Introduction

The development of Brain-Interface (BI) Footnote 1 technology is a relatively young research field that has grown substantially over the last decade and continues to attract new researchers from multiple disciplines. The aim of this research is to develop an effective and reliable machine interface that is controlled by signals measured directly from a person’s brain.107,125,209

This paper presents the results of a comprehensive survey of all BI technology designs published prior to January 1, 2006. The first objective of this study was to gather and classify BI technology design information in order to (1) determine which designs are directly comparable and (2) demonstrate that meta-analysis is possible and valuable to the field. The authors were motivated by two factors. First, researchers have identified a critical need for objective methods to compare BI technologies.108,125,199,208,209 However, direct comparison of technologies assumes individuals can identify which designs are directly comparable. Currently, this does not seem to be the case, as research groups have had difficulty interpreting the myriad of reported technology designs and determining which designs can be compared.108,199,208 This situation is not surprising given the diversity of perspectives and language used by the professions that comprise this field (including, neuroscience, psychology, engineering, computer science, assistive technology, rehabilitation and other technical and health-care disciplines). As direct, objective comparison is necessary for cross-group validation of findings (the only method to challenge reported performance results), the ability to identify comparable technologies is critical. The second motivating factor was the recent publication of a general framework for high-level BI design and related alterations and extensions.123,125,129 Prior to its publication, only simple classification schemes had been proposed, each based primarily on the neurological phenomenon used to operate the interface, (e.g., P300 response to an oddball stimulus) or the particular sensor technology used (e.g., EEG, ECoG or implanted microelectrode arrays).107,200,209 As such, these schemes lacked the necessary depth and breadth to classify the range of design details seen in the literature. The referential models and taxonomy provided in the new framework allowed us to describe all existing BI designs using a common language. Within this context, we were able to identify which technologies existed and which system and subsystem designs could be directly compared. (For readers unfamiliar with this framework, the principle models and taxonomy are reported in the “Review of Design Framework" section.) We want to emphasize that this survey strictly focused on BI technology design and did not incorporate any aspect of performance evaluation.

The second objective of this project was to provide a detailed, historical reference where researchers could (1) determine which design approaches have been proposed, (2) identify design approaches that have received little or no attention, (3) determine which applications areas (users, activities, environments) the existing technology has been designed for, and (4) characterize how well designs have been reported in the literature.

The methods used to conduct this study are detailed in “Methods” Section and the results are presented in “Results and Discussion” Section. Prior to presenting this material, we provide a short review of the design framework that was used as a basis for this survey.

Review of Design Framework

The functional models and taxonomy of the BI design framework proposed by Mason et al.123,125,129 are presented in Fig. 1 and Table 1.

FIGURE 1.
figure 1

Functional models of BI technology design: (a) model of person with a functional limitation able to perform desired activity in the environment with an AT. Icons on the right represent appliances, people or objects in the environment; (b) functional model of 2-component BI AT; (c) functional model of 3-component BI AT, (d) model of an (endogenous) BI Transducer illustrating the series of components used to translate spontaneous brain activity into control signals (note, two other models related to exogenous and modulated-response transducer architectures exist and can be found at Ref. 123); (e) functional model of 2-component demonstration system, where some form of Demo Device (e.g., a cursor or robotic arm) is used to demonstrate transducer control; (f) functional model of 3-component demonstration system.

TABLE 1. BI design taxonomy proposed by Mason et al.123,125,129

The framework is based on the model of assistive technology drawn in Fig. 1a. In this model, a person with a functional limitation is depicted on the left. This person desires to perform an activity in his or her environment, which may be to move their body or interact with objects, appliances, physical structures, or other people in the environment. The person’s functional limitation results in a gap (shown as the white “ability gap”) between the person’s abilities and those abilities required by the activity. The resulting inability to perform the desired activity is referred to as a disability. An assistive (or augmentative) technology (AT) can provide the additional functionality that the person requires to bridge the ability gap and perform the desired activity. Like other AT, BI AT provides additional functionality to a target population in order to perform specific (target) activities in certain (target) environments.

