1 Introduction

Eddy current testing (ECT) is one of the most widely used non-destructive testing (NDT) method for flaw detection and characterisation, thickness measurement, and estimating internal structural changes in electrically conducting materials and components [1]. The sensitivity of the eddy current (EC) coil is largely influenced by the coil parameters namely, number of turns, spacing between the windings, dimensional aspect ratio of the coil and the internal ferrite core length [2]. Due to the limited penetration of the EC governed by the skin depth (\( \delta_{s} \)) effect, ECT is commonly used for inspecting defects and corrosion in few mm thick metallic structures and components [3,4,5]. Besides defect detection, ECT is proposed for bulk material inspection such as evaluating the equi bi-axial compressive stress in non-ferromagnetic materials [6], evaluating the extent of carburization in reformer tubes [7], and assessing the weld quality of spot resistance welding technique [8].

In electrically thick conducting specimens, the induced EC and the magnetic field variation in the specimens are primarily confined to the material surface and near sub surface regions. Thus, increasing the penetration depth by lowering the excitation frequency lowers the signal to noise ratio as the voltage induced in the sensing coil is proportional to the rate of change of the magnetic field. An alternate to the traditional single or multi frequency excitation coils that are operated either in absolute or differential mode is the pulsed eddy current (PEC) testing. The broad frequency spectrum allows the detection of surface and near surface defects in PEC testing [9, 10]. Due to the skin depth limitation of the pickup coil [11, 12], detection of deep sub surface defects in electrically thick specimens remains a challenge. Thus, there is an increased interest in using sensitive magnetic field sensors such as Giant Magneto Resistor (GMR) [13, 14], Hall effect sensor [15] and Superconducting Quantum Interference Device (SQUID) sensor [16] for deep sub surface defects.

An electrically thick conducting specimen of thickness, \( t \) satisfies the criterion, \( t/\delta_{s} > 1 \). Thus, the specimen thickness is several times the standard penetration depth. As the frequency dependent skin depth, \( \delta_{s} \) obtained for plane wave excitation of an infinitely extended plate is greater than the probe dependent skin depth, \( \delta_{t} \) [17], the criterion, \( t/\delta_{t} > 1 \) also defines an electrically thick conducting specimen. In electrically thick conducting specimens, the sensitivity of the EC probe for sub surface and deep sub surface defects vary with the defect location. Thus, it is important to understand the role of probe dependent penetration depth for defect detection in electrically thick conducting specimens. In this paper, we present 3D numerical analysis of ECT to understand the scope and limitations of \( \delta_{s} \) and \( \delta_{t} \) for sub surface and deep sub surface defect detection using a pancake coil in electrically thick conducting specimen. Thus, simulations are reported for centrally located defect in a large plate and the influence of lift-off variation on probe impedance is not covered. The paper is organised as follows: Sect. 2 presents the 3D numerical model and study of the influence of probe size on optimal excitation frequency for varying plate dimensions, and the influence of defect location on the optimal test frequency for detection. Section 3 presents the experimental verification of the numerical analysis on an electrically thick aluminium plate with machined defects of varying depths. Section 4 concludes this work.

2 3D Numerical Model

A 3D model of a pancake coil above an electrically thick conducting specimen was developed in COMSOL® Multiphysics, a Finite Element Method (FEM) based simulation software to calculate the coil impedance for varying excitation frequency, plate dimensions and defect depth. Figure 1a shows the zoomed in 2D cross section of the 3D computational domain implemented in COMSOL. A pancake coil of height, \( h \) with inner and outer radii, \( r_{1} \) and \( r_{2} \) respectively placed at a distance, \( d \) above a planar electrically thick conducting specimen of dimensions, \( l \times w \times t\) was modelled. The numerical model solves the governing equation,

Fig. 1
figure 1

Cross sectional view of the numerical model

$$ \nabla^{2} \vec{A} = - \;\mu_{0} \vec{J}_{s} + \mu_{0} \sigma \left( {\frac{{\partial \vec{A}}}{\partial t}} \right), $$
(1)

where \( \vec{A} \) is the magnetic vector potential, \( \vec{J}_{s} \) is the source current density maintained by the coil at the excitation frequency, \( \mu_{0} \) and \( \sigma \) are the magnetic permeability and bulk electrical conductivity of the material respectively.

