Introduction

Parkinson’s disease is a neurological disorder associated with the loss of dopaminergic neurons from the substancia nigra and with an important dopamine depletion in the striatum [1]. A number of 123I agents have been developed for single-photon emission computed tomography (SPECT) imaging of the dopaminergic system to study the presynaptic dopamine transporter (DAT) binding. Commercially available pharmaceuticals labelled with 123I, such as β-2-carbomethoxy-3-(4′-iodophenyl)tropane (β-CIT) and fluoropropyl-CIT [2], have shown a high affinity for the DAT. Quantification of DAT SPECT imaging is suitable for discriminating Parkinsonian syndromes from other movement disorders, facilitating an early diagnosis of the disease [3], following up the progression of the disease [46] and for assessing the effects of treatment strategies [7].

To objectively assess striatal DAT binding, quantification is mandatory [8]. Nevertheless, the different degradations that are involved in the reconstruction process affect the quantification output. Thus, image-degrading effects such as attenuation, the spatially variant point spread function (PSF), scatter and the partial volume effect (PVE) have to be compensated to achieve an accurate quantification [913].

As regards to scatter, correction methods that model the three-dimensional (3D) scatter distribution through Monte Carlo (MC) simulation have been shown to perform an accurate quantification [14, 15]. We previously developed an absolute quantification method (AQM) for neurotransmission studies, which included a 3D iterative reconstruction algorithm with MC-based scatter compensation [12]. This method yielded good results in studies using 99mTc-labelled radioisotopes. However, in the 123I decay scheme, there are a few high-energy photons that have a non-negligible contribution to the final image [13, 16]. The effect of this high-energy contamination has to be compensated to improve the image quality and quantification.

Generally, PVE causes an underestimation of activity [17]. This effect is especially severe in small structures such as the basal ganglia. Given its significance, PVE correction methods have been developed in parallel by a number of research groups [1823].

Brain neurotransmission SPECT imaging with 123I is often performed with low energy high resolution (LEHR) parallel and fan-beam collimators [8]. In this work, a quantitative comparison of two imaging systems equipped with a parallel and a fan-beam collimator that employ different data acquisition protocols is made. To this end, an extension of the previous AQM was developed to include a MC-based scatter compensation for the high-energy contamination. This updated reconstruction algorithm uses a modelling of the high-energy PSF (hPSF) to accelerate the MC simulation. A handy post-reconstruction PVE compensation was also performed. The AQM with post-reconstruction PVE correction was used to assess the differences in the quantitative estimates between brain SPECT studies. As the quantification results had to be evaluated to test their reliability and accuracy, MC simulations with numerical phantoms were employed. Ideally, this quantification method should result in similar quantification values for both imaging configurations after correcting for all the degradations.

Materials and methods

The quantification method designed was evaluated through MC projections simulated for the following two camera/collimator combinations: Siemens E.CAM with LEHR parallel-hole collimator and Elscint HELIX with LEHR fan-beam collimator.

Numerical Phantom

Simulated projections were generated by using a numerical phantom obtained from a CT scan of the anthropomorphic striatal brain phantom (Radiological Support Devices, Long Beach, CA). The CT scan image consisted of 256 × 256 × 196 cubic voxels with a voxel size of 0.89 × 0.89 × 0.89 mm3. Brain tissue and bone were automatically segmented by thresholding the CT image, while the striatal cavities were manually drawn over the corresponding slices.

The non-uniform attenuation map was obtained by setting the appropriate attenuation coefficients to brain and bone depending on the energy of the simulated photons. Thirty different activity distributions were considered. For all these distributions, a constant value of 14 kBq/mL was assigned to the brain tissue. The striatal nuclei had a variety of values ranging from 15 to 98 kBq/mL to simulate 30 random values of the specific uptake ratio (SUR), which modelled normal and pathological distributions. SUR was defined as:

$${\text{SUR}} = \frac{{\overline {A_{\text{s}} } - \overline {A_{\text{o}} } }}{{\overline {A_{\text{o}} } }}$$
(1)

where \(\overline {A_{\text{s}} } \) is the mean activity concentration in the striatal region and \(\overline {A_{\text{o}} } \) is the mean activity concentration in a reference region located in the occipital area. Figure 1 shows one section of the activity distribution and the attenuation map used in the simulations.

