Abstract
Background
Modular tibial trays have been utilized in TKA for more than 20 years. However, concerns have been raised about modular implants and it is unclear whether these devices are durable in the long term.
Questions/purposes
We determined (1) survival, (2) relationship of age and polyethylene thickness with revision, (3) function, and (4) radiographic lucencies and osteolysis in patients having a single TKA implant at 20-year followup.
Methods
We prospectively followed 75 patients implanted with 101 Press-Fit Condylar® (Johnson and Johnson Professional, Inc, Raynham, MA, USA) posterior cruciate-retaining TKAs (with modular tibial trays) between 1988 and 1991. At 20 years, 59 patients were deceased. We clinically evaluated the living 16 patients (22 knees) and contacted the relatives of all deceased patients to confirm implant status. We clinically assessed 14 of the 16 patients with the Knee Society score, WOMAC, and UCLA and Tegner activity level scores. Radiographically, we determined lucencies, component migration, and osteolysis. We performed survival analysis including all original patients. Minimum followup was 20 years (mean, 20.6 years; range, 20–21.8 years).
Results
Six reoperations were performed in five patients (6% rate of revision) over the 20-year followup. All revisions were related to polyethylene wear and occurred at least 10 years after the primary procedure. Survivorship with revision for any reason as the end point was 91% (95% CI, 0.83–0.97) at 20 years. Average Knee Society clinical and functional scores were 90 (range, 60–100) and 59 (range, 30–87), respectively.
Conclusions
Our data demonstrate the durability of this posterior cruciate-retaining TKA design. The data provide a standard for newer designs and newer bearing surface materials at comparable followup.
Level of Evidence
Level IV, therapeutic study. See Instructions for Authors for a complete description of levels of evidence.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
Condylar-type TKA has been performed for more than 35 years. Several studies [28, 31, 33] have reported findings at or near a mean of 20 years. These studies show revision rates for aseptic loosening of between 0.8% and 3.3% at a mean 19 to 20 years, with survivorship with revision for any reason as the end point of 83.2% at 20 years, 77% at 21 years, and 91% at 23 years, respectively.
The introduction of modular tibial trays has allowed for greater operative versatility, allowing for cases of polyethylene exchange as an option in revision knee surgery, which allows for a potentially less involved (less destructive) and more cost-effective option. However, modularity has not been without controversy. Backside wear has been one major concern [13, 14, 16, 30]. Some have cautioned against simple liner exchanges, especially in revisions for accelerated polyethylene wear, particularly when there are issues with alignment and instability [2, 15]. Yet, in cases of well-fixed and well-aligned knees, some report polyethylene exchange in the revision knee situation is a reasonable option even in patients with extensive osteolysis [7, 23]. Followup studies of 10 to 17 years of TKA using modular tibial trays [21, 32, 34, 36] have shown survivorship of 87% to 92.8% at 12 to 17 years. We previously reported a 100% survivorship at 10 years [20] and 91% at 15 years [29] in patients with a modular tibial tray (no revisions at 10 years and six revisions at 15 years). Whether these findings are durable at longer followup is unclear.
We therefore determined (1) survival, (2) the relationship of age and polyethylene thickness with revision, (3) function, and (4) radiographic lucencies and osteolysis in patients treated by a single surgeon with a posterior cruciate-retaining device with a modular tibial tray at a minimum 20-year followup.
