Abstract
Background
Uncemented press-fit cups provide bone fixation in primary THA, but the use of screws is sometimes necessary to achieve primary stability of the socket. However, it is unclear whether and when screws should be used.
Question/Purposes
We analyzed the factors related to screw use with a press-fit uncemented cup and assessed whether screw use is associated with the same rates of loosening and revision as a press-fit technique.
Methods
We retrospectively reviewed 248 patients who underwent THA using the same prosthetic design. Eighty-eight hips had screws to achieve primary cup fixation (Group 1), and 189 did not (Group 2). Mean age was 50 years (range, 14–73 years). We analyzed factors related to the patient, acetabular type, and reconstruction of the rotation center of the hip. Minimum followup was 5 years (mean, 8.9 years; range, 5–12 years).
Results
We found higher screw use in women, patients with less physical activity, Acetabular Types A or C, and a distance from the center of the prosthetic femoral head to the normal center of rotation of more than 3 mm. There were four revisions in Group 1 and five in Group 2. Eight hips had radiographic loosening in Group 1 and nine in Group 2. Cups with a postoperative abduction angle of more than 50° had a higher risk for loosening.
Conclusions
Press fit was achieved less frequently in women and patients with Acetabular Types A or C and less physical activity; a closer distance to the normal center of rotation decreased screw use. Screw use to augment fixation achieved survival similar to that of a press-fit cup.
Level of Evidence
Level II, prognostic study. See Instructions for Authors for a complete description of levels of evidence
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
Porous-coated and other hemispherical designs are associated with high rates of stable fixation in primary THA [13, 14], with rates of aseptic loosening at 10 years of followup of less than 5% [3, 12, 42]. Although controversial [9], screws have been recommended for additional fixation [25, 36]. Their disadvantages include the risk of injury to major neurovascular structures [1, 38] or possible access for wear particle-producing osteolysis and the appearance of radiolucent lines around screws [28, 30].
Various factors influence the primary fixation of hemispherical uncemented cups. The geometry of the socket is altered in the arthritic hip due to the degenerative process, which may affect cup fixation [6, 34]. Bone quality and female sex have been postulated as factors influencing primary uncemented fixation [8]. When a hemispherical uncemented cup is used with a press-fit technique without screws, the acetabular bone preparation and loading at the cortical rim of the acetabulum are critical [8, 9, 23]. In a long-term study of 4289 uncemented cups, Engh et al. [9] observed a survivorship of higher than 90% at 15 years for six different designs; the highest survival rates were for press-fit components. Although they did not describe the technical differences between cups with spikes and those with or without screws, they observed a lower survival for the cups with spikes, probably due to an incomplete seating of the component in the acetabulum. Dorr et al. [6] reported the acetabular structure is different in the arthritic hip and this could be related to cup fixation, as well as the depth of reaming and the importance of loading of the cortical rim of the acetabulum. However, it is unclear whether and when screws should be used.
We therefore first analyzed the possible relationship between factors related to patient demographics (age, sex, physical activity, weight, diagnosis), their preoperative radiographic characteristics (acetabular geometry and bone quality), surgical technique according to the reconstruction of the rotation center of the hip on the postoperative radiographs, and the primary fixation using a press-fit technique of an uncemented hemispherical hydroxyapatite (HAP) cup in primary THA. Second, we compared the postoperative pain, function, and ROM and the rates of revision and radiographic loosening of the cups implanted with or without screws to augment primary fixation over a minimum followup of 5 years.
Patients and Methods
We prospectively followed all 257 patients with 289 primary THAs using the same prosthetic design between January 1999 and January 2006. A minimum time of 5 years from the operation (including radiographs) was required for inclusion in this followup study. Nine patients (12 hips) were lost to followup and were excluded. The remaining 277 hips (248 patients) formed the basis of the followup study. The patients were divided into two groups depending on whether screws were used to achieve intraoperative primary fixation of the cup: Group 1, who had screws (89 hips); and Group 2, who did not (189 hips). Mean age was 50.6 years (range, 14–73 years). The minimum clinical and radiographic followup at the last evaluation was 5 years in nonrevised hips (mean, 8.9 years; range, 5–12 years). Oral and written informed consent for participation in the study was obtained from all patients.
