Abstract
Tumor-to-normal ratio (T/N) measurement of 18F-FBPA is crucial for patient eligibility to receive boron neutron capture therapy. This study aims to compare the difference in standard uptake value ratios on brain tumors and normal brains using PET/MR ZTE and atlas-based attenuation correction with the current standard PET/CT attenuation correction. Regarding the normal brain uptake, the difference was not significant between PET/CT and PET/MR attenuation correction methods. The T/N ratio of PET/CT-AC, PET/MR ZTE-AC and PET/MR AB-AC were 2.34 ± 0.95, 2.29 ± 0.88, and 2.19 ± 0.80, respectively. The T/N ratio comparison showed no significance using PET/CT-AC and PET/MR ZTE-AC. As for the PET/MRI AB-AC, significantly lower T/N ratio was observed (− 5.18 ± 9.52%; p < 0.05). The T/N difference between ZTE-AC and AB-AC was also significant (4.71 ± 5.80%; p < 0.01). Our findings suggested PET/MRI imaging using ZTE-AC provided superior quantification on 18F-FBPA-PET compared to atlas-based AC. Using ZTE-AC on 18F-FBPA-PET /MRI might be crucial for BNCT pre-treatment planning.
Similar content being viewed by others
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.Introduction
Boron Neutron Capture Therapy (BNCT) has become a promising cancer treatment scheme that uses the neutron capture and fission reaction of the 10B boron isotope (10B). The fission reaction is induced by an epithermal neutron and yields a α particle with high linear energy transfer (LET) and a 7Li particle1. To achieve an expected treatment response and to mitigate the adverse effect, BNCT requires precise estimation of 10B concentration not only in the target tumor but in neighboring normal tissues2. Presently, the theranostic combination of the 10B boron-carrier “boronophenylalanine (BPA)” and the imaging probe [18F]-fluoro-BPA ([18F]-FBPA) have been used in most contemporary clinical treatments. The FBPA image by PET/CT scanner is implemented for treatment plans and uptake measurements3 and is the modality of choice to monitor the 10B boron-carrier distribution before BNCT treatment.
To ensure sophisticated treatment plans for BNCT, physicians need to determine precisely the 10B concentrations in targeted lesions and the surrounding normal tissues4 with a tumor-to-normal ratio(T/N ratio) ≥ “2.5”5–9. However, the CT in PET/CT provides brain anatomical images inferior to those of the magnetic resonance (MR) technique in PET/MR; therefore, the limited spatial resolution from PET/CT increases the difficulty in tumor delineation for precision BNCT planning, especially in brain regions10. Consequently, PET/MR has become more commonly used in the era of contemporary molecular imaging because it provides comprehensive anatomical information without the ionizing radiation exposure that occurs with CT. The simultaneous morphological and metabolic imaging of PET/MR fulfilled the requirements for intricate BNCT planning in brain tumors, especially for tumor boundary recognition and segmentation. Moreover, PET/MR is also promising for normal tissue determination since a previous report found that the delineation of normal tissue significantly influenced the T/N ratio measurement and treatment management.
Alternatively, to measure the 18F-FBPA standardized uptake value (SUV) and SUV ratio (SUVR) in PET/MR, it is necessary to perform photon-attenuation correction (AC) for optimized photon quantification. Photon attenuation in the body is mainly caused by soft tissues, bone and MRI hardware components11,8,13. In practice, this attenuation remains even with consistent agreement between SUVRs obtained from PET/CT and PET/MR. However, PET/MR AC remains debatable14. Existing PET/CT and PET/MR data for [18F]-fluorodeoxyglucose ([18F]-FDG) PET imaging indicated that MR-based AC is accurate and reliable in most tissue types, with SUVR deviations generally less than 10%15. With the development of hybrid modalities, there are four different AC approaches available in clinical use: PET/CT AC (CT-AC), atlas-based AC (AB-AC), ultrashort echo-time AC (UTE-AC), and zero echo-time AC (ZTE-AC). Because the image principle of segmentation-based AC will cause a considerable loss of bone signals16, especially in postoperative patients with deformed skull17, an alternative AC scheme for better clinical applications is required. With the advent of the ZTE technique, PET/MR AC has been advocated to overcome limitations for whole-body18 and whole brain19 studies. Although a slight underestimation might occur in the whole-brain fields17,20, 21, existing data indicates that ZTE-AC has been proven to offer various benefits for radiotherapy planning in the brain16,19. Previous FDG studies also have found that ZTE-AC provided an advanced quantitative evaluation in brain PET18,21, 22. However, to our knowledge, no report has focused on 18F-FBPA brain PET/MR yet. Thus, this study aimed to evaluate and compare the effects of ZTE-AC and AB-AC on the power of PET/MR quantification with those of the standard CT-AC in normal-brain and tumor-brain tissues.
