconstruction. Thus, the intraobserver variability in the navigation group still showed reasonable deviations between the planned tunnel positions of 5 mm (SD 2.4; range 1.8–9.6 mm) at the tibia and 4.6 mm (SD 2.1; range 2.1–8.4 mm) at the femur [37]. Schep explained the variability of tunnel positions in the image-free ACL navigation group with several existing, different but appropriate tunnel positions.
Although different aspects, such as translation and rotation measurements as well as benefits related to isometry analysis, have been validated and analysed, neither the above-mentioned studies [3, 5, 16, 21, 22, 25, 26, 29, 30, 33, 34] nor—to our best knowledge—other studies have so far analysed the variability of tunnel positions in image-free ACL navigation. Therefore, the results of the presented study have to be considered, discussing the accuracy and reliability of tunnel positioning in image-free ACL navigation. Corresponding results could be found in navigated total knee arthroplasty. However, navigation technology is accepted as a precise method to improve accuracy in navigated total knee arthroplasty, different studies report on limitations related to the variability of landmark identification. Determining a reasonable landmark identification variability in navigated total knee arthroplasty, Robinson et al. [35] concluded that the variability in landmark identification represents a source of method error. Beyond Amanatullah et al. [11] recently showed that deviations in landmark identification of more than 2 mm influence the precision and outcome in navigated total knee arthroplasty.
One of the main shortcomings of this study is that it was only possible to analyse the variability in one knee. Further, it was only possible to allow thirteen surgeons to identify the necessary landmarks. The number of surgeons was limited by the condition of the cadaver knee, which had to be the same for all surgeons during the test and the retest. Therefore, also the statistical impact of the subgroup analysis is limited.
Finally, the results of the presented study addressed the variability of landmark identification in image-based ACL navigation. The variability of the calculated tunnel positions by image-free navigation systems was not analysed. Nevertheless, it is obvious that the variability of tunnel positions in image-free ACL navigation is influenced by the variability of the intraarticular landmark identification, since the tunnel positions are calculated based on an algorithm.
Conclusion
The results of the presented study indicate reasonable deviations for landmark identification in the inter- and intraobserver analysis that have to be considered as a certain potential for tunnel position miscalculation in image-free ACL navigation. Surgeon’s applying image-free ACL reconstruction system’s should therefore be aware that the reliability of tunnel positioning depends on the quality of information the systems receives.
References
Amanatullah DF, Di Cesare PE, Meere PA, Pereira GC (2013) Identification of the landmark registration safe zones during total knee arthroplasty using an imageless navigation system. J Arthroplast 28(6):938–942
Amis AA, Beynnon B, Blankevoort L, Chambat P, Christel P, Durselen L, Friederich N, Grood E, Hertel P, Jakob R et al (1994) Proceedings of the ESSKA scientific workshop on reconstruction of the anterior and posterior cruciate ligaments. Knee Surg Sports Traumatol Arthrosc 2(3):124–132
Angelini FJ, Albuquerque RF, Sasaki SU, Camanho GL, Hernandez AJ (2010) Comparative study on anterior cruciate ligament reconstruction: determination of isometric points with and without navigation. Clinics 65(7):683–688
Bernard M, Hertel P (1996) Intraoperative and postoperative insertion control of anterior cruciate ligament-plasty. A radiologic measuring method (quadrant method). Unfallchirurg 99(5):332–340
Colombet P, Robinson J, Christel P, Franceschi JP, Djian P (2007) Using navigation to measure rotation kinematics during ACL reconstruction. Clin Orthop Relat Res 454:59–65
Cross MB, Musahl V, Bedi A, O’Loughlin P, Hammoud S, Suero E, Pearle AD (2012) Anteromedial versus central single-bundle graft position: which anatomic graft position to choose? Knee Surg Sports Traumatol Arthrosc 20(7):1276–1281
Degenhart M (2004) Computer-navigated ACL reconstruction with the OrthoPilot. Surg Technol Int 12:245–251
