Introduction

Besides the standard myocardial perfusion imaging (MPI) with single photon emission computed tomography (SPECT), noninvasive assessment of absolute myocardial blood flow (MBF) with positron emission tomography (PET) in milliliters per minute per gram and myocardial flow reserve (MFR) derived from stress to rest flow ratio has been continuously emerging as a clinical tool to detect the early stage of coronary atherosclerosis, stratify risks for cardiac events, and predict patient outcomes [17]. It has been reported that adding PET flow information to the relative perfusion imaging provides incremental diagnostic value for detection of coronary artery disease (CAD) [810] as PET flow itself can be an independent indicator of multivessel CAD [11, 12], representing a comprehensive tool to evaluate microvascular dysfunction with or without conventional cardiac risks [13, 14].

Although PET flow quantitation has been clinically marked as a powerful tool for diagnosis and prognosis of CAD at various clinical stages [1517], in reality, it is not generally available in many areas around the world (e.g., Asian countries). The main challenge is posed by the fact that the production of PET tracers demands either high start-up costs to install an on-site cyclotron within the hospital (15O-water and 13N-ammonia) which may only be economically affordable by large medical centers or relatively high ongoing costs to purchase commercial generators on a monthly basis (82Rb) which may not be economically realistic for clinical utilization. As a result, PET flow imaging has not been embraced by healthcare systems in these areas. Myocardial perfusion SPECT with 99mTc-labeled myocardial perfusion tracers (99mTc-sestamibi and 99mTc-tetrofosmin) remains the clinical standard for MPI worldwide [18]. Thus, flow quantitation with SPECT, when available, may be a practical solution to overcome challenges confronted by PET imaging and to warrant a widespread clinical utilization. In fact, modern SPECT and SPECT/CT instrumentation has been improved to attain high temporal resolution for dynamic data acquisition, although this unique capability in the past was not well investigated to design relevant protocols for clinical utilization. Furthermore, iterative reconstruction techniques with full physical corrections [19, 20] and the past experience of PET flow imaging collectively affirm development of a quantitative method for MBF quantitation with SPECT imaging.

This work presents a new method to achieve SPECT flow quantitation with a widely used SPECT/CT scanner by utilizing information derived from quantified dynamic SPECT (DySPECT) images and a compartment flow model for flow calculation comparable to the established PET methods [2123]. An integrated SPECT imaging protocol was developed to acquire both flow and perfusion data with the same tracer injections. The diagnostic performance of SPECT flow and perfusion was assessed by comparing both parameters to luminal stenosis confirmed by invasive coronary angiography (CAG).

Materials and methods

DySPECT data acquisition

DySPECT data and CT images for attenuation correction (AC) were acquired on a commercially available dual-head SPECT/CT scanner (Symbia T, Siemens Medical Solutions, Malvern, PA, USA). The scanner was equipped with 0.95-cm thickness of NaI(Tl) crystal for gamma photon detection and functioned to perform fast back-and-forth or continuous gantry rotation with a maximum speed of 10 s per rotation. In a DySPECT scan, the L-shaped detectors were rotated from the patient’s right anterior oblique (RAO) to left posterior oblique (LPO) positions to acquire a full set of projections with 64 views within a 180°degree arc and returned from LPO to RAO to acquire another set of projection data, and so forth. A DySPECT scan consisted of 10 × 10 + 5 × 20 + 4 × 60 + 1 × 280 s for a duration of 12 min, and in total, 20 back-and-forth rotations were acquired to produce 1,280 projection views by using acquisition parameters of continuous mode, circular orbit, low-energy high-resolution (LEHR) collimator, 64 × 64 matrix, 1.78 zoom factor, 5.39 mm pixel size, and multiple energy windows (126–154 keV and 109–125 keV). CT images for AC were acquired with a low radiation dose CT scan (120 keV, 35 mA, 12 s).

