Abstract
A method for the simultaneous analysis of nucleosides and nucleotides in infant formula using reversed-phase liquid chromatography–tandem mass spectrometry is described. This approach is advantageous for compliance testing of infant formula over other LC-MS methods in which only nucleotides or nucleosides are measured. Following sample dissolution, protein was removed by centrifugal ultrafiltration. Chromatographic analyses were performed using a C18 stationary phase and gradient elution of an ammonium acetate/bicarbonate buffer, mass spectrometric detection and quantitation by a stable isotope-labelled internal standard technique. A single laboratory validation was performed, with spike recoveries of 80.1–112.9 % and repeatability relative standard deviations of 1.9–7.2 %. Accuracy as bias was demonstrated against reference values for NIST1849a certified reference material. The method has been validated for the analysis of bovine milk-based, soy-based, caprine milk-based and hydrolysed milk protein-based infant formulae.
Similar content being viewed by others
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.Avoid common mistakes on your manuscript.
Introduction
The structure of nucleosides and nucleotides and their importance to infant nutrition have been described previously [1–3]. In view of their physiological benefits, nucleotides are routinely supplemented into infant formulae as sodium salts of adenosine 5′-monophosphate (AMP), cytidine 5′-monophosphate (CMP), guanosine 5′-monophosphate (GMP), inosine 5′-monophosphate (IMP) and uridine 5′-monophosphate (UMP) [4]. Although nucleosides are not supplemented into infant formulae, dephosphorylation of nucleotides to the corresponding nucleosides—adenosine (Ado), cytidine (Cyd), guanosine (Guo), inosine (Ino) and uridine (Urd)—can occur under certain processing conditions [5].
Analytical methods for nucleosides and nucleotides in infant formulae and milk have previously been reviewed [6]. Sample preparation of infant formulae is frequently achieved by acidic precipitation of casein proteins from the reconstituted sample [5, 7]. Alternatively, centrifugal ultrafiltration has also been reported [8] and offers a simple mechanism to remove high-molecular-weight material. Further cleanup of sample extracts using ion exchange solid phase extraction and a phenylboronate affinity gel has been reported [9–11].
Liquid chromatography, i.e. reversed-phase liquid chromatography (RPLC), ion pair RPLC, ion exchange liquid chromatography and hydrophilic interaction liquid chromatography, with UV detection is commonly used for the quantitation of nucleotides in milk products [5, 7, 8, 12–15]. RPLC is easily adapted for the analysis of nucleosides, although the retention of nucleotides is more challenging. However, at the appropriate mobile phase pH, polar nucleotides are retained on a C18 column and an organic solvent gradient is able to remove late-eluting nucleosides.
The use of mass spectrometry (MS) offers potential advantages with respect to accuracy and simplicity by incorporating the addition of stable isotope-labelled (SIL) internal standards, whilst the selectivity of tandem MS reduces the need to remove other components that often compromise UV analyses [16–18]. The aim of this study was, therefore, to develop an LC-MS/MS method for the simultaneous analysis of nucleosides and nucleotides in infant formulae. The method described involves a simple centrifugal ultrafiltration procedure followed by high-performance liquid chromatography (HPLC) with detection and quantitation by tandem MS. Confidence in analytical accuracy is assured through the use of a SIL standard for each analyte. This technique has been validated for a range of bovine milk-based, caprine milk-based, soy-based and hypoallergenic infant formulae.
Experimental
Reagents
Ammonium acetate (NH4CH3COO), ammonium bicarbonate (NH4HCO3), AMP sodium salt, CMP disodium salt, GMP disodium salt, IMP disodium salt, UMP disodium salt, and corresponding nucleosides were obtained from Sigma-Aldrich (St. Louis, MO, USA). SIL nucleoside standards—13C5 Ado, 13C9 15N3 Cyd, 15N5 Guo, 15N4 Ino and 13C9 15N2 Urd—were purchased from Cambridge Isotope Laboratories (Andover, MA, USA). SIL nucleotide standards—13C10 15N5 AMP, 13C9 15N3 CMP, 13C10 15N5 GMP and 13C9 15N2 UMP—were purchased from Sigma-Aldrich. 13C10 15N4 IMP was purchased from Medical Isotopes (Pelham, NH, USA).
Potassium dihydrogen phosphate (KH2PO4), acetic acid, orthophosphoric acid, potassium hydroxide and methanol were supplied by Merck. Water was purified with resistivity ≥18 M Ω using an E-pure water system (Barnstead, Dubuque, IA, USA).
