Abstract
Previously developed robotic colonoscopes have shown low locomotion performance since they have not had reliable clamping modules or a long stroke. Therefore, a clamping module consisting of six legs and two triggers is installed on the front and rear of the robot, enabling the robot to clamp effectively onto the colon. Locomotion capability is then improved via the on-time folding and unfolding of the legs according to the elongation and contraction of the bellows of the robot body. In addition, one pneumatic line based locomotive mechanism, which had been developed previously for in-pipe inspection, is adopted to reduce the friction force between the pneumatic lines and the locomotion environment. In order to evaluate locomotion performance, the robot and robot control system are constructed and are tested in an acrylic pipe and under condition of in-vitro. The robot can move regardless of slope in the case that the locomotion path has more than 25 mm of radius. In straight and vertical paths, it travels 33 mm/s and 12.1 mm/s, respectively. Next, a locomotion test under in-vivo condition is carried out with the optimized configuration and maneuvering parameters. Conclusively, a pneumatic bellows based locomotive mechanism with a dependable clamping module shows the reliable locomotion performance of about 8.5 mm/sec under in-vivo condition.
Article PDF
Similar content being viewed by others
Avoid common mistakes on your manuscript.
References
Kim, K. D., Lim, H., Kim, B. K., Park, J. O., and Hong, Y. S., “A Locomotive Mechanism for Colonoscope,” Transaction of the KSME, Vol. 26, No. 7, pp. 1296–1301, 2002.
Kim, K. D., Lee, S. J., Kim, B. K., and Park, J. O., “Radial Type Locomotive Mechanism with Worm for Robotic Endoscope,” Transaction of the ICASE, Vol. 8, No. 3, pp. 220–225, 2002.
Kim, B. K., Lee, J. H., Lim, Y. M., Park, J. O., Kim, S. H., and Hong, Y. S., “Locomotive Colonoscope,” Proceedings of the 32nd International Symposium on Robotics, pp. 1829–1833, 2001.
Menciassi, A., Park, J. H., Lee, S., Gorini, S., Dario, P., and Park, J. O., “Robotic Solutions and Mechanisms for a Semi-Autonomous Endoscope,” Proceedings of IEEE/RSJ International Conference on Intelligent Robots and Systems, pp. 1379–1384, 2002.
Park, S. H., Park, H. J., Park, S. J., and Kim, B. K., “A Paddling Based Locomotive Mechanism for Capsule Endoscopes,” Journal of Mechanical Science and Technology, Vol. 20, No. 7, pp. 1012–1018, 2006.
Kim, B. K., Lim, H. Y., Park, J. H., and Park, J. O., “Inchworm-Like Colonoscopic Robot with Hollow Body and Steering Device,” JSME Int. J., Vol. 49, No. 1, pp. 205–212, 2006.
Lee, J. S., Kim, B. K., and Hong, Y. S., “A Flexible Chain-based Screw Propeller for Capsule Endoscopes,” Int. J. Precis. Eng. Manuf., Vol. 10, No. 4, pp. 27–34, 2009.
Hong, Y. S., Kim, J. Y., Kwon, Y. C., and Song, S. Y., “Preliminary Study of a Twistable Thread Module on a Capsule Endoscope in a Spiral Motion,” Int. J. Precis. Eng. Manuf., Vol. 12, No. 3, pp. 461–468, 2011.
Hoeg, H. D., Slatkin, A. B., and Burdick, J. W., “Biomechanical modeling for the small intestine as required for the design and operation of a robotic endoscope,” IEEE International Conference on Robotics and Automation, pp. 1599–1606, 2000.
Valdastri, P., Webster, R. J., Quaglia, C., Quirini, M., Menciassi, A., and Dario, P., “A New Mechanism for Mesoscale Legged Locomotion in Compliant Tubular Environments,” IEEE Transactions Robotics, Vol. 25, No. 5, pp. 1047–1057, 2009.
Hu, H., Wang, P., Zhao, B., Li, M., and Sun, L., “Design of a novel snake-like robotic colonoscope,” Proceedings of the IEEE International Conference on Robotics and Biomimetics, pp. 1957–1961, 2009.
Peirs, J., Reynaerts, D., and Van Brussel, H., “Design of miniature parallel manipulators for integration in a selfpropelling endoscope,” Sensors and Actuators, Vol. 85, pp. 409–417, 2000.
Dario, P., Carroza, M. C., Lencioni, L., Magnani, B., and D’Attansio, S., “A Micro-robotics System for Colonoscopy,” Proceedings of the IEEE International Conference on Robotics and Automation, pp. 1567–1572, 1997.
Ozaki, K., Wakimoto, S., and Suzumori, K., “Novel design of rubber tube actuator improving mountability and drivability for assisting colonosocope insertion,” Proceedings of the IEEE International Conference on Robotics and Automation, pp. 3263–3268, 2011.
Trovato, G., Shikanai, M., Ukawa, G., Kinoshita, J., Murai, N., Lee, J. W., Ishii, H., Takanishi, A., Tanoue, K., and Ieiri, S., “Development of a colon endoscope robot that adjusts its locomotion through the use of reinforcement learning,” International Journal of Computer Assisted Radiology and Surgery, Vol. 5, No. 4, pp. 317–325, 2010.
Phee, L., Accoto, D., Menciassi, A., Stefanini, C., Carrozza, M. C., and Dario, P., “Analysis and development of locomotion devices for the gastrointestinal tract,” IEEE Transactions on Biomedical Engineering, Vol. 49, No. 6, pp. 613–616, 2002.
Lim, J. W., Park, H. J., An, J. M., Hong, Y. S., Kim, B. K., and Yi, B. J., “One pneumatic line based inchworm-like micro robot for half-inch pipe inspection,” Mechatronics, Vol. 18, pp. 315–322, 2008.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Park, Hj., Kim, D. & Kim, B. A robotic colonoscope with long stroke and reliable leg clamping. Int. J. Precis. Eng. Manuf. 13, 1461–1466 (2012). https://doi.org/10.1007/s12541-012-0192-2
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s12541-012-0192-2