Introduction

The carcinoembryonic antigen (CEA) is a complex glycoprotein and one of the most widely used tumor markers [1]. CEA serum levels are associated with tumor development, and CEA assays have been applied to a number of clinical questions related to the diagnosis, staging, monitoring and prognosis of cancer [2]. Therefore, a high-sensitivity and high-specificity CEA detection method for research and clinical diagnosis would be very useful.

Numerous methods for detecting CEA have been reported, such as radioimmunoassay, fluorescence measurement, column chromatography and enzyme immunoassay, but these methods have several disadvantages, such as complex sample preprocessing and handling operations, easy contamination, high cost and long analytic time [3,4,5,6]. Electrochemical immunosensors are based on the antigen-antibody combination reaction and have been widely used in clinical diagnosis and bioassays with high selectivity and specificity [7]. Compared with other methods, electrochemical immunosensors have attracted much attention due to their advantages of high efficiency, simple operation, fast response time, miniaturization and low cost [8].

Nanomaterials have been widely used as immunosensing substrates because of their large surface area, excellent biocompatibility, excellent electrical conductivity, high receptor loading and catalytic ability [9]. Gold nanoparticles (AuNPs) are the most commonly used nanomaterial, as they not only have the characteristics of the above materials but also have good redox activity [10]. Recently, cuprous oxide (Cu2O) nanomaterials have attracted significant attention from researchers due to their outstanding electrocatalytic performance toward H2O2 and have been widely applied in the immobilization of biological receptors on the surface of electrodes [11,12,13]. Fengyuan’s group reported the use of a Cu2O/graphene nanocomposite and its application in an electrochemical sensor for dopamine [14]. Research concerning the surface and facet-dependent catalytic properties of Cu2O nanoparticles shows that octahedral Cu2O nanocrystals have a larger surface area and better electrocatalytic ability than other shapes [15]. Nanocomposites doped with a noble metal show unique characteristics compared to their monometallic single metal oxide counterparts [16]. Thus, we synthesized Cu2O-Au nanocomposites consisting of an octahedral Cu2O nanocrystal precursor directly covered with gold nanoparticles (AuNPs), and this structure showed a further increased surface area, good stability, enhanced ability to immobilize electroactive substances and biomolecules, and excellent electrocatalytic activity toward the reduction of H2O2.

The immobilization of antibodies is an important factor affecting the sensitivity of immunosensors. Conventional antibody immobilization methods, such as physical adsorption and covalent bond immobilization, have low ability and efficiency to immobilize antibodies due to nonspecific adsorption and steric hindrance [17,18,19,20]. In contrast, the immobilization of antibodies using biological carriers increase the amount of immobilized antibodies [21]. Staphylococcal protein A (SPA) is a protein from the cell walls of Staphylococcus aureus [22]. SPA has four Fc binding domains that can specifically bind to the Fc regions of an antibody, and these properties greatly improve the antibody capture efficiency [23,24,25,26,27]. Hence, to further enhance the sensitivity of our sensor and reduce the analytic time, we introduce SPA as an antibody carrier.

In this work, octahedral Cu2O-Au nanocomposites as a signal-amplifying immunosensing substrate, were first applied in an electrochemical immunosensor for the ultrasensitive direct detection of CEA. As a signal-amplifying immunosensing substrate, the Cu2O-Au nanocomposites not only possess large surface area and high loading capacity but also display remarkable ability to catalyze H2O2 and excellent conductivity. Furthermore, SPA was introduced by loading on the Cu2O-Au nanocomposites through Au-N bonds to efficiently capture anti-CEA. Experimental studies showed that this designed electrochemical immunosensor has excellent performance and exhibits high sensitivity and extraordinary specificity for the quantitative detection of CEA in human serum, showing practicability in clinical applications.

Experimental methods

Reagents and apparatus

CEA and anti-CEA were purchased from Shanghai Linc-Bio Science Co., Ltd. (Shanghai, China, http://www.linc-bio.cn). Recombinant staphylococcal protein A was purchased from Hangzhou Neuropeptide Biological Science and Technology Inc., Ltd. (Hangzhou, China, http://www.nuptec.com/cn). Bovine serum albumin and gold (III) chloride tetrahydrate (HAuCl4·4H2O) were purchased from Sigma-Aldrich Chemical (USA, http://www.sigmaaldrich.com). Anhydrous copper (II) chloride (CuCl2) was purchased from Sangon Biotech (Shanghai, China, http://www.sangon.com/). Hydrazine monohydrate (N2H4·H2O, 98%) and polyvinylpyrrolidone (PVP K29–32) were purchased from Aladdin. (Shanghai, China, http://www.aladdin-e.com). Phosphate buffered saline (PBS) containing 0.1 M Na2HPO4, 0.1 M KH2PO4 and 0.1 M KCl was used as the working buffer. All reagents were of analytical grade, and all solutions were prepared using Millipore-Q water (≥18.2 MΩ·cm).

