Summary
The position of the patellar component and patellofemoral symptoms were assessed in a series of 72 cemented polyethylene patellar components in total condylar arthroplasties. A small (25 mm) component was used. The minimum follow up was 2 years and maximum 3 years. A lateral retinacular release was carried out in 22 knees. Clinical results were assessed by questionnaire and 86% had no patellofemoral symptoms. The mean postoperative patellar height was not changed compared to the preoperative height. Lateral placement of the patellar component was associated with both increased medial tilt and lateral subluxation. The component should be inserted so that its centre is slightly medial to the centre of the bone.
Résumé
Etude de la fonction fémoro-patellaire et de la position de la prothèse rotulienne dans une série de 72 prothèses totales de genou à pièce rotulienne en polyéthylène cimentée (prothèse sans conservation des croisés de Johnson and Johnson). Dans la plupart des cas, un petit composant (25 mm) a été utilisé. Le minimum de suivi est 2 ans. Une section de l'aileron rotulien externe a été faite pour 22 genoux (30%). Les résultats fonctionnels ont été étudiés au moyen d'un questionnaire. 86% des patients ont une articulation fémoropatellaire asymptomatique. L'épaisseur moyenne de la rotule n'a pas été modifiée de façon significative par l'intervention. L'inclinaison frontale moyenne du composant patellaire est de moins 2°, l'inclinaison de la coupe osseuse est de moins de 5° et l'inclinaison de l'ensemble osprothèse est en moyenne de 3°. La position latérale du composant patellaire s'accompagne d'une façon significative d'une augmentation de l'inclinaison patellaire interne et d'un déplacement externe de la rotule. La prothèse patellaire doit donc être placée avec son centre légèrement plus médial que le centre de la rotule. L'utilisation d'une prothèse patellaire de petite taille n'est pas un facteur de douleur fémoro-patellaire.
Article PDF
Similar content being viewed by others
Avoid common mistakes on your manuscript.
References
Aglietti P, Insall JN, Walker PS (1975) A new patellar prosthesis. Clin Orthop 107: 175
Bartel DL, Bicknell VL, Wright TM (1986) The effect of conformity, thickness, and material on stresses in ultrahigh molecular weight components for total joint replacement. J Bone Joint Surg [Am] 68: 1041–1051
Boyd AD, Ewald FC, Thomas WH, Poss R, Sledge CB (1993) Long-term complications after total knee arthroplasty with or without resurfacing of the patella. J Bone Joint Surg [Am] 75: 674–681
Cameron HU, Cameron GM (1987) The patellar meniscus in total knee arthroplasty. Orthop Rev 26: 170–175
Gomes LSM, Bechthold JE, Gustillo RB (1988) Patellar prosthesis positioning in total knee arthroplasty. Clin Orthop 236: 72–80
Goodfellow J, Hungerford DS, Zindel M (1976) Patellofemoral joint mechanics and pathology. J Bone Joint Surg [Br] 58: 287
Insall JN (1993) Surgery of the knee, Churchill Livingstone New York
Marmor L (1988) Technique for patellar resurfacing in total knee arthroplasty. Clin Orthop 230: 166–167
McNamara JL, Collier JP, Mayor MB, Jensen RE. A comparison of contact pressures in tibial and patellar total knee components before and after service in vivo. Clin Orthop 299: 104-113
Merchant AC, Mercer RL, Jacobsen RH (1974) Roentgenographic analysis of patello-femoral congruence. J Bone Joint Surg [Am] 56: 1391–1996
Ranawat CS (1986) The patellofemoral joint in total condylar knee arthroplasty. Clin Orthop 205: 93–98
Scott WN, Rubinstein M (1986) Posterior stabilized knee arthroplasty: six year experience. Clin Orthop 205: 138–146
Scuderi GR, Insall JN (1992) Total knee arthroplasty. Current clinical perspectives. Clin Orthop 276: 26–32
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Nelissen, R.G.H.H., Weidenheim, L. & Mikhail, W.E.M. The influence of the position of the patellar component on tracking in total knee arthroplasty. International Orthopaedics 19, 224–228 (1995). https://doi.org/10.1007/BF00185227
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF00185227