Methods
Participants and tumor VOI (volume of interest) selection
Between July 2019 to January 2022, a total of 28 patients with confirmed brain tumors (17 females and 11 males, mean age = 56 ± 15.97 years old) who were potential candidates for BNCT were enrolled (Table 1). All received PET/CT and PET/MR examinations were performed on the same day. The patients’ tumor types included 14 glioblastomas, 3 atypical meningiomas, 2 astrocytomas, 3 anaplastic astrocytomas, 1 oligoastrocytoma, 3 anaplastic oligodendrogliomas, and 3 metastatic cancers. The tumor regions of all datasets were delineated manually by an experienced radiographer and double-checked by an experienced nuclear medicine physician. This study was approved by the Institutional Review Board of Taipei Veterans General Hospital (IRB: 2022-11-003AC).
Data acquisition
Data were acquired on a PET/CT scanner (Discovery MI DR; GE Healthcare, Milwaukee, WI, USA) and a 3T PET/MR scanner (SIGNA PET/MR; GE Healthcare, Milwaukee, WI, USA) using a geometry-embracing method (GEM) head-and-neck unit for signal detection and a whole-body coil for radio-frequency excitation. The following settings were used to acquire the helical CT scans in PET/CT: 120 kVp; 30–130 mA; slice thickness of 3.75 mm; field of view (FOV) 512 × 512. In PET/MR imaging, some clinical images were acquired for diagnostic purposes before PET scan, including fast-SE (FSE) with inversion-recovery axial plane T1-weighted imaging with a repetition time (TR) of 1800 ms, echo time (TE) of 23 ms, inversion time (TI) of 750 ms), FSE coronal plane T2-weighted imaging (TR, 4545 ms; TE, 110 ms), and T2-fluid-attenuated inversion recovery imaging (TR, 9000 ms; TE, 92 ms; TI, 2472 ms). Single echo-planar-imaging based diffusion-weighted images (TR, 6000 ms; TE, 77 ms) were also acquired with a b-value of 1000 s/mm2. Then, the ZTE sequence (TR/TE = 4.048/0.0167 ms; FOV: 60 × 60 cm; matrix size: 128 × 128; slice thickness: 2.78 mm; flip angle: 5°) was used for ZTE-AC, with one structural imaging dataset, liver imaging with volume acceleration flexible (LAVA Flex), and the Dixon method. The LAVA Flex parameters were TR/TE of 5.35/1.936 ms, FOV of 40 × 40 cm, matrix size of 512 × 512, slice thickness of 2 mm, and flip angle of 12°. The Dixon method parameters were TR/TE of 4/1.1 and 2.2 ms, FOV of 50 × 37.5 cm, matrix size of 256 × 128; slice thickness of 5.2 mm, and flip angle of 5° (Table 2).
The PET/CT and PET/MR scans were acquired at 34.2 ± 2.3 min and 68.4 ± 4.6 min after intravenous injection, respectively. In both modalities, PET data were accumulated for 15 min to reconstruct the 18F-FBPA PET images. The average radioactivity at the start of the 18F-FBPA PET scanning was 159 ± 34.9 MBq in PET/CT (dose range: 109–260 MBq) and 128 ± 26.8 MBq in PET/MR (dose range: 86 -188 MBq) as shown in Table 1. In PET/MRI, two AC maps were generated by the ZTE-AC and AB-AC methods. The reconstruction algorithm was used with the following settings: 2 iterations, 28 subsets, 5-mm filter cutoff, 192 × 192 matrix size, and 60-cm display field-of-view (DFOV). In PET/CT, the reconstruction algorithm was used with 2 iterations, 24 subsets, 5-mm filter cutoff, 256 × 256-matrix size, and 70-cm DFOV.