DySPECT image reconstruction

DySPECT images were reconstructed with ordered subsets expectation maximization (OSEM, 3 iterations/16 subsets) [24] incorporated with full physical corrections for scatter, attenuation, isotope decay, collimator blur, and image noise [2529]. Prior to image reconstruction, scatter correction (SC) was performed by subtracting estimated scatter count density (counts in pixel) to generate scatter-corrected projection data with the formula as follows [25]:

$$ {S}_P=\frac{W_p}{2{W}_S}\cdot {P}_S $$
(1)

where S P is the estimated scatter count density in photopeak energy window (126–154 keV), and P S represents count density in the scatter window (109–125 keV). W P and W s are their energy window width. Each view of projection data was then normalized with the duration of their corresponding rotation and an exponential factor depicting isotope decay as follows:

$$ \frac{1}{D}\cdot ex{p}^{\left[0.693\times {t}_{R, A}/{T}_{T c99 m}\right]} $$
(2)

where D is the duration of a rotation, and t R,A is the time stamp relative to the starting point of the DySPECT scan and indexed by the rotation and projection angle. T Tc99m is the physical half-life of the 99mTc isotope. For AC, CT images were converted to 140-keV μ-map using the conversion formula as follows [26]:

$$ \begin{array}{c}\hfill {\mu}_{140}={\mu}_{water}\cdot \left(\frac{HU}{1000}+1.0\right)\ for\ HU<0\hfill \\ {}\hfill {\mu}_{140}={\mu}_{water}\cdot \left(0.665\cdot \frac{HU}{1000}+1.0\right)\ for\ HU\ge 0\hfill \end{array} $$
(3)

where HU is Hounsfield unit in CT images, and μ water  = 0.154 cm −1 is the attenuation coefficient of water for 140 keV photons. To implement AC, an attenuation matrix functioned to record photon attenuation for image pixels pointing to all projection angles was pre-calculated and then applied to pre-updated images in the forward projection step of OSEM. Prior to AC, μ-map and DySPECT images were fused to verify transmission/emission misregistration by searching unregistered heart locations in both images (e.g., any part of myocardium locations in lung fields of the μ-map). Manual realignment was performed to shift the μ-map for misregistration correction before generating the attenuation matrix. For resolution recovery (RR) of images, depth-dependent point spread functions were physically measured for the SPECT/CT system and applied in the forward projection step of OSEM [27]. Noise suppression (NS) for images was achieved by implementing a statistical method to model Poisson noise in the reconstruction process [28, 29]. For the remaining DySPECT images, the residual activity in the myocardium from the stress injection was estimated from images of the first frame and propagated to subtract the residual myocardial uptake in the subsequent frames.

Assessment of image resolution

To verify the spatial resolution of reconstructed images, seven point sources with 21.1 MBq 99mTc activity on average for each source were lined up 5 cm apart to create angularly varied distances (17.2–53.9 cm) during the DySPECT acquisition. Prior to image reconstruction, dynamic projection views were summed to create a single set of static projection data. Images were then reconstructed with OSEM + RR. To measure the spatial resolution, images were resampled to a smaller pixel size of 1.35 mm. Line profiles of 99mTc activity were then obtained by drawing straight lines across the center of point sources on orthogonal planes in resampled images. For each point source, line profiles were fitted with a Gaussian function to measure image resolution presented by the full-width at half-maximum (FWHM) in millimeters.

Determination of conversion factor to becquerels per milliliter

The conversion factor serves to translate the unit of counts per second per pixel (cps/pixel) in reconstructed images to the physical unit of becquerels per milliliter. An assessment of camera linearity was performed with a cylindrical phantom (radius 16 cm, height 20 cm) initially filled in 1,731.6 MBq 99mTc water solution. During 24.7 h, 14 DySPECT scans were acquired within every 1–3 h with the count rate dropping from 150 to 10 kcps in both detectors. Each dynamic projection datum was processed with SC and normalization and then integrated to create static projection data by weighting the duration of the individual rotation to the total time of the DySPECT scan. Images were reconstructed with additional corrections of AC, RR, and NS as previously described. A cylindrical region of interest (ROI) was drawn across five image slices (slice thickness 5.39 mm) from the center of the phantom as 102.7 cm3 and placed in phantom images to generate mean pixel values in the unit of counts per second per pixel. They were compared with known activity concentrations in the unit of becquerels per milliliter. A linear regression was performed to obtain the conversion factor as the slope of two variables.