We collected patient demographic data in terms of sex, age, weight, activity level, acetabular shape, preoperative diagnosis, and clinical status. The preoperative activity level was classified according to Devane et al. [5]. We used the classification of Dorr et al. [6] to assess acetabulum morphology according to the shape of the radiolucent triangle superior to the acetabulum (Table 1): Type A is an isosceles-triangle-shape seen more frequently in hips with avascular necrosis and inflammatory arthritis; Type B shows extension of the triangle into the teardrop seen more frequently in osteoarthrosis of the hip; and Type C is a right triangle seen more frequently in dysplastic hips. Preoperative radiographs were measured for Types 1, 2, and 3 femoral bone [7] so as to evaluate the influence of osteoporosis on bone fixation.
A Cerafit®-Triradius press-fit cup (Ceraver, Roissy, France), including a fully HAP-coated TiAl6V4 rough shell with optional screws and fitted with an Al2O3 liner, was implanted in every hip. The cup was paired with an uncemented fully HAP-coated, tapered, straight Cerafit®-Multicone stem (Ceraver). According to the manufacturer, the alumina had 99.8% purity, 3.98 density, 2-μm grain size, and was manufactured by hot isostatic pressing in all cases. The HAP coating was sandblasted to increase the implant surface for osseointegration, a relative crystallinity of 37%, a thickness of 80 ± 20 μm, a bond shear strength of 15 to 25 MPa, and a mean roughness of 3 μm; the arithmetic rugosity (Ra) beneath the HAP was 3 to 4 μm [11]. The femoral head size was 28 mm for cups sized 46 or 48 mm in diameter (20 cups) and 32 mm for cups sized 50 mm or greater (257 cups). We use this implant for patients aged 65 years or older and active patients (Activity Level 4 or 5 according to Devane et al. [5]).
Two surgeons (EGC, ACP) performed all operations with the same operative technique using a posterolateral approach. In all patients, we prepared the acetabular bone by removing the central osteophyte and reaming the articular cartilage until subchondral bone started bleeding. We attempted to maintain the subchondral bone [41]. The reaming was made line-to-line due to the oversized cup following the manufacturers′ recommendations. We first tried a press-fit implantation of the cup. The intraoperative stability of fixation was confirmed according to Udomkiat et al. [35] by trying to pull the metal shell out of the acetabulum using the attached insertion tool (pull-out test). Two screws were used if there was any movement after the mentioned technique and/or where autograft was employed.
After surgery, patients walked with crutches with toe-touch partial weightbearing for 3 weeks, after which they were allowed to walk using two crutches for the next 6 weeks. Cefazolin (or vancomycin in allergic patients) was administered for antibiotic prophylaxis during the anesthetic induction and continued for 24 hours, with low-molecular-weight heparin to decrease the risk of thromboembolic disease in all patients for 5 weeks postoperatively.