A standardising buffer (KH2PO4, 0.25 M, pH 3.5) was made by dissolving 34.02 g of KH2PO4 in 900 mL of water, adjusting the pH to 3.5 with orthophosphoric acid and then making the solution to 1 L. Mobile phase A (NH4CH3COO, 10 mM; NH4HCO3, 5 mM, pH 5.6) was made daily by dissolving 0.771 g of NH4CH3COO and 0.395 g of NH4HCO3 in 950 mL of water, adjusting the pH to 5.6 with acetic acid solution (10 %, w/v) and then making to 1 L with water. Mobile phase B consisted of 100 % methanol.
Standard solutions
SIL nucleoside and nucleotide stock standards were prepared by accurately weighing 50 mg each of 13C5 Ado, 13C9 15N3 Cyd, 15N5 Guo, 15N4 Ino, 13C9 15N2 Urd, 13C10 15N5 AMP, 13C9 15N3 CMP, 13C10 15N5 GMP, 13C10 15N4 IMP and 13C9 15N2 UMP into separate 50-mL volumetric flasks. To each flask, 40 mL of water was added and then shaken (with gentle warming if necessary) until the standard was completely dissolved before water was added to volume. Aliquots (∼1.5 mL) of SIL stock standards were immediately dispensed into individual cryogenic vials and frozen at −15 °C for later use. Prior to analysis, cryogenic vials containing each SIL nucleoside and nucleotide stock standard were allowed to thaw to room temperature.
Non-isotopically labelled (NL) nucleoside and nucleotide stock standards were prepared in a similar manner by accurately weighing approximately 50 mg of each into separate 50-mL volumetric flasks and making to volume with water. These were refrigerated at 4 °C for up to 1 month.
Estimation of the moisture content of nucleosides was performed using the oven moisture method (102 ± 2 °C, 4 h) and the concentration was calculated on a dry weight basis. Extinction coefficients at UV absorbance maxima were then determined for each nucleoside. These were compared with the values previously determined for nucleotides [5], with correction for molecular weight. The values obtained for each nucleoside were in close agreement with those for the corresponding nucleotide. Mean extinction coefficient values (nucleoside and corresponding nucleotide) were calculated by adjusting for molecular weight and are reported in Table 1. The concentration of each nucleoside and nucleotide stock standard was determined by adding 500 μL of each stock standard into separate 25-mL volumetric flasks, diluting with standardising buffer and measuring the absorbance at the appropriate λ max.
A mixed SIL intermediate standard was prepared by diluting 2.0 mL of each SIL stock standard into a 25-mL volumetric flask and making to volume with water. A mixed NL intermediate standard was made by adding 1.0 mL of each NL stock standard into a 25-mL volumetric flask and making to volume with water.
Four calibration standards were prepared by pipetting 1.0, 1.0, 0.5 and 0.2 mL of SIL intermediate standard and 2.0, 4.0, 5.0 and 8.0 mL of NL intermediate standard into 50-, 50-, 25- and 10-mL volumetric flasks, respectively. The calibration standards were then made to volume with water and mixed thoroughly.
Samples
A range of different infant formula samples were evaluated during the validation of the method. These included a partially hydrolysed bovine milk-based powder, a partially hydrolysed soy-based powder, an infant elemental powder, a bovine milk-based powder, a soy-based powder and a caprine milk-based powder.
Sample preparation
Approximately 5.0 g of infant formula powder was weighed accurately into a 50-mL polypropylene centrifuge tube (Biolab, Auckland, New Zealand) and dissolved in 25 mL of water. To this was added 1.0 mL of the SIL intermediate standard and the tube was capped and vortex-mixed. The sample was allowed to stand for 10 min to ensure complete hydration before dilution to a final volume of 50 mL with water.
A 4.0-mL aliquot of sample solution was added to an Amicon Ultra-4 3000 MWCO centrifugal ultrafiltration unit (Millipore, Billerica, MA, USA) and centrifuged at 3,500×g for 60 min. The filter was then removed and discarded and a 1-mL aliquot of filtrate was transferred to an HPLC vial ready for analysis.
