Abstract
Background
In general surgery, minimally invasive laparoscopic procedures have been steadily increasing over the last decade. The application of advanced bipolar and ultrasonic energy devices for sealing and cutting of blood vessels plays a vital role in routine clinical procedures. The advantages of energy-based instruments are enhanced sealing capability combined with both fast sealing time and minimal thermal injury. The purpose of this study was to compare the safety and efficacy profiles of nine laparoscopic sealing and cutting devices in a porcine model, with a new scoring system.
Methods
Comparative studies in a porcine model were performed to assess vessel sealing, burst pressure, thermal spread, maximum heat, sealing/cooling time, and compression strength over the full jaw. Nine different devices from five manufacturers were tested in this study. The sealing and cutting devices (SCD) score has been developed to enable standardized comparisons of various devices. For this purpose, the most important parameters were identified through a consensus approach.
Results
All sealed vessels with different devices could withstand a median pressure of more than 300 mmHg (range 112–2046 mmHg). The time for the sealing procedure was 7.705 s (range 5.305–18.38 s) for the ultrasonic and 7.860 s (range 5.08–10.17 s) for the bipolar devices. The ultrasonic instruments reached a median temperature of 218.1 °C (range 81.3–349.75 °C) and the bipolar devices a temperature of 125.5 °C (range 94.1–133.35 °C). The tissue reached a median temperature of 61.9 (range 47.1–80.6 °C) after ultrasonic sealing and 76.7 °C (range 63.1–94.2 °C) after bipolar sealing. The median SCD score was 10.47 (range 7.16–13.72).
Conclusion
All the instruments used seemed safe for use on the patient. The SCD score allows an indirect comparability of the instruments.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
In gynecological surgery, minimally invasive laparoscopic procedures have been steadily increasing over the last decade [1, 2]. The application of advanced bipolar and ultrasonic energy devices for sealing and cutting of blood vessels plays a vital role in routine clinical procedures [3, 4]. The challenge lies in combining multiple functions in the same instrument for coagulation and dissection without worsening the handling. The underlying mechanism is to achieve thermal hemostasis by protein denaturation, which occurs at a temperature of 45 °C onward [5]. The sealing with bipolar instruments can withstand high intraluminal pressures and is an alternative to the classical vessel occlusions such as ligature or clips [6,7,8,9,10,11,12]. The sealing takes place within the vessel wall structure and cannot dislodge [8, 13]. The closure is not always optimal and can be adversely affected for many different reasons [2, 11, 14]. Advantages of state-of-the-art energy-based instruments are enhanced sealing capability combined with both fast sealing time and minimal thermal injury [15]. The increasing popularity of these instruments has led to a wide arsenal of devices with different energy bases. Despite the safety- and efficacy-related aspects, the likely disadvantages are mainly lateral thermal spread, relatively high disposal cost, and variable burst pressures [16]. Excessive energy delivery to the tissue may result in larger necrosis and thus a higher risk of secondary bleeding and more frequent damage to sensitive structures such as the intestines or ureters [2, 17,18,19]. Previous studies identified homogeneous compression on the sealing area, controlled energy on the vessel, and a strong vascular sealing as important parameters to avoid collateral damages and comorbidity such as necrosis-associated adhesions [20,21,22,23]. Based on the studies of Wallwiener et al. and Soderstrom et al., we generated an easily reproducible porcine model to assess the effects of sealing and cutting devices [24, 25].
The purpose of this study was to compare the safety and efficacy profiles of nine laparoscopic sealing and cutting devices in a porcine model with the new sealing and cutting devices (SCD) score, which was developed to enable standardized comparisons of various devices. Instruments were compared in regard to thermal spread, burst and compression pressure, and sealing and cooling time.
