Introduction

Pulmonary embolism is a serious medical condition associated with high mortality, ranging from 3 to 6 % [1, 2]. Computed tomography pulmonary angiography (CTPA) is now widely accepted as the standard test for diagnosing pulmonary embolism (PE) [3], despite its relatively high radiation dose, which varied from 13 to 40 mSv in 2008 [4], but slowly decreased to 3 to 5 mSv nowadays [3].

In order to reduce the radiation dose delivered by CTPA, several strategies have been introduced, including lowering the tube potential [5, 6] and lowering the tube current–time product [7, 8]. However, low dose examinations are associated with increased image noise that can now be reduced by newer reconstructions technologies such as iterative reconstructions [911].

Model-based iterative reconstruction (MBIR, Veo®, GE Healthcare, Milwaukee, Wis) is a novel fully iterative [12] reconstruction algorithm allowing reduction of the dose without loss of image quality [1315]. It has already been demonstrated that chest CT with an x-ray dose equivalent to a posteroanterior and lateral chest x-ray is possible [16].

The primary aim of this study was to demonstrate equivalency (in terms of signal-to-noise ratio (SNR) in the pulmonary arteries) between standard FBP-CTPA and low mA (20 mA) MBIR-CTPA. Subjective image quality was also evaluated.

Materials and methods

This prospective equivalency trial was approved by the institutional review board of the Geneva University Hospital and each patient gave a written informed consent.

Sample size [17]

Necessary sample size was calculated for SNR as the primary outcome. If there is truly no difference between standard FBP-CTPA and low mA (20 mA) MBIR-CTPA, then 138 patients are required in order to be 80 % sure that the limits of a two-sided 90 % confidence interval will exclude a difference in SNR means in the main pulmonary artery of more than 10, assuming a standard deviation of 20.

Patients

From January to May 2013, 167 consecutive patients, scheduled for an enhanced chest CT, were examined on a multidetector CT system (Discovery 750 HD, GE Healthcare, Milwaukee, Wis). Of these 167 consecutive patients, two patients refused to participate and one changed his mind and asked afterwards to be excluded from the study; thus 164 patients were included in the study. In half of the patients (n = 82), a clinical suspicion of PE was the reason for performing CT. These patients were examined with a routine CTPA protocol consisting of intravenous injection of contrast material (60 ml of Iohexol 350, containing 350 mg I/ml, GE Healthcare) at a rate of 3.5 ml/s, followed by a 30-ml saline flush at the same flow rate. The patients were asked to hold their breath after mid-inspiration. A region of interest (ROI) was drawn in the main pulmonary artery and imaging started 5 s after a threshold of 150 HU was reached. The acquisition parameters were as follows: 0.6-s gantry rotation time, 100 kVp, 0.984:1 beam pitch, 40-mm table feed per gantry rotation, a z-axis tube current modulation was used, with a noise index (NI) of 28 (min/max mA, 100/500) and a 64 × 0.625-mm detector configuration. All routine CTPA were reconstructed with the FBP algorithm (FBP-CTPA).

In the other half of the patients (n = 82), the reasons for chest CT were (1) follow-up of a tumour (n = 51), (2) search for complication of pneumonia (n = 16), (3) follow-up of interstitial pneumonia (n = 11), (4) miscellaneous (n = 4). In these patients, an enhanced chest CT was performed 60 s after contrast media administration (60 ml of Iohexol 350, GE Healthcare at a rate of 3.5 ml/s, followed by a 30-ml saline flush at the same flow rate) to answer the clinical questions. A low mA (20 mA) dose chest CT angiography was added for each patient just prior to the enhanced chest CT (an ROI was placed in the main pulmonary artery, imaging started 5 s after a threshold of 150 HU was reached). The patients were also asked to hold their breath after mid-inspiration. In order not to increase the dose received by the patient, the dose of the enhanced chest CT was decreased by 10 % by increasing the NI. Modification of the NI changes the mA (and only the mA) used to acquire the image. As previously published, increasing the NI by one unit will result in approximately 10 % decreased dose to the patient [18]. The x-ray dose saved was hence used to acquire the low mA MBIR-CTPA. The low mA MBIR-CTPA protocol was as follow: 0.6-s gantry rotation time, 100 kVp, 0.984:1 beam pitch, 40-mm table feed per gantry rotation, 20 mA without tube current modulation and a 64 × 0.625-mm detector configuration. All low mA MBIR-CTPA were reconstructed with the model-based iterative reconstruction algorithm.