The computational domain was enclosed by a spherical air domain with magnetic insulation boundary condition, \( \hat{n} \times \vec{A} = 0, \) on the external boundary surface. Coil excitation was specified at a single frequency using surface source current density, \( \vec{J}_{s} \) given by, \( \left( {NI/A_{c} } \right)\vec{e}_{coil} \), where \( I \) is the current flowing through the coil, \( A_{c} \) is the wire cross sectional area, \( N \) is the number of turns, and \( \vec{e}_{coil} \) is the current direction in the coil. The computation domain was discretised into tetrahedral elements and numerical solution was obtained using Flexible Generalised Minimal Residual (FGMRES) iterative solver. The best practices for solving 2D/3D EC problems in COMOSL were implemented in the numerical simulations for model convergence [18]. With the knowledge of the magnetic vector potential (\( \vec{A} \)), the electric and magnetic fields were computed everywhere in the computational domain. The impressed current, \( I(\omega ) \) and voltage induced in the coil, \( V_{ind} (\omega ) \) calculated from the electric field were used to calculate the coil impedance,

$$ Z\left( \omega \right) = \frac{{I\left( \omega \right)R_{DC} - V_{ind} (\omega )}}{I\left( \omega \right)}, $$
(2)

where \( R_{DC} \) is the DC coil resistance and \( \omega = 2\pi f \) is the angular frequency. Simulation results of the EC numerical model implemented in COMSOL were validated with benchmark problems with good agreement between model and measurements [18].

3 Numerical Analysis

Numerical simulations were carried out for a defect free 10 mm thick (\( t \)) aluminium plate with electrical conductivity, \( \sigma \) = 30 MS/m, magnetic permeability, \( \mu_{0} \), plate dimensions, 100 mm × 100 mm × 10 mm (\( l \) = 100 mm, \( w \) = 100 mm), coil parameters, \( r_{1} \) = 1.5 mm, \( h \) = 3 mm and \( N \) = 530, and 0.5 mm lift off \( (d) \). A sinusoidal coil current of 20 mA (\( I \)) was fed to the coil with wire gauge of 38.

3.1 Influence of Coil Size on Test Frequency

The coil outer radius was swept from 6 to 9 mm in steps of 0.5 mm. The sweep range for \( r_{2} \) was selected such that the coil diameter (\( D_{2} \) = 2 \( r_{2} \)) is less than the sample lateral dimensions (\( l \times w \)) to avoid plate edge effects in the simulations. Figure 2 shows the zoomed in surface mesh in the coil and sample domains. A non-uniform mesh with higher mesh density in the coil domain and plate volume beneath the coil was selected to capture the fast decaying induced EC and the secondary magnetic field with reduced computational overheads. The skin depth at 100 Hz is approximately equal to the plate thickness i.e., \( t \approx \delta_{s} . \) Thus, coil excitation frequency was swept from 100 to 1500 Hz so that \( t/\delta_{s} \) varied over 1–3 satisfying the criterion for an electrically thick conducting specimen. Numerical simulations were carried out for varying \( r_{2} \) and \( f \), and the induced current density, \( \vec{J}_{e} \) calculated at the plate bottom defined at (\( x = \) 0, \( y = r_{2} \), \( z = - \;t/2 \)) was used to study the influence of coil size on the excitation frequency for ECT.

Fig. 2
figure 2

Surface plot of the finite element mesh in coil and specimen domains (zoomed view)