All electrochemical measurements, including cyclic voltammetry (CV), electrochemical impedance spectroscopy (EIS) and chronoamperometry (i-t), were performed on an AUTOLAB PGSTAT302 N electrochemical workstation from Metrohm Autolab B.V. (Switzerland, http://www.metrohm-autolab.com). All electrochemical measurements adopted the conventional three-electrode system, using a platinum wire as the counter electrode, Ag/AgCl (with 3 M KCl) as the reference electrode and a modified 3-mm-diameter glassy carbon electrode (GCE) as the working electrode; all electrodes were purchased from Gaossunin Technology Co., Ltd. (Nanjing, China, http://www.gaossunin.com). Field emission scanning electron microscopy (FE-SEM) images were obtained using a Hitachi S4800 scanning electron microscope (Hitachi Limited, Japan). Energy dispersive X-ray photoelectron spectroscopy (EDS) was carried out with a VG Scientific ESCALAB 250 spectrometer (Thermoelectricity Instruments, USA). All measurements were carried out at room temperature under optimum conditions.

Fabrication of the electrochemical immunosensor

The fabrication procedure of the electrochemical immunosensor is outlined in Scheme 1. Prior to use, the glassy carbon electrode was polished sequentially with to a mirror with Al2O3 powder of 0.3 μm and 0.05 μm to a mirror. Then, the electrode were successively subjected to ultrasonic cleaning in ultrapure water and anhydrous ethanol for 5 min and allowed to dry at room temperature. Afterwards, the electrode was immersed in 0.5 M H2SO4 for activation under a cycling electrode potential from −1.0 to 1.0 V at a scan rate of 100 mV s−1 until stabilization.

Scheme 1
scheme 1

Schematic illustration of the preparation procedure of electrochemical immunosensors for CEA

Synthesis of the Cu2O-Au nanocomposite

The octahedral Cu2O-Au nanocomposite was synthesized following a previously reported method with minor modifications [28, 29]. 2 mg of the Cu2O-Au nanocomposites was dispersed in 1 mL ultrapure water, and the suspension was sonicated for 30 min. The solution of the Cu2O-Au nanocomposite (5 μL, 0.2 mg·mL−1) was dropped onto the GCE surface and air-dried.

Antibody-oriented immobilization

SPA (5 μL, 3 μg·mL−1) and anti-CEA (5 μL, 0.2 μg·mL−1) were sequentially dropped on the Cu2O-Au nanocomposite-modified electrodes and incubated at 37 °C for 1 h. Then, a BSA solution (1%,w/v) was dropped onto the modified electrodes and incubated at 37 °C for 1 h to block the remaining active sites and eliminate non-specific binding effects. Ultimately, CEA at different concentrations was dropped on the prepared immunosensors. The electrodes were washed with PBS after every modification step to remove unbound SPA, anti-CEA, BSA and CEA. After all the steps, the electrodes were ready for measurement.

Electrochemical measurement

All electrochemical measurements were performed in a conventional electrochemical cell using a three-electrode system at room temperature. The immunosensor was immersed in the [Fe(CN)6]3−/4- solution (3 mL, 5 mmol·L−1) containing 0.1 M KCl for CV and EIS analysis. CV measurements were performed at a scan rate of 100 mV s−1 over a voltage range of −0.2 to 0.6 V. The EIS measurements were executed with a frequency sweep from 0.1 to 10^5 Hz and a 10 mV amplitude. The immunosensor was immersed in 8 mL phosphate buffered saline (PBS) (pH 6.4) for chronoamperometry analysis, the amperometric i-t curves were recorded at −0.4 V (vs.Ag/AgCl), 10 μL, H2O2 solution (15%,v/v) was added to PBS after the background current was stable, and the electrochemical signal changes were recorded.