Within this context, BI AT can be modeled as a series of components: a BI Transducer and an Assistive Device (as shown in Fig. 1b) or with a Control Interface (as depicted in Fig. 1c). Each of these design components can be represented by a more detailed model. For example, the BI Transducer can be modeled as the series of components in Fig. 1d that transforms brain activity into a control signal. According to the framework documentation, these models can represent all possible designs. However, readers may be more familiar with other functional terms, such as noise reduction, filtering, preprocessing or spike detection/sorting (which are all forms of feature extraction) and classifier or “decoding method” – or something similar – (which are both terms that describe a specific class of Feature Translator). In this work, we will use the generic component names proposed by the framework to avoid bias toward any design approach.

For many studies, researchers do not employ a full AT model, but use some reduced system in order to demonstrate basic control (as depicted in the Demonstration System models of Fig. 1e and f). Others test only BI Transducer designs.

All of these systems, components and subcomponents can be described in terms of design attributes. The taxonomy listed in Table 1, outlines the principal design attributes proposed by Mason et al.

Methods

The designs selected for this survey were chosen from journal and conference papers found in Pubmed, IEEE Xplore or ISI Web of Science databases that met the following criteria: (1) the keywords BI, BCI, BMI, DBI, ABI, or the corresponding descriptions, appeared in their title, abstract or keyword list; and (2) the paper described one or more of the principal design components listed in Table 1 (minimum technology content was a BI Transducer). Papers which presented tutorials, descriptions of new bio-recording technologies (such as new sensor and amplifier designs), overviews of research activities, or neuroanatomy and neurophysiology discussions were not considered. In addition, well-known historical BI-related works such as Vidal201 and Sutter190 were also included. Each paper was assigned to a research group using the affiliation of the first author at the time of publication. A research group was defined as the individuals working within a single laboratory or institution or in a group of laboratories or institutions working on a single project. (Refer to Appendix A for a list of research groups.) Papers from a research group that described previously reported designs were not included to avoid counting duplicates. For example, if a research group published two papers that reported different tests with the same technology design, the later paper was not included in our test set.

As we were not evaluating the quality of these designs, that is, how well they worked, but rather collecting design ideas or strategies, we weighted all reported designs equally. Thus designs from journal papers were not stressed more than conference papers in our frequency analysis. Likewise, designs from established laboratories with multiple publications did not carry more weight than those from new researchers with few publications.

A classification template based on the BI design taxonomy listed in Table 1 was constructed for this survey. We extended the taxonomy by adding sub-categories to each design attribute. The attribute sub-classes, as these sub-categories were named, were initially proposed by our team based on our experience with BI technology and related human–machine interaction research and then refined after an initial pass through the selected papers. To illustrate the concept of attribute sub-class, Table 2 depicts the attribute sub-classes for BI TransducerBio-recording Technology attribute. This example also defines the text style used in this paper to identify attribute names. Specifically, design attribute categories such as BI Transducer, are written in a bold, italic, sans-serif, font, design attribute names (like Bio-recording Technology) are written in a serif, bold type and attribute sub-classes (e.g., ECoG) are represented in serif, bold-italic type. The full classification template with all the design attributes and attribute sub-classes is listed in the left two columns of Table 3.

TABLE 2. Example of attribute sub-classes for the BI-TransducerBio-recording Technology design attribute.
TABLE 3. Results of BI technology design survey. Papers from the same research group are grouped within square brackets, [...], and conference papers are in normal type, e.g., [17], while journal papers are in bold type, e.g., [17]. Design attribute names are written in bold type and attribute sub-classes will be represented in bold-italic type. The column on the right summarizes the number of research groups (g), journal papers (jp) and conference papers (cp) within each attribute sub-class.

Each paper was reviewed by at least two individuals from a team of five reviewers which included the first four authors and a contracted medical student. Reviewers categorized each design by selecting the closest attribute sub-class for each design attribute defined in the classification template. Design attributes that were not reported were recorded as such. For the papers that reported multiple designs or design options, multiple classifications were recorded per paper. The reviewers classified designs based only on what was reported in each paper or referenced works. No personal knowledge of an author’s work was used in the classification.

