Abstract
Background
This study aims to compare proton density weighted magnetic resonance imaging (MRI) zero echo time (ZTE) and head atlas attenuation correction (AC) to the reference standard computed tomography (CT) based AC for 11C-methionine positron emission tomography (PET)/MRI.
Methods
A retrospective cohort of 14 patients with suspected or confirmed brain tumour and 11C-Methionine PET/MRI was included in the study. For each scan, three AC maps were generated: ZTE–AC, atlas-AC and reference standard CT-AC. Maximum and mean standardised uptake values (SUV) were measured in the hotspot, mirror region and frontal cortex. In postoperative patients (n = 8), SUV values were additionally obtained adjacent to the metal implant and mirror region. Standardised uptake ratios (SUR) hotspot/mirror, hotspot/cortex and metal/mirror were then calculated and analysed with Bland–Altman, Pearson correlation and intraclass correlation reliability in the overall group and subgroups.
Results
ZTE–AC demonstrated narrower SD and 95% CI (Bland–Altman) than atlas-AC in the hotspot analysis for all groups (ZTE overall ≤ 2.84, − 1.41 to 1.70; metal ≤ 1.67, − 3.00 to 2.20; non-metal ≤ 3.04, − 0.96 to 3.38; Atlas overall ≤ 4.56, − 1.05 to 3.83; metal ≤ 3.87, − 3.81 to 4.64; non-metal ≤ 4.90, − 1.68 to 5.86). The mean bias for both ZTE–AC and atlas-AC was ≤ 2.4% compared to CT-AC. In the metal region analysis, ZTE–AC demonstrated a narrower mean bias range—closer to zero—and narrower SD and 95% CI (ZTE 0.21–0.48, ≤ 2.50, − 1.70 to 2.57; Atlas 0.56–1.54, ≤ 4.01, − 1.81 to 4.89). The mean bias for both ZTE–AC and atlas-AC was within 1.6%. A perfect correlation (Pearson correlation) was found for both ZTE–AC and atlas-AC compared to CT-AC in the hotspot and metal analysis (ZTE ρ 1.00, p < 0.0001; atlas ρ 1.00, p < 0.0001). An almost perfect intraclass correlation coefficient for absolute agreement was found between Atlas-, ZTE and CT maps for maxSUR and meanSUR values in all the analyses (ICC > 0.99).
Conclusions
Both ZTE and atlas-AC showed a good performance against CT-AC in patients with brain tumour.
Similar content being viewed by others
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.Background
In neuro-oncology, positron emission tomography/magnetic resonance imaging (PET/MRI) combines Positron Emission Tomography (PET) functional metabolic information with the magnetic resonance imaging (MRI) morphological appearance, allowing simultaneous data acquisition with the potential to overcome the intrinsic limitations of MRI [1]. Among the existing radiopharmaceuticals used in PET imaging, 11C-methionine (MET) is considered one of the most suitable amino acid tracers for brain tumours [2]. The uptake of MET by transmembrane transport via sodium-independent L-transporters reflects the concentration gradient and cellular proliferation related to protein synthesis within the tumour. The distribution of 11C-methionine has potential to characterise primary brain tumour/metastases, assess the efficacy of oncological treatment and differentiate radionecrosis from tumour recurrence [3,4,5]. Due to these characteristics, 11C-methionine provides a higher selectivity for the identification of brain gliomas, especially for hypo- or isometabolic lesions on 18F-Flurodeoxyglucose (FDG) PET [6, 7] and a diagnostic accuracy similar to 18F-fluoroethyl-l-tyrosine (FET) [8]. For accurate quantification of the radioactivity concentration measured by PET, correction for photon attenuation is needed. Conversion of the MRI signal into attenuation values is challenging, hampering accurate AC in PET/MRI [9]. Several methods aiming to improve vendor specific MR-AC have been described [10,11,12]. Nonetheless, head CT scan is still needed for accurate AC, leaving PET/MRI dependent upon a separate CT and additional radiation exposure [13,14,15]. In comparison to PET/CT, PET/MRI lacks a clinically accepted standard method to directly obtain AC maps.
The unsolved AC issues in PET/MRI relate to the fact that proton density in MRI does not directly correlate to radiodensity [16, 17]. PET/MRI AC is especially challenging close to bony structures [18], air and metal implants [19]. Metal implants are encountered in postoperative brain tumour patients, introducing risk for errors in the AC near the postoperative site. Several attempts to improve MR-AC close to metal in PET/MRI has been made, for example using time-of-flight (TOF) [20, 21].
