Abstract
We are describing immunochromatographic test strips with smart phone-based fluorescence readout. They are intended for use in the detection of the foodborne bacterial pathogens Salmonella spp. and Escherichia coli O157. Silica nanoparticles (SiNPs) were doped with FITC and Ru(bpy), conjugated to the respective antibodies, and then used in a conventional lateral flow immunoassay (LFIA). Fluorescence was recorded by inserting the nitrocellulose strip into a smart phone-based fluorimeter consisting of a light weight (40 g) optical module containing an LED light source, a fluorescence filter set and a lens attached to the integrated camera of the cell phone in order to acquire high-resolution fluorescence images. The images were analysed by exploiting the quick image processing application of the cell phone and enable the detection of pathogens within few minutes. This LFIA is capable of detecting pathogens in concentrations as low as 105 cfu mL−1 directly from test samples without pre-enrichment. The detection is one order of magnitude better compared to gold nanoparticle-based LFIAs under similar condition. The successful combination of fluorescent nanoparticle-based pathogen detection by LFIAs with a smart phone-based detection platform has resulted in a portable device with improved diagnosis features and having potential application in diagnostics and environmental monitoring.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
Constant improvements are being made in diagnostic tests to obtain sensitive, specific, user friendly, low cost and reliable methodology under field conditions in resource limited situations [1]. Simple immunoassay protocols have become increasingly common at resource – limited sites, particularly in the form of lateral flow immunoassay (LFIA) for testing pregnancy, drug of abuse, diagnosis of various diseases and environmental monitoring [2]. However, the inherent limitation of traditional LFIA towards pathogen detection in terms of low sensitivity has necessitated research in the development of alternative labels to improve the detection limit of the assay [3]. In the past, several methods with different fluorescence labels have been tested to increase the sensitivity [4–7]. However, the direct use of flurophores has certain limitations such as photo bleaching, pH-dependence of the fluorescent signal or limited fluorescence intensity. To overcome these limitations, this study employs successful use of dye doped silica nanoparticles (SiNPs) as signaling probes which gives strong signal and have improved stability. In addition to this, wide compatibility of silica with biological materials and ease in biofunctionalization [8] render them ideal candidate for such applications. To note, manual interpretation of LFIA signal could be challenging for the untrained users who attempt to interpret the results. Hence, LFIA demands the development of a simple user friendly device to fully utilize its benefits. Earlier attempts to carryout quantitative detection of conventional LFIA signal have led to the use of special kind of reflectance photometer, necessitating skilled manpower and complex instrumentation [9]. It is pertinent to note that, over the past few years there has been consistent effort by various research groups [10–12] and commercial companies (Urisys 2400 and Miditron, Roche Diagnostics) to develop specialized assays and dipstick devices which led to increased cost and time. To address this gap, the available smart phone applications are utilized for pathogen and various biomarker diagnoses [13–17]. Smartphone cameras have been previously shown to be a portable microscope for monitoring pathogens, inexpensive spectrophotometer for educational purpose [18, 19]. Ozcan research group has demonstrated a prototype diagnostic test device on android platform for rapid detection of gold nanoparticles signal in LFIA [20]. In this scenario, the work presented here compliment the utilization of smart phone for interpretation and quantification of fluorescence LFIA signal using simple optical module attached to the camera of the smart phone. A novel approach was introduced combining sensitivity of fluorescence LFIA assay with portability of smart phone based device for point – of – care detection of food borne pathogen.
Experimental
Bacterial strains and reagents
The Salmonella spp. Escherichia coli, Klebsiella pneumoniae, Citrobacter freundii, Proteus mirabilis and Pseudomonas aeruginosa reference strain was obtained from Microbial Type Culture Collection, India. The 10 strains of Salmonella spp. isolated from typhoid fever patients were obtained from Sundaram Medical Foundation hospital laboratory. The clinical isolate of pathogenic Escherichia coli O157 was obtained from Christian Medical College, Vellore, India. All the organisms were cultured on nutrient broth at 37 °C, 195 rpm for 16–18 h. All the reagents were of analytical grade and purchased from Sigma (www.sigmaaldrich.com).
