Computational Forward Models in Transcranial Direct Current Stimulation

Introduction to Dose Definition and Selection

Dose, as defined in the context of transcranial stimulation with magnetic or electrical fields (E-fields), includes all the controllable parameters of the stimulation device that affect the electromagnetic field induced in the body (Peterchev et al. 2012). In the case of tDCS, this includes parameters such as the size, geometry, position, orientation and number of electrodes, the current intensity and polarity in each electrode, the duration of the applied current and the duration of the ramp-up/down period. Other parameters related to the experimental protocol and the skin preparation techniques used can also be included in this definition of dose to the extent they influence electromagnetic fields in the body – though they are always important in the broader content of protocol reproducibility. Replicating all of these dose parameters across subjects does not guarantee, however, that the subjects’ response will be the same. This has become increasingly evident by studies indicating that the responses to tDCS can vary substantially across subjects within the same protocol (Lopez-Alonso et al. 2014; Wiethoff et al. 2014). One cause for this difference is the fact that the E-field distribution in the head during tDCS can be substantially different across subjects due to individual differences in head geometry. Since direct in vivo measurement of the E-field distribution during tDCS is not possible, except in special cases where implanted electrodes are present (Datta et al. 2016; Dymond et al. 1975; Opitz et al. 2016), computational models remain the only practical tool available to predict E-field in the brain for a given tDCS dose. Information from these models can be used to adjust dose parameters to induce comparable E-field across subjects, or it can be used to optimize these parameters to maximize the effects in a specific target cortical region (Dmochowski et al. 2011; Ruffini et al. 2014).

Methods for Generation of Computational Models

The first models to predict the E-field in brain stimulation techniques relied on analytical solutions of the underlying physical equations (Eaton 1992; Rush and Driscoll 1968). The complexity of such approaches often demanded drastic simplifications at the level of head geometry, number of tissues involved and electrode/coil geometry. Numerical computational approaches were soon identified as promising alternatives, but the limited computational resources available at first imposed similar limitations to the models (Roth et al. 1991; Tofts 1990). As such, many of the first computational studies of E-field distribution induced in tDCS adopted spherical head models (Miranda et al. 2006) or computer-aided design (CAD) generated simplified geometries (Wagner et al. 2007). The advent of powerful computational resources has made it possible to build increasingly realistic head and electrode models (Datta et al. 2009a; Oostendorp et al. 2008).

Table 9.1 summarizes several published studies that have employed computational models of tDCS. Most of the recent studies have used realistic head models, and they employ similar techniques to generate such models. A generalized pipeline to create computational models is shown in Fig. 9.1. It should be mentioned that although the studies usually employ the same steps as those highlighted in the figure, the methods employed at each step can be substantially different across studies. The description of the studies presented in Table 9.1 shows the wide variety of electrode configurations and geometries that has been studied (see columns ‘Electrodes’ and ‘Montages’). It also shows that the studies can be used in a variety of applications, such as investigating the fundamental properties of the induced E-field, studying the distribution of the E-field in montages typically used in clinical settings, optimizing montages to target specific cortical areas and/or studying the properties of the E-field induced in susceptible populations (more on this in section “Use of Computational Models in Clinical Practice”).

Table 9.1 List of all studies involving computational models in tDCS
Fig. 9.1
figure 1

Modeling steps involved in the creation of a computational high-resolution model of the E-field induced in tDCS

The majority of studies shown in Table 9.1 employ a numerical technique known as the finite element method (FEM) to solve the governing physical equations that determine the E-field induced in tDCS (i.e., Laplace’s equation) (Johnson 1997). The FEM yields an approximate numerical solution to a specific equation (or set of equations) that specifies a physical problem in a geometry. In the FEM, the geometry is subdivided into a number of finite elements of a specific shape that are connected by nodes. The latter comprise the finite element’s mesh. Within each element of the mesh, the solution to the equation can be written in terms of an interpolation function whose parameters depend on the values of the dependent variable in the nodes of each element. In the case of tDCS, the dependent variable is V, the electrostatic potential, and its gradient yields the electric field vector, E. By approximating the solution to the equation with an interpolation function in every element of the mesh, the original equation can be rewritten as a linear equation of the form AV = b, where A is a matrix (whose values depend on the geometry of the mesh and the type of elements used), V is the column vector with the values of the dependent variable (electrostatic potential) in each node of the mesh and b is another column vector whose values depend on the boundary conditions of the problem (for instance the current injected at each electrode). In order to obtain the column vector V, it is necessary to invert matrix A, which can be performed using a number of numerical algorithms.

