1 Introduction

Potable water is vital to life, and that it is clean and safe to use is of utmost importance. However, millions of people still suffer from the threats of unhealthy drinking water around the world, especially those living in third-world countries and remote locations. Copper is a naturally occurring metal—often classified as a heavy metal—found in rocks, soil, plant life, a number of organisms, and water [1]. Many of these organisms’ diets—including humans—require copper in trace amounts; however, excessive or certain forms of copper have also been shown to cause gastrointestinal distress, liver disease, and kidney damage [2,3,4]. The highest level of copper in drinking water limited by the United States Environmental Protection Agency (U.S. EPA) is 1.3 mg/L (1.3 ppm) [5,6,7]. Therefore, it is crucial to be able to identify and detect the amount of copper in drinking water, if it is above the U.S. EPA limit.

Many aqueous elemental detection techniques have been developed previously. Conventional methods, including atomic absorption spectroscopy (AAS), atomic emission spectroscopy (AES), inductively coupled plasma mass spectrometry (ICP-MS), and inductively coupled plasma optical emission spectrometry (ICP-OES), have performances with excellent sensitivity, accuracy, and selectivity [8,9,10,11,12]. However, they require sophisticated instruments that are costly, time consuming, require expertise, and lack portability. Many new tools have occurred since the start of the twenty-first century, such as colorimetric sensors [13,14,15,16], “turn-on” or “turn-off” fluorescent probes [17,18,19,20], biosensors [21,22,23], square wave anodic stripping voltammetry [24,25,26], differential pulse voltammetry [27,28,29], and potentiometric measurements based on ion-selective electrodes [30,31,32]. They are not only more economic, portable, sensitive, and selective but also able to rapidly detect the concentrations of copper on site. Nevertheless, the processes to fabricate them are usually tedious, and particular chromogenic and fluorescent receptors, biological recognition elements, and electrode materials are often required to prepare.

Here, we propose a simple dip-coating procedure to fabricate a fabric amperometric sensor for the selective detection of Cu(II) ions in drinking water for the first time. The prepared Cu(II) sensor consists three layers. The bottom layer is nylon-6 nanofibers that are used as scaffolds to support the other two layers. The middle layer is composed of multi-walled carbon nanotubes (MWCNTs) which are utilized to yield a conductive sensing platform due to their superb electronic properties. The top layer is 2,2′:5′,2″-terthiophene molecules that have a high adsorption and selectivity towards Cu(II) ions [33]. Figure 1 illustrates the schematic diagram (macroscopic top view and nanoscale side view) of the Cu(II) sensor (2,2′:5′,2″-terthiophene/MWCNTs/nylon-6 nanocomposites) and the sensor response to the Cu(II) ions. When a voltage is applied to the Cu(II) sensor, a baseline current will be detected ascribed to the movement of charge carriers (electrons) in MWCNTs. As a tiny amount of water containing Cu(II) ions being added, the Cu(II) ions will be captured by the 2,2′:5′,2″-terthiophene molecules. Once that occurs, an impeded current can be observed, and the current impedance correlates with the concentration of Cu(II) ions in water. It is believed that the impedance results from the attraction of positively charged Cu(II) ions to the negative electrons in MWCNTs [34]. In this work, the sensitivity and selectivity of the Cu(II) sensor are carefully studied.

Fig. 1
figure 1

Mechanism of detecting Cu(II) ions by 2,2′:5′,2′′-terthiophene/MWCNTs/nylon-6 nanocomposites

2 Material and methods

Materials

Nylon-6 pellets (polycaprolactam, 3 mm), acetic acid (ACS, ≥ 99.7%), formic acid (ACS, ≥ 98%), Triton X-100 (4-(1,1,3,3-tetramethylbutyl)phenyl-polyethylene glycol, 625 g/mol), and silver nitrate (AgNO3, ACS, ≥ 99%) were purchased from Sigma-Aldrich. Multi-walled carbon nanotubes (MWCNTs, outer diameter 10–20 nm, inside diameter 3–5 nm, length 10–30 μm, specific surface area 233 m2/g, > 95%) were supplied by Cheap Tubes, Inc. 2,2′:5′,2″-terthiophene (≥ 98%) was obtained from Ark Pharm, Inc. Tetrahydrofuran (THF, certified) was purchased from Fisher Chemical. Copper chloride (CuCl2, 99%) and ferrous chloride (FeCl2, 97%) were purchased from Acros Organics. Cadmium chloride (CdCl2, 99.9%), lead nitrate (Pb(NO3)2, reagent), and mercury chloride (HgCl2, ACS) were purchased from Alfa Aesar.

