Abstract
Aim
To compare the safety and efficacy of four energy-based vascular sealing and cutting instruments.
Methods
Blood vessels of various types and diameters were harvested from four pigs using four instruments: Harmonic ACE™ (Ethicon Endo-Surgery, Cincinnati, OH), LigaSure™ V and LigaSure Atlas™ (Valleylab, Inc., Boulder, CO; a division of Tyco Healthcare), and EnSeal™ vessel fusion system (SurgRx, Inc. Redwood City, CA). The diameters of the vessels, speed and adequacy of the cutting and sealing process, and bursting pressures were compared. An additional set of specimens was sealed and left in situ for up to 4 h after which the vessels were harvested and histopathologically analyzed for the degree of thermal injury.
Results
The bursting pressures were significantly higher with EnSeal™ compared to all other instruments (p < 0.0001). The sealing process was significantly shorter with Harmonic ACE™ and significantly longer with LigaSure Atlas™ (p <0.0001). The mean seal width was larger with the LigaSure Atlas™ compared to the other instruments, and it was smaller with EnSeal™ and Harmonic ACE™. Less radial adventitial collagen denaturation was present with EnSeal™ and LigaSure™ V than with the other two instruments; there were no significant differences in collagen denaturation although proximal thermal injury to the smooth muscle in the media of the vessel wall was less common with LigaSure Atlas™ than with the other instruments; however, the numbers were too small for statistical analysis.
Conclusions
The bursting pressures with EnSeal™ were significantly higher than with all the other instruments. Harmonic ACE™ was the fastest sealing instrument and LigaSure Atlas™ was slowest. EnSeal™ created less radial thermal damage to the adventitial collagen of the vessels and LigaSure Atlas™ created less thermal damage to the media of the vessels. The clinical significance of these findings is unknown.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Recent advances in surgical technology include the use of various energy sources for sealing, coagulating, and cutting blood vessels as opposed to performing these procedures mechanically by tying, suturing, and even clipping or stapling them. The use of energy-based instruments has become even more popular in laparoscopic surgery because the traditional techniques of surgical hemostasis (pressure, tying, suturing) are not as easily laparoscopically applied. The efficacy and reliability of various energy-based vascular sealing instruments have been reported to be equivalent to the results with metallic clips and silk ties [1–3]; however, other researchers have demonstrated that energy-based devices produced either inferior [4] or superior [5] results compared to mechanical sealing and cutting techniques. Several other comparisons of various instruments [6–10] demonstrated a wide variety of results.
The aim of this study was to compare the safety and efficacy of four commercially available energy-based vascular sealing and cutting instruments, and to assess the histopathologic effect of each instrument with regard to the thermal effect it produced.
Materials and methods
Four 40–45 kg domestic pigs (Oakhill Genetics, Ewing, IL) were sedated with Telazol (4 mg/kg), ketamine (2 mg/kg) and xylazine (2 mg/kg) intramuscularly, intubated, and maintained under general anesthesia using Isoflurane (2–4%). The study was conducted in accordance with the guidelines set forth by the Guide for the Care and Use of Laboratory Animals [11], and was approved by the Washington University animal studies committee. Following surgical isolation using blunt/sharp dissection, blood vessels of various types (peripheral, visceral, arteries, and veins) and diameters were harvested using four instruments: Harmonic ACE™ (Ethicon Endo-Surgery, Cincinnati, OH), LigaSure™ V and LigaSure Atlas™ (Valleylab, Inc. Boulder, CO; a division of Tyco Healthcare Group) and EnSeal™ vessel fusion system (SurgRx, Inc. Redwood City, CA). The diameters of the vessels, speed and adequacy of the cutting and sealing process, and bursting pressures were compared. The diameters of the vessels were measured in vivo before application of the instruments. The speed of the sealing process was measured with a standard digital stopwatch. Since with the LigaSure instruments, the sealing process is separate from the cutting act, the time measured was the time elapsed to the termination of the sealing cycle as indicated by a longer beep. The adequacy of the seal was graded as follows: 1 – no seal; 2 – oozing that required additional intervention; 3 – oozing that stopped spontaneously; 4 – dry seal. The bursting pressure of the sealed end of the vessel was measured with a specialized saline infusion machine manufactured specifically for this experiment (Fig. 1): the open end of the harvested vessel segment was secured to the insertion tool with a silk tie; normal saline solution was infused at a constant rate into the vessel, and the pressure was digitally recorded and displayed on the controller readout. The highest recorded pressure measurement before the sealed end of the vessel burst was determined as the bursting pressure. If a segment had side branches identified during the pressure measurement, that specimen was disqualified and excluded from the analysis.
