Introduction

Nitrite (NO2 ) is an important source of nitrogen in green plants, and its complete reduction is achieved in nature by the nitrite reductase enzyme [1]. It is used as an additive in some types of food, and its occurrence in soils, waters, foods, and physiological systems is prevalent. The presence of excess nitrite in vegetables, drinking water, and food products is a serious threat to human health [2]. Nitrite promotes the irreversible oxidation of hemoglobin and reduces the blood capacity to transport oxygen. Also, it may interact in the stomach with amines and amides forming highly carcinogenic N-nitrosamine, many of which are known to be carcinogens. Therefore, NO2 determination is of great significance for environment security and public health. Monitoring variations of nitrite is important for understanding their biogeochemical processes in aquatic environments and for developing better methods of managing water quality [311].

Many analytical methods have been proposed for nitrite detection, mainly based on spectrophotometry, chromatography, capillary electrophoresis, chemiluminescence, and electrochemistry [1216]. Most of these methods are often affected by the presence of coexisting cationic or anionic species, and they are either time-consuming or needing sophisticated instruments. Owing to their rapid response, cheap, safe, and simple use, the electrochemical methods have often been employed for nitrite detection. In doing so, chemically modified electrodes have proven to be particularly efficient, in terms of eliminating the effects of interference species and/or by providing suitable electrocatalytic properties likely to improve nitrite oxidation processes, which was notably achieved by modifying the electrode surface with redox mediators [311, 1729]. Various redox mediators (e.g., dyes) were used for to build amperometric sensors for nitrite [30, 31]. On the other hand, the interest of electrodes based on nanocomposites [32, 33] or carbon nanotubes for the detection of bioorganic and inorganic compounds, e.g. nitrites, has been the subject of several previous reports (see, e.g., [3440]).

In this context, the present paper describes the coverage of glassy carbon (GC) electrodes with a polymeric film of Toluidine Blue O (GC/poly-TBO), as obtained by electropolymerization, and the subsequent modification of GC/poly-TBO with single-walled carbon nanotubes (SWCNT) by immobilization onto the polymeric film to give a composite electrode (GC/poly-TBO-SWCNT). The aim of using carbon nanotubes was to investigate their possible interest to increase the sensitivity for nitrite detection as expected from improved electron transport issues. Actually, the catalytic behavior of Toluidine Blue with respect to nitrite oxidation is known [30, 41] and the combined use of Toluidine Blue and carbon nanotubes has been reported for bioelectrocatalytic applications [4245], but, to date, poly(Toluidine Blue) and carbon nanotubes composite films were not yet described for nitrite detection. The electrochemical and electrocatalytic behavior of both GC/poly-TBO and GC/poly-TBO-SWCNT electrodes toward nitrite oxidation was studied comparatively. This was made by investigating the electrochemical stability and the influence of some experimental parameters on the voltammetric response (pH, scan rate). The obtained electrodes were tested as amperometric sensors for nitrite detection in synthetic and real samples. The nitrite concentration in a real sample (wastewater) was determined using the standard addition method and compared to a standardized procedure (Griess assay).

Experimental

Reagents

Toluidine Blue O (TBO) was purchased from Aldrich (Steinheim, Germany). Single-walled carbon nanotubes (SWCNTs) and sodium tetraborate were obtained from Sigma (St. Louis, MO, USA), while sodium nitrite was purchased from Reactivul Bucharest (Romania). Na2HPO4·2H2O and NaH2PO4·H2O were purchased from Merck (Darmstadt, Germany).

All reagents were of analytical grade and used as received. The supporting electrolyte was 0.1 M phosphate buffer solution, obtained by mixing solutions of Na2HPO4·2H2O and NaH2PO4·H2O in appropriate ratios. When necessary, pH was adjusted in the interval of 2–10, using H3PO4 or NaOH solutions. The nitrite stock solution used throughout this study was made of 0.1 M NaNO2, which was then diluted to reach the desired concentrations.

The electropolymerization solution (pH 9.1) used to modify the electrodes was composed of 5·10−4 M TBO, 0.1 M NaNO3, and 0.01 M sodium tetraborate [30].

