Abstract
Purpose
To compare image quality of abdominal arteries between full-iodine-dose conventional CT and half-iodine-dose virtual monochromatic imaging (VMI).
Materials and methods
We retrospectively evaluated images of 21 patients (10 men, 11 women; mean age, 73.9 years) who underwent both full-iodine (600 mg/kg) conventional CT and half-iodine (300 mg/kg) VMI. For each patient, we measured and compared CT attenuation and the contrast-to-noise ratio (CNR) of the aorta, celiac artery, and superior mesenteric artery (SMA). We also compared CT dose index (CTDI). Two board-certified diagnostic radiologists evaluated visualisation of the main trunks and branches of the celiac artery and SMA in maximum-intensity-projection images. We evaluated spatial resolution of the two scans using an acrylic phantom.
Results
The two scans demonstrated no significant difference in CT attenuation of the aorta, celiac artery, and SMA, but CNRs of the aorta and celiac artery were significantly higher in VMI (p = 0.011 and 0.030, respectively). CTDI was significantly higher in VMI (p = 0.024). There was no significant difference in visualisation of the main trunk of the celiac artery and SMA, but visualisation of the gastroduodenal artery, pancreatic arcade, branch of the SMA, marginal arteries, and vasa recta was significantly better in the conventional scan (p < 0.001). The calculated modular transfer function (MTF) suggested decreased spatial resolution of the half-iodine VMI.
Conclusion
Large-vessel depiction and CNRs were comparable between full-iodine conventional CT and half-iodine VMI images, but VMI did not permit clear visualisation of small arteries and required a larger radiation dose.
Key Points
・Reducing the dose of iodine contrast medium is essential for chronic kidney disease patients to prevent contrast-induced nephropathy.
・In virtual monochromatic images at low keV, contrast of relatively large vessels is maintained even with reduced iodine load, but visibility of small vessels is impaired with decreased spatial resolution.
・We should be aware about the advantages and disadvantages associated with virtual monochromatic imaging with reduced iodine dose.
Similar content being viewed by others
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.Avoid common mistakes on your manuscript.
Introduction
Computed tomography (CT) angiography is widely used to evaluate vessels throughout the body, and safe surgical or catheter intervention requires good understanding of the distribution of small arteries and anomalies in the branching of vessels around a lesion [1,2,3].
Visualisation of the small peripheral vessels in CT angiography necessitates the injection of a sufficient dose of iodine-based contrast medium. However, in patients with chronic kidney disease (CKD), the development of contrast-induced nephropathy (CIN) has been correlated with the excessive administration of contrast medium containing iodine, and minimal use of contrast has been recommended [4,5,6].
Recently, image reconstruction utilising virtual monochromatic imaging (VMI) in dual-energy CT has been adopted in an effort to reduce the amount of iodine-based contrast material needed in CT angiography. VMI allows an increase in intraluminal CT values at lower levels of energy, nearer those of the K-edge of iodine (33.2 keV) [7], which maintains vascular contrast even when the iodine dose is reduced. Indeed, past studies have reported comparable contrast-to-noise ratios (CNRs) in VMI-reconstructed images of CT angiography acquired with a reduced dose of iodine-based contrast medium and images of conventional CT angiography obtained using a full dose [8,9,10]. Nevertheless, the spatial resolution of the reconstruction method also affects the visualisation of small arteries [11], and we believe no study of CT angiography with VMI has examined spatial resolution and the visualisation of such peripheral arteries, though some researchers have focused on the CNR of relatively large vessels [8,9,10].
In emergent situation, the dose of iodine is not the main issue [12], and in preoperative situations, the need for better image quality, especially visualisation of small arteries, possibly takes priority to reduction of iodine dose to perform a safe procedure.
We therefore compared image quality between full-iodine-dose conventional CT and half-iodine-dose virtual monochromatic imaging in CT angiography to ascertain advantages and disadvantages of these two protocols.