Results and Discussion

The detailed classification results of the survey are summarized in Table 3. The references listed in the third column represent all the papers that contain designs with the listed attribute sub-classes. For instance, all the references listed beside Target Activitycommunication with people (synth. speech, text or drawing) are papers with technology designs aimed to help individuals communicate. As such, this table provides a comprehensive historical reference of all BI designs up to January 2006, which can be an extremely valuable resource for technology designers or new researchers entering the field. For example, if one is interested in finding all the BI technology designs that have used movement related potentials in EEG generated by movement imagery, the references to relevant journal and conference papers can be found under BI TransducerNeurological Phenomenonmovement-related potentials from imagined movements. Alternatively, if one wants an existing virtual keyboard design, they can find the Control Interface designs that emulate a keyboard under Control Interface-Interface Paradigms.

In Table 3, papers from the same research group are collected within square brackets separated by a semicolon. Conference papers are displayed in normal type, e.g., [17] while journal papers are shown in bold type, e.g. [17]. The column on the right summarizes the number of research groups, journal papers and conference papers within each attribute sub-class. As this work covers technology design and not technology evaluation, there has been no attempt to rate how reliably or accurately the published designs function; that is an issue for manuscript reviewers and readers to interpret once the design details and target applications are understood.

There are many interesting observations or comments which could be made from the results in Table 3. We will first focus our discussion on which BI designs can be compared to each other. Then we will specifically look at what technology designs have been proposed, which target applications have been pursued, which design approaches have received little or no attention and, finally, how well designs are reported. Each of these topics is presented in a separate subsection below.

Which Designs Are Comparable?

The main goal of this work was to identify which BI designs can be directly compared to each other. Researchers may choose to compare BI technology designs in terms of many potential factors; Accuracy, response time, throughput, the amount of mental attention required, ease of use or other usability factors such as, physical size and sensor application requirements are a few factors of interest to researchers. Since only technology that performs the same function can be directly compared, we will first discuss comparability of function.

Within the context of the chosen framework, technology comparisons can be made at various levels. BI Transducer designs that produce the same output are directly comparable (regardless of other factors such as bio-recording technology, neurological phenomenon or feature extraction and feature translation methods). For example, all transducer designs producing the same type of 2-state, discrete output can be compared. This principle also applies for direct comparisons of BI technology to non-BI technology like chin switches or eye tracking systems. Likewise, BI Transducer and Control Interface (CI) combinations producing the same output are directly comparable, and could be compared with BI Transducers that produce similar high-dimensional outputs. Systems with Assistive Devices (AD) that perform the same function are directly comparable, where function may be strictly defined as communication via a text display for example, or more broadly defined as communication (or information transfer) in general whether through a text display, drawing tablet, or speech synthesizer for instance. Now what may not be obvious to all readers is that BI technologies can only be compared if they produce the same output (i.e., perform the same function at some level). The focus here is on output and not the other design attributes. As stated above, this could be the BI Transducer output, the Control Interface output, the Demo Device output or the Assistive Device output. The other design attributes are not important in determining comparability (assuming that the designs are suitable for the target application-an issue that is discussed in the next section). As an example, even though many BI researchers tend to group technology by the bio-recording technology or sensor design (e.g., EEG, ECoG, implanted arrays, or more generally as invasive/non-invasive), the bio-recording technology is irrelevant in determining if two technologies can be compared at a functional level. For instance, an EEG-based transducer can be directly compared with one based on an implanted microelectrode array if the signal processing produces the same type of output.

Within our results we encountered certain types of transducer outputs that are not seen in common interface technologies like keyboards, mice and joysticks. These differences are important to this discussion so we will elaborate on them here. In general, there are three types of transducer output: discrete, continuous and spatial reference – these categories apply to all interface technologies. For brain interfaces, we have recognized that researchers use three types of discrete output, two types of continuous output and ranges of spatial reference output, all of which are unique and thus cannot be directly compared. The first type of discrete output we encountered was in line with our expectation of discrete transducer output. That is, the output signal had one or more states that corresponded to when the user is intentionally controlling the transducer plus a “do nothing” or “idle” state that corresponded to when the person was in a No Control (NC) state.22,128 This is the behavior one would expect from standard interface devices such as keyboards and switches. We created the attribute sub-class discrete (with 1 NC state) to refer to this type of output signal. The second type of transducer produced a discrete output that corresponded only to Intentional Control (IC) states, i.e., there was no output state that corresponded to the user’s NC state. These transducers assumed that another mechanism would handle the time periods when a user was in an NC state. This type of output is not seen in normal interface devices. We referred to this type of output signal as discrete (all IC states). The third class of discrete transducer produced an “unknown state” in the output to represent periods where the transducer lacked enough evidence (or confidence) to generate an output state classification corresponding to an IC or NC state. We used the attribute sub-class discrete (with 1 unknown state) if there was no NC output state or discrete (with 1 unknown and 1 NC state) if there was an NC output state. Note, none of the later class was found in the survey set.