An additional set of specimens was sealed and left in situ for approximately 2 h (range: 0.5–4 h) after which the vessels were harvested and histopathologically analyzed for the seal quality and thermal injury. Each specimen was embedded in an individual paraffin block. The blocks were faced and four 5 μm sections were prepared equidistant across the entire seal’s width. All sections were hematoxylin and eosin stained for evaluation. Each vascular seal was assessed for the following parameters: perpendicular width of tissue seal (Fig. 2); contacting artery wall layers forming tissue seal (Fig. 3a); radius of adventitial collagen denaturation proximal to tissue seal (Fig. 3b); presence of tissue homogenization (cellular/tissue architecture loss) in tissue seal (Fig 4a); presence of gas formation (tissue boiling) in tissue seal (Fig. 4b); presence of tissue arterial wall dissection in tissue seal (Fig. 5); presence of blood pockets in tissue seal; and injury to the smooth muscle in the proximal media.
Statistical analysis was performed using GraphPad Instat™, GraphPad software V2.04 (San Diego, CA), using Student’s t-test and analysis of variance (ANOVA) when appropriate.
Results
The number of specimens harvested with each device, diameters of vessels, quality and speed of the seal, and bursting pressures are summarized in Table 1. Individual comparisons of the speed of the sealing process and bursting pressures are summarized in Table 2. The diameters of the vessels harvested with Harmonic ACE™ were significantly smaller than the diameters of vessels harvested with LigaSure Atlas™ (p = 0.0006). This difference depicts a selection bias since we deliberately chose larger vessels to be sealed with LigaSure Atlas™ and smaller vessels to be sealed with Harmonic ACE™ according to the recommendations of the manufacturers of these instruments. There was no significant difference in the quality of the seal among the four devices. The bursting pressures were significantly higher with EnSeal™ compared to all other instruments (p < 0.0001). There was no difference in the bursting pressures among the other three instruments; furthermore, the lowest mean bursting pressure that was recorded in the entire cohort of specimens was more than three times higher than normal systolic pressure. The sealing process was significantly shorter with Harmonic ACE™ and significantly longer with LigaSure Atlas™ (p < 0.0001) compared to the other instruments.
All the histological findings are summarized in Table 3. The seal’s width was larger with LigaSure Atlas™ and smaller with EnSeal™ and Harmonic ACE™. Radial adventitial collagen denaturation was less common with EnSeal™ and LigaSure™ V than with the other two instruments; EnSeal™ caused somewhat less collagen denaturation than LigaSure™ V. Proximal thermal injury to the smooth muscle in the media of the vessel wall was least common with LigaSure Atlas™; however analysis with this instrument was performed only on three of the four specimens since one of the specimens lacked sufficient tissue proximal to the sealed end. Thermal injury to the smooth muscle in the vessel’s media was most common with EnSeal™ (present in four out of the five specimens). In the specimens treated with LigaSure™ V and Harmonic ACE™, this finding was present in three out of the four specimens. Gas pockets depicting tissue boiling were most commonly observed in specimens treated with Harmonic ACE™.
