Abstract
The aim of this study was to investigate the usefulness of saline chaser in 16-row multislice CT (16-MSCT) coronary angiography. Forty-two patients were divided into two groups for contrast material (CM) administration: group 1 (140 ml at 4 ml/s) and group 2 (100 ml at 4 ml/s followed by 40 ml of saline chaser at 4 ml/s). All patients underwent retrospectively ECG-gated 16-MSCT coronary angiography. The attenuation at the origin coronary vessels was assessed. Three regions of interest (ROIs) were drawn throughout the data set: (a) ascending aorta (ROI 1); (b) descending aorta (ROI 2); and (c) pulmonary artery (ROI 3). The attenuation in the superior vena cava was recorded (ROI 4). The average attenuation and the slope were calculated in each ROI and differences were assessed with a Student’s t test. The average attenuation in the coronary vessels was not significantly different in the two groups. The average attenuations in ROI 1 were 325 and 327 HU, in ROI 2 were 328 and 329 HU and in ROI 3 were 357 and 320 HU, for groups 1 and 2, respectively (p>0.05). The slopes in ROI 1 were −0.2 and 1.1, in ROI 2 were 2.8 and 2.1 (p>0.05) and in ROI3 were 3.9 and −9.0 (p<0.05), for groups 1 and 2, respectively. The average attenuations in ROI 4 were 927 and 643 HU (p<0.05), for groups 1 and 2, respectively. One hundred milliliters of CM with 40 ml of saline chaser provides the same attenuation as 140 ml of CM (35% less) with decreased hyper-attenuation in the superior vena cava.
Similar content being viewed by others
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.Avoid common mistakes on your manuscript.
Introduction
Multislice-CT scanners with 16 rows of detectors (16-row MSCT) and increased spatial and temporal resolution have been introduced [1, 2, 3]. The increased number of detector rows and gantry rotation speed reduce the time needed to scan the coronary arteries to <20 s [3]. Early experiences reported improved results in the visualisation of coronary arteries and in the detection of coronary artery disease [3, 4, 5, 6].
The shorter scan time, provided by 16-row MSCT, allows to decrease the volume of contrast material (CM) needed for CT coronary angiography (CTA) [7]. Moreover, bolus-tracking technique can be easily applied which allows a better synchronisation of the CTA scan with contrast material passage [4, 7, 8, 9, 10, 11, 12].
The use of a saline solution injected intravenously immediately after the CM main bolus, also known as bolus chaser, has been reported to allow a significant reduction of CM volume with vascular attenuation comparable with the one obtained with larger volumes of CM [13, 14, 15]. It is expected that the use of bolus chaser also allows CM volume reduction in MSCT coronary angiography [16, 17, 18].
The aim of this study was the comparison of a “conventional” CM protocol without bolus chaser with a “low-volume” protocol with bolus chaser in non-invasive 16-row MSCT coronary angiography.
Materials and methods
Patient population
In November and December 2002, 42 patients (30 men and 12 women; mean age 59 years, age range 34–79 years), undergoing non-invasive MSCT coronary angiography for suspected coronary artery disease were prospectively enrolled in the study. Exclusion criteria for coronary CTA were irregular heart rates, previous allergic reaction to iodine contrast media, renal insufficiency (serum creatinine >120 mmol/l), pregnancy, respiratory impairment, unstable clinical status, or marked heart failure. The Institutional Review Board approved the study and patients gave informed consent.
After enrollment, patients were randomly divided into two groups with different protocols for intravenous contrast material (Visipaque 320 mg I/ml, Amersham Health, Little Chalfont, UK) administration: group 1 (conventional) 140 ml administered at 4 ml/s and group 2 (low-volume) 100 ml at 4 ml/s followed by 40 ml of saline at 4 ml/s. In each patient, age, body weight and heart rate were recorded.
Multislice-CT scan
Prior to the examination, the patients’ heart rate (HR) was measured. Patients with a pre-scan HR ≥65 bpm were given 100 mg of metoprolol per os 1 h before the scan.
