Abstract
Background
Apnea test is a key component to confirm brain death. For patients receiving extracorporeal membrane oxygenation (ECMO), apnea test remains challenging. Brain death (BD) diagnosis is often made without apnea test.
Case
We report the case of a 29-year-old man presenting clinical signs of BD while treated with ECMO therapy for refractory cardiogenic shock. Decreasing the ECMO sweep gas flow from 3 to 1 L/min and increasing oxygen delivery to 100 % on ECMO during the apnea test have allowed increasing the PaCO2 of more than 20 mmHg without decreasing PaO2.
Discussion
In order to diagnose BD, neurological examination should be complete, including apnea testing, which can be not possible in patients receiving ECMO due to CO2 removal from the membrane. Decreasing sweep gas rate allows reduction in CO2 diffusion through the membrane. However, decreasing the ECMO gas flow to zero could be insufficient to maintain normoxemia. Decreasing (but not stopping) the sweep gas flow to 1 L/min and increasing the oxygen delivery through the ECMO have allowed performing the apnea test safely.
Conclusion
To assess brain death in patients on ECMO, apnea test can be performed without compromising oxygenation by decreasing (but not stopping) the sweep gas flow and increasing oxygen delivery through the membrane.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
Extracorporeal membrane oxygenation (ECMO) is a bedside modified heart–lung machine, used to support acute, severe and reversible cardiac and/or pulmonary failure. It has been successfully used in various causes of cardiac failure [1]. Nevertheless, ECMO is associated with a high rate of deaths and neurological injuries [2]. Thiagarajan et al. [3] reported an incidence of 21 % of brain death (BD) associated with ECMO therapy. Due to carbon dioxide (CO2) removal from the membrane, BD diagnosis, and especially apnea test, remains challenging in patients receiving ECMO [4].
Case
A 29-year-old male has been admitted in our hospital for chest pain with signs of acute non-ST-elevation myocardial infarction. The echocardiogram revealed a severe aortic stenosis linked to a bicuspid valve. The patient experienced a third degree atrioventricular block followed by a ventricular fibrillation, treated with electrical cardioversion and epinephrine for a total of 50 min of cardiopulmonary resuscitation. A veno-arterial ECMO was placed to assist the refractory cardiogenic shock, and the aortic valve was surgically replaced in emergency. The patient was afterward admitted in the intensive care unit under sedation (propofol 80 mg/h) and mechanical ventilation.
On day one after surgery, a bilateral mydriasis associated with Glasgow coma score at 3 and diabetes insipidus appeared. Propofol was stopped for neurological assessment. The neurological examination found a coma with absence of brainstem response. Brain CT scan was not possible due to his clinical condition. A first electroencephalogram was performed revealing persistent low-voltage frontal activity. A repeat neurological examination was performed 6 h later. The patient was still unresponsive with absence of brainstem responses.
In order to assess brain death diagnosis, an apnea test was performed. Before starting the apnea test, body temperature was 36.5 °C, mean arterial pressure (MAP) 70 mmHg, heart rate 100 bpm with norepinephrine infusion at 2.2 mg/h and epinephrine at 0.4 mg/h. ECMO parameters were ECMO blood flow = 4.1 L/min; inspired fraction of O2 (FiO2) = 40 %; sweep gas flow = 3 L/min. Mechanical ventilation parameters were tidal volume = 580 mL, respiratory rate = 16, FiO2 = 21 % and positive end expiratory pressure = 6 cmH2O. FiO2 was increased to 100 % on ECMO, whereas FiO2 was kept at 21 % on the ventilator. After 10 min of preoxygenation, on initial arterial blood gas analysis, arterial partial pressure of oxygen (PaO2) was 256 mmHg, arterial partial pressure of carbon dioxide (PaCO2) 47 mmHg and pH 7.43. Mechanical ventilation was interrupted, and a T-piece oxygen delivery at 9 L/min was connected to the tracheal tube. In the same time, sweep gas flow was decreased from 3 L/min to 1 L/min. During this 10 min apnea test, respiratory movements were absent, hemodynamic parameters were stable and catecholamine infusions remained unchanged; after 10 min of apnea test, arterial blood gas revealed a PaO2 of 353 mmHg, PaCO2 of 67 mmHg and pH of 7.28. The apnea test was considered as positive, and the mechanical ventilation was restarted. BD was confirmed by two electroencephalograms as required by the French law, but the organ donation was refused by the family.
Discussion
Clinical assessment of brain death is made on several criteria: irreversible coma with a known cause, loss of brainstem reflexes and positive apnea testing [5]. Prerequisites include the absence of cerebral nervous system-depressant drugs or neuromuscular blocking agents; no severe electrolyte, acid–base or endocrine disturbances; achievement of a normal core temperature (>36 °C) and a normal systolic blood pressure (>100 mmHg) [6]. In France, a confirmatory test must afterward be performed. Legal ancillary tests include cerebral angiography, computed tomography angiography or two electroencephalograms performed with a 4 h interval [7]. Nevertheless, BD is first a clinical diagnosis, and apnea testing should be performed.