Instrumentation
The HPLC system used consisted of a CBM20A system controller, two LC20ADXR pumps for high-pressure gradients, a CTO20AC column oven and a SIL20ACXR autosampler equipped with a 50-μL injection loop (Shimadzu, Kyoto, Japan). Chromatographic separation was achieved using a Gemini column, 5 μm, 4.6 × 250 mm (Phenomenex, Torrance, CA, USA), with a high-pressure gradient elution programme as described in Table 2.
The MS/MS system consisted of a 3200 QTRAP quadrupole mass spectrometer with a Turbo V ion source equipped with an electrospray ionisation (ESI) probe. Analyst 1.5.1 software was used for instrument control and data processing (ABSciex, Foster City, CA, USA). The mass spectrometer was operated in ESI+ mode with nitrogen utilised as the drying and collision gas. The instrumental parameters were set as follows: curtain gas at 30 psi, nebuliser gases GS1 and GS2 at 50 and 70 psi, respectively, desolvation temperature at 700 °C, collision-induced dissociation gas at medium and ion spray voltage at 5,500 V. Instrument settings and multiple reaction monitoring (MRM) transitions for the generation of product ions for nucleosides and nucleotides are given in Table 3.
Method validation
Six mixed nucleoside and nucleotide solutions covering the expected working range were prepared and linearity was evaluated by least-squares regression analysis of the SIL-corrected response (ratio of NL/SIL analyte peak area versus ratio of NL/SIL analyte concentration). A minimum value of 0.997 for the correlation coefficient (r 2) was deemed to be suitable. Plots of standard residuals were visually assessed as a further test of linearity.
Repeatability was estimated by analysing replicate pairs (n = 9 pairs) of a bovine milk-based infant formula and NIST 1849a. Intermediate precision was determined from replicate analyses (n = 6) of a bovine milk-based infant formula and NIST 1849a tested on three different days. Method detection limits (MDLs) were estimated in accordance with US Environmental Protection Agency procedures [19].
The robustness of the method was assessed by conducting a Plackett–Burman trial [20], as described previously [15]. The seven factors assessed were: initial sample water volume (27 and 23 mL), vortex time (40 and 20 s), wait time (14 and 6 min), centrifuge volume (4.2 and 3.8 mL), centrifuge speed (4,000 and 3,000×g), centrifuge time (70 and 50 min) and a dummy factor.
Method accuracy was determined as both recovery and bias. Recovery of both nucleosides and nucleotides was evaluated by spiking a range of sample matrices at 50 and 150 % of the concentration levels typically found in infant formulae. Bias was evaluated by performing a paired t test for nucleotides both against reference values of a NIST 1849a powder and against values for a bovine milk-based infant formula tested using AOAC Official Method 2011.20 [21].
Results and discussion
Method optimisation
The simultaneous chromatographic analysis of both nucleosides and nucleotides in infant formulae using LC-UV has previously been described [5]. However, the mobile phase contained a 0.1 M phosphate buffer, which is unsuitable for use in LC-MS. In this study, ammonium acetate (10 mM, pH 5.6) was initially chosen to buffer the mobile phase because of its compatibility with electrospray mass spectrometric detection and a pH buffering range (∼3.8–5.8) consistent with nucleoside and nucleotide pK a values. However, significant peak tailing for nucleotides was observed when this buffer was used.
Conventional LC-UV nucleotide analyses typically contain phosphate in the mobile phase and no significant peak tailing is observed [5, 7]. Unfortunately, the use of non-volatile buffers such as phosphate in LC-MS is generally not recommended because of contamination of the ion source leading to a decrease in sensitivity. Furthermore, the interaction of phosphorylated compounds with metal surfaces in liquid chromatographic applications resulting in peak tailing has been reported [22–26]. Pretreatment of the chromatographic system using phosphoric acid prior to switching to a non-phosphate eluent during analysis [26, 27], the substitution of polyether ether ketone tubing for stainless steel, the use of a high pH mobile phase [28] and the addition of EDTA to the mobile phase [25] have all been employed to overcome this problem.
A number of mass spectrometer manufacturers have evaluated phosphate buffers for use with their instruments and have demonstrated that modern source designs can tolerate the use of non-volatile buffers [29–32]. A phosphate-based ion pair RPLC-MS method was successfully applied to the quantitative analysis of intracellular nucleotides utilising a microbore column to reduce the amount of phosphate introduced to the ion source [33].
