1 Introduction

Among the many applications of frequency combs (FC) in metrology (e.g., optical clocks and low noise frequency synthesis) [1,2,3], combs have long been recognized as superior light sources for absorption spectroscopy [4,5,6]. FC lasers generate a large stable array of equally spaced, phase-coherent, narrow-band spectral lines. In excess of 105 comb lines can be produced with mode-locked short pulse lasers, with line spacing in the region of many 10 s of MHz (depending on the comb laser’s repetition rate [1]). The comb lines’ narrow bandwidth is determined by the coherence time of the laser.

In many cases, the spectral resolution in absorption experiments is, however, limited by the spectrometer used for detection. In these cases FCs are essentially broadband lasers which act as a “spectral ruler” [1]. To exploit the inherently high resolution of FCs, new detection schemes using virtually imaged phase array spectrometers [7,8,9], comb-cavity vernier approaches [10, 11] and Fourier transform-based dual-comb spectroscopy [12,13,14,15,16,17] have been established. Some of these methods were recently reviewed [18]. To enhance the sensitivity in absorption experiments, applications of FCs in conjunction with optically stable cavities covering a spectral range that is limited by the high reflectivity range of the cavity mirrors have been developed [19, 20]. The key challenge of cavity-enhanced FC absorption spectroscopy is the coupling of the FC to the cavity by matching the comb lines to an appropriately designed mode structure of a high finesse cavity—different approaches to coupling and locking of FC lines to high-finesse optical cavities have also been reviewed [21].

In contrast to developing ultimate instrumental limits, the approach to cavity-enhanced absorption detection in this work aims at experimental simplicity. The principle suitability of a custom-built gain-switched frequency comb (GSFC) laser in conjunction with a very modest optical cavity was examined without spectrally stabilizing or mode-matching the cavity. GSFCs are tunable FCs that are simple in design and significantly cheaper than mode-locked femtosecond FC lasers [22]. They, however, do not deliver the finely spaced large number of comb lines and span a significantly smaller wavelength region (~ typically a few nm). GSFCs are generated by the direct modulation of commercially available laser diodes, as a distributed feedback laser, with a radiofrequency (RF) sine wave. External optical injection in a master–slave laser configuration can be employed to generate narrow line widths and low-intensity noise [23]. GSFCs can also be integrated as a monolithic devices [24] (see Sect. 4). The approach is simple and cost-efficient, and enables wavelength tunability of the comb and the line spacing, but with a rather limited number of comb lines [22]. GSFC lasers are commonly used as benchtop units for optical telecommunication in the C-band [22]. Recently, a dual-comb GSFC setup was reported based on gain switching and optical injection locking [25], with potential applications of the dual-comb architecture in the mid-IR or near UV. In Ref [25] the absorption feature of H13CN in the 1.5 µm region was reported.

The objective of this study was to simply tune more than one laser comb line to absorption features of an appropriate target gas (hydrogen sulfide, H2S) and to achieve sensitivities below the combustion limit of the hazardous gas in the parts per thousand (ppTv) regime using a low-finesse (f ~ 450) optical cavity without mode-matching schemes.