The first step for creating computational tDCS models is the specification of the geometry of the volume conductor, i.e., the head and electrodes. In most recent studies this is achieved by segmentation of structural magnetic resonance (MR) images, thus obtaining masks, i.e., the set of voxels labeled as belonging to a specific tissue. The choice of MR over other image modalities, such as computed tomography (CT), is mainly related to the fact that it does not involve ionizing radiation. It also offers a higher contrast between soft tissues, such as grey matter (GM) and white matter (WM) and the possibility for molecular phenomena to be observed (e.g. diffusion weighted imaging) (Bashir et al. 2015). MR also has some limitations, however, like the fact that the skull emits a very weak MR signal, therefore making its reconstruction difficult (Windhoff et al. 2013). Most studies use T1-weighted MR images with an isotropic resolution of 1 mm (lower resolutions are sometimes used), although other studies have employed T2 and even PD-weighted images to allow for a better segmentation of the CSF and skull (Miranda et al. 2013; Windhoff et al. 2013). A variety of algorithms and software are described in the literature in order to perform the automatic segmentation of the images. However, manual inspection of the resulting masks and manual segmentation of some structures is often reported as well. The majority of studies produce segmentation masks for the skin, skull, CSF compartments, GM and WM, but others include many more tissues such as muscle, subdermal fat, and eye sclera (Shahid et al. 2013). A number of studies segment the skull into compact and spongy bone (Laakso et al. 2015; Opitz et al. 2015; Rampersad et al. 2012; Saturnino et al. 2015; Wagner et al. 2014a, b). MR images do not usually allow for these two tissues to be segmented, so manual methods (Opitz et al. 2015) or custom MR sequences (Rampersad et al. 2014) are employed.

Generation of the finite element mesh from the segmentation masks is performed using several possible software and algorithms. Three aspects related to the mesh are noteworthy. The first one is its resolution, i.e. the mean distance between the nodes in the mesh. The mesh must have sufficient resolution to accurately predict the spatial distribution of the E-field. The second aspect is the type of element, i.e. the geometry of the finite element. Tetrahedra are often used because they allow for easier compliance to complex curved geometries. However, meshes with hexaedra are described in some studies (Laakso et al. 2015; Sadleir et al. 2010). Tetrahedral meshes are more time consuming to produce, since they require that smooth triangulated surfaces be built for each mask. These surfaces are then used to generate the volume meshes. Hexahedral meshes can be built directly from the masks, but they do not allow for smooth tissue boundaries to be represented. The only study that was found that compares these two types of meshes (Indahlastari and Sadleir 2015), focused on differences between mean values of the current density in predefined volumes. The differences on the direction of the E-field, however, remain unclear. The third aspect is the quality of the elements in the mesh, which is related to the shape of the elements. Elements with low quality, i.e. elongated elements with very small or very large angles (Windhoff et al. 2013), may lead to numerical instabilities in the FEM. The impact of mesh size and quality is of paramount importance to the obtained E-field results, but remains largely unaddressed in studies published up until now. This is in part due to the complexity of the head geometry, which makes a systematic study of the impact of the mesh challenging.

The stimulating electrodes are accommodated in the model by modifying the scalp’s volume mesh. The latter can have different geometries and can be modeled with different degrees of complexity (see Table 9.1). Most models represent the electrodes as homogeneous patches with uniform conductivities (Miranda et al. 2013; Sadleir et al. 2010; Wagner et al. 2007) or as two layers of materials with different conductivities (metal on top of conductive gel) (Datta et al. 2009a). However, recent studies have shown the importance of accurately modeling the geometry of the electrodes, including the presence of a rubber pad inside the saline soaked sponge (sock-electrodes) and accounting for the position of the metal connector in the electrode (Saturnino et al. 2015).

