Introduction

Electroporation is a phenomenon that takes place when cells are exposed to intense electric fields, which dramatically change the transmembrane potential and lead to the creation of nanopores on the cell membrane.1,2 In clinical settings, non-lethal pulses have been administered to increase the permeability of cell membranes, allowing for the passage of large macromolecules through the membrane of targeted cells. These treatment modalities are referred to as electrochemotherapy3,4 or electro-gene therapy,1,2,5,6 depending on the adjunctive macromolecule. These therapies aim to minimize the lethality of the electrical pulses to optimize the biological effect of the injected therapy. In contrast, irreversible electroporation (IRE) protocols are designed to permanently disrupt cell membranes within a targeted region.7,8 IRE mechanisms are independent of thermal processes and clinical protocols are designed to minimize temperature rises in the tissue. This makes IRE an ideal treatment option for tumors near critical blood vessels and nerves where thermal ablative techniques, including radio-frequency and microwave ablation, may cause collateral damage or may be limited by the heat-sink effect.9,10 Recent clinical trials have demonstrated that IRE is highly efficacious for tumors where surgical removal or thermal ablation are contraindicated.11

Clinically, IRE is accomplished via introduction of percutaneous needle electrodes into or around the target tissue and a series of high voltage pulses, ranging from 50 to 100 µs in duration, are delivered.1113 Typical IRE protocols deliver 80–100 pulses, each separated by an approximately 1 s delay. When implemented clinically, pulses are synchronized with the patient’s heartbeat to minimize triggering of irregular heartbeats and a robust anesthetic protocol including a neuromuscular blockade is implemented to prevent the stimulation of muscle contractions.14 The NanoKnife clinical ablation system (AngioDynamics, Inc., Latham, NY) utilizes a voltage-to-distance ratio of 1500–1700 V/cm and the output voltage is adjusted based on physician-reported measurements of electrode spacing.15

A positive review highlighting the safety and efficacy of these treatments in a clinical setting was recently published by Scheffer et al.16 and IRE appears to be ideally suited for the treatment of tumors less than 3 cm in each dimension with a success rate between 93%17 and 98%.18 IRE is currently being evaluated as a treatment modality against a number of oncological disorders including pancreatic,15,19,20 lung,21 brain,22 kidney,2327 and liver28,29 cancers. Long-term follow-up on these patients indicated that IRE has the potential to leave critical blood vessels, nerves, and the more general healthy stroma unharmed.3032

While the past decade is marked by numerous positive results from IRE trials, a prospective clinical trial for lung tumor treatment was stopped when investigators observed that only 30% of treatment subjects (n = 7) showed complete tumor regression at the 12-month follow-up.33 The reason for this inconsistency is unclear; however, it is potentially due to the large degree of heterogeneity in lung tissue. The presence of air pockets may also insulate and/or protect regions of the tumor, possibly diminishing the overall effectiveness of standard IRE treatments.34 The complexity of treating tumors in these highly heterogeneous tissues and the need to administer a neuromuscular paralytic to avoid intense muscle contractions apparent in standard IRE treatments necessitate the research and development of novel IRE pulse protocols to mitigate these challenges.

High frequency IRE (H-FIRE) is a relatively new variant of IRE which replaces monopolar 100 µs pulses with a burst of shorter duration (0.25–5.0 µs) alternating positive and negative (bipolar) pulses separated by a short inter-pulse delay (Fig. 1). Theoretical models of H-FIRE predict that this therapy is capable of penetrating heterogeneous tissues and producing more predictable ablations than typical IRE pulses.35 For example, blood vessels in the treatment area can influence the electrical field of standard IRE pulses if they are not properly accounted for.36 This is due to the fact that IRE pulses are relatively long in duration (≥50 μs) and therefore have lower frequency components (<20 kHz), which are susceptible to the effects of heterogeneity (Supplemental Fig. 1a). At these frequencies, the impedance of muscle, fat, stroma, and other tissues can vary significantly from healthy or tumorous tissue.37 At higher frequencies the differences in impedance between these tissue types are minimized (Supplemental Fig. 1b) and the tissue appears more electrically homogeneous and therefore an electric field distribution which more closely matches the analytical solution is produced.38 Arena et al. recently demonstrated that bursts of 1 or 2 µs pulses can be used to ablate tissue while additionally eliminating the muscle contractions associated with IRE pulses, further contributing to the efficacy and potential of H-FIRE in the tumor ablation field.39 Siddiqui et al. recently showed that H-FIRE can be safely and predictably delivered next to critical hepatic anatomy such as vascular and biliary structure.10

