Abstract
The middle ear poses unique challenges when finding suitable materials for ossicular reconstruction, primarily because of its link to the external environment via the eustachian tube and, hence, its greater exposure to infectious agents. In this study, the biocompatability of titanium was examined in the middle ear of rabbits by using light and scanning electron microscopy. Implants were placed as middle ear prostheses or as free implants. These were inspected at 28 days, 84 days, 168 days, 336 days and 504 days following implantation for mucosal coverage, percent epithelization and any sign of foreign-body reaction. After 28 days, the prostheses were covered by regular mucosa. Although a majority of the free implants took up to 336 days for complete epithelialization, some of the free implants were not epithelialized even at day 504. There were no inflammatory cells observed on the surface of the material, nor were unusual amounts of fibrous tissue seen. In addition, the titanium material exhibited an affinity toward bone. The results of this animal experiment indicate that titanium is a favorable material for ossicular replacement prostheses.
Article PDF
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Author information
Authors and Affiliations
Additional information
Received: 20 January 1998 / Accepted: 19 March 1998
Rights and permissions
About this article
Cite this article
Schwager, K. Titanium as a biomaterial for ossicular replacement: results after implantation in the middle ear of the rabbit. European Archives of Oto-Rhino-Laryngology 255, 396–401 (1998). https://doi.org/10.1007/s004050050086
Issue Date:
DOI: https://doi.org/10.1007/s004050050086