A known allergy to iodinated contrast media, glomerular filtration rates less than 45 ml min−1 m2, pregnancy and being younger than 18 years old were considered as exclusion criteria.

Parameters of both protocols, as well as patient demographics, are summarized in Table 1.

Table 1 Protocols and demographic

Quantitative image analysis

Axial images were reconstructed at 0.625 mm slice thickness using a soft tissue kernel. All images were evaluated at standard CTPA window settings (window width 350, window level 40).

ROIs were drawn by XM in the (1) main pulmonary artery (MPA), (2) right pulmonary artery (RPA), (3) left pulmonary artery (LPA), (4) right upper lobe artery (RUA), (5) right middle lobe artery (RMA), (6) right lower lobe artery (RLA), (7) left upper lobe artery (LUA), (8) lingular artery (LIA) and (9) left lower lobe artery (LLA). The ROIs were sized to cover almost all the vessel diameters, taking care to avoid artefacts. The mean values (mean attenuation) of these nine ROIs served to calculate the SNR and contrast-to-noise ratio (CNR). Three ROIs were also drawn in each side of the patient (outside the patient) as well as on the anterior part of the patient to assess background noise. The ROIs were sized around 3 cm2 to cover as much air as possible, without including external objects surrounding the patient. The mean values of the three measurements served as background noise (mean background noise). ROIs (2 cm2) were also drawn in the paraspinal and subscapular muscle. The mean values of the two muscles served for contrast-to-noise calculation (mean muscle attenuation):

  • SNR = mean attenuation/mean background noise

  • CNR = (mean attenuation − mean muscle attenuation)/mean background noise

CT dose index volume (CTDIvol) and dose length product (DLP) were recorded for each examination.

The calculation of the effective dose (E) was based on the “European Guidelines on quality criteria on computed tomography” using the following formula: E = E DPL × DLP, where E is the effective dose in mSv and E DPL the dose length coefficient for the chest, which is equal to 0.017 [19].

The size-specific dose estimate (SSDE) was also calculated following the recommendation of American Association of Physicists in Medicine (AAPM) report n°204. Briefly, the anteroposterior (AP) and the lateral diameter of each patient were measured at the level of the tracheal bifurcation and summed. The sum of the lateral and AP dimension allows one to find a conversion factor (in the AAPM report n°204, based on a 32-cm-diameter PMMA phantom), which is used to multiply the CTDIvol reported by the scanner to obtain the SSDE [20, 21].

Subjective image quality

All the images were evaluated independently by two radiologists (XM with 10 years and ALH with 7 years of post-fellowship experience). Images were anonymised, aggregated in folders in a random way and evaluated in a standard CTPA windows setting (window width 350, window level 40). A five-point scale [5] was used to assess the overall image quality, defined as follows: 1, good image quality at the level of the main pulmonary artery (MPA); 2, good image quality at the level of the pulmonary arteries (RPA and LPA); 3, good image quality at the level of the lobar arteries; 4, good image quality at the level of the segmental arteries; 5, good image quality at the level of the subsegmental arteries.

Good image quality was defined as good visual homogeneous enhancement, without respiratory artefacts, allowing one to confidently exclude PE. In case readers did not reach a score of 5, they were asked to specify why image quality was insufficient to confidently exclude PE at a given level. The cause may be breathing artefacts, insufficient attenuation in the pulmonary vessels due to either bad timing of injection or a too high noise level.

Scores 1–3 were considered insufficient for diagnostic purposes, whereas scores of 4–5 were considered sufficient.

Such a score was used because there is still controversy in the literature as to whether subsegmental pulmonary emboli should be treated or not [3].

Ten cases of routine CTPA, not included in the study, were used to train the readers on image quality.

