Introduction

Pathogenic bacteria are considered to be serious threat to human health in developed, developing, and underdeveloped countries worldwide [1, 2]. Awareness is growing that aqueous environment, especially water, is susceptible to be contaminated by various bacteria and induces the outbreak of waterborne disease [3]. Evidences have shown that bacteria such as Escherichia coli O157:H7 (E. coli O157:H7), Staphylococcus aureus (S. aureus), Listeria monocytogenes (L. monocytogenes), and Pseudomonas aeruginosa [4,5,6,7] have been found in groundwater, surface water, and domestic sewage [8,9,10]. These harmful waterborne pathogens cause life-threatening infectious diseases including hemorrhagic colitis, hemolytic uremic syndrome, septicopyemia, and septicemia [11,12,13] via directly drinking of contaminated water or indirectly consumption of contaminated fruits and vegetables irrigated by contaminated water. A consensus has been reached that control of waterborne diseases can be greatly improved if bacterial infections could be timely detected and identified [14, 15]. As a result, developing of rapid and accurate detection methods with low cost and simplicity for simultaneous detecting multiple pathogenic bacteria in water is of great significance to reduce or avoid the potential risk of human infection or losing lives.

Generally, the most matured techniques for pathogenic bacteria detection are conventional culture- and molecular amplification-based methods [16, 17]. Although the applications of these methods have achieved impressive progresses for pathogens identification and analysis, the culture methods generally need lengthy culture at least 24 h to several days to obtain the final conclusive results, which lead to undesirable delays in completion of the analysis and life safety threat. Besides, the identification of pathogens is usually single target analysis with the specific culture medium, and different pathogens should be cultured independently. Molecular amplification methods including polymerase chain reaction [18] (PCR), real-time PCR [19] (RT-PCR), loop-mediated isothermal amplification [20,21,22] (LAMP), and recombinase polymerase amplification [23] (RPA) that can amplify specific regions of nucleic acids are still facing performance-limited issues such as false-positive amplification and cross contamination especially when facing with multiple pathogens. Besides, the most representative PCR and RT-PCR also require skilled technicians for careful handling of the specialized instruments and a certain level of expertise for the interpretation of final results, inhibiting the popularization of molecular methods. Immune principle-based methods have also been adopted for bacteria determination. Previous reported methods based on immunorecognitions [24, 25] needs complex interface construction and multistep immobilization, recognition, and washing. They are not user-friendly in terms of operation simplicity, sensitivity, and efficiency. Additionally, it is widely recognized that water samples are usually contaminated with different pathogenic bacteria. Current researches mainly focus on detecting only one kind of target bacteria, which cannot meet the demand of accurate estimating pathogen categories and concentrations in water [26, 27]. In this respect, design of new analytical strategies for overcoming abovementioned limitations and simultaneous monitoring multiple target pathogens is vital to acquire the evaluation information of water quality. Fortunately, surface-enhanced Raman scattering (SERS) is an ultrasensitive and noninvasive molecular spectroscopy technique [28,29,30]. With well-recognized characteristics of high sensitivity, high photo-stability and detection efficiency, low environmental interference, and less reagent consumption, SERS-assisted biosensors are ideal candidates for pathogenic bacteria detections [31,32,33]. More importantly, the narrow bands/fingerprint of Raman molecules offers SERS biosensing strategies a powerful multiplex screening capability and therefore is very suitable for simultaneous analyzing multiple analytes.

In 2015, a piece of breaking news about The drinkable book has greatly aroused the interest of public [34]. The book invented by “Theresa Dankovich” is actually a filtration kit equipped with a filter box and a book with pages that offer both sanitation advice and function as filter sheets. The waterborne pathogenic bacteria can be attached and adsorbed onto each utilized book sheet, while the water can be filtrated through the paper page, leaving it safe for drink. Inspired by the idea of drinkable book and taken above into consideration, in this study, we have concisely engineered a portable double-layer filtration device for bacteria detection and water quality control. The two layered filtration membrane can take effect to remove the interferants and retain the target pathogens for analysis. Different SERS-Tags as Raman signal reporting units were adopted and incubated with water samples for fast recognition, separation, concentration, and identification of multiple waterborne pathogens including E. coli O157:H7, S. aureus, and L. monocytogenes in lake water with this designed two-membrane device. Each of the SERS-Tag (antibody-conjugated AuTag@Ag) consists of an Au@Ag nanoparticle as Raman enhanced substrate, an adsorbed Raman reporter dye (CVa, R6G, and MB) to produce characteristic SERS signal, and a specific antibody to recognize and bind with target pathogen.

