Introduction

Electrochemiluminescence (ECL), as a kind of electrochemistry-merged luminescence [1], has aroused increasing concerns in bio-research and environmental monitoring because of its high sensitivity, low background interference, electrochemical controllability, and excellent selectivity [2, 3]. The ECL technology has been widely applied for detecting the protein, nucleic acid, and small biomolecules, such as dopamine and glucose. For example, our group constructed an ECL ratiometric biosensor to achieve a highly sensitive determination of dopamine [4]. The Khalilzadeh’s group developed an ECL aptamer-based biosensor for the determination of thrombin [5]. Besides, Liu and co-workers applied a bi-pedal DNA walker signal amplification strategy to construct an ECL biosensor for detecting miRNA-21 [6]. The ECL has provided a sensitive strategy for biological analysis.

The ECL emission usually involved two pathways, namely, the annihilation-based pathway and coreactant-based pathway. The annihilation-based ECL emission usually produced a low ECL efficiency, which made a trace analysis difficult to be implemented. Multitudinous ECL systems for biological analysis usually relied on a coreactant-based pathway to obtain enough detection sensitivity since the coreactant could remarkably enhance the ECL emission of luminophores [7,8,9]. However, as we know, a coreactant-dependent ECL detection would encounter various confusions no matter what strategy is used to apply the coreactant. For example, the direct addition of exogenous coreactant in a testing solution may destroy the detection microenvironment, and lead to a poor detection stability and reproducibility [10]. The immobilization of coreactants at the sensing interface may introduce complex operations. Furthermore, the immobilized coreactants may encounter a leakage from the sensing interface. Regarding the dissolved oxygen as coreactant, the uncertainty of dissolved oxygen concentration also would make the system lack the stability and repeatability. Considering the confusions from the application of coreactants, constructing a self-ECL system excluding the application of external coreactants and dissolved oxygen is essential.

In order to construct a self-ECL system for an actual analysis, the selection of self-ECL emitter and suitable ECL signal switching mode is crucial. Traditional ECL luminophores, such as organic emitters [11], metallic complex [12], quantum dots [13], and metal nanoclusters [14] generally needed external coreactants or dissolved oxygen to achieve high luminescence intensity. The polyfluorene, as a new kind of luminophore, exhibited a remarkable brightness, high emission rate, and extraordinary photostability, thus receiving an extensive interest in the ECL biosensing and imaging [15, 16]. Based on the application of coreactants H2O2, C2O42−, or tripropylamine (TPrA), the highly sensitive detection of choline [17], melamine [16], and cancer biomarkers [18] was achieved. More importantly, the polyfluorene has been recently found to exhibit an extraordinary self-ECL property without any external species and dissolved oxygen as conreactants, which endowed the polyfluorene an extensive application prospect in the construction of ECL sensing. For instance, our group proposed an ECL strategy to detect the nucleic acid by coupling carboxyl-functionalized PFBT dots as a self-ECL emitter and black hole (BHQ) as an effective quencher [19]. Ma and co-workers achieved the detection of nitroaromatics with poly [9,9-bis(3′-(N,N-dimethylamino) propyl)-2,7-fluorene]-alt-2,7 -(9,9-dioctylfluorene)] (PFN) as the self-ECL emitter and aluminum tris(8-quinolinolate) as the quencher [20]. However, the ECL resonance energy transfer (ECL-RET) between the self-ECL emitter and quencher was involved in the signal switching process, namely the ECL-RET between PFBT and BHQ, and between PFN and aluminum tris(8-quinolinolate). The ECL-RET-based signal switching mode would limit the application of above self-ECL systems due to the distance dependence of ECL-RET. Thus, developing a self-ECL system excluding the ECL-RET was of great significance.

Lead ions (Pb2+), as a major environmental pollutant, are harmful to the human body even at extremely low levels (100 mg/L in blood), including the hematological, cardiovascular, neurological, gastrointestinal, and renal system [21]. Thus, the rapid and accurate detection of Pb2+ in environmental and biological samples is particularly important and urgently demanded. Many traditional methods for Pb2+ detection have been developed, including high-performance liquid chromatography, atomic absorption/emission spectrometry, inductively coupled plasma-mass spectrometry, and inductively coupled plasma-atomic emission spectrometry, which have an acceptable sensitivity and accuracy [22,23,24,25]. Nevertheless, the ECL technology has received an extensive attention in Pb2+ detection due to its superiority in overcoming the expensive instruments, sophisticated sample preparation, or time-consuming procedures of the above-mentioned methods [26,27,28]. For example, Zhu et al. applied the Ru(bpy)32+-TPA system to achieve an ECL detection of Pb2+ [26]. Li et al. used CdS quantum dots/H2O2 ECL system to construct an ECL sensor for Pb2+ detection [28]. Although the ECL technology offers more possibilities for the sensitive detection of Pb2+, the ECL systems reported so far generally involved external species or dissolved oxygen as coreactants and the toxic and instability may be resulted from their application [11, 18]. Therefore, developing a self-ECL-sensing platform for Pb2+ excluding ECL-RET was very valuable.