Figure 1
figure 1

H-FIRE protocols replace the monopolar IRE pulses with a burst of bi-polar pulses. (a) Standard IRE protocol delivers 50–100 µs mono-polar pulses separated by a 1 s inter-pulse delay; (b) experimental 1.5 kV 50 µs IRE pulse; (c) H-FIRE protocols deliver a high frequency burst of 1–10 µs bi-polar pulses. Each burst of pulse cycles is separated by a 1 s delay. Energized time per burst or dose rate, represents the total charged time per burst; (d) experimental 1.5 kV H-FIRE burst with 26 × 2 µs pulses and a 10 µs delay between alternating polarity pulses; and (e) the first three alternating pulses within an H-FIRE burst.

H-FIRE pulses are situated in a relatively unexplored region of the pulse-duration space40 and the lethal electric field thresholds and long term biological responses for these pulse waveforms are comparatively unknown. The transmembrane potential does not rise instantaneously during exposure to a pulsed electric fields41 and the charging time for mammalian cell membranes is approximately 1 µs.42 As result, a number of interesting biological response arise when H-FIRE pulses are delivered. The processes of pore formation and expansion appear to decrease for short pulse durations (0.25–5 µs) as compared to longer duration pulses in the 100 µs to 10 ms range and permeabilization of the cell membrane may not be as robust or long lasting for short pulses.43

Current H-FIRE protocols mimic clinical IRE delivery procedure by energizing each burst for 50–100 µs and applying a series of 80–100 bursts at a 1 Hz repetition rate. Recent studies have shown that the lethal electric field threshold increases significantly as pulse duration is decreased.41 An H-FIRE protocol consisting of 80 bursts of 1–5 µs pulses was shown to reduce the viability of pancreatic cancer cells in suspension to less than 20% for fields strengths of 3000 V/cm or more. In contrast, it was shown that 1250 V/cm is for equivalent energy 100 µs IRE protocols to reduce the viability of these cells to the same extent.44 For cells in suspension, H-FIRE protocols utilizing 250 and 500 ns pulses appear to be relatively ineffective even at field strengths up to 4000 V/cm. However, when cells are grown in collagen tumor mimics, lethal thresholds of 2000, 1700, 1100, 760 V/cm were found for H-FIRE protocols with 0.25, 0.5, 1.0, and 2.0 µs pulses, respectively.45 Similar studies found that IRE protocols had a lethal threshold of 500 V/cm in collagen tumor mimics44 and 300–640 V/cm in liver tissue.4648 This discrepancy between cells in suspension, those grown in a physiologically mimicking environment, and in vivo highlights the need for further investigation of these novel protocols and their resulting thresholds.

In this work, potato tissue phantoms are used as a surrogate to enable the testing of a wide parameter array including the number of bursts and individual pulse duration in search of a potentially optimized treatment protocol. We investigated the lesion volumes produced by a clinical prototype system which can deliver bipolar H-FIRE pulses up to 2500 V in amplitude. We show that when delivering the same amount of energy, there is a logarithmic relationship between pulse width (PW) and ablation size. Finally, we show that there is no statistical difference in the ablation volumes created by delivering 40 through 200 bursts across two needle electrodes. This work highlights the need to understand the effect of unique parameters associated with H-FIRE not present in IRE to result in clinically useful application of H-FIRE for tissue ablation.

Materials and Methods

Electroporation Setup

Experiments were performed using potatoes as a tissue phantom to enable the evaluation of a large number of parameters. Potato tissue is relatively homogeneous and exhibits a significant increase in conductivity after electroporation, mimicking the response seen in vivo. 2,49 While potato does not recapitulate all of the properties of an in vivo tumor, it is a reasonable model for assessing trends and enables comparison of ablations to numerical models to determine the lethal thresholds as is done in vivo. 47,50,51 Lesions created in the tissue become visible 12–24 h post treatment due to the production of melanin52 and are easily measured using calipers.