Once the finite element mesh is obtained, the electric conductivity of each tissue and the boundary conditions must be set. The skin and CSF compartments are usually modeled as homogeneous and isotropic. In most models, the GM, WM and cerebellum are also modeled as such, but several studies use diffusion weighted (DW) MR to estimate the diffusion tensor and assign a conductivity tensor for these tissues (Metwally et al. 2015; Oostendorp et al. 2008; Opitz et al. 2015; Rampersad et al. 2014; Sadleir et al. 2010; Salvador et al. 2015; Schmidt et al. 2015; Shahid et al. 2013, 2014; Suh et al. 2012; Wagner et al. 2014a, b), thus allowing for the anisotropy to be taken into account. The skull can either be modeled as isotropic (single layer or three layers, when spongy bone and compact bone are segmented) or as anisotropic (with a single layer) (Rampersad et al. 2012). Conductivity values for DC or low frequencies can be found in a number of recent studies in the literature (Gabriel et al. 1996; Wagner et al. 2014a, b). However, the results are sometimes in disagreement with older data (more on this in the next section). The conductivity of the electrodes/conductive gel/conductive rubber in the electrode models is also often unknown. For homogeneous electrodes this seems to have a limited effect on the results (Opitz et al. 2015), but for more complex electrode models, this can significantly affect the E-field distribution (Saturnino et al. 2015). Other important boundary conditions specify the currents injected by each electrode. This is performed by adjusting the difference between the electric potential of the electrodes (upper boundary of the electrodes in simple electrode models or the metal connector in more realistic models), until the desired current is obtained.

The next step in these models involves inverting the matrix A therefore obtaining the values of the potential at each node of the mesh. Since these models can have more than ten million degrees of freedom (nodes), iterative procedures are usually employed (Barrett et al. 1993). These have been shown to provide solutions that have a small error with respect to analytical solutions (Faria et al. 2011). This form of validation, i.e. comparing the results of models with the numerical solution of the E-field, can only be performed in simpler geometrical models, when analytical solutions can be obtained (Ferdjallah et al. 1996; Rush and Driscoll 1968). It also does not allow for a validation of the model per se, since many simplifications are introduced related to the geometry of the volume conductor (head) and the electric properties of the tissues. The latter likely affect the predictions of these models much more than numerical inaccuracies.

The final step of the pipeline illustrated in Fig. 9.1 involves the visualization of the E-field distribution and its analysis. This is a crucial step as it will directly influence any clinical decision made about dose. These types of models allow for volume or surface data to be analyzed. Volume data can be processed to yield mean/maximum E-field magnitude values in regions of interest (ROI) . Mean E-field values are preferred because they are less sensitive to variations in element size or low-quality elements in the mesh. Surface data can be used to extract information regarding the orientation of the field, particularly in directions perpendicular to the surface (normal component) and tangential to it (tangential component). Analysis of surface data must be performed with care, as the normal component of the E-field is discontinuous across interfaces between tissues (Miranda et al. 2013). In order to further aid the visualization of the results, methods of cortical surface inflation have also been proposed (Laakso et al. 2015; Opitz et al. 2015), since they allow for an easier identification of the E-field maxima at the bottoms of sulci (Miranda et al. 2013). Another challenge lies in the comparison of E-field distributions across subjects, as inter-individual anatomical variability render direct comparisons non-trivial. One interesting solution that has been proposed is to register the individual models to a common atlas (Laakso et al. 2015).

Limitations of State-of-the-Art Computational Models

Computational models based on segmented structural MR images remain the only viable alternative to predict the E-field as a function of the dose parameters. However, they present limitations that can significantly affect the accuracy of their predictions. The first limitation is related to the inherent difficulties associated with segmentation of certain tissues in MR images . As mentioned previously, the skull and the CSF are particularly difficult to segment from T1/T2 images, and thus require manual corrections, or the use of specific image acquisition parameters that are not commonly used in a clinical setting. However, correctly modeling these tissues is critical to achieving accurate predictions, particularly because the largest change in electrical conductivity occurs precisely at the skull-CSF border. Many studies have shown the impact of the skull modeling (Opitz et al. 2015; Rampersad et al. 2012) and of the thickness of the CSF layer on the electric field in the brain (i.e., cortical regions where the CSF is thin have a significantly higher E-field magnitude, (Opitz et al. 2015)).