Sensor preparation

The nylon-6 nanofibers were fabricated via electrospinning. First, nylon-6 pellets were dissolved in 1: 1 (w/w) acetic acid:formic acid solution at 20 wt%. Then, the nylon-6 solution was transferred into a 5-mL syringe and loaded into the syringe pump. The nylon-6 nanofiber mat was electrospun at a solution feed rate of 9.1 μL/min for 4 h. A bias of 30 kV was applied between the syringe needle and rotating drum collector, between which was a distance of 8 cm. After that, the electrospun nylon-6 nanofiber mat was dried under vacuum at room temperature for 24 h.

A dip-coating procedure was employed to load MWCNTs and 2,2′:5′,2′′-terthiophene molecules onto the nylon-6 nanofibers [34, 35]. The fabricated nylon-6 nanofiber mat was cut into rectangular pieces with sizes of 1 × 2 cm and dipped into a bath in which MWCNTs were dispersed in a 0.3 wt% Triton X-100 aqueous solution. Triton X-100 was used as a surfactant to stabilize the MWCNTs in water. The concentration of MWCNTs aqueous solution was optimized at 0.25 g/L for the uniform dispersion on the surface of nylon-6, according to our previous work [36]. After 30 s, the nylon-6 nanofiber pieces were removed and placed in a water bath for 1 min to remove the loosely bound MWCNTs. The MWCNTs/nylon-6 nanocomposites were dried under ambient conditions, and then immersed in a 2.5 g/L 2,2′:5′,2′′-terthiophene solution using THF as the solvent for 24 h to ensure the fully load of molecules. The Cu(II) sensors, MWCNTs/nylon-6 nanocomposites loaded with 2,2′:5′,2′′-terthiophene, were prepared after they were dried under ambient conditions.

Sensor evaluation

All the electrochemical experiments were carried out utilizing the Solartron Analytical CellTest System which consisted of a 1470E multi-channel potentiostat unit and multiple 1455A series frequency response analyzers. The Cu(II) sensor was fixed on a glass slide with two nickel-plated steel alligator clips (each at one short side of the 1 × 2 cm sensor) connected to channel cables of the 1470E unit. To measure the resistivity, 30 μL of deionized water was dropped onto the Cu(II) sensor, and then a voltage ranging from − 3 to 3 V in 2 min was applied to it. The changes in current with voltage were collected.

To characterize the responses of Cu(II) sensors to Cu(II) ions, CuCl2 aqueous solutions in which the concentrations of Cu(II) ions were 0, 0.13, 0.65, 1.3, 6.5, 13, 39, and 65 ppm, respectively, were prepared. A constant voltage of 1 V was applied to the Cu(II) sensors, and 30 μL of those prepared CuCl2 aqueous solutions were dropped after 30 s. The changes in currents within 5 min were recorded.

Cd(II), Fe(II), Pb(II), Hg(II), and Ag(I) ions at their MCLs (maximum contaminant levels in drinking water enforced by the U.S. EPA) were employed to investigate their interferences on the response of the Cu(II) sensor to Cu(II) ions. More specifically, 1.3 ppm Cu(II) aqueous solutions containing 0.005 ppm Cd(II), 0.3 ppm Fe(II), 0.015 ppm Pb(II), 0.002 ppm Hg(II), and 0.1 ppm Ag(I), respectively, were prepared. A constant voltage of 1 V was applied to the Cu(II) sensors, and 30 μL of the prepared above solutions was dropped after 30 s. The currents were compared with those corresponding to 1.3 ppm pure Cu(II) aqueous solutions. The MWCNTs/nylon-6 nanocomposites were used for controlled experiments, and the procedures to investigate their resistance and response to Cu(II) ions were the same.

Material characterization

The prepared nylon-6 nanofibers, MWCNTs/nylon-6 nanocomposites, and Cu(II) sensors (2,2′:5′,2′′-terthiophene/MWCNTs/nylon-6 nanocomposites) were imaged by a JEOL 6500F scanning electron microscope (SEM). A thin layer of gold was sputter coated on their surfaces and the accelerating voltage was 15 kV. Fourier-transform infrared spectroscopy (FT-IR) was performed on a Bruker ALPHA FT-IR spectrometer. A resolution of 2 cm−1 in the range of 4000–500 cm−1 with 30 scans was applied. Thermogravimetric analysis (TGA) curves were acquired at a heating rate of 10 °C/min from room temperature to 600 °C under a N2 atmosphere through a TA Instruments Q500 thermogravimetric analyzer.