Materials and methods
Comparative studies were performed to assess vessel sealing and burst pressure (BP), thermal spread (TS), maximum heat, sealing/cooling time (ST/CT), and compression strength over the full jaw. Nine different devices from five manufacturers were tested in this study. During these in vitro experimental procedures, some instruments were either updated (Ethicon Harmonic ACE was updated to Harmonic ACE + 7®) or withdrawn (Aesculap Caiman® was withdrawn by the manufacturer from the study) and were thus not taken into account in the further evaluation. The energy-based devices used in this study are shown in Table 1.
The tissues for TS and BP experiments were taken from German domestic pigs weighing 40–60 kg. Immediately after the slaughter of the pigs in a slaughterhouse, the carotid arteries and small intestine were removed (OP organization and animal welfare, the competence center “MEDIZIN IM GRUNEN” in Wendisch Rietz, Germany). After skeletonizing the vessels and bowels, both were placed in normal saline (0.9%) and frozen at −20 °C. Before the experiments were started, the tissue was thawed to 4 °C and then warmed to body temperature (37 °C) in heat baths. In a previous study, it was demonstrated that freshly thawed frozen vessels reacted similar to fresh vessels [26].
Sealing and burst pressure
Burst pressure measurements are important endpoints and evaluation parameters, because they reflect the sealing quality of the devices. The normal blood pressure is 140/90 mmHg and reaches values >160/100 mmHg in hypertensive patients. In male athletes, the mean blood pressure reaches 311/284 mmHg during double-leg press sets at 85 and 100% of maximum with closed glottis Valsalva. The highest measured blood pressure in an individual during these tests was 370/360 mmHg [27]. During and after the exercises, values of more than 300 mmHg are not reached; since a safety distance must be maintained to the limits actually attainable, in this work the limit was set at 250 mmHg.
The sealing and BP experiments were conducted on the carotid arteries (diameter: 4–7 mm). The vessel diameter was measured with a digital caliper with a filled vessel and a pressure of 100 mmHg. The burst pressure measurements were carried out with a measuring device from Medimotec, Germany, which was provided by Ethicon, Germany. The experiments were controlled and recorded by a computer program. At the beginning of each day, the measuring instrument had to be rinsed with a 70% solution of 2-propanol to dissolve any possible gluing in the lines and check the continuity. The system was then rinsed with a 0.9% sodium chloride solution so that no propanol could subsequently come into contact with the vessels and eventually damage them. The following steps had to be done with each measurement: (1) Connecting vessel, (2) Vessel cleaning, (3) Leakage test, (4) Sealing and cutting the vessel, (5) Placing cover, (6) Cleaning the buffer, and (7) Activation. The bipolar instruments gave a feedback to the point of completion of the seal. For the ultrasonic instruments, the end of the seal was equated with the time of the vessel dissection. After each test, the sealed zone was removed with a surgical scissor while maintaining sufficient safety margins and the vessel was reused. In case of a defect on the vessel or insufficient length, it was replaced by a new carotid artery.
At least six measurements per device were carried out; in case of obvious failures such as leakage or slipping of the vessel, the experiment was repeated.
Maximum heat and sealing and cooling time
For the measurement of maximum heat and cooling time, the “testo 885” camera (Serial number: 02323152) from Testo (Testo SE & Co. KGaA, Lenzkirch, Germany) was used. The intestine was cleaned of residual fecal matter with physiological sodium chloride solution at 37 °C. In order to minimize unwanted reflections and thus inaccurate measurements, all reflecting areas were covered. All measurements were carried out uniformly with a camera distance of 28 cm. The intestine was fixed to avoid tension during the sealing and cutting procedure. The recording was stopped as soon as the tissue cooled down to 38 °C. The tissue temperature was measured directly on the dissection edge after removal of the device. The devices were carefully cleaned after each measurement. The scan area of the camera was in the range of 0–350 °C. To evaluate video recordings, the Testo software IR Soft 3.2 and 3.6 were used. Maximum heat, duration of coagulation, and cooling to 38 °C were measured. We defined 38 °C as nearly normal body temperature without any further damage to the collateral tissue. In order to determine the exact moment of the sealing and the time course of the temperature, the respective video was scanned in approximately 4-hundredths of a second step. The sealing time was defined as the time between the start of sealing until the bipolar instruments gave a feedback or the ultrasonic devices divided the vessels. In devices with dissimilar jaws (ultrasonic devices), which have an active blade and a passive jaw, both sides were separately measured.