Fig. 1
figure 1

A central slice of the activity distribution (left) and the attenuation map (right) of the striatal phantom. The striatal nuclei can be observed in the activity distribution

SimSET simulation

The SimSET MC code [24] was employed to simulate the SPECT projections. 123I photons with 159 keV (low-energy) and gamma-rays with higher energies (high-energy) were simulated separately and, finally, added together to obtain the total projections. For each camera/collimator system, the PSFs for the low-energy photons were calculated using the Gaussian function derived in [25], while the high-energy contamination effect was modelled using the hPSFs as described in [26]. These PSFs and hPSFs were included in SimSET to adapt the code to the 123I-labelled radioligands simulation case. Table 1 summarises the simulation parameters used for both imaging systems. The differences in the setting that can be observed in Table 1 are due to the different acquisition protocols used for each gamma camera. Energy windows were selected according to the values proposed by the manufacturers.

Table 1 Simulation parameters used for both SPECT systems

A total of 30 noisy projections for each SPECT gamma camera were generated by employing the randomly distributed values of SUR ranging from 0.1 to 6. All the simulated studies were considered to have approximately 3 million counts, thus mimicking real studies obtained with these two imaging systems.

Reconstruction algorithm

The AQM consists of a 3D iterative reconstruction algorithm based on the ordered subsets expectation maximisation (OSEM) algorithm [27], which includes a MC-based simulator as a forward projector for modelling the scatter component. The voxel values λ i are updated in each step of the reconstruction process using the expression [1214, 28]:

$$\lambda _i^{k + 1} = \frac{{\lambda _i^k }}{{\sum\limits_{j = 1}^{{\text{nbin}}} {t_{ji} } }}\sum\limits_{j = 1}^{{\text{nbin}}} {\frac{{t_{ji} p_j }}{{\sum\limits_{m = 1}^{{\text{nvox}}} {t_{jm} \lambda _m^k + s_j } }}} $$
(2)

where p j are the original projections, k is the iteration number, t ji is the transition matrix and s j is the scatter contribution estimated in bin j. The attenuation and the PSF are incorporated into the transition matrix, while the scatter contribution is calculated as:

$$s_j = s_j^{{\text{LE}}} + s_j^{{\text{HE}}} $$
(3)

where \(s_j^{{\text{LE}}} \) and \(s_j^{{\text{HE}}} \) are the low- and high-energy photons, which have mostly suffered scattering and have been finally detected. The scatter estimate s j is calculated using the reconstructed image of the original projections without any scatter compensation. This image is the input activity distribution of the SimSET simulator weighted by the corresponding emission yields of each ray. Once the scatter distribution is estimated, it is replaced in Eq. 2 to generate the reconstructed image. One remarkable fact is that in the AQM, the scatter fraction is included in the forward projector, which means that the scatter correction does not remove counts from the original projections. As a consequence, noise is not increased after the scatter correction, which it is an advantage of the AQM over other scatter correction methods used in clinical routine.

As a result, the AQM provides an image that is corrected by attenuation, PSF and scatter. Table 2 summarises the reconstruction parameters used for each camera.

Table 2 Reconstruction parameters used for both SPECT systems

Given that the convergence of the reconstruction process is not the same for the two collimators, the number of iterations needed for each camera is different. As an objective criterion, an additional iteration was calculated whenever the recovered SUR values increased at least by 1%.

Projections had to be filtered with masks before reconstruction because of high-energy contamination. Thus, the values located outside the field of view of the phantom were discarded to prevent them from interfering in the reconstruction process. Masking was performed by applying a geometric projector to the attenuation map and giving a value of 1 inside the resulting projections and 0 outside their limits.

Quantification of striatal uptake ratio with PVE correction

The size of the volumes of interest, particularly for small volumes such as the caudate and putamen, has a direct impact on the measurement of the activity concentration. Thus, correction for PVE should be performed to obtain the accurate quantification of the striatal nuclei. An approach to correcting the PVE in neuroreceptor imaging has been described by Fleming et al. [29, 30]. In the present work, we extend their approach by using 3D regions of interest (ROIs). Thus, our methodology is essentially based on the measurement of the total activity in the striatum and on the calculation of the exact volume of interest for each region and study. In clinical studies, the striatal ROIs could be obtained by segmenting MRI images registered with the SPECT images. In this high-resolution space, the total activity in the striatum is calculated by using automatically expanded ROIs. These are large enough to ensure the inclusion of all the activity that has spread outside the physical volume of the structures because of PVE. The advantage of using an automatic expansion of these ROIs is a reduced operator-introduced variability in their positioning.