Figure 3a shows the norm of the eddy current density, \( \left| {\vec{J}_{e} } \right| \) induced at the plate bottom for 12 mm and 18 mm coil outer diameter (OD) over 100–1500 Hz i.e., \( f \) corresponding to 1–3 \( \delta_{s} \). It can be observed that the maximum eddy current density (\( \left| {\vec{J}_{e} } \right| \)) is induced at the plate bottom for larger coil OD (18 mm), and the coil excitation frequency corresponding to the maximum eddy current density (\( \left| {\vec{J}_{e} } \right| \)) decreases with increase in coil OD, \( \varvec{D}_{2} \). The optimal test frequency, \( f_{opt} \) corresponding to the maximum \( \left| {\vec{J}_{e} } \right| \) is 450 Hz and 550 Hz for 18 mm and 12 mm coil ODs respectively. Figure 3a clearly indicates that there exists a probe dependent optimal test frequency \( (f_{opt} ) \) below which even though the skin depth increases, the induced EC at the plate bottom weakens. Figure 3b shows the induced current density in logarithmic scale at 350 Hz corresponding to \( t/\delta_{s} \) of 2 along the specimen depth indicated by cutline AB defined at \( (x = 0 \), \( y = r_{2} \)). In Fig. 3b, the true penetration depth, \( \delta_{t} \) calculated at 350 Hz for \( \varvec{D}_{2} \) of 18 mm and 12 mm is 3.5 mm and 2.8 mm respectively. This value is far less than the standard skin depth, \( \delta_{s} \) of 4.9 mm at 350 Hz. It can also be observed from Fig. 3b that at \( \varvec{z} = \delta_{s} \), \( \left| {\vec{J}_{e} } \right| \) is far less than 37%. As the EC current density is more for the 18 mm coil OD than 12 mm coil OD, pancake coil with \( r_{1} \) = 1.5 mm, \( r_{2} \) = 9 mm, \( h \) = 3 mm, and \( N \) = 530 turns was selected to study the influence of plate thickness, \( t \) and width, \( w \) on \( f_{opt} \) and \( \delta_{t} \).

Fig. 3
figure 3

a Eddy current density at plate bottom defined at \( (x = 0, y = r_{2} ,z = - \;t/2) \) for 18 mm and 12 mm coil ODs; b normalized eddy current density in logarithmic scale for 18 mm and 12 mm coil ODs at 350 Hz (\( t = \) 2 \( \delta_{s} \)) along the specimen depth indicated by cutline AB shown in the figure inset

3.2 Influence of Plate Dimensions on Excitation Frequency

Numerical simulations of defect free aluminium specimen were carried out for 10 mm thick plate of 100 mm (\( l \)) length and varying plate width (\( w \)) to study the influence of plate edges on the test frequency. Simulations were carried out for a pancake EC probe of 18 mm coil OD and 0.5 mm lift off \( (d) \). Figure 4 shows the normalised eddy current density calculated at the plate bottom for varying plate widths (\( w: \) 30–100 mm). For \( w \) > 3 \( D_{2} \), the optimal test frequency, \( f_{opt} \) approached 450 Hz which corresponds to \( f_{opt} \) for a large specimen without plate edge effect i.e., (\( l \times w \)) of 100 mm × 100 mm. From Fig. 4 it can be concluded that the plate width does not significantly influence the optimal test frequency, \( f_{opt} \) for a finite size plate.

Fig. 4
figure 4

Influence of plate width on coil excitation frequency (\( l \) = 100 mm). Normalized current density at the bottom of a 10 mm thick plate

Simulations were also carried out for varying plate thickness (\( t \)) of a wide specimen (\( l \) = 100 mm, \( w \) = 100 mm) excited by a pancake coil of 18 mm coil OD. Figure 5a shows the normalised EC density induced at the plate bottom for varying plate thickness. In Fig. 5a, as plate thickness increased, the optimum frequency (\( f_{opt} \)) corresponding to the maximum \( \left| {\vec{J}_{e} } \right| \) at the plate bottom decreased monotonically. The induced EC strength at the plate bottom also decreased with increase in plate thickness, \( t \) as shown in Fig. 5b. It should be noted that the plate thicknesses in Fig. 5 satisfy the criterion, \( t > \delta_{s} \), where \( \delta_{s} \) is computed at the \( f_{opt} \) determined from FEM simulations for the respective thickness, \( t \). From Fig. 5b, it can be observed that plate thickness, \( t \propto 1/\sqrt {f_{opt} } \) for an electrically thick conducting plate. The simulation results are in agreement with the empirical relationship reported by Nagendran et al. for deep sub surface defects [16] Furthermore, the simulation results in Figs. 4 and 5 indicate that for a given coil design, plate thickness, \( t \) significantly influences the optimal test frequency, \( f_{opt} \) than the plate lateral dimensions (\( l \times w \)).