Structurally, this two-membrane filtration device is a remolded syringe with an upper layer filtration membrane with pore size of 30 μm (intercepting membrane) and a lower layer filtration membrane with pore size of 220 nm (enrichment membrane). When utilized for multiple SERS decoding assay, as shown in Scheme 1, the mixed SERS-Tags (SERS-TagCVa, SERS-TagR6G, and SERS-TagMB) would firstly be incubated with water samples to bind to their respective target pathogenic bacteria (E. coli O157:H7, S. aureus, and L. monocytogenes) via the specific immune recognition. This incubated water sample was made to flow through the double-layer filtration device quickly. During this process, objects with large size including sand, plankton etc. are entrapped by the upper intercepting membrane (pore size of 30 μm), while three target pathogens can be intercepted and concentrated by the lower enrichment membrane (pore size of 220 nm). Simultaneous and direct SERS measurement of multiple target pathogens is realized on the enrichment membrane. This two-membrane device holds great potential for water quality control in resource-limited settings.

Scheme 1
scheme 1

Schematic illustration of the portable double-layer filtration device with SERS-Tags as Raman signal reporting units for detecting multiple bacteria in water samples

Materials and methods

Materials and instruments

Chloroauric acid, L-ascorbic acid, trisodium citrate dihydrate, sliver nitrate (AgNO3), bovine serum albumin (BSA), rhodamine 6G (R6G), agarose, dNTP, DNA marker, and Taq DNA polymerase were obtained from Sangon Biotech (Shanghai, www.sangon.com/) Co., Ltd. Methylene blue (MB)and cresyl violet acetate (CVa) were purchased from Aladdin (Shanghai, China, www.www.aladdin-e.com/). Anti-E. coli O157:H7 antibody (Anti-E-Ab), Anti-S. aureus antibody (Anti-S-Ab), and Anti-L. monocytogenes antibody (Anti-L-Ab) were purchased from Abcam (Beijing, China, www.abcam.cn). Portable SEED3000 Raman spectrometer was provided by OCEANHOOD (OCEANHOOD, Shanghai, China) and adopted for on-site Raman measurements with a 785-nm (100 mW) laser and an acquisition time of 10 s.

Bacterial strains

All bacterial strains including E. coli O157: H7, S. aureus, L. monocytogenes, Cronobacter sakazakii (C. sakazakii), Escherichia coli (E. coli), Pseudomonas aeruginosa (P. aeruginosa), and Salmonella enteritidis (S. enteritidis) were from American Type Culture Collection (ATCC) and China Microbiological Culture Collection (CMCC). See details in Table S1. They were cultured overnight in Luria–Bertani broth (Sangon Biotech Co., Ltd., Shanghai, China) at 37 ℃ and counted by the standard flat colony counting method [35].

Preparation of gold nanoparticles

Preparation of AuNPs was conducted according to previous reported method with minor modification [36]. Briefly, 750 µL of 5 g L−1 chloroauric acid was added to 50 mL of deionized water under vigorous stirring. After being heated to boiling, 800 µL of 1% trisodium citrate was added immediately. Then, the solution was kept in boiling for additional 15 min and then cooled to room temperature.

Preparation of various AuTag for multiple bacteria

To prepare AuTag (AuCVa, AuR6G, and AuMB), a certain amount of Raman signal molecule (10 µL 100 µM CVa for AuCVa, 10 µL 100 µM R6G for AuR6G, and 50 µL 10 µM MB for AuMB) was mixed with 1 mL AuNPs. The mixture was shaken for 10 min at room temperature to make Tag molecules fully bind with AuNPs. Obtained AuTag was purified by centrifugation at 8000 g and resuspended in 1 mL sterile water.

Preparation of AuTag@Ag

The preparation of AuTag@Ag (AuCVa@Ag, AuR6G@Ag, and AuMB@Ag) was performed based on the reducing of AgNO3 on the surface of AuTag [37, 38]. In brief, 100 µL ascorbic acid was added into 1 mL AuTag and stirred for 10 min. This was followed by quick adding 300 µL 1% AgNO3 under ultrasonic condition. Of note, the ultrasonic condition was maintained until the solution color changed from purple-red to orange-yellow. The final AuTag@Ag solution was centrifuged at 10,000 g and resuspended in 1 mL sterile water.