Compression strength
For the measurement of compression forces of the sealing and cutting devices, a sensor (Tekscan 5027@500psi™, Tekscan Inc., CMV Hoven GmbH, Germany) was placed between two thin layers of a silicone membrane (each side: 0.8 mm). Closed-jaw pressures were recorded with I-Scan™ software (Tekscan™ Inc., CMV Hoven GmbH, Germany).
No unit was used to measure the compression strength; the results represent the relative clamping pressure values of the instruments (along the closed jaw, from the tip to the base, Fig. 1). The compression pressure of the individual instruments must be seen in relation to the other instruments. Each device was measured with the maximum possible closing pressure, and this corresponded to the clicked close of some instruments. Each device was tested only once.
Necrosis zone
By measuring the necrosis zone, the cell damage can be measured lateral to the branches. Coagulated and dissected vessels were prepared for further histochemical hematoxylin and eosin staining. The necrosis zone was then microscopically measured on the stump of the vessel. For the two non-dissecting instruments BiClamp® fenestrated and BiClamp® Maryland (ERBE Elektromedizin GmbH, Germany), the measured necrosis zone was used to indicate half the distance in order to be able to make a comparison with the other instruments.
SCD score
The sealing and cutting devices (SCD) score was developed to enable standardized comparisons of various devices. For this purpose, the most important parameters ST, BP, and maximum temperature of the jaws of each device (T) were identified through a consensus approach, and the median value scaled. Next, the scaled values (T s, BPs, STs) were added and then divided by the number of parameters. Because BP is the most important parameter for clinical application, its value was doubled.
The respective span width of the parameters was divided by 20, and the values were converted into a score. For BP, a higher score was assigned for a higher value, whereas for ST and T a lower score was assigned for higher values. Devices with a bursting force below 250 mmHg are not recommended for use owing to a poor safety profile. The relevant parameters were scaled as follows:
The range was determined for the possible measured values of the respective measuring instruments. The minimum value of the score was 0, while the maximum value was 20.
Statistical analysis
This is an explorative analysis and hence the analysis of data was of a descriptive nature. Depending on the scale level of the variable and with respect to the small number of measurements, median and range or frequencies were provided. Data were analyzed with the SPSS software (IBM SPSS Statistics 24, Ehningen, Germany).
Results
The vessel caliber and anatomic origin of the vessel or the intestine did not have any influence on the experimental outcomes. The median vessel diameter was 5 mm (range 4–10 mm).
Sealing and burst pressure
A summary of the measured BP is shown in Table 2. All devices were measured until at least six coagulations were evaluable. The main reason for non-evaluable measurements was errors in the measuring device and individual vessel ramifications. We observed that all sealed vessels with different devices could withstand a median BP of more than 300 mmHg (range 112–2046 mmHg).
Maximum tissue temperature and cooling time
The time for the sealing procedure was 7.705 s (range 5.305–18.38 s) for the ultrasonic (including THUNDERBEAT) and 7.86 s (range 5.08–10.17 s) for the bipolar devices. The ultrasonic instruments reached a median temperature of 218.1 °C (range 81.3–349.75 °C) and the bipolar devices a temperature of 125.5 °C (range 94.1–133.35 °C). The tissue reached a median temperature of 61.9 °C (range 47.1–80.6 °C) after ultrasonic sealing and 76.7 °C (range 63.1–94.2 °C) after bipolar sealing. The median time till the tissue cooled down to 38 °C was 14.50 s (range 4.445–31.025 s) for the ultrasonic and 38.27 s (range 30.35–82.15 s) for bipolar sealing. The sealing and cooling down times and the temperature of the tissue and the devices are shown in Table 3.