2 Experiment

The schematic of the experimental setup is shown in Fig. 1. The main components of the GSFC laser were a tunable master laser (HP-8168F), a distributed feedback slave laser (Thorlabs, ITC-502), an radio frequency (RF) signal generator (HP-83751A), a biased RF amplifier (Aldetec, ALP-0618S430), a polarization controller, and a circulator. The slave laser was kept at a stable operating temperature by employing a thermoelectric cooler and a thermistor in the laser package. The (tunable) frequency of the master laser was slightly detuned (to higher energies) from that of the slave laser. Polarization controlled light from the master laser was injected via a circulator (ports 1 and 2) into the slave laser. The optical injection power was ≈ 3.1 mW (5 dBm). For a DC bias current of 45.1 mA the gain switching was achieved by continuously driving the slave laser above and below its laser threshold using an amplified sinusoidal RF signal. This created a train of pulses corresponding to an optical comb in the frequency domain. The (tunable) frequency of the RF generator hence determined the free spectral range of the FC, which in the present experiment was typically maintained at ~ 10 GHz (see Fig. 3). The absolute center frequency of the GSFC was controlled through the frequency of the master laser, which was tunable external cavity diode laser. In this experiment, wavelength of the master laser was typically maintained at ~ 1574.8 nm (~ 6349.4 cm−1); cp. Fig. 3. Overall, the GSFC produced 20 phase-coherent, equally spaced comb lines with an individual spectral width of ~ 300 kHz [23] covering a range from 6346 to 6354 cm−1 (1575.8…1573.8 nm) in the near-infrared with an overall quasi-continuous integrated power close to ≈ 4 mW (6 dBm). For monitoring the GSFC emerging from port 3 of the circulator (Fig. 1), the light was split into two beams (ratio 90:10) and the less intense fraction of the beam was guided to an optical spectrum analyzer (OSA, Advantest Q8384). The main beam (90%) was collimated with a plano-convex (f ~ 25 mm) and a biconvex lens (f ~ 75 mm). The collimated light was guided into a stable optical cavity consisting of two plano-concave dielectric mirrors (diameter 25 mm, r = − 200 cm, Layertec GmbH) at a distance of d = 66.0 ± 0.1 cm. The modest reflectivity of the cavity mirrors of R = 0.993 ± 0.001 was measured with a double-beam UV/Vis/NIR spectrometer (PerkinElmer, Lambda-1050). The optical cavity was enclosed by a stainless steel vacuum tube (diameter ~ 40 mm) which was fitted with an access port for evacuation and sample injection as schematically shown in Fig. 1. Before experiments, the static gas cavity was evacuated by a rotary pump to ~ 0.1 mbar. The light exiting the cavity was collected with an achromatic lens and focused into a multimode light guide (diameter 1.5 mm) which was connected to the entrance aperture of a Fourier transform spectrometer (FTS, Vertex 80 with CaF2 beam splitter, Bruker Ltd.). In the spectrometer, an InGaAs photodiode was used to detect the light transmitted by the cavity. Using the Bruker OPUS software spectra were recorded with a spectral resolution of 0.08 cm−1 in an integration time of 120 s (corresponding to four scans per spectrum). This integration time was based on the optimum measured signal-to-noise ratio of the FC, established in a series of intensity stability measurements using the FTS and OSA.

Fig. 1
figure 1

Schematic of the experimental setup. Mirrors M1 and M2 are high-reflectivity dielectric mirrors (R ~ 0.993) forming a medium-finesse optical cavity with a length of 66 cm. L: lenses used for collimation and light collection, P: absolute pressure gauge, V: valves, RF: radio frequency (~ 10 GHz), ADC: analog-to-digital converter, OSA: optical spectrum analyzer

Fig. 2
figure 2

The near-IR absorption spectrum of H2S at 3 mbar recorded using FT-IBBCEAS with a resolution 0.08 cm−1. The region between 6000 and 6600 cm−1 is part of the v = 2.5 polyad. The two dashed vertical lines (red) in the R-branch of the ν1 + ν2 + ν3 and 2ν1 + ν2 combination bands of H2S indicate the spectral region where the GSFC (20 lines) is located. Other band centers of weaker overtones in the region are also indicated

Hydrogen sulfide (H2S) (BOC Industrial Gases, > 99.9%) was used as target gas in this proof-of-principle experiment. It is a poisonous and corrosive gas that can occur as a pollutant in industrial settings, such as oil refineries, production facilities involving petro-chemistry, gas processing plants and gas fields. Since it is heavier than air it can accumulate in badly ventilated areas and constitute a fire or explosion hazard due to its flammable nature [26]. The main source of H2S as a natural atmospheric pollutant is from volcanic activity. The changes in the H2S release rate from dormant volcanos in comparison to other trace species are used to monitor volcanic activity before potential eruptions [26]. Thus, the spectroscopic detection of H2S in industrial and environmental applications is of high relevance and motivated its usage in this GSFC application. High detection sensitivity was, however, not expected a priori, due to the comparatively small absorption cross sections of H2S in the spectral target region at ~ 6350 cm−1 and the modest mirror reflectivity.