Fig. 5
figure 5

Influence of plate thickness on coil excitation frequency (\( l \) = 100 mm, \( w \) = 100 mm). a Normalized current density at plate bottom, b dependence of \( f_{opt} \) and surface current density on plate thickness, \( t \)

3.3 Influence of Plate Thickness on Skin Depth

Figure 6 shows the normalized true (\( \delta_{t} /t \)) and standard (\( \delta_{s} /t) \) skin depths calculated at \( f_{opt} \) for varying plate thickness, \( t:[3, 10] \) mm with lateral plate dimensions, \( l \) = 100 mm and \( w \) = 100 mm. In Fig. 6, the material dependent standard penetration depth, \( \delta_{s} \) is computed at the \( f_{opt} \) determined from FEM simulations for the respective thickness, \( t \). From Fig. 6, it can be noted that the graph can be divided into two regions: region I, where standard skin depth is more than the true skin depth (\( \delta_{t} < \delta_{s} \)), and region II, where \( \delta_{t} > \delta_{s} \). In Fig. 6, the standard skin depth at \( f_{opt} \) computed based on plane wave theory, and true skin depth numerically calculated based on the coil design at \( f_{opt} \) are equal for \( t \) = 4.5 mm (\( t_{p} \)). For \( t \) < \( t_{p} \) (region II), the induced EC density at \( f_{opt} \) shown in Fig. 5b attenuates slower than the exponential decay rate for a plane wave source. Thus, \( \delta_{t} > \delta_{s} \) in region II. As plate thickness increased beyond \( t_{p} \) (region I), the EC density induced by the pancake coil at \( f_{opt} \) decayed faster than the plane wave source. Thus, deep sub surface defect detection in electrically thick plates (\( t/\delta_{t} > 1 \)) must be based on the probe dependent penetration depth, \( \delta_{t} \) [19], and the scope of \( \delta_{s} \) for defect detection must be limited to thin plates which lie in region II. In Fig. 6, \( \delta_{t} /t \) varies over 0.876–0.343 for plate thickness, \( t:[ \) 3, 10] mm. The ratio, \( \delta_{t} /t \) of 1/3 corresponds to 5% of \( |\overrightarrow {{J_{e} }} | \) at the plate bottom \( (z = - \;t/2) \) which is relatively weak to induce a measurable change in the coil impedance. Thus, sensitivity of the EC coil for deep sub surface defect detection in electrically thick plate (\( t/\delta_{t} > 1 \)) decreases with increase in \( t \). As a result, defect detection using an absolute coil can only be improved by increasing either the coil diameter, \( D_{2} \) or excitation current, \( I \). Increasing \( D_{2} \) increases \( \left| {\vec{J}_{e}^{ max} } \right| \) at the plate bottom, lowers \( f_{opt} \) and reduces the spatial resolution of EC coil for defect localization. Increasing coil current, \( I \) leads to coil heating which increases the ohmic loss and causes measurement instability. To avoid these issues, deep sub surface defect detection is typically carried out using either a sensitive secondary coil [20] or GMR as the pick-up sensor [21].

Fig. 6
figure 6

Normalized skin depths, \( \delta_{s} /t \) and \( \delta_{t} \)/t at the optimal test frequency (\( f_{opt} \)) in electrically thick conducting plate. Plate length and width are 100 mm each