Electrode preparation

Glassy carbon electrodes modified with poly(Toluidine Blue) (GC/poly-TBO)

Prior to electrochemical experiments, GC electrodes (surface area of 0.07 cm2) were polished with aqueous 0.3 μm alumina slurry on a polishing cloth, followed by subsequent washing for 5 min in an ultrasonic bath with ethanol and water in order to remove any trace impurities. Then the polymer film electrode was prepared by scanning successive cyclic voltammograms of TBO on GC electrode in the potential range between −0.6 and 1.1 V vs. Ag/AgCl/KClsat, at 50 mV s−1, during 20 cycles, in 5·10−4 M TBO solution.

Glassy carbon electrodes modified with poly(Toluidine Blue) and carbon nanotubes (GC/poly-TBO-SWCNT)

The SWCNTs were immobilized onto the GC/poly-TBO electrode by using water as the dispersing agent. The casting solution was prepared by introducing 1 mg of SWCNTs into 200 μL deionized water.

The SWCNT were actually oxidized (giving rise to an oxygen-to-carbon atomic ratio of 6 %, as determined by X-ray photoelectron spectroscopy, exhibiting the typical narrow C1s sp2 main signal at 284.1 eV and a small contribution of C-O moieties at 285.9 eV), bearing enough hydrophilic groups to be quite easy to disperse. Strong sonication was applied and, if aggregates were still present, a rather homogeneous SWCNTs dispersion was achieved. Then 20 μL aliquot of this sonicated suspension was placed directly onto the GC/poly-TBO electrode surface and left to dry, after which the electrode was ready for use. In this way, the GC/poly-TBO-SWCNT electrode was obtained. A microscopic characterization of the proposed modified electrode by SEM revealed deposits with a dendritic-like morphology (Fig. 1a), containing quite regular substructures of submicron size (Fig. 1b–d) into which the SWCNTs are expected to be deposited (not visible by SEM). It is noteworthy that such configuration “electrode/polymer/nanotubes” might be unexpected by comparison to the more conventional “electrode/nanotubes/modifier” configuration, but this latter gave rise to lower sensitivities for the analysis of nitrite (by ca. 20 %), so that the sandwich configuration GC/poly-TBO/SWCNT was preferred here.

Fig. 1
figure 1

SEM micrographs of GC/poly-TBO-SWCNT at various magnifications

All presented results are the average one of at least three identically prepared electrodes, if not otherwise mentioned.

Electrochemical measurements

Cyclic voltammetry was performed in a conventional three-electrode system. A platinum wire was used as a counter electrode, Ag|AgCl/KClsat as a reference electrode, and the modified glassy carbon electrodes (3 mm diameter) as working ones. All electrochemical experiments were carried out using an Autolab electrochemical analyzer (Autolab-PGSTAT10, Eco Chemie, Utrecht, Netherlands). All measurements were performed at room temperature.

Batch amperometric measurements at various nitrite concentrations were carried out at an applied potential of +960 mV vs. Ag/AgCl/KClsat, under magnetic stirring, using 0.1 M phosphate buffer solution as supporting electrolyte. The current-time data were collected using the above mentioned electrochemical analyzer.

Spectrophotometric analysis of nitrite was carried out with the UV–Vis spectrophotometer Jasco 615.

Monitoring pH of the phosphate buffer solutions was made using an HI255 pH meter (Hanna Instruments, Romania), with a combined glass electrode.

Results and discussions

Basic electrochemical characterization

Typical cyclic voltammograms recorded for TBO electropolymerization on GC electrode (Fig. 2) are characterized by well-defined voltammetric peaks. From multisweep potential scanning of the GC electrode in the TBO solution, it can be observed that the anodic and cathodic peak currents increased continuously during cycling (Fig. 2), with shape variations in agreement with those previously observed for electrodeposited poly(Toluidine Blue) films [30]. This proves the progressive deposition of a TBO film on the GC surface, as also confirmed by the indigo color specific to TBO on the electrode surface. The application of 20 potential scans for TBO electropolymerization was the optimum condition giving rise to the highest voltammetric signals, in agreement with previous observations made for the electropolymerization of other dyes of related structure (e.g., 25 cycles for brilliant cresyl blue [31]).