Discussion
The use of sophisticated energy sources for dissection and hemostasis has greatly advanced and facilitated complex laparoscopic procedures. Due to their mode of action, traditional monopolar and bipolar cautery devices generate significant heat and result in inconsistent vessel sealing with substantial thermal spread and charring [3]. These drawbacks prompted the development of more advanced instruments such as the ultrasonic shears and feedback-monitored bipolar forceps. The desired end result of the action of every energy-based sealing/cutting instrument is the same regardless of the energy employed: a safely sealed vessel with minimal collateral thermal damage. The mode of action of ultrasonic instruments such as the Harmonic ACE™ is based on an active blade that vibrates at 55,500 Hz, which creates simultaneous coagulation and cutting of tissues. These instruments operate in lower temperatures and cause less smoke and charring due to the fact that electrical current is absent; however, the Harmonic ACE™ is recommended only for sealing vessels of up to 5 mm in diameter [12]. The LigaSure™ instruments utilize a high-current, low-voltage bipolar radiofrequency (RF) energy, in combination with a feedback-controlled response system that automatically delivers and disrupts the power according to the composition and impedance of the tissue between the jaws of the instrument [13]. These instruments can seal vessels of up to and including 7 mm in diameter. The EnSeal™ tissue sealing and hemostasis system (SurgRx, Inc. Redwood City, CA) is one of the newer products available for sealing blood vessels. It is a bipolar instrument that combines a high-compression jaw with a tissue-dynamic energy delivery mechanism. Due to the configuration and composition of the jaws of the instrument, each tissue type within the jaws receives a different energy dose that is constantly changing as the tissue is being sealed and its impedance changes. The instrument has an I-shaped blade, which is advanced as the tissue is being sealed, simultaneously cutting the sealed tissue [14]. This instrument is also recommended for vessels of up to and including 7 mm in diameter.
The results of the current experiment showed that all four instruments were efficacious and safe in sealing and cutting blood vessels. All four instruments created good seals with supraphysiological bursting pressures. Vessels sealed with EnSeal™ had statistically significantly higher bursting pressures compared to all the other instruments; however, even the lowest bursting pressures that were recorded were in the supraphysiological range, which makes the clinical significance of these differences unclear.
All four vascular seal groups showed variable degrees of histologic heterogeneity. This heterogeneity in the tissue seals precluded definitive determination of a primary representative histology for each device and their subsequent comparison. This heterogeneity may be due to inherent differences in the individual treated vessel walls (muscle/elastic media composition, wall thickness, vessel location, vessel diameter, for example), operator experience, device design, and/or other factors. Comparisons of acute cell injury (<4 h in vivo) are also limited as the cytological manifestations of cell/tissue injury do not fully develop for 36–48 h post in vivo treatment. The preliminary data suggest that the LigaSure™ instruments create a wider tissue seal compared to the other instruments; the EnSeal™ device caused variably less denaturation of the adventitial collagen, which may indicate lower temperatures in the outer layers of the vessels, but it appears to potentially have more thermal injury to the proximal inner layers of the vessels. Superheated tissue, as indicated by the presence of gas pockets, was more commonly observed in the specimens that were treated with the Harmonic ACE™. This finding indicates that, in spite of operating in lower temperatures, ultrasonic energy can create substantial tissue heat.
Within the seal region, local disruption of the vessel wall appeared similar with all four instruments. As the histological evaluation of nonsurvival study tissues for thermal injury is limited, 7-day acute and 30-day subacute follow-up studies are necessary to confirm, better characterize, and understand the apparent immediate thermal injury patterns associated with these devices. The clinical significance and implications of these immediate histological changes (<4 hour in vivo survival) cannot be determined with certainty.