Preparation of SERS-Tag

To synthesize SERS-Tags (SERS-TagCVa, SERS-TagR6G, and SERS-TagMB), namely, anti-E. coli O157:H7 antibody-conjugated AuCVa@Ag (AuCVa@Ag@Anti-E-Ab), anti-S. aureus antibody-conjugated AuR6G@Ag (AuR6G@Ag@Anti-S-Ab), and anti-L. monocytogenes antibody-conjugated AuMB@Ag (AuMB@Ag@Anti-L-Ab), the pH of AuCVa@Ag, AuR6G@Ag, or AuMB@Ag was firstly adjusted to ~ 8.0 by K2CO3. Thereafter, 4 µL of 1 mg mL−1 anti-E. coli O157:H7 antibody, anti-S. aureus antibody, or anti-L. monocytogenes antibody was mixed with 1 mL AuCVa@Ag, AuR6G@Ag, or AuMB@Ag, respectively. After reacting for 1 h at room temperature, the nano-conjugates were further blocked with 20 µL 10% BSA for an additional 1 h and centrifuged at 10,000 g for 10 min to remove free antibodies and BSA. The final products were re-suspended in 1 mL stock solution (10 mM PBS) for subsequent utilization. The mixture of different SERS-Tags was added into the water samples and incubated for 10 min, and the incubated water samples were quickly treated with the constructed double-layer filtration device.

Fabrication of the double-layer filtration device

The framework of the device composed with a syringe, a hollow polystyrene filtration tube, and two cellulose membranes with different pore sizes (30 μm and 220 nm). The adoption of the pore size of different membranes was based on the consideration of practical application conditions and existing commercial filtration membrane. Filtration membrane with pore size of 30 μm can guarantee the interception of most macro particles while the lower filtration membrane with pore size of 220 nm can retain most of the bacteria on the membrane. To fabricate the device, both two cellulose membranes were cut into disc shape with the diameter of 15 mm to cover the water inlet (diameter of 6 mm) and outlet (diameter of 4 mm). The hollow polystyrene tube was mechanically installed between the syringe and the needle.

SERS decoding measurement

To perform the SERS measurement, 20 µL of SERS-TagCVa, 20 µL of SERS-TagR6G, and 20 µL of SERS-TagMB were added to 10 mL of multiple bacteria-contaminated lake water samples with known bacteria concentrations and incubated for just 10 min. Then, the mixture was transferred to the constructed filtration device. To maximally avoid the non-specific absorption and removal impurities, the above operation was repeated and filtrated for 3 times. The second filtration membrane (enrichment membrane) with pore size of 220 nm was taken down and placed onto a glass slide for the collection of Raman signals. For the collection of Raman signals on the membrane, at least 10 different spots on the membrane should be performed to get the average values. Then, for the detection of each concentration, three parallel tests were conducted to obtain the average value, which was adopted for the construction of calibration plot for quantitative analysis. It is worth pointing out that the adopted water samples have all been pre-validated by both the culture method and PCR to ensure the existence of the target bacteria or not.

Bacterial DNA extraction and PCR amplification

The extraction of bacterial DNA and PCR amplification were conducted based on the reported method [17, 39]. For DNA extraction, the lower enrichment membrane with intercepted and concentrated bacteria was immersed in 700 µL of lysate solution (10 mM Tris-HCl, 200 mM NaCl, 5 mM EDTA, 1% SDS, pH 8.0). After adding with 30 µL 20 mg mL−1 Proteinase K and incubated at 60 ℃ for 10 min, the mixture was further added with 100 µg carboxylated MNPs and kept at room temperature for 5 min. The MNPs absorbed with extracted genomic DNA were collected by magnetic separation. Finally, the bacterial genomic DNA was eluted with ethanol and resuspended in 100 µL of TE buffer (10 mM Tris-HCl, 1 mM EDTA, pH 8.0). For PCR amplification, a 25 µL reaction mixture containing 1 µL extracted DNA template, 10 mM 10×PCR buffer, 0.2 µM forward primer, 0.2 µM reverse primer, 0.2 mM dNTPs, 1.75 mM MgCl2, and 1 U Taq DNA polymerase was reacted in the PCR device. The amplification conditions were set as follows: pre-denaturation at 95 °C for 5 min, 30 cycles of amplification consisting of denaturation at 94 °C for 30 s, annealing at 58 °C for 30 s, and extension at 72 °C for 30 s, and the final extension is at 72 °C for 5 min. The primer sequences for E. coli O157:H7 [19], S. aureus [16], and L. monocytogenes [40] were obtained from GENERAL BIOL (Chuzhou, China) and listed in detail in Table S2. PCR amplicons were confirmed by agarose gel electrophoresis.

