Introduction

Reduced radiotracer uptake and retention in the left ventricular apex frequently occur in myocardial perfusion imaging (MPI) using single photon emission computed tomography (SPECT) and positron emission tomography (PET). This phenomenon commonly referred to as apical thinning and usually considered to be a normal variant1,2 has repeatedly raised interest among investigators as it poses a diagnostic challenge in nuclear cardiology.3,4,5 However, as of today the cause for this phenomenon remains to be elucidated, and the scarce data available are solely based on SPECT studies with no data on its prevalence and etiology in PET-MPI. A variety of potential causes have been proposed, including partial volume effects,6 physiological inhomogeneities in myocardial radiotracer distribution,7 artifacts induced by attenuation correction (AC), 5,8,9,10,11,12 or time-of-flight (TOF) PET data acquisition.13 Alternatively, it may be hypothesized that apical thinning as observed in PET-MPI is attributable to true anatomical thinning, which has been documented in some studies based on either autopsy or cardiac computed tomography (CT).14,15,16,17 Against this background, we have used hybrid imaging combining contrast-enhanced cardiac CT and PET-MPI with 13N-ammonia to assess the relation between apical thinning and true anatomical thinning of the left ventricular myocardial apex.

Methods

Study Population

We retrospectively screened all patients who had undergone both 13N-ammonia PET-MPI and contrast-enhanced cardiac CT at our institution between January 2005 and December 2015 (n = 187, Figure 1). Patients in whom the scans were acquired with an interval of > 120 days were excluded, as were patients with suspected or known history of apical myocardial infarction or with obstructive coronary artery disease in an epicardial coronary artery subtending the apex or the peri-apical region (as assessed by fused hybrid PET/coronary artery CT angiography), patients with a history of coronary artery bypass grafting (CABG) of the vessel subtending the left ventricular apex or those with a history of urgent revascularization of said vessel. Further exclusion criteria were dilated and hypertrophic cardiomyopathies.

Figure 1
figure 1

Flowchart illustrating the number of screened, excluded, and included patients. PCI, percutaneous coronary intervention; CABG, coronary artery bypass grafting

PET Image Acquisition

Patients underwent a standard 13N-ammonia gated PET-MPI protocol at rest and during adenosine-induced stress (140 μg/min/kg over 7 minutes). Acquisition was performed on either a Discovery RX or Discovery STE PET/CT scanner (both GE Healthcare, Waukesha, WI, USA). 13N-ammonia was injected 3 minutes into pharmacologic stress. Rest images were acquired 50 minutes after completion of the stress scan. Mean administered activities were 713 ± 312 MBq for stress and 804 ± 412 MBq for the rest scans. Dynamic 3D (n = 42) or 2D (n = 15) PET acquisition was started in parallel with the injection and consisted of 9 × 10-s, 6 × 15-s, 3 × 20-s, 2 × 30-s, and 1 × 120-s frames, followed by a static acquisition over 12 minutes. PET data acquired in 3D mode were reconstructed with ordered subset expectation maximization (OSEM; 3 iterations and 16 subsets) while those obtained in 2D were reconstructed using a filtered back projection (FBP) algorithm with an 8-mm Hanning filter and standard decay scatter sensitivity corrections (voxel size 2.34, 2.34, 3.27). Attenuation maps were derived from low-dose non-contrast enhanced CT scans, and correct alignment with PET datasets was visually verified.

CT Image Acquisition

Contrast-enhanced cardiac CT scans were acquired on a 64-slice (LightSpeed VCT or Discovery CT750 HD) or a 256-slice (Revolution CT, all GE Healthcare, Waukesha, WI, USA) CT scanner. Image acquisition was performed with prospective electrocardiography (ECG) triggering at 75% of the R–R interval. Slice thickness was set to 0.625 mm. Up to 30 mg of metoprolol (Beloc Zok, AstraZeneca, Zug, Switzerland) was administered intravenously to achieve a target heart rate of ≤ 65 bpm. All patients received 2.5 mg isosorbiddinitrate (Isoket, CPS Cito Pharma, Uster, Switzerland) sublingually prior to the scan. Contrast agent (Visipaque 320, GE Healthcare, Opfikon, Switzerland) volume (30 to 105 mL) and flow rate (3.5 to 5 mL/min) were adapted to body mass index.