Fig. 2
figure 2

Cyclic voltammograms recorded at 50 mV s−1 for TBO electropolymerization on GC electrode (solution: 5·10−4 M TBO, 0.1 M NaNO3, and 0.01 M sodium tetraborate)

Analysis of the obtained poly-TBO-modified electrodes without (GC/poly-TBO) or with carbon nanotubes (GC/poly-TBO-SWCNT) by cyclic voltammetry in phosphate buffer (pH 7, 10 mV s−1, see plain curves in parts A and B of Fig. 3) reveals the existence of peak pairs characterized by quasi-reversible 2e transfer processes (∆E p = 0.14 V for GC/poly-TBO and 0.08 V for GC/poly-TBO-SWCNT). Values of formal standard potentials, E 0′, of such adsorbed mediator (evaluated as average of anodic and cathodic peak potentials) are −0.01 V vs. Ag/AgCl/KClsat for GC/poly-TBO and −0.04 V vs. Ag/AgCl/KClsat for GC/poly-TBO-SWCNT. The amount of electroactive TBO moieties was slightly larger in the presence of carbon nanotubes, resulting in higher apparent surface coverage (Γ app = 6.8 × 10−10 mol cm−2 for GC/poly-TBO and 15.3 × 10−10 mol cm−2 for GC/poly-TBO-SWCNT).

Fig. 3
figure 3

Cyclic voltammograms obtained at 10 mV s−1 using GC/poly-TBO (a at pH 7), GC/poly-TBO-SWCNT (b at pH 7; c at pH 3) and unmodified GCE (d), respectively, without (plain curves) and with increasing concentrations of nitrite in the medium

Actually, the values of surface coverage are related to the electropolymerization conditions. We even obtained, in other cases, smaller surface coverages (of 10−12 order), as in [46]. Here, the coverage corresponds to ca. 5–10 TBO units per square nanometer, so that we do have a very thin coating layer of few monolayers.

It can be concluded from these results that using SWCNT has a beneficial effect on the electrochemical behavior of the modified electrodes (lower ∆E p value and a higher amount of electroactive TBO). Also, it can be observed that the GC/poly-TBO-SWCNT-modified electrodes exhibit a negative shift of the E 0′ value, which can be attributed to a specific feature of CNT-modified electrodes [34].

As expected for surface-confined redox species, the cyclic voltammograms recorded at various potential scan rates (in the range of 0.01–6.4 V s−1) presented a linear dependence of peak current intensity (I p) on scan rate (v) (results not shown). The slopes of log I p vs. log v dependencies were close to the theoretical value of one, confirming the existence of adsorbed species. From such measurements performed at different scan rates, the heterogeneous electron transfer rate constants, k s, and transfer coefficients, α, were calculated using the method proposed by Laviron [47], from the variation of the peak potentials with the potential scan rate and based on the Eqs. (1) and (2). The values for k s increases in the sequence: 17.4 s−1 (pH 3) < 73.9 s−1 (pH 7) < 94.9 s−1 (pH 9) for GC/poly-TBO. For GC/poly-TBO-SWCNT, the k s value was 97.4 s−1 at pH 7.

It can be observed that the value of k s constant for GC/poly-TBO-SWCNT electrode is slightly higher than for GC/poly-TBO in phosphate buffer at pH 7.

$$ {E}_{\mathrm{pc}}-{E}^{0\prime }=\frac{\mathrm{RT}}{\alpha nF}\cdot 2.3 \log \frac{\alpha nF}{\mathrm{RT}{k}_s}+\frac{2.3\mathrm{RT}}{\alpha nF} \log v $$
(1)
$$ {E}_{\mathrm{pa}}-{E}^{0\prime }=\frac{\mathrm{RT}}{\left(1-\alpha \right) nF}\cdot 2.3 \log \frac{\left(1-\alpha \right) nF}{\mathrm{RT}{k}_s}+\frac{2.3\mathrm{RT}}{\left(1-\alpha \right) nF} \log v $$
(2)

The values of k s appear rather high in contrast to the CV curves presented in Fig. 3, but these values depend also on the surface coverage. Note that values for k s as high as 135 s−1 have been reported for a system that was not reversible (see [48]).