SPECT flow quantitation

To quantify SPECT flow, we utilized the quantitative method built in a PET software package (FlowQuant, University of Ottawa Hospital, Ottawa, ON, Canada) [30]. The method employed a one-tissue compartment flow model to calculate kinetic parameters (K1 and k2) from the time-activity curve (TAC) obtained in regions of the left ventricle and atrium and myocardial tissue as follows:

$$ {C}_{myo}(t)= FBV\cdot {C}_a(t)+\left(1- FBV\right)\cdot {K}_1{e}^{-{k}_2 t}\otimes {C}_a(t) $$
(4)

where K 1 (ml/min per g) and k 2 (min −1) characterize the rate of tracer uptake from blood to myocardial tissue and the rate of tracer washout from the tissue, respectively. C myo (t) is the measured activity concentration in the myocardium obtained from DySPECT images, assumed to consist of arterial blood input C a (t) and true myocardial uptake as a convoluted function of K1, k2, and Ca(t). FBV refers to the fractional blood volume in Cm(t) coming from Ca(t), and (1-FBV) is the rest of the fraction contributed from the true myocardial uptake. By applying the process of curve fitting, K1, k2, and FBV were solved to numerical values. MBF (ml/min per g) was then calculated from K1 with additional compensation for tracer extraction, E, in the myocardium as follows:

$$ {K}_1= MBF\cdot E(MBF)= MBF\cdot \left(1- a\cdot {e}^{-\beta / MBF}\right) $$
(5)

where α and β are physiological parameters derived from the effective capillary permeability surface (PS) area product (ml/min per g) counting E as a function of MBF [31, 32]. This study utilized the 99mTc-sestamibi (MIBI) extraction fraction curve reported from an animal study (α = 0.87 and β = 0.44) [23, 33]. To compensate for the physiological variance at rest stage, rest MBF was additionally corrected for the baseline heart rate and blood pressure by a factor of rate pressure product/10,000 bpm × mmHg [35]. MFR was calculated as the ratio of stress MBF (SMBF) to rest MBF (RMBF). Flow values of the entire myocardium were then presented in polar maps to calculate global MBF and MFR and for regional MBF and MFR in three coronary territories.

Patient population

This study included 21 patients with suspected or known CAD who had received DySPECT and ECG-gated SPECT (GSPECT) scans and CAG procedure (13 patients with ≥50 % stenosis in any vessel and 8 with patent arteries or <50 % stenosis). Each research subject agreed to sign a written consent approved by the Institutional Research Ethics Board in Show Chwan Memorial Hospital, Changhua, Taiwan.

Integrated SPECT imaging protocol

This study utilized the same MIBI injections for DySPECT and GSPECT scans. After proper patient preparation and giving the patient 500 ml water, a CT scout scan was performed to position the patient’s heart inside the SPECT field of view. Dipyridamole vasodilator was infused for 4 min to achieve hyperemic myocardial flow with a dose of 0.56 mg/kg. After waiting for an additional 3 min, DySPECT was started and followed by a 370 MBq MIBI dose injection within 10 s. A low-dose CT for AC then followed for stress AC. After the completion of the stress DySPECT scan, the patient got off the scanner and returned for the routine clinical stress GSPECT within 45–60 min. After 3 h, at physiological rest state, the DySPECT scan was repeated with the same imaging setup and an injection of 1,000 MBq MIBI. A second low-dose CT scan then followed for rest AC. The routine clinical rest GSPECT was performed within 45–60 min thereafter. The integrated SPECT imaging protocol is illustrated in Fig. 1.