3 Results and discussion

The SEM images of the prepared nylon-6 nanofibers, MWCNTs/nylon-6 nanocomposites, and Cu(II) sensors are exhibited in Fig. 2. The nylon-6 nanofiber has a uniform diameter of about 250 nm and smooth surface, as shown in Fig. 2a. This demonstrates the electrospinning process is steady and consistent. The electrospun nylon-6 mat with non-woven fibrous morphology is free of defects and can perform as an ideal supportive scaffold. It can be seen from Fig. 2b that the MWCNTs are successfully loaded and well dispersed onto the surface of nylon-6 nanofibers. There is no obvious difference between the Fig. 2b, and c and d, which means that the MWCNTs are still firmly anchored to the surface of nylon-6 nanofibers after the introduction of 2,2′:5′,2′′-terthiophene through dip-coating process. The copper binding agent 2,2′:5′,2′′-terthiophene molecules are in molecular level thus cannot be observed in SEM images.

Fig. 2
figure 2

SEM images of a nylon-6 nanofibers, b MWCNTs/nylon-6 nanocomposites, and c, d Cu(II) sensors

Figure 3 shows the FT-IR spectra of nylon-6 nanofibers, MWCNTs/nylon-6 nanocomposites, Cu(II) sensors, and 2,2′:5′,2′′-terthiophene, respectively. The nylon-6 nanofibers and MWCNTs/nylon-6 nanocomposites exhibit the same spectra, which suggests the MWCNTs are bonded onto the surface of nylon-6 electrospun fiber through physical interactions (Van der Waals force) rather than covalent or ionic bonds. (The overlapped spectra are shown in Fig. S1, as well as the peak positions given in Table S1. The differences on two spectra are negligible as the IR absorbances are well matched and the peak shifts are within 2 cm−1.) Compared with the samples without functional molecules, there are three additional peaks within a range of 875 to 625 cm−1 for the Cu(II) sensors. These are attributed to the carbon-hydrogen out-of-plane vibrations in thiophene rings, which are also observed in the FT-IR spectrum of 2,2′:5′,2′′-terthiophene. Thus, the 2,2′:5′,2′′-terthiophene molecules are successfully loaded onto the MWCNTs/nylon-6 nanocomposites. By further coupling with the previous SEM images, the successful fabrication of Cu(II) sensors containing MWCNTs and 2,2′:5′,2′′-terthiophene are confirmed.

Fig. 3
figure 3

FT-IR spectra of nylon-6 nanofibers, MWCNTs/nylon-6 nanocomposites, Cu(II) sensors, and 2,2′:5′,2′′-terthiophene molecules

TGA is employed to investigate the amount of MWCNTs and 2,2′:5′,2′′-terthiophene molecules adsorbed onto the nylon-6 nanofibers, and their TGA curves under nitrogen atmosphere are exhibited in Fig. 4. It shows that the pure electrospun nylon-6 nanofibers are thermally stable until 330 °C. Subsequently, from 330 through 460 °C is the degradation stage of the polymer, leaving the carbonaceous residue about 0.7 wt%. The observed thermogram of pure nylon-6 under nitrogen atmosphere in present study is consistent with the literature [37]. Since MWCNTs are kept intact after the pyrolysis process (as the MWCNTs are thermally stable up to ~ 1300 °C under the inert environment) [38], the MWCNTs/nylon-6 nanocomposites and Cu(II) sensors have a slightly greater weight percent on residue (see TGA curves of inset in Fig. 4). The difference on residue weight percentages is about 0.2 wt% which is ascribed to the intact MWCNTs. Compared with nylon-6 nanofibers, a lower onset temperature of the degradation is observed for MWCNTs/nylon-6 nanocomposites and Cu(II) sensors (280 °C as opposed to 330 °C for pure nylon-6), which can be partially attributed to the high thermal conductivity of MWCNTs (2 × 102 W/m K) [39]. The faster and easier local heat transfer readily degrades the nylon-6 through the breakages of amide bond in nylon-6 backbone chains leading to a lower onset decomposition temperature [40]. Other than the thermal conductivity, a trace amount of water in the MWCNTs will also aid the hydrolysis and degradation of the nylon-6 [41]. The defects and acid groups such as –COOH on the MWCNTs caused by the acid treatment during the synthesis process as well as the residual surfactant Triton X-100 might also impair the thermal stability of nylon-6 which induces the degradation. There is a small weight loss observed at 180 °C for the TGA curve of Cu(II) sensors, which is relating to the evaporation of the 2,2′:5′,2′′-terthiophene molecules. The 2,2′:5′,2′′-terthiophene has a boiling point of 160 °C, and it increases to some extent as a result of the interactions with the nylon-6 and MWCNTs. These TGA results further confirm that the MWCNTs and 2,2′:5′,2′′-terthiophene are successfully loaded onto the nylon-6 nanofibers scaffold and the relative weight percentages are determined to be 0.2 wt% and 2 wt%, respectively.