Compression strength
The median compression area for the ultrasonic devices was 67 mm2 (range 50–63 mm2) and that for the bipolar devices was 98 mm2 (range 77–102 mm2). The compression force decreased with the ultrasound devices by 50% in the median after a centimeter, and by 60% with the bipolar devices (Table 4). Figure 1 demonstrates the relation of the compression strength to distance of the jaws.
Necrosis zone
The necrosis zone ranged from 320.30 µm (Sonicision high) being the lowest to 5024.50 µm (Harmonic ACE + 7® Level 5) being the highest (Table 2). Figure 2 shows the histological sections of Olympus THUNDERBEAT (ultrasonic and bipolar, A), ERBE BiClamp® (bipolar without dissection function, B), Covidien LigaSure™ (bipolar, C), and Ethicon Harmonic ACE + 7® Level 3 (ultrasonic, D).
SCD Score
The median SCD score was 9.04 (range 5.73–12.24). The values calculated for the score are shown in Table 5.
Discussion
The number of sealing and cutting devices for laparoscopic procedures is constantly increasing; however, in vitro models for evaluation of efficacy and safety parameters are scarce and only limited to individual parameters. Moreover, properties have been compared only in small groups. In this study, we compared nine different laparoscopic devices from four different categories. The aim of this work was to compare the safety and efficacy profiles of nine laparoscopic sealing and cutting devices in a porcine model with a new scoring system. Each device was tested for compressive pressure, thermal development and energy release, BP after vessel sealing, and the resulting necrosis zone.
The devices used in recent times undergo continuous development, as surgeons demand these instruments for more difficult and complex procedures [28, 29]. The porcine model is suitable as the calibers of the sealed porcine vessels (carotid arteries [median: 5 mm]) were comparable with the human uterine artery, which ranges from 3 to 5 mm in diameter [30]. The purified porcine small bowels have a homogeneous and evenly thick tissue that is most suitable for experiments on thermal measurement. Hence, it can be concluded that all instruments evaluated in this study safely sealed the vessel, as no median burst values below 300 mmHg were recorded. In accordance to our findings, the BP measurements showed reliable sealing results on explanted porcine arteries. Our results are in the range of reported values in the literature [3, 4, 8, 21, 31,32,33,34,35]. An important clinical aspect is the collagen–elastin ratio in vessels, which predicts the BP of arteries using bipolar sealing [36].
In this work, the thermal behavior of the instruments and the energy output to the tissue were measured. Even if one of the instruments reached temperatures above 350 °C, the temperature of the tissue did not exceed 94 °C. Sealed tissue takes up to 80 s to reattain body temperature. The closing pressures are higher than that on bipolar instruments.
Tissue necrosis due to thermal damage is an important factor for vessel sealing. The detection and evaluation of the thermal damage is reportedly difficult [37, 38].
Not only the maximum temperature but also the duration of heat application is responsible for the extent of permanent damage [39].
In literature, the process and quality of the sealing are described as strongly dependent on the pressure distribution and the actual pressure between the jaws [21, 22]. Another issue worth mentioning is the closure pressure on the instruments.
All instruments showed similar closing pressure and a decrease of this pressure from the base to the tip. To our best knowledge, this study evaluates the largest number of different instruments and measurements. We acknowledge that the significance of this study is presently limited, as no statistical association could be derived owing to the low number of cases. However, all properties were measured with high reproducibility. Additionally, the aim of the study was to demonstrate that all the instruments could reach values that are safe for everyday clinical practice. The choice of the instrument depends on the user and convenience of handling. The lateral thermal spread in the tissue remains a major issue for clinical handling and shall be investigated in future studies that address the impact of the devices’ proximity to sensitive organs and structures. It is important to consider the moisture levels in the tissue during these tests, because water vapor, which is produced during the sealing, spreads laterally and can cause collateral damage.
