Abstract
Effective photon energy absorption (EABFeff) and exposure buildup factors (EBFeff) have been calculated based on the effective energy concept, for some dosimetric materials such as water, polymethyl methacrylate (PMMA), polystyrene, solid water (WT1), RW3 (Goettingen Water 3), and ABS (acrylonitrile butadiene styrene), for MV X-rays and 60Co gamma rays. Firstly, the equivalent atomic numbers (Z eq) of the given materials have been determined using the effective photon energies (E eff). Then, the five-parameter geometric progression (G-P) fitting approximation has been used to calculate both EABFeff and EBFeff values. Since the G-P fitting parameters are not available for the E eff values of the given materials, a linear interpolation in which a function of the logarithm of the variable is used has been performed, in order to calculate the parameters in each E eff, which will be further used for the determination of EABFeff and EBFeff. In the present paper, water equivalence properties of the given materials are also discussed based on the effective buildup factors. In this study, special emphasis is placed on the calculation of EABFeff and EBFeff values of different materials for photons that are not monoenergetic but heterogeneous in energy, to obtain an initial and prior knowledge of the probable energy and buildup of photons at locations of interest, i.e., to understand whether the real absorbed dose occurs at the surface or somewhere inside the medium of interest.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
In the field of medical physics, gamma ray buildup factors indicating photon buildup in a medium represent important parameters used to estimate the distribution of photon flux and radiation dose received by biological molecules (Chilton et al. 1984). This is so because the buildup of photons that could give rise to secondary radiation important, for example, in nuclear physics experiments, shielding design or dose estimation, is an important issue that has to be taken into account. For proper shielding and estimation of dose, the buildup factors should be determined for the materials under consideration. When photons penetrate through any material, they interact with the atoms of the target material, thus resulting in their buildup in the medium. This buildup which is due to secondary radiation can be characterized by the “buildup factor” (Sidhu et al. 2000). Because the medium in which photons interact can be the human body, buildup factors in human tissue are of importance. Studies regarding photon buildup factors in different materials have been widely made using well-known methods such as G-P fitting, invariant embedding (IE), using generalized feed-forward neural networks (GFFNNs), and Monte Carlo codes such as MCNP (Kurudirek and Ozdemir 2011a, b; Kurudirek et al. 2011a, b; Singh et al. 2008, 2010; Manohara et al. 2010; Sidhu et al. 1999; Sardari et al. 2009; Sardari and Baradaran 2010; Sardari et al. 2011; Mann et al. 2012a, b; Mann and Sidhu 2012; Sakamoto and Tanaka 1988; Shimizu 2002; Shimizu et al. 2004; Harima et al. 1986; Kucuk 2010; Kurudirek and Topcuoglu 2011). In those studies, the incident photons have been assumed as monoenergetic. However, there are almost no studies on the calculation of EABFeff and EBFeff values for clinical MV X-rays and 60Co gamma rays, which are not monoenergetic but heterogeneous in energy. In accelerator laboratories including, for example, LINACs, radiation monitoring (area and personnel) is a challenging task as no instruments are standardized for such high-energy radiation environments. In particular, it is necessary to have a prior knowledge of the probable energy of the photons at locations of interest around such accelerators, for a subsequent evaluation of absorbed dose (Bakshi et al. 2008). In the measurement of hard X-ray radiation of Bremsstrahlung with an extremely broad energy spectrum, X-ray radiation is generally characterized by the effective photon energy determined by an attenuation curve in certain suitable materials (Price 1958). For example, water is a commonly used phantom material for dosimetry measurements in radiation therapy since it is the main ingredient of the human body and because it has similar physical properties as soft tissue. Due to some practical restrictions when used in liquid form, solid water-equivalent materials would be more preferable because they can be can be modified in shape and size and therefore easily adjust to any required experimental geometry. To be used as a water phantom, any materials considered should have similar properties as water when exposed to ionizing radiation. This means that the photon interaction parameters such as effective atomic number and effective electron density should be as close as possible to those of water (Khan 2003; Baldock et al. 2010). In order to validate the calculation algorithms used, it could be reasonable to compare the calculated doses with measured doses in different phantoms (Shivaramu et al. 2001). As a sequel to the author’s previous work (Kurudirek 2013), EABFeff and EBFeff values for water and some water-equivalent materials have been calculated for 6 MV, 15 MV X-ray, and 60Co gamma beams. The water-equivalent materials used are PMMA, polystyrene, solid water (WT1), RW3, and ABS. The physical properties of the water-equivalent materials as well as those of water are given elsewhere (Kumar et al. 2010).