Synthesis of dye doped silica nanoparticles
The FITC doped SiNPs were prepared by Stoeber’s method [21]. Briefly, about 6 mg of FITC was dissolved in 2 mL of ethanol and 160 μL of 3-aminopropyltrimethoxysilane was added under magnetic stirring. To that was added 150 μL of tetraethyl orthosilicate, distilled water (1 M) and the mixture was made to 20 mL with ethanol followed by addition of 50 μL of ammonium hydroxide. Reaction was allowed to proceed overnight at room temperature under magnetic stirring resulting in silica encapsulated FITC nanoparticles. The obtained nanoparticles were washed several times with ethanol, vacuum dried and stored at room temperature. The Ru(bpy) doped SiNPs were synthesized by microemulsion method [22]. Briefly, 7.5 mL of cyclohexane was added to 1.8 mL of n-hexanol and 1.77 mL of Triton X-100 followed by addition of 480 μL of 20 mM Rubpy dye in water and mixed vigorously to form a microemulsion. About 100 μL of tetraethyl orthosilicate was added to above mixture under magnetic stirring. The reaction mixture was stirred continuously for 20 min followed by addition of 60 μL of 30 % ammonium hydroxide, and stirring continued for next 24 h at room temperature. After this period, an equal amount of acetone by volume of the reaction mixture was added and vortexed vigorously to break the microemulsion. The solution was centrifuged at 3,000 xg for 10 min and the particles obtained were washed several times with vortex and sonication in 95 % ethanol. The final particles were dried in vacuum and stored at room temperature for later use. The particles were characterization by SEM, fluorescence spectroscopy and FT – IR spectroscopy (Supplementary data).
Biofunctionalization of nanoparticles
The anti – Salmonella monoclonal antibody was developed against whole cell Salmonella antigen and conjugated to dye doped SiNPs as described previously by our group [23]. About 30 mg of nanoparticles were reacted with 20 mL of 1 % aminopropyl trimethoxysilane in 1 mM acetic acid under continuous mechanical stirring for 30 min. The particles were allowed to react with 10 % succinic anhydride in dimethylformamide for overnight at room temperature under nitrogen atmosphere. After incubation, the obtained carboxyl modified dye doped SiNPs were washed three times with water. Then 1 mL of 1 mg.mL−1 1-Ethyl-3-(3-dimethylaminopropyl)-carbodiimide and N- Hydroxysuccinimide in water was added and incubated at room temperature for 30 min. The activated nanoparticles were then immobilized with 30 μg of anti - Salmonella monoclonal antibody for 24 h at 4 °C in a rotary shaker. After washing thrice with PBS buffer, the free N-Hydroxysuccinimide esters were blocked by adding 40 mM ethanolamine and 1 % Bovine serum albumin in 1 mL of PBS for 2 h at room temperature. The resulting antibody conjugated dye doped SiNPs were resuspended in 1 mL of phosphate buffered saline containing 0.1 % Bovine serum albumin at 4 °C until further use. Similar protocol was followed for conjugation of Escherichia coli O157 polyclonal antibody (Kirkegaard & Perry Laboratories Inc, www.kpl.com) to the dye doped SiNPs. The preparation of antibody conjugated gold nanoparticles was carried out on 20 nm colloidal gold nanoparticles (Sigma, www.sigmaaldrich.com) using the previously described procedure [24].