Patients were followed postoperatively at 6 weeks, 3, 6, and 12 months, and annually thereafter. We evaluated preoperative and postoperative pain, function, and ROM according to the six-level scale described by Merle d’Aubigné and Postel [20]. Patients were also asked about the location of pain when reported. Standard AP radiographs of the pelvis and lateral radiographs of the hip were made preoperatively, immediately after the operation, at 6 weeks, at 3, 6, and 12 months, and annually thereafter following the same protocol. The patient was positioned supine with his/her feet together. The x-ray tube was positioned over the symphysis pubis 1 m from and perpendicular to the table with a symmetric obturator foramen and visible lesser trochanter and iliac crest [19]. A single author (EGR) who had not been involved in the surgery made all measurements, which were repeated three times for every radiograph. Cup position was assessed according to the acetabular abduction angle, the height of the center of the hip (as measured from the center of the femoral head to the interteardrop line), and the horizontal distance of the cup (measured from the center of the femoral head to the Kohler line) [15]. The reconstruction of the hip rotation center was evaluated according to the Ranawat triangle [26]. The true acetabulum region was the area enclosed by a right triangle with a height and width equal to 20% of the height of the pelvis on the AP radiograph. The midpoint of the hypotenuse coincides with the approximate center of the femoral head (AFHC) and is the center of rotation of the hip. The AFHC was used as the reference point to measure distance to the center of the prosthetic femoral head (CPFH). This distance was recorded in the preoperative and postoperative radiographs to assess the reconstruction actually achieved. The distribution of any radiolucent gaps on the initial postoperative radiograph and of radiolucent lines or osteolysis at the acetabular bone-prosthesis interface on the subsequent radiographs was recorded in the three zones described by DeLee and Charnley [4]. The current method of determining radiographic stable bone fixation into an acetabular component is by indirect inference based on the absence of the two classic signs of loosening: radiolucent lines and cup migration [21].
We expressed nominal data as frequencies or percentages and quantitative data as mean ± SD. We compared qualitative data for hips with and without screws using the chi square test or Fisher’s exact test and quantitative data using Student’s t-test. Pearson’s chi square test was used to compare the demographic qualitative data of patients between groups. Univariate and multivariate Cox regression models were used to assess risk factors for the use of screws. We used Kaplan-Meier survivorship analysis [16], with 95% CIs, to estimate the cumulative probability of not having a revision of one or both prosthetic components and to estimate the cumulative probability of not having cup loosening. The differences in survival with and without screws were assessed using the log-rank test to compare the Kaplan-Meier curves. We performed statistical analysis using SPSS® software (Version 9.0; SPSS Inc, Chicago, IL, USA).
Results
We found differences in sex, weight, activity level, and acetabular and femoral types between patients receiving screws (Group 1) and those not receiving screws (Group 2) (Table 2). Postoperative radiograph analysis showed a difference between groups in only mean CPFH to AFHC distance (Table 3). Multivariate analysis showed higher screw use in women (odds ratio [OR], 2.39; 95% CI, 1.1–5.21), patients with lower physical activity (OR, 0.54; 95% CI, 0.37–0.79), Acetabular Type A (OR, 3.46; 95% CI, 1.72–6.93) or C (OR, 5.19; 95% CI, 1.5–17.97), and a CPFH to AFHC distance of 3 to 5 mm (OR, 3.53; 95% CI, 1.75–7.12) or greater than 5 mm (OR, 34.78; 95% CI, 11.46–105.56) (Table 4).
We observed lower (p = 0.047 and p = 0.011, respectively) mean postoperative ratings for pain and ROM in Group 1 (Table 5). The survival rate for revision surgery for any cause at 12 years was 95.4% (95% CI, 91%–99%) in Group 1 and 97.8% (95% CI, 95%–99%) in Group 2 (p = 0.41) (Fig. 1). There were four revisions in Group 1 and five in Group 2; all revisions were performed due to aseptic loosening of the cup. Seventeen patients (eight in Group 1, nine in Group 2) had radiographic cup loosening (Table 6). The probability of not having radiographic loosening at 12 years was 90.9% (95% CI, 84%–96%) in Group 1 and 95.1% (95% CI, 92%–98%) in Group 2 (p = 0.14) (Fig. 2). Cups with an acetabular abduction angle of higher than 50° were associated with (p = 0.042; hazard ratio, 3.177; 95% CI, 1.045–9.661) radiographic loosening in both groups. The survival rate for radiographic loosening at 12 years was lower (p = 0.002) for these cups (Fig. 3). We also observed a tendency in men for a lower risk for loosening (p = 0.071) and a greater risk in cups with a CPFH to AFHC distance of more than 5 mm on the postoperative radiograph (p = 0.092). The Cox model showed no differences when acetabular type was assessed. Other factors related to screw use for primary cup fixation and the appearance of radiographic loosening are shown (Table 7).