Materials and methods
The computation steps required to obtain the photon buildup factors of the investigated dosimetric materials are described in the following.
Calculation of effective atomic cross section σ eff and effective photon energy E eff.
Equation (1) has been used to calculate the effective atomic cross section of the investigated material:
where \({{(\mu } \mathord{\left/ {\vphantom {{(\mu } \rho }} \right. \kern-0pt} \rho })_{\text{m}}\) is the mass attenuation coefficient of the sample, \(w_{\text{i}}\) is the weight fraction of the element \(i\), and \(A_{\text{i}}\)refers to the atomic weight.
In some cases, assigning an effective energy to a heterogeneous photon radiation could be useful in calculations (Khan 2003). In order to obtain the effective photon energy (E eff), the effective atomic cross-sectional values for the materials of the investigated phantoms obtained using Eq. 1 were interpolated using the attenuation cross sections calculated with the WinXCom (Gerward et al. 2001, 2004) program as follows (Kurudirek and Çelik 2012):
where σ 1 and σ 2 are total atomic cross sections (barns/atom) defining the interval in which the effective atomic cross section σ of the considered material lies, and E 1 and E 2 are photon energies corresponding to the cross sections σ 1 and σ 2, respectively.
As an example, in the following, calculation of E eff for a 15-MV photon beam for water (H2O) is described: The effective atomic cross section of water is 0.3022 barn, which lies between the atomic cross sections 0.3024 and 0.2763 barn corresponding to the energies E 1 = 5 MeV and E 2 = 6 MeV, respectively. Then, using these values in Eq. (2), a value for E eff of 5.005 MeV is obtained. Variation of the total effective atomic cross section of water with photon energy up to an energy of 15 MeV is shown in Fig. 1 as an example. Detailed information on the calculation of Eeff values for the given materials can be found elsewhere (Kurudirek 2013).
Calculation of the equivalent atomic number Z eq
The equivalent atomic number, Z eq, of a material defines the same ratio, \({{(\mu } \mathord{\left/ {\vphantom {{(\mu } {\rho )}}} \right. \kern-0pt} {\rho )}}_{\text{Compton}} /({\mu \mathord{\left/ {\vphantom {\mu {\rho )}}} \right. \kern-0pt} {\rho )}}_{\text{Total}}\) (R), of that material at a specific energy (E eff) corresponding to the ratio of an element at the same energy. For this purpose, a large attenuation coefficient pool including the Compton partial mass attenuation coefficient,\({{(\mu } \mathord{\left/ {\vphantom {{(\mu } {\rho )}}} \right. \kern-0pt} {\rho )}}_{\text{Compton}}\), and the total mass attenuation coefficients, \(({\mu \mathord{\left/ {\vphantom {\mu {\rho )}}} \right. \kern-0pt} {\rho )}}_{\text{Total}}\), of the elements of Z=4–40 and as well as those for the investigated materials at E eff values has been constructed using the WinXCom computer program (Gerward et al. 2001, 2004) (initially developed as XCOM (Berger and Hubbell 1999)). Then, interpolation to calculate Z eq has been done using the following relation (Sidhu et al. 1999):
where Z 1 and Z 2 are the atomic numbers of the elements corresponding to the ratios R 1 and R 2, respectively, R is the ratio of Compton mass attenuation coefficient to total mass attenuation coefficient for the given materials at a specific energy.
Let water again be the example, now to illustrate calculation of Z eq at an E eff value of 2.039 MeV, which corresponds to 6 MV X-rays. The ratio \({{(\mu } \mathord{\left/ {\vphantom {{(\mu } {\rho )}}} \right. \kern-0pt} {\rho )}}_{\text{Compton}} /({\mu \mathord{\left/ {\vphantom {\mu {\rho )}}} \right. \kern-0pt} {\rho )}}_{\text{Total}}\) of water is 0.9914, which lies between the ratios 0.9909 and 0.9922 corresponding to the elements Z 1 = 7 and Z 2 = 6, respectively. Then, using these values in Eq. (3), Z eq is equal to 6.62. The variation of the ratio of incoherent to total mass attenuation coefficients for elements with atomic number Z between 4 and 20 is given in Fig. 2.