PET Image Analysis

From the rest scans, normalized semi-quantitative myocardial PET tracer uptake summed across the entire cardiac cycle, end-diastolic (ED) and end-systolic (ES) uptake, quantitative myocardial blood flow (MBF), and the k2 rate constant (i.e., tracer washout), based on a two-compartment model, were computed using QPET (Version 2015, Cedars-Sinai Medical Center, Los Angeles, CA, USA) and PMOD (Version 3.7, PMOD Technologies, Zurich, Switzerland) and visualized on a 17-segment polar map. Segment 17 was defined as the apex.1 Apical thinning was defined as a semi-quantitative uptake value of < 70% in segment 17.18 Additionally, a segmental rest score (RS) for the apex, as automatically calculated from the QPET software solution19 and based on a normal database (Figure 2), was recorded according to the 5-point scale ASNC semiquantitative scoring system (0 = no defect, 1 = mildly reduced count density, 2 = moderately reduced, 3 = severely reduced, 4 = absent activity).1 Patients with normal or only mildly reduced apical counts (RS < 2) were compared to individuals with moderate or severe count reduction or absence of counts (RS ≥ 2). Left ventricular ejection fraction (LVEF) was obtained from gated PET, and heart rate (HR) during PET acquisition was recorded. Additionally, we calculated the base-to-apex gradient for summed tracer uptake, ED uptake, ES uptake, MBF and k2 expressed as the ratio between the respective values at the apex and the base as previously described.20,21,22,23

Figure 2
figure 2

Reference polar map of the normal database used by QPET to calculate segmental rest perfusion scores

CT Image Analysis

Cardiac CT datasets were analyzed on a dedicated workstation (AW 4.7, GE Healthcare). Images were reconstructed as vertical and horizontal long axis views. Window width and level were adjusted for optimal delineation of the myocardial wall. Anterior, inferior, and lateral left ventricular wall thicknesses were defined as the mean of three consecutive measurements at 30 mm intervals, starting at the base (Figure 3). The ventricular septum was excluded from measurement due to poor delineation from the right ventricular cavity. Apical thin point thickness was defined as the mean of the visually identified minimal myocardial wall thickness in both vertical and horizontal reconstructions. Apical wall thickness was averaged from the thin point thickness and adjacent measurements taken at a distance of 10 mm from the thin point in all four directions (Figure 3). Additionally, the measurements of the thin point and the apex were adjusted for interindividual variation in global myocardial wall thickness by expressing them as a fraction of the averaged anterior, inferior, and lateral wall thickness (referred to as “indexed” values). Additionally, we calculated the base-to-apex gradient expressed as the ratio between apical wall thickness and the mean of measurements A2, I2 and L2 (Figure 3).

Figure 3
figure 3

Schematic of left ventricular myocardial wall thickness measurement sites on vertical (A) and horizontal (B) long axis CT reconstructions. TP, thin point; AP1-4, apex measurements; A1-3, anterior wall; I1-3, inferior wall; L1-3, lateral wall

Statistics

All statistical analyses were performed using IBM SPSS Statistics 25 (IBM Corporation, Armonk, NY, USA). Continuous variables are expressed as mean ± standard deviation (SD) and paired or unpaired Student’s t-test was used for comparison. Categorical variables are listed as frequencies (%) and were compared with the Pearson χ2 test. Pearson correlation coefficients were used to measure the relation between myocardial wall thickness and PET-derived values. A P-value of < .05 was considered significant.

Ethics

The study was approved by the local ethics committee (BASEC-Nr. 2016-00177) and conforms to the Declaration of Helsinki.

Results

Fifty-seven patients were included in the final analysis (Figure 1). Patient baseline characteristics are given in Table 1.

Table 1 Patient characteristics

PET-MPI

Mean semi-quantitative normalized apical tracer uptake was 58.6 ± 7.1% and 53 patients (93%) had apical thinning (i.e., an apical uptake of < 70%). Mean rest MBF at the apex was 0.68 ± 0.28 mL/g/min and mean k2 rate constant was 0.29 ± 0.18/min. When confining analysis of normalized apical tracer uptake to ED, all 57 patients (100%) exhibited apical thinning compared to 43 patients (73.7%) at ES. Furthermore, apical tracer uptake increased significantly from ED to ES (43.0 ± 5.9% vs 62.9 ± 9.2%, P < .001). When compared to the normal database, perfusion at the apex was deemed normal (RS = 0) in 31.6% of cases (n = 18), whereas 22.8% (n = 13) were classified as having a mild (RS = 1), and 45.6% (n = 26) a moderate uptake reduction (RS = 2). Mean semi-quantitative normalized apical tracer uptake for patients with an apical RS < 2 (n = 31) and those with an apical RS ≥ 2 was 62.5 ± 6.3% and 53.9 ± 4.8%, respectively (P < .0001). In patients with an apical RS ≥ 2, lower apical MBF was measured than in patients with RS < 2 (0.60 ± 0.20 mL/g/min vs 0.75 ± 0.32 mL/g/min; P = .044). Apical k2 constants were 0.30 ± 0.22/min in patients with RS ≥ 2 and 0.29 ± 0.15/min in patients with RS < 2 (P = .892), respectively.