Results and discussion

Investigation of Raman spectra for multiple analysis and Raman enhancement effect

To realize simultaneous multiple detection with SERS, it is the prerequisite to select a group of Raman reporter molecules without inter-spectral interference. The Raman signals of CVa, R6G, MB, AuCVa, AuR6G, AuMB, AuCVa@Ag, AuR6G@Ag, and AuMB@Ag were firstly measured in the suspension dropped onto the glass slide. As measured in Fig. 1A, CVa, R6G, and MB show their distinct characteristic peaks at 586 cm−1, 1501 cm−1, and 1614 cm−1, respectively. Since these Raman reporters have their own characteristic finger spectrum separated from each other at least 100 cm−1, it is technically possible to achieve simultaneous multiple detection with these reporters. The inset image gives the structural models of CVa, R6G, and MB, respectively. Secondly, the SERS-Tags should be offered with a strong Raman enhancement effect so that their greatly enhanced Raman signals can benefit for the improvement of sensitivity. For the verification of SERS effect, results shown in Fig. 1B demonstrate that direct Raman measurement of CVa can just produce the weak signal while a remarkable increase of Raman signal can be observed when CVa is attached on the surface of Au@Ag as the AuCVa @Ag. The extraordinary enhancement effect confirms the basis for accurate and sensitive detection of target bacteria. Similarly, the verification results of enhancement effect to the R6G and MB are also demonstrated in Fig. 1C, D, respectively. All these results strongly provide the solid foundation for simultaneous detection of multiple pathogenic bacteria in drink water with the designed strategy.

Fig. 1
figure 1

A Typical Raman signals of CVa, R6G, and MB with selected characteristic peak at 586 cm−1 (CVa), 1501 cm−1 (R6G), 1613 cm−1 (MB), respectively. Inset gives their structural formulas. BD The Raman spectra of single Raman reporter, AuTag, and SERS-Tag for CVa, R6G, and MB, respectively. The stepwise intensified Raman intensities show the superior Raman enhancement effect

Characterization of functional gold nanoparticles and gold nanoparticles @silver nanoparticles

The size and morphology of AuNPs and Au@Ag (exemplified by AuCVa@Ag) are characterized by TEM and UV-vis as shown in Fig. 2. Panel a in Fig. 2A shows the example of highly monodispersed spherical AuNPs with average size of 20 nm. After reduced with AgNO3 on its surface, in panel b of Fig. 2A, it can be clearly observed that the surface of AuNPs is uniformly coated with a layer of silver shell with the thickness of about 2 nm. And the corresponding UV-vis measurements in Fig. 2B exhibits the treatment of AuNPs with AgNO3 which led to a distinct blue-shift of peak absorption from 524 nm to 403 nm, indicating the formation of Ag shell. Inset images show the wine-red colored AuNPs changed to orange-yellow colored AuCVa@Ag. And it is also noted that there is a weak shoulder at ca. 500 nm of the Au@Ag, which could be ascribed to the potential aggregation of the prepared Au@Ag. All these comparison results suggest the successful preparation of functional AuNPs and Au@Ag for SERS measurements. The conjugation of Ab and BSA onto the surface of Au@Ag for recognition and blocking respectively were also characterized with UV-vis. Detailed results in Figure S1 indicate that the modification of Ab and BSA induce the red-shift of 3 nm and 17 nm, respectively, indicating the successful conjugation of Ab and BSA onto the Au@Ag.