Patients and Methods
Between November 1988 and January 1991, one of the authors (SSK) performed 101 cemented TKAs with a modular posterior cruciate-retaining Press-Fit Condylar® (PFC®) prosthesis (Johnson and Johnson Professional, Inc, Raynham, MA, USA) in 75 patients. During this interval, the surgeon performed seven additional TKAs: two posterior-stabilized Insall-Burstein II® (Zimmer, Inc, Warsaw, IN, USA), four posterior-stabilized PFC® prostheses when an insufficient PCL was identified intraoperatively, and a hybrid PFC® with an uncemented femoral component in a young patient. Coumadin was routinely used for deep venous thrombosis prophylaxis postoperatively. The average age at the time of surgery was 71.2 years (range, 52.2–88.8 years). Study participants included 44 women (59 knees) and 31 men (42 knees). There were 52 (51%) right knees and 49 (49%) left knees. Twenty-six (35%) patients required bilateral TKAs. Of these, six patients had the bilateral TKAs performed under one surgical procedure. Diagnoses included primary osteoarthritis in 86 (85%) knees, rheumatoid arthritis in 13 (13%), avascular necrosis in one (1%), and posttraumatic osteoarthritis in one (1%). Previous knee procedures had been performed in 11 patients before the index TKA, including six arthroscopies, three meniscectomies, and two patellectomies. At a minimum of 20 years after the primary procedure, 16 patients (22 knees) were living and 59 patients (79 knees) were deceased. No patients were lost to followup. After locating living patients, we obtained consent for study participation as per the protocol previously approved by our institutional review board.
All components were cemented. The modular metal-backed tibial tray was titanium and came in six sizes. The femoral component was a cobalt-chrome alloy. A three-peg, all-polyethylene patella was used in all procedures except two. In these two procedures, the surgeon did not resurface the patella because a prior patellectomy had been performed. A posterior-lipped polyethylene insert with minimally raised anterior and posterior borders and a large radius of curvature was used in all procedures. Gamma-irradiated-in-air polyethylene and all-polyethylene patella components were utilized in all cases. All cases were fixed with cement and prospectively followed at 5-year intervals [20, 29]. During the period of these procedures, a more conforming curved tibial insert did become available but was not used in these patients. The thickness of the polyethylene insert was 8 mm (minimum thickness, 5.3 mm) in 78 (77%) knees, 10 mm (minimum thickness, 8 mm) in 17 (17%), 12.5 mm (minimum thickness, 10 mm) in five (5%), and 15 mm (minimum thickness, 13 mm) in one (1%). Beginning in 1991, the manufacturer increased the minimal thickness of the 8-mm insert from 5.3 to 6 mm, but we exclusively used the thinner implant in this patient cohort. All components were gamma irradiated in air.
Two investigators (MWB, DWH) not involved in the surgical procedures and not receiving compensation from the implant manufacturer contacted all patients. For all deceased patients, we made contact with a surviving relative. In all cases, the survival of the prosthesis at the time of death was confirmed, and no pending revisions or revision indications were identified at the time of death. Of the 16 patients (22 knees) known to be living at the time of followup, all were located, and the survivorship or need for revision of the prosthesis was confirmed. Due to dementia, two patients (three knees) were unable to participate in the full telephone assessment of clinical status as described above. Therefore, 14 patients (19 knees) completed the full questionnaire evaluation. Clinical measures included the pain and functional components of the Knee Society scoring system [26], WOMAC [4], and UCLA [1] and Tegner [35] activity level scores. We obtained radiographs, including standing AP, lateral, and Merchant views, in the standard fashion. We obtained 20-year radiographic followup for 12 patients (17 knees, 77% of knees in living patients) with an average followup of 20.6 years (range, 20.0–21.8 years). Among the four patients (five knees) living at the time of this study who did not return for radiographic followup, we evaluated radiographs of three patients (four knees) at an average of 15.3 years from the index procedure. The final living patient who chose not to return for followup radiographs had 0 years of radiographic followup but did participate in the questionnaire evaluation via telephone and had not been revised. The average radiographic followup of all living patients was 19.5 years (range, 0–21.8 years). Two authors (DWH, JJC), not directly involved in the surgery or subsequent clinical care of the patients, evaluated all radiographs using the Knee Society protocol [17]. This included the evaluation of radiolucent lines around the components, component position change, and evidence of osteolysis measuring greater than 1.0 cm2. We have demonstrated good interobserver agreement with this approach [13].
