Introduction

Surgical correction of adult spinal deformity (ASD) often requires a multilevel fusion ranging from the proximal level of L2 or higher to the pelvis. If fusion is not successful, substantial biomechanical loading, particularly at the proximal or distal ends, can result in rod fracture (RF), screw pullout, or other failure. Literature reports that RF occurs in 6.8–9.3% of ASD patients, and for those who underwent pedicle subtraction osteotomy (PSO), the total incidence of RF was as high as 22% [1, 2]. In the case of lumbopelvic fixation after long construct fusion for ASD, Cho et al. [3] reported a 34.3% incidence of overall failure, with 11.9% of patients experiencing clinically significant major failure requiring a revision. As there was no consensus on the definition of lumbopelvic failure in the primary author’s previous work, failure was defined as either major or minor failure [3]. Major failure included rod breakage between L4 and S1, failure of S1 screws (breakage, halo formation, or pullout), or prominent iliac screws which required revision surgery. Minor failure included rod breakage between S1 and iliac screws, or failure of iliac screws excluding prominence of iliac screws and did not require revision surgery.

Rod failure may be attributed to many factors such as: (1) rod material and diameter [2, 4]; (2) rod contouring techniques and achieved curvature [2, 4]; (3) patient pre-operative sagittal alignment, such as pelvic tilt (PT), sacral slope (SS), pelvic incidence (PI), and sagittal vertical axis (SVA) [2, 3, 5]; and (4) surgical technique-related factors such as choice of osteotomy or choice of distal foundation [2,3,4]. Although there is no agreement among researchers on which factor had the most impact on rod fractures, it is well known that sagittal alignment is a key factor in reducing abnormal loading on the spine, lessening the risk of RF. Sagittal alignment is usually assessed by a plumb line method drawn on full-length radiographs. Other methods, such as T1 and T9 spinopelvic inclination [5] or cranial sagittal vertical axis [6, 7] also have been used. Many studies have investigated the association between sagittal alignment and clinical outcomes [8,9,10]. There is overall agreement among those retrospective studies that positive sagittal alignment (SVA > 0) is associated with pain and disability. Emami et al. [11] demonstrated that patients with poor sagittal balance after long fusions to the sacrum had increased pain. Glassman et al. [8] reported that patients with positive sagittal imbalance expressed worse self-assessment in the SRS-22 questionnaire’s pain, function, and self-image subscores. In the past, force-plate analysis has been used to capture the spinal and pelvic offsets from the standing foot position and to help evaluate sagittal imbalance [12]. The axis of the femoral shaft and the plumb line has also been evaluated for the imbalance prior to surgical correction [13].

Even though major failure of lumbopelvic constructs occurred more often in patients with suboptimal lumbar lordosis and sagittal balance, there has been no biomechanical validation of this classification. Questions remain regarding mechanical risk factors, or if there is any relationship between implant type, spinopelvic parameters, and failure to achieve fusion. The purpose of this study is to biomechanically validate the classification of lumbopelvic fixation failure using an in silico model.

Materials and methods

The development of the finite element models involves three steps. In the first step, an L4–L5 FSU (functional spinal unit) finite element model was first developed and validated. The spine geometry was modified from a CAD model (Digimation, Inc., Lake Mary, FL, USA). Intervertebral disc was simplified as flat endplate. The cancellous core, posterior elements of the vertebrae, the annulus fibrosus, and the nucleus were modeled as three-dimensional isotropic four node tetrahedral solid elements (C3D4). Thin shell elements (S3R) with thicknesses of 0.5 mm and 0.3 mm were used to model the cortical shell and endplate, respectively. Material properties were obtained from several references and are listed in Table 1. The annulus fibrosus was modeled by a hyperelastic constitutive law for the ground substance and by nonlinear springs oriented at ± 30° to the horizontal for the annulus fibers [14]. Coefficients of the third-order Ogden hyperelastic formulation were determined from experimental data and used [15]. Facet cartilage was modeled as shell elements with a thickness of 0.3 mm [16]. Sliding interaction between facets was modeled as a frictionless contact. Six spinal ligaments were represented by three-dimensional unidirectional spring elements. A stepwise calibration was performed in the same sequence as reported by Schmidt et al. [17]. Material properties of each added component were modified to match the corresponding in vitro data point.

Table 1 Element type and material properties of the vertebrae used in the finite element model of the motion segment

In the second step, the validated L4–L5 FSU finite element model was then extended to T12–pelvis by adding more vertebrae and the pelvis. The sacroiliac joint was modeled as articular cartilage contacts surrounded by six types of strong ligaments (anterior sacroiliac, interosseous sacroiliac, long posterior sacroiliac, short posterior sacroiliac, sacrospinous, sacrotuberous) [18], depicted as spring elements. The range of motion of the intact T12–pelvis spine unit was then validated against an in-house cadaveric test of six specimens. A maximum moment of ± 10 Nm was applied in flexion/extension, lateral bending, and axial rotation. The motion captured was comparable to the experimental measurements (Fig. 1).