For transducers that produce continuous control, there were two general classes of output signal: continuous control relative to a fixed reference (like a volume control on a stereo or a joystick) and continuous control relative to previous values (like the wheel on a wheel mouse or the two-dimensional mouse position control). The first class, which we called continuous (fixed reference), produces a sequence of absolute continuous amplitude values while the second, which we referred to as relative continuous (no reference), produces a sequence of relative amplitude changes.

The transducers that produced a spatial reference (similar to eye tracking, a touchscreen or stylus mechanism), either produced a small number of selectable regions (≤ 5) or a large number of regions (> 10). As such, we arbitrarily grouped these types of transducers into devices that could reference five or less regions and those that could reference more than five regions.

A summary of the number of journal papers, conference papers and research groups related to comparable BI Transducer designs is presented in Fig. 2. From this figure we can see that 27 groups produced 68 designs for 2-state, discrete (all IC states) – 41 of these designs were reported in journals and 27 in conference proceedings. The specific papers are cited in Table 3 under BI TransducerOutputdiscrete (all IC states)-2-state. All designs in this category are directly comparable. As another example, all 25 transducer designs that produced a 1-D relative continuous (no reference) output are directly comparable. Direct comparison between the transducer designs in the different output categories is either not possible (e.g., discrete versus continuous versus spatial reference) or difficult to test and interpret, such as comparing the designs classified as 3-state discrete (with 1 unknown state) and 2-state discrete (all IC states).

FIGURE 2.
figure 2

Histogram of BI TransducerOutput grouped by output category. Refer to text in “Which Designs Are Comparable?” Section for definition of terms.

When we switched focus to Control Interface outputs, we observed that all 12 of the designs surveyed produced discrete symbols (alphanumeric characters or commands) and thus are directly comparable even though the underlying transducer, interface paradigm or CI translation methods are different. We also see that within the Control InterfaceInterface Paradigm attribute, five designs (36%) were based on a single, direct selection (as in Farwell and Donchin’s P300 system menu48), three (22%) were single, indirect (pointer-based) selections (like a point and dwell interface), two (14%) were multiple, direct selections (e.g., the multi-level menu system proposed by Scherer et al.179) and four (28%) were multiple, indirect (pointer-based) selections (like the binary keyboard in Birbaumer’s Thought Translation Device107).

The Full AT system designs reported in Table 3 made use of the following Assistive Devices (AD): 11 papers (55%) from seven groups reported a visual text display, three papers (17%) from three groups had a speech synthesizer, two papers (11%) from one group presented an appliance interface/remote control, and six papers (17%) from one group reported various designs using a limb control neuroprosthetic. The designs within each of these AD categories can be compared and possibly the text display systems could be compared with those that employed a speech synthesizer on a higher-level study of communication ability.

What Technology Designs Have Been Proposed?

From the survey data we can determine what technology designs have been proposed up to January 2006. Due to space limitations, we will restrict the discussion on overall trends with specific examples.

From Fig. 3 we can see that the majority of publications (56%) reported only a BI Transducer design. Of the remainder, 36% reported a Demonstration System to demonstrate control with feedback, and only 8% have proposed Full AT solutions – 11 (5%) of which used a 3-component system (Transducer-CI-AD combination) and three (3%) used a 2-component system (Transducer-AD). Refer to Fig. 1 for system definitions if required.

FIGURE 3.
figure 3

Distribution of BI Technology System Design Models.