Kaplan-Meier survivorship analysis with 95% CI [27] was performed using SPSS® 13.0 software (SPSS Inc, Chicago, IL, USA) with revision of any component for any reason, revision of any component for aseptic loosening, and component-specific revision as the end points. Additionally, we calculated Kaplan-Meier survivorship, with patient survival as an end point, based on age at the time of index surgery. We utilized the Cox proportional-hazard regression analysis in assessing for correlation of patient age and polyethylene implant thickness with need for revision.
Results
The survivorship (Fig. 1) of any component with revision for any reason as the end point was 90.8% (95% CI, 0.83–0.97) at 20 years. The survivorship of the femoral component with revision for any reason as the end point was 95.3% (95% CI, 0.87–0.99) at 20 years. The survivorship of the tibial component with revision for any reason as the end point was 95.2% (95% CI, 0.87–0.99) at 20 years. The survivorship of the patellar component with revision for any reason as the end point was 95.3% (95% CI, 0.89–0.99) at 20 years. The survivorship of the liner for revision for any reason as the end point was 90.8% (95% CI, 0.83–0.97) at 20 years. The survivorship of any component with revision due to aseptic loosening as the end point was 93.8% (95% CI, 0.87–0.98) at 20 years. The survival rates (Table 1) with revision for any reason as the end point in this cohort of patients were 100% at 10 years, 91% at 15 years, and 91% at 20 years. Six reoperations were performed in five patients (Table 2). One patient had both knees revised. At the time of revision, we assessed the fixation of the tibial, femoral, and patellar components. If the components were well fixed and without obvious malalignment, revision consisted of simple polyethylene exchange. In the presence of additional component loosening or osteolysis, the tibial, femoral, and/or patellar components were revised. Revision for any reason occurred in six of 101 (6%) knees at a minimum of 20 years. All revisions were related to polyethylene wear (four resulting in extensive osteolysis and aseptic component loosening and two related to wear of the polyethylene), and all occurred more than 10 years after the index procedure. There were no new revisions since the previously reported minimum 15-year followup.
We found an association (p = 0.030) between need for revision and age at the time of surgery. Among patients requiring revision, the average age at the time of the index procedure was 63.5 years (range, 60.8–68.6 years). The average age of the overall cohort at the time of surgery was 71.6 (range, 52.2–88.8 years). All six revisions occurred in patients with 8-mm polyethylene inserts. Those with 8-mm inserts tended (p = 0.051) to need revision. None of the four revised knees in three living patients had clinical or radiographic signs of failure at an average of 7.3 years (range, 5.5–8.9 years) after revision. The other two revised knees in two deceased patients at 20-year followup had not required additional revision surgery at the times of their death, which were 2.8 and 7.5 years after revision.
For living patients, other than the two infirmed patients, the average WOMAC scores (corrected to higher score = higher outcome) were 94 of 100 (range, 75–100) for the pain component, 84 (range, 25–100) for the stiffness component, and 79 (range, 53–100) for the functional component. The average Knee Society pain score was 90 of 100 (range, 60–100), and the average function score was 59 of 100 (range, 30–87). The average patient age at the time of latest clinical followup was 82.0 years. The average UCLA and Tegner activity level scores were 3.9 and 2.9, respectively, which correlated to light labor and mild activities, such as walking, limited housework, and limited shopping. Functional limitations were attributed primarily to other musculoskeletal or systemic morbidities, not the operative knee(s).
