Introduction

18F-fluoro-2-deoxy-d-glucose (18F-FDG) positron emission tomography (PET) has been widely used for detecting malignant tumors, differential diagnosis, determination of therapeutic strategy, monitoring treatment response and also in follow-up examinations [1,2,3,4,5,6,7,8,9,10]. Standardized uptake value (SUV) is a parameter used to evaluate 18F-FDG accumulation semiquantitatively. In addition, quantitative accuracy and reproducibility of SUV are especially important for differential diagnosis and for monitoring treatment response. Therefore, routine inspection and quality assessment of PET scanners are necessary to maintain accurate quantification during the PET examinations [11, 12]. Noise equivalent count (NEC) has been used as a standard metric to assess the performance of a PET scanner [13,14,15,16]. NEC describes the equivalent coincidence counts that would have the same noise properties as the net true counts, corrected for random and scattered coincidences. NEC reflects the quality of acquired raw data, and thus does not take into account the impact of reconstruction algorithms or correction methods. Recently, the incorporation of additional information, such as point-spread function and time-of-flight information, into iterative reconstruction algorithm markedly improved the quality of the PET image [17]. Signal-to-noise ratio in the liver (SNRliver) is used as a metric to evaluate image quality obtained from reconstructed images. SNRliver has been reported to be useful for determining the injection dose of 18F-FDG in relation to the patient’s body weight and the optimization of the acquisition time [18,19,20,21,22].

However, a poor correlation has also been reported between SNRliver and the result of the visual evaluation of clinical images [14, 15]. Because the 18F-FDG uptake in the liver is not homogeneous, the placement of the regions-of-interest (ROIs) for the liver was considered to affect the stability and reproducibility of the SNRliver [23]. Although the guidelines for the oncology 18F-FDG-PET/CT data acquisition protocol recommended placing a circular ROI with 3 cm diameter in the right lobe of the liver in a coronal image, various other methods have been also reported [17,18,19,20,21,22, 24, 25]. Furthermore, a spherical volume-of-interest (VOI) with 3 cm diameter for measurement the SUV of the liver has also been reported to be highly reproducible [26]. Therefore, the most appropriate method for placing an ROI, which provides a stable and reproducible SNRliver needs to be established.

The aim of this study is to compare the effects of different ROIs and VOIs on measurement of SNRlivers, and to determine the most reproducible method to measure SNRliver.

Materials and methods

Subjects

This retrospective study was approved by the institutional review board of our hospital, and a written informed consent was waived. This study included 108 patients (67 men and 41 women, mean age 57.2 ± 11.0 years, mean body weight 65.1 ± 12.9 kg, mean body mass index (BMI) 23.7 ± 3.2 kg/m2) who underwent 18F-FDG-PET/CT scan for the purpose of cancer screening. Patients with abnormal accumulations in the liver, those with a high blood-sugar level more than 110 mg/dL, those with liver dysfunction and those with a liver too small to place ROIs were excluded from this study.

Scanner description

PET/CT data were acquired using a True Point Biograph 40 scanner (Siemens AG, Munich, Germany). This scanner is comprised of 39 rings, with a total of 144 lutetium oxyorthosilicate (LSO) detectors, covering an axial field of view (FOV) of 16.2 cm and a transaxial FOV of 68.3 cm in diameter. Each LSO crystal is 4 × 4 × 20 mm3.

18F-FDG-PET/CT protocol

The patients fasted for approximately 6 h before 18F-FDG administration. The dose of 18F-FDG was 3.7 MBq/kg for patients weighing 70 kg or less and 259 MBq for patients heavier than 70 kg. After the administration of 18F-FDG, the patient rested in a dimmed room for an hour before the PET/CT scan was started. The patients were scanned with their arms down. Scan duration of one bed position was 2 min. The overlap ranges of PET data acquisition were 28%. The CT scan for attenuation correction was performed according to the following parameters: 120 kVp, 80 mAs, collimation 24 × 1.2 mm, pitch 0.8, 0.5 s per one rotation. The diameter of the FOV of CT images was 70 cm. The slice thickness of the reconstructed CT image was 5 mm with a 3 mm interval.

Scatter correction was performed using the single scatter simulation method. PET acquisition data were reconstructed using the method of Fourier rebinning two-dimensional ordered subset expectation maximization (FORE–OSEM) with two iterations and eight subsets, and were then smoothed with a Gaussian filter with a full width at half maximum (FWHM) of 6 mm. The slice thickness of PET images was 5 mm and the interval of slices was 3 mm.

Data analysis

For data analysis, the VOXBASE II fusion viewer (J-mac system, Sapporo, Japan) was used. Circular ROIs with a diameter of 3 cm (ROI3; 6.72 cm2) and of 4 cm (ROI4; 13.43 cm2) were placed on the right lobe of the liver on five consecutive coronal images (i = 1–5) (Fig. 1a). ROIs were carefully placed on the center of the right lobe of the liver, taking care not to include the hepatic portal region and subphrenic region. Spherical VOIs with a diameter of 3 cm (VOI3; 13.82 cm3) and of 4 cm (VOI4; 32.83 cm3) were placed on the right lobe of the liver in 3 dimensional volume data (Fig. 1b). The VOI placement was repeated five times (j = 1–5). The average and standard deviation of SUV within each ROI and VOI were calculated as SUVmean and SUVSD, respectively.