Assembly of test strip
The test strip (75 × 3 mm dimension) containing nitrocellulose membrane with pore size of 10 μm, absorbent pad, conjugate pad and sample pad required for dipstick preparation were purchased from Advanced Microdevices Pvt Ltd (www.mdimembrane.com). Initially, the capture antibody was immobilized in the designated test region and the control region of the nitrocellulose membrane [24]. LFIA relies on the migration of liquid across the surface of the nitrocellulose membrane. The bacteria present in the test sample reacts with the antibody conjugated nanoparticles in the conjugate pad which is placed beneath the sample pad and the immune complex were trapped in the test line immobilized with capture antibody to result in the fluorescent signal. The free antibody conjugated nanoparticles further move and gets trapped in the control line containing the anti-goat IgG for Escherichia coli O157 test strip and anti- mouse IgG for Salmonella test strip. The control line was used to confirm active migration across the membrane. The fluorescence signal caused by the accumulation of dye doped silica nanoparticles in the test line and control line provided the basis for detection of the pathogen. For immobilization in the test strip, 100 μL (1 mg.mL−1) of antibody was mixed with 8 μL of 20 % sucrose solution prepared with 50 mM potassium dihydrogen phosphate buffer (pH 7.5) and 8 μL of isopropanol. About 1 μL spot of Salmonella polysera and goat anti – mouse IgG (Bangalore Genei, www.bangaloregenei.com) was used for immobilization in the test line and the control line respectively for preparation of Salmonella LFIA strip. After immobilization of antibody, the membrane is blocked with 1 % Bovine serum albumin in 50 mM boric acid buffer (pH 8.5) for 30 min at room temperature to prevent non-specific binding. After incubation, the membrane was washed twice with PBS containing 0.05 % Tween 20. The test strips were allowed to dry at room temperature for 12 h. About 2 μL of antibody conjugated dye doped SiNPs was added to the conjugate pad and allowed to dry for 5 min. The adsorbent pad and the conjugate pad were mounted on an adhesive coated backing sheet. The prepared LFIA strip was stored in air tight container at 4 °C until use. Similarly, Escherichia coli O157 polyclonal antibody and rabbit anti-goat IgG was immobilized in the test line and control line respectively for preparation of Escherichia coli O157 LFIA strip.
Construction of fluorescence device
To enable the utility of fluorescence LFIA in low-resource settings, a simple optical design (Fig. 1a) was assembled. It consists of 3.7 V battery powered LED light source (Osram, in.rsdelivers.com) for illumination, a fluorescence filter cube (Chroma, www.chroma.com), a plano convex lens (Thorlabs, www.thorlabs.us) of 50 mm focal length and a sample tray for inserting the test strip. The band pass filter (460–490 nm) allows excitation of both dyes and also eliminate the high wavelength tail of the LED emission. The barrier filter (520 nm) enable simultaneous observation of fluorescence signals from both dye doped SiNPs without the need to change the filter set combinations, allowing multiplex detection. However, the module is designed to enable easy exchange of the fluorescence filters if required. The optical system is enclosed in a light proof plastic box to ensure improved signal to noise ratio by reducing ambient light. The entire optical module weighing ~ 40 g (Fig. 1b) is attached in front of a camera (5 mega pixel) of a smart phone (Sony Live with Walkman) to capture the fluorescence signal from LFIA strip. As observed in Fig. 1c, selected region of interests were analyzed using smartphone resident application which exhibited fluorescence intensity maxima corresponding to the test and control spots of the LFIA strip.
Fluorescence lateral flow immunoassay
The overnight grown cells were serially diluted in sterile PBS to attain 107– 104 cfu.mL−1. The samples were centrifuged at 8,000 xg for 2 min and resuspended in 200 μL of sterile PBS. The 200 μL was determined to be the optimum running volume required for the assay. For evaluation, all the samples were taken in 96 well microtiter plates. The test strip was placed sample pad end down into the well allowing at least 10 min for adequate absorption and migration of the liquid along the strip. The strips were then inserted into the strip tray of the smartphone attached fluorescence LFIA device for image capture. The reproducibility of the assay was analysed by performing the test in triplicates. Subsequently, the assay was performed on 10 clinical isolates of Salmonella spp. to demonstrate the applicability in clinical diagnosis. The fluorescence LFIA results were further compared with conventional gold nanoparticles based LFIA and the sensitivity of the assay was evaluated. The specificity of the assay was determined against Escherichia coli O157, Pseudomonas aeruginosa, Proteus mirabilis, Citrobacter freundii and Klebsiella pneumoniae. Adopting the same procedure the detection limit and cross reactivity of the Escherichia coli O157 LFIA strip was determined against Escherichia coli, Klebsiella pneumoniae, Salmonella spp, Citrobacter freundii, Proteus mirabilis and Pseudomonas aeruginosa.
