Abstract
An efficient and stable preconcentration method for ionic and ionizable analytes, named radical electric focusing solid phase extraction (REFSPE), was proposed here for enriching trace Cu(II), Ni(II), and Co(II) from water samples. In REFSPE, radical electric field was introduced into solid phase extraction system to enhance the mass transfer rate of heavy metal ions in aqueous solution and shorten the extraction time obviously. The concentration of mixed heavy metal ions were predicted using partial least squares (PLS) regression by analyzing ultraviolet and visible spectrum efficiently. By optimizing chromogenic agent, pH values, extraction voltage, extraction time, and PLS parameters, the proposed method had higher figures of merit. The comparison of the determination data and the calculation results showed that the proposed method can provide a favorite quantitative precision with expanded measuring range. The limit of detection (LOD) for Cu(II), Ni(II), and Co(II) were 0.10, 0.13, and 0.15 μg L−1, respectively. The reliability was confirmed by the RSD lower than 5% and the recovery of 95–102%. The novel enrichment method has a great potential for applications in detecting different ionic and ionizable analytes with the help of spectrometry determination and chemometrics calculation.
Similar content being viewed by others
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.Avoid common mistakes on your manuscript.
1 Introduction
With the rapid development of industries such as metal plating facilities, mining operations, fertilizer industries, tanneries, batteries, paper industries, waste disposal, and pesticides, heavy metal wastewaters are directly or indirectly discharged into the environment increasingly, especially in developing countries (Fu and Wang 2011; Qiao et al. 2013; Duruibe and Ogwuegbu 2007; Bailey et al. 1999). Pollution of the natural aquatic environment by heavy metals is a problem worldwide, and heavy metals have been paid much attention due to their toxicity, persistence, non-biodegradation, and bioaccumulation in food webs (Facchinelli et al. 2001; Li et al. 2015; Ayenimo et al. 2010). Therefore, accurate detection and monitoring of heavy metal pollution in different water environments to protect the people and the environment is extremely important.
As we have seen, the methods including atomic absorption spectroscopy (Kumar et al. 2001; Trindade et al. 2015), inductively coupled plasma mass spectrometry (ICP-MS) (Silva and Curtius 2000; Nardi et al. 2009), inductively coupled plasma optical emission spectrometer (Losev et al. 2015; Massadeh et al. 2016), ultraviolet and visible spectrophotometry (UV-Vis) (Feng et al. 2009; Zanjanchi et al. 2006), stripping voltammetry (Tukur et al. 2015), liquid chromatography (Peñalver et al. 2002), X-ray fluorescence spectrometry (Sitko 2015), and sensor (Forzani et al. 2005) are usually used to detect heavy metal ions. In spite of their widespread use, these techniques have more or less limitations including high cost, complex pre-treatment procedures, large and heavy analytical instrument requirements, skilled personnel, or expensive to run and maintain (Wei et al. 2014; Zhan et al. 2015). Accordingly, a simple, sensitive and economical method for simultaneous detection of multiple heavy metal ions is desirable (Huang et al. 2013).
In view of the abovementioned background, UV-Vis method is the most widespread and most appealing analytical technique because of its simplicity, reliability, cheap price, and ease-of-operation. However, it is difficult for the traditional UV-Vis approach in two situations: when simultaneous determination of multiple components is required and when the sample concentration is super low. As for the quantitation of different heavy metal ions simultaneously, chemometrics methods can be introduced into the data processing of spectrogram (Jia et al. 2014; Nakajima et al. 2015; Zeiner et al. 2010). In recent years, a variety of multicomponent concentration was determined with the combination method of chemometrics and spectrophotometric analysis (Juan and Tauler 2003; Jiang et al. 2002; Ni et al. 2006). The combination method can be applied to the analysis of multicomponent system quickly and accurately. As for the quantification of trace level samples accurately, new sample pre-treatment methods will achieve twice the result with half the effort. Nowadays, various techniques have been applied for the preconcentration of trace heavy metals from environmental samples, such as adsorption, liquid–liquid extraction (Anthemidis et al. 2004), cloud point extraction (Svegl and Ogorevc 2000; Talio et al. 2009), solid phase extraction (SPE) (Guo et al. 2017; Su et al. 2015), and co-precipitation (Feist and Mikula 2014). Considering the advantages and disadvantages of various methods, SPE as effective extraction technique become a preferred candidate (Mehdinia and Aziz-Zanjani 2013; Puig et al. 2008) except that it is kind of time-consuming and labor-intensive (Hennion 1999). In order to accelerate the mass transfer rate of analytes in aqueous samples and shorten the operation time, the radical electric focusing solid phase extraction (REFSPE) was proposed here. An electrostatic field, created by direct-current power supply in REFSPE, can drive charged ions to migrate orientationally. Charged ions were centralized to the membrane placed at the center of extractor and the preconcentration process of the trace heavy metals in the water completed.