This parameter could also be inserted into the SCD score. Future studies should aim to perform more measurements during the thermal imaging stage. Using the SCD score described herein, instruments can be easily compared with each other and their safety represented in a comprehensible manner.
Conclusion
To our best knowledge, this is the largest comparison of laparoscopic cutting and sealing devices that showed that all the instruments used are safe for daily clinical routine. The new SCD scoring system allows an indirect comparison of the instruments.
References
Philosophe R (2003) Avoiding complications of laparoscopic surgery. Fertil Steril 80(Suppl 4):30–39 quiz 54-36
Tulikangas PK, Smith T, Falcone T, Boparai N, Walters MD (2001) Gross and histologic characteristics of laparoscopic injuries with four different energy sources. Fertil Steril 75(4):806–810
Seehofer D, Mogl M, Boas-Knoop S, Unger J, Schirmeier A, Chopra S, Eurich D (2012) Safety and efficacy of new integrated bipolar and ultrasonic scissors compared to conventional laparoscopic 5-mm sealing and cutting instruments. Surg Endosc 26(9):2541–2549. doi:10.1007/s00464-012-2229-0
O´Keeffe K, Fuchs K (2013) Laparoscopic-assisted vaginal hysterectomy with bipolar coagulation cutting forceps (ENSEAL(r) trio device) versus suture technique vaginally: a comparative analysis. J Gynecol Surg 29(3):131–134
Martin B (2006) Elektrochirurgie Handbuch. KLS Martin Group, Martin Brothers GmbH & Co KG V1.1, Umkirch
Harold KL, Pollinger H, Matthews BD, Kercher KW, Sing RF, Heniford BT (2003) Comparison of ultrasonic energy, bipolar thermal energy, and vascular clips for the hemostasis of small-, medium-, and large-sized arteries. Surg Endosc 17(8):1228–1230. doi:10.1007/s00464-002-8833-7
Heniford BT, Matthews BD, Sing RF, Backus C, Pratt B, Greene FL (2001) Initial results with an electrothermal bipolar vessel sealer. Surg Endosc 15(8):799–801. doi:10.1007/s004640080025
Kennedy JS, Stranahan PL, Taylor KD, Chandler JG (1998) High-burst-strength, feedback-controlled bipolar vessel sealing. Surg Endosc 12(6):876–878
Levy B, Emery L (2003) Randomized trial of suture versus electrosurgical bipolar vessel sealing in vaginal hysterectomy. Obstet Gynecol 102(1):147–151
Novitsky YW, Rosen MJ, Harrell AG, Sing RF, Kercher KW, Heniford BT (2005) Evaluation of the efficacy of the electrosurgical bipolar vessel sealer (LigaSure) devices in sealing lymphatic vessels. Surg Innov 12(2):155–160. doi:10.1177/155335060501200215
Pietrow PK, Weizer AZ, L’Esperance JO, Auge BK, Silverstein A, Cummings T, Preminger GM, Albala DM (2005) PlasmaKinetic bipolar vessel sealing: burst pressures and thermal spread in an animal model. J Endourol 19(1):107–110. doi:10.1089/end.2005.19.107
Tamussino K, Afschar P, Reuss J, Perschler M, Ralph G, Winter R (2005) Electrosurgical bipolar vessel sealing for radical abdominal hysterectomy. Gynecol Oncol 96(2):320–322. doi:10.1016/j.ygyno.2004.09.021
Katkhouda N, Mavor E, Friedlander MH, Mason RJ, Kiyabu M, Grant SW, Achanta K, Kirkman EL, Narayanan K, Essani R (2001) Use of fibrin sealant for prosthetic mesh fixation in laparoscopic extraperitoneal inguinal hernia repair. Ann Surg 233(1):18–25