In this work, an ECL biosensor was constructed for detecting Pb2+ with C-PFBT Pdots as self-ECL emitter and Fc as quencher. As exhibited in Scheme 1, C-PFBT Pdots–modified glassy carbon electrode (GCE) produced a strong ECL signal. C-PFBT Pdots served as a matrix for immobilizing the substrate DNA and further captured the DNA 1 labeled with Fc to obtain a “signal-off” state since Fc and C-PFBT Pdots were brought into close proximity for making the ECL signal of C-PFBT Pdots be quenched by Fc. With the introduction of DNAzyme, the DNAzyme chain hybridized with the substrate chain to form a stable duplex, accompanied with the release of Fc-labled DNA 1. In the presence of Pb2+, the DNA duplex formed by the DNAzyme chain and substrate chain selectively recognized Pb2+. Then the ribonucleotides (rA) in the substrate chain were cleaved and Pb2+ was released. Meanwhile, the DNAzyme chain would hybridize with the adjacent complete substrate chain. Once again, the released Pb2+ was identified by the DNA duplex to circularly activate the next DNAzyme. As a result, the DNAzyme chain continuously walked along the substrate chain modified on the electrode and the Fc-labeled DNA 1 was released, namely forming a DNA walking machine, which recovered the ECL signal of C-PFBT Pdots to obtain a “signal-on” state, thus achieving the detection of Pb2+. The designed self-ECL strategy exhibited a good performance in the detection of Pb2+ in MB-231 and A-549 cells. Such a self-ECL assay excluding ECL-RET would provide an attractive ECL platform for the biological analysis.

Scheme 1
scheme 1

Schematic illustration of a the preparation of C-PFBT Pdots and b the fabrication of the biosensor

Experimental

Reagents, chemicals, and apparatus

The information on reagents, chemicals, and apparatus in this work has been described in the Supplementary Materials.

Preparation of the C-PFBT Pdots

Firstly, PFBT and poly (styrene-co-maleicanhydride) (PSMA) were dissolved in THF to achieve a stock solution (1.0 mg/mL), respectively. Subsequently, the prepared PFBT (4 mL) and PSMA (800 μL) solutions were mixed and sonicated for producing a homogeneous solution. The resulting solution was quickly added into 10 mL ultrapure water. Ultimately, the THF was evaporated under vacuum with a constant stirring to obtain C-PFBT Pdots. The synthesis process of C-PFBT Pdots is shown in Scheme 1a.

Fabrication of the ECL biosensor

After polishing and cleaning the glassy carbon electrode (GCE, Φ = 4.0 mm) [29], 10-μL dispersion of C-PFBT Pdots was coated on the surface of GCE. After dried in air, the obtained C-PFBT Pdots/GCE was incubated with 0.02 g/mL of EDC and 0.01 g/mL of NHS for 30 min. Then, the electrode was incubated with 2 μL of substrate DNA (50 μM) at 4 °C for overnight and further incubated with 2 μL of DNA1 (50 μM) at room temperature for 2 h. Finally, 2 μL of DNAzyme (50 μM) and Pb2+ with various concentrations were incubated onto the electrode surface at room temperature for 140 min. The construction of ECL biosensor is depictured in Scheme 1b.

Native polyacrylamide gel electrophoresis

The operation of PAGE analysis is displayed in the Supplementary Materials.

Results and discussion

Characterization of the C-PFBT Pdots

Figure 1a exhibits the SEM images of C-PFBT Pdots, and a clear spherical structure was observed, which is similar to the previous report [19], demonstrating the successful synthesis of C-PFBT Pdots. Figure 1b presents the size distribution of C-PFBT Pdots, and the average particle diameter of C-PFBT Pdots was about 68 nm.