Of the 12 nonrevised knees with 20-year followup radiographs (average radiographic followup, 20.4 years; range, 20.0–21.8 years), four had incomplete radiolucent lines on radiographic evaluation. Femoral radiolucencies consisted of two cases in Zone 1, one in Zone 2, and two in Zone 4. Lateral tibial radiolucencies consisted of two cases in Zone 1. On the AP projection of the tibia, there was one radiolucency each in Zone 1 and Zone 4. There were no radiolucent lines greater than 1 mm identified and no circumferential radiolucencies. We did not see any patellar radiolucencies on followup radiographs. In the same 12 nonrevised knees, osteolysis was evident in three patients (three knees), and in each case, the osteolysis totaled 1 cm2 or less and was limited to a single tibial zone. Two of these knees had shown no osteolysis at 15-year followup, and the osteolytic lesion in the third knee had not progressed since the 15-year followup radiograph. Of the six revised knees, four had osteolytic lesions (two in both femur and tibia, one in femur only, and one in tibia only) at the time of revision. Hence, for the entire cohort, 10 knees (10%) had osteolysis of at least 1 cm2 over the course of the 20-year followup.
Other complications in this series were relatively rare. There were no postoperative infections. Three patients required manipulation under anesthesia for poor postoperative ROM. One patient underwent repair of a quadriceps muscle rupture, related to a fall 4.5 months postoperatively. One patient developed a postoperative deep venous thrombosis, which was treated with anticoagulation. Patellar complications were also rare. Two knees had postoperative patellar instability; however, in both cases, this instability resolved without surgical intervention.
Discussion
Condylar-type TKA has been performed for more than 35 years, with relatively few studies [28, 31, 33] reporting findings at or near a mean of 20 years. Subsequently, the introduction of modular tibial trays and their associated potential benefits of a less destructive and more cost-effective revision have given surgeons an alternative. Although there have been a few studies reporting the results of TKA cohorts out to 20 years or longer [28, 31, 33], none have reported on modular tibial trays at this length of followup. To determine the long-term durability of a modular TKA construct, we prospectively followed a cohort of 75 patients with 101 PFC® cruciate-retaining prostheses at 5-year intervals. We previously reported 100% survivorship in this cohort at 10 years [20] and 91% at 15 years [29]. We therefore determined (1) survival, (2) the relationship of age and polyethylene thickness with revision, (3) function, and (4) radiographic lucencies and osteolysis in patients treated by a single surgeon with a posterior cruciate-retaining device with a modular tibial tray at a minimum 20-year followup.
We note the following limitations of our study. First, only 16 of the original 75 patients were alive. However, we had followup of the entire cohort of patients because of the prospective followup of the patients at 5-year intervals and no patients were lost to followup. We had a minimum 20-year radiographic evaluation in 77% of the living patients. However, the average radiographic followup was 19.5 years for all living patients and 14.7 years for the entire cohort. Our radiographic followup was more complete than most retrospective studies. Second, there was some selectivity of the implant in that seven other TKAs were performed during the time of the study, although this represented less than 10% of the total primary implants during the period of the index surgery. Third, this cohort of patients was older (average age at index surgery, 71.6 years). This may have accounted for lower revision and loosening rates than reported in other studies.
We found the PFC® prosthesis had high survivorship (Fig. 2). All revisions occurred greater than 10 years after the index procedure. All six failures were related to polyethylene wear leading to osteolysis and loosening. When comparing these results to the long-term results of this and other devices, survivorship was comparable (Table 3). This study also showed the potential benefit of tibial component modularity in terms of a less extensive revision because, in two of the six revisions, only liner exchange was necessary, and in another with polyethylene wear, the tibial component was retained at the time of femoral component revision. Hence, in 50% of revision cases that occurred over the 20-year followup interval, the modularity allowed for retention of the tibial component. The close-interval followup of this group of patients allowed for modularity to be optimally utilized.