CT Myocardial Thickness Measurements

Myocardial wall thickness measurements are given in Table 2. We observed a minimal but statistically significant difference in apical wall thickness between patients with RS < 2 and RS ≥ 2 (4.7 ± 0.7 mm vs 4.3 ± 0.7 mm; P = .043), while all other parameters did not differ between the groups.

Table 2 Myocardial thickness measurements

Base-to-Apex Gradients from PET and CT

All PET-derived values (i.e., end-diastolic, end-systolic, and summed uptake, MBF and k2), as well as CT-derived myocardial wall thickness, showed a significant base-to-apex gradient (Table 3). While tracer uptake, MBF and wall thickness declined from base to apex, k2 rate constants were higher at the apex compared to the base. The inter-individual range for each gradient is depicted in Figure 4. Of note, gradients for summed tracer uptake, ED, and ES uptake as well as MBF were all significantly less inclining than the gradient of myocardial wall thickness. Finally, the gradient for tracer uptake was inclining more at ED than at ES (0.76 ± 0.07 vs 0.85 ± 0.09, P < .001).

Table 3 Base-to-apex gradient
Figure 4
figure 4

Base-to-apex gradients depicted as mean ± SD (error bars)

Correlation Between PET and CT Findings

No correlation was observed between apical semi-quantitative tracer uptake and the apical thin point thickness (r = − 0.159, P = .237). Likewise, there was no correlation between apical tracer uptake and apical wall thickness (r = − 0.080, P = .553), nor did rest MBF correlate with either parameter (thin point: r = − 0.203, P = .129; apical wall thickness: r = − 0.211, P = .115). Analysis of k2 washout yielded no significant correlation with either thin point thickness (r = − 0.168, P = .230) or apical wall thickness (r = − 0.023, P = .872). Lastly, indexed myocardial wall thicknesses did not correlate with apical tracer uptake (thin pointindexed: r = − 0.128, P = .343; apexindexed: r = − 0.015, P = .914), MBF (thin pointindexed:r = − 0.065, P = .630; apexindexed: r = 0.130, P = .334) , or k2 washout (thin pointindexed: r = − 0.169, P = .226; apexindexed: r = 0.062, P = .658). Similarly, we found no correlation between the wall thickness gradient and summed uptake gradient (r = − 0.037, P = .783), ED gradient (r = 0.016, P = .906), ES gradient (r = 0.140, P = .299), MBF gradient (r = 0.036, P = .791), or k2 gradient (r = − 0.021, P = .883).

Discussion

This is the first study to investigate the prevalence of apical thinning in PET-MPI and its possible association with true anatomical thinning of the left ventricular apical myocardium. In conjunction with partial volume effects, the latter has been suggested as a potential cause for apical thinning in SPECT-MPI. While this theory may be applicable in the setting of the lower spatial resolution in case of SPECT-MPI, only limited data exists for PET-MPI.24

Both autopsy and CT-based studies have reported a high prevalence of focal reduction in myocardial wall thickness at the left ventricular apex. Bradfield et al. measured a thin point of < 2 mm in 97% of post-mortem hearts and in 67% of cases minimal thickness values were even lower than 1 mm, resulting in a mean apical thin point of 1.3 ± 0.7 mm.14 CT measurements in patients without a history of cardiac disease confirmed these findings: Ferencik et al. reported a mean thin point of 1.7 ± 0.7 mm in vertical long axis reconstructions,16 and Johnson et al. measured 1.2 ± 1.1 mm.17 Using prospective ECG triggering and latest-generation CT technology with image acquisition at mid-diastole (i.e., at 75% of the R-R-interval), we observed slightly higher thin point thickness values (2.3 ± 0.8 mm).

In our study population, apical thinning in PET-MPI was very frequent with 93% of the patients having a normalized semiquantitative radiotracer uptake of < 70% at the apex despite the fact that we carefully excluded any patients with myocardial disease involving the apex. Automated scoring based on our historical normal database, which consequentially exhibits apical thinning as well, correctly classified only little more than half (54.4%) of the patients as having normal or only mildly reduced apical rest perfusion, rendering distinction between apical thinning and true apical scars for the interpreting physician difficult in clinical routine. The latter also strongly emphasizes the need for a continuous update of the normal database to match the acquisition protocols and processing algorithms in order to serve as a tool to reliably unmask apical thinning as an artifact. However, future studies directly comparing patients with true apical disease vs normal patients (preferably identified through anatomical imaging such as CCTA) may be needed to assess the true value of semi-automated scoring systems and potentially even recognize specific cut-off values for the apical region.