Image processing
The total intensity of the images obtained from the fluorescence LFIA strip was analyzed using an application in smart phone. The images were processed for the region of interest covering the test and control spot using Photo Editor Ultimate, downloaded from Google Play [25]. The processed images were saved and recalled in IJ_ mobile application [26] installed in the smart phone for analysis. The application provides the graphical readout of the total intensity profile along the region of interest. For comparison, the captured fluorescence images were also copied to a laptop and independently analyzed using ImageJ (Supplementary data). A rectangular region of interest was placed covering both the test and control spot to obtain the integrated fluorescence intensity using ImageJ.
Result and discussion
The organic dye FITC and the inorganic dye Rubpy doped silica nanoparticles were synthesized by Stoeber’s and microemulsion methods respectively. The synthesized dye doped SiNPs were analysed by SEM to study the morphology and size of the nanoparticles. As observed in Fig. 2a and b, well dispersed and spherical nanoparticles in the size range of ~ 150 nm and ~100 nm for FITC and Ru(bpy) doped SiNPs respectively were obtained. Spectrofluorometric measurements of the dye doped SiNPs show emission at 525 nm for FITC doped SiNPs and 625 nm for Ru(bpy) doped SiNPs upon excitation at 475 nm in aqueous solution (Supplementary data).
Salmonella detection
The affinity of the antibody is essential for efficient binding of the cell–SiNPs immunoconjugate in the LFIA strip. The nanoparticles conjugated with mouse monoclonal antibody to Salmonella spp. were incubated with the Salmonella cells. The SEM image shows that nanoparticle conjugates adhered to the surface of Salmonella spp. (Fig. 3a and b). Thus the developed antibody was found specific to the surface antigens of the whole cells. In addition, numerous dyes doped SiNPs were found bound to single bacterium which help improve the sensitivity of detection.
To determine the detection limit of the assay, the prepared fluorescent LFIA strips were tested on known concentration of Salmonella spp. and the signal was obtained through optical module attached to the smart phone. The red/green fluorescence signal could be detected by visual inspection of the image obtained from the smart phone device. Subsequently, the images of the fluorescence LFIA strips were analyzed using the smartphone resident application that readout the total intensity along the region of interest. As shown in Fig. 4, significant intensity peak was recorded from the test spot for sample containing ≥105 cfu.mL−1 of bacteria. Consistent results were obtained during repeated screening (n = 3). The fluorescence LFIA strip can detect the pathogens in the sample within few minutes using the simple attachment to the smart phone. Thus the smart phone based fluorescence diagnosis can be a simple, user friendly screening tool in the hands of a end user. Further, the captured fluorescence images were copied on to a laptop and the total intensity along the width of the spot was measured using ImageJ software for comparison. Observation of similar intensity profile (Supplementary data) validates the reliability of smart phone based measurements. The optimized assay was then validated on 10 independent clinical isolates of Salmonella spp. using the similar protocol. Successful detection was observed for all the 10 samples when assayed at the concentration of 105 cfu.mL−1.
Further, the specificity of Salmonella fluorescence LFIA strips was determined against Escherichia coli O157, Pseudomonas aeruginosa, Proteus mirabilis, Citrobacter freundii and Klebsiella pneumoniae at 105 cfu.mL−1. There was no significant fluorescence observed in the test spot proving zero cross reactivity of the assay. As observed in Fig. 5, the Escherichia coli O157 LFIA also showed successful detection of Escherichia coli O157 from samples containing ≥105 cfu.mL−1 and zero cross reaction was observed when tested against six negative strains (Supplementary data). The above results convincingly prove that the fluorescence LFIA strip can constitute a sensitive and specific diagnostic platform for bacterial pathogens.