Fig. 1
figure 1

Comparison of range of motion of the intact spine unit between the in vitro biomechanical test and simulation result of finite element model under 10 Nm moment in flexion/extension (FE), lateral bending (LB), and axial rotation (AR)

In the last step, final T10–pelvis model was created to match the average spinal-pelvic parameters (PT, SS, and lumbar lordosis) of two cohorts of patients reported in the literature [3], major failure and non-failure groups (Fig. 2). The non-failure cohort had a mean post-operative SVA of 49.6 mm, placing it into the sagittal-forward alignment group (0–50 mm), and a mean gravity line (GL) offset from the heels of 110 mm [12]. The major failure group had a post-operative SVA of 70.3 mm, placing it into the sagittal-forward alignment group (> 50 mm), and a similar GL offset as the non-failure group. As shown in Fig. 2, the distance from the center of the hip joint to the center of the superior endplate of T10 was 57.7 mm and 59.0 mm in the non-failure and major failure groups in the sagittal plane, respectively. For the sagittal-forward group to maintain balance of the spinopelvic axis, pelvic retroversion (posterior shifting of the pelvis toward the heels to maintain a balanced mass distribution) has to occur [5]. In our model, we assumed that the pelvis was shifted toward the heel by 10 mm for the major failure group, which is quite conservative compared to the 30 mm pelvic shift for the sagittal-forward group observed in the literature [5]. If we assume the average human weight is 75 kg and the upper body weight above T10 is about 40% of the whole body weight (~ 300 N), then the moment created by the upper body on the spine is this weight multiplied by the distance (D) between the GL and the center of the T10 superior endplate (moment = 300 N × D). Since the pelvis was shifted toward the heel by 10 mm in the major failure group, the moment arm was 69.0 mm compared to 57.7 mm in the non-failure group (Fig. 1). As the result, the non-failure group underwent 17.3 Nm moment loading while the major failure group underwent 20.7 Nm moment loading. The corresponding moment was applied to the superior endplate of T10 to simulate the static gravimetric loading in a standing position. No cyclic loading was simulated.

Fig. 2
figure 2

Finite element models for major failure and non-failure cohorts; values listed in the table are extracted from the retrospective study conducted by Cho et al. [3]. PT = the angle between the vertical and the line through the midpoint of the sacral plate to femoral heads axis; SS = the angle between the horizontal and the sacral plate; SVA = the offset from C7 plumb line and the posterosuperior corner of S1

Results

Despite a fixed GL position relative to the heels, differences in spinopelvic parameters resulted in a neutral sagittal alignment in the non-failure spine model, but produced a sagittal-forward alignment in the major failure model. As a result, the bending moment was approximately 17.3 Nm in the non-failure group and 20.7 Nm in the major failure group, representing a 20% increase. Differences in loads produced 14 mm of translation and 4.9° of rotation for the major failure group—18% and 14% higher, respectively, compared to the non-failure group (11.9 mm translation and 4.3° rotation) (Fig. 3). For comparison purposes, the non-failure model was loaded with the same amount of moment applied to the major failure model (20.7 Nm) to isolate the effect of loading and alignment independently. The achieved translation and rotation were 13.2 mm and 4.7°, respectively—6% and 4% lower than the major failure group, respectively.

Fig. 3
figure 3

Load and displacement comparison between major failure and non-failure groups

Rod stresses were highest at L1–L2 and L4–L5 in both cases. In the major failure group, maximum stress (138.3 MPa) was observed between L4 and L5. In the non-failure group, maximum stress (115.4 MPa) was at the surface between L1 and L2 (Fig. 4). High stress (141.0 MPa) was also observed in S1 screws in the major failure group. This value was 42% greater than the stress (98.9 MPa) observed in the non-failure group.

Fig. 4
figure 4

Rod and screw stress contours under the gravitational loading; values shown in the figure are average surface von Mises stresses. Units are MPa

As with stress, high strain was also observed at L1–L2 and L4–L5 in both models (Fig. 5). The maximum strain was detected on the right rod surface between L4 and L5 in the major failure group and on the left rod surface between L4 and L5 in the non-failure group (0.13% and 0.11%, respectively). The average maximum rod surface strain was about 22% higher in the major failure group in comparison to the non-failure group. Additionally, noticeably high strain in the S1 screw was observed. Results were 0.13% for the major failure group—42% higher than the S1 screw strain (0.09%) observed in the non-failure group.

