Introduction

Interbody cages with bone graft are currently successfully used for spinal fusion [2, 30]. Using cages help maintaining lordosis and neuroforaminal distraction and reduce donor site morbidity when compared to tricortical bone graft [24, 27, 31, 32, 51]. However, reported complications of the use of interbody cages include migration, extrusion, pseudarthrosis, and implant subsidence [21, 25, 51, 54]. Being permanent implants, intervertebral cages remain at the risk of secondary infection, possibly requiring difficult removal. In addition, titanium alloy and stainless steel cages have specific disadvantages including stress-shielding due to excessive rigidity, which leads to bone resorption and osteopenia [19, 43]; corrosion, wear debris due to slight motion before achievement of fusion; and possible allergic reactions [16]. Carbon fibers are radiolucent, MRI-compatible, and better match the elasticity modulus of the bone [9]. However, small debris particles have been reported with carbon fiber cages [38]. Cases of rupture of these cages have also been reported [41].

Bioresorbable polymer implants aim at avoiding these complications with self degrading once their mechanical role is over, after achievement of solid bone fusion [5]. Resorbable implants would also facilitate imaging for assessing fusion [28, 39, 47, 48] and allow progressive transfer of the load to the fusion mass [3]. Currently, the most common resorbable biomaterial used in orthopedic and trauma surgery is polylactic acid (PLA). With limited foreign body reaction [12], biocompatibility with neural tissue [46, 22], nerves [13], bone-like low elastic modulus [11], and possibility of association with osteoinductive molecules such as BMP [40], PLA devices represent a promising class of bioresorbable materials with widespread potential in spinal surgery. In addition, the incidence of infections was comparable between resorbable implants and usual metal implant devices [4].

PLA has two forms: poly-(l-lactide), which can be pure (PLLA form), and poly-(d-lactide), which is systematically associated with poly-(l-lactide) (PDLLA form). Their combination constitutes a stereocopolymers poly-(l,dl lactide) (PLDLLA), with varying ratios of each [50], usually 70/30. Kinetics of resorption and strength retention have been investigated [7, 18, 26, 33, 35, 42, 47]. PLDLLA are considered to degrade faster with more tissue reaction as compared with PLLA implants, for which the degradation process has been described to last for several years. The degree of the material’s crystallinity also interferes in determining the occurrence of late degradation [7, 29]. In addition, the biological (vascularization) and the mechanical environment at the implantation site are also important parameters [16].

The goal of this study was to evaluate a new PLDLLA cage, where the L/DL ratio was set at 96/4, in a large animal model for lumbar interbody fusion. The main hypothesis was that the increased proportion of the PLLA form, when compared to 70/30 PLDLLA, would be a good compromise for degradation and fusion kinetics at the lumbar spine. The sheep model was chosen because of similarities with human spine in term of lamellar bone growth rate and biomechanics [52]. Histological and radiological bone formation and implant degradation as a function of time were investigated.

Materials and methods

All operative procedures and the animal care complied with the regulations of the legislation for animal research. Sixteen Pre Alps ewes (7 months-old, and 42 kg in average) underwent a left retroperitoneal lumbar approach for single level implantation of the bioresorbable 96/4 PLDLLA cage (PHUSIS-FRANCE) at L3–L4 (Table 1). Muscles of the abdominal wall were opened longitudinally in lateral position. The peritoneum and all retroperitoneal elements were moved medially together for exposuring the lumbar spine. The complete discetomy preserved the posterior longitudinal ligament and the vertebral endplates were reamed until the subchondral bone. The parallelepipedal cage was 22.2 mm long, 9 mm wide, and 9 mm high at one end and 11 mm high at the other end (lordotic cage of 6° angle). The wall thickness was 2.5 mm. It was filled and surrounded with autologous cancellous bone from the iliac crest. The cage was implanted obliquely in order to preserve the common anterior vertebral ligament and to avoid anterior migration. Six groups of two or three animals were sacrificed at 3, 6, 9, 12, 24, and 36 months post-implantation. Oxytetracyclin (30 mg/kg, IV) was administered 15 days and 5 days before sacrifice for labeling bone apposition.