Calculation of geometric progression (G-P) fitting parameters at E eff values of the investigated materials
The G-P fitting parameters have been calculated using a similar interpolation procedure as used for the equivalent atomic number. The G-P fitting parameters for the elements present in the ANSI/ANS-6.4.3 (1991) database have been used for this purpose. However, for these elements the G-P fitting parameters at E eff values are not available in the database. Thus, Eq. (4) has been used to interpolate the G-P fitting parameters (Harima 1986):
Then, Eq. (5) has been applied to obtain the G-P fitting buildup factor coefficients of the materials:
Where C 1 and C 2 are the values of the G-P fitting coefficients for the elements of atomic numbers Z 1 and Z 2, respectively, and Z eq is the equivalent atomic number of the material.
Calculation of energy absorption and exposure buildup factors
Finally, energy absorption and exposure photon buildup factors have been calculated using the parameters obtained above with the help of the G-P fitting formula (Eqs. 6–8) (Harima et al. 1986):
where
where E is the incident photon energy, x is the penetration depth in mean free paths (mfp), a, b, c, d, and \(X_{k}\) are the G-P fitting parameters, and b also corresponds to the value of buildup factor at 1 mfp.
Results and discussion
Effective photon energy
The values of the effective photon energy (E eff) for the investigated water-equivalent materials as well as for water, for different MV X-rays and 60Co gamma rays, are given in Table 1. In order to validate the used E eff concept for MV beams, the E eff values of 60Co gamma radiation have also been determined. It is well known that after the decay of 60Co photons of two main energies, i.e., 1.173 and 1.332 MeV, with branching ratios 99.97 and 99.99 %, respectively, they are emitted. Thus, the effective photon energy for the 60Co gamma radiation corresponds to a mean value of 1.25 MeV. Table 1 shows that except for the WT1 phantom, the E eff values of the given materials for 60Co gamma radiation are close to 1.25 MeV. Greenhouse et al. (1967) used the effective photon energy concept based on half-value layers (HVL) for photons of keV energies ranging from 60 to 1,000 keV, for different materials (Greenhouse et al. 1967). The validity of the Beer–Lambert law (\(I = \,I_{0} e^{ - \mu x}\)), which is generally used to compute photon attenuation coefficients, has already been tested and verified for describing the attenuation of a wide Bremsstrahlung X-ray beam in solid water for 6 MV photons (Midgley et al. 1998). The violation of Beer–Lambert law due to photon buildup and its subsequent modified version is explained in (Singh et al. 2008). In the previous work of the author, it has been shown that whether one calculates effective atomic numbers for MV X-rays using Auto-Z eff (Taylor et al. 2012) or using the method based on effective photon energy that is adopted in the present study, the resulting differences are in general less than 1 % except for water for which the difference is 2 % for 6 MV photons (Kurudirek 2013).