Data preprocess and analysis
All data preprocessing steps were performed in the SPM12 software. First, the SUV of 18F-FBPA was calculated for each voxel of PET/CT, PET/MR ZTE-AC, and PET/MR AB-AC. Second, the SUV acquired by PET/CT and the structural LAVA images of each patient were co-registered into the spatial orientation of each SUV image dataset by PET/MR. Third, spatial segmentation was applied in the registered structural images for extraction of white matter, gray matter, and cerebrospinal fluid by normalizing the Montreal Neurological Institute space to each individual. Fourth, the partial-volume effect of all image data was corrected by the PETPVE12 toolbox. Fifth, to acquire precise tumor- and normal-tissue delineation, a senior physician determined the accurate delineation. Sixth, to evaluate the differences between different AC methods, the VOI-wise analysis was equipped with the Automated Anatomical Labeling template version 3 (AAL3)23 (Fig. 1).
Following the clinical usage, the core area of a tumor region was generated manually by volume delineation, which covered the whole-tumor contour in three dimensions and was exploited by applying a 50% SUV threshold. On the basis of the previous literature10, the SUVs of tumor regions were divided into SUVs of cerebellar lobule Crus I-II for the following T/N ratio with normal tissue is cerebellum analysis. In group-level analysis, the significant differences in T/N ratio levels between CT-AC, MRI ZTE-AC, and MRI AB-AC were evaluated by performing paired t-tests. Moreover, to measure and visualize the regional differences in SUVRs between PET/CT, PET/MR ZTE-AC, and PET/MR AB-AC, anatomical volumes of interest analyses by AAL323 were performed to separate participants’ brains into eight major categories: the frontal region in light green, the limbic region in light blue, the occipital region in light yellow, the parietal region in light gray, the striatal region in. light orange, the temporal region in dark purple, the cerebellar region in dark gray, thalamic region in green and the anterior cingulate in yellow, (Fig. 3). To compare the gold standard PET/CT AC, Eq. (1) was applied to determine the comparisons of PET/MR AB-AC and PET/MR ZTE-AC with PET/CT AC individually. In the individual PET images of each patient, all manually-selected tumor regions were excluded on the basis of the MR images by an experienced physician who performed BNCT planning for > 12 years. Subsequently, the AAL3 template was applied to the residual brain regions to complete the normal-tissue delineation assessment. The VOIs in only ≤ 18 patients (< 2/3 of the enrolled subjects) were excluded in the following group analysis. Subsequently, to normalize the variation among patients, the SUVR was calculated from the SUV of each VOI divided into the average level of the SUV of the bilateral cerebellar lobule Crus I–II. In group-leveled data analysis, a paired t-test was also performed for the VOI-wise SUVR box plotting analyses between different AC approaches by PET/CT, PET/MR-ZTE, and PET/MR AB-AC (Fig. 4).
Ethical approval
This study was performed in line with the principles of the Declaration of Helsinki. Approval was granted by the Institutional Review Board of Taipei Veterans General Hospital (IRB: 2022-11-003AC).
Consent to participate
The requirement for informed consent was waived due to the nature of retrospective study and was approved by the Institutional Review Board of Taipei Veterans General Hospital.
Results
The FBPA-PET image processing with CT-AC, MR ZTE-AC, or MR AB-AC showed no obvious differences in both tumor and normal brain uptakes by the visual inspection (Fig. 2a–d). Use of the AAL3 template showed that the whole brain could be separated into 166 sub regions, and those containing < 30 voxels were excluded in the following analysis. In the normal tissue determination using parametric methods, the differences in SUVRs between PET/CT and PET/MR AC methods ranged from − 15 to + 15%, and almost all VOIs showed no significant difference by visual inspection between SUVRs in CT-AC, ZTE-AC, and AB-AC (Fig. 3a). Only the SUVRs of the right cerebellar lobule VIII by PET/MR-ZTE were significantly higher than those by PET/CT (Bonferroni corrected p < 0.05) (Fig. 3b). To clarify the performance between two PET/MR AC methods, we normalized all the individual AC maps (with tumor regions removing) into MNI-template and acquire the group-levelled AC maps. The result showed that ZTE-AC shows less difference with the CT-AC than AB-AC (Fig. 5).