The geometry of the head models is usually obtained from MR images of a single individual. This can significantly bias the obtained E-field distribution, since great anatomical variations can exist between individuals (Huang et al. 2016). The effects of these differences have been addressed in a number of studies (Datta et al. 2012; Laakso et al. 2015). One alternative is to create individualized models for each participant in a given study, but this is difficult due to the high computational resources and specific know-how required to build such models. One other option is to use a volume conductor geometry that is obtained from averaged MR images of a number of individuals (Huang et al. 2016). This makes the particular features of the geometry of the volume conductor less susceptible to individual features of a single head but might be less representative for a specific individual (i.e., patients or research subjects).

One large difference between studies lies in the way the electrical conductivity of tissues is modeled, and which values are assigned to it. This can be clearly seen by the sometimes great disparity between the conductivity values assigned to the different tissues (see the second column of Table 9.1). Since this greatly affects the predictions of computational models, some studies have conducted sensitivity analyses on the values of the isotropic conductivities or the effects of the presence of skull and/or WM anisotropy (Laakso et al. 2015; Metwally et al. 2015; Schmidt et al. 2015; Shahid et al. 2013, 2014; Suh et al. 2012; Wagner et al. 2014a, b). No “gold standard” exists nowadays for a way to model the electrical properties of tissues in the DC range of frequencies, so care must be taken when comparing predictions between studies that employ different settings.

Another important limitation of computational models is that the information they provide is hampered by the lack of knowledge about the precise mechanisms of interaction of the E-field with the neurons. In other words, much is still unclear about how the information obtained from these computational models about the magnitude and direction of the E-field translates into modulation of the electrical activity of neurons. At first this may sound paradoxical, since the fundamentals of the E-field interaction with single neurons have been known for a long time (see Chap. 2 and (Roth 1994) for a review). In summary, polarization will occur in regions of maximum value of the activation function (gradient along the neuron of the E-field parallel to the neuron’s trajectory) (Rattay 1986). At the scale of cortical neurons, the gradient of the E-field along the neuron is small, but strong polarizations can occur in regions where the axon or dendritic tree processes bend or terminate (Amassian et al. 1992; Nagarajan et al. 1993) as well as on the soma (Rahman et al. 2013).

The existence of several potential interaction mechanisms, together with the fact that the actual geometry of the neurons is rather complex, makes predictions about sites of activation and the components of the E-field more likely to influence them hard to make. One generally adopted approximation that has been proposed to solve this is to assume that the polarization of the neurons will be proportional to the E-field’s magnitude. This is termed the “quasi-uniform” assumption (Bikson et al. 2013) and similar approximations have been proposed before in other forms of stimulation, such as TMS (Ruohonen 1998). The E-field’s component in a direction either perpendicular (radial) or tangential to the cortical sheet may also be explicitly computed (Faria et al. 2011) and might be the relevant factor in determining the field’s interaction with neurons. The radial component of the E-field optimally polarize pyramidal cells in the cortex with dendrites that extend normal to the surface (Bikson et al. 2004; Datta et al. 2008). It also easily explains the polarity dependent nature of the neuro-modulatory effects elicited by tDCS that is observed up to a certain value of current intensity (Merlet et al. 2013; Nitsche and Paulus 2000). Tangential E-fields would optimally polarize axon afferents projecting along the surface (Rahman et al. 2013). The quasi-uniform assumption does not model neuron specific effects (Radman et al. 2009) or consider the role of changes in electric fields, as notably occurs across the grey-white matter interface (Miranda et al. 2003, 2007; Salvador et al. 2011).