Table 1 96/4 PLDLLA cages characteristics

Harvested spinal blocks were X-rayed, and had CT scan (General Electric, volumic acquisition and rebuilding from 1 mm thickness cuts, matrix 512 × 512) and MRI (SIGNA 1 Tesla General Electric, T1 and Fast T2 Fat Sat sagittal and frontal, cuts 3 mm thickness with 0.3 mm gap. Matrix 512 × 512, Fov 20 cm) imaging. Definite fusion was defined as the presence of bone bridges on the CT scan.

The harvested specimens were fixed in 10% buffered formalin, dehydrated in a graded alcohol series, and embedded in polymethylmetacrylate. Two ventro-dorsal parasagittal ground sections were prepared from each specimen using the Donath method for the Exakt system [14]. Histological sections were stained using the modified Paragon method. Qualitative and semi-quantitative analyses of histological inter-body fusion, bone and cell density at the implantation site, degree of tissue differentiation in contact with the cage, mineralization, remodeling, cage resorption, and inflammatory parameters were performed. The assessment of the bone mineralization activity in the interbody space was performed using epifluorescence examination. Observed parameters were quoted 0 to 4 on each histological section (0 = absent, 1 = mild, 2 = moderate, 3 = marked, 4 = very marked or complete) and averaged for each group of animals. The histological analysis was performed according to the NF EN ISO 10993–6 recommendations [1].

Qualitative histopathological examinations (Safranin–Hematoxylin–Eosin-stain) of local and remote tissues (paravertebral muscles and lymph nodes, liver, kidney, and spleen) were performed to determine possible reactions for the degradation products and/or resorption of the material.

Results

All sheep recovered from anaesthesia uneventfully without signs of neurological deficits. Histological results are summarized in Table 2. Three months after implantation, the cages displayed morphological deformations showing compression of the internal space of the implant by 10–40%, slight undulation of the walls, and limited fragmentation of the material. There was no sign of significant material resorption. However, active peri-implant bone neoformation associated with fibro-cartilaginous tissue was observed. Early centripetal bone in growth ranging from 0 to 60% (Fig. 1) occurred in the inner space of the cage, although no complete bone fusion had taken place. The neuronal and muscle tissues close to the implantation site did not exhibit signs of lesions. Six months following implantation, osteogenesis and bone remodeling around the implants increased. The cages deformations were the same as previously observed and were associated with direct bone-to-implant. Signs of bone fusion were visible outside the implant in all cases. Limited signs of fusion were also observed within the cage. No inflammatory, necrotic, or osteolytic reactions were visible. Fluorescence labeling clearly showed signs of mineralization in direct contact with the cage (Fig. 2). Nine months after implantation, the implant material gradation developed actively. Macro- and micro-particles were visible among macrophages. Fusion occurred at the periphery of the cage in the two animals while noticeable bony growth was seen within the cage in one case. In one ewe, the cage was implanted too much dorsally with direct contact with the dura mater. No local significant inflammatory reaction was reported (Fig. 3). At 12 months post-implantation, deformation and signs of extensive degradation of the implant (fragmentations and erosions) were detected. Approximately 30% of the cage was resorbed and replaced by trabecular bone tissue with structural characteristics similar to those of physiological neighboring bone tissue. The fibro-cartilaginous peri-implant tissue was distinctly thinned. Osteointegration of the cage was associated with pronounced bony growth within the implant without signs of significant inflammatory reaction. Signs of complete interbody arthrodesis were seen within and outside the cage in one animal, and only outside the cage in two animals. At 24 months, resorption of the implant proceeded with more fragmentation. Approximately 40% of the device was resorbed and replaced by trabecular bony tissue. Numerous direct bone-material or fibro-cartilaginous contacts tissue were seen around the implant (Fig. 4). Fusion outside the cage was seen in all cases. Two of the three ewes displayed bone fusion inside and outside the implant. One cage, that was implanted too much dorsally, in direct contact with the dura mater, had a localized and limited granulomatous reaction. No signs of local intolerance were seen around the implant, including the neighboring muscle. At 36 months post-implantation, the two cases exhibited histological fusion in regions of the device, as well as in the periphery of the cage (Fig. 5). In one case, the cage was completely reabsorbed; it was replaced by a fibro-bone marrow tissue containing mesenchymal cells. In the other case, a few remants of the cage were seen, associated with a fibrovascular tissue and bony bridges occupied the space left by the reabsorbed implant. A few macrophages showing blue-stained cytoplasmic inclusions were visible. Within and outside the former cage, the trabecular bone tissue was mature and of similar density as compared to the rest of the vertebral body. Signs of cortical bone formation were seen in one case. Epifluorescence analysis demonstrated mineralization activity in the implanted site, similar to that seen in the rest of the vertebral body. No significant signs of local intolerance reaction were seen either on the implanted site or in the neighboring neuronal or muscle tissues.