Discussion
Although uncemented press-fit cups obtain primary bone fixation [8, 9], there are some conditions that may preclude obtaining a primary stability of the implant [35]. We analyzed the possible relationship between factors related to patients and the reconstruction of the rotation center of the hip with press fit for uncemented hemispherical cup in primary THA; we also compared the postoperative pain, function, and ROM and rates of revision and radiographic loosening after a minimum followup of 5 years between cups implanted with or without screw fixation.
There are several limitations to our study. First, although the mean followup was 8.2 years, we know in vivo effects of screws (eg, contributing to osteolysis from wear debris) in augmenting cup fixation might appear later. Second, we had small numbers in our subcohorts. For example, the small number of Acetabular Type C hips provided some apparent trends that, in a larger cohort, could be different. The factors analyzed in the Cox model for screw use and aseptic loosening such as sex or acetabular reconstruction might have been clearer. Third, the surgical technique, although used by the same team following the same principles according to the pull-out test [35], may not be reproducible, leading to somewhat differing criteria for screw use. Fourth, this was not a prospective randomized trial, and the preoperative differences between groups may have influenced the comparisons.
Women had a higher rate of screw use, perhaps owing to the association with a greater frequency of osteoporotic bone [35]. Screw use was also more frequent in patients with a cylindrical femur [8]. Press fit was less frequently achieved in patients with less physical activity, usually with osteoporotic bone or a diagnosis other than primary osteoarthritis [6, 10, 11]. Acetabular Types A and C had a higher rate for screw use than Type B. The thinner medial wall of Acetabular Type A, frequent in hips with avascular necrosis and in inflammatory arthritis, is associated with higher stress forces, so impaired stability has been reported [2]. In Acetabular Type C, frequent in dysplastic hips, complete coverage of the cup is often difficult, so good contact between cup and bone in the cortical rim is sometimes difficult to obtain [11]. Recently, Takao et al. [33] reported no aseptic loosening in a series with a press-fit cup without screws in dysplasia. They additionally stated superior bone-cup contact is important for the stability of uncemented cups because the weightbearing force is concentrated on the superior portion of the acetabular cup, and they noted press-fit-only fixation in a dysplastic acetabulum may be challenging because of the deficient bone [33]. Our series is a relatively young population in whom we used an alumina-on-alumina uncemented THA and the patient′s acetabular types are different from other reports [2, 6, 32]. The most important factor we observed was the postoperative position of the cup; a CPFH to AFHC distance of more than 3 mm was associated with a higher rate of screw use. The acetabulum must be reamed at the anatomic site to restore the acetabular anatomy and the anatomic center of hip rotation so as to obtain stable fixation of the prosthetic components [24, 37, 41]. Since a good postoperative position of the cup is desirable, this finding suggests the importance of surgical technique.