3.4 Influence of Defect Depth on Coil Impedance

Numerical simulations were carried out to study the influence of defect depth, \( t_{df} \) on coil impedance over 100–2000 Hz covering \( f_{opt} \). Two dimensional notch of length, 100 mm, breadth, 2 mm and height, \( t_{d} \) was introduced at the bottom of a 100 \( \times \) 100 \( \times \) 10 mm3 plate. Figure 7 shows the change in coil impedance, \( \Delta Z = \,\Delta R \) + \( j\Delta X \) for varying defect depth in the 10 mm thick plate, where \( \Delta R = R_{D} - R_{ND} \), \( \Delta X = X_{D} - X_{ND} \), \( R_{D} \) and \( R_{ND} \) are coil resistances calculated in the presence and absence of the notch respectively, and \( X_{D} \) and \( X_{ND} \) are coil reactances calculated in the presence and absence of the notch respectively. The resistance and inductance values of the simulated coil are 37 Ω and 2.4 mH respectively. The coil location above the specimen with 2D notch is illustrated in Fig. 7. From Fig. 7, it can be observed that the coil impedance, \( Z \) at \( f_{opt} \) varies with notch height, and the phase change rotates to the 1st quadrant as the notch moves closer to the coil. Table 1 lists the change in coil impedance, \( \Delta Z \) for varying notch height, \( t_{d} \) in an electrically thick plate (\( t/\delta_{t} \) = 2.9, \( \delta_{t} \) = 3.4 mm, \( f_{opt} \) = 450 Hz). From Fig. 7 and Table 1, it can be concluded that the change in coil impedance at \( f_{opt} \) can be used to detect and also estimate the location of the defect.

Fig. 7
figure 7

Simulated change in coil impedance, \( \Delta Z = \Delta R \) + \( j\Delta X \) for varying defect height, \( t_{d} \) over 100–2000 Hz. Inset: Coil location simulated for impedance calculation

Table 1 Coil impedance, \( \Delta Z \) at \( f_{opt} \) for a 2 mm wide 2D notch at the bottom of an electrically thick aluminium plate (\( t/\delta_{t} \) = 2.9, \( \delta_{t} \) = 3.4 mm, \( f_{opt} \) = 450 Hz) for varying notch height, \( t_{d} \). Change in the impedance magnitude \( (|\Delta Z|) \), and phase \( \angle ({\Delta }Z) \) were calculated with reference to the coil impedance above a defect free plate

4 Experimental Validation

4.1 Test Specimens and Measurement Set Up

Two aluminium plates of 10 mm thickness were used as test specimens. Specimen A is a calibration block of dimension 100 \( \times \) 30 \( \times \) 10 mm3 made of 2024-T3 aluminium alloy with a centrally located 2D notch and specimen B is a rectangular aluminium plate of dimension 400 \( \times \) 100 \( \times \) 10 mm3 containing three equally spaced 2D notches located 120 mm apart. The dimensions of the notches fabricated using Electro Discharge Machining (EDM) are summarized in Table 2. A pancake coil with \( r_{1} \) = 1.5 mm, \( r_{2} \) = 9 mm and \( h \) = 3 mm was fabricated using a 38 gauge copper wire and characterised using a LCR meter (E4980A, Agilent Technologies USA). The coil inductance and resistance values were measured as 2.72 mH and 25 Ω respectively. Figure 8 shows the experimental set up for ECT.

Table 2 Dimenions of the 2D EDM notches in aluminium specimens
Fig. 8
figure 8

ECT experimental set up

The impedance of the EC probe was measured using a LCR meter as the probe was scanned in steps of 1 mm with the help of a XY scanner controlled using LabVIEW® programming interface. The coil impedance measured by the LCR meter was logged at excitation frequencies in the neighbourhood of the optimal test frequency, \( f_{opt} \) calculated in Sect. 3.1. The coil lift-off and current was maintained as 0.5 mm and 20 mA respectively. The probe impedance above the defect free region, \( Z_{ND} \) was taken as the reference to calculate the change in coil impedance, \( \Delta Z_{n} = \Delta R_{n} + j\Delta X_{n} , \) at the nth scan position, where \( \Delta R_{n} = R_{n} - R_{ND} \), \( \Delta X_{n} = X_{n} - X_{ND} \), and \( R_{n} \) and \( X_{n} \) are the coil resistance and reactance at the nth scan position respectively.