Therefore, the combination of REFSPE, UV-Vis detection, and improved MATLAB® partial least squares (PLS) regression was used in the quantification of trace heavy metal ions mixture in aqueous sample. All the influence factors such as chromogenic agent, pH values, extraction voltage, extraction time, and PLS parameters were studied and optimized. The enrichment performance of various heavy metal ions was validated by the comparison of the proposed method and ICP-MS.
2 Materials and Methods
2.1 Reagents and Samples
Stock standard solutions for Ni(II), Co(II), and Cu(II) (1000 mg L−1) were prepared by dissolving analytical reagent grade Ni(NO3)2·6H2O, Co(NO3)2·6H2O, and CuSO4·5H2O (Tianjin Kermel Co., Ltd., China) into deionized water separately. 0.05 mg L−1 solution of Cu(II), Ni(II), and Co(II) prepared from the abovementioned stock standard solutions was used for electric strengthen adsorption. Buffer of acetic acid-sodium acetate (CH3COOH/CH3COONa) was prepared by mixing appropriate volume of pure acetic and sodium acetate to pH 5–6. As chromogenic reagent, 0.3% xylenol orange (XO) was prepared by dissolving appropriate XO in deionized water. Polyvinylidene fluoride (PVDF) membranes, which are with high porosity and high surface to volume ratio, were purchased from Tianjin Motian technology Co., Ltd., China. Deionized water was used for all the dissolution and dilution process in the experiments. All the glassware were carefully cleaned with specific HNO3 solution and rinsed with distilled water before use.
2.2 Instruments
The direct-current power supply of the preconcentration process was KIKUSUI PMC18–5 (Kikusui Electronics Corporation, Japan), and the 0.5-mm platinum electrode was from a hardware supply store. Ultrasonic cleaning machine (Shanghai Crown Ultrasonic Instrument Co., Ltd., China) was used for desorption. The EVOLUTION 300 UV-Vis spectrometer (spectra range from 200 to 800 nm), which was selected as the detection tool for all the samples, was purchased from Thermo Fisher Scientific Co., Ltd., USA. The PHS-W series laboratory pH meter was purchased from Shanghai Bante Instruments Co. Ltd., China. Agilent 7900 ICP-MS from Agilent Technologies was used for validating the proposed detection method in the paper.
REFSPE, which was carried out on a self-made instrument, adopted a radial electric field formed by a cylindrical electrode surrounding a single electrode in a central location creatively. In this structure, there is a stronger and uniform electric field force, which can drive ions to migrate faster and shorten the extraction time significantly, in the extraction system without any mixing device. At the same time, an automatic precision three-dimensional lifting platform was used in the experiments in order to ensure that each experiment operating conditions are exactly the same. As a result, the detection accuracy and repeatability increased obviously. The device diagram of REFSPE was shown in Fig. 1.
2.3 The Preparatory Work
Some hollow PVDF membrane pieces of 2.5 cm were activated by dipping them in 2.0 mol L−1 anhydrous ethanol for 30 min at room temperature before rinsed by distilled water. After that, the platinum electrode, which was inserted into the PVDF-activated membrane, was connect with the negative electrode of the power supply and placed on the true center of the extractor during the extraction process.