Use of Computational Models in Clinical Practice

Optimizing Efficacy

Since the early days of tDCS, many options regarding dose parameters have been chosen on a trial-and-error basis. As an example, consider the original study by Nitsche and Paulus (Nitsche and Paulus 2000), which demonstrated that tDCS could elicit long lasting excitability changes in the motor cortex, by placing one electrode over the primary motor cortex and the other over the contralateral supraorbital area (M1-SO montage). In that study, the authors reported several other bipolar electrode configurations that were tested and that failed to elicit the same results. While in some follow-up studies these canonical results were interpreted as suggesting that current flow in the M1-SO montage affected mostly the motor region- and by extension that tDCS could be focalized to any region by placing a large pad electrode over it – it is import to recognize that such an interpretation was not implied by the authors of the original study.

The recent computational studies that have emerged provide useful insights into the E-field distribution in the brain that can aid in optimizing the efficacy in clinical applications of tDCS. Many of the initial computational studies aimed at clarifying the E-field distribution in realistic representations of electrode montages used in clinical practice in a series of applications (Datta et al. 2009a; Miranda et al. 2013; Sadleir et al. 2010). In contrast to earlier models based on concentric spheres (Miranda et al. 2006; Rush and Driscoll 1968) or abstracted geometry (CAD, (Wagner et al. 2007)), a critical feature of models applied since ~2009 is the incorporation of “gyri-precise” resolution based on precise anatomical MRI scans, and attention to continuity of CSF involving smoothing beyond scan resolution. This level of detail resulted in several key predictions that challenged prevailing views on dose design: (1) significant current flow occurs between (rather than simply under) electrodes; (2) current is clustered in hot-spots whose locations depend on idiosyncratic anatomy; (3) both electrodes are functional regardless of position (including extracephalic or “supra-orbital) and current under each electrode is influenced by the position of the others (Bikson et al. 2010). It should be noted that even as modeling technology has continued to evolve (e.g. addition of anisotropy; (Oostendorp et al. 2008; Suh et al. 2009)), and despite variation in methods (e.g. conductivity, (Opitz et al. 2015)), these basic predictions remain largely intact over the last 7 years of intensive modeling studies. In fact, is it precisely because current flow is not intuitive that models are important tools in the interpretation and design of tDCS studies.

These high-resolution studies proved useful in determining which regions were targeted by these configurations and what was the direction of the induced E-field in different regions. Based on the latter, a newer batch of studies has appeared geared towards optimizing multi-electrode montages to achieve a target E-field distribution and orientation in pre-defined cortical regions of interest (Dmochowski et al. 2011; Ruffini et al. 2014; Sadleir et al. 2012). They use the superposition principle, i.e., the principle that the E-field distribution in the head induced by a given set of electrodes can be obtained by the weighted sum of the E-fields induced by bipolar electrode configurations with a common return electrode. The weights used to sum the bipolar induced E-field configurations are the currents set in each electrode of the multi-electrode configuration. These types of models then determine the current intensity applied at each electrode (placed in a predefined array of positions) such that the induced E-field best approximates a pre-defined target E-field distribution in the brain. This optimization procedure can be further constrained by imposing maximum values for the current on each electrode and the total injected current.

The main difference between the implementations in each study is related to the components of the E-field which were optimized (pre-determined E-field with any desired direction, as specified by the user (Dmochowski et al. 2011), E-field component radial to the cortical sheet in (Ruffini et al. 2014) or magnitude of the E-field (Sadleir et al. 2012)). Other differences are the algorithms used to minimize the difference between the induced E-field and the target field, and the type of electrodes and pre-defined electrode positions. These studies have shown that it is possible to induce E-fields in target regions with higher focality and/or magnitude that those using conventional approached (i.e. bipolar “pad-like” electrodes) (Dmochowski et al. 2011). Moreover, they suggest the possibility of using data from functional imaging techniques (electroencephalography, positron emission tomography and functional magnetic resonance imaging) to derive cortical activation maps that may serve as the target E-field distribution that served as input to these studies (Ruffini et al. 2014). Finally, they offer the possibility of avoiding unwanted stimulation of certain regions by minimizing the E-field induced in the latter (Sadleir et al. 2012).