The scan parameters for MSCT coronary angiography (Sensation 16, Siemens, Forchheim, Germany) were: number of detectors 12; individual detector width 0.75 mm; gantry rotation time 420 ms; 120 kV; 400–500 mAs; feed/rotation 2.8 mm; and scan direction cranio-caudal. The heart rate during the scan was also recorded.
Contrast material administration and synchronisation protocols
The iodinated CM was administered intravenously using a prototype of double-head power injector (Stellant, MedRAD, Pittsburgh, Pa.) through an 18-G venflon in the antecubital vein. Two different injection protocols were applied.
Synchronisation between the passage of CM and data acquisition was achieved with real-time bolus tracking (CARE bolus, Siemens, Forchheim, Germany) using an ROI in the ascending aorta for monitoring a threshold of +100 HU above the baseline attenuation to trigger the scan.
Data collection
Two data sets were reconstructed using retrospective ECG gating with time window starting 400 ms before the next R wave. The first data set was reconstructed for the purpose of coronary artery attenuation assessment with effective slice width 1 mm, reconstruction interval 0.5 mm, field of view (FOV) 160 mm and convolution filter medium smooth. The second data set was reconstructed for the purpose of great thoracic vessel assessment with effective slice width 3 mm, reconstruction interval 3 mm, FOV 200 mm and convolution filter medium smooth.
Coronary artery attenuation
Axial slices in the data set were scrolled to find the best location to measure the attenuation at the origin of the main coronary arteries (Fig. 1): left main coronary artery; left anterior descending; circumflex artery; and right coronary artery (RCA). The ROI was drawn as large as possible and calcifications, plaques and stenosis were carefully avoided.
Bolus geometry of great vessels
The attenuation in Hounsfield units was measured in three arteries drawing a ROI in consecutive slices (at intervals of ~1 s.) through the data set (Fig. 2): (a) the ascending aorta–left ventricle (ROI 1); (b) the descending aorta (ROI 2); and (c) the pulmonary artery–right ventricle (ROI 3). In addition, the attenuation inside the superior vena cava (ROI 4) at the beginning of the scan (e.g. first slice of the data set) was recorded. Time-related contrast measurements for each vessel, which were available in all patients, were included in the study to maintain homogeneous results.
Data analysis
The attenuation values obtained from the three ROIs in each patient were averaged at each time point to generate the average time/density curves. Bolus geometry was described by two parameters, which represented quantitatively, the amount of CM present in the vessel during the scan (average attenuation) and the pattern of enhancement in the vessel (slope of attenuation). Three additional parameters were considered descriptive of bolus geometry: the attenuation value at the beginning of the scan (time 0), the maximum enhancement value (MEV; the peak of attenuation) and the time to reach the MEV (tMEV). Significant differences between the two groups were assessed using Student’s t test. A p value <0.05 was considered significant.
Results
Scans and bolus timing procedures were successfully completed in all patients. No significant adverse reactions to CM were observed. Age, weight, mean heart rate during the scan, mean scan delay and mean scan time were not significantly different in the two groups (Table 1).
Coronary artery attenuation
The attenuation at the origin of the four main coronary vessels was higher in group 1, but there was not a significant difference between the two groups (Fig. 3A; Table 2).
Ascending aorta–left ventricle
The average attenuation and the slope were slightly higher in group 2 but not significantly different (p>0.05; Fig. 3B; Table 3). The attenuation values at time 0, the MEV and the tMEV were slightly higher in group 2 but not significantly different (p>0.05).
Descending aorta
The average attenuation and the slope were slightly higher in group 2 but not significantly different (p>0.05; Fig. 3C; Table 3). The attenuation values at time 0 and the MEV were slightly higher in group 2 but not significantly different (p>0.05). The tMEV was significantly higher in group 1 (p<0.05).
Pulmonary artery–right ventricle
The average attenuation was higher in group 1 but not significantly different (p>0.05; Fig. 3D; Table 3). The slope, instead, was significantly higher in group 1 (p<0.05). The attenuation values at time 0 and the MEV were not significantly different between the two groups (p>0.05). The tMEV was significantly higher in group 1 (p<0.05).