Fig. 2
figure 2

A TEM Characterization of AuNPs (a) and Au@Ag (b). B UV-vis spectra of AuNPs and Au@Ag. The inset shows photograph of AuNPs and Au@Ag

Morphology characterization of the enrichment membrane with SEM

To validate the enrichment ability of the lower enrichment membrane, the morphology of the enrichment membrane is observed before and after utilized for target bacterium enrichment by SEM and demonstrated in Fig. 3. Results in Fig. 3A illustrate the high porosity of the virgin enrichment membrane before utilization. When the water confirmed without bacterium is treated with this enrichment membrane, results in Fig. 3B display the well-shaped pore structure of membrane even though some tiny particles are attached. In contrast, results of filtration of water containing 1 × 105 CFU mL−1E. coli O157:H7 are shown Fig. 3C, demonstrating numerous bacteria entrapped on the surface of the enrichment membrane. To zoom in one collected bacterial cell with TEM, in Fig. 3D, one can see clearly that many small-sized SERS-TagCVa are tightly and densely distributed on the surface of E. coli O157:H7 via antigen-antibody recognition. These morphology characterization results strongly demonstrate the feasibility of this designed two-layer filtration device for effective separation and enrichment of target pathogens on membrane for simultaneous multiple detection. Additionally, the entrapping of S. aureus at the concentrations of 1 × 101 CFU mL−1, 1 × 103 CFU mL−1, 1 × 105 CFU mL−1 on the surface of the enrichment membrane is also characterized by SEM and shown in Figure S2.

Fig. 3
figure 3

SEM images of A clean enrichment membrane treated with B no bacteria-contaminated lake water and C E. coli O157:H7 contaminated lake water. D TEM image of the surface of one E. coli O157:H7 cell to show copious amount of SERS-TagCVa attached

Assay performance investigation on multiple bacteria detection

After feasibility investigation and confirmation, we firstly test the sensing ability of the designed filtration device with only one kind of pathogenic bacterium, respectively. The detailed recorded Raman signals and corresponding linear response against the concentration of target bacterium are depicted in Fig. 4. Typically, in Fig. 4A, for E. coli O157:H7 detection, with the increase of E. coli O157:H7 concentration in water samples, the number of bacteria collected on the enrichment membrane increases accordingly. And the intensity of the characteristic peak (at 586 cm−1) of CVa is also increased with the increase of amount of E. coli O157:H7 in water samples (Fig. 4A). The linear response relationship is also constructed to be YCVa= 1515.96 lgX-21.72 with a correlation coefficient of 0.9929, in which YCVa and X stand for the Raman intensity of CVa at 586 cm−1 and the logarithmic concentration of the target E. coli O157:H7, respectively (Fig. 4B). Similarly, the detection results of S. aureus and L. monocytogenes are also shown in Fig. 4C–F. And the linear responses are YR6G= 609.99X + 9.94 (R2 = 0.9854) and YMB= 981.61X + 79.89 (R2 = 0.9942), for S. aureus and L. monocytogenes, respectively. For the detection results of all these target bacteria, the signal of 1 × 101 CFU mL−1 can be well distinguished from the blank group, indicating the excellent sensitivity for detection of bacteria in water. This satisfied detection performance can be ascribed to the effective remove of potential interferents with the upper large pore-size membrane and efficient retainment of target bacterium on lower small pore-size membrane for Raman measurement. Comparison results with other reported methods for pathogen detections and also with other SERS methods are described in Table S3. Comparatively, the reported method in this study is much better than others in items of sensitivity, linear range, target species, detection time, and even the detection cost.

Fig. 4
figure 4

Dose– response SERS spectra for E. coli O157:H7 (A), S. aureus (C), and L. monocytogenes (E) detection. The target concentrations from low to high are 0, 1 × 100, 1 × 101, 1 × 102, 1 × 103, 1 × 104, 1 × 105, and 1 × 106 CFU mL−1. Plot of SERS intensity at peak 586 cm−1 (B), 1501 cm−1 (D), and 1614 cm−1 (F) as a function of logarithm of E. coli O157:H7, S. aureus, and L. monocytogenes, respectively. Error bars are from three independent tests

Furthermore, any two kinds of the target bacteria are mixed to prepare the spike water samples. Different combinations including the E. coli O157:H7/S. aureus, E. coli O157:H7/L. monocytogenes, and S. aureus/L. monocytogenes are all tested with this designed method, respectively. Detailed SERS results are demonstrated in Fig. 5. Typically, for detection of the combination of E. coli O157:H7 and S. aureus in the water sample, two characteristic peaks at 586 cm−1 (for E. coliO157:H7) of CVa and 1501 cm−1 (for S. aureus) of R6G can be observed simultaneously in the Raman results in Fig. 5A. Of note, the intensities of these two peaks show the obvious concentration-depended properties of corresponding target bacterium. The recovery rate is calculated according to the constructed linear response relationship in Fig. 4. Detailed recovery results in Table S4 indicate that the quantity of the target bacterium can be well and accurately determined with satisfied recoveries and RSDs in the complicated samples without any interference from other coexisted bacteria. These studies strongly prove the capacity of this designed filtration device for simultaneous determination of multiple bacteria in the same water samples.