In spite of the usefulness of computational models, their implementation can be a complex task for research groups, particularly those with clinical applications. Working in close collaboration with a computational modeling group might be an alternative, and indeed, many studies nowadays seek to incorporate the information from these models as a rationale to dose parameter decisions and/or as support for some of the conclusions of the study (see references marked as type (5) in Table 9.1). The creation of tools to allow for easier implementation of these models (Jung et al. 2013; Windhoff et al. 2013) and of databases of models (Truong et al. 2014) can also allow for a wider implementation of computational models as part of the pipeline for experiment design.

Safety and Tolerability Considerations

It is not trivial to translate the predictions of computational E-field calculations to statements about safety or tolerability of a specific dose selection in tDCS. This occurs because, while the E-field distribution in tDCS can be predicted from these models in different tissues, the relation between the field and potential unwanted side-effects are not well known. None-the-less, under the assumption that increasing the E-field in a given target increases the theoretical risk of injury, even by providing comparative electric field across montages and subject populations (e.g. pediatric, stroke, injury), models are a valuable tool to assess risk.

One aspect related to tolerability, for instance, are the reported tingling and itching sensations under the electrodes and the acute erythema that has been associated with vasodilation (Woods et al. 2016). Computational models allow for the E-field in the skin-electrode interface to be calculated, and complex electrode models incorporating the gel and conductive rubber pads within saline soaked electrodes have been produced (Saturnino et al. 2015). These calculations, however, are of limited application since no model exists at the moment to relate the E-field to the reported undesired effects.

Regarding safety issues, attention has focused on predicting the electric field (current density) threshold at which injury may occur in the brain, skin or other structures. This work has, in turn, relied on injury thresholds proposed by animal models – which generally suggest thresholds more than one order of magnitude above clinical tDCS intensities. Computational models are essential to make scaling of intensities between animal models and humans more precise for this purpose, because for the same applied current or current density at electrodes, the resulting electric fields in rodent are much higher than those in human. For instance, injuries linked to possible tissue damage caused by heating (joule heating) may be relevant to tDCS. This has been addressed in a study (Datta et al. 2009b) which has shown that application of tDCS with a current to electrode area ratio of 142.9 A/m2, a value that has been shown to lead to lesions in in vivo rat models (Liebetanz et al. 2009), caused a small maximum temperature increase in the brain (0.55 °C) but a significant scalp temperature increase (14.68 °C). However, these models require validation and many important factors, like sweat gland response in the scalp, were not modeled.

Another important safety issue is related to the usage of extracephalic return electrodes , which have been proposed to avert unwanted effects of tDCS in cortical areas away from the target region (Moliadze et al. 2010). However, modeling studies showed that use of extracephalic electrodes does not “cancel” the role of the return electrode, but rather creates extensive current flow through the foramen magnum, thereby affecting deep and mid-brain structures (Datta et al. 2011). It was speculated that current flow produced by extracephalic electrodes might interfere with other excitable tissues in the brainstem or the heart, which can lead to severe complications. Studies addressing this (Parazzini et al. 2013a, b) calculated current density magnitudes in these tissues and compared them to reported threshold values for cardiac fibrillation or with values induced in the brainstem during conventional bipolar montages, as these do not show unwanted brainstem neuromodulatory effects. These approaches to assess the lack of unwanted effects of the induced E-field still have some potential shortcomings, however, since the direction of the E-field is also a factor influencing neuronal stimulation and this aspect is not taken into account when analyzing only the magnitude of the field. One alternative to the use of extracephalic electrodes is the use of the multi-electrode optimization approaches that were mentioned in the last section, which may allow for a reduction of the unwanted secondary activations in unwanted brain areas. The latter also allows for the current in each electrode and the total overall current to be limited to specified maximum values.

The use of tDCS in subjects with skull defects (Datta et al. 2010) or in children (Gillick et al. 2014; Kessler et al. 2013; Minhas et al. 2012; Parazzini et al. 2014a) is also a matter that warrants caution, since the E-field distribution in these populations may be significantly different from that in healthy adult subjects. The presence of skull holes, for instance, was found to significantly increase the E-field if the electrode is placed directly on top of it. Regarding pediatric tDCS applications, the E-field induced in the child brain was found to be on average higher than that induced in adult brain using the same dose parameters. These results support the importance of carefully studying the E-field distribution in these models and leveraging them as a means to adjust dose parameters.