References
Kennedy JS, Stranahan PL, Taylor KD, Chandler JG (1998) High-burst-strength, feedback-controlled bipolar vessel sealing. Surg Endosc 12:876–878
Harold KL, Pollinger H, Matthews BD, Krecher KW, Sing RF, Heniford BT (2003) Comparison of ultrasonic energy, bipolar thermal energy, and vascular clips for the hemostasis of small-, medium-, and large-sized arteries. Surg Endosc 17:1228–1230
Landman J, Kerbl K, Rehman J, Andreoni C, Humphrey PA, Collyer W, Olweny E, Sundaram C, Clayman RV (2003) Evaluation of a vessel sealing system, bipolar electrosurgery, harmonic scalpel, titanium clips, endoscopic gastrointestinal anastomosis vascular staples and sutures for arterial and venous ligation in a porcine model. J Urol 169:697–700
Bubenik LJ, Hosgood G, Vasanjee SC (2005) Bursting tension of medium and large canine arteries sealed with ultrasonic energy or suture ligation. Vet Surg 34:289–293
Marcello PW, Roberts PL, Rusin LC, Holubkov R, Schoetz DJ (2006) Vascular pedicle ligation techniques during laparoscopic colectomy. A prospective randomized trial. Surg Endosc 20:263–269
Matthews BD, Pratt BL, Backus CL, Krecher KW, Mostafa G, Lentzner G, Lipford EH, Sing RF, Heniford BT (2001) Effectiveness of the ultrasonic coagulating shears, LigaSure vessel sealer, and surgical clip application in biliary surgery: a comparative analysis. Am Surg 67:901–906
Carbonell AM, Joels CS, Krecher KW, Matthews BD, Sing RF, Heniford BT (2003) A comparison of laparoscopic bipolar vessel sealing devices in the hemostasis of small-, medium-, and large-sized arteries. J Laparoendosc Adv Surg Tech 13:377–380
Richter S, Kollmar O, Schilling MK, Pistorius GA, Menger MD (2006) Efficacy and quality of vessel sealing: Comparison of a reusable with a disposable device and effects of clamp surface geometry and structure. Surg Endosc 20:890–894
Diamantis T, Knotos M, Arvelakis A, Syroukis S, Koronarchis D, Papalois A, Agapitos E, Bastounis E, Lazaris AC (2006) Comparison of monopolar electrocoagulation, bipolar electrocoagulation, Ultracision, and Ligasure. Surg Today 36:908–913
Targarona EM, Balaque C, Marin J, Neto RB, Martinez C, Garriga J, Trias M (2005) Energy sources for laparoscopic colectomy: A prospective randomized comparison of conventional electrosurgery, bipolar computer-controlled electrosurgery and ultrasonic dissection. Operative outcome and costs analysis. Surg Innov 12:339–344
National Research Council (1996) Guide for the Care and Use of Laboratory Animals. National Academy, Washington (DC)
Available at: http://www.jnjgateway.com/home.jhtml?loc=USENG&page=viewContent&contentId=09008b9880a2d1f7&parentId=09008b9880a2ba17. Accessed on March 26, 2007
Available at: http://www.ligasure.com/pages/seal.htm. Accessed on March 26, 2007
Available at: http://www.surgrx.com/technology.html. Accessed on March 26, 2007
Disclosures
–Research grant provided by SurgRx Inc., Redwood City, California
–SurgRx Inc. provides educational grant support to the Department of Colorectal Surgery at Cleveland Clinic Florida
–Ethicon Endosurgery Inc. provides educational grant support to the Department of Colorectal Surgery at Cleveland Clinic Florida
–Ethicon Endosurgery Inc., SurgRx Inc., and Tyco Healthcare Inc. provide financial support for Cleveland Clinic Florida educational programs
–Dr. Talcott is a consultant and owns stock at SurgRx Inc.
Author information
Authors and Affiliations
Corresponding author
Additional information
Accepted for podium presentation at the SAGES meeting, April 2007, Las-Vegas
Rights and permissions
About this article
Cite this article
Person, B., Vivas, D.A., Ruiz, D. et al. Comparison of four energy-based vascular sealing and cutting instruments: A porcine model. Surg Endosc 22, 534–538 (2008). https://doi.org/10.1007/s00464-007-9619-8
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00464-007-9619-8