Materials and methods
Calculation of the modulation transfer function
We calculated the modulation transfer function (MTF) using a phantom (Fig. 1) consisting of an elliptically shaped acrylic cylinder (diameter, about 30 × 20 cm; height, about 10 cm) with a columnar hole at its centre (diameter, about 4 cm; height, about 8 cm) that we filled with either a 20-fold dilution of iopamidol contrast medium (Oiparomin 300, Fuji Pharma Co., Ltd.) for conventional scanning or a 40-fold dilution for dual-energy scanning. For dual-energy scanning, we reconstructed virtual monochromatic images at 52 keV, setting scan parameters for the respective imaging systems to yield a similar CT dose index (CTDI; LightSpeed VCT, 120 kVp [GE Healthcare]; 34.99 mGy, Discovery CT750 HD 120 kVp [GE Healthcare]; 35.43 mGy; Discovery CT750 HD 52 keV; 33.96 mGy). We obtained the edge spread function at the edge of the iodine contrast material filling the cylinder in the acrylic phantom, differentiated the edge spread function to determine the line spread function, and applied fast Fourier transformation of the line spread function to obtain the MTF [13,14,15].
Patient population of the clinical study
Our institutional review board approved this retrospective single-centre study and waived the requirement for informed consent from participants.
We searched our CT database to identify consecutive patients who underwent VMI using the half-iodine-dose (300 mg/kg) protocol because of impaired kidney function (most recent estimated glomerular filtration rate less than 45 mL/min/1.73 m2) between July 1 and September 30, 2016 and found 33 patients. We then further narrowed this group of 33 to 21 who also underwent conventional CT utilising the full-iodine-dose (600 mg/kg) protocol within the past 5 years (September 30, 2011 to September 30, 2016), and excluded the other 12. On average, the conventional scans were obtained 1.6 years before VMI acquisition. The study population comprised 21 patients (10 men, 11 women; mean age, 73.9 years; average body weight, 56.3 kg) who underwent CT screening for metastasis of colon cancer (n = 10), renal cancer (n = 6), cystic cancer (n = 3), uterine cancer (n = 1), and gastric cancer (n = 1). The scan ranges varied between the scan types and the individuals evaluated. The range was from the chest to the pelvis in 19 of the 21 patients for half-iodine VMI and 12 of the 21 for full-iodine conventional CT and from the upper abdomen to the pelvis in the other 2 for half-iodine VMI and 9 for full-iodine conventional CT.
Scan protocol
Our conventional scan protocol called for the injection of 600 mg/kg of iodine contrast medium in 30 s, and the VMI protocol utilised half that dose, 300 mg/kg, injected in 30 s. The scan of the CT angiography was begun at 40 s after the start of contrast material injection.
For the full-dose protocol, parameters were: collimation, 64 × 0.625 mm; pitch, 1.375; gantry rotation, 0.4 second; scan field of view (FOV), 50 cm; and tube voltage, 120 kVp. Tube current was automatically changed with auto exposure control (AEC) with a noise index of 12. Nine of the 21 patients were scanned using the LightSpeed VCT system and 12 with the Discovery CT750 HD system.
For the half-iodine VMI protocol, parameters were: collimation, 64 × 0.625 mm; pitch, 1.375; scan FOV, 50 cm; tube voltage, fast kilovoltage switching between 80 and 140 kVp. Gantry rotation and tube current were determined depending on the patient’s body size to achieve a noise index close to that of the conventional scan. All 21 patients were scanned with the Discovery CT750 HD system.
We employed a standard reconstruction kernel to reconstruct images of a section of 1.25-mm thickness and interval, and applied 40% adaptive statistical iterative reconstruction (ASiR) to reduce image noise. For VMI protocol, we reconstructed virtual monochromatic images at 52 keV.