Spivak H, Richardson WS, Hunter JG (1998) The use of bipolar cautery, laparosonic coagulating shears, and vascular clips for hemostasis of small and medium-sized vessels. Surg Endosc 12(2):183–185
Lyons SD, Law KS (2013) Laparoscopic vessel sealing technologies. J Minim Invasive Gynecol 20(3):301–307. doi:10.1016/j.jmig.2013.02.012
Weld KJ, Dryer S, Ames CD, Cho K, Hogan C, Lee M, Biswas P, Landman J (2007) Analysis of surgical smoke produced by various energy-based instruments and effect on laparoscopic visibility. J Endourol 21(3):347–351. doi:10.1089/end.2006.9994
El-Banna M, Abdel-Atty M, El-Meteini M, Aly S (2000) Management of laparoscopic-related bowel injuries. Surg Endosc 14(9):779–782
Harrell AG, Kercher KW, Heniford BT (2004) Energy sources in laparoscopy. Semin Laparosc Surg 11(3):201–209
Tulikangas PK, Beesley S, Boparai N, Falcone T (2001) Assessment of laparoscopic injuries by three methods. Fertil Steril 76(4):817–819
Eick S, Loudermilk B, Walberg E, Wente MN (2013) Rationale, bench testing and in vivo evaluation of a novel 5 mm laparoscopic vessel sealing device with homogeneous pressure distribution in long instrument jaws. Ann Surg Innov Res 7(1):15. doi:10.1186/1750-1164-7-15
Wallwiener CW, Rajab TK, Zubke W, Isaacson KB, Enderle M, Schaller D, Wallwiener M (2008) Thermal conduction, compression, and electrical current–an evaluation of major parameters of electrosurgical vessel sealing in a porcine in vitro model. J Minim Invasive Gynecol 15(5):605–610. doi:10.1016/j.jmig.2008.05.003
Reyes DA, Brown SI, Cochrane L, Motta LS, Cuschieri A (2012) Thermal fusion: effects and interactions of temperature, compression, and duration variables. Surg Endosc 26(12):3626–3633. doi:10.1007/s00464-012-2386-1
De Wilde RL, Brolmann H, Koninckx PR, Lundorff P, Lower AM, Wattiez A, Mara M, Wallwiener M (2012) The Anti-Adhesions in Gynecology Expert P (2012) Prevention of adhesions in gynaecological surgery: the 2012 European field guideline. Gynecol Surg 9(4):365–368. doi:10.1007/s10397-012-0764-2
Soderstrom RM, Levy BS, Engel T (1989) Reducing bipolar sterilization failures. Obstet Gynecol 74(1):60–63
Wallwiener CW, Rajab TK, Kramer B, Isaacson KB, Brucker S, Wallwiener M (2010) Quantifying electrosurgery-induced thermal effects and damage to human tissue: an exploratory study with the fallopian tube as a novel in vivo in situ model. J Minim Invasive Gynecol 17(1):70–77. doi:10.1016/j.jmig.2009.09.007
Stemper BD, Yoganandan N, Stineman MR, Gennarelli TA, Baisden JL, Pintar FA (2007) Mechanics of fresh, refrigerated, and frozen arterial tissue. J Surg Res 139(2):236–242. doi:10.1016/j.jss.2006.09.001
Narloch JA, Brandstater ME (1995) Influence of breathing technique on arterial blood pressure during heavy weight lifting. Arch Phys Med Rehabil 76(5):457–462
Uzunoglu FG, Bockhorn M, Fink JA, Reeh M, Vettorazzi E, Gawad KA, Bogoevski D, Vashist YK, Tsui TY, Koenig A, Mann O, Izbicki JR (2013) LigaSure vs. conventional dissection techniques in pancreatic surgery–a prospective randomised single-centre trial. J Gastrointest Surg 17(3):494–500. doi:10.1007/s11605-012-2107-z
Ikeda M, Hasegawa K, Sano K, Imamura H, Beck Y, Sugawara Y, Kokudo N, Makuuchi M (2009) The vessel sealing system (LigaSure) in hepatic resection: a randomized controlled trial. Ann Surg 250(2):199–203. doi:10.1097/SLA.0b013e3181a334f9