Fig. 4
figure 4

TGA curves of nylon-6 nanofibers, MWCNTs/nylon-6 nanocomposites, and Cu(II) sensors (heating rate 10 °C/min; atmosphere N2)

Figure 5a shows the current-voltage characteristics of the MWCNTs/nylon-6 nanocomposites and Cu(II) sensors in contact of deionized water. The MWCNTs exhibit typically semiconducting behaviors, as observed from the nonlinear data points of the MWCNTs/nylon-6 nanocomposites, while the Cu(II) sensors had more linear data points and metallic characteristics due to the addition of 2,2′:5′,2′′-terthiophene. In order to easily compare their resistances, an approximately linear relationship between the current and voltage is assumed for both the MWCNTs/nylon-6 nanocomposites (R2 = 0.97) and Cu(II) sensors (R2 = 0.99), which means that their resistances are considered stable and constant within a range of − 3 to 3 V as pure resistors. The slope of the current-voltage line for the Cu(II) sensors is greater than that for the MWCNTs/nylon-6 nanocomposites, which clearly indicates that the Cu(II) sensors have a lower resistance and a higher electrical conductivity than the MWCNTs/nylon-6 nanocomposites, which can be ascribed to the conductive nature of π stacking molecular structure from the solid state 2,2′:5′,2′′-terthiophene formed on MWCNTs [42,43,44].

Fig. 5
figure 5

a Current-voltage characteristics of MWCNTs/nylon-6 nanocomposites and Cu(II) sensors wetted by 30 μL of deionized water (measuring time 2 min). b Responses of MWCNTs/nylon-6 nanocomposites and Cu(II) sensors wetted by 30 μL of Cu(II) aqueous solution at a concentration of 1.3 ppm. c Responses of MWCNTs/nylon-6 nanocomposites and Cu(II) sensors wetted by 30 μL of Cu(II) aqueous solution at different concentrations. d Responses of Cu(II) sensors wetted by 30 μL of various metal ion aqueous solution (Cu(II) 1.3 ppm; Cu(II)-Cd(II) 1.3–0.005 ppm; Cu(II)-Fe(II) 1.3–0.3 ppm; Cu(II)-Pb(II) 1.3–0.015 ppm; Cu(II)-Hg(II) 1.3–0.002 ppm; Cu(II)-Ag(I) 1.3–0.1 ppm). For Fig. 5b–d, the applied voltage is 1 V.

The same amount of 30 μL Cu(II) aqueous solution at its MCL (1.3 ppm) was dropped onto the MWCNTs/nylon-6 nanocomposites and Cu(II) sensor after 30 s, and the monitored currents gradually became stable after 2 min. The final current is determined by averaging the current readings from 2 to 5 min. As shown in Fig. 5b, the Cu(II) sensor has a significant response to Cu(II) ions. The current in the MWCNTs/nylon-6 nanocomposites is about 1.27 times greater than that of the Cu(II) sensor. Note that the Cu(II) has a higher electrical conductivity in deionized water without Cu(II) ions than the MWCNTs/nylon-6 nanocomposites. In other words, the monitored stable current greatly decreases after replacing the MWCNTs/nylon-6 nanocomposites with Cu(II) sensor. This phenomenon is due to the binding of Cu(II) ions to the 2,2′:5′,2′′-terthiophene molecules on the surface of the Cu(II) sensor which greatly hinders the mobility of carrier charges.