Fig. 1
figure 1

A comprehensive imaging protocol integrating DySPECT with clinical GSPECT using the same rest and stress MIBI injections

Analysis of perfusion images

Perfusion images were processed without and with AC (combined with SC, RR, and NS) using the standard platform provided by the vender (syngo, Siemens, USA). Non-AC and AC image sets were interpreted simultaneously by two experienced nuclear medicine physicians to score perfusion defects with a 5-point scale (0 = normal, 1 = mildly reduced, 2 = moderately reduced, 3 = severely reduced, and 4 = absent uptake) for 17 segments [35]. Summed stress score (SSS) and summed rest score (SRS) were calculated by adding individual segmental scores and utilized to derive summed difference score (SDS) as SSS − SRS.

Invasive coronary angiography

CAG was performed according to the standard clinical procedure for all study subjects within 21 ± 18 days (0–73). The percentage of luminal diameter stenosis in coronary arteries was quantitatively measured with a commercial software (Innova 2100-IQ, GE Healthcare, Wauwatosa, WI, USA) operated by an experienced cardiologist. Positive CAD was defined as at least one lesion causing ≥50% luminal narrowing (obstructive lesion) in any vessel of left anterior descending (LAD), left circumflex (LCX), and right coronary artery (RCA) territories, otherwise non-CAD as patent or only <50 % luminal narrowing (nonobstructive lesion) in all of the three coronary arteries.

Statistical analysis

Continuous variables from image analysis were expressed as mean ± standard deviation, tested by Student’s t test and analyzed with receiver-operating characteristic (ROC) statistics. Noncontinuous variables from patient clinical characteristics were expressed in number and percentage. Using CAG findings as the reference standard, patient-based (PB) and vessel-based (VB) ROC analyses were performed to retrospectively analyze the area under the curve (AUC) for SMBF, MFR, SSS, and SDS. The associated cutoff value to separate normal and abnormal groups for each index was then determined to generate largest AUC. Pairwise comparison for flow and perfusion AUCs was conducted to evaluate the difference of diagnostic performance between quantitative flow and perfusion. A p value <0.05 was considered statistically significant.

Results

When images were reconstructed with OSEM only, the image resolution (FWHM) for seven point sources were 17.1–17.8 (mean = 17.4), 11.9–24.3 (18.5), and 15.7–21.8 mm (18.7) in x, y, and z directions as highly distance-dependent and non-isotropic. With OSEM + RR, the image resolution was improved to being less dependent on distance and more isotropic as 8.6–10.1 (9.6), 9.1–10.9 (9.7), and 7.7–8.8 mm (8.2) in three directions, respectively (Fig. 2). Line profiles for point sources obtained from OSEM and OSEM + RR images overall had excellent correlation coefficients (R 2 > 0.99) as regressed with the Gaussian function.

Fig. 2
figure 2

Top: Schematic illustration of SPECT scan with camera gantry rotated around 7 point sources from positions of RAO, LAO, anterior (ANT), lateral (LAT) to LPO and returned from LPO to RAO, and so forth. d1 and d7 are distances of point 1 and point 7 from detector surfaces. Distances of other sources are in between. Middle: The reconstructed images with OSEM or OSEM + RR for 7 points. Bottom: Corresponding image resolution presented in FWHM (mm)

The relation of count rate of detectors and 99mTc activities in the phantom was completely linear (R 2 = 0.999) as no count loss was verified up to 150 kcps count rate (Fig. 3). Mean pixel values in reconstructed images were perfectly linear to activity concentrations in the phantom (R 2 = 0.999) to give the conversion factor as 60,264 Bq/ml per cps/pixel.

Fig. 3
figure 3

Top: Linear regression of SPECT camera count rate and 99mTc activity in the phantom. Bottom: Linear regression of pixel value in reconstructed images in cps/pixel and true 99mTc concentration in Bq/ml in the phantom

With the injection of 370 MBq MIBI for stress and 1,000 MBq for rest, the average count density in the left ventricular region during the phase of MIBI activity filling into the left ventricular cavity was 8.4 ± 3.2 cps/pixel for stress and 24.4 ± 8.1 cps/pixel for rest. The peak count rate reported from the SPECT camera system was 36.2 ± 9.8 kcps for stress and 103 ± 28.6 kcps for rest. As myocardial uptake was stabilized in the last frame, average count density in the myocardial region was 1.75 ± 0.17 cps/pixel for stress and 5.17 ± 0.80 cps/pixel for rest as presented in Fig. 4. An example of rest and stress reconstructed DySPECT images in the vertical long axis view and corresponding TACs obtained from the left ventricular cavity and the overall myocardium can be found in Fig. 5.