Results and discussion

Choice of materials

In order to obtain excellent performance, the following materials were carefully selected for immunosensor construction. The octahedral Cu2O nanocrystals were used as precursor due to their excellent electrcatalytic performance toward H2O2, and the unique octahedral shape provides a larger surface area to immobilize biomolecules and better electrocatalytic ability. The Cu2O-Au nanocomposite selected as immunosensing substrate for immobilization of SPA, by considering that metallic alloy nanomaterials have better catalytic properties than monometallic counterparts, and the strong and stable electrochemical signal in the chronoamperometry study was derived from the catalysis of H2O2 by the Cu2O-Au nanocomposite. On this constructed immunosensor, SPA was adopted for oriented immobilization of antibodies to ensure antibodies being anchored on substrate surface through their Fc portion and thus the binding sites of antibodies remaining free and accessible for binding with antigens.

Characterizations of different nanomaterials

The size and morphology of the as-prepared nanomaterials were measured experimentally using FE-SEM. As Fig. 1a shows, most of the Cu2O nanocrystals present an obvious octahedral shape with an average diameter of approximately 250 nm. In addition, Fig. 1b shows that the average diameter of the AuNPs is approximately 30–50 nm and that they evenly covered the surface of the octahedral Cu2O nanocrystals. This indicates that we have successfully synthesized the Cu2O-Au nanocomposite with an octahedral structure.

Fig. 1
figure 1

FE-SEM image of the Cu2O nanocrystals (a) and FE-SEM image of the Cu2O-Au nanocomposites (b). EDS of the Cu2O-Au nanocomposites (c)

To further confirm the successful synthesis of the Cu2O-Au nanocomposite, EDS was used to investigate the chemical composition. As seen from the EDS image (Fig. 1c), there are significant peaks corresponding to Cu2O and Au, and agree with previous reports [28, 29]. In conclusion, these results demonstrate that the Cu2O-Au nanocomposites were successfully synthesized.

Characterization of the stepwise-modified electrodes

CV and EIS measurements were used to characterize the interface properties of the modified electrodes. In the CV scan of bare GCE (Fig. 2a, curve a), an obvious reversible redox peak was recorded. With the immobilization of Cu2O-Au nanocomposites on the electrode, the redox peak visibly increased (curve b), which indicates that the Cu2O-Au nanocomposites enhanced the effective electrochemical surface and promoted electron transfer. Then, the redox peak decreased after SPA was incubated on the immobilized electrode surface (curve c), as the non-electroactive SPA hinders electron transfer on the electrode. Next, the redox peak further declined after anti-CEA was captured (curve d), as the non-electroactive anti-CEA further increases the resistance of the electrode surface. Similarly, when the electrode was blocked by BSA, the redox peak obviously decreased (curve e). Finally, after the addition of CEA, the redox peak was reduced to a minimum because of the high-sensitivity specific binding between anti-CEA and CEA (curve f). In addition, the EIS (Fig. 2b) results agree with the CV results. In the EIS spectra, the semicircle diameter verifies assembly, which leads to a change in the electron transfer resistance (n = 3). Both the CV and EIS results support the successful construction of an immunosensor.

Fig. 2
figure 2

a CV and (b) EIS of modified electrode recorded in 5 mM [Fe(CN)6]3−/4- solution containing 0.1 M KCl: (a) bare GCE, (b) Cu2O-Au/GCE, (c) SPA/Cu2O-Au/GCE, (d) anti-CEA/SPA/Cu2O-Au/GCE, (e) BSA/anti-CEA/SPA/Cu2O-Au/GCE, and (f) CEA/BSA/anti-CEA/SPA/Cu2O-Au/GCE

Optimization of the experimental conditions

The sensitivity, selectivity, and electrochemical signal of immunosensors are influenced by many experimental parameters. To enhance the performance of the assay, the following parameters were optimized: (a) concentration of the Cu2O-Au nanocomposites, (b) concentration of SPA, (c) concentration of anti-CEA, (d) incubation time for anti-CEA, and (e) pH value of the supporting electrolyte. The results are presented in the Electronic Supporting Material (Fig. S1). The following experimental conditions yielded the optimal results: (a) Cu2O-Au nanocomposite concentration of 0.2 μg·mL−1, (b) SPA concentration of 0.3 μg·mL−1, (c) anti-CEA concentration of 0.2 μg·mL−1, (d) anti-CEA incubation time of 60 min, and (e) a pH value of 6.4.