Monopolar IRE pulses were generated using a commercial pulse generator (ECM 830, Harvard Apparatus, Holliston, MA) and bipolar H-FIRE pulses were generated using a custom pulse generator. Treatments were delivered to the tissue through two 1.0 mm diameter clinical IRE probes (NanoKnife™ Single Electrode Probes, AngioDynamics, Inc., Latham, NY) separated by 1.0 or 2.0 cm. These probes have an adjustable polyimide sheath which was retracted to expose 0.5 cm of the stainless steel electrodes. Voltage and current measurements were recorded using a custom high speed data acquisition system. To facilitate comparison between groups, a simplified electrical dose formula was used

$$Dose = V^{2} \;*\;T_{\text{p}} \;*\;n\;*\;N\;\left( {{\text{V}}^{2} \;{\text{s}}} \right),$$
(1)

where V is the applied voltage, T p is the pulse width, n is the number of pulses per burst, and N is the number of bursts per treatment. This calculation of dose was chosen as a simple way to compare the ablation outcomes between treatments with varied pulse parameters. It was recently shown by Sano et al. that when only voltage is varied between H-FIRE treatments, this calculation of equivalent dose results in equivalent lethal thresholds.45 Here, we compare treatments where voltage is held constant and other parameters are varied.

Pulse parameters used during this study are shown in Table 1 and the experimental set-up is shown in Fig. 2. 48 h post-treatment, the potatoes were sliced parallel to the electrode paths to expose the ablated region. Lesions were compared by measuring the optically visible lesions in potato tissue, which became distinctively darker in color, 48 h post-treatment.

Table 1 Treatment protocols evaluated.
Figure 2
figure 2

Experimental set-up. A custom pulse generator produces bipolar H-FIRE pulses (top left). The BTX ECM 830 pulse generator produces standard monopolar IRE pulses (top right). A custom measurement system (center) records the pulses and displays voltage, current, and resistance data on a laptop computer.

The visibly darkened tissue is the result of the oxidation process catalyzed by polyphenol oxidase released as the cell membranes lose integrity resulting in the generation of melanins.52,53 In a preliminary study, potatoes were sectioned incrementally between 6 and 72 h post-treatment and 48 h was determined to be the optimal time post-treatment for full coloration of the ablation zone to occur. Additionally, potatoes were sectioned at 30 h and the boundary of the ablation was monitored for 6 h. Over this time, the ablation volume remained constant (Supplemental Fig. 2), indicating that melanins were not diffusing out into untreated regions and confounding measurements. The characteristic lesion resulting from electroporation treatment displays a near-ellipsoid shape and lesion volume was calculated by representing the lesion as an ellipsoid with volume (v)

$$v = \frac{4}{3}\; \cdot \pi \cdot a \cdot b \cdot c {{\text{ [cm}}^{3}} \text{],}$$
(2)

where a, b, and c are one half of the length, width, and depth of the ablations. All values are reported as mean ± standard deviation.

Finite Element Analysis

In silico models for determining the lethal thresholds of IRE and H-FIRE treatments were generated using COMSOL Multiphysics (V4.3, COMSOL, Palo Alto, CA). The general geometry of the potato tissue was represented as an ellipsoid with the average dimensions of the tissue phantoms. Two 0.5 cm long cylindrical geometries with a 1 mm diameter were inserted into the center of the domain to mimic the experimental placement of the electrodes. The electric field distribution was found by solving the equation:

$$\nabla \cdot \left( {\sigma_{\text{d}} \nabla \varphi } \right) = 0\;[{\text{A}}/{\text{m}^3}],$$
(3)

where σ d represents the electrical conductivity of the tissue as a function of the applied electric field and φ is the electric potential.54 Boundary values for the electrode–tissue interface were set to φ = V 0 and 0, where V 0 is the experimental voltage. A sigmoidal changing conductivity function was incorporated into the model to simulate the changes due to electroporation by implementing a piecewise function in COMSOL which was smoothed to be second derivative continuous with a center point at 255 V/cm, the approximate IRE threshold of potato tissue.55 The initial conductivity of 0.039 S/m was chosen based on experimental measurements.

It was found that the conductivity of healthy untreated tissue was significantly higher than tissue within the experimental ablation zones (Fig. 3). Additionally, this change in conductivity was frequency dependent with a larger change occurring at 1 kHz than at 1 MHz. To account for these changes, the simulations were repeated for multiple values of conductivity change. The maximum conductivity used in the sigmoidal function set as a multiple (1×–15×) of the initial conductivity based on impedance measurements of healthy and ablated tissue (Fig. 3). Multipliers of 2×, 2×, 3×, and 9× were used for the 1, 2, 5, and 50 μs pulse duration groups, respectively. These values reflect the change in impedance between healthy and ablated tissue measured at 1 MHz, 500 kHz, 200 kHz, and 20 kHz, respectively and correspond to the center frequency of each pulse group. Multiplier values were rounded to the nearest integer. All remaining boundaries were defined as electrically insulating.

$${\mathbf{n}} \cdot {\mathbf{J}} = 0\;{\text{[A}}/{\text{m}^2{\text ],}}$$
(4)

where n is the normal vector to the surface, J is the electrical current density.