A recently developed template-based method, Zero Echo Time (ZTE), for AC in PET/MRI has been developed and tested in non-surgical patients, providing an accurate AC map when compared to attenuation correction with PET emission scan [10, 18, 22, 23]. The aim of this current study is to further evaluate ZTE–AC in pre- and postoperative patients with suspected or confirmed brain tumour for correct SUV quantification in the presence of surgical metal implants and clinical use of PET/MRI study of brain tumours. Correct AC in PET/MRI will improve diagnostic accuracy and guide treatment-based decisions in clinical practice.
To our knowledge, this is the first study evaluating ZTE–AC and atlas-AC compared to the reference standard CT-AC for brain MET PET/MRI.
Methods
The aim of the study is to retrospectively investigate ZTE–AC and atlas-AC compared to the reference standard CT-AC for MET PET/MRI in pre- and postoperative patients with suspected or confirmed brain tumour. Maximum and mean standardised uptake values (SUV) were measured in the hotspot, mirror region and frontal cortex. In postoperative patients (n = 8), SUV values were additionally obtained adjacent to the metal implant and mirror region.
Subjects
We evaluated 18 lesions from 14 consecutive patients (6 male, 8 female, median age 43 years, range 29–73) who had acquired MET PET/CT (not analysed further in this article) and 3 T PET/MRI (GE Healthcare, Waukesha, WI) at Karolinska University Hospital from January to June 2019. Inclusion criteria were 11C-MET PET/MRI acquisition, availability of atlas-AC map, proton density weighted ZTE–AC map and CT-based AC map as reference standard and suspected or confirmed brain tumour. The study was approved by the Swedish Ethical Review Authority (2019–01309), through which informed consent was waived.
Data acquisition and image reconstruction
Acquisition of PET/MRI
First, a MET PET/CT was acquired per clinical protocol 15 min after injection of 4 mBq/kg (max 400 mBq) 11C-methionine (MET), with an acquisition time of 15 min. A low dose CT (7.5 mAs 120 kVp) was acquired for both clinical PET/CT AC and PET/MRI CT-AC. After the standard PET/CT exam, subjects were moved to the PET/MRI facility, located in an adjacent building, and scanned in a GE Signa 3 T PET/MRI scanner using a 24-channel head-neck unit coil (GE Healthcare, Waukesha, WI). The average time from MET injection to PET acquisition at the PET/MRI scanner was 59 min (41–81), depending on the mobility of the subject. The mean duration of the MET-PET exam at the PET/MRI was 17.1 min (range 16–30 min). The MRI acquisition included the following scan protocol for AC: a 3D Liver Acquisition with Volume Acceleration-flexible (LAVA Flex) T1-weighted (GE Healthcare, Waukesha, WI) and a ZTE sequence were acquired. The LAVA Flex sequence was acquired for 14 s with 1 Number of excitations (NEX), 500 mm field of view (FOV), a 256 × 256 matrix and a 5.2 mm slice thickness and was used to generate water, fat and in- and out of phase echoes. ZTE sequence acquisition time was 42 s with 4 NEX, 264 mm FOV, a 110 × 110 matrix and 2.4 mm slice thickness. The clinical MRI protocol included T1, T2, T2*, postcontrast T1, DWI, ADC and Fractional Anisotropy for all the patients. Additional perfusion (ID 5–16) and spectroscopy (ID 10, 12, 14–18) sequences were also acquired.
Attenuation correction maps
For each subject, two different MR-based AC maps (ZTE–AC and atlas-AC) were generated and used for PET reconstruction and SUV calculation. For comparison, CT-AC was used as a reference standard method for PET AC.