Ferretti M, Doca D, Ingham SM, Cohen M, Fu FH (2012) Bony and soft tissue landmarks of the ACL tibial insertion site: an anatomical study. Knee Surg Sports Traumatol Arthrosc 20(1):62–68
Getelman MH, Friedman MJ (1999) Revision anterior cruciate ligament reconstruction surgery. J Am Acad Orthop Surg 7(3):189–198
Harner CD, Giffin JR, Dunteman RC, Annunziata CC, Friedman MJ (2001) Evaluation and treatment of recurrent instability after anterior cruciate ligament reconstruction. Instr Course Lect 50:463–474
Hart R, Krejzla J, Svab P, Kocis J, Stipcak V (2008) Outcomes after conventional versus computer-navigated anterior cruciate ligament reconstruction. Arthroscopy 24(5):569–578
Hefzy MS, Grood ES (1986) Sensitivity of insertion locations on length patterns of anterior cruciate ligament fibers. J Biomech Eng 108(1):73–82
Howell SM (1998) Principles for placing the tibial tunnel and avoiding roof impingement during reconstruction of a torn anterior cruciate ligament. Knee Surg Sports Traumatol Arthrosc 6(Suppl 1):S49–S55
Howell SM, Hull M, McAllister D (2012) Be sensible and cautious about criticizing tunnel placement in ACL reconstruction: commentary on an article by Duncan E. Meuffels, MD, PhD, et al.: computer-assisted surgery is not more accurate or precise than conventional arthroscopic acl reconstruction. a prospective randomized clinical trial. J Bone Joint Surg Am 94(17):133
Jaureguito JW, Paulos LE (1996) Why grafts fail. Clin Orthop Relat Res 325:25–41
Jenny JY, Abane L (2012) Navigated anterior cruciate ligament reconstruction: radiographic validation of a nonimage-based system. Orthopedics 35(10):18–21
Kamath GV, Redfern JC, Greis PE, Burks RT (2011) Revision anterior cruciate ligament reconstruction. Am J Sports Med 39(1):199–217
Kawakami Y, Hiranaka T, Matsumoto T, Hida Y, Fukui T, Uemoto H, Doita M, Tsuji M, Kurosaka M, Kuroda R (2012) The accuracy of bone tunnel position using fluoroscopic-based navigation system in anterior cruciate ligament reconstruction. Knee Surg Sports Traumatol Arthrosc 20(8):1503–1510
Khalfayan EE, Sharkey PF, Alexander AH, Bruckner JD, Bynum EB (1996) The relationship between tunnel placement and clinical results after anterior cruciate ligament reconstruction. Am J Sports Med 24(3):335–341
Klos TV, Habets RJ, Banks AZ, Banks SA, Devilee RJ, Cook FF (1998) Computer assistance in arthroscopic anterior cruciate ligament reconstruction. Clin Orthop Relat Res 354:65–69
Koh J, Koo SS, Leonard J, Kodali P (2006) Anterior cruciate ligament (ACL) tunnel placement: a radiographic comparison between navigated versus manual ACL reconstruction. Orthopedics 29(10 Suppl):S122–S124
Martelli S, Zaffagnini S, Bignozzi S, Lopomo N, Marcacci M (2007) Description and validation of a navigation system for intra-operative evaluation of knee laxity. Comput Aided Surg 12(3):181–188
Matsuo T, Mae T, Shino K, Kita K, Tachibana Y, Sugamoto K, Yoshikawa H, Nakata K (2013) Tibiofemoral relationship following anatomic triple-bundle anterior cruciate ligament reconstruction. Knee Surg Sports Traumatol Arthrosc. doi:10.1007/s00167-013-2646-1
McConkey MO, Amendola A, Ramme AJ, Dunn WR, Flanigan DC, Britton CL, Group MK, Wolf BR, Spindler KP, Carey JL, Cox CL, Kaeding CC, Wright RW, Matava MJ, Brophy RH, Smith MV, McCarty EC, Vida AF, Wolcott M, Marx RG, Parker RD, Andrish JF, Jones MH (2012) Arthroscopic agreement among surgeons on anterior cruciate ligament tunnel placement. Am J Sports Med 40(12):2737–2746
Muller-Alsbach UW, Staubli AE (2004) Computer aided ACL reconstruction. Injury 35(Suppl 1):S-A65–S-A67
Musahl V, Bedi A, Citak M, O’Loughlin P, Choi D, Pearle AD (2011) Effect of single-bundle and double-bundle anterior cruciate ligament reconstructions on pivot-shift kinematics in anterior cruciate ligament- and meniscus-deficient knees. Am J Sports Med 39(2):289–295
Musahl V, Burkart A, Debski RE, Van Scyoc A, Fu FH, Woo SL (2003) Anterior cruciate ligament tunnel placement: comparison of insertion site anatomy with the guidelines of a computer-assisted surgical system. Arthroscopy 19(2):154–160
Oettl GM, Imhoff AB (1998) Revision surgery in failed anterior cruciate ligament-plasty. Zentralbl Chir 123(9):1033–1039