To demonstrate the improved detection limit of fluorescence LFIA strips, conventional gold nanoparticles based LFIA assay was performed under similar setting. A visible red spot could be seen in the LFIA strips for sample containing ≥106 cfu.mL−1 of Salmonella spp. and Escherichia coli O157 (Supplementary data). Despite the gold nanoparticles based LFIA strip is being simple and the results can be readily read with the naked eye, they are one order of magnitude less sensitive in detection. Usually, sensitivity in gold nanoparticle assay is increased by silver enhancement, aptamers or modified assay probes. However, this leads to increased cost and time [27, 28]. Multiplex PCR has been reported for rapid and sensitive detection of food borne pathogens [29]. However, a short pre-enrichment of the test samples are needed for reliable assay performance and to enable direct sample detection [30]. The developed fluorescence nanoparticles based LFIA could detect pathogen at levels of 105 cfu.mL−1 without pre – enrichment of the samples. It has been reported that ingestion of 105 organisms cause enteric fever in healthy individuals [31]. The developed assay is well within this limit and hence can be a potential and rapid screening tool for detection of bacterial pathogens directly from the test sample to ensure food safety. The study demonstrates that both organic and inorganic dyes can be utilized as probes for detection of pathogens. The dye leakage from the silica nanoparticles is prevented by covalent attachment of the organic dye like FITC to the silica matrix by using the FITC-APTES as precursor during synthesis of dye doped SiNPs by using Stoeber’s method. To note, for the synthesis of inorganic dye like Ru(bpy) doped SiNPs, which possesses a mainly hydrophilic nature, the reverse microemulsion method is used to enable a more efficient dye encapsulation without dye leakage during subsequent handling of the nanoparticles. This increased sensitivity of bacterial detection by fluorescence LFIA strip is due to the fact that (i) very high ratio of dye are encapsulated inside the silica shell and (ii) numerous antibody conjugated nanoparticle probes bind to the surface of the single bacterium. The strong fluorescence signal from dye doped SiNPs can be easily excited with a simple diffused LED and captured using a smart phone or digital camera [32]. The affinity of the antibody determines the variability of the assay on different test samples. The monoclonal and polyclonal antibody may behave differently. However, in accordance to our assay consistent results were observed on repeated (n = 3) screening. Thus, our smart phone based fluorescence LFIA device satisfies the user friendly and portable screening applications without compromising on the sensitive detection.
Researchers have developed a rapid diagnostic test device to process the raw image of the gold nanoparticles based LFIA strip using a dedicated application running on the smart phone [16]. Recent reports also demonstrate a low cost fluorescent lateral flow assay [33, 34]. There have been efforts to establish computer screen shot assisted sensing using web/digital cameras for fluorescence biosensing [35]. In this work, we successfully demonstrate for the first time, the use of smart phone for both fluorescence image capture and its analysis. This can avoid the need of laptop and other computing devices for further interpretation of data. Thus, by integration of diagnostic test procedure with mobile phone an end-to-end solution for point – of – care detection in the hands of healthcare worker and clinicians has been evolved. This device can be adapted in various clinical testing applications which require rapid, bedside detection of pathogenic bacteria by simple customization of the module. These promising results form the basis of future diagnostics and create new opportunities in the examination of food, clinical and environmental samples
Conclusion
The specificity of antibodies was combined with the high signal enhancement and ease of surface functionalization offered by dye doped SiNPs to result in enhanced detection capabilities of conventional LFIA strips. Utilizing a simple smartphone based device, a reliable detection of pathogens directly in test sample at levels of 105 cfu.mL−1 has been demonstrated. This qualitative analysis of the test strip using a common smartphone may be of more utility to a clinician and affordable for routine use in developing countries.