ZTE based attenuation correction map
Following the method described in Wiesinger et al. [24] and using the MATLAB (version R2018b; The MathWorks) PET toolbox, a ZTE-based AC map was generated using:
-
Histogram-based bias correction and normalization
-
Tissue classification using a simple thresholding technique for soft tissue/bone and bone/air
-
Attenuating for bone (300–2000 HU) with a linear correlation between CT and ZTE MRI values
-
Classification of soft tissue fixed at 42 HU
Atlas-based attenuation correction map
Following the method described in Wollenweber et al. [25], an atlas-based AC map was generated using:
-
Enhancement of bone in the 3D FSPGR T1-weighted images using a Hessian filter
-
Rigid and non-rigid registration of bone-enhanced MRI to the CT skull atlas (provided by the manufacturer), creating a pseudo-CT
-
Transformation of the pseudo-CT to an AC map for 511 keV photons through standard energy conversion and resampling
-
Head coil and bed added to the AC map using a template
CT-based attenuation correction map
Following the method described in Burger et al. [26], CT-based AC map was generated using:
-
Rigid registration between the low dose head CT (7.5 mAs 120 kVp) and ZTE MR images using FMRIB’s Linear Image Registration (FLIRT) Tool Software Library v6.0, Oxford, UK. and in-house created MATLAB script
-
CT transformation following the same procedure as in the atlas-AC map
For each AC map, PET images were reconstructed using the GE PET toolbox (version MP26, GE Healthcare, Waukesha, WI) using TOF ordered subsets expectation maximization (OSEM) with three iterations, 16 subsets, and a 128 × 128 reconstruction matrix with a 256 mm FOV. A 3 mm Gaussian post-filter was applied to the reconstructed images.
ROI analysis
After MET PET/MRI acquisition, one of the authors (FDL) analysed the PET images using AW Server Volume Viewer 3D Viewer PET (GE Healthcare, Waukesha, WI). The analysis was blinded to the neuropathological diagnosis and clinical information. Results were analysed for the overall group and two subgroups (non-metal and metal), including pre- and postoperative patients respectively with suspected or confirmed brain tumour. Standardised uptake value (SUV) maps were calculated, normalized to body weight. A 27.4 mm2 elliptical region of interest (ROI) was delineated to include the PET hotspot with the help of gadolinium-enhanced T1weighted images from the PET/MRI or the prior MRI (Fig. 1a–d). For standardised uptake ratio (SUR) analysis, two different background regions were delineated:
-
Elliptic ROI (27.4 mm2) in the hotspot mirror region
-
Free-hand 2D ROI in the contralateral frontal cortex at the level of the basal ganglia
In the postoperative patients, all presenting with metal implants, SUV values were additionally obtained in the parenchyma tangential to the longitudinal axis of the adjacent metal implant and mirror region (Fig. 2a–d). In subjects with multiple surgical implants, the metal implant closest to the postoperative cavity and/or lesion was selected. SUR values – maxSURhotspot/mirror and maxSURhotspot/cortex; meanSURhotspot/mirror and meanSURhotspot/cortex; maxSURmetal/mirror and meanSURmetal/mirror—were calculated.
Statistical analysis
SUR values were analysed with the Bland–Altman (mean bias, SD, 95% CI), the Pearson correlation test (p < 0.05) and intraclass correlation reliability (ICC > 0.99) for absolute agreement.
Bland–Altman and Pearson correlation tests were performed using GraphPad Prism 8.4.3 (GraphPad Software, San Diego, California). Intraclass correlation reliability was performed using Statistical Package for Social Sciences (SPSS, IBM Corp. Released 2019. Version 26.0. Armonk, NY: IBM Corp.).
Mean bias was calculated as follows [27]:
Outlier analysis was performed using Robust regression and Outlier removal (ROUT) test [28].
Results
Fourteen patients with 18 lesions were included in the study. Seventeen lesions had a hotspot at PET/MRI and the diagnoses included 6 gliomas (WHO grade IV, n = 1; grade III, n = 3; grade II, n = 2), 9 metastases (melanoma, n = 3; breast, n = 4; kidney, n = 1; lung, n = 1) per WHO classification (2, 3) and two suspected brain tumours (two suspected low grade astrocytomas). The non-metal subgroup included 10 preoperative lesions. Seven out of seventeen lesions occurred in postoperative patients with metal implant. In the metal analysis (n = 8), one additional postoperative patient with astrocytoma WHO grade II without a hotspot on PET/MRI was included [29, 30]. The majority of the metal implants involved CranioFix titanium craniotomy clamps (B. Braun Medical, Melsungen, Germany) or Low-profile titanium plates or screws (DePuy Synthes, Warsaw, Indiana, United States). The majority of the lesions (n = 16) were treated using currently available treatment options in neuro-oncology and/or: surgery, chemotherapy, gamma knife radiosurgery or radiotherapy. Table 1.