Picard F, DiGioia AM, Moody J, Martinek V, Fu FH, Rytel M, Nikou C, LaBarca RS, Jaramaz B (2001) Accuracy in tunnel placement for ACL reconstruction. Comparison of traditional arthroscopic and computer-assisted navigation techniques. Comput Aided Surg 6(5):279–289
Piefer JW, Pflugner TR, Hwang MD, Lubowitz JH (2012) Anterior cruciate ligament femoral footprint anatomy: systematic review of the 21st century literature. Arthroscopy 28(6):872–881
Piefer JW, Pflugner TR, Hwang MD, Lubowitz JH (2012) Anterior cruciate ligament femoral footprint anatomy: systematic review of the 21st century literature. Arthroscopy 28(6):872–881
Plaweski S, Cazal J, Rosell P, Merloz P (2006) Anterior cruciate ligament reconstruction using navigation: a comparative study on 60 patients. Am J Sports Med 34(4):542–552
Plaweski S, Rossi J, Merloz P, Julliard R (2011) Analysis of anatomic positioning in computer-assisted and conventional anterior cruciate ligament reconstruction. Orthop Traumatol Surg Res 97(6 Suppl):S80–S85
Ramme AJ, Wolf BR, Warme BA, Shivanna KH, Willey MC, Britton CL, Magnotta VA, Grosland NM (2012) Surgically oriented measurements for three-dimensional characterization of tunnel placement in anterior cruciate ligament reconstruction. Comput Aided Surg 17(5):221–231
Robinson M, Eckhoff DG, Reinig KD, Bagur MM, Bach JM (2006) Variability of landmark identification in total knee arthroplasty. Clin Orthop Relat Res 442:57–62
Sati M, Staubli H, Bourquin Y, Kunz M, Nolte LP (2002) Real-time computerized in situ guidance system for ACL graft placement. Comput Aided Surg 7(1):25–40
Schep NW, Stavenuiter MH, Diekerhof CH, Martens EP, van Haeff CM, Broeders IA, Saris DB (2005) Intersurgeon variance in computer-assisted planning of anterior cruciate ligament reconstruction. Arthroscopy 21(8):942–947
Shafizadeh S, Balke M, Wegener S, Tjardes T, Bouillon B, Hoeher J, Baethis H (2011) Precision of tunnel positioning in navigated anterior cruciate ligament reconstruction. Arthroscopy 27(9):1268–1274
Shafizadeh S, Huber HJ, Grote S, Hoeher J, Paffrath T, Tiling T, Bouillon B (2005) Principles of fluoroscopic-based navigation in anterior cruciate ligament reconstruction. Oper Tech Orthop 15:70–75
Siebold R, Ellert T, Metz S, Metz J (2008) Femoral insertions of the anteromedial and posterolateral bundles of the anterior cruciate ligament: morphometry and arthroscopic orientation models for double-bundle bone tunnel placement: a cadaver study. Arthroscopy 24(5):585–592
Staubli HU, Rauschning W (1994) Tibial attachment area of the anterior cruciate ligament in the extended knee position. Anatomy and cryosections in vitro complemented by magnetic resonance arthrography in vivo. Knee Surg Sports Traumatol Arthrosc 2(3):138–146
Trojani C, Sbihi A, Djian P, Potel JF, Hulet C, Jouve F, Bussiere C, Ehkirch FP, Burdin G, Dubrana F, Beaufils P, Franceschi JP, Chassaing V, Colombet P, Neyret P (2011) Causes for failure of ACL reconstruction and influence of meniscectomies after revision. Knee Surg Sports Traumatol Arthrosc 19(2):196–201
Tsuda E, Ishibashi Y, Fukuda A, Tsukada H, Toh S (2007) Validation of computer-assisted double-bundle anterior cruciate ligament reconstruction. Orthopedics 30(10 Suppl):S136–S140
Tsuda E, Ishibashi Y, Fukuda A, Yamamoto Y, Tsukada H, Ono S (2010) Tunnel position and relationship to postoperative knee laxity after double-bundle anterior cruciate ligament reconstruction with a transtibial technique. Am J Sports Med 38(4):698–706
Wright RW, Huston LJ, Spindler KP, Dunn WR, Haas AK, Allen CR, Cooper DE, DeBerardino TM, Lantz BB, Mann BJ, Stuart MJ (2010) Descriptive epidemiology of the Multicenter ACL revision study (MARS) cohort. Am J Sports Med 38(10):1979–1986
Zaffagnini S, Klos TV, Bignozzi S (2010) Computer-assisted anterior cruciate ligament reconstruction: an evidence-based approach of the first 15 years. Arthroscopy 26(4):546–554
Zantop T, Wellmann M, Fu FH, Petersen W (2008) Tunnel positioning of anteromedial and posterolateral bundles in anatomic anterior cruciate ligament reconstruction: anatomic and radiographic findings. Am J Sports Med 36(1):65–72
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Shafizadeh, S., Balke, M., Hagn, U. et al. Variability of landmark acquisition affects tunnel calculation in image-free ACL navigation. Knee Surg Sports Traumatol Arthrosc 23, 1917–1924 (2015). https://doi.org/10.1007/s00167-014-2963-z
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00167-014-2963-z