Table 2 Histological results
Fig. 1
figure 1

Histological section at 3 months post-implantation. Original magnification 2×. Paragon staining. Sagittal section of the cage in the intervertebral position showing active bone neoformation

Fig. 2
figure 2

Histological section at 6 months post-implantation. Original magnification 10×. Epifluorescence labeling showing the mineralization front: the activity is characterized by direct apposition of bone in contact with the implant

Fig. 3
figure 3

Histological section at 9 months post-implantation. Original magnification 10×. Paragon staining. Peripheral fragmentation of the material

Fig. 4
figure 4

Histological section at 24 months post-implantation. Original magnification 10×. Paragon staining. Replacement of the material by bony tissue, direct bone apposition

Fig. 5
figure 5

Histological section at 36 months post-implantation. Original magnification 2×. Paragon staining. Complete intervertebral arthrodesis. The peri-implant bone trabecular bone density is comparable to the bone density of the adjacent vertebral bodies

Radiological results showed a continuous evolution toward fusion, which was following the histological curse. At 3 months post-implantation, radiological fusion was partial in all cases on standard radiographs; implants remained clearly visible in all cases. CT-scan section showed no bone bridges. At 6 months, radiological intervertebral fusion was definite in all cases on radiographs and CT-scan sections. No specific signal was detected on the MRI in the adjacent bone (Fig. 6). The radiological aspect was similar at 9 months after implantation. At 12 months, the radiological fusion was also definite in all cases in and outside the cage. This was associated with considerable resorption of the cage (Fig. 7). Radiological fusion was definite in all cases after 24 months with confirmed interbody fusion on the CT-scan (Fig. 8). At 36 months, fusion was complete on radiographs and CT scans, with calcification at the site of the degraded cage. The MRI aspect still showed a variation of signal at the location of the former cage; however, adjacent vertebral endplates were physiological (Fig. 9).

Fig. 6
figure 6

Imaging at 6 months post-implantation. Intrevetebral bone demonstrates interbody fusion on the radiograph (a) and CT-scan (b) and MRI (c)

Fig. 7
figure 7

CT-scan sections at 12 months post-implantation. Radiological aspect of interbody fusion is similar to the one at 6 months. a Sagittal view, b axial view

Fig. 8
figure 8

CT-scan sections at 24 months post-implantation. Signs of complete fusion with evidence of cortical bone ventrally. a Sagittal view, b axial view

Fig. 9
figure 9

Imaging at 36 months post-implantation. All the interbody space is fused with mature bone replacing the fully degraded 96/4 PLDLLA cage (a radiographs, b CT scan); the cage site still had an hydrophilic signal on the MRI (c) with no change of signal in peripheral tissue when compare to adjacent levels