In this work, striatal and occipital ROIs were defined by using a CT image of the anthropomorphic striatal brain phantom. To perform the quantification of the studies, all reconstructed images were re-sampled to the high-resolution space where the ROIs were defined. The exact volumes of the ROIs were calculated using the 3D ROI map. Compensation for the PVE was carried out by expanding the original striatal ROI and calculating the mean activity concentration inside the expanded ROI. The total activity, A s, inside the original striatal ROI was calculated by removing the non-specific uptake from the total activity in the expanded ROI:

$$A_{\text{s}} = \overline {A_{{\text{s'}}} } V_{{\text{s'}}} - \overline {A_{\text{o}} } \left( {V_{{\text{s'}}} - V_{\text{s}} } \right)$$
(4)

where \(\overline {A_{{\text{s'}}} } \) and \(\overline {A_{\text{o}} } \) are, respectively, the mean activity concentrations inside the expanded striatal ROI and the occipital region and V s and \(V_{{\text{s'}}} \) are the volumes of the original and the expanded striatal ROI, respectively.

Taking into account that the mean activity concentration inside the original striatal ROI is defined as \(\overline {A_{\text{s}} } = {{A_{\text{s}} } \mathord{\left/ {\vphantom {{A_{\text{s}} } {V_{\text{s}} }}} \right. \kern-\nulldelimiterspace} {V_{\text{s}} }}\) and substituting Eq. 4 in Eq. 1, the SUR value after PVE correction can be calculated as:

$${\text{SUR}} = \frac{{\left( {\overline {A_{{\text{s'}}} } - \overline {A_{\text{o}} } } \right)V_{{\text{s'}}} }}{{\overline {A_{\text{o}} } V_{\text{s}} }}$$
(5)

Note that this methodology assumes that the activity concentration of the background region is a good estimate of the non-specific uptake in the striatal ROI.

The expansion of the striatal ROI was established by calculating the resolution of the reconstruction method. This value was obtained by using SimSET to project a linear source in air with only low-energy photon emission and reconstructing the projections with PSF correction. The simulation and reconstruction were performed using the same parameters as those selected for the striatal phantom. The reconstructed image was fitted to a Gaussian function, and the standard deviation σ was calculated. Then, the striatal ROI was enlarged by 1σ. For the E.CAM parallel system, σ had a value of 1.90 mm (which leads to \(V_{{\text{s'}}} = 43.2{\text{ mL}}\)), and for the HELIX fan-beam system, σ had a value of 1.75 mm (which leads to \(V_{{\text{s'}}} = 38.8{\text{ mL}}\)).

Figure 2 shows one central section of the numerical phantom where the ROIs corresponding to the original striatum, the reference occipital region and the expanded striatum have been drawn.

Fig. 2
figure 2

A central slice of the phantom showing: a the considered ROIs for the striatal nuclei (red) and the occipital region (light blue), b the expanded striatal ROI (green) and c the overlay of a and b

Absolute quantification

The AQM was designed to provide an absolute volumetric activity at each voxel of the image. However, the values at each voxel of the reconstructed image (λ i ) obtained from Eq. 2 cannot be considered as real activity values. It is therefore necessary to find a factor that transforms these calculated values into real activity. To calculate this factor, the reconstructed image at the first iteration is selected as the input activity distribution of the SimSET simulator. The number of detected counts obtained in this simulation is compared with the total counts of the original projections. The factor that matches the counts of the simulated and the original projections is finally applied to the reconstructed image.

Results

The high-energy contamination contribution

Figure 3 represents for each of the cameras studied the contribution of the primary, low-energy scattered photons and high-energy scattered photons to the total projections. The images correspond to a noisy simulation of the striatal numerical phantom with a SUR value of 6. The numerical values of these contributions are also shown to emphasise the different contributions of the high-energy contamination when using two imaging systems equipped with a parallel and a fan-beam collimator and employing different acquisition protocols.

Fig. 3
figure 3

A simulated projection of the striatal phantom with SUR of 6 for a the E.CAM system and b the HELIX system. From top to bottom: primary, low-energy scattered photons, high-energy scattered photons and total photons detected. The numerical values of each contribution are also shown. Each image has been normalised to its maximum to improve its contrast

Relative quantification

To assess the effect of the corrections for the degrading phenomena in the quantification, results will be shown for different cases: (a) reconstruction without corrections (OSEM), (b) reconstruction with attenuation correction (OSEM-A), (c) 3D-reconstruction with attenuation and PSF corrections (OSEM-AP), (d) 3D-reconstruction with attenuation, PSF and scatter corrections (OSEM-APS) and (e) OSEM-APS plus PVE compensation (OSEM-APSV).