Figure 3
figure 3

Electrical impedance spectrum of potato tissue before and after ablation. (a) A five electrode impedance probe was used to (b) measure the electrical impedance of ablated and untreated regions of the tissue between 1 kHz and 1 MHz; (c) there is a significant difference in the conductivity of healthy and ablated tissue at all frequencies in this range. A maximum difference of 13.9× was observed at 1 kHz and a minimum difference of 2.3× was observed at 1 MHz; and (dg) numerical simulations showing the 125 V/cm iso-contours for models using 1×, 2×, 4×, 6×, 10×, and 15× changes in conductivity (inside to outside). Simulations were conducted with (d) 1.0 kV; (e) 1.5 kV; (f) 2.0 kV; and (g) 2.5 kV between the electrode pairs.

To improve the simulations, two strategies were employed. First, a two-step simulation was conducted where a stationary ‘pre-pulse’ was applied using a voltage of 50 V. This voltage does not result in electric fields sufficiently high enough to change the tissue conductivity and allows the simulation to determine a baseline electric field distribution. The solution to this simulation was used as the initial values of a second sequential stationary solver with the experimental voltage applied to the energized electrode. This process significantly reduces the simulation time. Second, the relative tolerance of each of these solvers was reduced from the default of 0.01 to 1E−12. The nonlinear method was changed from automatic (Newton) to automatic highly nonlinear (Newton) with a maximum number of iterations of 2500, an initial damping factor of 1E−8, and minimum damping factor of 1E−16. This forces the solver to search for a more accurate solution and eliminates finite element artifacts.

The electrodes were given a default ‘extra fine’ mesh and the main simulation geometry was defined as a ‘normal’ mesh. This mesh was refined once and the maximum error between additional successive refinements was 1.8% and the average error was 0.3%. The simulation solved 867,257 degrees of freedom using a mesh with 636,893 domain elements, 39,994 boundary elements, and 3122 edge elements. The simulations for all voltage and conductivity values were completed in approximately 145 h on an Intel i7 quad core processor with 32 GB RAM. Results of these simulations with 2500 V applied between the electrodes are shown in Fig. 4.

Figure 4
figure 4

Finite element modeling of tissue phantom to determine lethal thresholds. (a) A dynamic conductivity map was used to model the instantaneous effects of electroporation on the tissue’s electrical properties; and (b) electric field distribution when 2500 V was applied between the two energized electrodes.

Determination of Electric Field Threshold for Cell Death

48 h post-electroporation, the measured width, height, and depth of the darkened region of the potato tissue were recorded (Supplemental Fig. 3). These measurements were matched to an electric field magnitude in the x, y, and z axis obtained from the finite element analysis. All experimental groups reported were repeated a minimum of three times (N = 3) and the mean ± standard deviation for each group were determined from the electric field strengths determined in all three dimensions (e.g., a minimum of nine values).

Results

Equivalent Energy: Pulse Width Analysis

To investigate the effects of constituent PWs on lesion volume, four treatment groups receiving equivalent energy doses with varying PWs were investigated. Electrical dose was held constant at 11,250 V2 s by energizing the electrodes to 1500 V and delivering 100 bursts, each energized for 50 µs. Bursts with 1, 2, and 5 µs bipolar pulses or 50 µs monopolar constitutive pulses were evaluated with 50, 24, 10, and 1 pulses per burst, respectively. This method for defining dose was based on the assumption that H-FIRE treatments will be constrained to the on-time (50–100 µs) and pulse repetition rates (1 Hz or cardiac synchronization) currently employed in the clinic.

We found that lesion volumes increased as constituent pulse duration increased despite delivering equivalent doses of integrated energy (Fig. 5a). Bursts with 1, 2, 5, and 50 µs pulses produced lesion volumes with the following means and standard deviations: 1.75 ± 0.36, 3.14 ± 0.34, 5.56 ± 0.15, and 12.57 ± 1.82 cm3, respectively. The lethal threshold for these same respective groups were 312.3 ± 41.8, 257.3 ± 7.6, 222.0 ± 7.0, and 164.4 ± 10.4 V/cm (Fig. 5b). The lethal thresholds and ablation volumes were found to be statistically significantly different between all groups (α ≤ 0.05).

