Introduction

Cementless total hip arthroplasty (THA) shows excellent long-term implant survival, with a mean of 94.7 % after 16 years [1]. However, the main reasons for revision are based on aseptic loosening [1], caused by factors such as missing primary stability, stress shielding and wear-particle-induced osteolysis [2]. An implanted hip stem may change bone structure in the proximal femur [3]. High intrafemoral stresses and nonphysiological loading pattern may determine the bone remodelling process. A reduced load transfer to the femur may induce a reduction of bone density (atrophy) around the stem [4]. Consequently, increased loading may increase bone density (hypertrophy) [5]. These remodelling processes are influenced by many implant-related parameters, such as implant material properties [6] or design [7]. The aim of a modern cementless hip stem is to generate a metaphyseal fixation to reach a load transfer in the subtrochanteric area closely comparable with physiological conditions. Proximal load transfer therefore reduces proximal stress shielding [8, 9], which can lead to implant loosening [10]. Another essential condition in cementless THA is an optimal press fit (primary fixation) with low micromotions at the implant–bone interface for sufficient bone ingrowth (secondary fixation) [11]. Long-term implant fixation requires a balance between decreasing primary and increasing secondary fixation. Consequently, an imbalance between the two may result in implant loosening [12].

In our ongoing radiological follow-up investigation after THA using the Fitmore® stem (Zimmer), we sometimes noticed the formation of cortical hypertrophy in the distal part of the stem after one year of follow-up (Fig. 1). Comparable cases of cortical hypertrophies could be found [13]. Although we have yet to confirm whether or not the observed hypertrophies of the Fitmore® stem affect long-term stem stability, we analysed possible biomechanical and implant-related reasons for the formation of these distal hypertrophies.We hypothesised that (1) the rotational stability and (2) bending behaviour of the Fitmore® stem is different to that of the established CLS® stem, which is known for excellent clinical long-term results [14].

Fig. 1
figure 1

ad Fitmore® stem, size 8 B Ext. Offset, in a 68-year-old man immediately postoperatively (a) and 1 year after surgery (b) . Fitmore® stem, size 11 B Ext. Offset, in a 63-year-old man immediately postoperatively (c) and 1 year after surgery (d)

Materials and methods

Eight synthetic femora (composite bone fourth generation (3406), Sawbones® Europe, Malmö, Sweden) were osteotomised using a standardised femoral-neck resection. Four Fitmore® stems (size B10 extended, Zimmer GmbH, Winterthur, Switzerland) and four CLS® stems (size 12.5, Zimmer) were chosen based on X-ray template planning. An experienced surgeon reamed and rasped the synthetic femora. After preparation, the stems were implanted into the femora with 25 axial load cycles of 2,000 N and an additional 25 load cycles of 4,000 N using a material testing device (Frank- Universalprüfmaschine 81816/B, Karl Frank GmbH, Weinheim-Birkenau, Germany) to achieve a press-fit situation similar to the fit achieved in vivo during hip surgery [15]. The implant fit was monitored by X-ray examination.

Using a well-established implant-stability measuring device [1618], different load situations were applied on the stem-bone compound using two linear actuators. A rope system was used to minimise counteractions from the actuators and therefore to eliminate actuator-guided motions of the implant. During load application, six linear variable displacement transducers (LVDTs) (differential transducers type P2010, Mahr GmbH, Göttingen, Germany) with a resolution of 0.1 μm were attached to each measuring point (Fig. 2) to determine the micromotions in all degrees of freedom [17].