We can also see from Fig. 4a and b, 166 endogenous transducer designs (those driven from internally generated brain activity) were reported by 46 groups. This corresponded to 81% of the BI TransducerTransducer Design Model attribute sub-class. The remainder was 35 exogenous transducer designs (those driven from external stimuli to evoke a brain response) from 18 groups contributing 17%, and four modulated-response transducers designs (those designs where the user modulated their brain response to external stimuli) from two groups representing the last 2%. None of these design architectures is inherently better than the others. Each has its own merits and the choice of architecture depends on the target application and the type of interactive control required (e.g., direct selection, object positioning or spatial navigation). For instance, exogenous transducers have the benefit that no user training is required43 and they are the only design architecture that offers direct spatial referencing, which is a powerful interaction modality.129 The down side is they require a constant commitment of a sensory modality200 and repetitive sensory stimulation (where the long-term tolerance and safety are not yet established). They also must be paired with a Control Interface (such as a menu system) in order to produce a discrete or continuous output which may not be appropriate for many target applications. Endogenous transducers can produce discrete or continuous control and do not require an external sensory stimulator. Both are desirable features for many applications. However, the use of spontaneous components of the EEG entails the need to recognize when control is intended and when it is not. These designs also require a certain degree of user training which can range from a few sessions to many sessions spanning several weeks. Certain target applications, such as those with dedicated users, may tolerate long training periods, others, like short term use for out patient rehabilitation, will not. The design architecture used in modulated-response transducers provides discrete or continuous control but requires an external stimulator to produce a steady state response in the user’s brain which the user modulates to generate control. As these types of output can be provided by endogenous designs without an external stimulator, the benefit of this approach remains to be demonstrated.

FIGURE 4.
figure 4

Distribution of (a) BI Transducer Design Models and (b) bio-recording technologies.

As displayed in Fig. 4b, the BI TransducerBio-recording Technology attribute is heavily biased towards EEG (83%) with only 5% on ECoG, 12% on implanted microelectrode arrays, and fraction of 1% on functional Near Infrared. (fNI). When we analyzed this data over time as in Fig. 5, we get a clear view of the predominance of EEG in the context of the field’s relatively short history. The relatively large proportion of EEG-based designs compared to more invasive designs and fNI does not represent superior quality of control but rather reflects the relatively low barrier to entry for this older technology (e.g., bio-recording equipment is commercially produced, costs are reasonable and consenting subjects are readily available). The recent increase in more invasive methods, which seem to promise higher-dimensional control, are expected to continue as these technologies are proven to be stable, tolerant and safe and eventually become commercially available. The presence of fNI illustrates how future BI solutions may not be based solely on the measurement of electrical signals, but may include other signals such as metabolic, magnetic, chemical, thermal or mechanical changes related to brain activity as discussed in Ref. 199.

FIGURE 5.
figure 5

Histogram of bio-recording technoloy over time.

As with the design architectures discussed above, it is impossible to say that a bio-recording approach is superior to another without considering the target application. For example, some researchers may question the practical value of an EEG-based neuroprosthetic that uses external FES device to restores hand grasp function. We would concur if the target application was for permanent functional replacement in a parapalegic with a complete spinal lesion, but not so if the application was to use BI technology to assist the rehabilitation of out patients who were recovering from a stroke.108 From our perspective, EEG or other externally applied technologies seem the most appropriate choice when the target application is for short-term use (such as the rehabilitation example given above) or where the user, their medical insurer and/or government will not tolerate the costs and risks of implantation. Implanted technologies seem the most appropriate for individuals who require permanent or prolonged functional replacement or augmentation. But there are always special cases and the opinion that really matters is not ours but the end user, their insurer, governments and society in general. This later point was emphasized in many discussions at the BCI2005 workshop on clinical issues and applications summarized in Ref. 108. This group viewed the appropriateness of a bio-recording technology as a cost–benefit analysis; one that needs to be performed on a case by case basis and will probably change in the future. To illustrate, the costs of implantation (equipment, surgery, user impact, and social impact) are currently large, but imagine if the cost of a technology reduces to a relatively inexpensive day surgery with minimal recovery and the technology is proven to be stable and safe for one’s life time. This type of change would shift what people consider appropriate and ethical. Note, our comments assume that researchers will be able to resolve the outstanding technical issues identified by the BI community.199,211 For EEG, two key issues are user discomfort and application time related to “wet” electrodes. For implanted technology the main issues relate to the stability, tolerance and safety of the recording and transmission electronics.