Finally, the samples with the mixture of three kinds of bacteria are also evaluated with this designed filtration assisted SERS method. Results in Fig. 6A demonstrate that the corresponding characteristic peak of each target bacterium can be accurately decoded and measured for the quantitative analysis. The recovery studies in Table S5 of each target bacterium are all in the acceptable range. Besides, the collected bacteria in the enrichment membrane are also confirmed with the classical PCR. Results in Fig. 6B also further confirm the accuracy of detection results of our developed method. From all these results, it can come to the conclusion that this designed two membrane-assisted SERS method can be applied for the direct simultaneous analysis of multiple target bacteria in aqueous samples.

Fig. 5
figure 5

Representative SERS spectra collected from the target mixtures of E. coli O157:H7/S. aureus (A), E. coli O157:H7/L. monocytogenes (B), and S. aureus/L. monocytogenes (C)

Fig. 6
figure 6

A Simultaneous SERS decoding of E. coli O157:H7, S. aureus, and L. monocytogenes at different concentrations from lake water. B Gel electrophoresis analysis of PCR amplicons from bacteria on the surface of the enrichment membrane. The concentration of each bacterium is 103 CFU/mL; C Raman intensities at characteristic peaks caused by target bacteria and non-targets. D Detection of 103 CFU/mL E. coli O157:H7, S. aureus, and L. Monocytogenes from lake water, deionized water and liquid beverage. Error bars indicate the standard deviations from at least three measurements

Specificity evaluation

The specificity of our designed filtration-assisted SERS method is also evaluated. As gathered in Fig. 6C, positive Raman signals at their characteristic peaks can only be observed when target E. coli O157:H7, S. aureus, and L. monocytogenes exist. Compared with the Raman intensity results of blank control samples (the right group in Fig. 6C), the results of other four non-target bacteria including E. coli, P. aeruginosa, and S. enteritidis are also the same as the blank control group, indicating no Raman signal produced of these non-target bacteria.

For the three target bacteria (E. coli O157:H7, S. aureus, and L. monocytogenes), the obvious Raman signals can be easily distinguished from each other compared with both the non-target bacteria and blank control group, indicating an excellent specificity of this device for rapid and multiple bacterial detection and identification. More than that, we also exploited the generality of this device by comparing the multiple pathogens profiling results of lake water samples with deionized water and liquid beverage samples at the same concentration of various target bacteria. Of note, all aqueous samples have been pre-verified by the classic and gold standard PCR to ensure it did not contain the target bacteria by the Anhui Province Institute of Product Quality Supervision & Inspection. The Raman signal values in lake water are defined as 100% in Fig. 6D. Calculated relative responses of Raman signal in both deionized water and liquid beverage are in great accordance with that achieved in lake water, suggesting the universality and well anti-matrix effect of this designed filtration-assisted SERS method.

Conclusions

In summary, we have designed and built a portable filtration assisted SERS device for simultaneous detection of multiple target bacteria in water samples. And we have demonstrated its successful use for simultaneous SERS-decoding of three waterborne pathogens. In this study, depending on the design of the device with two membranes with different functions as well as the employment of the SERS-Tags, three kinds of target bacteria are recognized by different SERS-Tag, concentrated onto the enrichment membrane, and measured directly by Raman. We have demonstrated that this device worked well in lake water with unprecedented detection efficiency, ease of use, and sensitivity. By simple incubation (10 min), injection (10 min), and Raman measurement (5 min) steps, as few as 10 CFU mL−1 of E. coli O157:H7, S. aureus, and L. monocytogenes can be detected in less than 30 min. The method also shows a broad linear range from 101 to 106 CFU mL−1, an excellent specificity, and functions equally well for aqueous beverages, broadening its wide acceptance. With these characteristics, this straightforward approach offers a new guide for rapid determination of bacteria in identifying infection sources, surveying food chains, and biodefense.