Fig. 2
figure 2

Defined measuring points of the implant (1, 2) and of the bone (0, 3–6) at different measuring levels

Two different cyclic load applications were performed to characterise loading behaviour in vitro:

  • Setup A: To classify rotational implant stability, a cyclic torque of ±7 Nm was applied around the stem axis. Micromotions were measured at different levels along the stem axis between bone and implant, as shown in Fig. 2. Two measuring points were defined at the stem: proximal shoulder (1) and distal tip of the stem (2). Five measuring points were located at the femur: two femoral points (4, 6) were located directly corresponding to the stem points (1, 2), and the level of the lesser trochanter (0) served as a global reference. Additionally, two points, located eight centimetres (3) and 20 cm (5) distal to the lesser trochanter, were placed to achieve a higher resolution of femoral motion. Comparing the motions of stem and femur at different levels allowed the relative micromotions at the bone–implant interface to be calculated.

  • Setup B: To evaluate bending behaviour, we rearranged the measuring system to apply a varus–valgus torque of ±3.5 Nm. In this case, the absolute micromotions of the implants (measuring points 1 and 2) were compared with respect to magnitude and direction (medial or lateral) for both designs.

Statistics

An analysis of variance (ANOVA) was used to compare implant designs regarding location of their main fixation (Setup A). A second ANOVA determined whether there were differences between implant measuring levels. A least significant difference test (LSD) was calculated as a post hoc test. For Setup B, Student’s t test was used to demonstrate whether differences between measuring levels appeared to compare both designs in terms of flexibility. A p value <0.05 was considered significant. Data are expressed as means ± standard deviation (SD).

Results

Setup A: rotational torque around the stem axis: rotational primary stability

In Fig. 3 micromotions are compared between implant designs during rotational torque application and showed no significant difference (p = 0.82). However, within each design, significant differences between measuring levels were found (p < 0.01). The largest micromotions were measured at the distal tip of the stems (Fitmore® 11.91 ± 2.58 mdeg/Nm vs. CLS® 12.32 ± 3.68 mdeg/Nm; with no difference between designs p = 0.77), whereas the smallest micromotions were measured more proximally at the lesser trochanter region (Fitmore® 7.18 ± 0.79 mdeg/Nm vs. CLS® 6.49 ± 0.87 mdeg/Nm; with no difference between designs p = 0.63). Micromotions significantly differed from the distal tip to the trochanter minor region (Fitmore® p < 0.01 and CLS® p < 0.01) and the distal tip to the proximal shoulder (Fitmore® p < 0.01 and CLS® p < 0.01) within each implant design, indicating a proximal fixation of both stems (Fig. 4). Comparison of relative micromotions of the proximal shoulder (Fitmore® 7.50 ± 0.65 mdeg/Nm vs. CLS® 7.21 ± 1.21 mdeg/Nm; p = 0.84) with the lesser trochanter level revealed no significant differences within either stems (Fitmore® p = 0.82 and CLS® p = 0.62).

Fig. 3
figure 3

Rotational stability of the Fitmore® and CLS® stems in mdeg/Nm

Fig. 4
figure 4

a, b Mediolateral bending behavior of the Fitmore® stem (a) compared with the CLS® stem (b)

Setup B: varus–valgus torque: bending behavior of the stem

The Fitmore® stem showed absolute micromotions of 0.26 ± 0.10 mdeg/Nm at the proximal shoulder and −0.01 ± 0.08 mdeg/Nm at the distal tip of the stem (Fig. 4). Proximal motion of the Fitmore® stem was directed medially but changed to lateral at the distal part of the stem. In contrast, the CLS® stem reacted with medial motions at both measuring points (proximally 0.55 ± 0.09 mdeg/Nm and distally 0.40 ± 0.36 mdeg/Nm). Based on these differences in micromotions, the CLS® stem showed significantly higher flexibility than the Fitmore® stem (CLS® 0.95 ± 0.31 mdeg/Nm vs. Fitmore®0.25 ± 0.03 mdeg/Nm; p < 0.01).