Quantitative analysis of arteries
In axial images, circular regions of interest (ROIs) of 10-mm diameter were placed on the aorta (at the level between the orifice of the celiac artery and the SMA) and right psoas muscle, and circular ROIs of 2-mm diameter were placed on the main trunks of the celiac artery and SMA. CT attenuation was measured as the average CT values in these ROIs. Image noise was defined as the standard deviation (SD) of the CT values in the ROIs placed on the right psoas muscle. CNR was calculated as: CNR = (μA – μB) /ρ, in which μA is the signal intensity of the aorta, celiac artery, or SMA; μB is the signal intensity of the psoas major muscle, and ρ is the image noise.
We compared CTDI and dose-length product (DLP) between the VMI and conventional scans based on the dose report.
Subjective evaluation of proximal and peripheral arteries
We reconstructed coronal maximum intensity-projection (MIP) images (10-mm thickness, 1-mm interval) on the workstation (Advantage Workstation Version 4.6, GE Healthcare) and two board-certified diagnostic radiologists (observer A, with 8 years’ experience; observer B, 18 years’ experience) evaluated them. The images were displayed on a commercially available diagnostic monitor in random order without clinical information or identification of the reconstruction method. Observers were asked to score visualisation of the arteries using a 4-point scale (4, completely visible; 3, partially invisible or slightly irregular appearance; 2, partially visible; 1, invisible), evaluating the main trunk of the superior mesenteric artery (SMA), its branches (right colic, middle colic, ileocolic, jejunal, and ileal arteries), the marginal arteries and vasa recta, main trunk of the celiac artery, gastroduodenal artery (GDA), and arteries of the pancreatic arcade.
Statistical analysis
We analysed statistics using SPSS software (version 25) and tested significant difference in parameters for VMI and conventional images using Mann–Whitney U test.
For subjective image analysis, we added the scores of the two observers, applied the Mann–Whitney U test to assess significant difference between their scores in each category for VMI and conventional CT images. P < 0.05 indicated a statistically significant difference.
To evaluate interobserver agreement, unweighted kappa coefficients were calculated for the scores of subjective analyses of conventional images and VMI images.
Results
Calculation of the modulation transfer function
Figure 2a and b shows the magnified part of the circular edge of the iodine contrast material filling the cylinder in the acrylic phantom, and Fig. 2c shows the results of MTF calculation. Higher MTFs were obtained utilising the conventional 120-kVp scan mode with both the LightSpeed VCT and Discovery CT750 HD imaging units, and the 2 scanners demonstrated no obvious difference in this mode.
Results of quantitative analysis
Table 1 shows the results of quantitative analysis. The conventional and VMI images showed no significant difference in CT attenuation of the aorta, celiac artery, SMA, and psoas major muscle. The VMI images demonstrated significantly less image noise (p < 0.001) and significantly higher CNRs of the aorta (p = 0.011) and celiac artery (p = 0.030) as well as a higher CNR of SMA, though the difference was not significant. In addition, the average CTDI of the VMI (13.40 ± 4.58 mGy) was significantly higher than that of the conventional scan (9.84 ± 4.31 mGy; p = 0.024). The average DLP of the VMI (920.0 ± 358.1 mGy·cm) was also significantly higher than that of the conventional scan (577.7 ± 279.6 mGy·cm; p = 0.006).
Results of subjective evaluation of the proximal and peripheral arteries
Table 2 and Figs. 3 and 4 show the results of evaluations by observers A and B. In most cases, both observers gave a maximum score of 4 points for visualisation of the main trunks of the celiac artery and SMA for almost all cases, and the two scan modes showed no significant difference. However, scores of the GDA, pancreatic arcade, SMA branch, marginal artery, and vasa recta were significantly higher in the conventional scan mode (p < 0.001). The unweighted kappa coefficients were 0.66 for the conventional images and 0.802 for the VMI. Good interobserver agreement was obtained [16]. Figs. 5 and 6 are representative images used in the subjective evaluation.
