Introduction

In patients with asymptomatic abdominal aortic aneurysm (AAA) the fate of any individual patient is hard to predict [1]. Increasing aortic wall stiffness with a subsequent change in the wall's capacity to withstand tension changes of the pressure peaks, as a consequence of the aortic wall elastin fiber loss, could be predictive of subsequent rupture [2, 3]. In this work aortic distensibility is evaluated with ECG-gated CT [4, 5] to investigate whether distensibility would be higher in the aneurysmatic than in a non-affected segment, as postulated by Wilson [6], or rather lower, according to our own assumption. Furthermore, the difference of distensibilities for small (<5 cm) and large (>5 cm) aneurysms was investigated, a diameter which by current standards constitutes a major rupture risk and warrants repair.

Materials and methods

Sixty-seven consecutive patients with known infrarenal abdominal aortic aneurysm who had presented to the Department of Vascular Surgery were studied. The decision to subsequently operate or not upon a patient in this study was entirely at the discretion of the surgeons in charge. Approval of the local ethics committee and informed consent were obtained properly. Twenty-three patients presented an aneurysm diameter of less than 5 cm (mean, 3.9; ±0.9 cm); 44 presented with aneurysms of more than 5-cm diameter (mean 5.6; ±0.6 cm). The clinical characteristics of the population collected for this study are listed in Table 1. Mean age for the patients with larger sized aneurysms was 67±9 years including 35 men and 9 women. Mean age for the patients with smaller sized aneurysms was 69±9 years including 21 men and 2 women. Mean heart rate during examination was 67±12 bpm; no medication was used to reduce the heart rate during the examination.

Table 1 Patients’ clinical characteristics

Data acquisition

Examinations were performed using a four detector-row CT system (SOMATOM Volume Zoom; Siemens Medical Solutions, Erlangen, Germany). To reconstruct time-resolved CT data, separate retrospective ECG-gating was added to our standard CT angiography protocol. There was no adaptation of dose to morphology for the individual patient. Each patient's DLP (dose length product) is registered routinely. The acquired raw data set was used to obtain two image sets. First, standard angiography images without ECG synchronization were calculated. Second, time-resolved images were obtained with an ECG-gating algorithm developed for separate image reconstruction [4, 7] because at the time of the study no multi-phasic image reconstruction algorithm with sufficient temporal resolution was commercially available. Temporal resolution depends on the number of rotations used for interpolation, heart rate and the scanner’s rotation time [4, 8], [9]. For an interpolation range of four rotations, i.e., a 2-s time window, as used in this study, data from four consecutive rotations were used. The temporal window that corresponded to this range depended on the heart rate: for example, a heart rate of 60 bpm to the window was two heart beats long. In this study the temporal resolution ranged between 90 and 250 ms. The patient’s abdomen was scanned after an i.v. pump infusion of 130 ml Iomeprol 300 (Imeron, Altana Pharma, Konstanz, Germany) at a flow rate of 4 ml/s. The automated bolus tracking program implemented on the scanner was applied, and a ROI was placed in the abdominal aorta using a threshold of 110 HU.

The scans of the four detector-row CT scanner covered the whole abdomen and were performed with a rotation time of 0.5 s at 120 kV peak voltage, a current-time product of 120 mAs (effective), a pitch of 0.375, an acquisition slice thickness of 2.5 mm and thus a collimation of 10 mm (i.e., 4×2.5 mm). Reconstruction thickness was 2.5 mm. A collimation of 10 mm (4×2.5 mm) was used to cover the whole examination region within an image acquisition time of about 30 s and thus to acquire all images during a single breath hold period. The protocol was the same as our standard CT-angiography protocol.

From the so derived data set, 20 time frames per heart cycle were calculated at each position, i.e., images were reconstructed at 0%, 5%, 15%...95% of the RR interval by half scan reconstructions. This temporal resolution is required to define the phases of minimum and maximum aortic distension. Aortic cross-sectional area changes were determined at two positions of the aorta, immediately above the infrarenal AAA and at the level of its maximal diameter (Fig. 1). The positions were selected by a senior radiologist experienced in vascular CT.