Hotspot analysis
ZTE–AC demonstrated narrower SD and 95% CI than Atlas-AC in the hotspot analysis for all groups (ZTE overall ≤ 2.84, − 1.41 to 1.70; metal ≤ 1.67, − 3.00 to 2.20; non-metal ≤ 3.04, − 0.96 to 3.38; Atlas overall ≤ 4.56, − 1.05 to 3.83; metal ≤ 3.87, − 3.81 to 4.64; non-metal ≤ 4.90, − 1.68 to 5.86). Mean bias was within 2.4% for both ZTE–AC and atlas-AC compared to CT-AC. In the subgroup of patients with metal implants, atlas-AC demonstrated a narrower mean bias range—closer to zero—in three out of four analyses (atlas − 0.23 to 1.63, ZTE − 1.80 to 0.66). Results for the analysis of maxSURhotspot/cortex and maxSURhotspot/mirror are presented in Fig. 3a–f. A perfect correlation was found for both ZTE–AC and atlas-AC compared to CT-AC in the hotspot analysis (ZTE ρ 1.00, p < 0.0001; atlas ρ 1.00, p < 0.0001). Higher absolute values of correlation were found when the mirror region was chosen as background. The ROUT outliers test was performed and no outliers were identified for the analyses. Intraclass correlation reliability using absolute agreement between Atlas-, ZTE and CT maps for maxSUR and meanSUR values (hotspot/cortex; hotspot/mirror) was conducted and showed an almost perfect intraclass correlation coefficient (ICC > 0.99) in all the analyses. Detailed results for Bland–Altman, Pearson correlation and intraclass correlation reliability can be found in Additional file 1: Table S1, S3, Fig. S1–S3.
Metal analysis
In the metal region analysis, ZTE–AC demonstrated a narrower mean bias range—closer to zero—SD and 95% CI (ZTE 0.21–0.48, ≤ 2.50, − 1.70 to 2.57; Atlas 0.56–1.54, ≤ 4.01, − 1.81 to 4.89) (Fig. 4a, b). Both ZTE–AC and atlas-AC showed a perfect correlation compared to CT-AC (ZTE ρ 1.00, p < 0.0001; atlas ρ 1.00, p < 0.0001). ROUT outliers test was performed and no outliers were identified for the analysis. Intraclass correlation reliability using absolute agreement between Atlas-, ZTE and CT maps for maxSUR and meanSUR values (metal/mirror) was conducted and showed an almost perfect intraclass correlation coefficient (ICC > 0.99) in both analyses. Detailed results for Bland–Altman, Pearson correlation and intraclass correlation reliability can be found in Additional file 1: Table S2–S3, Fig. S4. Representative PET/CT images of CT-AC, PET and fused PET/CT are shown in Additional file 1: Fig. S5.
Discussion
In the present study, both ZTE–AC and atlas-AC showed a good performance against CT-AC.
Recently, Schramm and Ladefoged described the current state of the art in metal artefact correction strategies for PET/MRI [31]. Several studies have investigated AC methods in the presence of dental implants [32,33,34,35], whereas only one has focused on MR-AC in postoperative patients with brain tumour and surgical metal implants [11].
Previous studies in the field investigated ZTE–AC and atlas-AC compared to gold standard 68GE transmission scan and reference standard CT-AC [10, 22, 36]. Those studies demonstrated a better performance of ZTE–AC when compared to the reference standard used in the study.
In the present study, no consistent superiority between the ZTE–AC and atlas-AC was found for brain PET/MRI AC in the hotspot region and postoperative site. Both methods presented mean bias within ± 5% across the regions, as previously described in a multi-centre study for MR-AC in brain PET/MRI [37]. Our findings are in line with a previous work comparing ZTE–AC and atlas-AC using brain 15O-labeled water (H215O) PET/MRI [38]. The study showed no significant differences between AC for the regional values for cerebral blood flow. Notably, while the measurement of tracer uptake using H215O PET/MRI is affected by small variations of cerebral blood flow, semi-quantitative measurement using MET PET may be robust enough to achieve clinical objectives. Part of the design of the present work differs from previously published studies regarding the selection of reference method [10], patient selection [36,37,38,39] and lack of metal implants [10, 22, 36, 38, 39].
In the present work, a 2D elliptical ROI was used for measurement of SUR values. Notably, previous works have used a volume of interest (VOI) for delineation [36]. Takano et al. [40] recently described the value of both 2D and 3D ROI in PET analyses showing a better performance for 2D ROIs in FDG PET, apparent diffusion coefficient (ADC) map, FA (fractional anisotropy), and a comparable p value for 2D ROIs and 3D ROIs (p value = 0.0056; p value = 0.0050, respectively) in MET PET. Based on these results, and for practical clinical usefulness in the radiological work-flow, a 2D ROI was chosen.