Performance of the assay

The sensitivity and quantitative range were evaluated by amperometric i-t curves and quantitative detection with different concentrations of CEA under the optimal experimental conditions, at a typical working voltage of −0.4 V (vs. Ag/AgCl), the amperometric i-t curves were recorded by adding 10 μL of H2O2(15%, v/v) to 8 mL PBS (pH 6.4) at 100 s. As shown in Fig. 3a, the relationship between the electrochemical signal and the logarithmic values of the CEA concentration have strong linear relationship in the range of 2 pg·mL−1 to 20 ng·mL−1 with a lower detection limit of 200 fg·mL−1 (based on the 3σ criterion). The regression equation is y = −25.055lgCCEA + 68.207 with a correlation coefficient of R = 0.9979. Fig. 3b intuitively shows the relationship between the electrochemical signal with respect to the reduction of H2O2 and CEA concentration. The results indicate that the electrochemical immunosensor can be used to quantitatively detect CEA. In addition, as shown in Table 1, compared with other existing CEA detection methods, the designed immunosensor has wider linear range and lower detection limit.

Fig. 3
figure 3

a The linear relationship between the current response and CEA concentration (n = 3) in the range of 0.002–20 ng·mL−1. b Amperometric response of the immunosensor to CEA concentrations (n = 3) of (a) 0.002 ng·mL−1, (b) 0.005 ng·mL−1, (c) 0.01 ng·mL−1, (d) 0.05 ng·mL−1, (e) 0.1 ng·mL−1, (f) 0.5 ng·mL−1, (g) 1 ng·mL−1, (h) 10 ng·mL−1, and (i) 20 ng·mL−1. c Amperometric response to CEA (0.5 ng·mL−1) compared with four nonspecific interferents including ascorbic acid (0.5 ng·mL−1), uric acid (0.5 ng·mL−1), dopamine (0.5 ng·mL−1), tryptophan (0.5 ng·mL−1), a blank sample and a mixture with above four nonspecific interferents (n = 3). d Electrochemical signal response for the detection of CEA (0.01 ng·mL−1) at different weeks (n = 3)

Table 1 Comparison of linearities and detection limits of the newly designed immunosensor and other CEA detection methods

Specificity, stability and repeatability of the electrochemical immunosensor

High specificity, stability and repeatability of the signal response are crucial for the performance of the electrochemical immunosensor. We confirmed the specificity of the designed immunosensor by comparing the response to CEA (0.5 ng·mL−1) with the response to four nonspecific interferents, such asascorbic acid (0.5 ng·mL−1), uric acid (0.5 ng·mL−1), dopamine (0.5 ng·mL−1), tryptophan (0.5 ng·mL−1) a mixture with above four nonspecific interferents and a blank sample. As Fig. 3c shows, there was no significant difference in the responses of the blank signal and the nonspecific interferents. However, the electrochemical signal obviously changed in the presence of CEA. These results indicate the high specificity of the electrochemical immunosensor.

The stability of the immunosensor was investigated by long-term storage. The fabricated immunosensors were stored at 4 °C before testing. As seen in Fig. 3d, there were no obvious changes after the first week the signals changed less than 3.7%. After 4 weeks, the electrochemical signal retained 87.13% of the initial signal, which indicated that the immunosensor had good stability.

The reproducibility was evaluated by recording the current response of three electrodes prepared at the same concentration. As shown in the Electronic Supporting Material (Table. S1), at three different concentrations, the relative standard deviation (RSD) ranged from 1.4% to 4.8%.

Application for the analysis of the serum samples

Ten-fold-diluted human serum samples (obtained from University-Town Hospital of Chongqing Medical University) were mixed with different CEA concentrations of 0.005, 0.05, 0.5 and 10 ng·mL−1, and amperometric i-t curves of the obtained samples were recorded. As shown in Table 2, the RSD ranged from 0.82% to 4.49%, and the recovery rates ranged from 96.8% to 104.4%.

Table 2 Recovery of CEA in human serum (n = 3) with the proposed immunosensor

Conclusions

In summary, an ultrasensitive electrochemical immunoassay for CEA was designed by using a signal-amplifying substrate composed of octahedral Cu2O nanocrystals functionalized with AuNPs. On one hand, the larger surface and excellent biocompatibility of the Cu2O-Au nanocomposite make them ideal nanocarriers to immobilize large amounts of SPA, which further enhances the antibody capture efficiency. On the other hand, the Cu2O-Au nanocomposites have outstanding electrocatalytic activity toward the reduction of H2O2 and can amplify the electrochemical signal and improve the sensitivity. Due to the above advantages, the assay exhibits high sensitivity, good repeatability, specificity and can be used for the detection of CEA in human serum samples. We hope this electrochemical immunosensor can be considered a valid alternative for CEA determination in clinical and diagnostic applications. However, there is a limitation that the method involves several time-consuming steps led to the whole analysis takes 3.5 h, which will be addressed in future work.