Discussion
Dual-energy CT has many clinical applications [17]. Virtual monochromatic images are produced by blending and reconstructing image data acquired at two different tube voltages. VMI permits acquisition of images at any energy level and increases the CT value of iodine at lower levels of energy that approach the K-edge energy of iodine (33.2 keV) [7]. In this way, CT attenuation in vessels can be maintained even when we reduce the dose of injected iodine in the acquisition of VMI at low energy levels [8,9,10]. On the other hand, image noise is known to increase as the energy level is decreased in VMI [7, 18]; so, techniques to reduce noise, such as iterative reconstruction, are commonly used in the acquisition of CT angiography with VMI using a reduced iodine dose [8,9,10].
In our study, CT values in the aorta, celiac artery, and SMA were maintained in VMI images despite the reduced dose of injected iodine, and the use of ASiR reduced image noise even at the low keV setting. The high CT values of vessels and reduced image noise led to higher CNRs than those obtained with conventional scanning using a full dose of injected iodine, a finding also reported in other studies [8,9,10].
CNR is commonly used to evaluate image quality in CT because it is easy to measure [19,20,21]. However, though it reflects image quality to some extent, other factors, such as spatial resolution and noise texture, also affect lesion detectability [14, 22]. It is therefore preferable to measure image quality using the method appropriate for given patient in a given situation, in other words, based on task-based analysis [22,23,24,25].
We observed no significant difference in visualisation of relatively large arteries, such as the celiac artery or SMA, but small arteries were much more difficult to detect in VMI acquired using half the conventional dose of iodine. The spatial resolution of the scan technique or reconstruction method is known to affect the visualisation of small vessels [11], and our MTF calculation suggests that spatial resolution is reduced in VMI compared with that of conventional scanning. We therefore attribute the difference in visualisation of relatively small arteries in our study to the reduced spatial resolution of VMI.
In patients with chronic kidney disease, the merit of reducing the injected dose of iodine to prevent contrast-induced nephropathy is tremendous. Thus, acquisition of VMI using a reduced dose of iodine is considered beneficial in the routine examination of these patients, such as for the recurrence or metastasis of disease after surgery. However, in other cases, such as for preoperative examination, we must know the precise condition of small arteries to evaluate tumour invasion or understand the vessel anatomy [1, 26]. In these cases, we have to consider the trade-off between reduction of iodine load and decreased spatial resolution.
In addition, CTDI was significantly higher in the half-iodine VMI group. One possible reason of increased radiation dose is usage of low keV level in VMI. When the dose of iodine contrast material is reduced, we have to use a low keV level to maintain contrast. Image noise is known to increase as keV level is decreased [7]; so this leads to increased radiation dose to maintain acceptable image noise. Another possible reason for increased radiation dose is unavailability of AEC in VMI. In the past study, the reduction of radiation dose by the use of AEC is reported [27]. In fast-kVp switching dual-energy CT, we cannot use AEC with the dual-energy scan mode. This possibly leads to increased radiation dose in VMI. DLP was also significantly different, but this is possibly affected by the difference of scan range between two groups.
This is a retrospective study, so the images of conventional CT angiography were obtained 1.6 years before half-iodine VMI on average. There is a possibility that the diameter of arteries became smaller because of arteriosclerotic change. Although this effect is considered to be minimal because of relatively short time span between these two scans, this is one of limitations of our study.
Another limitation of this study is that we used dual-energy CT units from a single vendor, and we believe comparison between different dual-energy systems is desirable. The decreased spatial resolution of VMI in dual-energy CT may result from misregistration between images acquired using low and high kVp settings. Misregistration is considered to be relatively mild when using a fast kVp switching technique compared to other dual-energy systems, such as the rotate-rotate technique and dual-source system, but theoretically, the dual-layer system has no misregistration [17]. The decreased spatial resolution may also be attributable to the method by which VMI images were reconstructed from two images obtained with different tube voltages. This reconstruction process is considered to be more complicated than that of conventional reconstruction and could lead to the deterioration of image quality, and different dual-energy systems offer different methods of reconstruction.
We also used only one noise reduction technique. In VMI in dual-energy CT, image noise is generally increased as the energy level selected decreases, so noise reduction is needed when the injected iodine volume is reduced and images at low keV levels are reconstructed. Some model-based iterative reconstruction algorithms are reported to improve spatial resolution as well as image noise [28, 29], and their availability for application to images acquired at low keV levels may improve the spatial resolution of VMI in dual-energy CT.