In the present study, a low ionic strength phosphate buffer (NH4H2PO4 0.08 mM, pH 5.6) was initially evaluated for compatibility with the mass spectrometer. The chromatographic parameter resolution, retention factor, peak area repeatability, retention time repeatability, plate number and asymmetry were evaluated, with acceptable results being obtained (data not shown). There was some loss of sensitivity as replicate analyses progressed, consistent with a small accumulation of phosphate on the cone. The method was applied to the analysis of nucleotides in infant formula samples in a validation study. Linear response was demonstrated for NL/SIL peak area versus NL/SIL analyte concentration (r 2 = 0.997–0.999). Accuracy and precision were evaluated, with both spike recovery (84.2–107.1 %) and repeatability relative standard deviation (1.5–3.1 %) deemed to be acceptable. Despite this performance, a limitation with this phosphate-based approach was that the number of samples within each analytical run was limited due to the requirement for regular maintenance of the source.
An alternative chromatographic system was evaluated based on the observations of Asakawa et al. [22], who found a beneficial chromatographic effect with a number of mobile phase additives. Of those evaluated, only ammonium bicarbonate is volatile and considered suitable for use in LC-MS and was therefore incorporated as an additive in the ammonium acetate eluent.
The optimisation of the MS conditions was performed by infusion of a standard of each nucleoside or nucleotide (∼10 μg mL−1) diluted in a mixture of mobile phases A and B (90:10). Initial development focused on ESI+ for nucleosides and ESI− for nucleotides. However, it was found that ESI+ gave superior response for both analytes, with the [M+H]+ ion most abundant and low levels of potassium adduct, thereby simplifying the analysis with the detection of all analytes in positive mode.
The conditions for MRM were optimised by selecting individual fragments and adjusting collision energies to maximise the product ion signal. The most abundant fragment ion observed for nucleosides and nucleotides was formed by cleaving of the glycosidic bond, resulting in the loss of ribose or ribose + phosphate group and the detection of the positively charged nucleobase. The exception to this was UMP, which underwent fragmentation and rearrangement to generate the m/z 97.0 ion. A similar fragmentation scheme has been reported for the generation of a product ion with m/z 81.0 from the fragmentation of deoxycytidine 5′-monophosphate [34].
Using the LC-MS/MS method developed, the simultaneous detection of nucleosides and nucleotides in a standard solution was achieved (Fig. 1).
Method performance
A high degree of selectivity is afforded by an MRM experiment; however, chromatographic separation is required for critical peaks with similar MRM transitions if accurate quantitation is to be achieved. Chromatographic performance was assessed by replicate analyses (n = 6) of a mixed nucleoside and nucleotide standard, with satisfactory resolution being obtained between IMP/AMP (6.7), Ino/Ado (6.8) and Cyd/Urd (4.3) critical pairs, compounds which differ in mass by <2 Da (Table 4).
Method validation experiments to determine linearity, detection limits and precision are summarised in Table 5. Linearity was evaluated by least-squares regression analysis, with acceptable values being obtained for the correlation coefficient and with standard residual plots showing no pattern and only a small amount of random variation. The detection limits were appropriate, as defined by the infant formula industry, with the exception of those for CMP and Urd [35]. Although the detection limits of CMP and Urd were higher than those specified, the MDL was two orders of magnitude lower than that found in unfortified milk powder [5]. Precision was evaluated as repeatability (1.9–7.2 %) and intermediate precision (2.9–14.4 %). The suitability of these results was demonstrated by a Horwitz (repeatability) ratio of 0.2–0.6 [36].
The method was found to be robust for the seven method performance parameters studied, with variances in the results obtained not being significantly different from those expected by chance. Given the method’s simplicity, two critical steps are required to ensure the accuracy of the results obtained: accurate measurement of the amount of sample weighed and accurate addition of the internal standard.
Accuracy determined as spiked recovery results measured in the six different product types were within the acceptable limits of 80–115 % at the 10-μg g−1 level, as suggested by the AOAC [36] (Table 6). Accuracy estimated as bias was evaluated against reference values for NIST 1849a CRM (Table 7) and against an LC-UV method for determining nucleotides in infant formula (AOAC method 2011.20; Table 8). Although there were statistically significant differences for some of the results, the differences were small enough (0–13 %) that they are unlikely to be of practical significance for compliance and labelling requirements.