Also, the aforementioned k s values suggest an influence of pH, which was especially studied for the GC/poly-TBO electrode (GC/poly-TBO-SWCNT being only used for comparison purpose at the optimum pH 7). Actually, not only electron transfer rate parameters were affected by pH but also the voltammetric peak potentials were pH dependent (compare the plain curves in parts B and C of Fig. 3). Within the pH range investigated here (from 3 to 9), the E 0′ vs. pH dependence was linear, with a slope of 0.038 V/∆pH (E 0' = 0.26 + 0.038 pH; R (correlation coefficient)/N (no. of experimental points) = 0.961/7). This suggests the existence of a redox process involving 2e/2H+ in accordance with literature data [30].

pH was also found to affect the apparent surface coverage (i.e., the amount of electroactive TBO species per surface unit) and the stability of the electrode response upon continuous potential scanning. Stability tests of investigated systems were realized in potentiodynamic conditions: the potential of modified electrodes was scanning in the potential range covering the redox activity of mediator, with the scan rate of 50 mV s−1, during 20 cycles, in phosphate buffer solution of pH 3, pH 5, pH 7, and pH 9 for GC/poly-TBO and in phosphate buffer solution of pH 7 for GC/poly-TBO-SWCNT. Actually, pH 7 was the best compromise between good sensitivity and acceptable stability. As observed for GC/poly-TBO, at pH 5 and pH 9, the peaks almost vanished after 20 cycles, whereas data obtained at pH 3 revealed a quite good stability but very low sensitivity (Γ app values 4–5 times lower than for the pH range of 5–9). Further experiments will be thus performed at pH 7. It should be also noted that if Γ app was 2–3 times higher for GC/poly-TBO-SWCNT than for GC/poly-TBO (as mentioned above), it tended to deactivate more rapidly. The deactivation rate constants, k deact, as calculated from the slope of Eq. 3 (variation of Γ app values with time of potential scanning, see Fig. 4), were 3.3 × 10−12 mol cm−2 s−1 and 1.1 × 10−11 mol cm−2 s−1, respectively, for GC/poly-TBO and GC/poly-TBO-SWCNT.

$$ {\varGamma}_{\mathrm{app}}={\varGamma}_{\mathrm{app}\ \left(t=0\right)}-{k}_{\mathrm{deact}}t $$
(3)
Fig. 4
figure 4

Linear dependence of surface coverage Γ vs. cycling time for GC/poly-TBO and GC/poly-TBO-SWCNT. Experimental conditions: starting potential, −600 mV vs. Ag/AgCl/KClsat; scan rate, 50 mV s−1; supporting electrolyte, 0.1 M phosphate buffer, pH 7

Based on the good electrochemical stability observed for GC/poly-TBO-SWCNT (expressed by a low value of deactivation rate constant: 1.1 × 10−11 mol cm−2 s−1), one can assume a rather durable immobilization of SWCNTs on the electrode surface.

Electrochemical response to nitrite

Cyclic voltammetry evidence of catalytic properties

Figure 3 presents the cyclic voltammograms corresponding to GC/poly-TBO (part A) and GC/poly-TBO-SWCNT (part B at pH 7 and part C at pH 3). As can be observed, in the presence of nitrite, an oxidation peak starts to grow at ca. +0.60 V to reach a maximum at +0.80 V, which significantly increases with nitrite concentration. By comparison to the unmodified GC electrode (Fig. 3d), these signals are slightly shifted toward lower overpotentials (i.e., the oxidation peak starts to grow at ca. +0.70 V on unmodified GC and reach a maximum at +0.90 V), suggesting a favorable electrocatalytic effect due to the presence of TBO onto the electrode surface. The rather small shift in the observed overpotentials with respect to the unmodified electrode is consistent with previous observations for similar mediators applied to nitrite detection [30, 31]. Actually, the oxidation of nitrite is quite complex at glassy carbon electrode and usually better defined in acidic medium [49], but all further investigations were preferably made in neutral pH here because this pH is more appropriate for amperometric nitrite detection in real samples. In addition, comparing the responses of GC/poly-TBO and GC/poly-TBO-SWCNT also features a slightly better electrocatalytic behavior for GC/poly-TBO-SWCNT (lower overpotentials by ca. 30 mV) as well as some increase in sensitivity for nitrite detection in the presence of SWCNTs (compare parts A and B in Fig. 3), yet with larger capacitive currents as expected from the increased electroactive surface area when CNTs are present onto the electrode surface. This latter effect should be overcome when working in amperometric conditions, as discussed in section “Amperometric detection of nitrite ions.” Note that this GC/poly-TBO-SWCNT configuration gave rise to larger electrocatalytic peak enhancement than the GC/SWCNT-poly-TBO one (compare Figs. 3b and 5a).