Conclusion
Large-vessel depiction and CNRs were comparable between full-iodine conventional CT and half-iodine VMI images, but VMI did not permit clear visualisation of small arteries and required a larger radiation dose.
Abbreviations
- AEC:
-
Auto-exposure control
- ASiR:
-
Adaptive statistical iterative reconstruction
- CIN:
-
Contrast-induced nephropathy
- CKD:
-
Chronic kidney disease
- CNR:
-
Contrast-to-noise ratio
- CT:
-
Computed tomography
- CTDI:
-
Computed tomography dose index
- DLP:
-
Dose-length product
- GDA:
-
Gastroduodenal artery
- HU:
-
Hounsfield units
- MIP:
-
Maximum intensity projection
- MTF:
-
Modular transfer function
- ROI:
-
Region of interest
- SD:
-
Standard deviation
- SMA:
-
Superior mesenteric artery
- VMI:
-
Virtual monochromatic imaging
References
Hong KC, Freeny PC (1999) Pancreaticoduodenal arcades and dorsal pancreatic artery: comparison of CT angiography with three-dimensional volume rendering, maximum intensity projection, and shaded-surface display. AJR Am J Roentgenol 172:925–931
Winter TC 3rd, Freeny PC, Nghiem HV et al (1995) Hepatic arterial anatomy in transplantation candidates: evaluation with three-dimensional CT arteriography. Radiology 195:363–370
Marti M, Artigas JM, Garzón G, Alvarez-Sala R, Soto JA (2012) Acute lower intestinal bleeding: feasibility and diagnostic performance of CT angiography. Radiology 262:109–116
McCullough PA, Wolyn R, Rocher LL, Levin RN, O’Neill WW (1997) Acute renal failure after coronary intervention: incidence, risk factors, and relationship to mortality. Am J Med 103:368–375
McCullough PA, Choi JP, Feghali GA et al (2016) Contrast-induced acute kidney injury. J Am Coll Cardiol 68:1465–1473
Thomsen HS, Morcos SK (2003) Contrast media and the kidney: European Society of Urogenital Radiology (ESUR) guidelines. Br J Radiol 76:513–518
Matsumoto K, Jinzaki M, Tanami Y, Ueno A, Yamada M, Kuribayashi S (2011) Virtual monochromatic spectral imaging with fast kilovoltage switching: improved image quality as compared with that obtained with conventional 120-kVp CT. Radiology 259:257–262
Yuan R, Shuman WP, Earls JP et al (2012) Reduced iodine load at CT pulmonary angiography with dual-energy monochromatic imaging: comparison with standard CT pulmonary angiography–a prospective randomized trial. Radiology 262:290–297
Lee JW, Lee G, Lee NK et al (2016) Effectiveness of adaptive statistical iterative reconstruction for 64-slice dual-energy computed tomography pulmonary angiography in patients with a reduced iodine load: comparison with standard computed tomography pulmonary angiography. J Comput Assist Tomogr 40:777–783
Ma CL, Chen XX, Lei YX et al (2016) Clinical value of dual-energy spectral imaging with adaptive statistical iterative reconstruction for reducing contrast medium dose in CT portal venography: in comparison with standard 120-kVp imaging protocol. Br J Radiol 89:20151022
Pontone G, Bertella E, Mushtaq S et al (2014) Coronary artery disease: diagnostic accuracy of CT coronary angiography—a comparison of high and standard spatial resolution scanning. Radiology 271:688–694
Hinson JS, Ehmann MR, Fine DM et al (2017) Risk of acute kidney injury after intravenous contrast media administration. Ann Emerg Med 69:577–586
Baer RL (2003) Circular-edge spatial frequency response test. Proc SPIE 5294:71–81. Available via http://kilopixel.net/publications/EI5294-11.pdf. Accessed 5 Aug 2018
Richard S, Husarik DB, Yadava G, Murphy SN, Samei E (2012) Towards task-based assessment of CT performance: system and object MTF across different reconstruction algorithms. Med Phys 39:4115–4122
Samei E, Richard S (2015) Assessment of the dose reduction potential of a model-based iterative reconstruction algorithm using a task-based performance metrology. Med Phys 42:314–323