2.4 The Preconcentration Process
At the beginning of the extraction experiment, 50 mL sample solution was enclosed into a homemade titanium plating ruthenium cylindrical container with an inner diameter of 4.8 cm and height of 8 cm, which served as a positive pole. Then, the extraction procedure was completed under the assistance of a direct current electrical field for some time. After that, the mixed ions adsorbed on hollow PVDF membrane were eluted by 2.0 mL NaAc-HAc buffer solution with the help of an ultrasonic cleaner.
2.5 Data Acquisition and Data Mining
After the desorption process, 1.0 mL color developing agent XO was added into the eluent to make UV-Vis detection much more sensitive. Then, the UV-Vis spectra data, obtained from EVOLUTION 300 UV-Vis spectrometer, were disintegrated into individual of each ion with the help of PLS. Upon the MATLAB® calculation, certain concentrations of each ion were acquired. The entire operation process was shown in Fig. 1.
3 Results and Discussion
In order to obtain an effective and sensitive detection for trace mixed heavy metal ions in aqueous sample, the conditions of enrichment and detection such as pH values, extraction time, operation voltage, elution time, and volume of XO were optimized, and the concentration of heavy metal ions was calculated from the enriched sample solution.
3.1 Effect of pH for Measured Sample
From the practical point of view, an appropriate pH can increase the sensitivity of metal ion complex and coordination agent. In order to find the optimal pH value, the absorbance values of heavy metal ions with the pH values from 5 to 6 were measured. The relationship between absorbance and pH data were displayed in Fig. 2. It was shown that the absorbance data had the highest level at the pH of 5.6. Therefore, pH = 5.6 was chosen as the detection pH value for all the experiments.
3.2 Effect of Volume of 0.3% XO Solution for Detection
Coordination agent XO was mixed with three kinds of ions and had an occurrence of obvious chromogenic reaction. The mixture of eluent solution and different volumes of 0.3% XO were detected by UV-Vis spectrometer. The absorbance data of different additive amounts of 0.3% XO solution were illustrated in Fig. 3. It was illustrated that with the increase of reagents, absorbance tends to be stable; however, the color effect became weak due to the dilution effect of the 0.3% XO solution if excessive 0.3% XO solution is used. Therefore, 1.0 mL 0.3% XO solution was chosen for the further experiments.
3.3 Effect of Adsorption Time and Voltage
The time profiles of REFSPE were depicted in Fig. 4 at different operation voltages. It is found that the enrichment factors were constant when the adsorption equilibrium was achieved. The results are also shown in Fig. 4 that the less adsorption equilibrium time and greater enrichment factor can be reached when the higher adsorption voltage in the range of 3–9 V was adopted. On the other hand, there were some pure metals of analyte coating on the extraction material, which could reduce the preconcentration result and had an obvious effect on the subsequent experiments, when the adsorption voltage was higher than 9 V. Therefore, the optimal adsorption time was chosen at 11 min with the operation voltage of 9 V.
3.4 Effect of Desorption Time
In order to collect the preconcentrated metal ions during the adsorption process, ultrasonic elution method was selected owing to its effective elution behavior. In the elution process, the hollow fiber membrane after adsorption was put into 2 mL buffer solution with pH = 5.6 for supersonic elution. The relationship of enrichment result and desorption time were shown in Fig. 5. According to the time profile, the stripped ion content increased until the elution content became constant after 3 min. Therefore, 3 min was chosen as the optimal desorption time in the following experiments.
3.5 Method Evaluation
In order to evaluate enrichment properties of REFSPE for heavy metal ion mixture, partial least squares of latent structure model were introduced into the calculation of mixed ions. The PLS model built for the quantification of heavy metal ions mixture was from UV-Vis spectra of 50 random samples ranged from 0.001 to 2.5 mg L−1. The root-mean-square error of cross validation (RMSECV), which was an important index for predictive value through using the optimal factor number of latent variables (LV), were calculated and shown in Fig. 6. On the basis of the computed result, RMSECV were 0.1417, 0.0855, and 0.0751, when the LV for Cu(II), Co(II), and Ni(II) in sample solutions were confirmed at 5, 12, and 12, respectively. The relationships between predictive values and truth values of each ion were all higher than 0.99.