Fig. 5
figure 5

Cyclic voltammograms obtained at 10 mV s−1 using GC/SWCNT/poly-TBO (a) and GC/SWCNT (b), respectively, without (plain curves) and with 10−3 M nitrite in the medium (dashed curves)

From the mechanistic point of view, it appears clearly that the electrocatalytic effect is due to the poly-TBO (i.e., no significant electrocatalytic effect of SWCNT, see control experiment in Fig. 5b) and that the presence of SWCNT only adds an additional effect of sensitivity enhancement, which might be explained by their high specific surface area and good conductivity (contributing thereby to improve the electrocatalytic behavior of poly-TBO). The possible electrocatalytic process is thus expected to operate as previously described [30] and can be expressed as follows:

$$ \begin{array}{l}\mathrm{poly}-{\mathrm{TBOH}}^{+}+{{\mathrm{NO}}_2}^{-}\to \mathrm{poly}-{\mathrm{TBOH}}^{+}\cdot {{\mathrm{NO}}_2}^{-}\hfill \\ {}\mathrm{poly}-{\mathrm{TBOH}}^{+}\cdot {{\mathrm{NO}}_2}^{-}+{\mathrm{H}}_2\mathrm{O}\to \mathrm{poly}-{\mathrm{TBOH}}^{+}\cdot {{\mathrm{NO}}_3}^{-}+2{\mathrm{H}}^{+}+2{\mathrm{e}}^{-}\hfill \\ {}\mathrm{poly}-{\mathrm{TBOH}}^{+}\cdot {{\mathrm{NO}}_3}^{-}\to \mathrm{poly}-{\mathrm{TBOH}}^{+}+{{\mathrm{NO}}_3}^{-}\hfill \end{array} $$

Amperometric detection of nitrite ions

The amperometric measurements were performed in stationary regime, at an applied potential of +0.92 V vs. Ag/AgCl/KClsat, at two pH values (3 and 7) in magnetically stirred solutions. The main analytical parameters are presented in Table 1. The detection limit was calculated as signal/noise ratio of 3, from a current vs. time curve corresponding to the analysis of nitrite at a concentration of 8 × 10−5 M (Fig. 6). Figure 6a illustrates typical amperometric signals recorded for the analysis of nitrite at increasing concentrations at pH 7 (as well as the corresponding calibration curves in Fig. 6b). The results clearly show much better sensitivity for GC/poly-TBO-SWCNT in comparison to GC/poly-TBO (compare curves “a” and “b” in Fig. 4), confirming again the improving role of carbon nanotubes in close contact to the polymer. Also, from successive calibration experiments in the presence of SWCNTs, it was observed that the analytical parameters remained almost at the same values (if ten repetitive measurements were made in 1 day, the sensor losts ∼2 % of the initial response). This confirmed once again that SWCNT adhered to the polymer, most probably via favorable hydrophobic interactions.

Table 1 Analytical parameters corresponding to amperometric calibration curves for nitrite detection, using GC/poly-TBO and GC/poly-TBO-SWCNT electrodes (pH was adjusted using 0.1 M phosphate buffer solutions)
Fig. 6
figure 6

It (a) and corresponding calibration (b) curves for nitrite, as measured using GC/poly-TBO (a) and GC/poly-TBO-SWCNT (b). Experimental conditions: applied potential, +920 mV vs. Ag/AgCl/KClsat; supporting electrolyte, 0.1 M phosphate buffer solution, pH 7. Every step in It curves corresponds to a final concentration of nitrite, and this is obtained by using the same calculation as in Fig. 4a, curve b, the last five steps (the previous steps for curve b are 1, 0.8, 0.6, 0.4, 0.2, 0.1, 0.08, 0.06, 0.04, and 0.02 mM, etc.)