3 Results and discussion

3.1 Near-IR overview spectrum of H2S using FT-IBBCEAS

Before matching frequency comb lines with absorption features of H2S, its near-IR spectrum was measured in the region between 6000 and 6600 cm−1 using Fourier transform incoherent broadband cavity-enhanced absorption spectroscopy (FT-IBBCEAS) [27,28,29]. FT-IBBCEAS is an approach to measure broadband cavity-enhanced spectra in the NIR region with high resolution [30,31,32]. The spectrum was measured at a static H2S pressure of ca. 3 mbar with an integration time of 60 min and a spectral resolution of 0.08 cm−1 (cavity length 644 cm, mirror reflectivity R ≈ 0.9998); for more details concerning the experimental setup, see Ref. [30]. The line positions obtained agreed with those in HITRAN [33] and the Pacific Northwest National Laboratory (PNNL) database [34] to within the resolution of the instrument. Figure 2 shows the overview spectrum of the 2.5 polyad of H2S (asymmetric top, point group C2v) containing features of 5 combination overtone bands ν1 + 3ν2, 3ν2 + ν3, 2ν1 + ν2, ν1 + ν2 + ν3, and ν2 + 2ν3. The spectrum is subject to resonance interactions between all states in this polyad. The same region was spectroscopically studied previously by Ulenikov et al. [35]. The ν1 + ν2 + ν3 (asymmetric) and 2ν1 + ν2 (symmetric) bands, centered at ~ 6289.17 and ~ 6288.14 cm−1 exhibits, are the strongest combination bands with clearly visible and overlapping P, Q and R ro-vibrational components; cf. Fig. 2. The bands ν1 + 3ν2 and 3ν2 + ν3 are significantly weaker and especially the very weak combination overtone ν2 + 2ν3 is negligible for detection applications. A transition in the R-branch at ~ 6339.868 cm−1 was used before by Modugno et al. [36] for detection of H2S at the parts per million level (~ 4 ppmv) using two-tone frequency modulation spectroscopy. More recently also Chen et al. [26] used R-branch lines of ν1 + ν2 + ν3 at ~ 6362.88 cm−1 for off-axis integrated cavity output spectroscopy.

3.2 GSFC application to cavity-enhanced H2S detection around 6350 cm−1

Owing to the limited tunability of the GSFC, ro-vibrational R-branch transitions of the ν1 + ν2 + ν3 combination band were used in this study to detect H2S in the region around 6350 cm−1, i.e., in the spectral region indicated by two vertical red lines in Fig. 2. Matching comb lines with H2S absorption features turned out to be cumbersome as the FWHM of the absorption lines (as well as the FC line widths) are smaller than the spectral resolution of the detection setup.

The HITRAN database [33] gives the H2S FWHM self-broadening and air-broadening coefficients at 296 K as ~ 0.316 and ~ 0.148 cm−1 atm−1, respectively (with negligible pressure shift). Since the overview spectrum in Fig. 2 was measured at ~ 3 mbar of H2S at room temperature, pressure broadening was negligible (< 0.001 cm−1) and the expected FWHM due to Doppler broadening at 296 K was fG ~ 0.013 cm−1. Thus, the observed FWHM of the absorption lines in Fig. 2 was limited by the resolution of the FTS of 0.08 cm−1.

The situation was similar for H2S detection with the GSFC in the static gas, where the same FTS was used in conjunction with the short optical cavity. The GSFC experiments were performed at ~ 82 mbar total pressure where the expected FWHM of absorption lines due to pressure broadening was fL ~ 0.012 cm−1 [33] (see also discussion below).Footnote 1 The value of fL is approximately the same size as the expected FWHM Doppler limit of 0.013 cm−1 leading to an anticipated Voigt FWHM of \({f_{\text{V}}} \approx 0.5346{f_{\text{L}}}+\sqrt {0.2166f_{{\text{L}}}^{2}+f_{{\text{G}}}^{2}} =0.021\;{\text{c}}{{\text{m}}^{ - 1}}\) [37]. Even though the comb lines possess a small band width of only ≈ 10−5 cm−1 [23], and even though the absolute comb position could be finely tuned in relative steps of ~ 0.004 cm−1, it was very cumbersome to overlap of comb lines with H2S absorption lines, because all cavity transmission spectra were measured with the limited spectral resolution of the FTS of 0.08 cm−1 (i.e., approximately 3–4 times the FWHM of the expected absorption lines).Footnote 2 Therefore, software was written that identifies the required FSR and optimum GSFC center position (arbitrarily chosen as the position of line 8—see Fig. 3) to match as many comb lines as possible with absorption features in the relevant spectral region (see Fig. 2). In the software, the GSFC spectrum was treated as a series of delta functions, owing to the small bandwidth of ≈ 10−5 cm−1 (300 kHz) per comb line [23]. Using the measured H2S spectrum in Fig. 2 the matching of up to three comb lines within experimental uncertainty was predicted for a center position of 6349.4 cm−1 (position of comb line 8) and an FSR of 10 GHz (~ 0.334 cm−1). The comb’s center position (line 8) was first coarsely tuned several times between ~ 6349.15 and ~ 6349.45 cm−1 (this range corresponds approximately to the FSR of the GSFC). Where the largest intensity reduction on comb line 3 was observed in the coarse tuning, the comb’s center position (line 8) was then finely tuned in steps of 0.004 cm−1 (0.001 nm), i.e., from 6349.375 to 6349.415 cm−1. For a comb center position at 6349.407 cm−1 (line 8) the best overlap between comb line 3 (at 6347.7487 cm−1) and comb line 9 (at 6349.7532 cm−1) with absorption features of H2S was obtained.