3.6 Analytical Performance
Under the optimal experimental conditions, the limit of detection (LOD) for Cu(II), Ni(II), and Co(II) were 0.10, 0.13, and 0.15 μg L−1, respectively. In order to estimate the reliability of the proposed method, the calculation results for the enriched sample from 0.05 mg L−1 solution for each ion were compared with the determination values of ICP-MS. The comparison data were shown in Table 1 with RSD (n = 5) lower than 5% and recovery values in the range of 95–102%.
4 Conclusions
Simultaneous spectrophotometric quantization method for trace heavy metal ions in water samples using REFSPE coupled with PLS approaches was proposed here. In REFSPE, radical electric field was introduced into an SPE system to enhance the mass transfer of heavy metal ions in aqueous solution and shorten the operation time of the sample enrichment process and the experimental results proved that REFSPE can be developed as a new effective sample preparation method for ionic and ionizable analytes. The concentration of mixed heavy metal ions can be calculated through PLS approaches by analyzing UV-Vis spectrum efficiently. The LOD for Cu(II), Ni(II), and Co(II) of the proposed combination method were found at 0.10, 0.13, and 0.15 μg L−1, respectively. The reliability was confirmed by the RSD lower than 5% and the recovery of 95–102%. The experimental study shows that REFSPE is a great potential for applications in detecting trace metal ion samples with the help of spectrometry determination and chemometrics calculation.
References
Anthemidis, A. N., Zachariadis, G. A., Farastelis, C. G., & Stratis, J. A. (2004). On-line liquid-liquid extraction system using a new phase separator for flame atomic absorption spectrometric determination of ultra-trace cadmium in natural waters. Talanta, 62(3), 437–443.
Ayenimo, J. G., Yusuf, A. M., Adekunle, A. S., & Makinde, O. W. (2010). Heavy metal exposure from personal care products. Bulletin of Environmental Contamination and Toxicology, 84(1), 8–14.
Bailey, S. E., Olin, T. J., Bricka, R. M., & Adrian, D. D. (1999). A review of potentially low-cost sorbents for heavy metals. Water Research, 33(11), 2469–2479.
Duruibe, J. O., & Ogwuegbu, M. O. C. (2007). Heavy metal pollution and human biotoxic effects. International Journal of Physical Sciences, 2(5), 112–118.
Facchinelli, A., Sacchi, E., & Mallen, L. (2001). Multivariate statistical and GIS-based approach to identify heavy metal sources in soils. Environmental Pollution, 114(3), 313–324.
Feist, B., & Mikula, B. (2014). Preconcentration of some metal ions with lanthanum-8-hydroxyquinoline co-precipitation system. Food Chemistry, 147(6), 225–229.
Feng, Q. Z., Zhao, L. X., Yan, W., Lin, J. M., & Zheng, Z. X. (2009). Molecularly imprinted solid-phase extraction combined with high performance liquid chromatography for analysis of phenolic compounds from environmental water samples. Journal of Hazardous Materials, 167(1–3), 282–288.
Forzani, E. S., Zhang, H., Chen, W., & Tao, N. (2005). Detection of heavy metal ions in drinking water using a high-resolution differential surface plasmon resonance sensor. Environmental Science & Technology, 39(5), 1257–1262.
Fu, F., & Wang, Q. (2011). Removal of heavy metal ions from wastewaters: a review. Journal of Environmental Management, 92(3), 407–418.
Guo, Y., Zhao, H., Han, Y., Liu, X., Guan, S., Zhang, Q., et al. (2017). Simultaneous spectrophotometric determination of trace copper, nickel, and cobalt ions in water samples using solid phase extraction coupled with partial least squares approaches. Spectrochimica Acta Part A: Molecular and Biomolecular Spectroscopy, 173, 532–536.