Fig. 4
figure 4

An example of 20 dynamic projection views acquired at RAO (a), ANT (b), ANT (c), and LPO (d) positions from a rest DySPECT scan. e Mean myocardial count density at the ANT position calculated for rest and stress DySPECT scans from all study subjects

Fig. 5
figure 5

An example of rest and stress TACs obtained from the left ventricular cavity and the overall myocardium in DySPECT images. Letters present time frames and their corresponding images in the vertical long axis view

Clinical characteristics of 21 study subjects are shown in Table 1 [34]. In the CAD group, 13 of them (62 %) presented ≥50 % luminal stenosis in any of the major coronary arteries as 5 one-vessel CAD, 4 two-vessel CAD, and 4 three-vessel CAD. There were a total of 38 obstructive lesions in 25 coronary artery territories as 21 lesions in LAD, 11 lesions in LCX, and 6 lesions in RCA. In the non-CAD group, only one nonobstructive lesion (stenosis = 30 %) was found in LAD, and the rest of the arteries were patent. The list of luminal stenosis and associated locations in vessels is shown in Table 2.

Table 1 Patient characteristics (n = 21)
Table 2 List of severity of luminal narrowing measured by CAG for CAD patients who had stenosis ≥50 % in any of the vessels with diameters ≥2.0 mm

Global kinetic, flow, and perfusion parameters for CAD and non-CAD groups are shown in Table 3. Global rest K1 and RMBF in these two groups were similar, while global stress K1, SMBF, and MFR in the CAD group were significantly lower than those of the non-CAD group (Fig. 6). SRS, SSS, and SDS for the CAD group were also lower, but the difference was not statistically significant. ROC curves of SMBF, MFR, SSS, and SDS for PB and VB analyses are displayed in Fig. 7. AUC and pairwise comparison between flow values and perfusion scores are shown in Table 4.

Table 3 Global kinetic parameters, flow values, and perfusion scores in CAD and non-CAD groups
Fig. 6
figure 6

Individual plot of global/vessel SMBF and MFR in non-CAD and CAD groups

Fig. 7
figure 7

PB and VB ROC curves of SMBF, MFR, SSS, and SDS to detect CAD using CAG as the reference standard

Table 4 Difference of ROC AUC between flow and perfusion for 21 patients with CAG

In PB analysis, MFR AUC was significantly greater than those of SSS and SDS. SMBF AUC was greater, but not statistically significant. The cutoff criterion to define a patient with CAD was ≤2.0 MFR with sensitivity (SEN) = 84.6 %, specificity (SPE) = 100 %, positive predictive value (PPV) = 100 %, and negative predictive value (NPV) = 80.0 % (power = 0.9). The cutoff criterion for SMBF was ≤1.96 ml/min per g with SEN = 84.5 %, SPE = 100 %, PPV = 100 %, and NPV = 79.6 % (power = 0.8). There was no statistical difference between the MFR AUC and SMBF AUC or SSS AUC and SDS AUC. The cutoff criterion for SDS was ≥2 with SEN = 69.2 %, SPE = 70.1 %, PPV = 78.9 %, and NPV = 58.3 % (power < 0.8). The cutoff criterion for SSS was ≥3 with SEN = 69.2 %, SPE = 62.5 %, PPV = 74.9 %, and NPV = 55.5 % (power < 0.8). Figure 8 shows an example of three-vessel CAD confirmed by CAG and uncovered as globally severe ischemic myocardium by stress MBF = 0.54 and MFR = 1.0, but defined as within normal limits by perfusion images.