Fig. 1
figure 1

Surface reconstruction of the CTA data showing the two positions above the branching of the renal arteries and at the level of the aneurysm (AAA) for functional image reconstruction indicated as horizontal lines (a) in a patient presenting a small aneurysm, (b) in a patient presenting a big aneurysm

In order to exclude major subclavian artery stenosis, bilateral brachial arterial blood pressure was registered non-invasively using an arm cuff before examination. Directly after image acquisition unilateral blood pressure was registered for distensibility calculation. To minimize the error in pressure estimation, an automatic pneumatic sphygmanometer was used (Maglife C, Schiller Medical SA, Wissembourg, France). The measurement accuracy of this system as given by the manufacturer is 2%.

Determination of vessel area

The vessel area was determined from the ECG-gated time-resolved image series of 20 images. Because of the high degree of operator-dependence with manual outlining of the aortic wall area, we preferred to use a semi-automatic segmentation algorithm, which detected the arterial wall in all CT images of the cardiac cycle. The algorithm has been described in the literature [10, 11] and tested previously [12]. Briefly, a seed point was defined near the center of the aorta. Next, the algorithm inflated a so-called active contour starting at the seed point like a balloon until the vessel wall was found. In order to better fit the boundary, the final contour was refined by an iterative process (Figs. 2 and 3).

Fig. 2
figure 2

Illustration of the processing steps required to calculate the vessel area. The example shows a single time frame of the functional images aortic diameter <5 cm of the patient presented in Fig. 1a (a) of the suprarenal aorta. The CT images are segmented using an active contour algorithm, setting a seed point and defining two ROIs in between which the vessel wall is found (b). The resulting area-versus-time curve (in percent of RR interval) is shown for the patient's area change in the aorta and at the level of aneurysm AAA (c)

Fig. 3
figure 3

Processing steps required to calculate vessel area, shown for an example measurement at aneurysm level Ø<5 cm. The example shows the functional image (a) of the aneurysmatic aorta of the patient presented in Fig. 1b. The reconstructed CT images are segmented using the active contour algorithm, setting a seed point and defining two ROIs in between which the vessel wall is found (b). The resulting area-versus-time curve (in percent of RR interval) is shown for the patient's area change at the AAA (c)

Elasticity quantification

The visco-elastic properties of the aortic wall are typically characterized by two parameters: dilation d [mm] and distensibility D [pa-1]. The dilation is defined as the maximum systolic-to-diastolic diameter change of the selected area [13]:

$$\rm {\text{d = d}}_{{{\text{sys}}}} {\text{ - d}}_{{{\text{dia}}}} $$
(1)

Distensibility D [pa-1] is defined as the relative change in vessel cross-sectional area A [mm2] that occurs during the cardiac cycle, divided by the corresponding change in blood pressure Δp [Pa].

$$ D = \frac{{\Delta A}} {{A_{0} \cdot \Delta p}} $$
(2)

A 0 is the minimum vessel area over the cardiac cycle, and ΔA is the difference between maximum and minimum area [14]; Δp can be estimated using sphygmomanometry [15].

In order to compare results with literature values, distensibility was also converted into pulse wave velocity PWV [m/s] [14]:

$$ PWV = \frac{1} {{{\sqrt {D \cdot \rho } }}} $$
(3)

Here \( \rho = 1055 \cdot 10^{3} \) [kg/m3] denotes the mass density of blood that is assumed to be constant.

For each time series the segmentation process was repeated five times, and the reproducibility coefficient of variation was determined. Distensibility values were calculated using Eq. 2. Paired t-test was performed to assess the statistical significance of differences at the AAA level and adjacent aorta; difference between large and small aneurysms were tested by t-test with SAS software (version 9.1; SAS Institute Inc., Cary, NC). Box plots were generated using R software (version 2.4.1. The R Foundation for Statistical Computing).

Results

With the active contour algorithm the cross-sections of the aorta could be segmented successfully in all 67 patients (Fig. 2), although some minor streak artifacts were visible due to the limited amount of data for reconstruction. Mean signal variation on the streak artifacts was on the order of 32 HU. Difference in density between the vessel lumen and the surrounding fatty tissue was 297 HU. Coefficient of variation for the five measurements in each patient was 8.6%. The mean patient DLP (dose length product) was 900±232 mGy x cm for the two helicals (native, arterial ECG-gated contrast phases). CTDIw for the applied protocol was CTDIw=41 mGy.