In the T/N ratio assessment, the corrected ratios by PET/CT AC, ZTE-AC, and AB-AC were 2.34 ± 0.95, 2.29 ± 0.88, and 2.19 ± 0.80, respectively. Compared with the T/N ratios from PET/CT AC, those from ZTE-AC were not significantly different, whereas there were statistically significant differences between those for AB-AC and for CT-AC and ZTE-AC (Bonferroni corrected p < 0.05) (Fig. 4a). There were no significant differences in the T/N ratios between ZTE-AC and PET/CT AC, but the values were slightly lower (− 0.94% ± 8.97%; p = 0.58) than those of PET/CT AC. The T/N ratio was significantly lower for the AB-AC method (− 5.18% ± 9.52%; p < 0.05). The difference in the T/N ratios between ZTE-AC and AB-AC was also statistically significant (4.71% ± 5.80%; p < 0.01) (Fig. 4b).
Discussion
Our data revealed that for [18F]-FBPA quantification in normal brain tissues, there were no significant differences in the T/N ratios among PET/MR, ZTE-AC, AB-AC, and the current-standard PET/CT-AC. In the brain tumors, the T/N ratios from PET/MR with ZTE-AC were similar to those for the current-standard PET/CT-AC, and significantly lower T/N ratios were observed for PET/MR AB-AC.
In the past decade, PET/CT has had an important role in the management of cancer patients because it not only offers accurate attenuation correction by low-energy CT, but also reduced the scanning time by > 40% relative to the time for whole-body scans by conventional PET study24. Moreover, in terms of whole-body scanning, PET/CT provides acceptable anatomic and morphologic information for diagnosis and treatment planning. However, for brain tumors, CT images from a PET/CT scanner render delineation of lesions and normal-brain tissues inferior to that from MRI images, so PET/CT images enable precise dosimetry planning before radiation therapy25.
The T/N ratio assessment of [18F]-FBPA-PET is crucial before BNCT to ensure suitable patient selection and treatment planning. However, in our clinical practice, we found some difficulties using PET/CT for measuring the T/N ratio in our brain tumor cohort, which was caused by determining tumor delineation, brain perifocal edematous effects, or anatomic structural changes following previous surgery or radiation therapy. Therefore, after the installation of a hybrid PET/MRI scanner, PET/MR gradually became the dominant modality of choice for patient evaluations before BNCT.
Although the anatomic and metabolic images from PET/MRI are favorable for BNCT evaluation, the reliability of PET/MR AC remains unclear. With the development of various MR AC methods, three prominent approaches, i.e., AB-AC, UTE-AC, and ZTE-AC, have been frequently adopted. The first PET/MR AC method is AB-AC. To generate the AB-AC, the segmentation-driven AC maps are classified by Dixon-based MR images and divided into different tissues, such as air, bone, and soft tissue, with standardized attenuation coefficient values26. Then, the AB-AC maps can be equipped by combining template and segmentation methods, including CT images, MRI models, and a brain mask27,24,25,30. The second method is UTE-AC (Biograph mMR PET/MRI from Siemens Healthcare), which utilizes the short T2* to acquire tissue signals as the attenuation map (μ-map). This method has established a linear relationship between the relaxation rate, R2*, stemming from 1/T2* and CT density which comes from the attenuation values in bone MRI images31,32. Currently, the ZTE-AC (SIGNA PET/MRI from GE Healthcare) method is based on the pseudo-CT attenuating property to develop the subject-specific AC map17. The ZTE-AC approach enables precise quantitative evaluation of brain PET according to previous FDG studies18,21, 22. BNCT is a type of α-particle radiotherapy. Precise delineation of the tumor lesion and separating it from normal tissue based on [18F]-FBPA PET result is crucial for successful treatment and minimization of side effects. However, there have been no reports that focused on optimizing [18F]-FBPA brain PET/MR imaging for the same purpose. The rationale for this study was to observe the imaging outcomes after applying different attenuation-correction methods, i.e., ZTE-AC and AB-AC, for [18F]-FBPA PET/MR quantification accuracy and compare them with the current benchmark of PET/CTAC.