While we found a statistically significant difference in apical wall thickness in patients with RS < 2 as compared to patients with RS ≥ 2, and while this difference may be considered negligibly small in absolute numbers (i.e., 0.4 mm), it has to be taken into account that the difference of observed tracer activity assessed by PET between two objects of varying size is influenced by differing recovery coefficients for these objects during image reconstruction. In our setting, however, we assume this relative change in observed tracer activity for a change in wall thickness of 0.4 mm to lie in the range of up to 3% which does not entirely explain the observed differences in semiquantitative normalized apical tracer uptake between both patient groups (i.e., 62.5 ± 6.3% vs 53.9 ± 4.8%). By contrast, our findings suggest that the partial volume effect may play a more important role in the creation of the apical thinning phenomenon as a comparison of ED and ES datasets revealed a significant increase in apical tracer uptake and a less prominent base-to-apex gradient at ES. However, limiting the analysis to ES did not eliminate apical thinning in the majority of patients as 74% still exhibited apical tracer uptake below the threshold of 70% (compared to 100% of patients on ED images). Interestingly, however, base-to-apex gradients were much less pronounced for the PET-derived parameters as compared to myocardial wall thickness, contrary to what would be expected from a very strong influence of the partial volume effect. Finally, apical myocardial wall thickness did not correlate with tracer uptake in absolute numbers nor when comparing their respective base-to-apex gradients on summed, ED or ES images. From this, we conclude that other factors than true anatomical wall thickness and/or the partial volume effect play a more prominent role in the creation of the apical thinning artifact. This conclusion is corroborated by the fact that in some cases, apical thinning and severe reduction in apical myocardial wall thickness coincide (Figure 5A), while in others such an association cannot be found (Figure 5B). Finally, in some cases marked apical thinning can even be observed in patients with a thin point thickness of up to 4 mm (Figure 5C).

Figure 5
figure 5

(A) Polar map display of semiquantitative tracer uptake (top) and vertical long axis view (bottom) of a patient exhibiting both visual apical thinning and true anatomical thinning (thin point thickness 1.4 mm). (B) example of anatomical thinning (thin point 1.4 mm) but only minimal visual apical thinning on PET-MPI. (C) occurrence of apical thinning in a patient without true anatomical thinning (thin point thickness 4.1 mm)

Apical thinning has been reported to be more pronounced on attenuation-corrected SPECT images than on non-corrected images (NC), with some authors suspecting an artifact induced by misalignment of the attenuation map,8,11 and with others assuming an unmasking of true anatomical wall thinning by AC.25,26,27,28,29,30 Specifically, it has been postulated that, on NC images, the apex shows falsely high counts as it is located close to the chest wall and therefore less affected by tissue attenuation. AC algorithms for SPECT usually incorporate additional scatter compensation and resolution recovery which may further accentuate apical thinning.2,25,28 In light of this, it may be perceived as a limitation that we did not assess NC images in our study and therefore cannot rule out AC as a confounder. However, in PET-MPI, AC is a mandatory process according to current guidelines.1 Additionally, all AC maps in the current study were manually checked for any perceivable misregistration and, while artifactual decreases in uptake because of AC misregistration are a proven concept, they are mostly encountered in the lateral and anterior wall rather than at the apex.8,9,10,31,32,33 Therefore, we would expect to see additional defects in other areas of the heart if AC was a major contributing factor to apical thinning. The same reasoning may also apply to respiratory and patient motion artifacts.27 Regional differences in metabolic trapping of 13N-ammonia at the apex, as previously described in the inferior and posterolateral wall7, are not supported by our findings. While washout at the apex was higher compared to the base (Figure 4), it was subject to a high level of variation and did not differ between patients with mild or pronounced apical thinning.

Limitations

It may be perceived as a limitation that we used CT acquired at mid- to end-diastole (i.e., at 75% of the R–R interval) to assess anatomical wall thickness as this may not accurately reflect the true end-diastolic myocardial wall thickness. However, while such potential overestimation of the apical wall thickness may have had an impact on our analyses based on absolute measurements, the effect on the gradient analyses, which are based on ratios, should be negligible. Furthermore, the fact that some datasets were acquired and reconstructed differently, may be seen as a limitation namely because of the resulting differences in spatial resolution. Hence, we have performed all statistical analyses separately for both groups of patients as well and did not find any discrepant results that would have changed our conclusions. Finally, the historical normal database used in the present study, which comprises patients with low likelihood of CAD imaged during different scanner eras with differing acquisition protocols (e.g., 2D and 3D acquisition) may be imperfect and normal databases from other institutions may yield differing results.

New Knowledge Gained

Apical thinning in 13N-ammonia PET-MPI has a very high prevalence.

Apical thinning as seen in PET-MPI is not solely attributable to true anatomical apical wall thickness and/or the partial volume effect.

Conclusion

Apical thinning is a highly prevalent finding during 13N-ammonia PET-MPI that is not solely attributable to true anatomical apical wall thickness or the partial volume effect. Other factors that yet need to be identified seem to have a more prominent impact.