Summary
The main shortcoming with the procedure to determine 226Ra in a gamma spectrum of an environmental sample by means of the 214Bi and 214Pb photopeaks is the likelihood of 222Rn leakage from the sample counting vial. An option to make such determination is to disregard the 226Ra gamma-contributions to the spectrum, other than 186.2 keV (3.5%), subtracting the 235U contribution to the 226Ra+235U peak at 186 keV. The use of this option to determine directly 226Ra activity concentrations in environmental samples and in NORM/TENORM matrices will be presented and discussed.
Article PDF
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Justo, J., Evangelista, H. & Paschoa, A. Direct determination of 226Ra in NORM/TENORM matrices by gamma-spectrometry. J Radioanal Nucl Chem 269, 733–737 (2006). https://doi.org/10.1007/s10967-006-0293-4