Fig. 1
figure 1

a SEM images of C-PFBT Pdots. b Size distribution of C-PFBT Pdots

ECL behavior of the self-ECL system

The effect of dissolved O2 on the ECL emission of C-PFBT Pdots was firstly studied. As seen from Fig. 2a, in both cases, namely in air-saturated PBS solution (0.10 M pH 7.4) (curve a) and N2-saturated PBS solution (0.10 M pH 7.4) (curve b), the ECL signals were almost identical, indicating that the dissolved O2 was excluded from the ECL reaction of C-PFBT Pdots.

Fig. 2
figure 2

a The ECL response of C-PFBT Pdots/GCE in PBS (0.10 M pH 7.4) with (a) air-satuarted and (b) N2-satuarted, (c) in MeCN solution under air-saturated, in PBS (0.10 M pH 7.4) containing (d) l-cys and (e) SOD. b ECL response of C-PFBT Pdots/GCE in PBS (0.10 M pH 7.4) with pH from 1 to 11. Scanning potential range, 0~1.25 V; scanning rate, 300 mV/s

Secondly, the effect of testing mediums including PBS solution and MeCN solution was also monitored. In the case of MeCN solution, TBAPF6 served as the supporting electrolyte. The ECL signal of C-PFBT Pdots/GCE was detected in PBS solution (0.10 M, pH 7.4) and MeCN solution, respectively. As seen from Fig. 2a, the modified electrode displayed a remarkably decreased ECL signal in MeCN solution (Fig. 2a, curve c), as compared with that in PBS solution (Fig. 2a, curve a), indicating a vital role of the aqueous solution in the ECL emission.

Since the vital role of aqueous solution was observed, the effects of reactive oxygen species (ROS) including the hydroxyl radical (OH·) and superoxide radical (O2·) in aqueous solution were explored. The l-cysteine (l-cys) and superoxide dismutase (SOD) were introduced to eliminate OH· and O2·, and Fig. 2a presents the results. As seen, the ECL signal at C-PFBT Pdots/GCE with l-cys (curve d) or SOD (curve e) was almost identical to that without l-cys and SOD (curve a), illustrating that OH· and O2· did not participate in the ECL reaction of C-PFBT Pdots.

To further elucidate the possible ECL mechanisms of C-PFBT Pdots without any exogenous coreactants, the effect of pH was explored and Fig. 2b depictures the results. As seen, the C-PFBT Pdots–modified electrode displayed a constantly increased ECL signal with increasing pH from 1 to 11, demonstrating the decisive role of OH in the ECL emission. The possible self-ECL mechanisms for C-PFBT Pdots may be summarized as follows. Firstly, the oxidation of C-PFBT Pdots occurred on the electrode surface, and the radical cation C-PFBT Pdots·+ was generated. The radical cation reacted with OH to produce the reducing intermediate radical C-PFBT Pdots·. Then, the excited state C-PFBT Pdots * was yield when C-PFBT Pdots· encountered with C-PFBT Pdots·+, thereby occurring the self-ECL emission, as reported by the previous literature [19].

Polyacrylamide gel electrophoresis analysis

The feasibility of the DNA reaction on the electrode was verified by polyacrylamide gel electrophoresis (PAGE). Fig. S1 illustrates the results. The distinct bands in lanes 1–3 corresponded to the DNA1, substrate DNA, and DNAzyme, respectively. With the hybridization between DNA1 and substrate DNA, a brightened band appeared (lane 4). After the DNAzyme was introduced, an additionally slower migrate band was observed (lane 5), suggesting that the hybrids of DNAzyme and substrate DNA were formed and the DNA1 was released. After Pb2+ was introduced, a new and obvious band appeared (lane 6), which corresponded to the short chain DNA released during the Pb2+ cleaving, and it migrated the fastest. The above PAGE results verified the formation of DNA walking machine during the detection of Pb2+ at the biosensor.

CV, EIS, and ECL characterization of the modified electrode

CV characterization of each modification step was conducted in PBS (0.10 M, pH 7.4) in the presence of K3[Fe(CN)6]/K4[Fe(CN)6] (5.0 mM) as the supporting electrolyte and Fig. 3a depictures the results. Compared with the bare GCE (curve a), the C-PFBT Pdots–modified GCE displayed an obviously decreased redox peak current (curve b), which was ascribed to the resistance of C-PFBT Pdots to the electron transfer. As the C-PFBT Pdots/GCE was successively incubated with the substrate DNA and Fc-labeled DNA 1, a further decreased redox peak current was observed (curve c). After the DNA walking machine was triggered by DNAzyme and Pb2+, the Fc-labeled DNA 1 was released from the electrode surface, and the redox peak current was enhanced again (curve d).