Wilson MJ, Lopez M, Vargas M, Julian C, Tellez W, Rodriguez A, Bigham A, Armaza JF, Niermeyer S, Shriver M, Vargas E, Moore LG (2007) Greater uterine artery blood flow during pregnancy in multigenerational (Andean) than shorter-term (European) high-altitude residents. Am J Physiol Regul Integr Comp Physiol 293(3):R1313–R1324. doi:10.1152/ajpregu.00806.2006
Santini M, Vicidomini G, Fiorello A, Laperuta P, Busiello L (2008) Electrothermal bipolar tissue sealing systems in lung surgery. Multimed Man Cardiothorac Surg. doi:10.1510/mmcts.2007.003111 mmcts 2007 003111
Newcomb WL, Hope WW, Schmelzer TM, Heath JJ, Norton HJ, Lincourt AE, Heniford BT, Iannitti DA (2009) Comparison of blood vessel sealing among new electrosurgical and ultrasonic devices. Surg Endosc 23(1):90–96. doi:10.1007/s00464-008-9932-x
Hruby GW, Marruffo FC, Durak E, Collins SM, Pierorazio P, Humphrey PA, Mansukhani MM, Landman J (2007) Evaluation of surgical energy devices for vessel sealing and peripheral energy spread in a porcine model. J Urol 178(6):2689–2693. doi:10.1016/j.juro.2007.07.121
Presthus JB, Brooks PG, Kirchhof N (2003) Vessel sealing using a pulsed bipolar system and open forceps. J Am Assoc Gynecol Laparosc 10(4):528–533
Berdah SV, Hoff C, Poornoroozy PH, Razek P, Van Nieuwenhove Y (2012) Postoperative efficacy and safety of vessel sealing: an experimental study on carotid arteries of the pig. Surg Endosc 26(8):2388–2393. doi:10.1007/s00464-012-2177-8
Sindram D, Martin K, Meadows JP, Prabhu AS, Heath JJ, McKillop IH, Iannitti DA (2011) Collagen-elastin ratio predicts burst pressure of arterial seals created using a bipolar vessel sealing device in a porcine model. Surg Endosc 25(8):2604–2612. doi:10.1007/s00464-011-1606-4
Pearce J, Thomsen S (1995) Rate process analysis of thermal damage. In: Welch S, von Gemert S (eds) Optical thermal response of laser irradiated tissue. Plenum Press, New York, pp 561–606
Ghadially F (1988) Ultrastructural pathology of the cell and matrix, 3rd edn. Butterworths, Boston
Campbell PA, Cresswell AB, Frank TG, Cuschieri A (2003) Real-time thermography during energized vessel sealing and dissection. Surg Endosc 17(10):1640–1645. doi:10.1007/s00464-002-8781-2
Acknowledgements
We thank Covidien, Ethicon, Olympus, Martin, and ERBE for providing the generator and the instruments and Mr. Glier (Ethicon) for helping with and providing the burst pressure measurement device. We would also like to thank Testo for supporting and providing the thermal imaging camera and Tekscan, CMV Hoven GmbH, for the pressure measurement device.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Disclosures
No financial support was obtained. Mrs. Brecht, Prof. Wallwiener, Prof. Sohn, Prof. Schuetz, Prof. Domschke, Prof. Fluhr and Drs. Sarah Schott, Dinkic, Golatta, Stenzinger, Kirchner, and Rom have no conflicts of interest or financial ties to disclose.
Rights and permissions
About this article
Cite this article
Brecht, L., Wallwiener, M., Schott, S. et al. Implementation of a novel efficacy score to compare sealing and cutting devices in a porcine model. Surg Endosc 32, 1002–1011 (2018). https://doi.org/10.1007/s00464-017-5778-4
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00464-017-5778-4