Fig. 3
figure 3

The thick black trace is a small spectral segment of the H2S reference spectrum in Fig. 2 which has been scaled at 6350.71 cm−1 to the HITRAN absorption cross sections (= blue solid dots) for a spectral resolution of 0.08 cm−1 (bold axis title, right axis). The thin black trace shows 18 comb lines of the GSFC laser (relative intensity; second title on the right axis). The red trace (left axis) shows the absorption coefficient measured on the 18 comb lines for a center position of the GSFC at 6349.407 cm−1 and a comb line spacing of 10 GHz. The red vertical arrows indicate comb lines which overlap significantly with H2S absorption features at 6347.74, 6349.74, and 6352.74 cm−1

Figure 3 illustrates the situation. The thick black trace in Fig. 3 is a small spectral segment of the H2S reference spectrum in Fig. 2, which has been scaled at 6350.71 cm−1 (ν1 + ν2 + ν3; 422–321) to the HITRAN absorption cross sections represented by the blue solid dots. The HITRAN cross section have been evaluated for a spectral resolution of 0.08 cm−1 and are only shown for the strongest lines in this spectral region. The cross-sectional values are shown on the right axis (bold axis title). The thin black trace shows the relative intensity of 18 comb lines of the GSFC, only the position of the comb lines is of relevance here. The red trace (left axis) shows the absorption coefficient measured on the 18 comb lines (discussed below) when the comb line spacing (FSR) was 10 GHz and the center position of the GSFC (line 8) was at 6349.407 cm−1. As mentioned above, for these conditions comb lines 3 and 9 overlap well with H2S absorption features at 6347.7487 cm−1 (ν1 + ν2 + ν3; 524–423) and 6349.7532 cm−1 (ν1 + ν2 + ν3; 717–606), respectively. Comb line 18 also overlaps significantly with an absorption feature at 6352.74 cm−1 (ν1 + ν2 + ν3; 827–808); these three lines are marked with red vertical arrows in Fig. 3.

The spectral separation between the peak absorptions at comb line 3 (at 6347.7487 cm−1) and comb line 9 (at 6349.7532 cm−1) is 2.00451 cm−1, which is slightly larger than 6*FSR = 2.00133 cm−1 of the comb. This leaves a potential discrepancy of 0.0032 cm−1, which is smaller than the smallest tuning step size of 0.004 cm−1 of the comb’s center and ca. three times smaller than the expected FWHM of the H2S absorption lines at the corresponding wavelengths. The discrepancy was within the measurement accuracy of the setup. Nevertheless, assuming the HITRAN line positions to be correct within 0.001 cm−1, a comb FSR of 0.334085 cm−1 (15.87 MHz larger than the 10 GHz used) would have been optimum for matching comb lines 3 and 9 with H2S absorption lines.

In comparison, the spectral separation between the peak absorptions at comb line 3 (at 6347.7487 cm−1) and comb line 18 (at 6352.7623 cm−1) is 5.0136 cm−1, which is slightly larger than 15*FSR = 5.0033 cm−1 of the comb. The discrepancy of 0.0103 cm−1 is more significant than that for lines 3 and 9. It corresponds to approximately the HWHM of the expected Voigt profile and is thus in agreement with the observation in Fig. 3, i.e., comb line 18 was not exactly positioned in the absorption line center. In conclusion, the position match of comb line 3, 9 and 18 with H2S absorption lines will never reach the theoretical optimum. The strongest absorption line at ~ 6351.08 cm−1 was unfortunately not overlapping well with comb line 13 (see Fig. 3) for the given center position of the comb and other multiple matchings were also not easily attained.