Superior vena cava
The average attenuation was significantly higher in group 1 (p>0.001; Table 3).
Discussion
With the new generation of MSCT scanners featuring 16 rows of detectors, the time needed to scan the heart for the purpose of coronary CTA has been reduced to ~20 s [1, 2, 3, 4, 5]; yet, no studies have been performed on the optimisation of CM administration in non-invasive coronary imaging with MSCT.
We compared a conventional CM protocol for non-invasive MSCT with a low-volume protocol adding the pushing and washout effects of a bolus chaser. The two protocols showed comparable attenuation values at the level of coronary arteries and similar bolus geometry in the ascending and descending aorta. A significant difference was observed at the level of pulmonary artery where more “pooling” of CM was present for the conventional protocol. This observation is confirmed by the significantly lower attenuation observed in the superior vena cava for the low-volume protocol and supports the conclusion that 100 ml of CM pushed by 40 ml of saline provide optimal coronary artery enhancement with less attenuation in the right cavities of the heart. This result can be beneficial for a better visualisation of the mid-tract (segment 2) of RCA that may suffer from beam-hardening artefacts when highly concentrated CM is present in the right atrium and right ventricle. This effect can be a drawback when cardiac masses need to be studied. In this case a conventional CM protocol must be used.
The attenuation at time 0 and the MEV were not significantly different between the two protocols. The tMEV, instead, was significantly longer in the descending aorta (~3.2 s) and significantly shorter in the pulmonary artery (~3.6 s), for the conventional protocol. We have no clear explanation for the differences detected at this level.
Previous experiences with bolus chaser support the evidence that the same results, in terms of vascular and parenchymal attenuation, can be achieved using less CM volume (up to 20–40% less) and with a concomitant reduction of the artefacts at the level of superior vena cava [13, 14, 15, 19].
The explanation may be that a saline chaser pushes the injected CM through the veins of the forearm which will give the same result as the injection of a larger contrast volume, and that the saline chaser prevents the decrease of the contrast material flow in the arm veins which may normally cause an increase in the CM concentration after the end of the contrast injection [7].
A limitation of this study is that the evaluation of coronary vessels could have been completed by the assessment of the length of coronary artery visualisation and/or the number of side branches visualised. We did not perform this evaluation because of the variable diameter of coronary arteries, heart rate and the degree of vessels with atherosclerotic disease (soft and calcified lesions, vessel stenosis and vessel occlusions). These parameters severely affect the capability of visualisation of the vessel regardless of the performance of the protocol for CM administration (e.g. the attenuation inside the vessel), and the number of patients enrolled in our study was not large enough to account for these variables.
Conclusion
In conclusion, the integration of bolus chaser in the CM administration protocol for non-invasive coronary artery angiography with 16-row MSCT allows to preserve the optimal intra-vascular attenuation, decreasing the volume of CM (35%) and the hyper-attenuating superior vena cava and right heart, with resulting cost savings and decreased risk of CM nephropathy.