It can be observed from data in Table 1 that the pH solution significantly influences the linear regression parameters for GC/poly-TBO and GC/poly-TBO-SWCNT. The best detection limit (0.1 μM) was obtained with GC/poly-TBO-SWCNT-modified electrode at pH 3, but the obtained results prove that both types of modified electrodes constitute quite good detectors for nitrite, owing to their relatively low detection limits. Good response time was also observed as the amperometric response reached 95 % of the steady-state value within ca. 4 s after the addition of nitrite. Compared with previous works [30], the present system enables to work at less anodic values (operating potential lower by ca. 200 mV here) but still suffers from lower sensitivity.

Table 2 compares the analytical performance of amperometric sensors for nitrite, such as sensitivity, linear range, and detection limit, based on other modified electrodes reported in the literature. As it can be seen from Table 2, all the analytical performance obtained for the GC/poly-TBO-SWCNT sensor presented here is comparable with that obtained in previous reports and sometimes even better.

Table 2 A comparison of analytical detection parameters obtained with various modified electrodes in the literature for electrocatalytic detection of nitrite

On the other hand, as expected for such modified electrodes, in the whole studied concentration range for reaction of nitrite electrocatalytic oxidation, the apparent reaction order toward nitrite was very close to unity (see the slope of log I vs. log C dependence from Table 1, as determined in the concentration range corresponding to linear domain of amperometric response). This proves that in all cases the nitrite sensors work under kinetic control, corresponding to reaction order 1 between nitrite and the immobilized mediator. It can be observed that the GC/poly-TBO-SWCNT offered the best performance, confirming again the beneficial effects of CNTs in enhancing reaction rates for nitrite oxidation.

The stability of the modified electrodes was finally investigated. Both modified electrodes present excellent long-term stability under continuous use. They lost ∼ 5 % of their initial response after storage for 4 weeks in 0.1 mM nitrite and ∼ 2 % after storage of a week in 0.1 mM nitrite.

Several chemical species were investigated with respect to their possible interference in the amperometric determination of nitrite. The results showed that 100-fold SO4 2−, CO3 2−, Cl, NO3 , Ca2+, and Mg2+ had no noticeable effect (<1 % for most species, ∼3 % for CO3 2−) on the current responses of 0.1 mM nitrite. These results showed that the obtained amperometric sensors possessed high selectivity.

Determination of nitrite in a wastewater sample

The GC/poly-TBO-SWCNT sensor was tested for the detection of nitrite in a wastewater sample obtained from a company of milk products, using the standard addition method. The real sample was diluted by 100 to fall within the linear range of the calibration curve, the applied potential was +0.92 V vs. Ag/AgCl/KClsat and pH was adjusted to 7. The obtained nitrite concentration in that sample was 60.6 mM. This value is very close to that obtained when analyzing the same wastewater sample using the Griess protocol [50] (i.e., 61.3 mM). The good agreement between the results obtained using the spectrophotometric method and those from the electrochemical method shows the possibility to use this sensor for nitrite determination in real samples.

Conclusions

This work has demonstrated that the immobilization of single-walled carbon nanotubes onto glassy carbon electrodes modified with a polymeric film of Toluidine Blue O contributes to improve the catalytic behavior of the mediator layer toward nitrite oxidation. Both pH effects (potential shifts, variations in sensitivity, and heterogeneous rate constants) and good stability have been evidenced. The obtained modified electrodes can be used as amperometric sensors for nitrite detection. Kinetic parameters were determined from the calibration curves, and the best sensitivity was obtained for GC-poly-TBO-SWCNT (5.85 mA M−1). The obtained sensors have been successfully applied to the determination of nitrite concentration in a wastewater sample, using the standard addition method, and the results were confirmed by comparison to the classical spectrophotometric method.