When we analyzed BI TransducerOutput versus Neurological Phenomenon (and the related Bio-recording Technology), as depicted in Fig. 6, we can see that overall the most predominant endogenous transducer designs are discrete (all IC states) based on movement related activity (whether temporal movement-related potentials (MRPs), shown as MRP–IM and MRP–AM, or power changes in particular frequency bands, shown as ERD/ERS or rhythm power). Since these observations are biased towards EEG as there were significantly more EEG-based designs in our review set, we can alternatively look within each bio-recording category. Here we see that the EEG, ECoG and fNI designs were predominantly discrete output (which is relatively easier to manifest and qualify) and the few continuous control designs predominantly offered only one-dimensional control. Within the Implanted category, the technology is more uniformly distributed across output type. A few of these papers claimed their continuous control technology produced an absolute continuous value (i.e., a continuous (fixed reference) output). However, this type of control is more difficult to manifest, and because many authors did not specifically demonstrate this ability, it was impossible to substantiate the authors’ claims. Only the exogenous transducers produced a spatial reference output as seen in Fig. 7. These designs were predominately related to a P300 response or steady-state visual evoked potential (SSVEP) response in EEG. Although we cannot predict which neurological phenomenon will be most suitable for a person or activity in general, characteristics such as response time, the customization (system training) process and the amount of user training required for a design (attributes not included in the current design taxonomy) all affect the practical usefulness of an approach. For instance, one could not use the current fNI system to control a vehicle as their response time is in the order of 10s of seconds.

FIGURE 6.
figure 6

Distribution of BI Transducer output for endogenous transducers in relation to bio-recording technology and neurological phenomena. In this figure, NIR = Near Infrared, implanted = electrode arrays implanted in the cortex, MRP-IM = Movement-related potentials – imagined movement, MRP-AM = movement-related potentials – attempted movement, cog. task = cognitive task, SCP = slow cortical potentials, ERD/ERD = event-related desynchronization/event-related synchronization, firing rate-AM = cellular firing rate related to attempted movement, firing rate-CT = cellular firing rate related to cognitive task, IC = Intentional Control, NC = No Control state. All other terms are defined in the text.

FIGURE 7.
figure 7

Distribution of BI Transducer output for exogenous transducers in relation to bio-recording technology and neurological phenomena. In this figure, NIR = Near Infrared, implanted = electrode arrays implanted in the cortex, P300 = P300 response, SSVEP = steady-state visual evoked potential, VEP = visual evoked potential, AEP = auditory evoked potential, firing rate-stim = cellular firing rate related to external stimulation, IC = Intentional Control, NC = No Control state. All other terms are defined in the text.

The majority of published Control Interface designs, were either multiple-selection virtual keyboards driven by a one-dimensional pointer (36%) like the binary keyboard in Birbaumer’s TTD, or single-selection menu systems with a spatial reference input (27%) like the menu used in Farwell and Donchin’s P300 system. All produced an N-state, discrete output (translating continuous control or spatial reference inputs into N-state, discrete output values). There is a wide variation in Control Interface design and most designs do not seem based on existing interface design principles.51,147,189 This issue requires further attention. Also, only a few research groups have proposed a method for an On mechanism33,86 or discuss the use of error correction158,194 – both of which are important issues for real world BI AT.

Within the Demo Device category, we see in Fig. 8 that the majority of papers employed a computer monitor with a range of virtual objects under control: 40% with cursor (eight groups), 1% with cursor + discrete functions (similar to a mouse) (1 group), 23% with virtual world (seven groups), 12% with discrete indicator (four groups), 10% with menu system (six groups), 9% with level indicator (three groups), and 3% with some other type of device (one group). The rest presented some form of physical device: 27% with vehicle simulator (one group), 18% with model vehicle (one group), 18% with robotic arm/hand (two groups), and 36% with some other type of device (three groups). The dominance of cursor control as a demonstration was not unexpected as many groups are looking to develop a computer interface. The observed Demo DevicesType of Control attribute broke down into 47% direct selection, 8% continuous value adjustment, 1% indirect selection and 44% object positioning. We found the breakdown of object positioning informative. We had identified four main types of object positioning, defined with increasing difficulty: (1) move object from fixed starting point(s) to 1 target per trial (in 1, 2 or 3D), (2) move object from fixed starting point(s) to 1 of N possible targets, (3) continuous positioning (path not important), and (4) continuous path following/ navigation /drawing (where path is important). The first type, which was used in the majority (50%) of these types of demonstration systems, exhibits basic control, but the reported error rates are optimistically exaggerated as there is no potential for moving to a wrong target. Also there is no practical use for this type of interaction, thus these types of demonstrations should be limited to initial proof of concept studies.