Fig. 8
figure 8

A 71-year-old male patient with hyperlipidemia and no history of CAD complained of chest pain unrelated to increased physical activity. He was ordered to receive GSPECT with DySPECT scans for CAD evaluation. Top: The GSPECT scan revealed no evidence of myocardial perfusion abnormality on the stress/rest images, and no transient ischemic dilatation was found. Middle: The quantitative flow values from his DySPECT scan were SMBF = 0.54 ml/min per g, RMBF = 0.53 ml/min per g, and MFR = 1.0, which were extensively lower than the CAD thresholds (SMBF = 1.96 and MFR = 2.0) determined in this study. Bottom: Because of the concern of continuous symptoms, CAG was performed 31 days after the SPECT imaging. CAG confirmed severe luminal stenosis in LAD, LCX, and RCA coronary arteries as severe three-vessel CAD

In VB analysis, MFR AUC was significantly greater than those of SSS and SDS. SMBF AUC was also greater but not significant. The difference of MFR AUC vs SMBF AUC or SSS AUC or SDS AUC was overall not significant. The cutoff criterion for MFR to determine a vessel with CAD was ≤1.70 with SEN = 76.0 %, SPE = 84.2, PPV = 76.0, and NPV = 84.2 % (power = 0.95). The cutoff criterion for SBMF was ≤1.51 ml/min per g with SEN = 80.0 %, SPE = 76.3 %, PPV = 69.0 %, and NPV = 85.2 % (power = 0.95). The cutoff criterion for SDS was ≥1 with SEN = 60.0 %, SPE = 52.6 %, PPV = 45.5 %, and NPV = 66.6 % (power < 0.8). The cutoff criterion for SSS was ≥2 with SEN = 56.0 %, SPE = 62.8 %, PPV = 49.7 %, and NPV = 68.4 % (power < 0.8).

Discussion

In this study, we presented a new method to quantify MBF and MFR with MIBI myocardial perfusion tracer, a conventional SPECT/CT system capable of DySPECT scan, iterative reconstruction with full physical corrections for quantitative images in becquerels per milliliter, and kinetic modeling for flow calculation. Unlike those previous SPECT studies [3640], our method to derive MFR from absolute MBF (in ml/min per g) was highly comparable with PET methods [2123, 30]; thus, the performance of flow calculation may be more consistent in terms of avoiding physical interferences. In this study, the spatial resolution of DySPECT images was improved to twofold (from 18.2 to 9.2 mm on average) with OSEM + RR which provided an optimal image resolution to extract arterial and myocardial TACs. However, this improved but imperfect resolution still caused a significant amount of spillover from the left ventricle to the myocardium when MIBI activity was filling the left ventricular cavity (Fig. 5). The spillover effect was parameterized as FBV in the flow model (Eq. 4). The value was reported in a range of 0.2–0.6, depending on the physical size of the myocardium. This study did not include an additional kinetic parameter, k3, to depict the interactive activity of the MIBI tracer and mitochondria using a two-tissue compartment model. In fact, the flow calculation often failed as k3 was introduced, possibly due to the uncertainty from image noise, but further investigation would be needed for clarification. In contrast to PET imaging, which is proficient in simultaneously acquiring all projection (or sinogram) views while the tracer is still rapidly circulating in the patient, the nature of DySPECT imaging as fast as 10 s per rotation may not be proficient in tracing fast tracer movement in all projections as illustrated by the second frame of the LPO view in Fig. 4. However, this inconsistency was previously verified by a separate Monte Carlo simulation study and confirmed <10 % error in flow calculation with such temporal resolution [41]. In our study, MIBI doses for DySPECT imaging utilized clinical doses for GSPECT (stress: 370 MBq, rest: 1,000 MBq). The doses were sufficient to generate good image accuracy to extract TACs for patients with 20.3–36.2 body mass index (BMI, average 25.7). For patients with a BMI greater than this range, BMI-adjusted MIBI dose may be considered if image accuracy is substantially degraded. Furthermore, because of using the same dose injections for DySPECT and GSPECT scans, the imaging protocol did not request giving extra radiation to patients, but only adding extra imaging time for the DySPECT scan while the patient was waiting; thus, the integrated imaging SPECT protocol can be efficient for clinical utilization. In DySPECT scans, elevated uptake in liver usually became detectable about 90–120 s after MIBI injection as shown in Fig. 4. However, the impact of liver spillover to TAC accuracy should depend on its relative location to the myocardium and was varied patient by patient in our study. Under extreme conditions, the liver and myocardial inferior wall might share the same image voxel, and their boundary may become not differentiable. A protocol of routinely giving patients a substantial amount of water may be implemented to enlarge the stomach size in order to propel the liver away from the myocardium to eliminate the liver interference. As shown in Fig. 5, typical rest and stress TACs reached optimal clearance of MIBI activity in the left ventricular cavity and stabilized myocardial uptake at 5 min after MIBI injections; thus, it may be feasible to shorten the DySPECT scan time to 5 min to still provide sufficient data, but the impact on flow calculation needs further investigation.