A limitation of this study is attributed to the relatively small number of patients included. This is explained by the high cost, limited use of PET/MRI, and the lack of universal validation in neuro-oncology [41].
Although the timing of PET/MRI was not standardised, the comparison between the AC methods might not be affected by the average post-injection time, since the acquisition for all the three AC maps was simultaneous. The aim of this study was not to render a diagnosis based on SUV values, which depend on acquisition timing, noise, tracers and tissue composition [42,43,44]. The evaluation of ZTE- and atlas-AC compared to CT-AC was performed on a patient-by-patient basis using SUV/SUR values, as previously reported in literature [11]. Further investigation will be necessary to explore the impact of the intrinsic differences between the AC maps on the accuracy of each AC-method compared to the reference gold standard, and to assess the severity of artefacts on MR-AC based on the size and appearance of the metal implant.
This is a single-centre, retrospective study. Future multi-centre prospective studies with larger cohorts, analysis of MRI biomarkers, single pathology brain tumours at the same treatment stage after diagnosis and analysis of MRI biomarkers will be needed to confirm our preliminary results. The study was performed using a single-reader assessment. Multi-reader assessment can be used in the upcoming study for ROI delineation, SUR analyses and inter-reader reliability. The study included all available patients from one institution during a specific time. Due to the retrospective nature of the study, no pre-study calculation of power was performed.
Conclusions
This study retrospectively evaluated 14 pre- and postoperative patients with suspected or confirmed brain tumour. Good performance of ZTE–AC and atlas-AC compared to reference standard CT-AC was found for all analyses. Our results indicate that both ZTE and atlas are feasible MET PET/MRI AC methods.
Availability of data and materials
All data used and/or analysed during the current study are available from the corresponding author, Francesca De Luca (francesca.de.luca@ki.se), on reasonable request.
Abbreviations
- 3D:
-
Three-dimensional
- AC:
-
Attenuation correction
- AW server:
-
Advantage workstation server
- CI:
-
Confidence intervals
- CT:
-
Computer tomography
- FET:
-
Fluoroethyl-l-tyrosine
- FDG:
-
Flurodeoxyglucose
- FOV:
-
Field-of-view
- FSPGR:
-
Fast spoiled gradient echo
- GE:
-
General electric
- H215O:
-
15O-labeled water
- ID:
-
Identity
- LAVA:
-
Liver acquisition with volume acceleration
- MET:
-
Methionine
- MR:
-
Magnetic resonance
- MRI:
-
Magnetic resonance imaging
- NEX:
-
Number of excitations
- OSEM:
-
Ordered subset expectation maximization
- PET:
-
Positron emitting tomography
- ROI:
-
Region of interest
- ROUT:
-
Robust regression and outlier removal
- SD:
-
Standard deviation
- SUR:
-
Standardised uptake value ratio
- SUV:
-
Standardised uptake value
- TOF:
-
Time-of-flight
- UTE:
-
Ultrashort echo time
- ZTE:
-
Zero echo time
References
Ehman EC, Johnson GB, Villanueva-Meyer JE, Cha S, Leynes AP, Larson PEZ, et al. PET/MRI: where might it replace PET/CT? J Magn Reson Imaging. 2017;46(5):1247–62.
Glaudemans AW, Enting RH, Heesters MA, Dierckx RA, van Rheenen RW, Walenkamp AM, et al. Value of 11C-methionine PET in imaging brain tumours and metastases. Eur J Nucl Med Mol Imaging. 2013;40(4):615–35.
Minamimoto R, Saginoya T, Kondo C, Tomura N, Ito K, Matsuo Y, et al. Differentiation of brain tumor recurrence from post-radiotherapy necrosis with 11C-methionine PET: visual assessment versus quantitative assessment. PLoS ONE. 2015;10(7):e0132515.
Tomura N, Kokubun M, Saginoya T, Mizuno Y, Kikuchi Y. Differentiation between treatment-induced necrosis and recurrent tumors in patients with metastatic brain tumors: comparison among. AJNR Am J Neuroradiol. 2017;38(8):1520–7.
Verma N, Cowperthwaite MC, Burnett MG, Markey MK. Differentiating tumor recurrence from treatment necrosis: a review of neuro-oncologic imaging strategies. Neuro Oncol. 2013;15(5):515–34.