Effective photon buildup factors
The Z eq values obtained here for the investigated materials at different E eff values correspond to the 6 and 15 MV X-rays, and the 60Co gamma rays are also given in Table 1. Additionally, the G-P EABFeff and EABFeff coefficients for the investigated materials are given in Table 2. It is worth noting that the word “effective” is used due to the polychromatic nature of the photon spectra. The EABF values are shown graphically at fixed penetration depths (Fig. 3) as well as at fixed energy values (Fig. 4) for water as an example. It has to be noted that EABF values for water have been taken from the ANSI database (1991). It can be observed that the EABF values show variations in the continuous energy region, depending on which of the different photon interaction processes dominates in the different energy regions (Fig. 3). At lower energies, the dominant photon interaction process is photoelectric absorption for which the atomic cross section is proportional to \(\tau \propto \frac{{Z^{4 - 5} }}{{E^{{{7 \mathord{\left/ {\vphantom {7 2}} \right. \kern-0pt} 2}}} }}\). At intermediate energies, the Compton scattering process starts dominating showing a linear atomic number dependence, while at higher energies the pair production process dominates with an atomic number dependence, which is in the order of Z 2. Among these interactions, the absorption processes (photoelectric absorption and pair/triplet production) result in a complete removal of photons, and this clearly explains the lower values of EABF and EBF at lower and higher energies. In contrast, the highest EABF and EBF values are observed at intermediate energies where Compton scattering dominates. For this process, the photons are not completely removed but their energies are only reduced. Hence, this process results in more multiple scattered photons that lead to an increase in the buildup of photons in the medium. It can be seen that at lower energies, i.e., at 0.015 MeV, the values of EABF remain constant and are close to unity for all penetration depths for water (Fig. 4). At higher energies, however, there is an increase in the EABF and EBF values for water, due to the increase in the number of multiple scattered photons. It can be seen from Figs. 5 and 6 that higher EABFeff and EBFeff values are observed for 60Co gamma rays that have an E eff value of approximately 1.25 MeV. On the other hand, lower EABFeff and EBFeff values are observed for 15 MV X-rays that have an E eff value of about 5 MeV. This can be explained due to the predominance of Compton scattering at 1.25 MeV rather than at 5 MeV for which the main dominant interaction process is pair production. Maximum differences up to 5 % are observed between EABF and EBF values for water (Fig. 7). Figures 8, 9, and 10 show significant differences between EABFeff and EBFeff values for the investigated materials for 6 MV and 15 MV X-rays, and for 60Co gamma rays. It should be noted that in general, the values for EBFeff are higher than the values for EABFeff (Figs. 8, 9, and 10). The absorption of photons in air is comparable with materials of low Z eq; hence, the EBFeff values could be higher than the EABFeff values. The Zeq value of air lies between 7 and 8, whereas the Z eq values of water and water-equivalent phantoms are between 5.32 and 6.94. The investigated materials show in general lower values of Z eq when compared to air. Thus, absorption in air contributes much more to EBFeff rather than to EABFeff. Figures 8, 9, and 10 show that the relative difference (%) between EABFeff and EBFeff is high (up to 37 % for 60Co and up to 17 % for 6 MV) in the intermediate energy region (~1.25 MeV for 60Co gamma rays, ~2 MeV for 6 MV X-rays), due to the build up of scattered photons; the difference is not significant (up to 2 %), however, at higher energies (~5 MeV for 15 MV X-rays), due to the absorption processes resulting in lower values of the buildup factor.
Water equivalence properties
The investigated materials have been studied with respect to the water equivalence in terms of effective buildup factors (Figs. 11, 12). For 6 MV photons, differences in EABFeff are <4 % and the minimum differences are observed for ABS (Fig. 11a). For 15 MV photons, differences in EABFeff are <3 % and the minimum differences are observed for WT1 up to 10 mfp, and for PMMA beyond 10 mfp (Fig. 11b). Finally, for 60Co gamma rays, differences in EABFeff are <5 % and the PMMA material shows better water equivalency than the other materials (Fig. 11c). When it comes to EBFeff, it should be noted that WT1 has better water equivalence properties for 6 MV X-rays, 15 MV X-rays (up to 15 mfp), and 60Co gamma rays. Differences are <1 % between 20 and 30 mfp, while beyond 30 mfp polystyrene seems to be the best choice of material for a water substitute.
Comparison with buildup factors for monoenergetic photons
For the purpose of the comparison, buildup factors for water have been also calculated for photons of 1.173 and 1.332 MeV separately. Then, the weighted mean buildup factor has been obtained by taking into account the branching ratios of the emission of gamma radiation emitted due to the decay of 60Co (99.974 % at 1.173 MeV and 99.986 % at 1.332 MeV). The results have been compared with the buildup factors of water, which were obtained using the effective energy approximation that is adopted in the present work. It can be seen from Table 3 that the relative differences are in the range of 0.1–1.8 % for exposure buildup factors and 0.03–0.46 % for energy absorption buildup factors.
Calculation uncertainty
The accuracy of the calculated values of effective photon energy is based on total atomic cross sections, thus on mass attenuation coefficients. The mass attenuation coefficients of the elements have been taken from the WinXCom (Gerward et al. 2001, 2004) database, which is the successor of the XCom (Berger and Hubbell 1999) database. The uncertainties in mass attenuation coefficients have been addressed in the previous reports (Hubbell 1999; Chantler 2000).