4.2 Measurements and Data Analysis

Figure 9 shows the change in the coil impedance as the probe was scanned over specimen A with a deep sub surface notch (N1). Measurements indicate maximum impedance change at 450 Hz which is close to the optimal test frequency, \( f_{opt} \) (470 Hz) based on the true penetration depth, \( \delta_{t} \) for the 30 mm wide specimen (Fig. 4a). The change in coil impedance for notch N1 at 450 Hz was measured as 1.60 mΩ \( \angle \) − 48.07 \( ^\circ \) with phase change close to − 45°. At frequencies below \( f_{opt} \), though the test frequency satisfy the criteria, \( t/\delta_{s} \) < 3, the coil response to deep sub surface defect at 150 Hz and 250 Hz is poor and noisy than at \( f_{opt} \). At frequencies above \( f_{opt} \), EC strength decreases due to decrease in probe penetration depth.

Fig. 9
figure 9

Change in coil impedance for the deep sub surface notch, N1 (\( t_{df} \) = 8.5 mm, \( t \) = 10 mm) specimen A

Figure 10 shows the change in coil impedance for specimen B as the EC probe was moved above the deep sub surface notches, N2 and N3 with defect location, \( t_{df} \) > \( t \)/2. The change in coil impedance at \( f_{opt} \) was measured as 5 mΩ \( \angle \) − 38.81° and 7.93 mΩ \( \angle \) − 34° for N2 and N3 notches respectively. As the defect depth from the plate surface (\( t_{df} \)) decreased, the phase change at \( f_{opt} \) moved away from − 45° as observed in the simulations (Fig. 7, Table 1). Furthermore, due to the close proximity of the sub surface defect, the change in coil impedance is more for N3 than N2.

Fig. 10
figure 10

Change in coil impedance for deep sub surface 2D notches. a \( t_{df} \) = 8 mm (N2), and b \( t_{df} \) = 7 mm (N3) in specimen B

Figure 11 shows the change in coil impedance for the near sub surface notch, N4 located 4 mm from the plate surface i.e., \( t_{df } \) < \( t \)/2. Figure 11 shows the measurements at the test frequencies used for the deep sub surface notches, N2 and N3 shown in Fig. 10. The rotation of the impedance plane trajectory in Fig. 11 confirms that the impedance change is due to the defect. The phase change at \( f_{opt} \)(450 Hz) is \( + \) 19° (1st quadrant) indicating that the defect is a near surface defect. Thus, the experimental observation is in agreement with the simulations (Fig. 7, Table 1).

Fig. 11
figure 11

Change in coil impedance for the sub surface 2D notch, N4 in specimen B. Impedance plane trajectory of the EC probe in the neighbourhood of FEM based \( f_{opt} \) of 450 Hz

5 Conclusion

Unlike a plane wave source, the true penetration depth of an EC test coil depends on the material property as well as the coil dimensions. Furthermore, the magnetic field induced by a finite size coil approaches the field induced by a plane wave source as the coil becomes infinitely large. Thus, a larger test coil induced more EC in the specimen. As a result, the true penetration achieved by a finite size coil, \( \delta_{t} \) is lower than the standard penetration depth, \( \delta_{s} . \) 3D FEM simulations presented here clearly indicate the importance of the true (\( \delta_{t} \)) penetration depth of a test coil for sub surface defect detection in electrically thick conducting specimens using ECT. The optimum test frequency, \( f_{opt} \) for deep sub surface defect detection in electrically thick specimens is given by \( \delta_{t} \) which is primarily determined by the outer diameter of the EC probe and not the material dependent standard penetration depth, \( \delta_{s} \). Thus, the scope of standard penetration depth, \( \delta_{s} \) for determining the test frequency is applicable for inspection of thin plates (\( t/\delta_{s} \) < 1). As the defect location from the plate surface decreases, the change in the phase of the complex coil impedance, \( \angle\Delta Z \) at \( f_{opt} \) could be used to estimate the defect location. The numerical findings were experimentally verified on electrically thick aluminium plates for the EC probe dimensions studied in the FEM simulations. It is shown that the plate thickness, \( t \propto 1/\sqrt {f_{opt} } \), and defect depth (\( t_{df} ) \) can be estimated from \( \angle\Delta Z \) measured at \( f_{opt} \). The close agreement between simulations and measurements indicate that the proposed approach could be used for selecting EC probe and the optimal test frequency for defect detection in electrically thick conducting specimens.