Patients with a press-fit cup without screws reported less pain and had better ROM; however, their pathologic preoperative conditions were not similar, so this finding must be interpreted with caution. We also observed the rate of revision and aseptic loosening at 12 years was similar regardless of screw use. One in vitro study [18] suggested the addition of supplemental screws did not improve the stability of uncemented cups under proper press-fit conditions. Press-fit acetabular implantation helps to distribute compressive forces to the periphery of the acetabulum, so most of these designs have a flattened pole area to facilitate complete seating and are oversized to optimize press fit [22, 27, 39]. Won et al. [40] observed in the laboratory the use of screws increases micromotion at the opposite side of the screw due to distractive cyclic loads, and all cyclic loads were compressive in the press-fit component. There have been a few comparative clinical studies [6, 9, 23, 29–31] (Table 8). In a radiostereometric study, Önsten et al. [23] reported no differences at 2 years when screws were used. Roth et al. [29] reported there were no advantages in using screws with a press-fit cup and fewer radiographic changes in cups without screws. They concluded there could be more micromotion in a hip with a cup implanted with screws. Schmalzried et al. [31] observed a press-fit technique is effective in reducing the incidence of radiolucency. Kress et al. [17], in a study with CT-assisted osteodensitometry (using the same designs as in our series), reported nonprogressive bone radiolucent lines that were not related to the appearance of loosening of this cup. Several studies [22, 42] suggest it is important to maintain the subchondral bone to reduce the risk of cup migration. Since a vertical cup with an acetabular angle of greater than 50° was the most important factor related to aseptic loosening and a tendency for a closer distance to the approximate center of the hip rotation was also observed, we emphasize the importance of the surgical technique.
We conclude the initial press fit of an uncemented cup depends on several factors. Women, patients with less physical activity, and patients with an acetabulum with a thin medial wall and dysplastic hips had higher rate of screw use. When comparing the survival rate for revision and aseptic loosening, we observed the use of screws in hemispherical cups, when primary stability is not obtained, provided a rate of loosening similar to that of cups implanted with a press-fit technique; although the screws are useful for the purpose of providing initial stability of the cup, in some cases, they were not enough to avoid loosening. Vertical cups were at greater risk for aseptic loosening, so we recommend a careful technique when using an uncemented cup in primary THA.
References
Barrack RL. Neurovascular injury: avoiding catastrophe. J Arthroplasty. 2004;19(suppl 1):104–107.
Bizot P, Nizard R, Hamadouche M, Hannouche D, Sedel L. Prevention of wear and osteolysis: alumina-on-alumina bearing. Clin Orthop Relat Res. 2001;393:85–93.
Clohisy JC, Harris WH. The Harris-Galante porous-coated acetabular component with screw fixation: an average ten-year follow-up study. J Bone Joint Surg Am. 1999;81:66–73.
DeLee JG, Charnley J. Radiological demarcation of cemented sockets in total hip replacement. Clin Orthop Relat Res. 1976;121:20–32.
Devane PA, Home JG, Martin K, Coldham G, Krause B. Three-dimensional polyethylene wear of a press-fit titanium prosthesis: factors influencing generation of polyethylene debris. J Arthroplasty. 1997;12:256–266.
Dorr LD, Bechtol CO, Watkins RG, Wan Z. Radiographic anatomic structure of the arthritic acetabulum and its influence on total hip arthroplasty. J Arthroplasty. 2000;15:890–900.
Dorr LD, Faugere MC, Mackel AM, Gruen TA, Bognar B, Malluche HH. Structural and cellular assessment of bone quality of proximal femur. Bone. 1993;14:231–242
Dorr LD, Wan Z, Cohen J. Hemispheric titanium porous coated acetabular component without screw fixation. Clin Orthop Relat Res. 1998;351:158–168.
Engh CA, Hopper RH Jr, Engh CA Jr. Long-term porous-coated cup survivorship using spikes, screws, and press-fitting for initial fixation. J Arthroplasty. 2004;19(suppl 2):54–60.
García-Cimbrelo E, Cruz-Pardos A, Cordero J, Sanchez-Sotelo J. Low-friction arthroplasty in patients younger than 40 years old: 20- to 25-year results. J Arthroplasty. 2000;15:825–832.
García-Rey E, Cruz-Pardos A, García-Cimbrelo E. Alumina-on-alumina total hip arthroplasty in young patients: diagnosis is more important than age. Clin Orthop Relat Res. 2009;467:2281–2289.
García-Rey E, García-Cimbrelo E, Cordero-Ampuero J. Outcome of hemispherical porous-coated acetabular component with a proximally hydroxyapatite-coated anatomical femoral component: a 12- to 15-year follow-up study. J Bone Joint Surg Br. 2009;91:327–322.