FIGURE 8.
figure 8

Distribution of demo device type.

To conclude this discussion, we will say a few words on which of these technologies are compatible. For components to be compatible, the input of one component must match the output of another. As an example, all five of the Control Interface designs require an input signal of type 1-D relative continuous (no reference). These are compatible with the 25 BI Transducer designs that produce the same type of output. Many other examples can be extracted from Table 3. This ability to identify compatible technology has the potential to increase cross-fertilization of technology between research groups.

Which Target Applications Have Been Pursued?

Of the papers that reported Target Population, 99% of the designs were targeted at individuals with some form of severe motor disabilities, some reports being more specific, e.g., 56% were targeted at individuals with paralysis (40% specified full paralysis, 2%, partial paralysis and 14% unspecified). Less than 1% of designs were aimed at able-bodied individuals (viz., pilots).

The papers that reported a specific Target Activity were focused on communication with people (15%), control of appliances (15%), control of paralyzed limbs (12%), control of vehicle (1%), personal mobility (1%) or control of objects in virtual reality (< 1%). The majority (56%) of papers were particularly vague about their target activity. Most papers simply included wording such as “BCIs are a device for communication and/or control” and did not indicate a specific activity. Although we have recorded all these papers as if they are targeting a general solution, we suspect that several of the authors have not given this aspect much consideration and treat the development of BI technology as an academic challenge with no real connection to their target community. This is a vulnerable position for this research community as funders have limited patience and may eventually withdraw support if real, practical solutions that address people’s needs do not begin to emerge.108,199

Which Design Approaches Have Received Little or No Attention?

Several of the attribute sub-classes shown in the above figures or tables had few or no observations. These approaches are potentially opportunities for technology advancement.

One of the most notable opportunities is the use of implanted technologies for exogenous transducers. Exogenous transducers produce a valuable mode of interaction: direct spatial referencing. Given the relative success with EEG-based exogenous transducers, one would expect greater accuracy and resolution from an implanted system as seen in Ref. 190. Combining this with a switch (discrete transducer) with NC support one could have a powerful point-and-click interface.

Within EEG-based exogenous transducers, there are many configurations of stimulation mechanism parameters that have not been tried. Alternative designs may provide increased response to a stimulus, resulting in higher recognition accuracy. The reader is directed to Ref. 129 for a more detailed discussion of stimulation mechanism configuration and examples.

Another important area that has not received much attention is the development of transducer designs that support No Control (NC) (the times when a user is not intentionally controlling the BI). This is a critical design feature as lack of NC support precludes self-paced (or asynchronous) AT operation, which is the most natural mode of interaction. Although the number of BI Transducer designs with NC support has grown modestly over recent years, more effort is required before we can claim to have useful, self-paced operation.128

In designs using field potentials, artifacts, particularly ocular, movement and muscular (electromyographic, EMG), can often disrupt or confuse a transducer. Given that this is a recognized problem in signals from externally applied sensors,199 it is surprising that less than 8% of the reported articles utilize some form of artifact processing and fewer actually demonstrate how well their artifact processing methodologies work. The lack of artifact processing or the use of a particular processing method is a major issue that may render a particular transducer design useless in real-world settings, and as such this issue deserves more attention from those working with externally applied sensors.

Although a few people have demonstrated 2-D or 3-D continuous control (where the movement path is not important), no one has reported continuous path following (where the path is important). This type of control is needed for drawing and navigation, which are powerful, highly desirable abilities, and should be attainable given the reported abilities of some of the 2- and 3-D continuous controllers. All that seems to be missing is for researchers with this technology to demonstrate this ability.

One last item would be the On mechanism. Only two groups have attempted to build an On mechanism into their AT. Much more work is needed in this area if the community is to realize useful, unsupervised assistive technology.