In the present study, we could not visually determine the differences between [18F]-FBPA PET images from CT-AC, MR ZTE-AC, or MR AB-AC (Fig. 2a–d). To reveal the precise uptake differences in normal brain tissues, we used AAL3 segmentation to perform automated labeling with subtle brain segmentations and to calculate the proportion of labeled brain sub-regions23. Small SUVR differences in all normal brain regions between the three attenuation-correction methods were observed, but only higher SUVR values in the right cerebellar lobule III were observed for ZTE-AC (Fig. 3a,b). Some regions, such as the thalamus and anterior cingulate, exhibited high agreement between CT-AC, MR ZTE-AC, and MR AB-AC.
Our study found no significant differences in the [18F]-FBPA T/N ratio between the current gold standard CT-AC and ZTE-AC, whereas a significantly lower T/N ratio of AB-AC (Fig. 4a,b) reflects a problem of underestimation. These results suggested that PET/MRI AB-AC may underestimate the [18F]-FBPA T/N ratio and could result in the failure of patient recruitment in the BNCT treatment management. The main reason for underestimating AB-AC in the T/N ratio may be generated by undervaluing the tumoral uptake relative to that of ZTE-AC and CT-AC. If ZTE-AC is used, the T/N ratio in [18F]-FBPA PET data might be closer to that in real-world settings that perform CT-AC by generating a pseudo-CT. This pseudo-CT not only provides a linear correlation with CT density of the bone in Hounsfield units (HU) but also a subject-specific AC map that provides flexibility and an advanced AC calculation17. Moreover, ZTE intensity in some soft-tissue regions with misclassification between air and bone, such as spongious temporal bones, frontal sinus region, mastoid region, and nasal-sinus cartilage, is closer to the phenomenon of the partial-volume effect in the same regions observed by PET/CT17,21. In the brain region, ZTE-AC uses classification along with the CT information in the (0,100) HU soft-tissue range21. Although previous studies have provided significant differences that may be seen in the neck that are related to the different procedures used in PET/CT and PET/MR17, all subjects recruited in this study were examined for brain tumors. Till date, no literature has discussed the effect of ZTE-AC and AB-AC on [18F]-FBPA PET/MRI. Our findings suggest that PET/MR imaging using ZTE-AC can provide [18F]-FBPA-PET quantification superior to that of AB-AC because the statistical results from PET/MR ZTE-AC are more similar to those in PET/CT AC. Using ZTE-AC on [18F]-FBPA, PET/MRI may have a crucial role in BNCT pretreatment planning in determining the T/N ratio.
Regarding the clinical impact, 12 patients can be included in BNCT treatments using PET/CT AC measurements. Sixteen patients cannot be included in because the standard of T/N ratio ≥ 2.5 was adopted. Similar results were observed with ZTE-AC PET/MRI, when T/N ratio ≥ 2.5 was found in 13 patients. However, only 9 subjects received T/N ≥ 2.5 using AB-AC PET/MRI. Three patients will be excluded from BNCT treatments attributed to the underestimation of T/N from AB-AC. Consequently, 25% of subjects is eligible and can be include in BNCT treatments by using PET/MR ZTE-AC and PET/CT AC.
Our study had some limitations that should be considered. First, the different spatial resolutions between two hybrid machines should be considered carefully. Another limitation is the small study sample size. However, a larger-scale prospective study might be difficult to conduct because (18F)-FBPA-PET and the subsequent BNCT therapy are part of a compassionate treatment combination with strict regulation from our institutional review board. We hope a further multi-center meta-analysis focusing on (18F)-FBPA-PET using hybrid PET/MRI would be able to validate our results. A third study limitation is that because the principles of ZTE are related to the use of pseudo-CT for AC, some uncertain factors might affect the ZTE-AC map. For example, in our study, we found two outliers shown in Fig. 4b. One patient was affected by a large pseudo-skull material following his previous surgery. This unknown material probably led to misleading characteristics of the material on the reconstruction and generated AC errors in ZTE-AC33. Finally, we observed a recurrent tumor sticking to the meninges and scalp, which may cause tumor delineation difficulties and AC errors. Further studies are warranted to better evaluate possible confounding factors in [18F]-FBPA-PET treatment planning using PET/MR.