Figure 5
figure 5

Constitutive pulse length has a large effect on lesion volumes when treatments deliver equivalent energy doses. Bursts with shorter constitutive pulses produced smaller ablations. The data shows a strong logarithmic correlation (y = 2.81 * ln(x) + 1.38) between pulse length and lesion volume (r 2 value of 0.995). All values between groups were found to be statistically significantly different for α = 0.05 (**) and α = 0.01 (***).

Burst Number Analysis on Lethal Thresholds

Typical IRE protocols prescribe the delivery of 80–100 pulses. To investigate how deviating from this affects the treatment volume, we investigated six additional protocols. The number of bursts was varied between 10 and 200. Applied voltage (2000 V), PW of (2 μs), and an energized time of per burst (100 μs) were held constant. This enabled the electrical dose to vary between 4000 and 80,000 V2 s. A higher voltage was used in this experiment to ensure that the 10× burst group produced contiguous lesions.

The data shown in Fig. 6, indicates that lesion volume increases between 10 and 40 bursts. Treatment volume plateaued for groups that received more than 40 bursts. Treatments with 10, 20, 40, 60, 80, 100, 150, and 200 bursts produced lesions which were 1.76 ± 0.68, 5.72 ± 1.91, 7.17 ± 0.87, 8.10 ± 1.38, 8.16 ± 1.01, 8.08 ± 0.47, 9.52 ± 1.12, 8.46 ± 0.13 cm3, respectively. The lethal threshold for these same respective groups were 417.8 ± 44.8, 301.5 ± 40.1, 268.7 ± 39.1, 261.8 ± 39.5, 258.4 ± 40.1, 256.0 ± 48.5, 246.6 ± 42.1, and 258.6 ± 32.8 V/cm. Representative ablations from these groups can be seen in Fig. 7. The ablation volumes for treatments with 10 bursts displayed statistically significantly differences compared to groups receiving 40–200 bursts (α = 0.01). The ablation volume created by 20 bursts was also statistically different from the volume created by 200 bursts (α = 0.01). The lethal thresholds for treatments with 10 bursts were statistically different than all groups (α = 0.01) and treatments receiving 20 bursts had statistically different thresholds than the groups receiving 60 or more bursts. The lethal thresholds for groups with 40 through 200 bursts were not found to be statistically significantly different.

Figure 6
figure 6

Ablations created with bursts of 2 µs pulses where each burst was energized for 100 µs. (a) Treatments with 10× bursts produced the smallest ablation volumes. Ablation volumes increased sequentially when 20×, 40×, and 60× bursts were delivered. However, no significant increases in ablation volume were found for trials where 60 pulses or more were administered; and (b) the lethal threshold decreased sequentially when 10×–60× bursts were delivered. However, no significant changes in lethal threshold were found between treatments receiving 40 or more bursts. Statistical significance between groups was determined using a one-sided Student’s T test with unequal variances and α = 0.01 (***).

Figure 7
figure 7

Ablations created with a 2 cm separation, 0.5 cm exposure, 2500 V amplitude pulses. (a) 10×; (b) 20×; (c) 40×; (d) 60×; (e) 80×; (f) 100×; (g) 150×; and (h) 200× bursts. Delivery of more than 40 bursts does not significantly increase lesion volume. Scale bar is 2 cm.

Discussion

H-FIRE is a relatively new tissue ablation technique and the effects of different protocols on resulting lesion volume have not been fully examined. When investigating the effects of PW, it was found that when electrical dose is held constant, bursts with longer pulses are significantly more effective at ablating tissue. Bursts with 50 µs constitutive pulses produced lesion volumes which were 7.2× larger than those with 1 µs constitutive pulses. This trend was highly correlated to a logarithmic curve [volume = 1.28·ln(PW) + 1.38 (cm3), r 2 = 0.995], where PW is the pulse width in μs.

The lethal threshold found for H-FIRE protocols with 50 µs bursts (Fig. 5b) was approximately 165 V/cm in our potato tissue phantom. Similar IRE protocols in liver and brain suggest a lethal threshold of approximately 500 V/cm for mammalian tissue,22,46 3× greater than the potato tissue phantom. We anticipate that H-FIRE protocols will follow a similar trend in vivo and require an electric field approximately three times greater than potato to ablate mammalian tissue. Here we found lethal thresholds of 257 and 227 V/cm for 2 µs H-FIRE protocols with 100 bursts which were energized for 50 and 100 µs, respectively. Applying this 3× assumption, these protocols should require approximately 680–770 V/cm to ablate mammalian tissue. Lethal thresholds for pancreatic cancer cells grown in 3D culture were found to be 855 and 718 V/cm for H-FIRE treatments with 2 µs constitutive pulses energized for 50 and 100 µs, respectively.45 Similar 2 µs H-FIRE treatment protocols delivered to in vivo rabbit liver corresponded to a lethal threshold of 861 V/cm which were 2.5–2.7× higher than matched 50 and 100 µs IRE treatments, respectively.56 This indicates that ablations in potato tissue can provide some limited guidance for the development of clinical protocols.