Many additional areas have not been pursued, and will probably be pursued only if the current approaches prove to be fruitful. Examples include (1) more complex Demo Devices and Control Interfaces, (2) targeting individuals with inconsistent motor control, such as cerebral palsy (CP) or multiple sclerosis MS, (3) targeting activities such as body control (posture, bladder), and (4) building BI Transducers that produce higher dimension outputs.

How Well Are Designs Reported?

The majority of the papers reviewed were found in journals and conference proceedings of a technical nature. This was not an unexpected result considering that much of the research in the BI area is still in an early technological development stage.

Overall, we were able to interpret the general technical characteristics of a design most reports, enough to categorize the designs using our classification template. However, many of the papers lacked significant detail on the signal processing methods that precludes replication of their work. This was not an issue of lack of common vocabulary and methods but rather a lack of full reporting. It is a serious scientific issue that the BI community needs to address quickly in order to establish and maintain rigor within this maturing field.

In general, the Target Application attributes (Target Population, Target Activity and Target Environment) were not well reported: 34% of Target Population, 17% of Target Activity, and 97% of Target Environment were not reported. The relatively poor reporting of targets is an important oversight in these works since the usefulness of a design cannot be interpreted or challenged without an explicit definition of the target application.

BI TransducerNC Support was also poorly reported – 88% of papers did not report this attribute. Most of these papers were testing BI TransducerOutputs in a synchronized test environment and as such overlooked the need to test and report how well their transducer designs functioned when a user was in the No Control (NC) state. Many may dismiss this as unimportant, but even in synchronized control environments one can not guarantee the user will always be in an Intentional Control (IC) state during the system-defined windows of control, especially in real-world (unsupervised) settings. As such, all BI Transducers should be evaluated and their response to NC reported in order to understand what types of errors can be expected from a proposed design.

Finally, there was confusion around the use of the terms “motor imagery” and “imagined movements” in the literature. From the surveyed papers, “motor imagery” or “imagined movements” was used on one hand to describe a type of mental activity in able-bodied subjects. In these cases, the subjects were likely using some form of motor imagery – although the experimental protocols rarely reported the specific controls (such as structured questionnaires) to ensure that the subjects were doing as requested. These terms were also used to describe a mental activity in individuals with severe movement disabilities, although with the lack of reported controls, it was possible that these people were actually using attempted movements instead of imagery. As attempted movements are different neurological mechanisms then imagining movement, it would help if the field could clarify the use of these terms and use appropriate controls to ensure subjects are using the neurological mechanism requested.

Conclusions

We have completed the first comprehensive survey of BI technology designs published up to January 2006. The results of this survey form a valuable, historical cross-reference from which we have discussed the following points: (1) which technologies are comparable, (2) which technology designs have been proposed, (3) the application areas (users, activities and environments) targeted by researchers, 4) the design approaches which have received little or no attention, which are possible opportunities for new technology, and 5) how well designs are reported. We have also demonstrated that this type of meta-analysis of BI design is possible and produces information that is valuable to the field.

This work directly addresses the need for methods to identify comparable technology. Like other AT, typical studies of BI technology cannot be evaluated using classical means with a “no treatment” control group due to factors such as high subject variability, small number of subjects, uncontrolled experimental factors (unbalanced subject groups); and functional limitations of the target population. As such, identifying superior technology can only be achieved by directly comparing BI technologies to each other within the same test environment. The presented results not only allow researchers to determine which BI technologies can be directly compared in this way, but it also illuminates which non-BI technologies may be compared as well.

In many ways, the discussion of the results presented is a sample of the type of analysis that can be performed. There are many other types of analysis that could be conducted, particularly analysis across multiple attributes, attribute development over time, and technology developed within a specific research group. A possible extension of this work may include an on-line design database that would allow the community to directly access these survey results and perform their own analysis.

One of our hopes is that this study will spawn further discussion of theoretical formalisms, leading to improved (and possibly multiple) models and taxonomies for BI technology. The attribute sub-classes defined in Table 3, for instance, represents a proposal of subcategories, not a final set, and we encourage others to revise or expand this initial set. For example, we have previously mentioned the opportunities to add design attributes related to customization, user training and response timing to the design taxonomy. Other examples include adding attribute sub-classes to more clearly delineate target populations or expanding the Feature Extraction and Feature Extraction design attributes to capture the increasing diversity of signal processing methods being reported.