Conclusion
These findings suggest that the routinely quantitative use of ZTE-AC on [18F]-FBPA-PET/MR for BNCT therapeutic planning of brain tumors might provide a closer result to the standard PET/CT AC.
Data availability
The datasets generated and/or analyzed during the current study are available from the corresponding author on reasonable request.
References
Barth, R. F., Soloway, A. H. & Fairchild, R. G. Boron neutron capture therapy for cancer. Sci. Am. 263, 100–106 (1990).
Coderre, J. A. & Morris, G. M. The radiation biology of boron neutron capture therapy. Radiat. Res. 151, 1–18 (1999).
Miyatake, S. I. et al. Boron neutron capture therapy for malignant tumors related to meningiomas. Neurosurgery 61, 82–90 (2007).
Isohashi, K. et al. Comparison of the image-derived radioactivity and blood-sample radioactivity for estimating the clinical indicators of the efficacy of boron neutron capture therapy (BNCT): 4-borono-2–18F-fluoro-phenylalanine (FBPA) PET study. EJNMMI Res. 6, 1–8 (2016).
Chen, Y.W. et al. Salvage Boron Neutron Capture Therapy for Malignant Brain Tumor Patients in Compliance with Emergency and Compassionate Use: Evaluation of 34 Cases in Taiwan. Biology (Basel). 10(4), 334. https://doi.org/10.3390/biology10040334 (2021).
Lan, T.L. et al. Advances in Boron Neutron Capture Therapy (BNCT) for Recurrent Intracranial Meningioma. Int J Mol Sci. 24(5), 4978. https://doi.org/10.3390/ijms24054978 (2023).
Fukumitsu, N. & Matsumoto, Y. Development of an Imaging Technique for Boron Neutron Capture Therapy. Cells. 10(8), 2135. https://doi.org/10.3390/cells10082135 (2021).
Hanaoka, K. et al. FBPA PET in boron neutron capture therapy for cancer: prediction of (10)B concentration in the tumor and normal tissue in a rat xenograft model. EJNMMI Res. 4(1), 70. https://doi.org/10.1186/s13550-014-0070-2 (2014).
Watanabe, T. et al. Comparison of the pharmacokinetics between L-BPA and L-FBPA using the same administration dose and protocol: a validation study for the theranostic approach using [18F]-L-FBPA positron emission tomography in boron neutron capture therapy. BMC Cancer. 16(1), 859. https://doi.org/10.1186/s12885-016-2913-x (2016).
Lo, Y. W. et al. The importance of optimal ROIs delineation for FBPA-PET before BNCT. Appl. Radiat. Isot. 163, 109219 (2020).
Paulus, D. H., Tellmann, L. & Quick, H. H. Towards improved hardware component attenuation correction in PET/MR hybrid imaging. Phys. Med. Biol. 58, 8021–8040 (2013).
Eldib, M. et al. Attenuation correction for flexible magnetic resonance coils in combined magnetic resonance/positron emission tomography imaging. Invest. Radiol. 49, 63–69 (2014).
Mehranian, A., Arabi, H. & Zaidi, H. Quantitative analysis of MRI-guided attenuation correction techniques in time-of-flight brain PET/MRI. Neuroimage 130, 123–133 (2016).
Yang, J. et al. Evaluation of sinus/edge-corrected zero-echo-time–Based attenuation correction in brain PET/MRI. J. Nuclear Med. 58, 1873–1879 (2017).
Seith, F. et al. Comparison of positron emission tomography quantification using magnetic resonance- and computed tomography-based attenuation correction in physiological tissues and lesions. Invest. Radiol. 51, 66–71 (2016).
Wiesinger, F. et al. Zero TE MR bone imaging in the head. Magn. Reson. Med. 75, 107–114 (2016).
Khalifé, M. et al. Subject-specific bone attenuation correction for brain PET/MR: Can ZTE-MRI substitute CT scan accurately?. Phys. Med. Biol. 62, 7814–7832 (2017).
Sekine, T. et al. Clinical evaluation of zero-echo-time attenuation correction for brain 18F-FDG PET/MRI: Comparison with atlas attenuation correction. J. Nuclear Med. 57, 1927–1932 (2016).