Conclusions
The optimisation and validation of an LC-MS/MS method for the analysis of nucleosides and nucleotides in infant formulae has been described. The use of SIL internal standards provides confidence in the accuracy of the results obtained. The method has been demonstrated to be precise and accurate and has been validated for the analysis of bovine milk-based, soy-based, caprine milk-based and hydrolysed milk protein-based infant formulae.
References
Hess JR, Greenberg NA (2012) The role of nucleotides in the immune and gastrointestinal systems: potential clinical applications. Nutr Clin Pract 27:281–294
Cosgrove M (1998) Perinatal and infant nutrition. Nucleotides Nutrition 14:748–752
Carver JD, Walker WA (1995) The role of nucleotides in human nutrition. J Nutr Biochem 6:58–72
Commission of the European Communities (1991) Second addendum to the reports of the Scientific Committee for Food concerning the essential requirements of infant formulae and follow-up milks based on cow’s milk proteins and the minimum requirements for soya-based infant formulae and follow-up milks. http://ec.europa.eu/food/fs/sc/scf/reports/scf_reports_28.pdf. Accessed 16 June 2010
Gill BD, Indyk HE (2007) Development and application of a liquid chromatographic method for analysis of nucleotides and nucleosides in milk and infant formulas. Int Dairy J 17:596–605
Gill BD, Indyk HE (2007) Determination of nucleotides and nucleosides in milks and pediatric formulas: a review. J AOAC Int 90:1354–1364
Perrin C, Meyer L, Mujahid C, Blake CJ (2001) The analysis of 5′-mononucleotides in infant formulae by HPLC. Food Chem 74:245–253
Inoue K, Obara R, Hino T, Oka H (2010) Development and application of an HILIC-MS/MS method for the quantitation of nucleotides in infant formula. J Agric Food Chem 58:9918–9924
Gill BD, Indyk HE, Manley-Harris M (2012) Determination of total potentially available nucleosides in bovine, caprine, and ovine milk. Int Dairy J 24:40–43
Leach JL, Baxter JH, Molitor BE, Ramstack MB, Masor ML (1995) Total potentially available nucleosides of human milk by stage of lactation. Am J Clin Nutr 61:1224–1230
Gill BD, Indyk HE, Manley-Harris M (2011) Determination of total potentially available nucleosides in bovine milk. Int Dairy J 21:34–41
Inoue K, Obara R, Akiba T, Hino T, Oka H (2008) Determination of nucleotides in infant formula by ion-exchange liquid chromatography. J Agric Food Chem 56:6863–6867
Ferreira IMPLVO, Mendes E, Gomes AMP, Faria MA, Ferreira MA (2001) The determination and distribution of nucleotides in dairy products using HPLC and diode array detection. Food Chem 74:239–244
Oliveira C, Ferreira IMPLVO, Mendes E, Ferreira M (1999) Development and application of an HPLC/diode array methodology for determination of nucleotides in infant formulae and follow-up milks. J Liq Chromatogr Relat Technol 22:571–578
Gill BD, Indyk HE, Kumar MC, Sievwright NK, Manley-Harris M (2010) A liquid chromatographic method for routine analysis of 5′-mononucleotides in pediatric formulas. J AOAC Int 93:966–973
Ren Y, Zhang J, Song X, Chen X, Li D (2011) Simultaneous determination of 5′-monophosphate nucleotides in infant formulas by HPLC-MS. J Chromatogr Sci 49(4):332–337
Viñas P, Campillo N, Melgarejo GF, Vasallo MI, López-García I, Hernández-Córdoba M (2010) Ion-pair high-performance liquid chromatography with diode array detection coupled to dual electrospray atmospheric pressure chemical ionization time-of-flight mass spectrometry for the determination of nucleotides in baby foods. J Chromatogr A 1217(32):5197–5203
Inoue K, Dowell D (2012) HILIC-MS/MS method for the quantitation of nucleotides in infant formula and adult nutritional formula: First Action 2011.21. J AOAC Int 95(3):603–605
EPA (1999) Guidelines establishing test procedures for the analysis of pollutants (40 CFR Appendix B to Part 136, Definition and procedure for the determination of the method detection limit). Environmental Protection Agency, Washington, DC
Youden WJ, Steiner EH (1975) Statistical manual of the AOAC. AOAC International, Arlington, VA