Distensibility, pulse wave velocity values and dilation were calculated (Tables 2, 3, 4). A significant difference in distensibility D (dilation d) between the adjacent aorta wall and the aneurysm was found for both aneurysm sizes. Upstream (AAA) means for both groups together were D above =(1.3±0.8)·[10−5 Pa−1], D AAA =(0.6±0.5)·[10−5 Pa−1], t-test large AAA: pD<0.0001, (pd=0.04) and t-test small AAA: pD<0.0001, (pd=0.045) visualized in the box plot in Fig. 4.

Table 2 Distensibility of small vs. large aneurysms
Table 3 Results-pulse wave velocity of small vs. large aneurysms
Table 4 Results-dilation of small vs. large aneurysms
Fig. 4
figure 4

Box plot: Comparison of the distensibilities in the aorta and at aneurysm level (AAA) for small and big aneurysm. The box stretches from the lower hinge (defined as the 25th percentile) to the upper hinge (the 75th percentile). The median is shown as a line across the box. The “H-spread” is defined as the difference between the hinges, and a “step” is defined as 1.5 times the H-spread. Inner fences are one step beyond the hinges. Outer fences are two steps beyond the hinges.There are two adjacent values: the largest value below the upper inner fence and the smallest value above the lower inner fence. For the data plotted in the figure, the minimum value is above the lower inner fence and is therefore the lower adjacent value. The maximum value is the inner fences so it is not the upper adjacent value. As shown in the figure, a line is drawn from the upper hinge to the upper adjacent value and from the lower hinge to the lower adjacent value. Every score between the inner and outer fences is indicated by an “o”

However differences between large and small AAA were not found to be significant. Distensibility values were very low for both groups. The upstream AAA distensibility range for both groups was D=(0.4–3.7) [10Pa-1]. At AAA the range was D=(0.1–3.6) [10Pa-1]. This corresponds to a pulse wave velocity (PWV) range of above AAA level PWV=(5.1–15.9) m/s and at AAA level of PWV=(5.1 to 29.9) m/s. Likewise dilation was very small at the AAA with a range of (0.1–2.8) mm and slightly higher above the AAA with a range of (0.4–2.5–29.9) mm.

Discussion

The main characteristic of an abdominal aortic aneurysm (AAA) is the replacement of the medial elastic fibers by connective tissue, with a loss of elastic properties, increase in diameter and, finally, wall rupture. Other factors, such as parietal calcification, intrasaccular thrombus and maximum diameter, may possibly modify the behavior of the aortic wall. Thus, distensibility might provide information about the constitution of the arterial wall.

Several techniques have been proposed to measure vascular distensibility non-invasively: These include pulse-wave velocity measurements employing either magnetic resonance velocity mapping [1518], or Doppler ultrasound [19], and methods that monitor the change of the vessel cross-sectional area during the cardiac cycle [20]. From the change in cross-sectional area between systole and diastole and an estimate of the corresponding intra-arterial pressure difference, the distensibility can be calculated. The latter has been done using magnetic resonance imaging [21, 22] and ultrasound [20, 23]. Ultrasound image data cannot image all parts of the aorta in a single view. MR pulse wave velocity measurements are unreliable in situations with complex flow patterns [24]; availability of MR systems is limited, the acquisition of several pulse sequences prolongs scan time, and monitoring of instable patients can be difficult. Finally, spatial resolution of MRI is currently inferior to that of CT. Therefore, in spite the radiation exposure, CT angiography still is the preferred method to screen aortic pathologies [2527].

In this study, the distensibility was assessed indirectly by measurement of the change over time of the vessel cross-sectional area, using ECG-gated MDCT and by measuring peripheral systemic blood pressure. The vessel cross-sectional area was determined from CT image data by a semiautomatic segmentation algorithm in AAA patients. It was tested whether substantial distensibility changes can be found above the AAA vs. AAA level for two aneurysm sizes, since smaller aneurysms are known to be less likely to rupture than larger aneurysms. The technique applied in this study has already been validated to derive cyclic cross-sectional area changes of the aorta [4, 28] and tested on angiography data sets from patients undergoing routine MDCT [29].

We found that the distensibility D was significantly lower in the aneurysms than in the proximal, non-aneurysmatic parts of the aorta and that there was no apparent difference between small and larger aneurysms. This finding supports our hypothesis that the normal vessel is more distensible and can better withstand systemic pressure peaks, whereas the wall of the AAA is stiffer. We believe that this is a plausible explanation for the pathomechanism that ultimately leads to an aneurysm in this area, since in a stiffened vessel wall, the impact of distending forces is more likely to cause damage than in a compliant wall.