Fig. 1
figure 1

Placement of an ROI and a VOI. a Circular ROIs with a diameter of 3 cm (dashed line) and 4 cm (solid line) were placed on the right lobe of the liver on consecutive five coronal images. b Spherical VOIs with a diameter of 3 cm (dashed line) and 4 cm (solid line) were placed on the right lobe of the liver in three-dimensional volume data five times

The SNRliver,i with an ROI and the SNRliver,j with a VOI were defined with the following formulae.

$$\begin{aligned} {\text{SN}}{{\text{R}}_{{\text{liver}},i}} &= \frac{{{\text{SU}}{{\text{V}}_{{\text{mean}},i}}}}{{{\text{SU}}{{\text{V}}_{{\text{SD}},i}}}},{\text{ in the case of circular ROIs,}} \\ {\text{SN}}{{\text{R}}_{{\text{liver}},j}}&= \frac{{{\text{SU}}{{\text{V}}_{{\text{mean}},j}}}}{{{\text{SU}}{{\text{V}}_{{\text{SD}},j}}}},{\text{ in the case of spherical VOIs}}, \\ \end{aligned}$$

where i and j represent the number of an ROI or a VOI on each patient.

SNRliver was calculated by the following formulae using the values of five ROIs or VOIs.

$$\begin{aligned} {\text{SN}}{{\text{R}}_{{\text{liver}}}}= & \frac{1}{5}\left( {\mathop \sum \limits_{{{\text{i}}=1}}^{5} {\text{SN}}{{\text{R}}_{{\text{liver}},i}}} \right),{\text{ in the case of circular ROIs}}, \\ {\text{SN}}{{\text{R}}_{{\text{liver}}}}= & \frac{1}{5}\left( {\mathop \sum \limits_{{{\text{j}}=1}}^{5} {\text{SN}}{{\text{R}}_{{\text{liver}},j}}} \right),{\text{ in the case of spherical VOIs}}, \\ \end{aligned}$$

SD of the SNRliver,i and SNRliver,j was also calculated, respectively.

Statistical analysis

Comparisons of the mean of SUVmean, SUVSD and SNRliver among different ROIs and VOIs were performed by one-way ANOVA, followed by a post hoc Tukey HSD test. Comparison of the SD of SNRliver,i and SNRliver,j among ROIs and VOIs was performed in the same way for the purpose of determining the stability of measurement of SNRliver in each ROI. We used a level of significance of p < 0.05.

Results

Figure 2 shows the comparison of the mean SUVmeans among two ROIs and two VOIs. SUVmeans were not significantly different among any ROIs or VOIs. On the other hand, Fig. 3 shows that the SUVSDs with ROI3 were significantly lower than those with others. SUVSD with ROI4 was also significantly lower than that with VOI4. SUVSDs with VOIs were not significantly different from each other.

Fig. 2
figure 2

Comparison of SUVmean among ROIs and VOIs. SUVmeans were not significantly different among any ROIs or VOIs (p = 0.94)

Fig. 3
figure 3

Comparison of SUVSD among ROIs and VOIs. The mean value of SUVSD with ROI3 was significantly lower than those with others. SUVSD with ROI4 was also significantly lower than that with VOI4 but was not significantly lower than that with VOI3 (p = 0.06). SUVSDs with VOIs did not significantly differ from each other

Figure 4 shows the comparison of SNRlivers among two ROIs and two VOIs. SNRlivers with two ROIs and two VOIs were significantly different from each other, and SNRlivers with ROI3 was the highest, followed by ROI4, VOI3 and VOI4.

Fig. 4
figure 4

SNRlivers of ROIs and VOIs. SNRlivers with two ROIs and two VOIs were significantly different from each other. SNRlivers with ROI3 was the highest followed by ROI4, VOI3 and VOI4

Figure 5 shows the comparison of the SDs of SNRliver,i and SNRliver,j among two ROIs and two VOIs. The SD of VOI4 was lower than those of VOI3, ROI4 and ROI3 in ascending order. Each SD had a significant difference from the others.

Fig. 5
figure 5

The standard deviations of SNRliver of ROIs and VOIs. The SD of VOI4 was lower than those of VOI3, ROI4 and ROI3 in ascending order. Each SD showed a significant difference from the others

Discussions

In this study, we investigated the influence of ROI or VOI determination on the stability and reproducibility of SNRliver to evaluate PET image quality. Although the SUVmean was not different among ROIs and VOIs, the SUVSD was small with a small ROI. Furthermore, a small ROI resulted in a large SNRliver and large SD of SNRliver.