References
Urdea M, Penny LA, Olmsted SS, Giovanni MY, Kaspar P (2006) Requirements for high impact diagnostics in the developing world. Nature 23:73–79
Posthuma-Trumpie GA, Korf J, Jvan Amerongen A (2009) Lateral flow (immuno) assay: its strengths, weaknesses, opportunities and threats. A literature survey. Anal Bioanal Chem 393:569–582
Pyo DJ (2007) Comparison of fluorescence immunochromatographic assay strip and gold colloidal immunochromatographic assay strip for detection of microcystin. Anal Lett 40:907–919
Zaytseva NV, Montagna RA, Lee EM, Baeumner AJ (2004) Multi-analyte single-membrane biosensor for the serotype-specific detection of Dengue virus. Anal Bioanal Chem 380:46–53
Khreich N, Lamourette P, Boutal H, Devilliers K, Créminon C, Volland H (2008) Detection of Staphylococcus enterotoxin B using fluorescent immunoliposomes as label for immunochromatographic testing. Anal Biochem 377:182–188
Yang H, Li D, He D, Guo Q, Wang K, Zhang XQ, Huang P, Cui DX (2010) A novel quantum dots-based point of care test for syphilis. Nanoscale Res Lett 5:875–881
Juntunen E, Myyryläinen T, Salminen T, Soukka T, Pettersson K (2012) Performance of fluorescent europium (III) nanoparticles and colloidal gold reporters in lateral flow bioaffinity assay. Anal Biochem 428:31–38
Cho EB, Volkov DO, Sokolov I (2011) Ultrabright fluorescent silica mesoporous silica nanoparticles: control of particle size and dye loading. Adv Funct Mater 2:3129–3135
Penders J, Fiers T, Giri M, Wuyts B, Ysewyn L, Delanghe JR (2005) Quantitative measurement of ketone bodies in urine using Reflectometry. Clin Chem Lab Med 43:724–729
Zeng Q, Mao X, Xu H, Wang S, Liu G (2009) Quantitative immunochromatographic strip biosensor for the detection of carcinoembryonic antigen tumor biomarker in human plasma. Am J Biomed Sci 1:70–79
Liu G, Lin Y, Wang J, Wu H, Wai CM, Lin Y (2007) Disposable electrochemical immunosensor diagnosis device based on nanoparticles probe and immunochromatographic strip. Anal Chem 79:7644–7653
Mao X, Baloda M, Gurung AS, Lin Y, Liu G (2008) Multiplex electrochemical immunoassay using gold nanoparticles probes and immunochromatographic strips. Electrochem Commun 10:1636–1640
Smith ZJ, Chu K, Espenson AR, Rahimzadeh M, Gryshuk A (2011) Cell-phone-based platform for biomedical device development and education applications. PLoS ONE 6:e17150
Choodum A, Kanatharana P, Wongniramaikul W, Nic Daeid N (2013) Using the iPhone as a device for a rapid quantitative analysis of trinitrotoluene in soil. Talanta 15:143–149
Wei Q, Qi H, Luo W, Tseng D, Ki SJ, Wan Z, Gorocs Z, Bentolila LA, Wu TT, Sun R, Ozcan A (2013) Fluorescent imaging of single nanoparticles and viruses on a smart phone. ACS Nano 7:9147–9155
Coskun AF, Nagi R, Sadeghi K, Phillips S, Ozcan A (2013) Albumin testing in urine using a smart-phone. Lab Chip 7:4231–4238
Zhu H, Sencan I, Wong J, Dimitrov S, Tseng D, Nagashima K, Ozcan A (2013) Cost-effective and rapid blood analysis on a cell-phone. Lab Chip 13:1282–1288
Wang SX, Zhou XJ (2008) Spectroscopic sensor on mobile phone. U.S. Patent 7420663
Breslauer DN, Maamari RN, Switz NA, Lam WA, Fletcher DA (2009) Mobile phone based clinical microscopy for global health applications. PLoS ONE 4:e6320
Mudanyali O, Dimitrov S, Sikora U, Padmanabhan S, Navruz I, Ozcan A (2012) Integrated rapid-diagnostic –test reader platform on a cellphone. Lab Chip 12:2678–2686