While longer duration pulses appear to be more effective at ablating tissue, muscle contraction intensity is highly correlated to pulse duration. In clinical settings, 50–100 µs pulse lengths require a very intense anesthetic protocol to minimize these muscle contractions14 and delivery with cardiac synchronization in order to avoid inducing arrhythmia.57 A critical advantage of the H-FIRE protocol is the use of ultra-short bipolar pulses which have the potential to eliminate these muscle contractions. Rogers et al. showed that the threshold for muscle contractions, of gastrocnemius muscles, increased from 1.83 to 112 V/cm when pulse duration was decreased from 100 to 1 μs, a 61× increase. In practical terms, this means that a significantly smaller volume of tissue will be exposed to muscle stimulating electric fields when 1 μs pulses are applied (Supplemental Fig. 4). In contrast, we show here that the lethal threshold for bursts of 1 μs pulses is only 1.9× higher than for mono-polar 50 μs pulses. The significant increase in muscle contraction threshold paired with a relatively small increase in lethal threshold indicates that clinically relevant ablations can be created without inducing the extreme muscle contractions seen in typical IRE procedures. This may also mitigate the likelihood of inducing cardiac arrhythmias and enable treatment without the need for chemical paralytics or cardiac synchronization.

The work presented here indicates that the use of a small number of high energy bursts may be beneficial as statistically equivalent lesion volumes were created by 40–200 bursts when pulse length, number of cycles, and voltage were held constant. Using IRE pulses, Bonakdar et al. observed a large change in the conductivity of potato tissue during the first 30 pulses and a significantly less dramatic change between 30 and 100 pulses55 indicating that most electroporation is occurring early on in the treatment. Typical clinical IRE protocols follow the guidance reported in early studies by Al-Sakere et al. for the delivery of 100 pulses per location and some treatment protocols have been reported with 270 total pulses per treatment location.58 The results reported here indicate that pulse delivery, beyond a critical threshold, may not significantly increase the ablation volume.

This data does not refute the clinical benefit that additional pulses may provide and it is possible that the optimal number of pulses is different for various healthy and tumorous tissue types which cannot be recreated in a potato model. However, further investigation in vivo is warranted as this could have significant translational effects through the reduction of operating room time and anesthetic dose required. Additionally, overall procedure costs climb the longer the patient is under anesthesia or in a perioperative state and therefore a systematic reduction in this time is financially beneficial. For example, when treating large tumors (>3 cm), it is common for physicians to bracket the tumor with four or more electrodes and deliver pulses between each probe combination. A hypothetical four electrode treatment would currently deliver 600 pulses. If it is found in vivo that ablation volumes do not increase past 40 pulses, than this would reduce the total treatment by 60% to 240 pulses. This has the potential to drastically shorten treatment times for very large tumors where the electrodes are pulled back to treat multiple locations.

There are a number of important limitations to this study. Potato tissue is far from an ideal model for investigating IRE. The lethal thresholds for potato tissue are drastically lower than mammalian tissue resulting in ablations which are substantially larger than those seen in vivo. Additionally, this tissue phantom model is not perfused and Joule heating could be greater in this model than observed in vivo. It is unclear how lethal thresholds observed in this model will translate to healthy and cancerous tissues in vivo and the thresholds presented here should not be used for clinical treatment planning. Despite these limitations, potato remains the one of the most viable models for comparing the effects of different H-FIRE protocols at clinically relevant voltages (1000–3000 V).

Conclusion

In this study, we used a potato tissue phantom model to confirm in vitro data on cells in suspension and 3D culture which showed that lethal thresholds for H-FIRE treatments increase as constitutive pulse duration decreases, even when equivalent energy doses are delivered. Data from these studies indicate that delivering over 40 bursts, while holding other parameters constant, does not significantly reduce the threshold for cell death or increase the total ablation volume. This implies that alternative strategies are necessary to increase the size of ablations created by H-FIRE. Future work should focus on optimizing protocols to deliver the fewest high-energy H-FIRE bursts possible while avoiding thermal damage to the tissue.