We arbitrarily used line 8 of the comb to define the comb’s center position, since this line also served as a good reference point for the intensity measurement, I0, without the sample. In the optimum comb position (at 6349.407 cm−1, see Fig. 3) line 8 does not overlap with any H2S absorption feature, and its peak intensity can be used as “reference intensity”, \(I_{0}^{{{\text{ref}}}}\), for all other comb lines. Prior to measurements with H2S the spectrum of the comb transmitted by an empty cavity (evacuated at ~ 0.1 mbar) was extensively measured and averaged. From this reference measurement intensity ratios rk for each comb line k and center comb line 8 can be calculated, \({r_{\text{k}}}={{I_{{0,{\text{k}}}}^{{{\text{ref}}}}} \mathord{\left/ {\vphantom {{I_{{0,{\text{k}}}}^{{{\text{ref}}}}} {I_{{0,8}}^{{{\text{ref}}}}}}} \right. \kern-0pt} {I_{{0,8}}^{{{\text{ref}}}}}}\). For a measurement with H2S at first the cavity transmission intensities Ik were stability corrected by dividing through the intensities recorded simultaneously by the OSA. Then the ratio rk was used to evaluate I0 on a line that matches a H2S absorption by multiplying rk by the peak intensity of comb line 8; I0,k = rk I0,8. The absorption coefficient αk on a matching comb line k can then be calculated from [27, 29]:

$${\alpha _{\text{k}}}=\frac{{1 - R}}{d}\left( {\frac{{{r_{\text{k}}}{I_{0,8}}}}{{{I_{\text{k}}}}} - 1} \right).$$
(1)

The absorption coefficients for all αk in Fig. 3 (red trace, left axis) were calculated using Eq. (2). For the two most relevant lines 3 and 9, absorption coefficients \({\alpha _3}(6347.7487\,{\kern 1pt} {\text{c}}{{\text{m}}^{ - 1}})=2.93 \times {10^{ - 5}}\,{\kern 1pt} {\text{c}}{{\text{m}}^{ - 1}}\) and \({\alpha _9}(6349.7532\,{\kern 1pt} {\text{c}}{{\text{m}}^{ - 1}})=1.06 \times {10^{ - 5}}\,{\text{c}}{{\text{m}}^{ - 1}}\) were determined (using R = 0.993 ± 0.001 and d = 66 cm). Since the H2S absorption cross section, \({\sigma _3}(6347.7487\,{\kern 1pt} {\text{c}}{{\text{m}}^{ - 1}})=1.17 \times {10^{ - 21}}\,{\text{c}}{{\text{m}}^2}\) and \({\sigma _9}(6349.7532\,{\kern 1pt} {\text{c}}{{\text{m}}^{ - 1}})=4.91 \times {10^{ - 22}}\,{\text{c}}{{\text{m}}^2}\) are known from the HITRAN database (converted to instrument resolution, blue dots in Fig. 3), the number density, \(n={\alpha _{\text{k}}}(\tilde {\nu })/{\sigma _{\text{k}}}(\tilde {\nu })\), can be calculated. For absorption features on comb lines 3 and 9 we found n3 = 2.50 × 1016 and n9 = 2.16 × 1016 cm−3, corresponding to a discrepancy of the number density of ~ 13.6%. There are several reasons for this discrepancy: (1) the H2S cross sections in the HITRAN database have a reported uncertainty of 10% [30] which on two different lines could lead to a max error of 20%. (2) Even though for this measurement the matching of the comb lines to the maximum of the absorption features on line 3 and 9 was satisfactory, tuning instabilities of the GSFC may cause the comb lines to be not exactly at the maximum of the absorption feature leading to an uncertainty in the expected relative absorption signals. When the GSFC was tuned in between measurements the reproducibility of the number densities on the different absorption lines is decreased due to this effect. The reason is that signal optimization is done at a spectral resolution of only 0.08 cm−1 (limited by the FTS). This impinges on the overall error of the measured number densities as further discussed in Sect. 3.2. (3) The comb intensity was recorded for ~ 4 h continuously and the absolute upper limit of the observed intensity fluctuations was found to be ≈ 3.3%. Systematic uncertainties that may impinge on the absolute value of the number density are the measured value of the mirror reflectivity (0.1%) and the length of the cavity (0.3%), which are considered negligible in comparison to the uncertainties listed above.