We demonstrated a correlation with younger patient age and revision in this study. The average age at primary surgery of patients requiring revision was 63.5 years as compared to 71.6 years among nonrevised patients. The average age of the overall cohort at the time of the index procedure was 71.2 years, which was older than most contemporary TKA cohorts. Our cohort consisted entirely of polyethylene inserts gamma irradiated in air, a process associated with increased rates of polyethylene wear [10, 12, 37], particularly in combination with increased shelf life of polyethylene before implantation [11, 12, 18]. We were unable to determine the shelf life of the polyethylene utilized in this cohort of patients. The catastrophic early failure observed in PFC® knees of the same design implanted after 1991 [18] was not evident in this cohort, which included only TKAs performed just before that time interval. In 78 of 101 knees, an 8-mm insert was implanted. However, taking into account the tibial baseplate, the actual minimum thickness of an 8-mm insert implanted was 5.3 mm. All failures occurred in patients with 8-mm inserts, but we could not show this related to revision, given the small number of thicker (10-, 12.5-, and 15-mm) inserts implanted. The PFC® implant studied also employed a relatively flat, nonconforming surface, particularly the posterior-lipped insert. In spite of the potentially deleterious effects of thin [3, 9, 24, 38], flat [5, 19] inserts on polyethylene wear, clinically significant wear rates in this study were low and occurred late. However, the low wear rates may have also been a reflection of an elderly, less active cohort. This was further evidenced by the fact that there were no new revisions after 15.6 years, which could be attributed to the aging and less active living cohort.
We found Knee Society clinical and function scores of 90 and 59, respectively. These clinical outcome measures were comparable to the 14- to 17-year followup of PFC® TKA reported by Rodricks et al. [32], as well as other long-term followups of TKA cohorts (Table 3) [6, 8, 21, 22, 25, 28, 31–34, 36].
Our results in terms of loosening are relatively comparable to those of the other series [6, 8, 21, 22, 25, 28, 31–34, 36]. Although the rates of osteolysis are not well defined in other studies, the prevalence of 10% in this study occurred late (after 10 years) and was associated with the knees requiring revision. We attributed the osteolysis to the first-generation modular capturing mechanism and thin gamma-irradiated-in-air polyethylene. Osteolysis has been recognized as the dominant mode of aseptic failure in TKA, particularly with modular designs [12, 18]. Because osteolysis can be asymptomatic for a relatively long period of time, close followup, especially after 10 years, is warranted.
This study should serve as a comparison for other designs utilized during this time, as well as for designs with better capturing mechanisms and better bearing surface materials (ie, polyethylene gamma irradiated in an inert environment and crosslinked polyethylene). In addition, Kaplan-Meier survivorship with patient survival as an end point (calculated by age at the time of surgery) demonstrated the need to follow younger patient cohorts to have a relatively high number of patients available for review at 20 years (Fig. 3).
References
Amstutz HC, Thomas BJ, Jinnah R, Kim W, Grogan T, Yale C. Treatment of primary osteoarthritis of the hip: a comparison of total joint and surface replacement arthroplasty. J Bone Joint Surg Am. 1984;66:228–241.
Babis GC, Trousdale RT, Morrey BF. The effectiveness of tibial insert exchange in revision total knee arthroplasty. J Bone Joint Surg Am. 2002;84:64–68.
Bartel DL, Bicknell VL, Wright TM. The effect of conformity, thickness, and material on stresses in ultra-high molecular weight components of total joint replacement. J Bone Joint Surg Am. 1986;68:1041–1051.
Bellamy N, Buchanan WW, Goldsmith CH, Campbell J, Stitt LW. Validation study of WOMAC: a health status instrument for measuring clinically important patient relevant outcomes to antirheumatic drug therapy in patients with osteoarthritis of the hip or knee. J Rheumatol. 1988;15:1833–1840.
Benjamin J, Szivek J, Dersam G, Perselin S, Johnson R. Linear and volumetric wear of tibial inserts in posterior cruciate-retaining knee arthroplasties. Clin Orthop Relat Res. 2001;392:131–138.
Buechel FF Sr. Long-term followup after mobile-bearing total knee replacement. Clin Orthop Relat Res. 2002;404:40–50.