Discussion

The current study investigated lumbar intervertebral fusion using 96/4 PLDLLA cages in an ovine model. In addition to the histological investigation, fusion and tissue response were assessed radiologically. We noted the obvious correspondence between histology and radiology and the absence of material related artifacts. We demonstrated that interbody fusion using 96/4 PLDLLA cages was rapidly achieved, starting around the cage. The “locked pseudarthrosis” area, described as the presence of fibrous tissue with collagen fibers in the fusion zone without enchondral bone formation, when using 70/30 PLDLLA polymer [39], was not encountered in our study. We did not observe a quiescent fibrous tissue layer surrounding the cages and persisting overtime in any of our cases. Direct bone apposition onto the implant was gradually observed. This is in accordance with some previous observations showing direct apposition of bone on the polymeric material with PLDLLA implants and a thin layer of connective tissue with PLLA [6, 23, 43]. By separately analyzing the inner and outer fusion, we observed that peripheral fusion came first, which is in accordance with the notion of radiological sentinel sign for intervertebral fusion [36]. It is likely that the microenvironment (blood supply) is more favorable around than within the cage. We acknowledge that placing autologous bone graft around the cage in addition to within the cage in the protocol made a difference with other animal studies (Table 3). However, the in and out graft is a common procedure in spinal surgery of the anterior column using autologous bone graft, and this had to be taken into account.

Table 3 Synoptic table for comparing histological results from in vivo studies concerning interbody cages in term of fusion

Concerning cage degradation and tissue reaction, we observed that the overall original geometry of the 96/4 PLDLLA cages was maintained at 3 and 6 months, with respect to a slight change in shape, that was probably due to the combination of mechanical constraint and swelling of degradation. At 12 months, approximately 30% of the cage was resorbed and no specific inflammatory reaction was seen. Thirty-six months after implantation, the site operated on displayed no signs of significant inflammatory, immunologic, or necrotic reaction.

Some intensive inflammatory tissue response was reported with the use of highly crystalline self-reinforced PGA implants [8, 10, 17, 49]. It was noticeable that the PLA cage in this study was produced with using hydrophilic zinc catalyst in place of the usual hydrophobic stannous octanoate [34], which provides highly crystalline matrix and degradation particles [35].

Chemical composition and L/DL ratio is also a prominent factor influencing tissue response and degradation. Other studies using PDLLA 70/30 cages [15, 20] showed fast degradation associated with an important inflammatory response without allowing definite fusion. Severe foreign body reaction has been using 70/30 PDLLA at the cervical spine in a sheep model [15, 20]. On the other hand, PLA stereoisomers have been shown to generate less soft-tissue reactions [8, 37, 53].

We noted that density of the new bone in the interbody arthrodesis using 96/4 PLDLLA cages was similar in all ways to that of the adjacent vertebral bodies, showing physiological path of mineralization. This is indirectly promising in terms of mechanical quality of the fusion mass; however, our study did not include biomechanical testing and do not allow to state on this point. Since the aim was purely histological with evaluating degradation and fusion, the mechanical issues related to violation of the endplates and cage subsidence were not investigated.

We acknowledge that the study was limited by the number of animals at each retrieval time; however, the number of time points and the long term follow-up did not allow to use more than three large animals per group. In spite of the differences in quality of bone between the sheep and humans, specifically the use of relatively young animals, the promising results in the short and long term incite developing investigations concerning the use of 96/4 PLDLLA as lumbar interbody implant. In addition to the differences in the mechanical environment between sheep and humans, the association with some anterior/posterior spinal instrumentation in human surgery has to be taken into account for interpreting the results of the stand-alone model presented here. However, using bone graft in and around the cage for filling the intervertebral space after discetomy has been inspired from human surgery and improved the realism of the experiment.

Conclusions

The use of 96/4 PLDLLA resorbable cages and autologous bone graft has been shown to lead to appropriate bone fusion at the lumbar spine in the sheep. Gradual resorption of 96/4 PLDLLA demonstrated low potential for foreign body reaction. Specifically, biocompatibility with the dura mater, nervous tissue, and distant organs was confirmed. The promising balance between fusion and degradation of 96/4 PLDLLA intervertebral implants suggests developing this resorbable polymer for spine surgery.