Chung JK, Kim YK, Kim SK, Lee YJ, Paek S, Yeo JS, et al. Usefulness of 11C-methionine PET in the evaluation of brain lesions that are hypo- or isometabolic on 18F-FDG PET. Eur J Nucl Med Mol Imaging. 2002;29(2):176–82.
Zhao C, Zhang Y, Wang J. A meta-analysis on the diagnostic performance of (18)F-FDG and (11)C-methionine PET for differentiating brain tumors. AJNR Am J Neuroradiol. 2014;35(6):1058–65.
Galldiks N, Lohmann P, Albert NL, Tonn JC, Langen KJ. Current status of PET imaging in neuro-oncology. Neurooncol Adv. 2019;1(1):vdz010.
Wagenknecht G, Kaiser HJ, Mottaghy FM, Herzog H. MRI for attenuation correction in PET: methods and challenges. MAGMA. 2013;26(1):99–113.
Sousa JM, Appel L, Engström M, Papadimitriou S, Nyholm D, Larsson EM, et al. Evaluation of zero-echo-time attenuation correction for integrated PET/MR brain imaging-comparison to head atlas and. EJNMMI Phys. 2018;5(1):20.
Ladefoged CN, Andersen FL, Kjær A, Højgaard L, Law I. RESOLUTE PET/MRI attenuation correction for O-(2-. Front Neurosci. 2017;11:453.
Benoit D, Ladefoged CN, Rezaei A, Keller SH, Andersen FL, Højgaard L, et al. Optimized MLAA for quantitative non-TOF PET/MR of the brain. Phys Med Biol. 2016;61(24):8854–74.
Merida I, Costes N, Heckemann R, Hammers A. Pseudo-CT generation in brain MR-PET attenuation correction: comparison of several multi-atlas methods. EJNMMI Phys. 2015;2(Suppl 1):A29.
Hofmann M, Pichler B, Schölkopf B, Beyer T. Towards quantitative PET/MRI: a review of MR-based attenuation correction techniques. Eur J Nucl Med Mol Imaging. 2009;36(Suppl 1):S93-104.
Pearce MS, Salotti JA, Little MP, McHugh K, Lee C, Kim KP, et al. Radiation exposure from CT scans in childhood and subsequent risk of leukaemia and brain tumours: a retrospective cohort study. Lancet. 2012;380(9840):499–505.
Keereman V, Mollet P, Berker Y, Schulz V, Vandenberghe S. Challenges and current methods for attenuation correction in PET/MR. MAGMA. 2013;26(1):81–98.
Martinez-Moller A, Souvatzoglou M, Delso G, Bundschuh RA, Chefd’hotel C, Ziegler SI, et al. Tissue classification as a potential approach for attenuation correction in whole-body PET/MRI: evaluation with PET/CT data. J Nucl Med. 2009;50(4):520–6.
Delso G, Wiesinger F, Sacolick LI, Kaushik SS, Shanbhag DD, Hullner M, et al. Clinical evaluation of zero-echo-time MR imaging for the segmentation of the skull. J Nucl Med. 2015;56(3):417–22.
Hargreaves BA, Worters PW, Pauly KB, Pauly JM, Koch KM, Gold GE. Metal-induced artifacts in MRI. AJR Am J Roentgenol. 2011;197(3):547–55.
Davison H, ter Voert EE, de Galiza BF, Veit-Haibach P, Delso G. Incorporation of time-of-flight information reduces metal artifacts in simultaneous positron emission tomography/magnetic resonance imaging: a simulation study. Invest Radiol. 2015;50(7):423–9.
Rezaei A, Schramm G, Willekens SM, Delso G, Van Laere K, Nuyts J. A quantitative evaluation of joint activity and attenuation reconstruction in TOF-PET/MR brain imaging. J Nucl Med. 2019;60(11):1649–55.
Sekine T, Ter Voert EE, Warnock G, Buck A, Huellner M, Veit-Haibach P, et al. Clinical evaluation of zero-echo-time attenuation correction for brain 18F-FDG PET/MRI: comparison with atlas attenuation correction. J Nucl Med. 2016;57(12):1927–32.
Delso G, Kemp B, Kaushik S, Wiesinger F, Sekine T. Improving PET/MR brain quantitation with template-enhanced ZTE. Neuroimage. 2018;181:403–13.
Wiesinger F, Sacolick LI, Menini A, Kaushik SS, Ahn S, Veit-Haibach P, et al. Zero TE MR bone imaging in the head. Magn Reson Med. 2016;75(1):107–14.