Figure 4 shows the calculated SURs against the nominal SURs for the 30 simulated studies and after reconstruction with different corrections. These values correspond to the eighth iteration for the E.CAM camera and to the sixth iteration for the HELIX camera. Fitting these plots to a straight line, we find correlation coefficients ranging from 0.997 to 0.998 for all the corrections applied, which demonstrates that there is a linear correlation between calculated and true values. This linear correlation may be described as:

$${\text{SUR}}_{{\text{calculated}}} = \alpha {\text{SUR}}_{{\text{theoretical}}} + \beta $$
(6)

where α is the slope of the regression line and β stands for the intercept at the origin. The dark solid line in Figure 4 corresponds to the identity line, whereas the light solid lines correspond to the linear fits calculated for each correction. The PVE compensation was applied by expanding the original striatum volume from V s = 21.5 to a volume of \(V_{{\text{s'}}} = 43.2{\text{ mL}}\) (ΔV = 21.7 mL) for the E.CAM system and from V s = 21.5 to \(V_{{\text{s'}}} = 38.8{\text{ mL}}\) (ΔV = 17.3 mL) for the HELIX system.

Fig. 4
figure 4

Calculated SURs obtained with different corrections for a the E.CAM system and b the HELIX system: without corrections (OSEM) (empty squares), OSEM-A (filled circles), OSEM-AP (empty circles), OSEM-APS (filled triangles) and OSEM-APSV (empty triangles)

Table 3 shows the mean and standard deviations for α and β, and the correlation coefficients of the linear fits for each acquisition system and for all the corrections implemented.

Table 3 Mean and standard deviations for α and β and correlation coefficients of the linear fits

In Figure 5, we can observe the reconstructed image of a central slice at the first iteration of the reconstruction algorithm for two cases: OSEM-AP and OSEM-APS. The improvement in the image quality can be clearly noted when the scatter correction is included. The first iteration is the most suitable for visual assessment, although the most accurate quantification results are achieved at a higher number of iterations.

Fig. 5
figure 5

Reconstructed central slice of the striatal brain phantom for the E.CAM system (top) and the HELIX system (bottom). For each camera, the images correspond to a nominal SUR of 4.7 and to a reconstruction using OSEM-AP (left) and OSEM-APS (right)

Absolute quantification

Figure 6 shows the bias between calculated and nominal mean activity concentration values as a function of nominal SUR. Quantification of the radiotracer uptake is less sensitive to PVE in the occipital region than in the striatum because of its size [31]. Therefore, PVE compensation is omitted for the occipital region and only the bias in the striatum after PVE compensation is reported.

Fig. 6
figure 6

The activity concentration bias (%) between calculated and nominal mean activity concentration values vs nominal SURs for (top) the striatum when OSEM-APS (empty squares) and OSEM-APSV (empty triangles) are used, and (bottom) the occipital region when OSEM-APS is used (empty squares). Biases are shown for: a the E.CAM system and b HELIX system. Horizontal lines represent a percent bias equal to 0

The bias in the striatum is shown when OSEM-APS and OSEM-APSV are used and in the occipital region when the OSEM-APS is applied. The bias from the true values was calculated as:

$$\overline{A} _{{{\text{bias}}}} {\left( \% \right)} = \frac{{\overline{A} _{{{\text{calculated}}}} - \overline{A} _{{{\text{nominal}}}} }}{{\overline{A} _{{{\text{nominal}}}} }} \times 100$$
(7)

As can be observed in Fig. 6, for the HELIX system and for low values of SUR (approximately between 0.5 and 1), there is a slight positive bias in the striatum when PVE is compensated and a slight negative bias in the occipital region. This causes an overestimation of the calculated SUR in patients with severe reduction in the DAT density.

Discussion

Figure 3 highlights the importance of the high-energy contamination when pharmaceuticals labelled with 123I are used in brain SPECT imaging. This contribution is strongly dependent on the acquisition system employed. When a LEHR parallel collimator is employed using the parameters shown in Table 1, more than 50% of the total photons is due to scattered photons. Indeed, high-energy photons account for 40%, which agrees with previous reported results [13] for the same imaging system. The high-energy photons degrade the image quality as they reduce the signal-to-noise ratio and the resolution in the projections. For the LEHR fan-beam collimator studied, the scattered photons contribute up to 43%, whereas high-energy contamination contribution is about 27%.