Hailer NP, Garellick G, Kärrlholm J. Uncemented and cemented primary total hip arthroplasty in the Swedish Hip Arthroplasty Register: evaluation of 170,413 operations. Acta Orthop. 2010;81:34–41.
Hallan G, Dybvik E, Furnes O, Havelin LI. Metal-backed acetabular components with conventional polyethylene: a review of 9113 primary components with a follow-up of 20 years. J Bone Joint Surg Br. 2010;92:196–201.
Johnston RC, Fitzgerald RH, Harris WH, Poss R, Müller ME, Sledge CB. Clinical and radiographic evaluation of total hip arthroplasty: a standard system of terminology for reporting results. J Bone Joint Surg Am. 1990;72:161–168.
Kaplan EL, Meier P. Nonparametric estimation from incomplete observations. J Am Stat Assoc. 1958;53:457–481.
Kress AM, Schmidt R, Vogel T, Nowak TE, Forts R, Mueller LA. Quantitative computed tomography-assisted osteodensitometry of the pelvis after press-fit cup fixation: a prospective ten-year follow-up. J Bone Joint Surg Am. 2011;93:1152–1157.
Kwong LM, O’Connor DO, Sedlacek RC, Krushell RJ, Maloney WJ, Harris WH. A quantitative in vitro assessment of fit and screw fixation on the stability of a cementless hemispherical acetabular component. J Arthroplasty. 1994;9:163–170.
Manning DW, Chiang PP, Martell JM, Galante JO, Harris WH. In vivo comparative wear study of traditional and highly cross-linked polyethylene in total hip arthroplasty. J Arthroplasty. 2005;20:880–886.
Merle d’Aubigné R, Postel M. Functional results of hip arthroplasty with acrylic prosthesis. J Bone Joint Surg Am. 1954;36:451–475.
Moore MS, McAuley JP, Young AM, Engh CA Sr. Radiographic signs of osseointegration in porous-coated acetabular components. Clin Orthop Relat Res. 2006;444:176–183.
Morscher EW. Current status of acetabular fixation in primary total hip arthroplasty. Clin Orthop Relat Res. 1992;274:172–193.
Önsten I, Carlsson AS, Sanzén L, Besjakov J. Migration and wear of a hydroxyapatite-coated hip prosthesis: a controlled of a roentgen stereophotogrammetric study. J Bone Joint Surg Br. 1996;78:85–91.
Pagnano MW, Hanssen AD, Lewallen DG, Shaughnessy WJ. The effect of superior placement of the acetabular component on the rate of loosening after total hip arthroplasty: long-term results in patients who have Crowe type-II congenital dysplasia of the hip. J Bone Joint Surg Am. 1996;78:1004–1014.
Parvizi J, Sullivan T, Duffy G, Cabanela ME. Fifteen-year clinical survivorship of Harris-Galante total hip arthroplasty. J Arthroplasty. 2004;19:672–677.
Ranawat CS, Dorr LD, Inglis AE. Total hip arthroplasty in protrusio acetabuli of rheumatoid arthritis. J Bone Joint Surg Am. 1980;62:1059–1065.
Ries M, Harbaugh M. Acetabular strains produced by oversized press fit cups. Clin Orthop Relat Res. 1997;334:276–281.
Röhrl SM, Nivbrant B, Snorrason F, Kärrholm J, Nilsson KG. Porous-coated cups fixed with screws: a 12-year clinical and radiostereometric follow-up study of 50 hips. Acta Orthop. 2006;77:393–401.
Roth A, Winzer T, Sander K, Anders JO, Venbrocks RA. Press fit fixation of cementless cups: how much stability do we need indeed? Arch Orthop Trauma Surg. 2006;126:77–81.
Schmalzried TP, Harris WH. The Harris-Galante porous-coated acetabular component with screw fixation: radiographic analysis of eighty-three primary hip replacements at a minimum of five years. J Bone Joint Surg Am. 1992;74:1130–1139.