Wollenweber SD, et al. Evaluation of an atlas-based PET head attenuation correction using PET/CT & MR patient data IEEE transactions on nuclear science. IEEE Trans Nucl Sci. 2013;60(5):3383–90.
Burger C, Goerres G, Schoenes S, Buck A, Lonn AH, Von Schulthess GK. PET attenuation coefficients from CT images: experimental evaluation of the transformation of CT into PET 511-keV attenuation coefficients. Eur J Nucl Med Mol Imaging. 2002;29(7):922–7.
Bland JM, Altman DG. Measuring agreement in method comparison studies. Stat Methods Med Res. 1999;8(2):135–60.
Motulsky HJ, Brown RE. Detecting outliers when fitting data with nonlinear regression—a new method based on robust nonlinear regression and the false discovery rate. BMC Bioinformatics. 2006;7:123.
Louis DN, Ohgaki H, Wiestler OD, Cavenee WK, Burger PC, Jouvet A, et al. The 2007 WHO classification of tumours of the central nervous system. Acta Neuropathol. 2007;114(2):97–109.
Louis DN, Perry A, Reifenberger G, von Deimling A, Figarella-Branger D, Cavenee WK, et al. The 2016 world health organization classification of tumors of the central nervous system: a summary. Acta Neuropathol. 2016;131(6):803–20.
Schramm G, Ladefoged CN. Metal artifact correction strategies in MRI-based attenuation correction in PET/MRI. BJRn|Open. 2019;1(1):20190033.
Ladefoged CN, Hansen AE, Keller SH, Fischer BM, Rasmussen JH, Law I, et al. Dental artifacts in the head and neck region: implications for Dixon-based attenuation correction in PET/MR. EJNMMI Phys. 2015;2(1):8.
Gunzinger JM, Delso G, Boss A, Porto M, Davison H, von Schulthess GK, et al. Metal artifact reduction in patients with dental implants using multispectral three-dimensional data acquisition for hybrid PET/MRI. EJNMMI Phys. 2014;1(1):102.
Burger IA, Wurnig MC, Becker AS, Kenkel D, Delso G, Veit-Haibach P, et al. Hybrid PET/MR imaging: an algorithm to reduce metal artifacts from dental implants in Dixon-based attenuation map generation using a multiacquisition variable-resonance image combination sequence. J Nucl Med. 2015;56(1):93–7.
Ladefoged CN, Andersen FL, Keller SH, Beyer T, Law I, Højgaard L, et al. Automatic correction of dental artifacts in PET/MRI. J Med Imaging (Bellingham). 2015;2(2):024009.
Sgard B, Khalifé M, Bouchut A, Fernandez B, Soret M, Giron A, et al. ZTE MR-based attenuation correction in brain FDG-PET/MR: performance in patients with cognitive impairment. Eur Radiol. 2020;30(3):1770–9.
Ladefoged CN, Law I, Anazodo U, St Lawrence K, Izquierdo-Garcia D, Catana C, et al. A multi-centre evaluation of eleven clinically feasible brain PET/MRI attenuation correction techniques using a large cohort of patients. Neuroimage. 2017;147:346–59.
Okazawa H, Tsujikawa T, Higashino Y, Kikuta KI, Mori T, Makino A, et al. No significant difference found in PET/MRI CBF values reconstructed with CT-atlas-based and ZTE MR attenuation correction. EJNMMI Res. 2019;9(1):26.
Schramm G, Koole M, Willekens SMA, Rezaei A, Van Weehaeghe D, Delso G, et al. Regional accuracy of ZTE-based attenuation correction in static [18F]FDG and dynamic [18F]PE2I brain PET/MR. Front Phys. 2019;7:211.
Takano K, Kinoshita M, Arita H, Okita Y, Chiba Y, Kagawa N, et al. Influence of region-of-interest designs on quantitative measurement of multimodal imaging of MR non-enhancing gliomas. Oncol Lett. 2018;15(5):7934–40.
Martin O, Schaarschmidt BM, Kirchner J, Suntharalingam S, Grueneisen J, Demircioglu A, et al. PET/MRI versus PET/CT in whole-body staging: results from a unicenter observational study in 1003 subsequent examinations. J Nucl Med. 2019.
Aki T, Nakayama N, Yonezawa S, Takenaka S, Miwa K, Asano Y, et al. Evaluation of brain tumors using dynamic 11C-methionine-PET. J Neurooncol. 2012;109(1):115–22.