Okazawa, H. et al. No significant difference found in PET/MRI CBF values reconstructed with CT-atlas-based and ZTE MR attenuation correction. EJNMMI Res. 9, 1–9 (2019).
Sekine, T. et al. Multi-atlas-based attenuation correction for brain 18F-FDG PET imaging using a time-of-flight PET/MR scanner: Comparison with clinical single-atlas-and CT-based attenuation correction. J. Nuclear Med. 57, 1258–1264 (2016).
Delso, G., Kemp, B., Kaushik, S., Wiesinger, F. & Sekine, T. Improving PET/MR brain quantitation with template-enhanced ZTE. Neuroimage 181, 403–413 (2018).
Blanc-Durand, P. et al. Attenuation correction using 3D deep convolutional neural network for brain 18FFDG PET/MR: Comparison with Atlas, ZTE and CT based attenuation correction. PLoS One 14, 1–12 (2019).
Tzourio-Mazoyer, N. et al. Automated anatomical labeling of activations in SPM using a macroscopic anatomical parcellation of the MNI MRI single-subject brain. Neuroimage 15, 273–289 (2002).
Blodgett, T. M. et al. PET/CT artifacts. Clin. Imaging 35, 49–63 (2011).
Decazes, P. et al. Trimodality PET/CT/MRI and radiotherapy: A mini-review. Front. Oncol. 10, 1–9 (2021).
Martinez-Moller, A. et al. Tissue classification as a potential approach for attenuation correction in whole-body PET/MRI: Evaluation with PET/CT data. J. Nuclear Med. 50, 520–526 (2009).
Wollenweber, S. D. et al. Evaluation of an atlas-based PET head attenuation correction using PET/CT & MR patient data. IEEE Trans. Nuclear Sci. 60, 3383–3390 (2013).
Burgos, N. et al. Attenuation correction synthesis for hybrid PET-MR scanners: Application to brain studies. IEEE Trans. Med. Imaging 33, 2332–2341 (2014).
Koesters, T. et al. Dixon sequence with superimposed model-based bone compartment provides highly accurate PET/MR attenuation correction of the brain. J. Nuclear Med. 57, 918–924 (2016).
Cabello, J. et al. Comparison between MRI-based attenuation correction methods for brain PET in dementia patients. Eur. J. Nuclear Med. Mol. Imaging 43, 2190–2200 (2016).
Keereman, V. et al. MRI-based attenuation correction for PET/MRI using ultrashort echo time sequences. J. Nuclear Med. 51, 812–818 (2010).
Cabello, J. et al. MR-based attenuation correction using ultrashort-echo-time pulse sequences in dementia patients. J. Nuclear Med. 56, 423–429 (2015).
Rezaei, A. et al. A quantitative evaluation of joint activity and attenuation reconstruction in TOF PET/MR brain imaging. J. Nuclear Med. 60, 1649–1655 (2019).
Acknowledgements
The authors appreciated the support from Melissa Lee Cancer Foundation.
Funding
This study received support from the grant provided by Melissa Lee Cancer Foundation.
Author information
Authors and Affiliations
Contributions
Y.-W.L., K.-H.L., and W.-S.H. are devoted to conceptualization. Y.-W.L. takes full responsibility for paper writing and part of paper editing. K.-H.L. is in charge of the experiment supervision, paper editing, and revising. C.-Y.L., Y.-W.C., L.-W.W., and Y.-H.W. have participated in data acquisition in clinics and scanner rooms. C.-W.L. and C.-Y.L. are dedicated to data statistics. K.-H.L. and W.-S.H. (wshuang01@gmail.com) contribute equally as corresponding authors.
Corresponding authors
Ethics declarations
Competing interests
CW Li and CY Lin are MR scientists at GE Healthcare. The rest of the authors declare no conflicts of interest in this study. Also, the authors have no personal, financial, or institutional interest in any drugs, materials, or devices in this study.
Additional information
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Lo, YW., Lin, KH., Lee, CY. et al. The impact of ZTE-based MR attenuation correction compared to CT-AC in 18F-FBPA PET before boron neutron capture therapy. Sci Rep 14, 13950 (2024). https://doi.org/10.1038/s41598-024-63248-9
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-024-63248-9
- Springer Nature Limited