Introduction

There are currently several near-infrared spectroscopy (NIRS) oximetry devices commercially available and used for detecting cerebral ischemia during cardiopulmonary bypass (CPB). These devices may help surgeons and anesthesiologists avoid poor cerebral perfusion, which can result in perioperative neurological adverse events. These devices operate on the same measurement principles; however, their algorithms of detection and calculation of regional cerebral oxygen saturation (rSO2) value are different, and it is inappropriate to compare values measured by such different devices directly. Thus, measured values and trends fluctuate depending on devices.

Our previous retrospective evaluation of 223 adult patients undergoing cardiac surgery has revealed that brain atrophy, poor left ventricular ejection function, anemia, and hemodialysis had significant correlations with low initial cerebral rSO2 values measured by INVOS 5100C (Medtronic, Minneapolis, MN, USA) [1]. In clinical settings, these fluctuations of measured values affect the decision making for intervention during CPB.

In this study, 70 patients were randomly assigned to 1 of 2 groups in which surgery was performed using 2 different devices: the INVOS 5100C and the FORE-SIGHT ELITE (CAS Medical Systems, Branford, CT, USA). We prospectively evaluated intraoperative rSO2 values in each group and the presence or absence of intervention for low rSO2 value during CPB.

Subjects

From January 2017 to August 2017, 96 patients underwent elective cardiac surgery at Kitasato University Hospital. Patients ≤ 12 years of age; patients who did not undergo preoperative brain magnetic resonance imaging; and patients with a medical history of acute cerebral infarction, intracranial hemorrhage or neurosurgery, surgery performed in a nonsupine position, off-pump coronary artery bypass surgery, and/or with left ventricular assist devices were excluded from the study. Finally, 70 patients were enrolled.

Methods

Each patient was randomly assigned to 1 of 2 groups utilizing either the INVOS 5100C (group I; n = 35) or the FORE-SIGHT ELITE (group F; n = 35). This research was conducted with the approval of the Kitasato University Medical Ethics Organization (approval number: KMEO B16-96). All patients gave informed consent before inclusion in the study.

Scalp–Cortex distance

Preoperative brain magnetic resonance imaging was routinely evaluated for detection of cerebrovascular disease. Scalp–Cortex distance (SCD) was measured on a T1-weighted image of the mid-sagittal plane as the distance from the surface of the scalp on the upper forehead, at 4 cm from the superciliary arch, to the frontal cerebral cortex (Fig. 1).

Fig. 1
figure 1

Scalp–Cortex distance (SCD) was defined as the distance between the scalp surface of the forehead at 4 cm from the superciliary arch and the frontal cerebral cortex in reference to a midsagittal plane of a T1-weighted magnetic resonance image

Cerebral rSO2 monitoring

From anesthetic induction to the end of surgery, continuous monitoring of cerebral rSO2 was carried out in each group. Optodes of each device were placed on the patient’s forehead, at 4 cm from the superciliary arch, to avoid the frontal sinus [2]. The values were recorded at anesthetic induction, CPB start, aortic cross clamp, and CPB end.

Interventions during CPB

Interventions for low rSO2 during CPB were defined as approaches to improve rSO2 values such as increasing perfusion flow rate, increasing inspired oxygen fraction, and blood transfusion. Increasing perfusion flow should be considered as primary intervention method, and the others as supplemental. Since there is no established value of rSO2 that should be maintained during CPB to avoid postoperative cognitive disorder, the thresholds for intervention were set as an absolute drop to below 40% or a 10% relative decrease from the initial baseline value. These thresholds were based on the recommendation by the manufacturer of INVOS 5100C, Medtronic, with clinical modifications.

Data collection

Patient medical records were reviewed for preoperative clinical and demographic characteristics according to the hospital information system.

Data analysis

Continuous variables were presented as mean ± standard deviation. For each group, odds ratios for the presence of intervention and correlations between rSO2 value and SCD were evaluated.

Differences with P < 0.05 were considered statistically significant. Statistical analyses were performed using JMP Software, version 12.0 (SAS Institute, Carey, NC, USA).

Results

Patient characteristics

Patient characteristics including age; sex; body surface area; medical history (stroke, hypertension, diabetes mellites, and hemodialysis); preoperative left ventricular ejection fraction; hemoglobin concentration (Hb); and SCD are shown in Table 1. There were no significant differences between the 2 groups.

Table 1 Patient characteristics

Operative outcomes

Surgical procedures, CPB time, perfusion index, minimum core temperature, blood transfusion during CPB, and postoperative neurological adverse events are shown in Table 2. In group I, 1 patient of aortic valve replacement experienced convulsions and 1 patient of aortic valve replacement, tricuspid valve annuloplasty, coronary artery bypass and maze procedure experienced delayed emergence. However, there were no significant differences between the 2 groups.

Table 2 Operative outcome

Intraoperative rSO2 values

The rSO2 values were recorded at anesthetic induction, CPB start, aortic cross clamp, and CPB end. The values for group F are significantly higher than those for group I. However, there were no significant differences in mean mixed venous oxygen saturation value in the 2 groups (Table 3).

