Introduction

Recently, Van Limbeck et al., in a systematic literature review, failed to identify the best method for a single-level anterior cervical interbody fusion [13]. With the introduction of different interbody fusion devices, such as cages, surgical strategies to achieve anterior cervical fusion have evolved. The first papers on such promising techniques have been published [4, 7, 10, 14]. However, Wigfield, in a thorough literature review, concluded that there was limited evidence supporting the use of a cervical interbody fusion device in place of autologous bone [15].

We therefore reviewed our first cases of stand-alone cervical cages with a specific emphasis on the subsidence behaviour of such cages. This was prompted by a clinical impression of settling of the implants during the follow-up of our patients

Materials and methods

We conducted a retrospective review of the first eight patients that we operated on with a stand-alone cervical cage. All patients initially presented with a cervical radiculopathy and concordant soft disc herniation that had persisted despite conservative management. Anterior discectomy and stand-alone cervical cage interbody fusion were performed in the C5-C6 segment in three cases and in the C6-C7 segment in six cases, using the same technique (one patient had a two-level cage insertion). Two independent observers reviewed hospital charts, operative reports and all radiographs. Follow-up time ranged from 12 to 18 months (average 15 months). The age of the patients ranged from 33 to 61 years (average age 43 years). Seven patients were male and one female.

Surgical technique

All patients were operated according to the same standardised technique. After insertion of a cervical spine distractor, a complete discectomy and a neural decompression were performed. Great care was taken to avoid damaging the cortical structures of the bony end plates. Using cage trials under fluoroscopy, the appropriate size of cage was assessed. The cage was then filled with cancellous bone from the anterior iliac crest that was harvested through a mini-incision. A standard convex SynCage-C anterior cervical interbody fusion implant (Synthes USA, Paoli, Pa., USA; Fig. 1) was then inserted under fluoroscopy for an exact placement, the cervical distractor was then removed and the cage tested for stability. Additional bone graft was applied in front of the cage. Patients were kept in a soft collar for 4–6 weeks.

Fig. 1.
figure 1

Cervical cage used in the study. The Syncage-C has a porous shaped box design with a superior and inferior central opening. The inferior opening is larger than the superior one, as it is used to fill the cage with cancellous bone. The two central openings permit bony ingrowth into and from the adjacent end plate, allowing the fusion to take place. The superior aspect of the cage is convex, the inferior aspect straight, to match the anatomy of the inferior and superior end plate of the cervical vertebra

Radiological evaluation

Plain radiographs in anterior-posterior and lateral views were performed postoperatively before discharge and at the different follow-ups. In order to assess subsidence or migration, three different measurements on the lateral radiographs were taken for each case, postoperatively and at the latest follow-up (see Fig. 2, Table 1). Subsidence was defined as any change in at least one of our parameters of 3 mm or more. When settling was present, it was then characterised as a superior migration, an inferior migration or a bi-directional migration. Segmental kyphosis or lordosis was measured by the angle between the upper and lower adjacent end plate line. Functional cervical radiographs in flexion and extension were used to assess segmental cervical motion. We assumed that stability was present if the difference on the flexion and extension radiographs was not greater than 2°.

Fig. 2.
figure 2

Measure of the anterior and posterior disc space height and anterior or posterior migration. Distance of the cage to the posterior wall of the vertebral body

Table 1. Radiological parameters (see Fig. 2 for explanation) (Ant DSH anterior disc space height, Post DSH posterior disc space height, Dist→C-PW distance of the cage to the posterior wall)

Results

There were no intra- or postoperative complications. Neurologic symptoms regressed initially in all patients.

Radiologically, only three patients out of the eight studied (or four cages out of the nine inserted) did not show any change in the studied parameters (Table 1, Fig. 3).

Fig. 3.
figure 3

Case 6: a 41-year-old man operated for a C6 radiculopathy. Postoperative and 15-months' follow-up radiographs show identical measurements with no sign of subsidence

A change in at least one of our defined radiological parameters of at least 3 mm was observed in five of nine segments (Table 1, Fig. 4). There was no case of anterior or posterior migration of the cage. Superior migration of the cage into the superior end plate was obvious on the lateral radiographs in two cases. Inferior cage migration into the inferior end plate was observed in two patients. Simultaneous superior and inferior end plate migration was observed in two cases. All the patients had functional radiographic evaluation, and no segmental motion or kyphosis was observed in any of these cases.

Fig. 4.
figure 4

Case 7: a 39-year-old woman with a recurrence of her radiculopathy at 6 months follow-up. A comparison of the postoperative lateral radiograph (left) with a radiograph obtained at 6 months' follow-up (right), reveals cage subsidence in the inferior and superior end plate

At the last follow-up, one patient had recurrence of her radiculopathy. This latter patient (patient 7) had experienced an initial regression of her neck pain and C7 radiculopathy following surgery, but her symptoms recurred again 6 months later. An obvious reduction of the intervertebral space with subsidence of the cage was observed on plain radiographs, and at 15 months' follow-up one could notice a fracture of the cage (Fig. 5). The patient required revision, removal of the cage, further cervical decompression, iliac crest bone graft and anterior plating.

Fig. 5.
figure 5

Same patient as in Fig. 4. At 15 months' follow-up the cage is broken (left). Right Aspect of the cage during its retrieval

Discussion

Uninstrumented fusion with graft shows collapse, extrusion or resorption, followed by neural compression, instability and pseudarthrosis [1]. To reduce these problems, anterior plating has been used in the past. In a recent meta-analysis by Floyd and Ohnmeiss, anterior cervical fusion with autograft was found to be superior to allograft as far as pseudarthrosis and disc space collapse were concerned [3]. More recently, reports have claimed advantages of stand-alone cervical cages over classic autologous bone graft with or without anterior plating [4, 7, 10, 14]. However, strong evidence of the superiority of cages is missing. Wigfield and Nelson claimed there was very little evidence to justify their use [15]. Vavruch et al., in a prospective randomised study comparing the Cloward procedure and a carbon-fibre cage, concluded that the clinical outcomes were similar, the anatomic alignment was better with cages, but the fusion rate was much higher with the Cloward procedure (86%) than with the carbon-fibre cages (62%) [14].

Our results demonstrated subsidence in five out of nine cages measured by a decrease in the disc space narrowing of at least 3 mm. One of the nine cages broke during the follow-up period. Our series, though small as yet, is very alarming, and we assume that subsidence in cervical stand-alone interbody cage fusion is a major problem and additional stabilisation is necessary to avoid this problem, as has already been reported in vitro [8, 11, 12].

The causes of subsidence may stem from the instability created by the discectomy, the postoperative cervical motion, the cage design, the end plate preparation, or the bone mineral density [2, 5, 6, 9, 16]. At any rate, subsidence is not classically observed after anterior cervical fusion with a plate.

Does cage subsidence or pseudarthrosis really matter clinically? In the study by Vavruch et al., the carbon-fibre cage group did as well as the group with the classic Cloward procedure, but the pseudarthrosis rate of their cage group (38%) is, to our mind, unacceptable in the long run [14]. Their exact rate and quantification of subsidence remains unknown.

Conclusions

The results of this small series of homogeneous patients, with the high incidence of cage subsidence and one cage breakage, indicates the need for caution with the use of stand-alone titanium cervical cages in anterior cervical surgery. We felt that, ethically, we could not continue using stand-alone cervical cages without the addition of anterior cervical plating. The subsidence that we observed does not seem to be specific to the particular cage design that we used, as other studies have significant subsidence with other cages, at least in vitro. This paper should represent a serious warning to those who want to use stand-alone cervical cages.