Schmalzried TP, Wessinger SJ, Hill GE, Harris WH. The Harris-Galante porous acetabular component press-fit without screw fixation: five-year radiographic analysis of primary cases. J Arthroplasty. 1994;9:235–242.
Sedel L, Kerboull L, Christel P. Alumina on alumina hip replacement results and survivorship in young patients. J Bone Joint Surg Br. 1990;72:658–663.
Takao M, Nakamura N, Ohzono K, Sakai T, Nishii T, Sugano N. The results of a press-fit-only technique for acetabular fixation in hip dysplasia. J Arthroplasty. 2011;26:562–568.
Teinturier P, Terver S, Jaramillo CV, Besse JP. [Biomechanics of the acetabulum] [in French]. Rev Chir Orthop Reparatrice Appar Mot. 1984;70(suppl 2):41–46.
Udomkiat P, Dorr LD, Wan Z. Cementless hemispheric porous-coated sockets implanted with press-fit technique without screws: average ten-year follow-up. J Bone Joint Surg Am. 2002;84:1195–2000.
Utting MR, Raghuvanshi M, Amirfeyz R, Blom AW, Learmonth ID, Bannister GC. The Harris-Galante porous-coated, hemispherical, polyethylene-lined acetabular component in patients under 50 year of age: a 12- to 16-year review. J Bone Joint Surg Br. 2008;90:1422–1427.
Vasu R, Carter DR, Harris WH. Stress distribution in the acetabular region. Part I. Before and after total joint replacement. J Biomech. 1982;15:155–164.
Wasielewski RC, Cooperstein LA, Kruger MP, Rubash HE. Acetabular anatomy and the transacetabular fixation of screws in total hip arthroplasty. J Bone Joint Surg Am. 1990;72:501–508.
Widmer KH, Zurfluh B, Morscher EW. Load transfer and fixation mode of press-fit acetabular sockets. J Arthroplasty. 2002;17:926–935.
Won CH, Hearn TC, Tile M. Micromotion of cementless hemispherical acetabular components: does press-fit need adjunctive screw fixation? J Bone Joint Surg Br. 1995;77:484–489.
Yoder SA, Brand RA, Pedersen DR, O’Gorman TW. Total hip acetabular component position affects component loosening rates. Clin Orthop Relat Res. 1988;228:79–87.
Zenz P, Stiehl JB, Knechtel H, Titzer-Hochmaier G, Schwagerl W. Ten-year follow-up of the non-porous Allofit cementless acetabular component. J Bone Joint Surg Br. 2009;91:1443–1447.
Author information
Authors and Affiliations
Corresponding author
Additional information
Each author certifies that he or she, or a member of his or her immediate family, has no commercial associations (eg, consultancies, stock ownership, equity interest, patent/licensing arrangements, etc) that might pose a conflict of interest in connection with the submitted article.
All ICMJE Conflict of Interest Forms for authors and Clinical Orthopaedics and Related Research editors and board members are on file with the publication and can be viewed on request.
Clinical Orthopaedics and Related Research neither advocates nor endorses the use of any treatment, drug, or device. Readers are encouraged to always seek additional information, including FDA approval status, of any drug or device before clinical use.
Each author certifies that his or her institution approved the human protocol for this investigation, that all investigations were conducted in conformity with ethical principles of research, and that informed consent for participation in the study was obtained.
About this article
Cite this article
García-Rey, E., García-Cimbrelo, E. & Cruz-Pardos, A. Cup Press Fit in Uncemented THA Depends on Sex, Acetabular Shape, and Surgical Technique. Clin Orthop Relat Res 470, 3014–3023 (2012). https://doi.org/10.1007/s11999-012-2381-9
Published:
Issue Date:
DOI: https://doi.org/10.1007/s11999-012-2381-9