Background. Traditionally, pediatric radiologists have been advocates of fluoroscopy systems that provide diagnostic images at the lowest possible radiation dose to the pediatric patient. Manufacturers of fluoroscopic equipment vary as to their claims of “low radiation“ exposures.¶Objectives. To obtain comparative data on radiation exposure and image quality from four pediatric hospitals, across variants of fluoroscopic equipment (such as pulsed versus continuous fluoroscopy).¶Materials and methods. Images were acquired from phantoms that simulated the size of a 3-year-old child. Phantom results, both stationary and rotating dynamic, were evaluated for radiation exposure and for image resolution of high- and low-contrast objects.¶Results. Radiation exposure from the four fluoro units varied widely; the lowest-dose selectable fluoro mode produced exposures varying between 34 and 590 mrads/min among the four fluoro units, and the highest-dose selectable fluoro mode produced 540–2230 mrads/min. The lowest radiation exposures were produced by pulsed fluoro units, and the very lowest radiation exposure was produced by a fluoroscope that had been especially optimized for pediatric imaging. There was only a small variation in image quality among the hospitals for visualization of stationary objects. A wide variability was noted for detection of objects on the moving phantom.¶Conclusions. The variability in the number of detected objects was considerably smaller than the variability in radiation exposure. Pulsed fluoroscopy provides improved resolution for moving objects. Optimization of one hospital's fluoroscope especially for pediatric imaging produced the best ratio of image quality to radiation exposure.
Article PDF
Similar content being viewed by others
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.Avoid common mistakes on your manuscript.
Author information
Authors and Affiliations
Additional information
Received: 29 June 1999/Accepted: 23 August 1999
Rights and permissions
About this article
Cite this article
Brown, P., Silberberg, P., Thomas, R. et al. A multihospital survey of radiation exposure and image quality in pediatric fluoroscopy. Pediatric Radiology 30, 236–242 (2000). https://doi.org/10.1007/s002470050729
Issue Date:
DOI: https://doi.org/10.1007/s002470050729