Kundel HL, Polansky M (2003) Measurement of observer agreement. Radiology 228:303–308
McCollough CH, Leng S, Yu L, Fletcher JG (2015) Dual- and multi-energy CT: principles, technical approaches, and clinical applications. Radiology 276:637–653
Sellerer T, Noël PB, Patino M et al (2018) Dual-energy CT: a phantom comparison of different platforms for abdominal imaging. Eur Radiol 28:2745–2755
Rose A (1948) The sensitivity performance of the human eye on an absolute scale. J Opt Soc Am 38:196–208
Korn A, Fenchel M, Bender B et al (2012) Iterative reconstruction in head CT: image quality of routine and low-dose protocols in comparison with standard filtered back-projection. AJNR Am J Neuroradiol 33:218–224
Gramer BM, Muenzel D, Leber V et al (2012) Impact of iterative reconstruction on CNR and SNR in dynamic myocardial perfusion imaging in an animal model. Eur Radiol 22:2654–2661
Christianson O, Chen JJ, Yang Z et al (2015) An improved index of image quality for task-based performance of CT iterative reconstruction across three commercial implementations. Radiology 275:725–734
Goenka AH, Herts BR, Obuchowski NA et al (2014) Effect of reduced radiation exposure and iterative reconstruction on detection of low-contrast low-attenuation lesions in an anthropomorphic liver phantom: an 18-reader study. Radiology 272:154–163
Jensen K, Andersen HK, Smedby Ö et al (2018) Quantitative measurements versus receiver operating characteristics and visual grading regression in CT images reconstructed with iterative reconstruction: a phantom study. Acad Radiol 25:509–518
Jensen K, Martinsen AC, Tingberg A, Aaløkken TM, Fosse E (2014) Comparing five different iterative reconstruction algorithms for computed tomography in an ROC study. Eur Radiol 24:2989–3002
Sahani D, Mehta A, Blake M, Prasad S, Harris G, Saini S (2004) Preoperative hepatic vascular evaluation with CT and MR angiography: implications for surgery. Radiographics 24:1367–1380
Lechel U, Becker C, Langenfeld-Jäger G, Brix G (2009) Dose reduction by automatic exposure control in multidetector computed tomography: comparison between measurement and calculation. Eur Radiol 19:1027–1034
Katsura M, Sato J, Akahane M, Misi Y, Sumida K, Abe O (2017) Effects of pure and hybrid iterative reconstruction algorithms on high-resolution computed tomography in the evaluation of interstitial lung disease. Eur J Radiol 93:243–251
Nishiyama Y, Tada K, Nishiyama Y et al (2016) Effect of the forward-projected model-based iterative reconstruction solution algorithm on image quality and radiation dose in pediatric cardiac computed tomography. Pediatr Radiol 46:1663–1670
Funding
The authors state that this work has not received any funding.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Guarantor
The scientific guarantor of this publication is Shigeru Suzuki.
Conflict of interest
The authors of this manuscript declare no relationships with any companies, whose products or services may be related to the subject matter of the article.
Statistics and biometry
No complex statistical methods were necessary for this paper.
Informed consent
Written informed consent was waived by the institutional review board.
Ethical approval
Institutional review board approval was obtained.
Methodology
• retrospective
• performed at one institution
Rights and permissions
About this article
Cite this article
Sugawara, H., Suzuki, S., Katada, Y. et al. Comparison of full-iodine conventional CT and half-iodine virtual monochromatic imaging: advantages and disadvantages. Eur Radiol 29, 1400–1407 (2019). https://doi.org/10.1007/s00330-018-5724-4
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00330-018-5724-4