A previous study by Asano and Sakamoto (Asano and Sakamoto 2007) indicated that the results of Monte Carlo calculations agree well with those obtained with the method adopted in the present study, but with slight differences. The reasons behind these differences might be that the ANSI/ANS database uses the method of moments (Eisenhauer and Simmons 1975) with a parallel beam source and the Monte Carlo code EGS4 with an isotropic emission source, and the differences might also be due to the low-energy photon treatment applied in EGS4 for K–X radiation, L–X radiation, and Bremsstrahlung. In another study, it has been shown that different methods such as invariant embedding, G-P fitting, and Monte Carlo simulation agree well for the materials of low atomic numbers (Shimizu et al. 2004). The Z eq values of the materials in the present work cover a range from 5 to 7, which means that they can be considered as low Z materials. It has also been shown that the maximum deviations in exposure buildup factors were 0.5–3, 0.4–9.3, 0.9–42.7, and 0.4–53.2 % for G-P fitting, three-exponential approach, Berger approach, and Taylor approximation, respectively (Harima et al. 1986).
Additionally, it should be noted that a revision has been made to the existing ANSI database to improve it and catch up with the latest developments in fundamental data (Ryman et al. 2008; Ruggieri and Sanders 2008). These improvements include the issues raised from the lack of self-consistency within the standard (ANSI/ANS-6.4.3. 1991) in that the mass attenuation coefficients tabulated in the standard were not necessarily used in the calculation of the buildup factors, correction factors for coherent scattering, and tissue dose. Moreover, more accurate attenuation coefficient data have become available (Chadwick et al. 2006) since the use of the standard database (Ryman et al. 2008). One of the other improvements is that photon buildup factors for high atomic number materials have been calculated (Ruggieri and Sanders 2008).
As a summary, photon buildup factors, which have to be taken into account, for example, prior to a radiotherapy application as well as prior to shielding calculations, have been calculated for heterogeneous radiation sources. Dosimetric materials such as water and water-equivalent materials have been considered, to estimate energy absorption as well as exposure buildup factors for MV photons and gamma radiation. The present work could be useful from the point of view that it can provide an initial estimation on how radiation could interact with a target especially in terms of photon buildup factors.
Conclusions
Studies regarding photon buildup factors in different materials have been widely made under the assumption that the incident photons are monoenergetic. In contrast, in the present study, photon buildup factors in water and some water-equivalent phantoms have been calculated for clinical MV X-rays and 60Co gamma rays that are not monoenergetic, to obtain an initial knowledge of the probable energy and the buildup of photons at locations of interest in water-equivalent media, i.e., to answer the question whether the absorbed dose occurs at the surface or somewhere inside the medium. The present study reports new data on photon buildup factors of some dosimetric materials for different multi-energetic radiation sources. In order to get a rough and quick estimation of photon radiation buildup in terms of exposure and energy absorption, which is important for therapy planning or shielding calculations, the effective photon buildup factors introduced in the present study will be useful. However, for final corrections and/or detailed planning, and to obtain more confirmation and a precise knowledge on the photon buildup factors for multi-energetic photons, Monte Carlo calculations should be applied, since they take into account not only all secondary effects but also any specific experimental arrangements.