Statistical analysis

Statistical analysis was performed with Prism (Prism, version 6b, 2012; GraphPad Software, San Diego, CA, USA). For continuous values, the results are presented as mean ± standard error.

Non-normally distributed datasets (established from Kolmogorov–Smirnov tests) were compared using Friedman test with Dunn post hoc test. Normally distributed data sets were compared using ANOVA test with Bonferroni post hoc test. Two-sided testing was used. Differences were considered significant at p < 0.05.

Results

Patient demographics

The FBP-CTPA group comprised 46 men and 36 women with a mean age of 64 ± 15 years. The low mA MBIR-CTPA group comprised 49 men and 33 women with a mean age of 60 ± 14 years (Table 1).

Dose estimates

The low mA MBIR-CTPA had a significantly lower CTDIvol (0.59 ± 0.003 vs 8 ± 1.8) and DLP (21 ± 2 vs 282 ± 64) than the routine FBP-CTPA protocols (Table 2). Estimated effective doses were 0.3 ± 0.03 vs 4.1 ± 1.1 mSv, p < 0.0001 (Table 2), whereas SSDE were 0.74 ± 0.14 vs 13.3 ± 3.7 mGy (p < 0.0001) for low mA MBIR-CTPA and routine FBP-CTPA, respectively.

Table 2 Comparison of low mA MBIR-CTPA and routine FBP-CTPA

Quantitative image analysis

The low mA MBIR-CTPA yielded the same attenuation in each of the nine arteries of the lung, a lower noise and a better SNR and CNR, despite a significant x-ray dose reduction, when compared to classical FBP-CTPA (Table 2).

Qualitative image analysis

Table 3 summarizes the image quality of the low mA MBIR-CTPA and of the routine FBP-CTPA protocol for each reader. The inter-reader agreements were excellent for both routine FBP-CTPA and low mA MBIR-CTPA, with kappa values of 0.921 and 0.865, respectively. In the routine FBP-CTPA group, there was only one examination (for reader 1) which was considered as non-diagnostic. This was attributed to a suboptimal injection time. The non-diagnostic examinations in the low mA MBIR-CTPA groups (n = 7 for both readers) were mostly due to patients with a BMI greater than 30 kg/m2 (n = 6) and due to suboptimal injection time in one case.

Table 3 Subjective image quality of low mA MBIR-CTPA and routine FBP-CTPA for each reader

Figure 1 illustrates a typical case of routine FBP-CTPA and low mA MBIR-CTPA for patients with BMI less than 20 kg/m2, BMI between 20 and 30 kg/m2, and BMI greater than 30 kg/m2.

Fig. 1
figure 1

Coronal reformatted thin slice CT (2 mm slice thickness) of routine FBP-CTPA (first row) and of low mA MBIR-CTPA (second row) of patients with a BMI less than 20 kg/m2 (a, d), with a BMI between 20 and 30 kg/m2 (b, e) and with a BMI greater than 30 kg/m2 (c, f). The representative images demonstrated an excellent image quality for all patients with a BMI less than 30 kg/m2

Figure 2 shows the presence of a non-occlusive PE in the right lower lobe artery on a low mA MBIR-CTPA, confirmed 3 days later by a routine FBP-CTPA.

Fig. 2
figure 2

Axial thin slice CT (0.625 mm slice thickness) demonstrating a pulmonary embolism (arrow) in the right lower lobe artery picked up on the low mA MBIR-CTPA (a), confirmed 3 days later (arrow) on a routine FBP-CTPA (b). Please note that the BMI of this patient was 36.8 kg/m2, explaining the low image quality of the low mA MBIR-CTPA. Despite this, the PE was clearly identified on the images

Discussion

Our prospective study shows that low mA MBIR-CTPA allows for a significant x-ray dose reduction of 93 %, while maintaining the same attenuation in each of the nine pulmonary arteries investigated. Moreover, as the noise was significantly lower when using MBIR, the SNR and CNR were significantly higher on the low mA MBIR-CTPA than on the routine FBP-CTPA protocol. In this paper, and despite the very low dose level achieved, we do not use the term “ultra-low dose CTPA” in accordance with the editorial by Bankier and Kressel [22], in which they state that the terms “low”, “ultra-low” and “even super-extra-nano-low-dose” should not be used anymore. Instead, we indicate the DLP, CTDIvol, effective dose and SSDE for objective comparison with other studies.