References
Flohr T, Stierstorfer K, Bruder H, Simon J, Schaller S (2002) New technical developments in multislice CT. Part 1: approaching isotropic resolution with sub-millimeter 16-slice scanning. Rofo Fortschr Geb Rontgenstr Neuen Bildgeb Verfahr 174:839–845
Flohr T, Bruder H, Stierstorfer K, Simon J, Schaller S, Ohnesorge B (2002) New technical developments in multislice CT. Part 2: sub-millimeter 16-slice scanning and increased gantry rotation speed for cardiac imaging. Rofo Fortschr Geb Rontgenstr Neuen Bildgeb Verfahr 174:1022–1027
Heuschmid M, Kuttner A, Flohr T, Wildberger JE, Lell M, Kopp AF, Schroder S, Baum U, Schaller S, Hartung A, Ohnesorge B, Claussen CD (2002) Visualization of coronary arteries in CT as assessed by a new 16-slice technology and reduced gantry rotation time: first experiences. Rofo Fortschr Geb Rontgenstr Neuen Bildgeb Verfahr 174:721–724 [in German]
Nieman K, Cademartiri F, Lemos PA, Raaijmakers R, Pattynama PM, de Feyter PJ (2002) Reliable noninvasive coronary angiography with fast submillimeter multislice spiral computed tomography. Circulation 106:2051–2054
Ropers D, Baum U, Pohle K, Anders K, Ulzheimer S, Ohnesorge B, Schlundt C, Bautz W, Daniel WG, Achenbach S (2003) Detection of coronary artery stenoses with thin-slice multi-detector row spiral computed tomography and multiplanar reconstruction. Circulation 107:664–666
Cademartiri F, Nieman K, Raaymakers RH, de Feyter PJ, Flohr T, Alfieri O, Krestin GP (2003) Non-invasive demonstration of coronary artery anomaly performed using 16-slice multidetector spiral computed tomography. Ital Heart J 4:56–59
Cademartiri F, van der Lugt A, Luccichenti G, Pavone P, Krestin GP (2002) Parameters affecting bolus geometry in CTA: a review. J Comput Assist Tomogr 26:598–607
Kopka L, Rodenwaldt J, Fischer U, Mueller DW, Oestmann JW, Grabbe E (1996) Dual-phase helical CT of the liver: effects of bolus tracking and different volumes of contrast material. Radiology 201:321–326
Kirchner J, Kickuth R, Laufer U, Noack M, Liermann D (2000) Optimized enhancement in helical CT: experiences with a real-time bolus tracking system in 628 patients. Clin Radiol 55:368–373
Mehnert F, Pereira PL, Trubenbach J, Kopp AF, Claussen CD (2001) Automatic bolus tracking in monophasic spiral CT of the liver: liver-to-lesion conspicuity. Eur Radiol 11:580–584
Mehnert F, Pereira PL, Trubenbach J, Kopp AF, Claussen CD (2001) Biphasic spiral CT of the liver: Automatic bolus tracking or time delay? Eur Radiol 11:427–431
Sandstede JJ, Tschammler A, Beer M, Vogelsang C, Wittenberg G, Hahn D (2001) Optimization of automatic bolus tracking for timing of the arterial phase of helical liver CT. Eur Radiol 11:1396–1400
Haage P, Schmitz-Rode T, Hubner D, Piroth W, Gunther RW (2000) Reduction of contrast material dose and artifacts by a saline flush using a double-power injector in helical CT of the thorax. Am J Roentgenol 174:1049–1053
Hopper KD, Mosher TJ, Kasales CJ, TenHave TR, Tully DA, Weaver JS (1997) Thoracic spiral CT: delivery of contrast material pushed with injectable saline solution in a power injector. Radiology 205:269–271
Bader TR, Prokesch RW, Grabenwoger F (2000) Timing of the hepatic arterial phase during contrast-enhanced computed tomography of the liver: assessment of normal values in 25 volunteers. Invest Radiol 35:486–492
Garcia P, Genin G, Bret PM, Bonaldi VM, Reinhold C, Atri M (1999) Hepatic CT enhancement: effect of the rate and volume of contrast medium injection in an animal model. Abdom Imaging 24:597–603
Han JK, Kim AY, Lee KY, Seo JB, Kim TK, Choi BI, Lhee CS, Han MC (2000) Factors influencing vascular and hepatic enhancement at CT: experimental study on injection protocol using a canine model. J Comput Assist Tomogr 24:400–406
Yamashita Y, Komohara Y, Takahashi M, Uchida M, Hayabuchi N, Shimizu T, Narabayashi I (2000) Abdominal helical CT: evaluation of optimal doses of intravenous contrast material—a prospective randomized study. Radiology 216:718–723
Dorio PJ, Lee FT Jr, Henseler KP, Pilot M, Pozniak MA, Winter TC III, Shock SA (2003) Using a saline chaser to decrease contrast media in abdominal CT. Am J Roentgenol 180:929–934
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Cademartiri, F., Mollet, N., van der Lugt, A. et al. Non-invasive 16-row multislice CT coronary angiography: usefulness of saline chaser. Eur Radiol 14, 178–183 (2004). https://doi.org/10.1007/s00330-003-2188-x
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00330-003-2188-x