Fig. 3
figure 3

a CV and b EIS diagrams of (a) bare GCE, (b) C-PFBT Pdots/GCE, (c) C-PFBT Pdots/GCE with substrate and Fc-labeled DNA 1, (d) C-PFBT Pdots/GCE with substrate, DNAzyme and Pb2+ in 5.0 mM [Fe(CN)6]3−/4− solution. The insert of b: the equivalent circuit. The effect of c pH of PBS solution and d incubation time of Pb2+ on the ECL response at the biosensor. Scanning potential range of − 0.2~0.6 V and scanning rate of 100 mV/s for CV detection. The init potential of 0.22 V, amplitude of 5 mV, and frequency of 0.1~105 Hz for EIS detection. Scanning potential range of 0~1.25 V and scanning rate of 300 mV/s for ECL detection

The EIS characterization was also performed with the [Fe(CN)6]3−/[Fe(CN)6]4− as a redox probe. The impedance spectra are pictured as the Nyquist plots (Fig. 3b), and their high-frequency regions exhibited a semicircle. The diameter of the semicircle was equal to the impedance of electron transfer (Ret). As seen, a relatively small Ret was detected at a bare GCE (curve a). When C-PFBT Pdots were modified onto the surface of GCE, the Ret evidently increased due to the nonconductive of C-PFBT Pdots nanomaterial (curve b). After the substrate DNA and Fc-labeled DNA 1 were successively incubated onto the surface of C-PFBT Pdots/GCE, a further increased Ret was observed because the biomolecules blocked the electron transfer (curve c). When the DNA walking machine was triggered by DNAzyme and Pb2+, a decreased Ret was detected (curve d). Additionally, the insert pictures are the equivalent circuit of impedance spectra. EIS results indicated a successful construction of the biosensor.

Additionally, ECL measurements of each modification step were further conducted in PBS (0.10 M, pH 7.4). No ECL signal was detected at a bare GCE (Fig. S2, curve a). As expectedly, the C-PFBT Pdots modified electrode exhibited a strong ECL response (Fig. S2, curve b). With the introduction of Fc-labeled DNA 1 (Fig. S2, curve c), the ECL signal was quenched by Fc, which was mainly ascribed to the electron transport between Fc and C-PFBT Pdots. Finally, when Fc was away from the surface of electrode with the generation of DNA walking machine triggered by target Pb2+, an increased ECL signal was obtained (Fig. S2, curve d).

Optimization of experimental conditions

The incubation time of Pb2+ and pH of PBS solution were optimized in the case of 1.0 × 10−4 M Pb2+. Firstly, the pH of PBS was optimized and Fig. 3c displays the results. The ECL intensity gradually increased as pH shifted from 5.5 to 7.4 and reached a maximum at pH 7.4. As pH exceeded 7.4, a gradually decreased ECL signal was observed with increasing pH from 7.4 to 9.0, which was mainly ascribed to the fact that the nucleic acid strand showed the maximum activity at a physiological pH. Therefore, the optimal pH of 7.4 was used in subsequent experiments. Secondly, the incubation time of Pb2+ was optimized and Fig. 3d presents the results. When the cleaving time of target Pb2+ ranged from 50 to 200 min, the ECL signal increased first and then decreased accordingly and reached the maximum at 140 min. Therefore, the incubation time of target Pb2+ in subsequent experiments was chosen as 140 min.

Analytical performance of the biosensor

The quantitative detection of Pb2+ was performed at the constructed biosensor under the optimized experimental conditions. As seen from Fig. 4a, as the concentration of Pb2+ ranged from 0.5 pM to 0.1 mM, the ECL intensity gradually increased. The increase in ECL signals was ascribed to the departure of the ECL quencher Fc labeled in the DNA 1 from the electrode’s surface, which was triggered by Pb2+. And a good liner relationship between the ECL intensity and the logarithm of Pb2+ concentration was obtained with a linear regression equation I = 12,988.28 + 1145.28 lg c (R = 0.9974) (Fig. 4b). Here, I and c stand for the ECL intensity and the concentration of Pb2+, respectively. Based on the IUPAC recommendations [30, 31], the limit of detection (LOD) of 0.17 pM was obtained. Our proposed biosensor was compared with other strategies for Pb2+ detection and the results are illustrated in Table 1. Compared with reported detection platforms, our constructed ECL system exhibited a higher sensitivity for quantifying Pb2+.