The difference in SUVmeans among ROIs and VOIs were not statistically significant. This suggests that a similar SUVmean of the liver can be obtained when using either the diameter of 3 or 4 cm circular ROIs or spherical VOIs. On the other hand, the SUVSD with ROI3 was smaller than those with other ROI and VOIs. This is presumably dependent on the region where the ROI is placed. Manual placement of a small ROI tended to be restricted to homogeneous regions unintentionally. Thus, it may not represent the general variability of the liver SUV. Therefore, the placement of a relatively large ROI is recommended to represent the general FDG uptake and distribution in the liver. SNRliver with ROI3 was the largest, and was followed by those with ROI4, VOI3 and VOI4 in descending order. As the SUVmean was not significantly different among all ROIs and VOIs, the difference of SNRliver among ROIs and VOIs was considered to result from the difference in SUVSD. In our study, the SD of SNRliver with ROI3 was also the largest and was followed by those with ROI4, VOI3 and VOI4. Thus, the SNRliver with VOI4 is considered to be the most reproducible among the ROIs and VOIs. McDermott et al. used a spherical VOI with a diameter of 5 cm and showed a small SD of SNRliver [25]. From the findings described above, we have concluded that the measurement of SNRliver with a smaller ROI is not reproducible, and that to obtain a reproducible SNRliver, it is necessary to place a sufficiently large VOI.

Concerning the manual placement of an ROI or a VOI, placing a circular ROI has a higher degree of freedom than a spherical VOI. Therefore, the variation of measured values should become larger with circular ROIs. Some methods such as increasing the number of measurements is recommended when using a circular ROI [17]. With a larger diameter of ROI or VOI, the degree of freedom when placing an ROI was low, and then the deviation of measurement of SUVSD became smaller. However, placing ROI or VOI with too large of a diameter may be difficult in some cases in which the patient has a small liver.

In this study, a single researcher performed data analysis. Thus, the variance indicates the intra-observer difference for ROI analysis. Our results showed that the SD of SNRliver, which was an intra-observer variance, was small in both large ROIs and large VOIs. Viner et al. examined the inter-observer agreement of SUVmean normalized with lean body mass in the liver [26]. They used a VOI with a 3 cm diameter and showed a good inter-observer agreement of SUVmean measurements. An inter-observer variance is generally considered to be larger than an intra-observer variance [27, 28]. This suggests that using a larger VOI with a 4 cm diameter can be expected to obtain better results. Although we did not examine the inter-observer variance of SNRliver at this time, we predict it would be larger but have a similar tendency to the intra-observer variance of SNRliver. Further examination is required to elucidate the effect of ROIs and VOIs on the inter-observer variance of SNRliver.

SNRliver itself is considered to be unstable as an index of the image quality. Primarily, the administered dose of 18F-FDG is a potential influencing factor in the reliability of SUV [29]. A relatively low administered dose generally results in impaired image quality due to low count statics. This effect also influences the SNR measurement. We suppose that the SD of SNR measured with a small ROI may be larger than with a larger ROI due to variance of SUV resulting from increasing image noise in the liver. Therefore, we infer that reproducibility of SNR measurement would also be improved with a large ROI in the case of a low administered dose. In addition, as the administered dose in this study was determined based on the patient body weight, the influence of the administered dose on the results should be considered negligible, with the exception of possible calibration errors by the dose calibrator. On the other hand, at institutions where 18F-FDG is supplied by venders, the administered dose cannot easily be adjusted, which may affect the reliability of SUV, as a result of SNR. Secondly, patients with a large BMI are reported to show high normal organ SUVs due to copious adipose tissue, which does not accumulate 18F-FDG [30,31,32]. A lean body mass is recommended when calculating regular tissue SUV due to potentially increased values in patients with a high BMI [30,31,32]. Moreover, in the case of an obese patient, image quality is deteriorated by increases in the number of random and scattered coincidences, and thus patient body weight influences the SNR measurement. However, the effect of body weight as a normalization factor for calculating SUV is negligible in SNR derived as the ratio of SUVmean to SUVSD. In addition, patient body weight was not considered to be a variable in this study, as no patient was of a weight that could be deemed as having an effect on regular tissue SUV. Third, underestimation of liver SUV near the diaphragm due to respiratory movement sometimes occurs in cases involving a small liver. The SUV in the dorsal region of the right lobe of the liver is frequently low, due to an artifact associated with arm motion. The ROIs or VOIs must be carefully placed to avoid such regions. Fourth, this study did not include patients with liver dysfunction, because the liver FDG uptake in patients with liver dysfunction has been reported to be decreased and heterogeneous, and so the SNRliver of such patients needs to be analyzed in separate investigations [23, 33].

Conclusion

To obtain a stable and reproducible SNRliver, a 4-cm-diameter spherical VOI is considered as superior to a 3-cm-diameter spherical VOI, a 3-cm-diameter and 4-cm-diameter circular ROI.