Hennion, M. C. (1999). Solid-phase extraction: method development, sorbents, and coupling with liquid chromatography. Journal of Chromatography A, 856(1–2), 3–54.
Huang, D., Niu, C., Wang, X., Lv, X., & Zeng, G. (2013). “Turn-on” fluorescent sensor for Hg2+ based on single-stranded DNA functionalized Mn:CdS/ZnS quantum dots and gold nanoparticles by time-gated mode. Analytical Chemistry, 85(2), 1164–1170.
Jia, J., Wu, A., & Luan, S. (2014). Spectrometry recognition of polyethyleneimine towards heavy metal ions. Colloids and Surfaces A: Physicochemical and Engineering Aspects, 449, 1–7.
Jiang, J. H., Berry, R. J., Siesler, H. W., & Ozaki, Y. (2002). Wavelength interval selection in multicomponent spectral analysis by moving window partial least-squares regression with applications to mid-infrared and near-infrared spectroscopic data. Analytical Chemistry, 74(14), 3555–3565.
Juan, A. D., & Tauler, R. (2003). Chemometrics applied to unravel multicomponent processes and mixtures : revisiting latest trends in multivariate resolution. Analytica Chimica Acta, 500(1), 195–210.
Kumar, M., Rathore, D. P. S., & Singh, A. K. (2001). Pyrogallol immobilized Amberlite XAD-2: a newly designed collector for enrichment of metal ions prior to their determination by flame atomic absorption spectrometry. Microchimica Acta, 137(3), 127–135.
Li, P., Lin, C., Cheng, H., Duan, X., & Lei, K. (2015). Contamination and health risks of soil heavy metals around a lead/zinc smelter in southwestern China. Ecotoxicology & Environmental Safety, 113, 391–399.
Losev, V. N., Buyko, O. V., Trofimchuk, A. K., & Zuy, O. N. (2015). Silica sequentially modified with polyhexamethylene guanidine and Arsenazo I for preconcentration and ICP–OES determination of metals in natural waters. Microchemical Journal, 123, 84–89.
Massadeh, A. M., Alomary, A. A., Mir, S., Momani, F. A., Haddad, H. I., & Hadad, Y. A. (2016). Analysis of Zn, Cd, As, Cu, Pb, and Fe in snails as bioindicators and soil samples near traffic road by ICP-OES. Environmental Science and Pollution Research, 23(13), 13424–13431.
Mehdinia, A., & Aziz-Zanjani, M. O. (2013). Advances for sensitive, rapid and selective extraction in different configurations of solid-phase microextraction. TrAC Trends in Analytical Chemistry, 51(11), 13–22.
Nakajima, H., Hara, K., Yamamoto, Y., & Itoh, K. (2015). Effects of Cu on the content of chlorophylls and secondary metabolites in the Cu-hyperaccumulator lichen Stereocaulon japonicum. Ecotoxicology & Environmental Safety, 113, 477–482.
Nardi, E. P., Evangelista, F. S., Tormen, L., Saint’Pierre, T. D., Curtius, A. J., Souza, S. S. D., et al. (2009). The use of inductively coupled plasma mass spectrometry (ICP-MS) for the determination of toxic and essential elements in different types of food samples. Food Chemistry, 112(3), 110–113.
Ni, Y., Lin, D., & Kokot, S. (2006). Synchronous fluorescence, UV–visible spectrophotometric, and voltammetric studies of the competitive interaction of bis(1,10-phenanthroline)copper(II) complex and neutral red with DNA. Analytical Biochemistry, 352(2), 231–242.
Peñalver, A., Pocurull, E., Borrull, F., & Marcé, R. M. (2002). Solid-phase microextraction coupled to high-performance liquid chromatography to determine phenolic compounds in water samples. Journal of Chromatography A, 953(1–2), 79–87.
Puig, P., Borrull, F., Calull, M., & Aguilar, C. (2008). Sorbent preconcentration procedures coupled to capillary electrophoresis for environmental and biological applications. Analytica Chimica Acta, 616(1), 1.