Dose Selection on an Individual Basis

Currently, the technical sophistication required to generate high-resolution computational models make implementation on a subject specific basis a very time-consuming process. Recent efforts have aimed to automate the individual segmentation process in a manner that does not compromise model precision (Huang and Parra 2015). Typically, however, computational studies involve the creation of one head model and the generalization of the results to other head geometries. This approach works well when the study’s goal is to determine basic principles of the spatial distribution of the E-field in tDCS. A few modeling studies have studied the impact of intersubject variability in the E-field distribution in tDCS (Datta et al. 2012; Laakso et al. 2015). These studies report significant differences arising from details in the geometry of the cortical sheet. One of these studies reports that the E-field in the hand motor area across 24 individual subjects followed a normal distribution with a standard deviation of about 20% of the mean (Laakso et al. 2015). Recently, a standardized head based on the MNI has been developed specifically for tDCS modeling, and has been show to provide reasonable predictions of individualized subject response (Huang et al. 2016).

Other examples of cases warranting individualized head models are studies involving stroke patients, for which the size of the lesion and its location and shape can significantly affect the distribution of the induced E-field (Datta et al. 2011). Other examples are the already mentioned studies involving children (Minhas et al. 2012) or studies involving obese subjects (Truong et al. 2013). In those cases, individualized models remain the only viable way to provide a realistic E-field distribution and thus optimize dose parameters (Dmochowski et al. 2013).

Examples of Application of Computational Modeling in Case Studies

Computational models of tDCS can be performed proactively or retroactively. Proactive modeling can influence montage selection by informing researchers of stimulation focality and intensity for a region of interest. In atypical case studies, safety concerns can be assessed and mitigated by proactively modeling the stimulation protocol and comparing to a typical subject. Examples of this strategy include pediatrics, stroke, and subjects with cranial defects (Bikson et al. 2016; Minhas et al. 2012). In Fig. 9.2, tDCS is simulated in a subject with and without idealized Deep Brain Stimulation (DBS) leads. As an extreme case, the burrhole defect typical in subthalamic nucleus DBS is allowed to be fluid filled and relatively conductive. Common sponge (conventional) and HD-tDCS montages for motor and cerebellar stimulation are compared. Fluid filled burrholes draw a greater amount of current density than what would normally exist with healthy tissue (dashed images). However, peak current density and electric field are minimally affected (less than twofold). In general, HD configurations exhibit lower electric field intensities in deep brain structures while exhibiting more focal field patterns.

Fig. 9.2
figure 2

Simulation of tDCS in subjects with idealized Deep Brain Stimulation (DBS) leads using conventional 5 × 7 cm sponge and HD (4 × 1) montages

While tDCS modeling tools are becoming more accessible (COMETS, Bonsai, SimNIBS) (Jung et al. 2013; Thielscher et al. 2015; Truong et al. 2014) many transcranial stimulation studies have been and are performed without the guidance of modeling. Retroactive modeling of a specific study’s stimulation parameters can help to resolve mechanisms of action or explain variance within or between studies. Fig. 9.3 represents a post-hoc analysis of common tDCS montages used in schizophrenia (Brunelin et al. 2012; Shiozawa et al. 2013). The electric field magnitude on the cortical surface depicts regions of maximum stimulation regardless of field orientation (A). The radial component of the electric field predicts the effects of stimulation on layer V pyramidal neurons aligned perpendicular to the cortical surface (B). Field orientation, anodal (red) or cathodal (blue), is commonly postulated to have excitatory or inhibitory effects on local regions (B). Meanwhile, tangential electric field magnitude is predicted to affect local connections oriented along the cortical surface (C).

Fig. 9.3
figure 3

Posthoc analysis of common tDCS montages used in schizophrenia. (a) electric field magnitude; (b) electric field normal to the cortical surface; (c) electric field tangential to the cortical surface (Brunelin et al. 2012; Shiozawa et al. 2013). (Adapted from Brunoni et al. (2014); with permission)