Abstract
The paper describes the new experimental setup assembled at the PTB for the absolute spectral responsivity measurement of radiation thermometers. The concept of this setup is to measure the relative spectral responsivity of the radiation thermometer using the conventional monochromator-based spectral comparator facility also used for the calibration of filter radiometers. The absolute spectral responsivity is subsequently measured at one wavelength, supplied by the radiation of a diode laser, using the new setup. The radiation of the diode laser is guided with an optical fiber into an integrating sphere source that is equipped with an aperture of absolutely known area. The spectral radiance of this integrating sphere source is determined via the spectral irradiance measured by a trap detector with an absolutely calibrated spectral responsivity traceable to the primary detector standard of the PTB, the cryogenic radiometer. First results of the spectral responsivity calibration of the radiation thermometer LP3 are presented, and a provisional uncertainty budget of the absolute spectral responsivity is given.
Article PDF
Similar content being viewed by others
Avoid common mistakes on your manuscript.
References
D.R. Taubert, J. Hartmann, J. Hollandt, J. Fischer, in Temperature, Its Measurement and Control in Science and Industry, vol 7, part 1, ed. by D. Ripple (AIP, New York, 2003), pp. 7–12
Taubert D.R., Friedrich R., Hartmann J., Hollandt J.: Metrologia 40, S35 (2003)
Werner L., Fischer J., Johannsen U., Hartmann J.: Metrologia 37, 279 (2000)
Yoon H.W.: Metrologia 43, S22 (2006)
A. Sperling, O. Larionov, U. Grusemann, S. Winter, in Proceedings of the 9th International Conference on New Developments and Applications in Optical Radiometry (NEWRAD), ed. by J. Groebner (PMODWRC, Davos, Switzerland, 2005), p. 93
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Anhalt, K., Zelenjuk, A., Taubert, D.R. et al. New PTB Setup for the Absolute Calibration of the Spectral Responsivity of Radiation Thermometers. Int J Thermophys 30, 192–202 (2009). https://doi.org/10.1007/s10765-008-0480-3
Published:
Issue Date: