This article focuses on the interpretation of new practical development in computed tomography (CT). In liver imaging, diagnostic accuracy, potential pitfall and limitation by perfusion CT, dual-energy CT, and volume shuttle helical scan are addressed. And dose reduction with these new CT technologies is also addressed.

Perfusion CT

CT can be used not only to image anatomical structure, but also to analyze liver function. Perfusion CT is performed by the acquisition of serial images after the administration of a bolus of iodinated contrast agent, enabling detailed analysis of liver hemodynamics. This CT technique is designed to calculate hepatic blood flow and portal blood flow using a color-encoded display of parameters from the liver time–density curve (TDC), with iodine contrast agent. Thus, liver functional maps are calculated by the input function based on regions of interest (ROIs) set on the aorta and portal vein.

Sahani et al. [1] reported perfusion CT as a feasible technique for quantifying tumor vascularity and angiogenesis in advanced HCC. Perfusion CT is obtained by monitoring the first pass through the vasculature after bolus injection of an iodinated contrast agent, and depicts tumor vascular physiology in patients with HCC. Following advances in imaging techniques and post-processing software, a deconvolution-based method has become widely used for liver perfusion CT imaging [2, 3]. There are two vascular inputs (hepatic arterial flow and portal flow) to the liver; however, the deconvolution-based method enables analysis of liver perfusion on CT. Enhancement of the abdominal aorta is used as a substitute for hepatic artery input, because the hepatic artery is too small to obtain a reliable TDC.

Hepatic perfusion CT enables analysis of liver function, and measurement of the following: tissue blood flow (TBF; mL/min/100 g); tissue blood volume (TBV; mL/100 g); mean transit time (MTT; s), which is the average time for blood elements to traverse the vasculature from arterial inlet to venous outlet (proportional to perfusion pressure); and hepatic arterial fraction (HAF; %), which is the ratio of arterial perfusion to total liver perfusion. The passage of contrast material through blood vessels enables the calculation of a TDC. The impulse residue function (IRF) of a localized bolus in hepatic blood flow is fundamental to perfusion CT. The passage of contrast agent enables the calculation of a TDC, and is shown as the IRF.

Mathematically, IRF is calculated as a tissue curve. This principle is shown as follows: [3].

$$ \begin{aligned} {\text{Q}}\left( t \right)& = {\text{ TBF}} \times {\text{I}}\left( t \right) \, \int {\text{R}}\left( t \right) \hfill \\ {\text{I}}\left( t \right) &= \, \alpha {\text{Ca}}\left( t \right) + \left( { 1- \alpha } \right){\text{ Cp}}\left( t \right)\\ \end{aligned} $$

where Q(t), TDC in the liver; TBF, tissue blood flow; I(t), iodine concentration flowing through the liver; ∫, convolution; R(t), impulse residue function; α, hepatic arterial fraction (HAF); Ca(t), iodine concentration of hepatic artery; Cp(t), concentration of contrast medium of portal vein.

Positron emission tomography (PET) also provides information regarding tissue perfusion, including that of the liver, but the spatial resolution is low [4]. The application of MRI is limited because of difficulty in quantifying the concentration of contrast agent in the tissue and its high sensitivity to respiratory movement [5, 6].

Xenon CT, which was developed during the late 1970s and 1980s, has been used to obtain information on cerebral blood flow following the inhalation of non-radioactive xenon gas [7, 8]. Xenon CT enables quantitative evaluation of hepatic arterial and portal-venous blood flow [911]; however, specialized equipment is required. Hashimoto et al. [12] reported a good correlation between hepatic TBF quantified by xenon CT and that by perfusion CT with iodine. In the clinical setting, perfusion CT has greater availability than xenon CT because dynamic CT with iodine is an established method for investigating liver tumors.

Clinical impact of hepatic perfusion CT

Analysis of the microcirculation by perfusion CT may be useful in the clinical situation; e.g., in the evaluation of response to therapy. Perfusion CT can be used for detection of tumor angiogenesis and in assessing response to antiangiogenic treatment for various cancers [1316]. The greatest impact of perfusion CT with regard to the liver is in assessment of patients with HCC, especially that of tumor response to antiangiogenic drugs [17, 18].

Standard chemotherapy is not a treatment method for advanced HCC because of its chemo-resistance. Sorafenib (Nexavar®, Bayer Schering Pharma AG, Berlin, Germany), an orally active multikinase inhibitor with effects on tumor-cell proliferation and tumor angiogenesis, is a promising therapy that demonstrates superior survival in advanced HCC [19]. Therapeutic response has been assessed based on tumor size measurement using the Response Evaluation Criteria in Solid Tumors (RECIST). This assessment for response to Sorafenib therapy by RECIST, however, has highlighted the limitations associated with morphologic measurement. Perfusion CT quantifies not only tumor biologic behavior, but also changes in liver parenchyma (without tumor) caused by Sorafenib therapy in patients with advanced HCC (Figs. 1, 2).

Fig. 1
figure 1

A 60-year-old man with an advanced hepatocellular carcinoma (HCC). A Advanced HCC pre-Sorafenib-therapy on dynamic CT. a Pre-enhancement CT, b arterial phase, c portal phase, d equilibrium phase. Dynamic liver CT showed the large HCC with hypervascularity at the arterial phase (a) and washout at the equilibrium phase (d) in right lobe and segment 4 of the liver. B Advanced HCC pre-Sorafenib-therapy on perfusion CT. a Tissue blood flow (TBF; mL/min/100 g), b tissue blood volume (TBV; mL/100 g), c mean transit time (MTT; s), d hepatic arterial fraction (HAF; %). Each parameter of HCC and liver parenchyma was TBF; 297, TBV; 28, MTT; 43, HAF; 80 and TBF; 179, TBV; 27, MTT; 12, HAF; 17, respectively. This case suggests the ability of perfusion CT to demonstrate advanced hypervascular HCC and liver parenchyma.

Fig. 2
figure 2

A 60-year-old man with an advanced hepatocellular carcinoma (HCC). A Advanced HCC post-Sorafenib-therapy (4 months) on dynamic CT. a Pre-enhancement CT, b arterial phase, c portal phase, d equilibrium phase. Early enhancement of large HCC in right lobe and segment 4 of the liver at the arterial phase (a) decreased after Sorafenib therapy, in comparison to pre-therapy (Fig. 1A). B Advanced HCC post-Sorafenib-therapy (4 months) on perfusion CT. a Tissue blood flow (TBF; mL/min/100 g), b tissue blood volume (TBV; mL/100 g), c mean transit time (MTT; s), d hepatic arterial fraction (HAF; %). Each parameter of HCC and liver parenchyma was TBF; 152, TBV; 14, MTT; 7, HAF; 87 and TBF; 84, TBV; 15, MTT; 16, HAF; 9, respectively. The decreases of TBV and TBF of HCC after Sorafenib therapy reflected antiangiogenic response, and was TBV of HCC was correlated with degree of arterial enhancement in liver dynamic CT.

For detecting and characterizing hepatic tumors in the clinical setting, perfusion CT can be followed by liver dynamic CT. Tissue enhancement curves can be mathematically fitted to a compartmental or deconvolution model, and quantitative parameters such as TBV, TBF, MTT, and HAF can be obtained. The linearity between tissue attenuation and iodine concentration contributes to the estimation of quantitative parameters.

Liver perfusion CT has generally been performed for a single section [12, 20, 21]; however, the high temporal resolution of volume helical shuttle (VHS) scanning enables whole liver perfusion imaging with multi-sections.

Liver cirrhosis and perfusion CT

Hepatic arterial perfusion is increased in patients with cirrhosis [21, 22]. Hashimoto et al. [21] reported the mean TBF, TBV, MTT, and HAF values of normal subjects by a deconvolution algorithm to be 103.9 ± 18 mL/min/100 g, 12.5 ± 2.0 mL/100 g, 11.1 ± 1.6 s, and 18.4 ± 5.6%, respectively. In contrast, the mean TBFs of patients with Child A, B, and C were 95.1 ± 24, 86.7 ± 29, and 75.5 ± 6.5 mL/min/100 g, respectively. Bader et al. investigated perfusion CT using a maximum gradient algorithm. The comparison between a deconvolution algorithm and a maximum gradient algorithm in perfusion CT of normal subjects is shown in Table 1. Evaluation of local liver function by perfusion CT may be important in some clinical situations, such as planning of surgery and therapy evaluation.

Table 1 Comparison between two perfusion CTs in normal subjects

Limitations and solutions for perfusion CT

Perfusion CT has the major drawback of radiation exposure, but Adaptive Statistical iterative Reconstruction (ASiR, GE Healthcare, Milwaukee, WI, USA) provides similar quality at lower dose compared with the usual filtered back-projection algorithm. If ASiR is used at the same dose, better image quality is obtained than that using the back-projection algorithm.

Perfusion CT has a limitation when scanning small targets because the obtained color-mapping image may not enable visualization of a small lesion. This limits the number of such studies permissible in the course of a clinical trial; however, implementation of low-dose scanning protocols may lead to more widespread and frequent use of this technique.

VHS

The VHS technique is a breakthrough technique for liver imaging that offers new clinical opportunities such as dynamic blood flow studies, leading to a new concept termed “four-dimensional (4D) CT.” In addition, when VHS is used with ASiR, the acquired images can be produced without artifacts. ASiR enables a reduction in radiation dose compared with the usual filtered back-projection algorithm; alternatively, when employed at the same dose as that for the filtered back-projection algorithm, ASiR provides better image quality. We performed the world’s first VHS CT scan for dynamic CT angiography and also for liver- and brain-perfusion studies, with high temporal resolution. VHS provides almost real-time hemodynamic change by shuttling the CT scanning cradle back and forth during scanning, and also enables wider coverage for complete organ imaging: >120 mm longitudinally. In contrast, conventional helical scans are performed at a constant table speed, meaning that the exposure is ineffective during acceleration of the table, and temporal resolution of conventional helical scan is lower than that of VHS.

The VHS technique requires no extra exposure because the acquisition of CT data is effective even with varying table speed. The whole liver can be scanned 24 times (12 round trips) during a 40-s breath hold, enabling the technique to be used for perfusion CT. Because VHS is a fast scanning technique (e.g., approximately 2 s/whole liver scan) with improved temporal resolution, even during table acceleration/deceleration; therefore, it can be used to investigate the relation between tumors and feeding arteries in detail. VHS could possibly be used as a navigation tool for therapy, such as transarterial chemoembolization (TACE) in HCCs; it could also be used to perform physiological assessment of the liver.

VHS has potential for use in 4D imaging; that is, producing a movie-like image, with a wide volume coverage. With VHS, the liver is repeatedly scanned in axial acquisitions, thereby providing coverage of the whole liver with reciprocating movement. Hypervascular HCC can be analyzed by using several whole liver scans, such as 12 phases in the arterial phase of dynamic CT (Fig. 3). Blood flow information, such as a blood flow direction, can also be obtained by VHS scanning.

Fig. 3
figure 3

A 50-year-old man with a small hypervascular hepatocellular carcinoma (HCC). Volume helical shuttle (VHS) scan was performed at 12 phases in the arterial phase of dynamic CT. Arterial enhancement pattern of HCC (arrow) in liver segment 4 was analyzed in detail.

The VHS technique can be used for perfusion studies of the liver and for 4D CT angiography. When used with CT angiography, the main abdominal vessels can be visualized to almost the same degree as with conventional digital subtraction angiography (DSA) (Fig. 4).

Fig. 4
figure 4

A 60-year-old man with multiple hypervascular hepatocellular carcinomas (HCCs). The VHS technique can be used for four-dimensional (4D) CT angiography. This image was obtained at a coronal plane. When used with CT angiography, the hepatic arteries and multiple HCCs (arrows) could be visualized to almost the same degree as with conventional digital subtraction angiography (DSA) examination. Arrow head shows stain of the gallbladder.

Dual-energy CT

Dual-energy CT is a promising technique used to obtain material-specific images. It enables differentiation of materials and tissues in images obtained based on the differences in iodine and water densities; the resulting CT images are reconstructed using two different energy spectra. Techniques based on the concept of dual-energy have been explored over the past 30 years [23]. Dual-energy CT can be performed with multidetector CT (e.g., 64-detector row CT) in current clinical settings.

An advantage of dual-energy CT is its ability to perform scanning at low tube voltages (kVp). Because of the photoelectric effect, the attenuation value of iodine is higher when low-voltage X-rays (such as 80 kVp) are used [24, 25]. Iodine contrast agents are more conspicuous at low than at high tube voltage settings, but may result in high image noise, particularly in large patients [26]. The increased noise of low-voltage X-ray scanning may be problematic because the density of the liver is relatively homogeneous. Therefore, maintaining an appropriate signal-to-noise ratio (SNR) in low-voltage raw data is an important factor in dual-energy scanning.

There are four types of dual-energy CT systems, although two of these have only recently become commercially available. These dual-energy techniques are used routinely in clinical abdominal imaging.

GE Healthcare incorporates two advanced technologies for dual-energy CT: a gemstone detector in the scanner, which uses a garnet-based detector material; and fast kV switching, in which the kVp levels are switched with every other view. Gemstone spectral imaging is designed to reduce afterglow in the detectors. Intermittently and alternately, 80 and 140 kVp X-rays change for each view during fast kV switching. Only one X-ray tube is used instead of two, which produces X-rays at two different energies (80 and 140 kVp) in alternative views; thus, dual-energy CT can be performed with “projection base” dual-energy reconstruction with beam-hardening reduction, to produce monochromatic images. This fast kV switching method requires specialized technology to control the high frequency and high voltage of the high voltage generator. Without this technology, dual-energy scanning is disadvantaged by limited energy separation.

The Siemens dual-energy CT technique has two X-ray sources and two X-ray detectors offset by 90°, with a single CT gantry. This system has the advantage of better energy separation compared with the fast kV switching method; however, the dual-source CT system has the disadvantage of cross-contamination by scattering, and “image base” dual-energy reconstruction is performed without beam-hardening reduction. Thus, the lower-energy blended images exhibit a greater beam-hardening effect compared with that for other techniques.

Philips and Toshiba employ a slow kV switching system for dual-energy CT at alternative rotation, although this system is not commercially available. Slow kV switching has the advantage of better energy separation compared with other methods, but has the disadvantage of difficulty in performing helical scans because of the alternating rotations of the X-ray tube, as well as problems concerning patient misregistration artifact.

Several previous studies have employed dual-energy CT for clinical use in the abdominal region [2729]. Dual-energy CT provides an accurate, noninvasive method of quantitating liver iron [30], offers more specific tissue characterization on CT, and can improve the assessment of vascular disease [31] (Fig. 5A). Dual-energy CT has been shown to have improved ability to detect contrast agent and to distinguish high-density substances created by iodine from those created by calcium or other substances, and thus is useful for the detection and characterization of renal stones [32]. Moreover, postprocessing algorithms enable subtraction of iodine maps from dual-energy CT data (e.g., subtraction of calcification) to create a virtual noncontrast image. This method may increase diagnostic accuracy, especially for detecting renal stones from the enhanced CT data alone.

Fig. 5
figure 5

A 60-year-old man with dual-energy CT imaging for hepatic vessels and two hepatocellular carcinomas (HCCs). A Assessment of hepatic vessels by dual-energy CT: monochromatic images with 55 keV (low tube voltage) and 80 keV (conventional tube voltage) in the arterial phase and the portal-venous phase of dynamic CT. a Arterial phase with 55 keV, b arterial phase with 80 keV, c portal-venous phase with 55 keV, d portal-venous phase with 80 keV. All images were obtained as matrix 512 × 512, 640 mA and 5 mm slice thickness. ad Enhancement effects of hepatic artery (arrow) and portal vein (arrow head) with 55 keV were higher than those with 80 keV in both the arterial phase and the portal-venous phase of dynamic CT. B Assessment of HCC by dual-energy CT: monochromatic images with 55 keV (low tube voltage) and 80 keV (conventional tube voltage) in the arterial phase and the equilibrium phase of dynamic CT. a Arterial phase with 55 keV, b arterial phase with 80 keV, c equilibrium phase with 55 keV, d equilibrium phase with 80 keV. All images were obtained as matrix 512 × 512, 640 mA and 5 mm slice thickness. a, b Enhancement effects of hypervascular HCCs in liver segment 5 (arrow) and 6 (arrow head) with 55 keV were higher than those with 80 keV in the arterial phase of dynamic CT. c, d Equilibrium phase of dynamic liver CT showed washout (defined as low density relative to surrounding liver parenchyma) of HCCs in liver segment 5 (arrow) and 6 (arrow head). The contrast between HCC and liver parenchyma on the low tube voltage (55 keV) image was visually higher than that on the conventional tube voltage (80 keV) image.

Fused images can be created for dual-energy CT by using a weighting factor or the blending method for contrast enhancement [33, 34], or monochromatic images can be reconstructed by dual-energy reconstruction. Blended or monochromatic images provide some improvement in contrast. Dual-energy material decomposition images can be successfully used for routine abdominal imaging: they offer improved visualization of vascular structures and abnormal calcifications such as renal and gall stones as well as depiction of soft tissue structures such as the kidneys, liver, and spleen. Material decomposition images have the potential to increase sensitivity and specificity in the detection of pathology.

HCC diagnosis by dual-energy CT

CT images at 80 kVp or the equivalent 55 keV (kilo-electron voltage) monochromatic images may show a higher contrast-to-noise ratio for hypervascular HCCs because the attenuation value of iodine increases with the use of low-voltage X-rays [35]. Pure 80 kVp data acquired from a dual-energy CT scanner produce greater differences in attenuation between hepatic lesions and surrounding liver, and potentially improve the assessment and detection of liver tumors. Finally, by utilizing a dual-energy CT system with dual-energy scanning, there exists the potential to improve the assessment of hypervascular liver tumors on lower keV monochromatic images (Fig. 5B).

Dose reduction

CT image quality is strongly proportional to radiation dose. The dose level should be known when CT scanning is performed. Several parameters, such as milli-amperage (mA), exposure time, peak kilo-voltage (kVp), and pitch, are adjusted based on the relation between the benefit and risk to patients. In general, CT characteristics are largely determined by peak tube voltage (kVp), which determines the upper limit of X-ray energy, and tube current (mA). Images obtained with low-voltage X-rays contain a high degree of noise, although this is alleviated by increasing the tube current. Beam-hardening artifact is greater on low-voltage than on 120 kVp X-ray images, so a technique with reduced beam hardening in the images is desirable, such as monochromatic images made by dual-energy scanning.

Conclusion

This article summarizes the current and potential applications for perfusion CT, dual-energy CT, and VHS. CT is a powerful tool for measuring bodily architecture and function. Perfusion CT offers functional imaging that can produce important data regarding blood flow, permeability, leakage space, and blood volume. Thus, perfusion CT is feasible for monitoring the response after antiangiogenic therapy, such as Sorafenib therapy.

The VHS technique can image dynamic blood flow, leading to the new concept of four-dimensional CT, which is a movie-like image with a large volume of coverage. It also provides blood flow information, such as blood flow direction. Therefore, this technique is particularly useful for obtaining information regarding back-flow in patients with vascular stenosis or occlusion.

Dual-energy CT may offer increased detection of hepatic lesions and specific characterization in the abdominal region, and can improve the assessment of vascular disease. The study of dual-energy technology has just started in clinical settings, therefore, this technique requires extensive studies for the confirmation of clinical usefulness.

Optimization of imaging protocols for perfusion CT, dual-energy CT, and VHS can improve the utility of these techniques. In particular, low-dose CT technique is required for functional imaging, when several scans of the target lesion and organ are performed. In the future, it will be necessary to reduce the radiation dose level of abdominal CT imaging for a variety of diagnostic purposes, although new advanced technologies enable larger volumetric datasets and promising functional data to be obtained using CT.