The averaged number density of 2.33 × 1016 cm−3 in this measurement corresponds to 0.92 ppTv. This mixing ratio, calculated from the absorption of sample inside the cavity, was a factor of ~ 35 lower than that initially expected from the partial pressure measured upon sample preparation. For the preparation of H2S/air gas mixtures a vacuum line with a mixing volume of ~ 10 dm3 made of Pyrex glass was set up inside a fume hood because of the hazardous nature of H2S. For the αk measurement in Fig. 3 mixtures of H2S in air were prepared at a ratio of 6:4 at room temperature (assuming no H2S losses, see below), i.e., the mixing volume was first filled with H2S at ~ 400 mbar and then air at a partial pressure of ~ 600 mbar was added. The mixture was collected for transfer to the cavity setup (Fig. 1) in a small detachable Pyrex glass tube which was connected on one side of the vacuum line via a Teflon valve. The volume of this “transport tube” was ~ 82 cm3. As the cavity volume was 998 cm3 the partial pressures of air and H2S in the cavity were expected to be 49.3 and 32.9 mbar, respectively (provided there are no losses of H2S when filling the transfer volume). The latter pressure value corresponds to a mixing ratio of 32.4 ppTv under standard conditions (1 atm, 296 K), which is ca. 35 times larger than the amount of H2S found in the absorption experiment (see above). The reason for this difference is the significant sample loss in the mixing chamber and transfer volume, as both consisted of Pyrex glass causing inherently large wall losses. Due to efficient chemisorption at metal ion sites in the glass, the wall losses in the mixing and transport volume are far from negligible. The fractional throughput of H2S through a Pyrex glass enclosure is known to be approximately 3% [38], corresponding to losses of a factor ~ 33 in agreement with the loss factor found in the absorption experiment. Although the data in Ref [38] are not universal for arbitrary experimental conditions, the low fractional throughput of Pyrex indicates that much less H2S was collected in the transfer volume. Note that the filling pressure was not measured in the transfer volume itself, but at a distance from it in the larger mixing volume, which was essentially filled with a large excess of air. After the transfer of the sample to the GSFC setup the pressure was measure in the cavity again.

To systematically study the effect of losses in the transfer of gas mixtures from the mixing volume to the cavity, the absorption coefficients were determined on laser lines 3 and 9 for five different H2S concentrations. For these measurements the mixing volume was initially filled with different partial pressures of H2S between 100 and 500 mbar, then filled up with air to 1000 mbar, and finally transferred to the cavity in the transfer volume. The absorption coefficients on lines 3 and 9 had an overall uncertainty of Δmax < 25% [see Table 1; this error is not to be confused with the difference (discrepancy) in the number densities based on measurements on comb lines 3 and 9]. The maximal uncertainty of 25% is associated with the reproducibility of the measured absorption coefficient per comb line. For each partial pressure of H2S the absorption coefficient was measured 3–4 times. Between each measurement the GSFC was finely tuned to optimize the signal and to obtain an uncertainty in the reproducibility for a given H2S concentration, which was evaluated based on a Student’s t distribution. Apart from contributions to Δmax that are associated with the uncertainty of the H2S cross sections in HITRAN (10%), the optimum position of the comb lines with respect to the absorption maximum was difficult to establish from measurement to measurement as a detection resolution of 0.08 cm−1 was used for signal optimization. This resolution is larger than the FWHM of the absorption line at the measurement conditions (~ 0.021 cm−1) and much larger than the comb line width (~ 10− 5 cm−1). The accuracy, therefore, relied strongly the tuning stability of the GSFC, causing a fluctuation in the absorption coefficient and a discrepancy in the relative absorption coefficients measured on line 3 and 9. The 3σ standard deviation of the absorption coefficient on lines that were not matched perfectly with H2S absorption features was ≈ 3.6 × 10−6 cm−1. This value of the minimum absorption coefficient in an integration time of 120 s corresponds to ≈ 680 ppmv, which can be taken as a conservative (upper) limit of detection (LOD) for this approach, because some of the laser lines always overlap to a certain extent with H2S absorptions, all of which were taken into account in the above-stated 3σ LOD. From the H2S absorption measurements the number densities in the cavity was again estimated based on HITRAN cross sections, which were systematically (expectedly) smaller than those estimated from the pressure in the mixing volume due to the losses of H2S in the transfer process. An expected loss factor of ca. 33 (averaged) was found for comb line 9, while the loss factor appeared to be larger (average ~ 50) for line 3. This difference corroborates that the comb lines 3 and 9 were not equally well matched to the maximum of a H2S absorption line. While line 9 was probably near the maximum of the absorption feature (owing to the smaller loss factor), line 3 was positioned off the center of the H2S absorption line. Within our spectral resolution of only 0.08 cm−1 the exact position cannot be independently verified. The fact that the ratio of loss factors is approximately constant for all concentrations indicates that the approach is reproducible within the above-stated limits, i.e., the laser position and mode structure enabling the light to be transmitted by the cavity is sufficiently stable in the current approach. The data are summarized in Table 1 and graphically represented in Fig. 4.