Fig. 4
figure 4

a ECL response of the biosensor towards different concentrations of Pb2+, from a to h: 5.0 × 10−10, 5.0 × 10−9, 5.0 × 10−8, 5.0 × 10−7, 1.0 × 10−7, 1.0 × 10−6, 1.0 × 10−5, and 1.0 × 10−4 M, respectively. b Linear calibration between the ECL intensity and the logarithm of Pb2+ concentration. Scanning potential, 0~1.25 V; scanning rate, 300 mV/s

Table 1 Comparison of this constructed ECL assay with other methods

Stability, selectivity, and reproducibility studies

The consecutive cyclic potential scans at the biosensor were performed to evaluate the stability of the biosensor. As depicted in Fig. 5a, the relative standard deviation (RSD) of 4.54% was obtained for the ECL intensity in the case of the consecutive cyclic potential scans for 9 cycles at 0.10 μM of Pb2+, demonstrating a good stability of the biosensor. Some possible interferences including Hg2+, Cu2+, Mg2+, Mn2+, Ni2+, Cd2+, and Co2+ were chosen to evaluate the selectivity of our proposed self-ECL assay. As seen from Fig. 5b, no obvious ECL response was observed in the case of 1.0 mM interferences (from b to h), as compared with the blank (a). However, in the case of 0.10 μM Pb2+ (i), the biosensor presented a strong ECL response. Moreover, the ECL response towards the mixture of 0.10 μM Pb2+ and 1.0 mM interferences (j) was almost the same as that of Pb2+ alone (i), suggesting a high selectivity of our proposed ECL biosensor for Pb2+ detection. The intra- and inter-assays were explored to evaluate the reproducibility of the biosensor, and Fig. 5c depicts the results. As seen, four biosensors from the different or the same batch did not exhibit any obvious change in the ECL response towards 0.10 μM Pb2+, indicating an acceptable reproducibility for Pb2+ detection.

Fig. 5
figure 5

a ECL response of the biosensor to 0.10 μM Pb2+ in 0.10 M PBS (pH 7.4) under the continuous cyclic scans. b ECL response of the biosensor towards the blank (a), 0.10 mM of Hg2+, Cu2+, Mg2+, Mn2+, Ni2+, Cd2+, Co2+ (from b to h), 0.10 μM Pb2+ (i), and the mixture of Pb2+ and interferences (j). c The inter-assay and intra-assay of four biosensors in the presence of 0.10 μM Pb2+ in 0.10 M PBS (pH 7.4). d Calibration curve of ECL intensity vs the logarithmic concentrations of intracellular Pb2+ from MB-231 (red line) and A-549 (blue line). Scanning potential, 0~1.25 V; scanning rate, 300 mV/s

Analysis of real samples

To estimate the practical applicability of our developed ECL biosensor, the MB-231 cells and A-549 cells were selected as models and were cultured in DMEM medium containing different concentrations of Pb2+. After being cultured with Pb2+, the MB-231 cells and A-549 cells were washed with the DMEM medium without Pb2+ for removing the Pb2+ adsorbed on the cells’ surface. Then, the cell lysates were prepared and further employed to perform the ECL measurements using our developed biosensors. The preparation of cell lysates was described in the Supplementary Materials. The Pb2+ in cell lysates of MB-231 cells and A-549 cells was firstly measured with the inductively coupled plasma-mass spectrometry (ICP-MS). As expected, for two different cell lines, an excellent linear relationship was observed between the ECL signal strength and the logarithm of intracellular Pb2+ concentration (Fig. 5d). Furthermore, the calibration plots for two different cell lines almost overlap, demonstrating a potential application of our designed coreactant-free ECL strategy.

Conclusion

A self-ECL biosensor was designed for the ultrasensitive detection of Pb2+ based on C-PFBT Pdots as the self-ECL emitter and Fc as the quencher. Firstly, such a self-ECL system successfully overcame the disadvantage of introducing exogenous coreactants. Secondly, the combination of excellent ECL performance of C-PFBT Pdots and high quenching efficiency of Fc endowed this signal on-off-on strategy a low background signal and high detection sensitivity for Pb2+. Thirdly, the specific cyclic cleaves of Pb2+ achieved target recycling and enhanced the selectivity of the biosensor. Moreover, the designed biosensor demonstrated a preliminary application in the determination of intracellular Pb2+. The self-ECL system provided a new strategy and platform for ECL sensing and imaging technologies without any exogenous coreactants and the ECL-RET.