Qiao, Y., Yang, Y., Gu, J., & Zhao, J. (2013). Distribution and geochemical speciation of heavy metals in sediments from coastal area suffered rapid urbanization, a case study of Shantou Bay, China. Marine Pollution Bulletin, 68(1–2), 140–146.
Silva, M. A. M. D., & Curtius, A. J. (2000). Determination of trace elements in water samples by ultrasonic nebulization inductively coupled plasma mass spectrometry after cloud point extraction. Spectrochimica Acta, Part B: Atomic Spectroscopy, 55(7), 803–813.
Sitko, R. (2015). Green approach for ultratrace determination of divalent metal ions and arsenic species using total-reflection X-ray fluorescence spectrometry and mercapto-modified graphene oxide nanosheets as a novel adsorbent. Analytical Chemistry, 87(6), 3535–3542.
Su, R., Ruan, G., Chen, Z., Du, F., & Li, J. (2015). Application of mercapto-silica polymerized high internal phase emulsions for the solid-phase extraction and preconcentration of trace lead(II). Journal of Separation Science, 38(24), 4262–4268.
Svegl, I. G., & Ogorevc, B. (2000). Soil-modified carbon paste electrode: a useful tool in environmental assessment of heavy metal ion binding interactions. Fresenius Journal of Analytical Chemistry, 367(8), 701–706.
Talio, M. C., Luconi, M. O., Masi, A. N., & Fernández, L. P. (2009). Determination of cadmium at ultra-trace levels by CPE-molecular fluorescence combined methodology. Journal of Hazardous Materials, 170(1), 272–277.
Trindade, A. S., Dantas, A. F., Lima, D. C., Ferreira, S. L., & Teixeira, L. S. (2015). Multivariate optimization of ultrasound-assisted extraction for determination of Cu, Fe, Ni and Zn in vegetable oils by high-resolution continuum source atomic absorption spectrometry. Food Chemistry, 185, 145–150.
Tukur, S. A., Yusof, N. A., & Hajian, R. (2015). Gold nanoparticles-modified screen-printed electrode for determination of Pb(II) ion using linear sweep anodic stripping voltammetry. IEEE Sensors Journal, 15(5), 2780–2784.
Wei, Y., Li, B., Wang, X., & Duan, Y. (2014). Magnified fluorescence detection of silver(I) ion in aqueous solutions by using nano-graphite-DNA hybrid and DNase I. Biosensors & Bioelectronics, 58(8), 276–281.
Zanjanchi, M. A., Noei, H., & Moghimi, M. (2006). Rapid determination of aluminum by UV-Vis diffuse reflectance spectroscopy with application of suitable adsorbents. Talanta, 70(5), 933–939.
Zeiner, M., Rezić, T., & Šantek, B. (2010). Monitoring of Cu, Fe, Ni, and Zn in wastewater during treatment in a horizontal rotating tubular bioreactor. Water Environment Research, 82(2), 183–186. doi:10.2175/106143009x442907.
Zhan, S., Xu, H., Zhan, X., Wu, Y., Wang, L., Lv, J., et al. (2015). Determination of silver(I) ion based on the aggregation of gold nanoparticles caused by silver-specific DNA, and its effect on the fluorescence of rhodamine B. Microchimica Acta, 182(7), 1411–1419.
Acknowledgments
This work was supported by the Project of Innovation and Entrepreneurship Training for College Students of China (No. 201410058050), the National Natural Science Foundation of China (No. 21405110), and the National Natural Science Foundation of China (No. 21476172).
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Guo, Y., Liu, X., Han, Y. et al. Effective Enrichment and Simultaneous Quantitative Analysis of Trace Heavy Metal Ions Mixture in Aqueous Samples by the Combination of Radial Electric Focusing Solid Phase Extraction, UV-Vis Spectrophotometric Determination and Partial Least Squares Regression. Water Air Soil Pollut 228, 317 (2017). https://doi.org/10.1007/s11270-017-3502-6
Received:
Accepted:
Published:
DOI: https://doi.org/10.1007/s11270-017-3502-6