Table 3 Intraoperative rSO2 values

Interventions for low rSO2 value

Interventions for low rSO2 during CPB for group I and group F were increase perfusion flow (13:5; P = 0.03), blood transfusion (7:1; P = 0.02), and both (6:1; P = 0.04), respectively. Interventions were significantly more frequent in group I, despite similar patient backgrounds (Table 4). The cases of high perfusion pressure of > 400 mmHg at the inlet of the artificial lung were managed only with blood transfusion without increasing perfusion flow.

Table 4 Intraoperative intervention for low rSO2 value

SCD and rSO2 values

In group F, there was no correlation between the mean rSO2 value and SCD. In contrast, the mean rSO2 value in group I had a significant negative correlation with SCD (Fig. 2). In group I, in which mean rSO2 and SCD were significantly correlated, the relationship between SCD and the presence or absence of intervention was examined. The mean SCD of interventions were significantly longer than that of noninterventions (17.1 ± 2.5 vs 15.1 ± 1.6 mm; P = 0.007). A cutoff value of SCD for group I was determined by ROC analysis. The area under the curve was 0.75, and a cutoff value for intervention was SCD > 17.6 mm (Table 5) (Fig. 3).

Fig. 2
figure 2

There is a significant negative correlation between mean rSO2 value and Scalp–Cortex distance (SCD) in group I. In group F, measured mean rSO2 value is not affected by SCD

Table 5 Intervention and Scalp–Cortex distance of Group I
Fig. 3
figure 3

Result of receiver operating characteristic analysis for the evaluation of decision making for intervention for desaturation during CPB in group I. Area under the curve (AUC) was 0.75. Cutoff values for intervention was Scalp–Cortex distance of > 17.6 mm

Discussion

Since CPB surgery was first introduced, perioperative stroke emerged as a serious adverse event. In 1965, Pillsbury et al. reported their 450 consecutive open-heart surgeries. The incidence of stroke was 2.0% (n = 9), and the mortality rate was 55.6% (n = 5) [3]. Recent meta-analysis revealed that the incidence of perioperative stroke after cardiac surgery was 2.03%. The operative and late mortality were 21.3 and 10.9%, respectively [4]. It is inappropriate to compare these studies with studies reported by clinicians with different historical and clinical backgrounds. However, despite much innovative progress, perioperative stroke remains a complication that concerns surgeons.

Real-time, noninvasive monitoring of tissue oxygenation using an NIRS oximetry device was first reported in 1977 [5]. Today, clinical utility of such devices has been established and commonly applied to detect cerebral ischemia during CPB surgery. In 184 Japanese medical centers in which cardiac surgery was performed, 155 (84%) used intraoperative NIRS oximetry monitoring, and 70 (38%) used it routinely [6]. Although several NIRS oximetry devices are commercially available, their algorithms of detection and calculation of rSO2 value are different. Thus, measurement results may vary, and factors that influence measurements must be weighed differently [6].

Fluctuations in measured rSO2 values with the INVOS 5100C during cerebral desaturation are reportedly significantly wider than those with the EQUANOX 7600 (Nonin Medical Inc, Plymouth, MN, USA), and a similar but nonsignificant trend was observed for the INVOS 5100C compared with the FORE-SIGHT ELITE [7]. Greenberg et al. reported that the FORE-SIGHT ELITE was significantly less affected by extracranial contamination such as dermal ischemia than the INVOS 5100C [8]. These reports suggest that vulnerability of noise-tolerance specifications of INVOS 5100C. In addition, recent report suggests that rSO2 alone is insufficient in describing cerebral metabolism and predicting neurological outcome [9].

In our study, intraoperative interventions for lower rSO2 value were performed more frequently in group I than in group F. Measured decreasing rSO2 values in group I were significantly affected by longer SCD, which means that brain atrophy may influence rSO2 values, depending on device selection. Inappropriate intervention for pseudo-lower rSO2 values may have been performed without the presence of actual ischemia. In particular, SCD of longer than 17.6 mm in group I may misguide perfusionists to perform unnecessary or excessive interventions. In such cases, INVOS 5100C should not be used. An appropriate application of the intervention threshold such as decreasing rSO2 value of 40% assumes that the value is less affected by non-ischemic parameters. It is unlikely that there were more cases of cerebral ischemia in group I than in group F in a similar clinical populations. Moreover, it is inappropriate to use the same intervention criteria for different NIRS oximetry devices during CPB. Further study will be required to determine optimal value of rSO2 to be maintained for each NIRS device.

Study limitations

The limitation of this study is small sample size (≤ 100 patients) in a single-center design. Moreover, thresholds for intervention against desaturation may vary depending on each clinical team. Additional multicenter studies should be carried out in larger patient populations to investigate device selection and measured cerebral rSO2 values during CPB surgery and the impact on the need for interventions.

Conclusion

We report a prospective study to investigate the relationship between NIRS oximetry device selection and measured cerebral rSO2 values during CPB and its influence on whether interventions to improve desaturation should be performed. In the INVOS 5100C group (group I) in our study, SCD longer than 17.6 mm may compel clinicians to perform excessive interventions. In such cases, INVOS 5100C should not be used. Moreover, it is inappropriate to use the same intervention criteria for different NIRS oximetry devices during CPB. Device selection affect intervention for desaturation during CPB. Appropriate device selection should be considered for safe and minimally invasive cardiac surgery.