References
ANSI, ANS-6.4.3. (1991) Gamma ray attenuation coefficient and buildup factors for engineering materials. American Nuclear Society, La Grange Park
Asano Y, Sakamoto Y (2007) Gamma ray buildup factors for heavy concretes. JAEData/Code 2007-006
Bakshi AK, Nayak MK, Haridas G, Chatterjee S, Kher RK (2008) Estimation of bremsstrahlung photon energy in the environment of high-energy electron accelerator using CaSO4:Dy based TL dosemeter. Radiat Prot Dosimetry 128:43–48
Baldock C, Deene YD, Doran S, Ibbott G, Jirasek A, Lepage M, Mcauley KB, Oldham M, Schreiner LJ (2010) Polymer gel dosimetry. Phys Med Biol 55:R1–R63
Berger MJ, Hubbell JH (1999) XCOM: photon cross sections database. Web Version 1.2, available at http://physics.nist.gov/xcom, National Institute of Standards and Technology, Gaithersburg, MD 20899, USA, August 1999. Originally published as NBSIR 87-3597, XCOM: Photon cross sections on a personal computer (July 1987) 1987–1999
Chadwick MB, Oblozinsky P, Herman M et al (2006) ENDF/B-VII.0: next generation evaluated nuclear data library for nuclear science and technology. Nucl Data Sheets 107:2931–3060
Chantler CT (2000) Detailed tabulation of atomic form factors, photoelectric absorption and scattering cross section, and mass attenuation coefficients in the vicinity of absorption edges in the soft X-ray (Z = 30–36, 60–89, E = 0.1–10 keV), addressing convergence issues of earlier work. J Phys Chem Ref Data 29:597
Chilton AB, Shultis JK, Faw F (1984) Principle of radiation shielding. Prentice Hall, Englewood Cliffs
Eisenhauer CM, Simmons GL (1975) Point isotropic gamma ray buildup factors in concrete. Nucl Sci, Eng 56
Gerward L, Guilbert N, Jensen KB, Levring H (2001) X-ray absorption in matter. Reengineering XCOM. Radiat Phys Chem 60:23–24
Gerward L, Guilbert N, Jensen KB, Levring H (2004) WinXCom-a program for calculating X-ray attenuation coefficients. Radiat Phys Chem 71:653–654
Greenhouse NA, Maillie HD, Mermagen H (1967) A thermoluminescent microdosimetry system for the measurement of photon quality. Radiat Res 32:641–650
Harima Y (1986) An approximation of gamma ray buildup factors by modified geometrical progression. Nucl Sci Eng 83:299–309
Harima Y, Sakamoto Y, Tanaka S, Kawai M (1986) Validity of the geometric- progression formula in approximating gamma ray build up factors. Nucl Sci Eng 94:24–35
Hubbell JH (1999) Review of photon interaction cross section data in the medical and biological context. Phys Med Biol 44:R1–R22
Khan FM (2003) The physics of radiation therapy, 3rd edn. Lippincott Williams & Wilkins, Philadelphia
Kucuk N (2010) Computation of gamma-ray exposure buildup factors up to 10 mfp using generalized feed-forward neural network. Expert Syst Appl 37:3762–3767
Kumar R, Sharma SD, Despande S, Ghadi Y, Shaiju VS, Amols HI, Mayya YS (2010) Acrylonitrile Butadiene Styrene (ABS) plastic based low cost tissue equivalent phantom for verification dosimetry in IMRT. J Appl Clin Med Phys 11:24–32
Kurudirek M (2013) Water equivalence study of some phantoms based on effective photon energy, effective atomic numbers and electron densities for clinical MV X-ray and Co-60 γ-ray beams. Nucl Instr Meth A 701:268–272
Kurudirek M, Çelik A (2012) A simple method to determine effective atomic numbers of some compounds for multi-energetic photons. Nucl Instr Meth A 689:75–78
Kurudirek M, Ozdemir Y (2011a) A comprehensive study on energy absorption and exposure buildup factors for some essential amino acids, fatty acids and carbohydrates in the energy range 0.015–15 MeV up to 40 mean free path. Nucl Instr Meth B 269:7–19
Kurudirek M, Ozdemir Y (2011b) Energy absorption and exposure buildup factors for some polymers and tissue substitute materials: photon energy, penetration depth and chemical composition dependence. Nucl J Radiol Prot 31:117–128
Kurudirek M, Topcuoglu S (2011) Investigation of human teeth with respect to the photon interaction, energy absorption and buildup factor. Nucl Instrum Meth B 269:1071–1081