Callaghan JJ, Reynolds ER, Ting NT, Goetz DD, Clohisy JC, Maloney WJ. Liner exchange and bone grafting: rare option to treat wear & lysis of stable TKAs. Clin Orthop Relat Res. 2011;469:154–159.
Callaghan JJ, Wells CW, Liu SS, Goetz DD, Johnston RC. Cemented rotating-platform total knee replacement: a concise followup, at a minimum of twenty years, of a previous report. J Bone Joint Surg Am. 2010;82:705–711.
Collier JP, Mayor MB, McNamara JL, Surprenant VA, Jensen FE. Analysis of the failure of 122 polyethylene inserts from uncemented tibial knee components. Clin Orthop Relat Res. 1991;273:232–242.
Collier JP, Sperling DK, Currier JH, Sutula LC, Saum KA, Mayor MB. Impact of gamma sterilization on clinical performance of polyethylene in the knee. J Arthroplasty.1996;11:377–389.
Collier MB, Engh CA Jr, Engh GA. Shelf age of the polyethylene tibial component and outcome of unicondylar knee arthroplasty. J Bone Joint Surg Am. 2004;86:763–769.
Collier MB, Engh CA Jr, McAuley JP, Engh GA, McAuley JP, Engh GA. Factors associated with the loss of thickness of polyethylene tibial bearings after total knee arthroplasty. J Bone Joint Surgery Am. 2007;89:1306–1314.
Conditt MA, Stein JA, Noble PC. Factors affecting the severity of backside wear of modular tibial inserts. J Bone Joint Surgery Am. 2004;86:305–311.
Engh GA, Dwyer KA, Hanes CK. Polyethylene wear of metal-backed tibial components in total and unicompartmental knee prostheses. J Bone Joint Surgery Br. 1992;74:9–17.
Engh GA, Koralewicz LM, Pereles TR. Clinical results of modular polyethylene insert exchange with retention of total knee arthroplasty components. J Bone Joint Surg Am. 2000;82:516–523.
Engh GA, Lounicia S, Rao AR, Collier MB. In vivo deterioration of tibial baseplate locking mechanisms in contemporary modular total knee components. J Bone Joint Surg Am. 2001;83:1660–1665.
Ewald FC. The Knee Society total knee arthroplasty roentgenographic evaluation and scoring system. Clin Orthop Relat Res. 1989;248:9–12.
Fehring TK, Murphy JA, Hayes TD, Roberts DW, Pomeroy DL, Griffin WL. Factors influencing wear and osteolysis in press-fit condylar modular total knee replacements. Clin Orthop Relat Res. 2004;428:40–50.
Feng EL, Stulberg SD, Wixson RL. Progressive subluxation and polyethylene wear in total knee replacements with flat articular surfaces. Clin Orthop Relat Res 1994;299:60–71.
Fetzer GB, Callaghan JJ, Templeton JE, Goetz DD, Sullivan PM, Kelley SS. Posterior cruciate-retaining modular total knee arthroplasty: a 9- to 12-year followup investigation. J Arthroplasty. 2002;17:961–966.
Gill GS, Joshi AB. Long-term results of kinematic condylar knee replacement: an analysis of 404 knees. J Bone Joint Surg Br. 2001;83:355–358.
Gill GS, Joshi AB, Mills DM. Total condylar knee arthroplasty: 16- to 21-year results. Clin Orthop Relat Res. 1999;367:210–215.
Griffin WL, Scott RD, Dalury DF, Mahoney OM, Chiavetta JB, Odum SM. Modular insert exchange in knee arthroplasty for treatment of wear and osteolysis. Clin Orthop Relat Res. 2007;464:132–137.
Hirakawa K, Bauer TW, Stulberg BN, Wilde AH, Borden JS. Characterization of debris adjacent to failed knee implants of 3 different designs. Clin Orthop Relat Res. 1996;331:151–158.
Hooper G, Rothwell A, Frampton C. The low contact stress mobile-bearing total knee replacement: a prospective study with a minimum followup of ten years. J Bone Joint Surg Br. 2009;91:58–63.