Kim DW, Jung SA, Kim CG, Park SA. The efficacy of dual time point F-18 FDG PET imaging for grading of brain tumors. Clin Nucl Med. 2010;35(6):400–3.
Spence AM, Muzi M, Mankoff DA, O’Sullivan SF, Link JM, Lewellen TK, et al. 18F-FDG PET of gliomas at delayed intervals: improved distinction between tumor and normal gray matter. J Nucl Med. 2004;45(10):1653–9.
Acknowledgements
The authors would like to thank the radiographers Yords Österman, Aleksandra Ramolli and Soroush Sohrabian for the technical help and assistance during the scan acquisition and the Departments of Nuclear Medicine and Neuroradiology for valuable assistance with the project. We also thank Mattia Russel Pantalone, MD, Department of Neurology, Karolinska University Hospital for the valuable feedback and technical support.
Funding
Open Access funding provided by Karolinska Institute. This work was supported by Åke Wibergs Stiftelse (reference number M18-0132) and Magnus Bergvalls Stiftelse (reference number 2018-02643).
Author information
Authors and Affiliations
Contributions
FDL, AFD and HM designed the study. FDL analysed the data and drafted the manuscript. MB performed the reconstruction. LB and CW contributed to the image analysis pipeline and critically evaluated the manuscript. AFD, HM and MB contributed to drafting of the manuscript. FDL and HM were responsible for the patients’ recruitment. All authors read and approved the final version of the manuscript.
Corresponding author
Ethics declarations
Ethics approval and consent to participate
The study was conducted in accordance to the declaration of Helsinki and approved by the Swedish Ethical Review Authority (2019-01309), through which informed consent was waived.
Consent for publication
Not applicable.
Competing interests
Nothing to disclose. Karolinska University Hospital has several research agreements with different radiology vendors. No money has been paid from any of these vendors directly to the authors for this study.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary information
Additional file 1:
PDF file containing Supplementary material. Table S1. Hotspot analysis. Correlation and agreement for hotspot analysis in the overall group, metal and non-metal subgroups. As background, contralateral frontal cortex and mirror ROI were respectively assessed. ZTE/Atlas-AC compared to reference gold standard CT-AC. Table S2. Metal analysis. Correlation and agreement for parenchyma analysis in the metal subgroup. As background, contralateral mirror parenchyma ROI was assessed. ZTE/Atlas-AC compared to reference gold standard CT-AC. Table S3. Intraclass correlation reliability. Intraclass correlation reliability for absolute agreement in hotspot and metal analysis. Comparison among ZTE, Atlas and CT for SUR values. Fig. S1. Pearson correlation for hotspot analysis in the overall group. Correlation analysis for hotspot analysis in the overall group. As background, contralateral frontal cortex and mirror for maxSUR (a,b) and meanSUR (c,d) ROI were respectively assessed. ZTE/Atlas-AC compared to reference gold standard CT-AC. Fig. S2. Pearson correlation for hotspot analysis in the metal subgroup. Correlation analysis for hotspot analysis in the metal subgroup. As background, contralateral frontal cortex and mirror for maxSUR (a,b) and meanSUR (c,d) ROI were respectively assessed. ZTE/Atlas-AC compared to reference gold standard CT-AC. Fig. S3. Pearson correlation for hotspot analysis in the non-metal subgroup. Correlation analysis for hotspot analysis in the non-metal subgroup. As background, contralateral frontal cortex and mirror for maxSUR (a,b) and meanSUR (c,d) ROI were respectively assessed. ZTE/Atlas-AC compared to reference gold standard CT-AC. Fig. S4. Pearson correlation for metal analysis. Correlation for parenchyma analysis in the metal subgroup. As background, contralateral mirror parenchyma for maxSUR (a) and meanSUR (b) ROI was assessed. ZTE/Atlas-AC compared to reference gold standard CT-AC. Fig. S5. PET/CT images. Representative PET/CT images of CT-AC, PET and fused PET/CT are shown for a postoperative patient with metal implants.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.
About this article
Cite this article
De Luca, F., Bolin, M., Blomqvist, L. et al. Validation of PET/MRI attenuation correction methodology in the study of brain tumours. BMC Med Imaging 20, 126 (2020). https://doi.org/10.1186/s12880-020-00526-8
Received:
Accepted:
Published:
DOI: https://doi.org/10.1186/s12880-020-00526-8