Kurudirek M, Doğan B, Ingeç M, Ekinci N, Özdemir Y (2011a) Gamma-ray energy absorption and exposure buildup factor studies in some human tissues with endometriosis. Appl Radiat Isot 69:381–388
Kurudirek M, Doğan B, Özdemir Y, Moreira AC, Appoloni CR (2011b) Analysis of some Earth, Moon and Mars samples in terms of gamma ray energy absorption buildup factors: penetration depth, weight fraction of constituent elements and photon energy dependence. Radiat Phys Chem 80:354–364
Mann KS, Sidhu GS (2012) Verification of some low-Z silicates as gamma-ray shielding materials. Ann Nucl Energy 40:241–252
Mann KS, Singla J, Kumar V, Sidhu GS (2012a) Investigations of mass attenuation coefficients and exposure buildup factors of some low-Z building materials. Ann Nucl Energy 43:157–166
Mann KS, Kurudirek M, Sidhu GS (2012b) Verification of dosimetric materials to be used as tissue-substitutes in radiological diagnosis. Appl Radiat Isot 70:681–691
Manohara SR, Hanagodimath SM, Gerward L (2010) Energy absorption buildup factors for thermoluminescent dosimetric materials and their tissue equivalence. Radiat Phys Chem 79:575–582
Midgley S, Millar RM, Dudson J (1998) A feasibility study for megavoltage cone beam CT using a commercial EPID. Phys Med Biol 43:155–169
Price WJ (1958) Nuclear radiation detection. McGraw-Hill, New York 1
Ruggieri LP, Sanders CE (2008) Update to ANSI/ANS-6.4.3. 1991 gamma ray buildup factors for high Z engineering materials (Part I). Trans. Am. Nucl. Soc. 99:618–620
Ryman JC, Alpan FA, Durani LA, Eckerman KF, Faw RE, Ruggieri L, Sanders CE, Xu XG (2008) Revision of ANSI/ANS-6.4.3. Trans Am Nucl Soc 99:613–614
Sakamoto Y, Tanaka S (1988) Interpolation of gamma ray buildup factors for point isotropic source with respect to atomic number. Nucl Sci Eng 100:33–42
Sardari D, Baradaran S (2010) Semi-empirical relationship for photon buildup factor in soft tissue and water. Rad Prot Dosim 142:209–212
Sardari D, Abbaspour A, Baradaran S, Babapour F (2009) Estimation of gamma- and X-ray photons buildup factor in soft tissue with Monte Carlo method. Appl Radiat Isot 67:1438–1440
Sardari D, Saudi S, Tajik M (2011) Evaluation of gamma ray buildup factor data in water with MCNP4C code. Ann Nucl Energy 38:628–631
Shimizu A (2002) Calculations of gamma-ray buildup factors up to depths of 100 mfp by the method of invariant embedding, (I) analysis of accuracy and comparison with other data. J Nucl Sci Technol 39:477–486
Shimizu A, Onda T, Sakamoto Y (2004) Calculations of gamma-ray buildup factors up to depths of 100 mfp by the method of invariant embedding, (III) generation of an improved data set. J Nucl Sci Technol 41:413–424
Shivaramu VR, Rajasekaran L, Ramamurthy N (2001) Effective atomic numbers for photon energy absorption of some low-Z substances of dosimetric interest. Radiat Phys Chem 62:371–377
Sidhu GS, Singh PS, Mudahar GS (1999) Energy absorption buildup factor studies in biological samples. Rad Prot Dosim 86:207–216
Sidhu GS, Singh PS, Mudahar GS (2000) A study of energy and effective atomic number dependence of the exposure buildup factors in biological samples. J Radiol Prot 20:53–68
Singh SP, Singh T, Kaur P (2008) Variation of energy absorption buildup factors with incident photon energy and penetration depth for some commonly used solvents. Ann Nucl Energy 35:1093–1097
Singh S, Ghumman SS, Singh C, Singh Thind K, Mudahar SG (2010) Buildup of gamma ray photons in flyash concretes: a study. Ann Nucl Energy 37:681–684
Taylor ML, Smith RL, Dossing F, Franich RD (2012) Robust calculation of effective atomic numbers: the Auto-Zeff software. Med Phys 39:1769–1778
Acknowledgments
The author wishes to thank Dr. Gerward for providing the WinXCom computer program as well as the Reviewers who contributed much to this work by their valuable comments.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Kurudirek, M. Photon buildup factors in some dosimetric materials for heterogeneous radiation sources. Radiat Environ Biophys 53, 175–185 (2014). https://doi.org/10.1007/s00411-013-0502-9
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00411-013-0502-9