Discussion

Bone remodelling, as with clinically observed formations of cortical hypertrophies when using the Fitmore® stem, is already well known in THA. In the nineteenth century, Julius Wolff precisely described remodelling processes in bone tissue [19]. Wolff’s law states that depending on the applied load, the bone structure remodels itself over time to become more resistant. This is followed by secondary changes to the external cortical portion of the bone. In simple terms, the law is “load it or lose it”. Implantation of a hip stem will fundamentally change the load transfer to the bone [20]. One evident example of altered bone remodelling after implantation is the Zweymüller stem (e.g. Zimmer Alloclassic®), which has been clinically shown to be hypertrophic distally and osteolytic metaphyseally [21]. One study using the same experimental setup as used in this study showed that the Zweymüller stem has a distal fixation (location of load transfer) [16]. As expected from this distal fixation, parts of the proximal metaphysis will be isolated from loading, which may explain the proximal atrophy found in clinical cases [16].

A major limitation of our study is the use of synthetic femora rather than human bones. Similar measurements in living humans are obviously not practicable. Human specimens would offer more realistic in in vitro conditions. However, synthetic femora were used to create standardised and comparable conditions. The use of primary stability measurements to perform secondary stability observations is another study limitation. Similar measurements, including bone ingrowth, are not yet possible to perform in vitro. However, it has been shown that a causal relationship between initial implant fit and osseointegration seems to exist [14].

The CLS® stem was chosen as a reference due to its excellent clinical long-term survival [14, 22]. Compared with the relatively new Fitmore® stem, the CLS® stem has a 28-year history, and cortical hypertrophies are uncommon [14, 23]. Both stems showed low micromotions within the proximal part of the stem during rotational torque application (Setup A), similar to other stems [16, 24]. Comparable rotational stability results might indicate a similar proximal anchoring behaviour for both designs. Considering the proximal fixation of both stem types and the explained correlations of distal fixation and hypertrophy, rotational stability of the Fitmore® stem might not be a plausible explanation for clinically observed formations of hypertrophies. Under mediolateral torque application (Setup B), the CLS® stem acted more flexibly than the shorter Fitmore® stem. Presumably, the CLS® closely follows bending of the bone, whereas the shorter Fitmore® stem is more rigid. This may be due to an enlarged axial cross section of the stem and therefore an enlarged implant stiffness of the shorter Fitmore®.

Many clinical studies have observed cortical hypertrophy (mainly diaphyseal within Gruen zones 3, 5 and 6) [21] and atrophy (mainly metaphyseal within Gruen zones 1 and 7) [20, 21]. Furthermore, stem stiffness is one important design variable for bone remodelling processes (e.g. rigid cobalt-chromium alloy vs. flexible Titanium alloy) [6, 8, 25, 26]. Stems with greater flexibility reduce proximal bone loss and prevent cortical hypertrophies at the distal part of the stem [26]. Other aspects are fixation type (cemented vs. cementless) [27] and location (diaphyseal vs. metaphyseal vs. hybrid) [25] and implant surface modifications (e.g. porous-coated, hydroxyapatite layers, etc.) [28]. Short stems were introduced to facilitate small incision surgery as well as to achieve proximal load transfer within the metaphysis and therefore reduce distal hypertrophies and stress shielding [9, 25, 29] compared with traditional stems [30]. However, distal hypertrophies and proximal atrophy might also occur within short stems [31]. We therefore believe that the rigid bending behaviour of the Fitmore® stem may induce higher intrafemoral stresses, which might lead to the formation of cortical hypertrophies.

In conclusion, both stems showed low relative micromotions at each site and a similar metaphyseal fixation within the rotational torque test in vitro. Therefore, both the Fitmore® and CLS® stems may be adequate with respect to primary fixation. However, the stems exhibit different biomechanical behaviour under varus–valgus torque. Stem rigidity or flexibility needs to be considered, as this may influence load transfer at the implant–bone interface and so possibly affect long-term clinical bone-remodelling processes. Shorter and thicker stems, in particular, are generally confronted with increased rigidity. Further clinical studies may be necessary to follow the Fitmore® stem with respect to patient- or surgery-related parameters, such as frequency and severity of cortical hypertrophies, and the clinical long- term outcome of this stem.