The results in Fig. 4 and Table 3 demonstrate that the recovered SUR values have a linear correlation with the true specific ratios for both cameras and for each correction applied. It should be pointed out that the relationship between measured and true SUR values is mainly linear with a non-zero intercept at the origin. For high nominal SURs, the slope (α) of the linear regression determines the SUR recovery factor. Nevertheless, for low nominal SURs, the value of the intercept at the origin β becomes a determinant parameter as it induces a bias in the recovery factor.

The values of α shown in Table 3 indicate that although both imaging systems achieve different recovery factors with the same corrections, there is a continuous improvement as the corrections are progressively included. For both cameras, the results in Table 3 show that the successive corrections raise the value of α from 36 to 97% for the E.CAM system (total increase of 61%) and from 42 to 100% (total increase of 58%) for the HELIX system. For the E.CAM camera, the total increase of 61% corresponded to 8% for the attenuation correction, 11% for the PSF correction, 14% for the scatter correction and 28% for the PVE compensation. For the HELIX camera, the total increase of 58% corresponded to 8% for attenuation, 10% for PSF, 15% for scatter and 25% for PVE. These values demonstrate that PVE compensation plays a major role in the recovery of the true specific ratios as it may induce an improvement of approximately 25–30% in α values depending on the imaging system used. The scatter correction also plays an important part as it increases α values by 15%. PSF and attenuation corrections are the least significant corrections with improvements of around 10 and 8%, respectively.

For all the cases with the exception of OSEM-APSV, the HELIX camera shows a value of α that is, on average, 6% higher than that of the E.CAM camera. On the other hand, the two systems differed in the behaviour of the β values depending on the correction applied. One important fact is that scatter correction causes β to fall close to 0. The addition of the PVE compensation induces a slight increase in β, which has no relevance in the recovery factor of both cameras. Furthermore, the PVE compensation minimises the differences in α, so that both imaging systems attain a recovery factor of around 100%. Our study suggests that recoveries close to 100% may be obtained by expanding the original striatal ROI in 1σ of the spatial resolution of the reconstruction. Thus, the enlargement of the original ROI by 1σ may be established as a general criterion for PVE correction in dopaminergic SPECT studies.

Overall, all these results indicate that although the correction for attenuation is considered to be mandatory in the clinical routine [8], it is necessary to correct all the image-degrading effects to achieve the nominal SURs. Thus, for a SUR of 4, the recovered SUR for the E.CAM system is 33% without corrections, 45% for OSEM-A, 56% for OSEM-AP, 68% for OSEM-APS and 97% for OSEM-APSV. For the HELIX system, the recovered SUR is 41% without corrections, 55% for OSEM-A, 65% for OSEM-AP, 75% for OSEM-APS and 102% for OSEM-APSV.

As the two gamma cameras equipped with completely distinct collimators and using different data acquisition protocols lead to similar SUR values when the AQM plus PVE compensation is employed, the proposed method may therefore be suitable for comparative studies using different gamma cameras or multicentric studies.

Furthermore, our results also highlight the importance of correcting for degradations to improve absolute quantification in neurotransmission SPECT studies with 123I-labelled radioligands. In this regard, Fig. 6 shows that when the 3D reconstruction includes attenuation, PSF and scatter correction, the calculated mean activity concentration in the striatum is underestimated by about 35% for the E.CAM system and about 20% for the HELIX system, whereas the maximum percent bias in the occipital region is about 5% for both cameras. PVE compensation proved to be effective in minimising the bias in the calculated mean activity concentration in the striatum, reducing errors to about 5% for both systems. These results indicate that PVE compensation in the striatum is necessary for absolute quantification. PVE compensation in the occipital region is not necessary as this area is large enough to influence the measurement of its mean activity concentration.

Conclusions

Our findings corroborate that quantitative results are dependent on the imaging system used in the acquisition when the reconstruction method does not include all the corrections for all the degrading phenomena [32, 33]. When all the corrections were incorporated into the quantification method, the differences practically disappeared, and both imaging systems reached approximately nominal values. The methodology presented in this paper establishes the corrections and the criteria to follow to compensate for all the image-degrading phenomena that affect neurotransmission 123I SPECT imaging, including a PVE compensation. We have sought to show that our quantification method leads to the theoretical quantitative values for two imaging devices equipped with different collimator systems and using different acquisition protocols.