The effects of the Cu(II) concentrations on the responses of the MWCNTs/nylon-6 nanocomposites and Cu(II) sensors were further investigated. The currents both increase with the increase of Cu(II) concentration due to the ionic nature of Cu(II) ions in aqueous solution, as shown in Fig. 5c. Moreover, there are two intersections between the current-concentration curves of the Cu(II) sensor and MWCNTs/nylon-6 nanocomposites. Since very few Cu(II) ions are present at the ultralow Cu(II) concentrations (< 0.2 mg/L), the current flow in the Cu(II) sensor is still greater than that in the MWCNTs/nylon-6 nanocomposites attributed to the lower inherent resistance of the Cu(II) sensor. More Cu(II) ions forming complexes with 2,2′:5′,2′′-terthiophene molecules as the Cu(II) concentration increases, and the current impedance caused by those complexes gradually becomes the main influencer. As a result, the current in the Cu(II) sensor gets smaller than that of the MWCNTs/nylon-6 nanocomposites. However, at high Cu(II) concentrations (> 40 mg/L), the slopes of the current-concentration curves for the Cu(II) sensor becomes higher than the MWCNTs/nylon-6 nanocomposites again. This is because the maximum adsorption capacity for Cu(II) ions of the Cu(II) sensor has been reached; thereafter the current is predominantly originated not from the movement of electrons in MWCNTs but from the conductivity of Cu(II) aqueous solution. Compared with the MWCNTs/nylon-6 nanocomposites, the Cu(II) sensor has strong hydrophobicity with the account of the hydrophobic 2,2′:5′,2′′-terthiophene molecules. Therefore, less Cu(II) aqueous solution is stuck in the Cu(II) sensor, and the current is larger than that of the MWCNTs/nylon-6 nanocomposites. Based on the above analysis, the Cu(II) sensor exhibits an obvious response to Cu(II) ions at Cu(II) concentrations between those two intersections that are considered to be the working range of the Cu(II) sensor, namely from 0.65 to 39 ppm. The 0.65 ppm is 50% lower than the MCL of 1.3 ppm enforced by the U.S. EPA for Cu(II) ions in drinking water, which means that the Cu(II) sensor in our work is suitable and sensitive enough for detecting Cu(II) ions in drinking water. Meanwhile, the upper limit of 39 ppm is ~ 30 times of the EPA MCL demonstrating the sensor is applicable for a wide range of the Cu(II) contaminated water.

When a voltage is being applied to the Cu(II) sensor, an electric current is generated as electrons flow through the MWCNTs. The Cu(II) ions that are captured by 2,2′:5′,2′′-terthiophene molecules can draw away some electrons from the MWCNTs due to their positive charges; therefore, a responsive change (a decrease) in current will be observed. Since the 2,2′:5′,2′′-terthiophene molecules have a much high selectivity towards Cu(II) ions over the other metal ions, the response of the Cu(II) sensor should remain the same with the existence of interfering metal ions. In other words, there should be no obvious change for the current when other metal ions are introduced into the aqueous solution. In order to be more practical, Cd(II), Fe(II), Pb(II), Hg(II), and Ag(I) at their MCLs set by the U.S. EPA were employed into the Cu(II) aqueous solutions to investigate their interferences on the response of the Cu(II) sensor to Cu(II) ions. It can be seen from Fig. 5d that the currents hardly change (less than 5%) when different interfering metal ion aqueous solutions, instead of pure Cu(II) aqueous solution, are used. Pb(II) and Fe(II) show slightly greater effects with current differences of 4.8% and 4.7%, respectively. Compared with the sensor sensitivity of Cu(II), the interferences of these metal ions are negligible. Thus, the Cu(II) sensor is able to selectively detect Cu(II) ions among various other interfering metal ions in water.

4 Conclusion

Detection of Cu(II) ions in drinking water is usually a bigger challenge than in other contaminated water as a result of its much lower MCLs. A sensitive, cost-effective, flexible, and portable amperometric sensor is successfully prepared via a simple dip-coating method in this work, which is composed of electrospun nylon-6 nanofibers, MWCNTs, and 2,2′:5′,2′′-terthiophene molecules. Owing to the selective binding of 2,2′:5′,2′′-terthiophene molecules with Cu(II) ions, the fabricated sensor shows a significant response to Cu(II) ions in a wide range of 0.65 to 39 ppm, which indicates that its detection limit is way lower than the MCL (1.3 ppm) for Cu(II) ions. Moreover, the Cu(II) sensor exhibits a superior selectivity towards Cu(II) ions over other metal ions, and the presence of Cd(II), Fe(II), Pb(II), Hg(II), and Ag(I) ions at their MCLs has no obvious interfering effects on its response to Cu(II) ions. Therefore, the Cu(II) sensor in this work shows promise in monitoring Cu(II) ions in drinking water in real time.