High sensitivity (83–100 %) and high specificity (89–98 %) have been reported for routine CTPA for diagnosing pulmonary embolism (PE) [2325]. However, because of the relatively low prevalence (9–35 %) of PE among patients evaluated by CTPA, one must be extremely cautious with the dose delivered to patients [8, 26, 27], as it has been estimated that 5 fatalities per 100,000 persons per mSv could be expected [28].

Previous reports investigated the image quality of CTPA after dose reduction due to decrease kVp or mA or both [5, 29, 30] or due to lowering of the z coverage [31, 32]. All these reports showed that PE could be diagnosed with a DLP varying from 40 to 110 mGy cm. Particularly, it was shown that the use of 100 kVp was possible, without significant loss of image quality [5]. Hence, 100 kVp was used in this study for both the routine FBP-CTPA and low mA MBIR-CTPA. Our study uses a lower DLP of around 21 mGy cm. This was made possible by the use of a second generation of iterative reconstruction, the so-called model-based iterative reconstruction. This type of reconstruction has already been applied to unenhanced chest imaging, where a dose of radiation close to the one delivered by a posteroanterior and lateral chest x-rays were used [16]. The MBIR algorithm makes fewer approximations than the FBP algorithm. MBIR takes into consideration a model of noise, a model of the object and a model of the optical properties of the CT. Taking into account all these models is computationally intense and around 40 min is needed for the reconstruction to be done. It is nowadays clearly a limitation, but the time needed for the reconstruction to be done should be significantly reduced with newer and faster calculators.

It has been reported that a diagnostic CTPA examination should have a contrast-enhanced blood attenuation of 200–250 HU with a noise level below 22 HU [28, 33, 34]. Low mA MBIR-CTPA reached both criteria with a mean attenuation of 340 ± 80 HU and a background noise of 6 ± 1. The subjective image quality of all patients with a BMI less than 30 kg/m2 was considered by both readers as diagnostic, except for one patient. The subjective image quality of all patients with a BMI greater than 30 kg/m2 was graded as insufficient for diagnostic purpose (in six patients). This corresponds to 43 % of potentially non-diagnostic CTs in this category of patients. The use of low mA MBIR-CTPA protocols in obese patients (i.e. a BMI greater than 30 kg/m2) should then be strongly discouraged. Future studies should be undertaken to investigate how far the dose could be reduced in accordance with patient BMI.

Study limitation

Two different cohorts of patients were investigated and therefore a direct comparison in the same patients between low mA MBIR-CTPA and routine FBP-CTPA was not possible. It was difficult, from an ethical point of view, to inject each patient twice during the same imaging session, which is why two groups of patients have been compared.

There is only one PE in the low mA MBIR-CTPA protocols (Fig. 2), which is clearly not enough to draw conclusions of the clinical efficacy of this new low dose protocol. Nevertheless, the primary objective of this study was to establish the possibility of acquiring low mA MBIR-CTPA and to assess its image quality. Dose reduction was obtained using reduction of mA without reduction of kVp, leaving the higher attenuation of iodine at low kVp unutilized. We were afraid of having too many artefacts in the shoulder region using low kVp and low mA. This should be studied in the future.

Moreover, as the patients included in the low mA MBIR-CTPA protocol were not suspected of having PE, the probability of finding a PE was low.

Since MBIR is one of the only fully iterative algorithms commercially available [12], it is unclear if our results are applicable to other iterative reconstruction algorithms.

Conclusion

Our prospective study showed that low mA MBIR-CTPA allows a significant dose reduction while improving SNR and CNR in the pulmonary vessels, as compared with routine FBP-CTPA. This first study was a necessary step to control image quality of low mA MBIR-CTPA. The results showed that low mA MBIR-CTPA can now be studied in the clinical context of suspected PE.