Table 1 Summary of measurement parameter concerning concentration-dependent measurements
Fig. 4
figure 4

H2S number density in the cavity as measured from the absorption coefficient on comb line 3 (blue circles) and line 9 (red circles) vs the alleged partial pressure of H2S in the mixing volume. The triangular symbols represent the number densities derived from the partial pressures on the horizontal axis, divided by a loss factor of 50 (blue) and 33 (red). They can then be compared with the number densities from the absorption measurement. Error bars refer to the uncertainties stated in column 4 of Table 1

Figure 4 shows, on the horizontal axis, the partial pressure of H2S as measured in the mixing volume. On the vertical axis the number densities inside the cavity are shown (circular symbols). They were determined from the absorption coefficient measured on line 3 (blue data points) and line 9 (red data points) for the respective mixing partial pressures. The error bars refer to the circular symbols. Additionally displayed are the number densities (triangular symbols, straight black line, no error bars) as they would be expected from the partial pressures, divided by the average loss factor for H2S for the corresponding line (50 for line 3 and 33 for line 9; cf. Table 1). Figure 4 illustrates the consistency of Table 1 for measurement at different concentrations.

The first value in Fig. 4 (100 mbar) is essentially the detection limit. The lower explosion limit (LEL) of H2S in air is ~ 4% while the upper explosion limit (UEL) is ~ 44% [39]. Since the LOD of ~ 0.7 per mille by volume is significantly smaller than the lower explosion limit (LEL) [39], the GSFC measurements presented here are well sufficient to detect H2S at its LEL even in the near-infrared.

Finally, an attempt was made to corroborate our results in Figs. 3 and 4 through the measurement of a high-resolution line spectrum, by recording absorption coefficients on line 3 while stepping the center of the GSFC over the optimum position at 6349.407 cm−1 (line 8 in Fig. 3) in steps of 0.004 cm−1. The problem with this experimental line width test was that all absorption measurements were taken with a resolution of 0.08 cm−1, and the true position of the laser with respect to the line center of the absorption line at 6347.74 cm−1 (line 3) could not be determined with sufficient accuracy. The test measurement relied solely on the tuning stability of master laser. Even though the spectrum of the line at 6347.74 cm−1 could be recorded by tuning the GSFC, the FWHM was ca. half the size of the anticipated Voigt FWHM (0.021 cm−1) according to the estimated pressure and Doppler broadening, see above. This discrepancy is likely to be a consequence of the electronic tuning stability and potential small drifts of the GSFC. The stability was investigated by independently tuning to the GSFC center position to the maximum absorption signal many times in the context of the I0 measurements. The absolute laser position, as solely determined from the tuning voltage, could be reproduced within 0.017 cm−1, which is 20% smaller than the anticipated FWHM of the expected H2S absorption line. The tuning to the maximum position is uncertain within that limit, explaining the different loss factors found for the laser comb lines 3 and 9.

4 Conclusions and outlook

Two comb lines (3 and 9) were matched to absorption lines of H2S in the near-infrared L-band without locking the laser to the cavity mode structure. Without locking the laser line, the overlap of the laser with the absorption features was accurate and stable within ≈ 80% of the FWHM of the absorption line. From the non-resonant comb lines, an rms noise equivalent absorption for this approach of approximately 3.9 × 10−5 cm−1 Hz−1/2 was found, which corresponds to a detection limit of ca. 700 ppmv of H2S in 120 s. This sensitivity is well below the lower explosion limit of H2S of 4% and was achieved with a very modest finesse (~ 450) cavity.

Future work will focus on making the setup more compact using custom-designed integrated combs and improving the detection resolution through a dual-comb approach. Alternative coupling scheme of the laser to the cavity will be looked into and the detection cell is to be developed into a flow cell. Other spectral regions are to be targeted.