In veno-arterial ECMO, blood is drained from a venous cannula placed near the right atrium and returned to a femoral arterial cannula. The external circuit carries venous blood to the membrane oxygenator, an artificial lung, to increase O2 and remove CO2, before reinjection. CO2 clearance depends on the rate of sweep gas flow through the oxygenator, on the ECMO blood flow, on the physical properties of the oxygenator (maximum CO2 exchange rate) and on the presence of CO2 in the inspired gas. In case of low sweep gas flow, if the other parameters are stable, CO2 removal is mainly dependent of the rate of sweep gas flow through the oxygenator [8].
Various techniques have been used for apnea tests [6]. We used the apneic oxygenation–diffusion with a T-piece oxygen delivery at 9 L/min, with respect of prerequisites (normotension, normothermia, euvolemia and absence of hypoxia) [6]. A moderate hypercapnia was present at baseline, which could have had been corrected by increasing sweep gas flow on ECMO (for an ECMO blood flow of 4 L/min, a sweep gas flow of 4 L/min should have been better to obtain a normocapnia). Apneic oxygenation–diffusion has been proven to be safe [9], even if respiratory acidosis and hypercarbia linked to apnea testing may be associated with cardiac arrhythmias or hypotension [5].
Apnea testing is considered positive if respiratory movements are absent, and PaCO2 is above or equal to 60 mmHg or increases for at least 20 mmHg [6]. In order to limit CO2 removal from the ECMO, the sweep gas flow has been decreased from 3 to 1 L/min. Decreasing sweep gas rate allows reduction in CO2 diffusion through the membrane. If the sweep gas rate is maintained at the same level, the interruption of mechanical ventilation is not sufficient to provide an increase in PaCO2. Yang et al. [11] proposed to turn the ECMO gas flow to zero during the apnea test. However, it could be insufficient to maintain normoxemia [4]. The use of a continuous positive airway pressure (CPAP) has been shown to be prevent hypoxemia [10]; In our case, the observed increase in PaO2 during apnea is probably related to the increased oxygen delivery through the ECMO, but also the result of maintaining (and not stopping) a sweep gas flow. We selected 1 L/min of sweep gas flow during apnea testing as an estimation of the best level between blood oxygenation and CO2 clearance for this patient.
In case of hypothermia, a lower sweep gas flow would probably be required to allow CO2 to rise because of reduced body metabolism and CO2 production. In the same way, hypotension would probably require a reduced gas flow. Morphologic parameters and gender also influence basal metabolism rate and CO2 production, but probably without enough differences to affect the required sweep gas flow.
Conclusion
Assessment of BD, especially apnea test, remains challenging in patients receiving ECMO. Here, we describe a simple and safe method of apnea testing in a patient on ECMO. Decreasing the sweep gas flow to 1 L/min and increasing oxygen delivery to 100 % on ECMO during the apnea test have allowed to significantly increase the PaCO2, and to conclude to a positive apnea test, without compromising oxygenation.
References
Aissaoui N, Luyt CE, Leprince P, et al. Predictors of successful extracorporeal membrane oxygenation weaning (ECMO) after assistance for refractory cardiogenic shock. Intensive Care Med. 2011;37:1738–45.
Mateen FJ, Muralidharan R, Shinohara RT, Parisi JE, Schears GJ, Wijdicks EF. Neurological injury in adults treated with extracorporeal membrane oxygenation. Arch Neurol. 2011;68:1543–9.
Thiagarajan RR, Brogan TV, Scheurer MA, Laussen PC, Rycus PT, Bratton SL. Extracorporeal membrane oxygenation to support cardiopulmonary resuscitation in adults. Ann Thorac Surg. 2009;87:778–85.
Muralidharan R, Mateen FJ, Shinohara RT, Schears GJ, Wijdicks EF. The challenges with brain death determination in adult patients on extracorporeal membrane oxygenation. Neurocrit Care. 2011;14:423–6.
Wijdicks EF. Determining brain death in adults. Neurology. 1995;45:1003–11.
Wijdicks EF, Varelas PN, Gronseth GS, Greer DM. Evidence-based guideline update: determining brain death in adults: report of the Quality Standards Subcommittee of the American Academy of Neurology. Neurology. 2010;74:1911–8.
Boulard G, Guiot P, Pottecher T, Tenaillon A. Management of subjects in a state of brain death and the preservation of organs. Ann Fr Anesth Reanim. 2005;24:836–43.
Hout MS, Hattler BG, Federspiel WJ. Validation of a model for flow-dependent carbon dioxide exchange in artificial lungs. Artif Organs. 2000;24:114–8.
Wijdicks EF, Rabinstein AA, Manno EM, Atkinson JD. Pronouncing brain death: contemporary practice and safety of the apnea test. Neurology. 2008;71:1240–4.
Levesque S, Lessard MR, Nicole PC, et al. Efficacy of a T-piece system and a continuous positive airway pressure system for apnea testing in the diagnosis of brain death. Crit Care Med. 2006;34:2213–6.
Yang HY, Lin CY, Tsai YT, Lee CY, Tsai CS. Experience of heart transplantation from hemodynamically unstable brain-dead donors with extracorporeal support. Clin Transplant. 2012;26:792–6.
Conflict of interests
The authors have not disclosed any potential conflicts of interests.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Smilevitch, P., Lonjaret, L., Fourcade, O. et al. Apnea Test for Brain Death Determination in a Patient on Extracorporeal Membrane Oxygenation. Neurocrit Care 19, 215–217 (2013). https://doi.org/10.1007/s12028-013-9845-y
Published:
Issue Date:
DOI: https://doi.org/10.1007/s12028-013-9845-y