Insall JN, Dorr LD, Scott RD, Scott WN. Rationale of the Knee Society clinical rating system. Clin Orthop Relat Res. 1989;248:13–14.
Kaplan EL, Meier P. Nonparametric estimation from incomplete observations. J Am Stat Assoc. 1958;53:457–481.
Ma HM, Lu YC, Ho FY, Huan CH. Long-term results of total condylar knee arthroplasty. J Arthroplasty. 2005;20:580–584.
Malin AS, Callaghan JJ, Bozic KJ, Liu SS, Goetz DD, Sullivan N, Kelley SS. Routine surveillance of modular PFC TKA shows increasing failures after 10 years. Clin Orthop Relat Res. 2010;468:2469–2476.
Parks NL, Engh GA, Topoleski LD, Emperado J. The Coventry Award. Modular tibial insert micromotion: a concern with contemporary knee implants. Clin Orthop Relat Res. 1998;356:10–15.
Pavone V, Boettner F, Fickert S, Sculco TP. Total condylar knee arthroplasty: a long-term followup. Clin Orthop Relat Res. 2001;388:18–25.
Rodricks DJ, Patil S, Pulido P, Colwell CW. Press-fit condylar design total knee arthroplasty: fourteen to seventeen-year followup. J Bone Joint Surg Am. 2007;89:89–95.
Rodriguez JA, Bhende H, Ranawat CS. Total condylar knee replacement: a 20-year followup study. Clin Orthop Relat Res. 2001;388:10–17.
Sextro GS, Berry DJ, Rand JA. Total knee arthroplasty using cruciate-retaining kinematic condylar prosthesis. Clin Orthop Relat Res. 2001;388:33–40.
Tegner Y, Lysholm J. Rating systems in the evaluation of knee ligament injuries. Clin Orthop Relat Res. 1985;198:43–49.
Weir DJ, Moran CG, Pinder IM. Kinematic condylar total knee arthroplasty: 14-year survivorship analysis of 208 consecutive cases. J Bone Joint Surg Br. 1996;78:907–911.
White SE, Paxson RD, Tanner MG, Whiteside LA. Effects of sterilization on wear in total knee arthroplasty. Clin Orthop Relat Res. 1996;331:164–171.
Williams IR, Mayor MB, Collier JP. The impact of sterilization method on wear in knee arthroplasty. Clin Orthop Relat Res. 1998;356:170–180.
Acknowledgments
The authors thank Steve S. Liu MD, Yubo Gao PhD, and Rhonda Chalus RN for their help in completion of this study.
Author information
Authors and Affiliations
Corresponding author
Additional information
One of the authors (JJC) has or may receive payments or benefits, in any 1 year, an amount in excess of $1,000,000 from DePuy Orthopaedics, Inc (Warsaw, IN, USA) and, in any 1 year, an amount in excess of $1000 from Lippincott Williams & Wilkins (Philadelphia, PA, USA).
Clinical Orthopaedics and Related Research neither advocates nor endorses the use of any treatment, drug, or device. Readers are encouraged to always seek additional information, including FDA approval status, of any drug or device before clinical use.
Each author certifies that his institution has approved the reporting of these cases, that all investigations were conducted in conformity with ethical principles of research, and that informed consent for participation was obtained.
This work was performed at the University of Iowa (Iowa City, IA, USA) and Des Moines Orthopaedic Surgeons (West Des Moines, IA, USA).
About this article
Cite this article
Callaghan, J.J., Beckert, M.W., Hennessy, D.W. et al. Durability of a Cruciate-retaining TKA With Modular Tibial Trays at 20 Years. Clin Orthop Relat Res 471, 109–117 (2013). https://doi.org/10.1007/s11999-012-2401-9
Published:
Issue Date:
DOI: https://doi.org/10.1007/s11999-012-2401-9