Abstract
A monochromator integrating-sphere-based spectral comparator facility has been developed to calibrate standard radiation thermometers in terms of the absolute spectral radiance responsivity, traceable to the PTB cryogenic radiometer. The absolute responsivity calibration has been improved using a 75 W xenon lamp with a reflective mirror and imaging optics to a relative standard uncertainty at the peak wavelength of approximately 0.17 % (k = 1). Via a relative measurement of the out-of-band responsivity, the spectral responsivity of radiation thermometers can be fully characterized. To verify the calibration accuracy, the absolutely calibrated radiation thermometer is used to measure Au and Cu freezing-point temperatures and then to compare the obtained results with the values obtained by absolute methods, resulting in T − T 90 values of +52 mK and −50 mK for the gold and copper fixed points, respectively.
Article PDF
Similar content being viewed by others
Avoid common mistakes on your manuscript.
References
D.R. Taubert, J. Hartmann, J. Hollandt, J. Fischer, in Temperature, Its Measurement and Control in Science and Industry, vol. 7, part 1, ed. by D.C. Ripple (AIP, New York, 2003), pp. 7–12
Anhalt K., Hartmann J., Lowe D., Machin G., Sadli M., Yamada Y.: Metrologia 43, S78 (2006)
Brown S.W., Eppeldauer G.P., Lykke K.R.: Appl. Opt. 45, 8218 (2006)
Anderson V.E., Fox N.P., Nettleton D.H.: Appl. Opt. 31, 536 (1992)
Y. Ichino, I. Saito, Y. Yamada, J. Ishii, Spectral Radiance Responsivity Calibration Facility for Thermodynamic Temperature Determination. Presented at TEMPMEKO 2007, 10th International Symposium on Temperature and Thermal Measurements in Industry and Science, Lake Louise, Canada (2007)
A. Sperling, O. Larionov, U. Grusemann, S. Winter, in Proceedings of NEWRAD 2005, ed. by J. Groebner (PMODWRC, Davos, Switzerland, 2005), p. 93
Anhalt K., Zelenjuk A., Taubert D.R., Keawprasert T., Hartmann J.: Int. J. Thermophys. 30, 192 (2009)
Ahtee V., Brown S.W., Larason T.C., Lykke K.R., Ikonen E., Noorma M.: Appl. Opt. 40, 4228 (2007)
Yoon H.W., Gibson C.E., Khromchenko V., Eppeldauer G.P., Bousquet R.R., Brown S.W., Lykke K.R.: Int. J. Thermophys. 29, 285 (2008)
T. Keawprasert, K. Anhalt, R.D. Taubert, A. Abd-Elmageed, A. Sperling, J. Hartmann, in Proceedings of NEWRAD 2008, 10th International Conference on New Development and Applications in Optical Radiometry, ed. by Dong-Hoon Lee (KRISS, Daejeon, Korea, 2008), p. 287
Hartmann J.: Phys. Rep. 469, 205 (2009)
H.C. McEvoy, The Examination of Base Parameters for ITS-90 Scale Realisation in Radiation Thermometry, EUROMET.T-S1 Main Measurement Report Final Report (NPL, United Kingdom, 2007)
Fischer J., Jung H.J.: Metrologia 26, 245 (1989)
Anhalt K., Wang Y., Yamada Y., Hartmann J.: Int. J. Thermophys. 29, 969 (2008)
Woodward J.T., Smith A.W., Jenkins C.A., Lin C., Brown S.W., Lykke K.R.: Metrologia 46, 277 (2009)
www.bipm.org/cc/CCT/Allowed/24/D13_rev_WG4_report_CCT_25_June_2008.pdf
Author information
Authors and Affiliations
Corresponding author
Additional information
Asterisks in the text represent references to commercial products and are provided for identification purposes only and constitute neither endorsement nor representation that the item identified is the best available for the stated purpose.
Rights and permissions
About this article
Cite this article
Keawprasert, T., Anhalt, K., Taubert, D.R. et al. Monochromator-Based Absolute Calibration of Radiation Thermometers. Int J Thermophys 32, 1697–1706 (2011). https://doi.org/10.1007/s10765-011-1031-x
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10765-011-1031-x