Fig. 5
figure 5

Rod and screw strain contours under the gravitational loading; values shown are average surface strains (in percentage format)

Discussion

In the current study, an in silico model was used to validate the major failure and non-failure classifications of lumbopelvic fixation described by Cho et al. [3]. Patients with major failure risk factors—including larger PI, decreased lumbar lordosis (LL), and much higher SVA—demonstrated a greater likelihood for sagittal imbalance when compared to the non-failure group. From a biomechanical point of view, the major failure group observed a 20% increase in loading due to the fixed gravity line, which led to a 14% and 18% increase in rotation and translation, as well as a 20% increase in rod stress/strain when compared to the non-failure group. For the major failure group to maintain balance, pelvic retroversion/shift had to occur [5]. Pelvic shift (in relation to the feet) is an important component in maintaining a somewhat fixed GL-heels offset, even in the setting of variable SVA and trunk inclination [12]. This mechanism uses muscle force to improve alignment, and thereby restore the GL position and horizontal gaze [19]. In this case, the pelvis shift was modeled at 10 mm. However, clinically, sagittal-forward patients with pelvic shifts greater than 30 mm have been observed [5], which would have increased the gravimetric load by up to 54%. In this worst-case scenario, dramatic increases in rotation, translation, and stress concentration on the rod and screw would be expected. Generally speaking, the more pelvic retroversion, the greater the shift of the pelvis toward the heel, resulting in a greater moment arm between the GL and the lumbopelvic spine, increased muscle activity, and greater load on the construct.

Assessing SVA alone may underestimate the severity of the deformity [20]. Schwab et al. [20] reported that the combination of a PI/LL mismatch (PI minus LL) and SVA can better predict patient disability and provide a guide for appropriate therapeutic decision-making. The difference in PI/LL mismatch between the major failure and non-failure groups was significant (22° vs. 9.3°) (Fig. 1) [3]. As shown in Fig. 3, even when the applied moment is the same, the major failure group observes an increase in rotation and translation. This indicates that the PI/LL mismatch also contributes to the biomechanical stability of the lumbopelvic spine. Surgical intervention to increase LL often involves an osteotomy. One of the most popular osteotomy procedures, PSO, can achieve greater correction at a single segment, with an average correction of 30°–40° [21]. Numerous articles have discussed the benefits and disadvantages of this technique [1, 2, 22]. From a biomechanical point of view, a PSO can improve sagittal balance, thereby decreasing the amount of loading on the spine. However, sharp bending at the apex level—as well as the anatomic nature of the apex relative to gravitational loading—increases the mechanical demand on the implants. Consequently, most RFs are found at the apex in PSO patients [1].

The current study observed high stress/strain concentration at the S1 screws for the major failure cohort. High strain on the S1 screws may result in complications such as screw pullout or pseudarthrosis at the lumbosacral junction for long-segment cases [23]. This result is also consistent with the high rate of S1 screw failure found in the major failure group [3]. In this case, the fixation distal level was at the pelvic region. The addition of iliac fixation screws serves as temporary scaffolding to allow for the maturation of bony fusion across the lumbosacral junction, which may lower the incidence of screw pull out, L5–S1 pseudarthrosis, and sacral insufficiency fracture at S1–2 [23]. One would expect high load demand on S1 when distal fusion ends at the sacrum [11, 24, 25].

There are some limitations of current model. First, rod contouring was not modeled. Studies have shown that excessive rod bending (e.g., at L5/S1 or at PSO level) are correlated to the high rate of rod fractures [1, 2]. Unfortunately, modeling of rod contouring is extremely difficult in finite element analysis; therefore, it was not included in current study. Second, due to the complexity involved, as well as to a lack of information, muscle forces are neglected in the current model. Very few studies have included muscle force into their models but these studies were still at the earlier stages of incorporating of muscle forces [26, 27]. It was extremely difficult to incorporate muscle force into already complicated models to study the long spinal fusion. However, with the more information and studies of muscle force available to the researchers, we expect that a more robust finite element model can be developed and studied.

This study only looked at the effects of sagittal alignment on the biomechanics of lumbopelvic fixation. This model can be easily modified to look at the effect of posterior bony fixation on the overall mechanical demands on the implants, the effect of anterior column support, rigidity of posterior or posterolateral fusion mass, the choice of distal fixation level, the effect of rod material or diameter, and the effect of surgical techniques (e.g., PSO, bilateral vs. unilateral). All of these studies would be out of the scope of the focus of the current study, and will be addressed in future analyses.

Conclusion

Due to compensatory differences in the alignment of spinopelvic parameters between normal and failed spines in the presence of a fixed GL, the major failure cohort in this study observed a 20% higher load and 18% greater instability. The higher load and instability increased mechanical demand on the posterior rods in the lower lumbar spine, further emphasizing the importance of proper sagittal alignment, while validating the classification of lumbopelvic fixation failure.