The results did not show any statistical difference in aneurysmal aortic wall mechanics in AAAs with a diameter of more than 5 cm vs. smaller aneurysms, which are generally treated conservatively. The number of individuals with aneurysms of less than 5 cm was quite small (n=23), which makes it difficult to rule out smaller differences between the groups.

The time-resolved images show streak artifacts that lead to artificial contour changes using the described algorithm. This systematic error fundamentally limits the achievable precision in the distensibility calculations. Streak artifacts will simulate vessel dilation in completely stiffened vessels.

Indirect measurement of arterial blood pressure may lead to a small systematic error of the obtained pressure values. However, in patient studies this error is consistent [30]. Furthermore, invasive blood pressure measurements would not be practicable in patients undergoing routine CT examinations.

In a recent study, an age-dependent loss of distensibility was found in patients without known vascular diseases, using the same method [29]. Averaging over all of our patients with aneurysm yielded a low mean distensibility: D=1.35 10−5 Pa−1 above and D=0.65 10−5 Pa−1 at the aneurysm level, whereas the expected distensibility (according to extrapolation of the literature data) for a 68.5-year-old human, representing the mean age group, not suffering from abdominal aorta aneurysm would have been D=1.9 10−5 Pa−1 above and D=1.06 10−5 Pa−1 below the renal arteries.

The pitch used in this study is not small enough for cardiac imaging; however, for aortic vessel area determination a lower temporal resolution is acceptable if a certain temporal smearing of the data is tolerated. To measure the maximum and minimum vessel area the exact temporal correlation within the heart cycle is not required, because the changes do not occur as rapidly as in the heart. Although low pitch values are necessary to reconstruct time-resolved images, relatively high pitch values of 0.375 were used here to limit radiation dose. These pitch values correspond to our routine non ECG-gated CT angiography protocol. With the mean DLP of 900±232 mGy x cm (dose length product) for the two helicals (native, arterial ECG-gated contrast phases), this scan protocol did not exceed the diagnostic reference value given by the German Federal Radiation Protection Authority for abdominal scan protocols, where a DLP of 1,500 mGy x cm for 2.5 helicals is indicated [31].

In this study arterial wall distensiblity was assessed in AAA patients by ECG-gated CT angiography. As CT angiography is routinely performed for many aortic diseases, the presented method can be integrated into a clinical protocol. The small additional effort is appropriate for routine clinical use. The image protocol provides correct morphological as well as functional information about the vessel distensibility. With a semiautomatic segmentation algorithm, vessel cross-sectional area was measured. Further improvements of the segmentation algorithm will facilitate the determination of distensibility as a by-product of CT angiography. Thus, the proposed method will benefit from the ongoing improvements in CT technology [32].

In the literature it is controversial whether a difference in aortic wall mechanics indicates those AAAs that subsequently rupture. Wilson et al. [6] applied an ultrasound B-mode echo-tracking device and found that baseline AAA compliance was significantly related to rupture. He also hypothesized that aneurysms with low stiffness are the ones with risk of rupture [33], whereas Moritake et al. hypothesized that the stiffest aneurysms might be the ones most prone to rupture [34]. Sonesson et al. [35] as well as Mac Sweeney et al. [3] so far failed to show any relationship between the stiffness of aneurysms and their size, which is in agreement with our results. Possibly the loss of elastic properties precedes the formation of the aneurysms, so that, once this has happened, the ensuing changes are only destructive ones, resulting in further dilatation, ultimately rupture, but only minimal alterations of the already deteriorated elastic properties. So far, we remain unable to determine the signs that herald an imminent rupture of an aneurym. However, stiffening of an aortic wall that otherwise appears to be normal, with respect to the patient’s age, might be a sign that this vessel is at risk of developing an aneurism in the future.

Conclusion

It is possible to determine distensibility non-invasively with ECG-gated CT of abdominal aortic aneurysms. The normal vessel is more distensible and can better withstand systemic pressure peaks, whereas the wall of the AAA is stiffer. However, reduction of distensibility within aneurysms compared to normal proximal aorta is subtle; the lack of difference between both small and large aortic aneurysms suggests that this reduction occurs early in the aortic aneurysm's development. Hence reduced aortic distensibility might be a predictive parameter in patients with high risk of aortic disease, e.g., diabetes mellitus or Marfan's syndrome.