Gill BD, Indyk HE, Kumar MC, Sievwright NK, Manley-Harris M, Dowell D (2012) Analysis of 5′-mononucleotides in infant formula and adult/pediatric nutritional formula by liquid chromatography: First Action 2011.20. J AOAC Int 95:599–602
Asakawa Y, Tokida N, Ozawa C, Ishiba M, Tagaya O, Asakawa N (2008) Suppression effects of carbonate on the interaction between stainless steel and phosphate groups of phosphate compounds in high-performance liquid chromatography and electrospray ionization mass spectrometry. J Chromatogr A 1198–1199:80–86
De Vijlder T, Boschmans J, Witters E, Lemière F (2011) Study on the loss of nucleoside mono-, di- and triphosphates and phosphorylated peptides to a metal-free LC-MS hardware. Int J Mass Spectrom 304:83–90
Kim J, Camp DG, Smith RD (2004) Improved detection of multi-phosphorylated peptides in the presence of phosphoric acid in liquid chromatography/mass spectrometry. J Mass Spectrom 39:208–215
Liu S, Zhang C, Campbell JL, Zhang H, Yeung KKC, Han VKM, Lajoie GA (2005) Formation of phosphopeptide–metal ion complexes in liquid chromatography/electrospray mass spectrometry and their influence on phosphopeptide detection. Rapid Commun Mass Spectrom 19:2747–2756
Wakamatsu A, Morimoto K, Shimizu M, Kudoh S (2005) A severe peak tailing of phosphate compounds caused by interaction with stainless steel used for liquid chromatography and electrospray mass spectrometry. J Sep Sci 28:1823–1830
Yamaoka N, Kudo Y, Inazawa K, Inagawa S, Yasuda M, Mawatari K, Nakagomi K, Kaneko K (2010) Simultaneous determination of nucleosides and nucleotides in dietary foods and beverages using ion-pairing liquid chromatography–electrospray ionization–mass spectrometry. J Chromatogr B 878:2054–2060
Tuytten R, Lemière F, Witters E, Van Dongen W, Slegers H, Newton RP, Van Onckelen H, Esmans EL (2006) Stainless steel electrospray probe: a dead end for phosphorylated organic compounds? J Chromatogr A 1104:209–221
Applied Biosystems (2006) Degradation and impurity product profiling with LC/MS/MS: the use of phosphate-containing mobile phases and fast data analysis using LightSight™ software. http://www3.appliedbiosystems.com/cms/groups/psm_marketing/documents/generaldocuments/cms_042526.pdf. Accessed 10 January 2013
Dionex Corporation (2001) Extending the performance of LC/MS to nonvolatile buffers with the AQA. http://www.dionex.com/en-us/webdocs/4740-TN519_LPN1274.pdf. Accessed 10 January 2013
Agilent Technologies (1998) The effect of nonvolatile buffers on the Agilent 1100 series LC/MSD. http://www.chem.agilent.com/Library/applications/5968-2690_016655.pdf. Accessed 10 January 2013
Waters Corporation (1998) A phosphate buffer study: performance comparison of various LC-MS API interfaces using phosphate buffer containing mobile phase. http://www.waters.com/waters/library.htm?cid=511436&lid=1526095. Accessed 10 January 2013
St Claire RL (2000) Positive ion electrospray ionization tandem mass spectrometry coupled to ion-pairing high-performance liquid chromatography with a phosphate buffer for the quantitative analysis of intracellular nucleotides. Rapid Commun Mass Spectrom 14:1625–1634
Curtis M, Minier MA, Chitranshi P, Sparkman OD, Jones PR, Xue L (2010) Direct analysis in real time (DART) mass spectrometry of nucleotides and nucleosides: elucidation of a novel fragment [C5H5O]+ and its in-source adducts. J Am Soc Mass Spectrom 21:1371–1381
Sullivan D (2012) Infant formula and adult/pediatric nutritional methods approved First Action using the AOAC voluntary consensus standards process. J AOAC Int 95:287–290
Horwitz W (2002) AOAC requirements for single laboratory validation of chemical methods. Draft 2002-11-07. AOAC International, Gaithersburg, MD
Acknowledgments
The financial assistance of Fonterra Co-operative Group Limited and the Tertiary Education Commission in providing an Enterprise Scholarship is gratefully acknowledged.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Gill, B.D., Indyk, H.E. & Manley-Harris, M. Analysis of nucleosides and nucleotides in infant formula by liquid chromatography–tandem mass spectrometry. Anal Bioanal Chem 405, 5311–5319 (2013). https://doi.org/10.1007/s00216-013-6935-9
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00216-013-6935-9