Wang L, Tan W (2006) Multicolor FRET silica nanoparticles by single wavelength excitation. Nano Lett 6:84–88
Santra S, Wang K, Tapec R, Tan W (2001) Development of novel dye-doped silica nanoparticles for biomarker application. J Biomed Opt 6:160–166
Padmavathy B, Vinoth Kumar R, Uttara S, Suvro C, Jaffar Ali BM (2013) Immunomagnetic nanoparticles based quantitative PCR for rapid detection of Salmonella. Microchim Acta 180:1241–1248
Tanaka R, Yukhi T, Nagatani N, Endo T, Kerman K, Takamura Y, Tamiya E (2006) A novel enhancement assay for immunochromatographic test strips using gold nanoparticles. Anal Bioanal Chem 385:1414–1420
https://play.google.com/store/apps/details?id=com.icecoldapps.photoeditorultimatefree. Accessed 18 Nov 2013
https://play.google.com/store/apps/details?id=com.ij_mobile. Accessed 18 Nov 2013
Shen G, Xu H, Gurung AS, Yang Y, Liu G (2013) Lateral flow immunoassay with the signal enhanced by gold nanoparticle aggregates based on polyamidoamine dendrimer. Anal Sci 29:799–804
Wang H, Zhang C, Xing D (2011) Simultaneous detection of Salmonella enterica, Escherichia coli O157:H7, and Listeria monocytogenes using oscillatory – flow multiplex PCR. Microchim Acta 173:503–512
Lui G, Xiaofeng Y, Xue F, Chen W, Ye Y, Yang X, Lian Y, Yan Y, Zong K (2012) Screening and preliminary application of a DNA aptamer for rapid detection of Salmonella O8. Microchim Acta 178:237–244
Roda A, Mirasoli M, Roda B, Bonvicini F, Colliva C, Reschiglian P (2012) Recent developments in rapid multiplexed bioanalytical methods for foodborne pathogenic bacteria detection. Microchim Acta 178:7–28
Ryan KJ, Ray CG (2004) Sherris medical microbiology: an introduction to infectious disease. McGraw-Hill, New York
Wang XD, Meier RJ, Wolfbeis OS (2013) Fluorescent pH-sensitive nanoparticles in an agarose matrix for imaging of bacteria growth and metabolism. Angew Chem Int Ed 52:406–409
Lee LG, Nordman ES, Johnson MD, Oldham MF (2013) A Low-cost, high-performance system for fluorescence lateral flow assays. Biosensors 3:360–373
Bamrungsap S, Apiwat C, Chantima W, Dharakul T (2013) Rapid and sensitive lateral flow immunoassay for influenza antigen using fluorescently – doped silica nanoparticles. Microchim Acta. doi:10.1007/s00604-013-1106-4
Filippini D, Alimelli A, Di Natale C, Paolesse R, D’Amico A, Lundstorm I (2006) Chemical sensing with familiar devices. Angew Chem 118:3884–3887
Acknowledgments
We acknowledge Central Instrumentation Facility, Pondicherry University for SEM imaging and FT-IR characterization. We thank Dr.T.Vaidehi, Sundaram Medical Foundation Hospital, Chennai for providing clinical isolates of Salmonella spp. and Dr. Gagandeep Kang, Christian Medical College, Vellore for providing Escherichia coli O157. BP thanks Council of Scientific and Industrial Research (CSIR) for the award of Senior Research Fellowship.
Author information
Authors and Affiliations
Corresponding author
Additional information
Vinoth Kumar Rajendiran and Padmavathy Bakthavathsalam equally contributed to this work.
Electronic supplementary material
Below is the link to the electronic supplementary material.
ESM 1
(PDF 621 kb)
Rights and permissions
About this article
Cite this article
Rajendran, V.K., Bakthavathsalam, P. & Jaffar Ali, B.M. Smartphone based bacterial detection using biofunctionalized fluorescent nanoparticles. Microchim Acta 181, 1815–1821 (2014). https://doi.org/10.1007/s00604-014-1242-5
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00604-014-1242-5