In this study, the diagnostic performance of flow and perfusion in ROC analyses utilizing CAG as the reference standard favored flow to detect CAD. This finding is concordant with several previous PET studies using 13N-ammonia or 82Rb as flow tracers [810]. Figure 8 demonstrated a case with completely homogeneous perfusion images and no evidence of transient ischemic dilatation, but SMBF and MFR were overall below the CAD threshold, and CAG confirmed three-vessel CAD. Several studies also highlighted that myocardial perfusion SPECT was very limited to detect multivessel CAD (two or three) [42, 43]; thus, low diagnostic sensitivity (0.48) and underestimated CAD severity were observed. In our study, there was a relatively high prevalence of patients (8 of the 13 CAD patients; 62 %) with multivessel CAD confirmed by CAG. More likely the extent and severity of the perfusion abnormality with perfusion scores in this population would also be underrated to report lower AUC values than those values (0.74–0.84) published in the literature [44, 45]. Due to the limited ability of perfusion images to differentiate between severe multivessel disease and normal perfusion, differences of SSS and SDS in the CAD and non-CAD groups were not statistically significant in our study. However, statistical differences of these two groups were achievable by the analysis of flow parameters (SMBF: p = 0.021; MBF: p < 0.001). Consequently, adding SPECT flow quantitation to the routine perfusion assessment may be substantially valuable to rule out an underestimated perfusion abnormality and potential false-negative diagnosis of multivessel CAD. In the future, clinical implementation of SPECT flow quantitation may be very important, particularly for Asian countries where myocardial PET tracers are unavailable or economically unaffordable and myocardial perfusion SPECT is still used as the main clinical tool for CAD diagnosis. The new method developed in this study for SPECT flow quantitation therefore may offer a simple and effective solution to fill in the gap.

Study limitation

This study adopted MIBI extraction from a nonhuman study to calculate MBF from K1 for patients due to the reason that human extraction has not been available. Although not ideal, flow parameters with animal MIBI extraction were still able to differentiate between CAD and non-CAD patients. To address this imperfectness, our group is currently pursuing a comparative study of MIBI K1 to 13N-ammonia PET flow in the same healthy volunteers to obtain human MIBI extraction fraction data. Besides, this study did not address the impact of intra-scan patient motion to affect the accuracy of flow calculation, and motion correction algorithms were not implemented. A future study should be pursued to address this important issue. This study only included a limited number of patients (n = 21) who had received MIBI DySPECT, GSPECT, and CAG for comparison. In fact, the statistical power of ROC analyses for this small group of patients still met the minimal criteria for statistical satisfaction (power ≥ 0.80). An extended study with large cohorts should be pursued to reconfirm the incremental value of SPECT flow quantitation found in this study.

Conclusion

Flow quantitation with quantitative dynamic SPECT is a superior method over myocardial perfusion SPECT for CAD detection as perfusion images often underestimate the extent and severity of perfusion abnormalities in the population with multivessel CAD. SPECT flow quantitation with the conventional SPECT/CT system is effective to enhance CAD detection without adding extra radiation burden to patients.